15
84 Dental Hypotheses Jul-Sep 2014 / Vol 5 | Issue 3 Electronic apex locator: A comprehensive literature review — Part I: Different generations, comparison with other techniques and different usages Hamid Mosleh, Saber Khazaei 1 , Hamid Razavian 2 , Armita Vali, Farzad Ziaei Dental Students’ Research Center, 2 Department of Endodontics, Torabinejad Dental Research Center, School of Dentistry, Isfahan University of Medical Sciences, Isfahan, 1 Department of Research, School of Dentistry, Kermanshah University of Medical Sciences, Kermanshah, Iran Introduction Ideal pulp treatment is defined as the removal of infected pulp and cleaning, shaping, and disinfecting the root canal system. [1] Subsequently, a three-dimensional filling can be provided. To achieve this goal, an essential stage is the assessment of the correct length of the root canals. [1,2] Working length is defined as the distance between the coronal/incisal reference point and the area that has been prepared and at which the filled canal should end. [1] Commonly, the minor apical foramen or apical isthmus is considered the end of the area for canal preparation and filling. The minor apical foramen is the border line between the dental pulp and periodontal area, which is approximately 0.5-1 mm from the anatomic apex. [1,3] Failure to determine the root canal length can result in both over- and underestimation of the root canal length. Overestimated working length can result in preparation beyond the apical isthmus, which can damage the peri-apical region. [4] Underestimated working length and inadequate debridement can cause unsuccessful treatment and dissatisfaction of both the patient and dentist. [5] Due to the pivotal role of working length determinations in root canal therapy, several methods have been introduced as follows. A: Tactile sensation and using the mean canal length and the application of paper cones are examples of experimental methods that are used by some clinicians due to their simplicity and relative efficiency. [6] These Corresponding Author: Dr. Hamid Razavian, Department of Endodoncs, School of Denstry, Isfahan University of Medical Sciences, Hezar Jerib St, Isfahan-81746-73461, Iran. E-mail: [email protected] SYSTEMATIC REVIEW ABSTRACT Introduction: To compare electronic apex locators (EAL) with others root canal determination techniques and evaluate other usage of this devices. Materials and Methods: “Tooth apex,” “Dental instrument,” “Odontometry,” “Electronic medical,” and “Electronic apex locator” were searched as primary identifiers via Medline/PubMed, Cochrane library, and Scopus data base up to 30 July 2013. Original articles that fulfilled the inclusion criteria were selected and reviewed. Results: Out of 402 relevant studies, 183 were selected based on the inclusion criteria. In this part, 108 studies are presented. Under the same conditions, no significant differences could be seen between different EALs of one generation. The application of EALs can result in lower patient radiation exposure, exact diagnosing of fractures, less perforation, and better retreatment. Conclusions: EALs were more accurate than other techniques in root canal length determination. Key words: Dental instrument, electronic apex locator, electronic medical, odontometry, tooth apex Access this article online Quick Response Code: Website: www.dentalhypotheses.com DOI: 10.4103/2155-8213.136744 [Downloaded free from http://www.dentalhypotheses.com on Friday, July 25, 2014, IP: 112.196.14.6] || Click here to download free Android application for this journa

Razavian H_Dental Hypot_2014.PDF (1)

Embed Size (px)

DESCRIPTION

Localizador de ápice

Citation preview

Page 1: Razavian H_Dental Hypot_2014.PDF (1)

84 Dental Hypotheses Jul-Sep 2014 / Vol 5 | Issue 3

Electronic apex locator: A comprehensive literature review — Part I: Different generations, comparison with other techniques and different usages Hamid Mosleh, Saber Khazaei1, Hamid Razavian2, Armita Vali, Farzad ZiaeiDental Students’ Research Center, 2Department of Endodontics, Torabinejad Dental Research Center, School of Dentistry, Isfahan University of Medical Sciences, Isfahan, 1Department of Research, School of Dentistry, Kermanshah University of Medical Sciences, Kermanshah, Iran

Introduction

Ideal pulp treatment is defined as the removal of infected pulp and cleaning, shaping, and disinfecting the root canal system.[1] Subsequently, a three-dimensional filling can be provided. To achieve this goal, an essential stage is the assessment of the correct length of the root canals.[1,2]

Working length is defined as the distance between the coronal/incisal reference point and the area that has been prepared and at which the filled canal should end.[1]

Commonly, the minor apical foramen or apical isthmus is considered the end of the area for canal preparation and filling. The minor apical foramen is the border line between the dental pulp and periodontal area, which is approximately 0.5-1 mm from the anatomic apex.[1,3] Failure to determine the root canal length can result in both over- and underestimation of the root canal length. Overestimated working length can result in preparation beyond the apical isthmus, which can damage the peri-apical region.[4] Underestimated working length and inadequate debridement can cause unsuccessful treatment and dissatisfaction of both the patient and dentist.[5] Due to the pivotal role of working length determinations in root canal therapy, several methods have been introduced as follows.

A: Tactile sensation and using the mean canal length and the application of paper cones are examples of experimental methods that are used by some clinicians due to their simplicity and relative efficiency.[6] These

Corresponding Author: Dr. Hamid Razavian, Department of Endodontics, School of Dentistry, Isfahan University of Medical Sciences, Hezar Jerib St, Isfahan-81746-73461, Iran. E-mail: [email protected]

SyStEmatic rEviEw

A B S T R A C T

Introduction: To compare electronic apex locators (EAL) with others root canal determination techniques and evaluate other usage of this devices. Materials and Methods: “Tooth apex,” “Dental instrument,” “Odontometry,” “Electronic medical,” and “Electronic apex locator” were searched as primary identifiers via Medline/PubMed, Cochrane library, and Scopus data base up to 30 July 2013. Original articles that fulfilled the inclusion criteria were selected and reviewed. Results: Out of 402 relevant studies, 183 were selected based on the inclusion criteria. In this part, 108 studies are presented. Under the same conditions, no significant differences could be seen between different EALs of one generation. The application of EALs can result in lower patient radiation exposure, exact diagnosing of fractures, less perforation, and better retreatment. Conclusions: EALs were more accurate than other techniques in root canal length determination.

Key words: Dental instrument, electronic apex locator, electronic medical, odontometry, tooth apex

Access this article online

Quick Response Code:Website: www.dentalhypotheses.com

DOI: 10.4103/2155-8213.136744

[Downloaded free from http://www.dentalhypotheses.com on Friday, July 25, 2014, IP: 112.196.14.6]  ||  Click here to download free Android application for this journal

Page 2: Razavian H_Dental Hypot_2014.PDF (1)

Mosleh, et al.: Electronic apex locator

85Dental HypothesesJul-Sep 2014 / Vol 5 | Issue 3

techniques can be inaccurate in some patients, however, due to open apex teeth and apical curvature.[6]

B: Radiography is a common method for determining the apical isthmus, which is 0.5-1 mm shorter than radiographic apex.[2] The radiographic method has its advantages, such as direct observation of the root canal system and the canal curvature and of the existence of peri-apical lesions,[2] but radiography cannot determine the apical isthmus, because it provides a two-dimensional picture of a three-dimensional object.[2,7] In addition, a disadvantage of radiation is that it can be dangerous to both patients and dental staff.

C: Owing to the advantages of electronic apex locators (EAL), such as the elimination of radiographic obstacles and EAL’s accuracy and convenience, the application of EAL has developed.[3,8] The principal design and development of the early apex locators dates back to Suzuki (1942)[9] whom investigated on dogs and found out that the electrical resistance between the periodontal membrane and the oral mucosa was a constant value. This point was introduced into clinical practice by Sunada[10] (1962) which almost measured the electrical resistance between oral mucosa and periodontal ligament.

Over the last decade, different versions of EAL have been released. Table 1 shows and compares the different versions of these devices, based on the functions of each generation. The aim of the present study was to investigate and compare the accuracy of EALs in determining the working length through root canal therapy.

Materials and Methods

Search strategy Electronic searches were performed using “tooth apex,” “dental instrument,” “odontometry,” “electronic medical,” and “electronic apex locator” as keywords. Moreover, cross-references were screened to identify further study that probably missed through the search strategy.

Electronic databasesElectronic searches were performed in Medline/PubMed, Cochrane library, and Scopus data base up to 30 July 2013. Four hundred and two articles were found.

Inclusion criteria The inclusion criteria were articles, clinical trials in humans and cohort and case-control studies in the

English language only. Systematic reviews, case reports, letter to editors, editorials, and congress abstracts were excluded. The title and abstract of each article were reviewed by three of the authors (HM, AV, and FZ), and articles that fulfilled the inclusion criteria were selected.

Data extraction Extraction of data from studies and assessment of validity was independently performed by two authors (HM and AV) and checked by a third author (FZ). In the case of disagreement between evaluators, it reassessed by discussion between two reviewers and a final consensus was agreed on (HR and SKh). Figure 1 provides information on the number of papers identified through the search strategy. Information of the authors, their institutions, and result of primary studies were removed before assessment of the validity. Information on the first author, year of publication, study design, study population and sample size, and the outcome measurements (main results) were extracted.

Results

Out of 402 articles, 183 studies were reviewed and 108 studies were selected for this part. The studies were categorized as follows.

Comparison of different EALsIn this part of the present study, 38 articles, consisting of 14 in vitro,[11-24] 9 ex vivo,[25-33] and 15 in vivo studies [34-48] were reviewed. Of the in vitro studies, four articles did not report significant differences between various devices[12,13,16,19] [Table 2]. All of the ex vivo studies showed significant differences between different devices, except for the study by Comin Chiaramonti et al.[31] and Baginska et al.[32][Table 2]. Among the

Figure 1: Flow chart of selected articles

[Downloaded free from http://www.dentalhypotheses.com on Friday, July 25, 2014, IP: 112.196.14.6]  ||  Click here to download free Android application for this journal

Page 3: Razavian H_Dental Hypot_2014.PDF (1)

Mosleh, et al.: Electronic apex locator

86 Dental Hypotheses Jul-Sep 2014 / Vol 5 | Issue 3

in vivo studies, Welk et al.[34] and Arora et al.[47] found a significant difference between EALs [Table 2].

Comparison of EALs with different working length determination methodsThirty-one studies evaluated different methods of working length determination. Among these studies, eleven compared different EALs and conventional and digital radiography as three different methods of working length determination.[7,35-46,49-59] Among the studies that compared EALs to radiographic method, six studies

did not report any significant differences, [7,35,36,43,53,54] three studies concluded that radiography method was more accurate,[49,55,59] and the remainder reported greater accuracy with the EAL [Table 3]. Three studies that compared digital radiography to EAL showed that EAL was more accurate[37,54,60] [Table 3]. Shanmugaraj et al.[6] compared three methods of measuring root canal length (apex locator, radiography, tactile sense) and reported that EAL was the most reliable method, but Ounsi et al.[61] showed that EAL and radiography method have

Table 1: Different generations of electrical apex locator devices and their operation base[86]

Generation Operation base Device samples

First Measurement of electrical resistance Root canal meter (Onuki medical Co.tokyo, Japan)Endodontic meter(Onuki medical Co.tokyo, Japan)Dentometer (Dahlin ectromedicine, Copenhagen, Denmark)Endo Radar (Electronica liarre, Imola, Italy)

Second Measurements of electrical impedance Sono-Explorer (Hayashi Dental Supply, Tokyo, Japan)Endo Cater (Yamaura Seisokushu, Tokyo, Japan)Digipex (Mada Equipment Co.Carlstadt, NJ, USA)Exact-A-Pex (Ellman International Hewlett, NY, USA)Formatron IV (Parkell Dental, Farmingdale, NY, USA)Endodontic Meter S II (Onuki Medical Co., Tokyo, Japan)Sono-Explorer Mark II (Hayashi Dental Supply, Tokyo, Japan).Sono-Explorer Mark II Junior (Hayashi Dental Supply, Tokyo, Japan)

Third Using two different frequencies at the same time in order to measure the difference or ratio between two currents

Endex/Apit (Osada Electrica Co. Tokyo, Japan)Root ZX (J.Morita, Tokyo, Japan)Neosono Ultima EZ (Satelec Inc, Mount Lourel, Nj, USA)TCM Endo V (Nouvag Ag, Goldach, Switzerland)Apex Pointer (MicroMega, Besanc¸on, France)Dat Apex (Dentsply Maillerfer, Ballaiques, Switzerland)Just or Justy II (Yoshida Co.Yokyo, Japan)Mark V Plus (Moyco/Union Broach, Bethpage, NY, USA)Apex pointer Endy 5000 (Loser, Leverkusen, Germany)Mini Apex Locator (Sybron Endo, Anaheim, CA, USA)Dentaport ZX (J. Morita MFG Corp., Osaka, Japan)Endo Analyzer Model 8005 (Analytic/Endo, Orange, CA, USA)Apex Finder AFA (Analytic Technologies, Redmond, WA)Mark V Plus (Moyco/Union Broach, Bethpage, New York, USA) Endox (Co. Lysis, Milan, Italy).Endy (Loser, Leverkusen, Germany) Apex Finder (Endo Analyzer 8001; AnalyticTechnology, Redmond, WA, USA) Foramatron D10 (Parkell Electronic Division,Farmingdale, New York, USA).

Fourth Using two or more non-simultaneous continuous frequencies in order to measure the difference or ratio between two currents

Bingo1020/Raypex4 (Foroum Engineering Technologies Rishon Lezion, Israel)Raypex4 (VDW, Munich, Germany)Element Dianostic Unit & Apex Locator (Sybron Endo, Anaheim, CA, USA)Neosono MC (Amadent Medical and Dental, Co.,Cherry Hill, New Jersey, USA). Propex (Dentsply Maillerfer, Ballaiques, Switzerland)Novapex (Foroum Engineering Technologies Rishon Lezion, Israel)Apex NRG XFR (Medic NRG Ltd, Tel Aviv, Israel)Apex DSP (Septodont, Saint-Maur des Fosse´s, Cedex, France)AFA Apex Finder, Model 7005 (Ana-lytic Endodontics, Orange, CA)iPex (NSK Ltd, Tokyo, Japan)Romi Apex D-30 (Romidan LTD, Kiryat-ono, Israel)

Fifth Measures the capacitance and resistance of the circuit separately

Propex II (Dentsply Maillerfer, Ballaiques, Switzerland)Top of FormBottom of FormApex Locator Joypex 5 (Henan, CBD Neihuan Road, Zhengzhou, China) I-ROOT (E-Magic Finder)(S-Denti SEoul, South Korea)Raypex 5 (VDW, Munich, Germany)

[Downloaded free from http://www.dentalhypotheses.com on Friday, July 25, 2014, IP: 112.196.14.6]  ||  Click here to download free Android application for this journal

Page 4: Razavian H_Dental Hypot_2014.PDF (1)

Mosleh, et al.: Electronic apex locator

87Dental HypothesesJul-Sep 2014 / Vol 5 | Issue 3

Table 2: Comparison of different electronic apex locators (EALs) in working length determinationAuthors Type of

studyPublication year

Sample (n) Types of studied ELA

Aim of study Main study result Type of generation

Comin Chiaramonti et al.[31]

Ex vivo 2012 40 teeth Bingo 1020Propex

Comparing operation of different apex locator devices

No significant difference

Fourth

Baginska et al.[32] Ex vivo 2012 40 teeth Raypex 5, Apex D.S.P and Locapex Five

Comparing operation of different apex locator devices

No significant difference

Fifth

Somma et al.[89] In vivo 2012 30 teeth Dentaport ZX, Raypex 5,ProPex II

Comparing operation of different apex locator devices

No significant difference

Third andFourth

Paludo et al.[48] In vivo 2012 100 rootCanals

Apex and iPex Comparing the accuracy of EALs in length measurement

There weren’t significant differences and both devices were accurate

Fourth

Jung et al.[15] In vitro 2011 104 teeth Root ZX,I-Root

Comparing operation of different apex locator devices

No significant difference

Third

Stober et al. [90] In vivo 2011 40 root canals

Raypex 5Mini Apex Locator

Comparing operation of different apex locator devices

No significant difference

Fourth

Stober et al.[87] In vivo 2011 40 root canals

Root ZX and iPex Comparing operation of different apex locator devices

No significant difference

Third andFourth

Miletic et al.[91] In vivo 2011 48 root canals

Dentaport ZX, RomiApex A-15 and Raypex 5

Comparing operation of different apex locator devices

No significant difference

Third andFourth

Silveira et al. [88] In vivo 2011 23 teeth Root ZX,Novapex

Comparing operation of different apex locator devices

No significant difference

Third andFourth

deVasconcelos et al. [28]

Ex vivo 2010 38 teeth Root ZX, RomiApex D-30, and Ipex

Comparing operation of different apex locator devices

Root ZX was more accurate

Third andFourth

Guise et al.[17] In vitro 2010 40 teeth Root ZX II, Elements AL, Precision AL

Comparing operation of different apex locator devices

Root ZX II was more accurate

Fourth

D’Assuncao et al. [33] Ex vivo 2010 31 teeth Root ZX-II, Novapex, Mini AL

Comparing operation of different apex locator devices

Root ZX II was more accurate

Fourth

Pascon et al. [92] In vivo 2009 831 root canals

DentaPort ZX, Raypex 5

Comparing operation of different apex locator devices

No significant difference

Third andFourth

Higa et al.[13] In vitro 2009 12 teeth Justylll, Dentaport,E-Magic Finder

Comparing operation of different apex locator devices

Justylll was more accurate

Third andFourth

Siu et al. [93] In vivo 2009 29 teeth Root ZX II, Apex NRGXFR, Mini Apex Locator

Comparing operation of different apex locator devices

No significant difference

Third

Pascon et al.[25] Ex vivo 2009 60 teeth Dentaport ZX, Raypex 5 ElementsDiagnostic Unit and A L

Comparing operation of different apex locator devices

ElementsDiagnostic Unit and A L was not accurate than others

Third andFourth

Ebrahim et al.[16] In vitro 2007 32 teeth Dentaport ZX, ProPex,Foramatron D10, Apex NRG,Apit 7

Comparing operation of different apex locator devices

Dentaport ZX, ProPex,Foramatron D10 were more accurate

Third andFourth

D’Assuncao et al.[18] Invitro 2007 40 teeth Mini AL, Root ZX II Comparing operation of different apex locator devices

No significant difference

Fourth

Bernardes et al.[12] Invitro 2007 40 teeth Root ZX, Elements Diagnostic Unit and A L, RomiAPEX D-30

Comparing operation of different apex locator devices

No significant difference

Third andFourth

(Continued )

[Downloaded free from http://www.dentalhypotheses.com on Friday, July 25, 2014, IP: 112.196.14.6]  ||  Click here to download free Android application for this journal

Page 5: Razavian H_Dental Hypot_2014.PDF (1)

Mosleh, et al.: Electronic apex locator

88 Dental Hypotheses Jul-Sep 2014 / Vol 5 | Issue 3

Table 2: (Continued)Authors Type of

studyPublication year

Sample (n) Types of studied ELA

Aim of study Main study result Type of generation

Wrbas et al.[94] In vivo 2007 20 teeth Root ZXRaypex5

Comparing operation of different apex locator devices

No significant difference

Third

Stavrianos et al.[24] In vivo 2007 80 root canals

Dentaport ZX,RayPex 4

Comparing operation of different apex locator devices

No significant difference

Third andFourth

Venturi et al.[26] Ex vivo 2007 60 root canals

Root ZX, Apex Finder

Comparing operation of different apex locator devices

Root ZX was more accurate

Third

Topuz et al.[19] In vitro 2007 47 teeth TCM Endo VRoot ZX

evaluating the accuracy of theapex-locating function of the TCM Endo V and to compare the results to these ofthe Root ZX

TCM Endo V proved to be as reliable as Root ZX but the use of thedevice to determine the working length was not easy as Root ZX

Third andFourth

Ebrahim et al.[27] Ex vivo 2006 36 teeth Root ZX, Foramatron D10, Apex NRG and Apit 7

Comparing operation of different apex locator devices

Root ZX and Foramatron D10 were more

Third andFourth

Plotino et al.[30] Ex vivo 2006 40 teeth Root ZX, Elements Diagnostic Unit and A L,ProPex

Comparing operation of different apex locator devices

ProPex was not accurate than others

Third andFourth

D’Assunco et al. [20] In vitro 2006 40 teeth NovapexRoot-ZX

to compare the accuracy of the Root-ZX andNovapex electronic apex locators (EALs) in locating the apical foramen

Root-ZX and Novapex areuseful and accurate devices for the apical foramen location

Third andFourth

Hor et al.[29] Ex vivo 2005 193 teeth Justy IIRaypex4

Comparing operation of different apex locator devices

Raypex4 was more accurate

Third andFourth

Haffner et al.[95] In vivo 2005 40 teeth Root ZXEndyJusty IIEndoxLysis

Comparing operation of different apex locator devices

No significant difference

Third

ElAyouti et al.[14] Ex vivo 2005 182 root canals

Raypex 4, Apex pointer, Root ZX

Comparing operation of different apex locator devices

Root ZX was more accurate

Third andFourth

Venturi et al.[96] Invivo 2005 64 teeth Apex Finder,Root ZX

Comparing operation of different apex locator devices

No significant difference

Third

Lucena-Martin et al.[11]

In vitro 2004 20 teeth Justy II,Root ZX,NeosonoUltima EZ

Comparing operation of different apex locator devices

No significant difference

Third

Hoer et al.[46] In vivo 2004 75 teeth Justy II, Endy 5000 Comparing operation of different apex locator devices

No significant difference

Third

Welk et al. [34] In vivo 2003 32 teeth Root ZXEndo Analyzer Model 8005

Comparing operation of different apex locator devices

Root ZX was more accurate

Third

De Moor et al. [21] In vitro 1999 15 single canal teeth

Apex Finder AFA Model 7005, Apex-Finder, NeosonoUltima EZ and Apit 2

The accuracy and operator dependency of four electroniccanal length measuring deviceswere compared under a set of specified conditions

Apex Finder AFA Model 7005 was the mostaccurate

Third

(Continued )

[Downloaded free from http://www.dentalhypotheses.com on Friday, July 25, 2014, IP: 112.196.14.6]  ||  Click here to download free Android application for this journal

Page 6: Razavian H_Dental Hypot_2014.PDF (1)

Mosleh, et al.: Electronic apex locator

89Dental HypothesesJul-Sep 2014 / Vol 5 | Issue 3

same accuracy and significantly were more accurate than tactile method. In the study by Subramaniam et al.[62], there were no significant differences among conventional radiography, digital radiography, and tactile sense in primary teeth. Janner et al.[63] compared cone beam computed tomography to EAL in root canal length determination and concluded that both two methods were comparable [Table 3].

Use of EAL in root perforations, fractures, and apical foramen widening Ten studies reported other uses of EALs, such as identification of root perforation sites and the location of horizontal and vertical fractures. EALs were only accurate in horizontal fracture diagnosis[64,65] although Topez et al.[66] reported that EALs were accurate in both vertical and horizontal root fractures. Furthermore, Goldberg et al.[67] studied the consistency of EALs in teeth with simulated horizontal root fractures, and they showed that EALs were accurate and consistent. Several studies showed that EALs were able to detect perforation sites [Table 4].[68-71] There were five studies that evaluated EALs’ ability to control apical foramen widening with rotary files, and they both concluded that use of EALs in root canals that were prepared with rotary instruments was not sufficiently accurate to control apical extensions.[72-75] Hoer et al.[76] evaluated the ability of the Justy II and Endy 5000 devices in

determining the canal length of 79 teeth (93 canals). They showed that these devices can specify the sites of the minor and major apical foramen, but they cannot determine apical constriction with sufficient accuracy. Also, Oishi et al.[77] investigated the accuracy of EALs in determining apical constriction, and they were accurate [Table 4].

Use of EAL in root canal retreatmentSix articles were published on this topic.[78-81] Two of them evaluated the accuracy of EALs before and after canal filling and showed that, in most cases, EALs were accurate in root canal retreatment.[78,80] In the study by Aggarwal et al.[80], the accuracy of Root ZX and Protaper devices was evaluated in the retreatment of filled canals with: 1) gutta-percha+ zinc oxide ogenol sealer; 2) gutta-percha+ AH plus sealer; and 3) Resilon+ Epiphany sealer. Both devices showed high accuracy in the first and second treatments, and no significant differences were reported in the presence of different filling materials. Uzun et al.[79] studied the precision of two apex locators (tri auto ZX TCM, locating handpieces endo apex) in root canal retreatment with root-end-cured teeth evaluated. They demonstrated that both devices could be used for determining apical area, but for root canal retreatment, accuracy of 0.0 mm is required, which these devices could not accomplish [Table 4].

Table 2: (Continued)Authors Type of

studyPublication year

Sample (n) Types of studied ELA

Aim of study Main study result Type of generation

Weiger et al. [22] In vitro 1999 41 teeth Root ZXApit

Two apex locators were compared regarding their ability to accurately locate theapical constriction in the presence of various canal fluids at different meterreadings.

In the presence of NaOCl, Root ZX provides the most accurateEWL measurements

Third

Lauper et al.[97] Ex vivo 1996 130 root canals

Apit and Odontometer

Evaluated the accuracy of EAL in length measurement

Apit was moreAccurate

First and Third

Arora et al. [47] In vivo 1995 61 root canals

ENDEX and RCM Mark II

Comparison the accuracy of EALs in canal length measurement in present of different contents(vital pulp, necrotic pulp, pus/exudates, sodiumhypochlorite, and water)

ENDEX was higher accuracy than the RCM Mark II

Third

Pallares and Faus [98] Ex vivo 1994 116 root canals

Odontometer and Endo Cater

Compared the accuracy of EAL in length measurement

Endo Cater was higher accuracy than the Odontometer

First and Second

Nahmias et al. [23] In vitro 1987 60 single root teeth

Sono-Explorer,C. L. Meter and Neosono-D

Compared the accuracy of EAL in length measurement

All EALs were accurate

Second and Third

[Downloaded free from http://www.dentalhypotheses.com on Friday, July 25, 2014, IP: 112.196.14.6]  ||  Click here to download free Android application for this journal

Page 7: Razavian H_Dental Hypot_2014.PDF (1)

Mosleh, et al.: Electronic apex locator

90 Dental Hypotheses Jul-Sep 2014 / Vol 5 | Issue 3

Table 3: Comparison of electronic apex locators (EALs) and different other techniques in working length determinationAuthors Type of

studyPublication year

Sample (n) Types of studied ELA

Aim of study Main study result

Orosco et al.[60] In vivo 2012 25 teeth Root ZX Comparison of EAL and radiography in determination of working length

Conventional and EAL were comparable and better than digital

Kishor[99] In vitro 2012 35 teeth Foramatron D10 Comparison of EAL and radiography, radiovisography in determination of working length

EAL are more accurate

Chougule et al.[58] In vivo 2012 13 primary teeth Dentaport ZX Comparison of EAL and radiography in determination of working length

EAL are more accurate

Zand et al.[100] In vivo/ Ex vivo

2011 75 teeth Root ZX Comparison of EAL and radiography in determination of working length

No significant difference

Jarad et al.[36] In vivo 2011 46 teeth Raypex 5 Comparison of EAL and radiography in determination of working length

No significant difference

Parekh and Taluja[101]

Ex vivo 2011 20 patients Root ZX Comparison of EAL and radiography in determination of working length

Radiography and EAL can be used together

Janner et al.[63] In vivo 2011 3 patients Root ZX Compared CBCT method to EAL Two methods were comparable

Neena et al.[54] In vivo 2011 30 teeth — Comparison of EAL and radiography in determination of working length

No significant difference

Singh et al.[57] In vivo 2011 20 teeth Propex II Comparison of EAL and radiography in determination of working length

EALs are more accurate

Real et al.[51] In vitro 2011 37 root canals Root ZXJust IIElements Diagnostic

Comparison of EAL and radiography in determination of working length

EALs are more accurate

Patino-Marin et al.[45]

In vivo 2011 61 root canals of primary teeth

Root ZX and ProPex

Comparison of EAL and radiography in determination of working length

EALs are more accurate

Kqiku et al.[7] Ex vivo 2011 30 teeth Root ZX Comparison of EAL and radiography in determination of working length

No significant difference

Vieyra et al.[38] In vivo/In vitro

2011 245 teeth (693 root canals)

Root ZX, Elements-Diagnostic, Precision ALand Raypex 5

Comparison of EAL and other methods in determination of working length

EALs are more accurate

Mancini et al.[44] Ex vivo 2011 120 teeth Endex, Propexll and Root ZX

Comparison of EAL and other methods in determination of working length

EALs are more accurate

Cianconi et al.[37] Ex vivo 2010 101 teeth EndexProPex IIRoot ZX

Comparison of EAL and radiography in determination of working length

EALs are more accurate

Sharma and Arora[102]

In vivo/ Ex vivo

2010 100 teeth Root ZX Comparison of EAL and radiography in determination of working length

EALs are more accurate

Vieyra et al.[46] In vivo 2010 160 teeth Root ZX and Elements-Diagnostic

Comparison of EAL and radiography in determination of working length

EALs are more accurate

Javidi et al.[40] In vitro 2009 30 teeth Root ZX Comparison of EAL and radiography in determination of working length

Radiography and EAL can be used together

Kim et al.[43] In vivo 2008 25 teeth Root ZX Evaluating EAL with or withoutradiography in determination of working length

No significant difference

Krajczar et al.[39] In vitro 2008 70 teeth ProPex Comparison of EAL and radiography in determination of working length

EAL are more accurate

Hassanien et al.[50] In vivo 2008 20 patients Root ZX Comparison of EAL and radiography in determination of working

EAL are more accurate

Shanmugaraj et al.[6]

In vivo/In vitro

2007 30 teeth Foramatron-IV Comparison of EAL and other methods in determination of working length

EAL are more accurate

Smadi et al.[51] In vivo 2006 151 root canals Tri Auto ZX Comparison of EAL and radiography in determination of working length

Use of EAL could reduce need for additional radiography

Subramaniam et al.[62]

In vitro 2005 20 teeth Formatron D10 Comparison of EAL and other methods in determination of working length

No significant difference

(Continued )

[Downloaded free from http://www.dentalhypotheses.com on Friday, July 25, 2014, IP: 112.196.14.6]  ||  Click here to download free Android application for this journal

Page 8: Razavian H_Dental Hypot_2014.PDF (1)

Mosleh, et al.: Electronic apex locator

91Dental HypothesesJul-Sep 2014 / Vol 5 | Issue 3

Table 3: (Continued)Authors Type of

studyPublication year

Sample (n) Types of studied ELA

Aim of study Main study result

ElAyouti et al.[41] In vitro 2002 30 teeth Root ZX Comparison of EAL and radiography in determination of working length

EAL are more accurate

Brunton et al.[42] In vitro 2002 50 teeth - Effect of apex locator on reduction X-ray exposure

Use of EAL reduced patients X-ray exposure

Martinez-Lozano et al.[35]

In vitro 2001 70 teeth Apit EM-S3 Comparison of EAL and radiography in determination of working length

No significant difference

Saad et al.[103] In vivo 2000 14 teeth Root ZX Effect of apex locator combiningRoot ZX and a digital imaging system (RadioVisioGraphy]on reduction X-ray exposure

This technique was useful

Ounsi et al.[61] In vitro 1998 37 teeth Endex Comparison of EAL and radiography, tactile sense method in determination of working length

Tactile sense method was inaccurate, but other two methods were same accurate

Himel et al.[55] In vivo 1993 96 root canals Formatron IV Comparing the accuracy of EALs and radiography method in length measurement

EALs weren’t accurate

Hembrough et al.[49]

In vivo 1993 26 maxillary molar teeth

Sono-Explorer Mark III

Compared the accuracy of EAL and radiography method in length measurement

EAL is useful only combined with radiography and couldn’t replace it

Frank et al.[56] In vivo 1993 185root canal

Endex Compared EAL with radiography method in length measurement

EAL was comparable with radiography method

Trope et al.[59] In vivo 1985 127 root canals Sono-Explorer Mark III

Comparison of EAL and radiography in determination of working length

EAL wasn’t accuracy same as radiography

Murakami et al.[104] In vivo 2002 66 infected canals

Sono-Explorer To retrospectively assess the success of endodontic treatment that had been guided by audiometric (electronic) measurement.

Use of the Sono-Exploreraided successful treatment of infected root canals

Stavrianos et al.[105]

In vivo 2007 85 teeth Raypex 5 Evaluated the accuracy of EAL in length measurement

EAL was accurate

Ounsi et al.[106] In vitro 1999 39single root teeth

Root ZX Evaluated the accuracy of EAL in length measurement

Root ZX couldn’t detect apical constriction and should only use to detecting major foramen

Vajrabhaya et al.[107]

In vivo 1997 20Single root teeth

Root ZX Evaluated the accuracy of EAL in length measurement

In clinical acceptable range its accurate

Shabahang et al.[108]

In vivo 1996 26 root canals Root ZX Evaluated the accuracy of EAL in length measurement

In clinical acceptable range its accurate

Wu et al.[109] In vivo 1992 20 single root teeth

Sono-Explorer type Y-III

Evaluated the accuracy of EAL in length measurement

In clinical acceptable range its accurate

Ricard et al.[110] In vivo 1991 37 teeth RCM Mark II Evaluated the accuracy of EAL in length measurement

In clinical acceptable range its accurate

McDonald et al.[111] In vivo 1990 47 teeth Endocater Evaluated the accuracy of EAL in detecting apical constriction

It was accurate

Berman et al.[112] In vivo 1984 24 matureand 5 immature root canals

Neosono-D Evaluated the accuracy of EAL in length measurement

EAL only in mature canal was accurate

Busch et al.[113] In vitro 1976 77 teeth (46 vital and 26 necrotic)

Sono-Explorer Evaluated the accuracy of EAL in length measurement

EAL was accurate in both groups

Use of EAL in patients with pacemakersWilson et al.[82] investigated the operation of the Endo Analyzer Model 8005 in patients with pacemakers and cardioverter/defibrillator devices. They demonstrated that there was no interference between the apex locator and pacemaker function.

Discussion

Different generations of EALs, with improved functions and greater clinical applications, have entered the market over these past few decades. Our results indicate that there are no significant differences between different

[Downloaded free from http://www.dentalhypotheses.com on Friday, July 25, 2014, IP: 112.196.14.6]  ||  Click here to download free Android application for this journal

Page 9: Razavian H_Dental Hypot_2014.PDF (1)

Mosleh, et al.: Electronic apex locator

92 Dental Hypotheses Jul-Sep 2014 / Vol 5 | Issue 3

Table 4: Other usage of electronic apex locators (EALs)Authors Type of

studyPublication year

Sample (n) Types of studied ELA

Aim of study Main study result

Aggarwal et al.[80] In vitro 2010 60 teeth Root ZX and ProPex

Application of apex locator devices in root canal retreatment

Both devices had high accuracy in retreatment

Stavrianos et al.[114] In vitro 2008 40 teeth Dentaport ZX, RayPex 5, Endo Master and Bingo-1020

Application of apex locator devices in root canal retreatment

All device were reliable, but Dentaport ZX and Endo Master were more accurate

Alves et al.[78] Ex vivo 2005 62 teeth Tri Auto ZX Application of apex locator devices in root canal retreatment

In most cases EAL was accurate in retreatment

Goldberg et al.[115] In vitro 2005 20 teeth ProPex, NovApex, and Root ZX

Application of apex locator devices in root canal retreatment

Third devices had high accuracy in retreatment

Uzun et al.[79] Ex vivo 2008 40 teeth TCM Endo V and Tri Auto ZX

Evaluation the accuracy of apex locator device along with rotary files in root canal length measurement in retreatment

Devices should be used with caution

Uzun et al.[81] In vitro 2007 40 root resected teeth

TCM Endo V and Tri Auto ZX

Evaluation the accuracy of apex locator device along with rotary files in root canal length measurement in retreatment

These devices are not appropriate for retreating.

Fadel et al.[74] In vivo 2012 30 single root premolar

Root ZX II Use of apex locator in controlled canal widening

Not appropriate

Jakobson et al.[73] In vivo 2008 24 teeth Root ZX II Evaluation of apex locator ability control apical fromen widening with rotary files

EAL with rotary instruments was not accurate to controlling apical extension

Felippe et al.[72] Ex vivo 2008 67 single root teeth

Root ZX II Use of apex locator in controlled canal widening

Not appropriate

Campbell et al.[75] In vitro 1998 60 teeth Tri Auto ZX to examine the apical extent of rotary canal instrumentation and the ability to maintain apical constriction with the Tri Auto ZX at different automated settings

Instrumentation with the automatic apical reverse feature set at 1 consistently approximated the apical constriction; however, the constriction wasfrequently enlarged

Goldberg et al.[67] In vitro 2008 20 teeth ProPexNovApexRoot ZXElementsAL

Evaluation apex locatordevice operation in diagnosis root fractures

All devices are reliable

Topuz et al.[66] In vitro 2008 40 teeth TCM Endo Vand Tri Auto ZX

Evaluation apex locator device operation in diagnosis root fractures

Both devices identified different root fracture in an acceptable range

Ebrahim et al.[65] In vitro 2006 90 teeth Root ZX, Foramatron D10, Apex NRG

Evaluation apex locatordevice operation in diagnosis root fractures

Device works accurately in teeth with horizontal root fractures

al Kadi et al.[116] In vitro 2006 100 teeeth Propex and Raypex-4

Evaluation apex locatordevice operation in diagnosis root fractures

Device works accurately in teeth with horizontal and vertical root fractures

Azabal et al.[64] In vitro 2004 64 teeth Justy II Evaluation apex locatordevice operation in diagnosis root fractures

Device works accurately in teeth with horizontal root fractures

Hoer et al.[76] In vitro/ In vivo

2004 93 root canals Justy II, Endy 5000 Evaluation apex locators ability in determination apical constriction

Devices cannot determine apical constriction

Oishi et al.[77] In vitro 2002 771 teeth ROOT ZX Evaluation apex locators ability in detremination apical constriction

Devicecan determine apical constriction

Pratten and Mc Donald [117]

In vitro 1996 — Apit Evaluation apex locators ability in detremination apical constriction

Devicecan determine apical constriction

(Continued )

[Downloaded free from http://www.dentalhypotheses.com on Friday, July 25, 2014, IP: 112.196.14.6]  ||  Click here to download free Android application for this journal

Page 10: Razavian H_Dental Hypot_2014.PDF (1)

Mosleh, et al.: Electronic apex locator

93Dental HypothesesJul-Sep 2014 / Vol 5 | Issue 3

EALs of the same generation. The first generation apex locator was supplied by single frequency of direct current in order to measurement of electrical resistance. Pain and discomfort were often felt with using this type of apex locator.[83] The second generation apex locator known as impedance apex locators was measured opposition to the flow of alternating current or impedance.[84] The disadvantage of this generation is that electro-conductive materials in canal affect on its accuracy.[83] The third generation apex locator (frequency dependent apex locators) was supplied by two frequencies to measure the impedance in the canal. The disadvantage of this generation sensitivity to canal fluid and the machine needs a fully charged battery.[85] The fourth generation apex locator measures the impedance characteristics using more than two frequencies.[3] The disadvantage includes needing to perform in relatively dry or in partially dried canals.[84] The fifth generation apex locator was developed in 2003 which measure the capacitance and resistance of the circuit separately.[86]

Many studies compared the ability of various generations of EALs in determining root canal length. Most of these studies showed that EALs were accurate for canal length measurement, within a clinically acceptable range of ± 0.5. Some studies indicated that the most recent generation of these devices had enhanced accuracy, better patient acceptance and greater ease of use for dentists.[31,87,88] but other studies mentioned that some EALs of the third generation were more accurate than those of the fourth generation.[20,27,28]

Although most of the previous studies reported that EALs were more accurate, compared to radiography, some of the studies noted no significant differences between the two methods due to small sample sizes. However, a recent randomized, controlled clinical trial study showed no significant differences between these two methods.[2] To consider the advantages of conventional radiography, such as the ability to observe the root canal system and the canal curvature directly and to determine the existence of peri-apical lesions, the decision of which method to use should be different in each case. It should be noted that EALs could decrease the patient’s radiation exposure.[42,51]

There is controversy in the diagnosis of the horizontal and vertical root fractures by EALs. Some studies have reported that EALs have the capacity to diagnose horizontal and lateral root fractures,[66,67] and others studies have indicated that horizontal fractures and perforation sites can be better diagnosed by EALs than vertical fractures.[64,65] Few studies have investigated the ability of apex locators to detect root fractures and perforations. Due to limited information on this subject, a general conclusion could not be achieved. More studies are required on this subject.

The present review has some limitations. First, only relevant articles were searched in Medline/PubMed, Cochrane library, and Scopus, which might have restricted the results. Second, our keywords were limited to “Tooth apex,” “Dental instrument,” “Odontometry,” “Electronic medical,” and “Electronic apex locator” to focus on EALs. More prospective, randomized clinical

Table 4: (Continued)Authors Type of

studyPublication year

Sample (n) Types of studied ELA

Aim of study Main study result

Keller et al.[118] In vivo 1991 99 canals Endocater Evaluated the ability of EAL in detecting apical constriction and cemento dentinal junction

Device was not accurate

Zmener et al.[71] In vitro 1999 40 teeth Tri Auto ZX Detection and measurement of endodontic root perforations using a newly designedapex-locating handpiece

The Tri Auto ZX detected and measured endodontic rootperforations within a range of clinically acceptable variations

Kaufman et al.[70] In vitro 1997 30 teeth with perforation in middle third

Root ZX , Sono Explorer Mark II Junior and Apit III

Evaluated the ability of EALs in locatin perforation

Regardless of the perforation size all EALs were accurate

Fuss et al.[68] In vitro 1996 32 teeth with perforation in midle third

Sono Explorer Mark 2 Junior and Apit 2

Evaluated the ability of EALs in locatin perforation

Both device were accurate

Hulsmann et al.[69] In vivo 1989 21 teeth Exact-A-Pex Evaluated the ability of EAL in control apical bridge formation in the treatment of teeth with incomplete root formation

Was accurate

[Downloaded free from http://www.dentalhypotheses.com on Friday, July 25, 2014, IP: 112.196.14.6]  ||  Click here to download free Android application for this journal

Page 11: Razavian H_Dental Hypot_2014.PDF (1)

Mosleh, et al.: Electronic apex locator

94 Dental Hypotheses Jul-Sep 2014 / Vol 5 | Issue 3

trials are needed to determine various conditions that affect EALs’ accuracy.

Conclusion

The results of the present study showed that EAL is an appropriate technique for root canal length measurements.

References

1. Renner D, Grazziotin-Soares R, Gavini G, Barletta FB. Influence of pulp condition on the accuracy of an electronic foramen locator in posterior teeth: An in vivo study. Braz Oral Res 2012;26:106-11.

2. Ravanshad S, Adl A, Anvar J. Effect of working length measurement by electronic apex locator or radiography on the adequacy of final working length: A randomized clinical trial. J Endod 2010;36:1753-6.

3. Duran-Sindreu F, Stober E, Mercade M, Vera J, Garcia M, Bueno R, et al. Comparison of in vivo and in vitro readings when testing the accuracy of the Root ZX apex locator. J Endod 2012;38:236-9.

4. Stoll R, Urban-Klein B, Roggendorf MJ, Jablonski-Momeni A, Strauch K, Frankenberger R. Effectiveness of four electronic apex locators to determine distance from the apical foramen. Int Endod J 2010;43:808-17.

5. Ozsezer E, Inan U, Aydin U. In vivo evaluation of ProPex electronic apex locator. J Endod 2007;33:974-7.

6. Shanmugaraj M, Nivedha R, Mathan R, Balagopal S. Evaluation of working length determination methods: An in vivo / ex vivo study. Indian J Dent Res 2007;18:60-2.

7. Kqiku L, Stadtler P. Radiographic versus electronic root canal working length determination. Indian J Dent Res 2011;22:777-80.

8. Mello-Moura AC, Moura-Netto C, Araki AT, Guedes-Pinto AC, Mendes FM. Ex vivo performance of five methods for root canal length determination in primary anterior teeth. Int Endod J 2010;43:142-7.

9. Suzuki K. Experimental study on iontophoresis. J Jap Stomatol 1942;16:411.

10. Sunada I. New method for measuring the length of the root canal. J Dent Res 1962;41:375-87.

11. Lucena-Martin C, Robles-Gijon V, Ferrer-Luque CM, de Mondelo JM. In vitro evaluation of the accuracy of three electronic apex locators. J Endod 2004;30:231-3.

12. Bernardes RA, Duarte MA, Vasconcelos BC, Moraes IG, Bernardineli N, Garcia RB, et al. Evaluation of precision of length determination with 3 electronic apex locators: Root ZX, Elements Diagnostic Unit and Apex Locator, and RomiAPEX D-30. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 2007;104:e91-4.

13. Higa RA, Adorno CG, Ebrahim AK, Suda H. Distance from file tip to the major apical foramen in relation to the numeric meter reading on the display of three different electronic apex locators. Int Endod J 2009;42:1065-70.

14. ElAyouti A, Kimionis I, Chu AL, Lost C. Determining the apical terminus of root-end resected teeth using three modern apex locators: A comparative ex vivo study. Int Endod J 2005;38:827-33.

15. Jung IY, Yoon BH, Lee SJ, Lee SJ. Comparison of the reliability of “0.5” and “APEX” mark measurements in two frequency-based electronic apex locators. J Endod 2011;37:49-52.

16. Ebrahim AK, Wadachi R, Suda H. In vitro evaluation of the accuracy of five different electronic apex locators for determining the working length of endodontically retreated teeth. Aust Endod J 2007;33:7-12.

17. Guise GM, Goodell GG, Imamura GM. In vitro comparison of three electronic apex locators. J Endod 2010;36:279-81.

18. D’Assuncao FL, de Albuquerque DS, Salazar-Silva JR, de Queiroz Ferreira LC, Bezerra PM. The accuracy of root canal measurements using the Mini Apex Locator and Root ZX-II: An evaluation in vitro. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 2007;104:e50-3.

19. Topuz O, Uzun O, Tinaz AC, Sadik B. Accuracy of the apex locating function of TCM Endo V in simulated conditions: A comparison study. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 2007;103:e73-6.

20. D’Assuncao FL, de Albuquerque DS, de Queiroz Ferreira LC. The ability of two apex locators to locate the apical foramen: An in vitro study. J Endod 2006;32:560-2.

21. De Moor RJ, Hommez GM, Martens LC, De Boever JG. Accuracy of four electronic apex locators: An in vitro evaluation. Endod Dent Traumatol 1999;15:77-82.

22. Weiger R, John C, Geigle H, Lost C. An in vitro comparison of two modern apex locators. J Endod 1999;25:765-8.

23. Nahmias Y, Aurelio JA, Gerstein H. An in vitro model for evaluation of electronic root canal length measuring devices. J Endod 1987;13:209-14.

24. Stavrianos C, Vladimirov SB, Vangelov LS, Papadopoulos C, Bouzala A. Evaluation of the accuracy of electronic apex locators Dentaport ZX and Ray Pex 4 under clinical conditions. Folia Med (Plovdiv) 2007;49:75-9.

25. Pascon EA, Marrelli M, Congi O, Ciancio R, Miceli F, Versiani MA. An ex vivo comparison of working length determination by 3 electronic apex locators. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 2009;108:e147-51.

26. Venturi M, Breschi L. A comparison between two electronic apex locators: An ex vivo investigation. Int Endod J 2007;40:362-73.

27. Ebrahim AK, Wadachi R, Suda H. Ex vivo evaluation of the ability of four different electronic apex locators to determine the working length in teeth with various foramen diameters. Aust Dent J 2006;51:258-62.

28. de Vasconcelos BC, do Vale TM, de Menezes AS, Pinheiro-Junior EC, Vivacqua-Gomes N, Bernardes RA, et al. An ex vivo comparison of root canal length determination by three electronic apex locators at positions short of the apical foramen. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 2010;110:e57-61.

29. Hor D, Krusy S, Attin T. Ex vivo comparison of two electronic apex locators with different scales and frequencies. Int Endod J 2005;38:855-9.

30. Plotino G, Grande NM, Brigante L, Lesti B, Somma F. Ex vivo accuracy of three electronic apex locators: Root ZX, Elements Diagnostic Unit and Apex Locator and ProPex. Int Endod J 2006;39:408-14.

31. Comin Chiaramonti L, Menini M, Cavalleri G. A comparison between two fourth generation apex locators. Minerva Stomatol 2012;61:183-96.

32. Baginska J, Ewa L, Bartlomiej B, Anna K. An ex vivo comparison of three electronic apex locators. Health Med BA 2012;10:3275-80.

33. D’Assuncao FL, Albuquerque DS, Salazar-Silva JR, Dos Santos VC, Sousa JC. Ex vivo evaluation of the accuracy and coefficient of repeatability of three electronic apex locators using a simple mounting model: A preliminary report. Int Endod J 2010;43:269-74.

[Downloaded free from http://www.dentalhypotheses.com on Friday, July 25, 2014, IP: 112.196.14.6]  ||  Click here to download free Android application for this journal

Page 12: Razavian H_Dental Hypot_2014.PDF (1)

Mosleh, et al.: Electronic apex locator

95Dental HypothesesJul-Sep 2014 / Vol 5 | Issue 3

34. Welk AR, Baumgartner JC, Marshall JG. An in vivo comparison of two frequency-based electronic apex locators. J Endod 2003;29:497-500.

35. Martinez-Lozano MA, Forner-Navarro L, Sanchez-Cortes JL, Llena-Puy C. Methodological considerations in the determination of working length. Int Endod J 2001;34:371-6.

36. Jarad FD, Albadri S, Gamble C, Burnside G, Fox K, Ashley JR, et al. Working length determination in general dental practice: A randomised controlled trial. Br Dent J 2011;211:595-8.

37. Cianconi L, Angotti V, Felici R, Conte G, Mancini M. Accuracy of three electronic apex locators compared with digital radiography: An ex vivo study. J Endod 2010;36:2003-7.

38. Vieyra JP, Acosta J. Comparison of working length determination with radiographs and four electronic apex locators. Int Endod J 2011;44:510-8.

39. Krajczar K, Marada G, Gyulai G, Toth V. Comparison of radiographic and electronical working length determination on palatal and mesio-buccal root canals of extracted upper molars. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 2008;106:90-3.

40. Javidi M, Moradi S, Rashed R, Raziee L. In vitro comparative study of conventional radiography and Root ZX apex locator in determining root canal working length. N Y State Dent J 2009;75:48-51.

41. ElAyouti A, Weiger R, Lost C. The ability of root ZX apex locator to reduce the frequency of overestimated radiographic working length. J Endod 2002;28:116-9.

42. Brunton PA, Abdeen D, MacFarlane TV. The effect of an apex locator on exposure to radiation during endodontic therapy. J Endod 2002;28:524-6.

43. Kim E, Marmo M, Lee CY, Oh NS, Kim IK. An in vivo comparison of working length determination by only root-ZX apex locator versus combining root-ZX apex locator with radiographs using a new impression technique. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 2008;105:e79-83.

44. Mancini M, Felici R, Conte G, Costantini M, Cianconi L. Accuracy of three electronic apex locators in anterior and posterior teeth: An ex vivo study. J Endod 2011;37:684-7.

45. Patino-Marin N, Zavala-Alonso NV, Martinez-Castanon GA, Sanchez-Benavides N, Villanueva-Gordillo M, Loyola-Rodriguez JP, et al. Clinical evaluation of the accuracy of conventional radiography and apex locators in primary teeth. Pediatr Dent 2011;33:19-22.

46. Vieyra JP, Acosta J, Mondaca JM. Comparison of working length determination with radiographs and two electronic apex locators. Int Endod J 2010;43:16-20.

47. Arora RK, Gulabivala K. An in vivo evaluation of the ENDEX and RCM Mark II electronic apex locators in root canals with different contents. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 1995;79:497-503.

48. Paludo L, Souza SL, So MV, Rosa RA, Vier-Pelisser FV, Duarte MA. An in vivo radiographic evaluation of the accuracy of Apex and iPex electronic Apex locators. Braz Dent J 2012;23:54-8.

49. Hembrough JH, Weine FS, Pisano JV, Eskoz N. Accuracy of an electronic apex locator: A clinical evaluation in maxillary molars. J Endod 1993;19:242-6.

50. Hassanien EE, Hashem A, Chalfin H. Histomorphometric study of the root apex of mandibular premolar teeth: An attempt to correlate working length measured with electronic and radiograph methods to various anatomic positions in the apical portion of the canal. J Endod 2008;34:408-12.

51. Smadi L. Comparison between two methods of working length determination and its effect on radiographic extent of root canal

filling: A clinical study [ISRCTN71486641]. BMC Oral Health 2006;6:4.

52. Fouad AF, Reid LC. Effect of using electronic apex locators on selected endodontic treatment parameters. J Endod 2000;26:364-7.

53. Real DG, Davidowicz H, Moura-Netto C, Zenkner Cde L, Pagliarin CM, Barletta FB, et al. Accuracy of working length determination using 3 electronic apex locators and direct digital radiography. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 2011;111:e44-9.

54. Neena I, Ananthraj A, Praveen P, Karthik V, Rani P. Comparison of digital radiography and apex locator with the conventional method in root length determination of primary teeth. J Indian Soc Pedod Prev Dent 2011;29:300-4.

55. Himel VT, Cain C. An evaluation of two electronic apex locators in a dental student clinic. Quintessence Int 1993;24:803-6.

56. Frank AL, Torabinejad M. An in vivo evaluation of Endex electronic apex locator. J Endod 1993;19:177-9.

57. Singh SV, Nikhil V, Singh AV, Yadav S. An in vivo comparative evaluation to determine the accuracy of working length between radiographic and electronic apex locators. Indian J Dent Res 2012;23:359-62.

58. Chougule RB, Padmanabhan MY, Mandal MS. A comparative evaluation of root canal length measurement techniques in primary teeth. Pediatr Dent 2012;34:53-6.

59. Trope M, Rabie G, Tronstad L. Accuracy of an electronic apex locator under controlled clinical conditions. Endod Dent Traumatol 1985;1:142-5.

60. Orosco FA, Bernardineli N, Garcia RB, Bramante CM, Duarte MA, Moraes IG. In vivo accuracy of conventional and digital radiographic methods in confirming root canal working length determination by Root ZX. J Appl Oral Sci 2012;20:522-5.

61. Ounsi HF, Haddad G. In vitro evaluation of the reliability of the Endex electronic apex locator. J Endod 1998;24:120-1.

62. Subramaniam P, Konde S, Mandanna DK. An in vitro comparison of root canal measurement in primary teeth. J Indian Soc Pedod Prev Dent 2005;23:124-5.

63. Janner SF, Jeger FB, Lussi A, Bornstein MM. Precision of endodontic working length measurements: A pilot investigation comparing cone-beam computed tomography scanning with standard measurement techniques. J Endod 2011;37:1046-51.

64. Azabal M, Garcia-Otero D, de la Macorra JC. Accuracy of the Justy II Apex locator in determining working length in simulated horizontal and vertical fractures. Int Endod J 2004;37:174-7.

65. Ebrahim AK, Wadachi R, Suda H. Accuracy of three different electronic apex locators in detecting simulated horizontal and vertical root fractures. Aust Endod J 2006;32:64-9.

66. Topuz O, Uzun O, Tinaz AC, Bodrumlu E, Gorgul G. Accuracy of two apex-locating handpieces in detecting simulated vertical and horizontal root fractures. J Endod 2008;34:310-3.

67. Goldberg F, Frajlich S, Kuttler S, Manzur E, Briseno-Marroquin B. The evaluation of four electronic apex locators in teeth with simulated horizontal oblique root fractures. J Endod 2008;34:1497-9.

68. Fuss Z, Assooline LS, Kaufman AY. Determination of location of root perforations by electronic apex locators. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 1996;82:324-9.

69. Hulsmann M, Pieper K. Use of an electronic apex locator in the treatment of teeth with incomplete root formation. Endod Dent Traumatol 1989;5:238-41.

70. Kaufman AY, Fuss Z, Keila S, Waxenberg S. Reliability of different electronic apex locators to detect root perforations in vitro. Int Endod J 1997;30:403-7.

71. Zmener O, Grimberg F, Banegas G, Chiacchio L. Detection and measurement of endodontic root perforations using a newly

[Downloaded free from http://www.dentalhypotheses.com on Friday, July 25, 2014, IP: 112.196.14.6]  ||  Click here to download free Android application for this journal

Page 13: Razavian H_Dental Hypot_2014.PDF (1)

Mosleh, et al.: Electronic apex locator

96 Dental Hypotheses Jul-Sep 2014 / Vol 5 | Issue 3

designed apex-locating handpiece. Endod Dent Traumatol 1999;15:182-5.

72. Felippe WT, Felippe MC, Reyes Carmona J, Crozoe FC, Alvisi BB. Ex vivo evaluation of the ability of the ROOT ZX II to locate the apical foramen and to control the apical extent of rotary canal instrumentation. Int Endod J 2008;41:502-7.

73. Jakobson SJ, Westphalen VP, da Silva Neto UX, Fariniuk LF, Picoli F, Carneiro E. The accuracy in the control of the apical extent of rotary canal instrumentation using Root ZX II and ProTaper instruments: An in vivo study. J Endod 2008;34:1342-5.

74. Fadel G, Piasecki L, Westphalen VP, Silva Neto UX, Fariniuk LF, Carneiro E. An in vivo evaluation of the auto apical reverse function of the Root ZX II. Int Endod J 2012;45:950-4.

75. Campbell D, Friedman S, Nguyen HQ, Kaufman A, Keila S. Apical extent of rotary canal instrumentation with an apex-locating handpiece in vitro. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 1998;85:319-24.

76. Hoer D, Attin T. The accuracy of electronic working length determination. Int Endod J 2004;37:125-31.

77. Oishi A, Yoshioka T, Kobayashi C, Suda H. Electronic detection of root canal constrictions. J Endod 2002;28:361-4.

78. Alves AM, Felippe MC, Felippe WT, Rocha MJ. Ex vivo evaluation of the capacity of the Tri Auto ZX to locate the apical foramen during root canal retreatment. Int Endod J 2005;38:718-24.

79. Uzun O, Topuz O, Tinaz C, Nekoofar MH, Dummer PM. Accuracy of two root canal length measurement devices integrated into rotary endodontic motors when removing gutta-percha from root-filled teeth. Int Endod J 2008;41:725-32.

80. Aggarwal V, Singla M, Kabi D. An in vitro evaluation of performance of two electronic root canal length measurement devices during retreatment of different obturating materials. J Endod 2010;36:1526-30.

81. Uzun O, Topuz O, Tinaz AC, Alacam T. Apical accuracy of two apex-locating handpieces in root canal retreatments of root-end resected teeth. J Endod 2007;33:1444-6.

82. Wilson BL, Broberg C, Baumgartner JC, Harris C, Kron J. Safety of electronic apex locators and pulp testers in patients with implanted cardiac pacemakers or cardioverter/defibrillators. J Endod 2006;32:847-52.

83. Suchde RV, Talim ST. Electronic ohmmeter. An electronic device for the determination of the root canal length. Oral Surg Oral Med Oral Pathol 1977;43:141-50.

84. Muthu MS, Sivakumar N. Accuracy of electronic apex locator in length determination in the presence of different irrigants: An in vitro study. J Indian Soc Pedod Prev Dent 2006;24:182-5.

85. Kobayashi C, Suda H. New electronic canal measuring device based on the ratio method. J Endod 1994;20:111-4.

86. Gordon MP, Chandler NP. Electronic apex locators. Int Endod J 2004;37:425-37.

87. Stober EK, Duran-Sindreu F, Mercade M, Vera J, Bueno R, Roig M. An evaluation of root ZX and iPex apex locators: An in vivo study. J Endod 2011;37:608-10.

88. Silveira LF, Petry FV, Martos J, Neto JB. In vivo comparison of the accuracy of two electronic apex locators. Aust Endod J 2011;37:70-2.

89. Somma F, Castagnola R, Lajolo C, Paterno Holtzman L, Marigo L. In vivo accuracy of three electronic root canal length measurement devices: Dentaport ZX, Raypex 5 and ProPex II. Int Endod J 2012;45:552-6.

90. Stober EK, de Ribot J, Mercade M, Vera J, Bueno R, Roig M, et al. Evaluation of the Raypex 5 and the Mini Apex Locator: An in vivo study. J Endod 2011;37:1349-52.

91. Mi le t ic V , Be l j ic - Ivanovic K , Ivanovic V . C l in ica l reproducibility of three electronic apex locators. Int Endod J 2011;44:769-76.

92. Pascon EA, Marrelli M, Congi O, Ciancio R, Miceli F, Versiani MA. An in vivo comparison of working length determination of two frequency-based electronic apex locators. Int Endod J 2009;42:1026-31.

93. Siu C, Marshall JG, Baumgartner JC. An in vivo comparison of the Root ZX II, the Apex NRG XFR, and Mini Apex Locator by using rotary nickel-titanium files. J Endod 2009;35:962-5.

94. Wrbas KT, Ziegler AA, Altenburger MJ, Schirrmeister JF. In vivo comparison of working length determination with two electronic apex locators. Int Endod J 2007;40:133-8.

95. Haffner C, Folwaczny M, Galler K, Hickel R. Accuracy of electronic apex locators in comparison to actual length — an in vivo study. J Dent 2005;33:619-25.

96. Venturi M, Breschi L. A comparison between two electronic apex locators: An in vivo investigation. Int Endod J 2005;38:36-45.

97. Lauper R, Lutz F, Barbakow F. An in vivo comparison of gradient and absolute impedance electronic apex locators, J Endod 1996;22:260-3.

98. Pallares A, Faus V. An in vivo comparative study of two apex locators. J Endod 1994;20:576-9.

99. Kishor KM. Comparison of working length determination using apex locator, conventional radiography and radiovisiography: An in vitro study. J Contemp Dent Pract 2012;13:550-3.

100. Zand V, Mokhtari H, Lotfi M, Reyhani MF, Sohrab A, Tehranchi P, et al. Accuracy of working length determination with root ZX apex locator and radiography: An in vivo and ex vivo study. Afr J Biotechnol 2013;10:7088-91.

101. Parekh V, Taluja C. Comparative study of periapical radiographic techniques with apex locator for endodontic working length estimation: An ex vivo study. J Contemp Dent Pract 2011;12:131-4.

102. Sharma MC, Arora V. Determination of working length of root canal. Med J Armed Forces India 2010;66: 231-4.

103. Saad AY, al-Nazhan S. Radiation dose reduction during endodontic therapy: A new technique combining an apex locator (Root ZX) and a digital imaging system (RadioVisioGraphy). J Endod 2000;26:144-7.

104. Murakami M, Inoue S, Inoue N. Clinical evaluation of audiometric control root canal treatment: A retrospective case study. Quintessence Int 2002;33:465-74.

105. Stavrianos C, Vasiliadis L, Stavrianou I, Kafas P. Clinical evaluation of the accuracy of an electronic tooth apex locator. J Med Sci 2007;7:1369-71.

106. Ounsi HF, Naaman A. In vitro evaluation of the reliability of the Root ZX electronic apex locator. Int Endod J 1999;32:120-3.

107. Vajrabhaya L, Tepmongkol P. Accuracy of apex locator. Endod Dent Traumatol 1997;13:180-2.

108. Shabahang S, Goon WW, Gluskin AH. An in vivo evaluation of Root ZX electronic apex locator. J Endod 1996;22:616-8.

109. Wu YN, Shi JN, Huang LZ, Xu YY. Variables affecting electronic root canal measurement. Int Endod J 1992;25:88-92.

110. Ricard O, Roux D, Bourdeau L, Woda A. Clinical evaluation of the accuracy of the Evident RCM Mark II Apex Locator. J Endod 1991;17:567-9.

111. McDonald NJ, Hovland EJ. An evaluation of the Apex Locator Endocater. J Endod 1990;16:5-8.

112. Berman LH, Fleischman SB. Evaluation of the accuracy of the Neosono-D electronic apex locator. J Endod 1984;10:164-7.

113. Busch LR, Chiat LR, Goldstein LG, Held SA, Rosenberg PA. Determination of the accuracy of the Sono-Explorer for establishing endodontic measurement control. J Endod 1976;2:295-7.

[Downloaded free from http://www.dentalhypotheses.com on Friday, July 25, 2014, IP: 112.196.14.6]  ||  Click here to download free Android application for this journal

Page 14: Razavian H_Dental Hypot_2014.PDF (1)

Mosleh, et al.: Electronic apex locator

97Dental HypothesesJul-Sep 2014 / Vol 5 | Issue 3

114. Stavrianos CH, Vladimirov ST, Vasiliadis L, Stavrianou I, Panayotov I, Pamporakis P. In vitro evaluation of the precision of four different electronic apex locators in determining the working length of teeth after removing root canal obturation materials. Res J Med Sci 2008;2:282-6.

115. Goldberg F, Marroquín BB, Frajlich S, Dreyer C. In vitro evaluation of the ability of three apex locators to determine the working length during retreatment. J Endod 2005; 31:676-8.

116. al Kadi H, Sykes LM, Vally Z. Accuracy of the Raypex-4 and Propex apex locators in detect ing horizontal and vertical root fractures: An in vitro study. SADJ

2006;61:244-7.117. Pratten DH, McDonald NJ. Comparison of radiographic and

electronic working lengths. J Endod 1996;22:173-6.118. Keller ME, Brown CE Jr, Newton CW. A clinical evaluation of the

Endocater an electronic apex locator. J Endod 1991;17:271-4.

Cite this article as: Mosleh H, Khazaei S, Razavian H, Vali A, Ziaei F. Electronic apex locator: A comprehensive literature review - Part I: Different generations, comparison with other techniques and different usages. Dent Hypotheses 2014;5:84-97.

Source of Support: Nil. Conflict of Interest: None declared.

[Downloaded free from http://www.dentalhypotheses.com on Friday, July 25, 2014, IP: 112.196.14.6]  ||  Click here to download free Android application for this journal

Page 15: Razavian H_Dental Hypot_2014.PDF (1)

Copyright of Dental Hypotheses is the property of Dental Hypotheses and its content may notbe copied or emailed to multiple sites or posted to a listserv without the copyright holder'sexpress written permission. However, users may print, download, or email articles forindividual use.