Upload
zion-haye
View
216
Download
1
Tags:
Embed Size (px)
Citation preview
David Joy
Dr. David C. JoyDistinguished ProfessorMaterials Science and Engineering
University of Tennessee Knoxville, TN
D.Phil., University of Oxford (UK): A Study of Magnetic Domains in the SEMHitachi payroll
A Clean Machine
The FEGSEM cannot operate except in a clean, ultra-high vacuum. (Numerous caveats for low vacuum and ESEM operation.) This requires items such as:
Scroll Pumps Turbo Molecular Pumps Ion Pumps as well as associated gauges and even with the cleanest of vacuums there is still The
Dark Side of SEM that must be faced….
Terminology
“Low vacuum” = High pressure
“High Vacuum” = Low pressure
Common Vacuum Units There are many varied
units that are used to specify pressures
The Torr, the Bar and the Pascal are in common use...
.. but the Pascal is the SI recommended unit for pressure and so is the best choice for documentation
1 Atmospheric pressure is 760 mm Hg = 1 Bar = 105 Pa
1 Torr = 1 mm Hg 1 Torr = 1/760 of an
atmosphere = 132 Pa 1 milliTorr = 0.13Pa = 1 μmHg
1mbar = 1/1000 Atm = 0.76 Torr = 100Pa
1 Pa = 7.6 milliTorr = 7.6 μmHg
Pressure: Units of Measure
Pressure exerted by a column of fluid:P ≡ F/A = mg/A = ghA/A = gh h 1 Atm (mean sea level) = 760 Torr =
1013 mBar = 1.01x105 Pa = 101.3 kPa = 14.7 psi = 34 ft. water
Average atmospheric pressure in SLC is about 635 Torr, 12.3 psi, 28.4 ft water…
“Kinds of Pressure”
Gauge Pressure: measured with respect to ambient.
Absolute pressure: measured with respect to vacuum
Car tires, basketballs, boilers, LN2 tanks, JFB/MEB compressed air supply…
Vacuum systems, cathode ray tubes, light bulbs, barometers
Mean Free Path in Gases
With sufficient accuracy for approximate calculations we may take:
λ = 7 x 10-3/p mbar-cmλ = 5 x 10-3/p Torr-cmλ = 5/p μmHg-cm
Qualitative Vacuum Ranges
Low vacuum (SEC) 760 to 1 Torr
Medium vacuum (SEC) 1 to 10-3 Torr
High vacuum (Chamber) 10-3 to 10-6 Torr
Very high (Column) 10-6 to 10-9 Torr
Ultra-high (Gun) 10-9 and lower
FEGSEMs contain examples of each vacuum level
laminar
molecular
Vacuum pumps For each of the vacuum
ranges identified earlier there is one or more type of pump that is best
Pumps are always used in combination - one pump is used to start the next
The sequencing of the pump down is crucial and so this is done under computer control
Scroll Pumps Scroll pumps are the
foundation of clean vacuum systems
They consist of two Archimedes’ screws machined into aluminum plates mounted so that the spirals interleave
One plate is held fixed while the other oscillates. Gas is trapped between the spirals and forced out to the exit port
Pumping speed is constant from Atmospheric pressure down to about 1000Pa and the ultimate pressure is about 10Pa
Scroll pumps are oil-free and require neither inlet nor outlet valves
The world’s oldest pump technology – Archimedes’ screw
Alternatively….
Roughing can also be carried out using a diaphragm pump
Oscillation of the diaphragm alternately pulls gas in one port and then expels it through the other.
Oil free pumps are clean but typically a factor of 3x slower, and 3x more expensive, than pumps containing oil – but worth the wait and expense
Rotary Vane Mechanical Pump
RobustInexpensiveOperates to
ambient pressure
Single stage and two stage
Turbomolecular pump Archimedian screw -
runs at 20000+ rpm Needs electronic
protection / control for the bearings in case of loss of power
Produces a clean, oil-free, high vacuum down to 10-6 Pa (10-8 Torr)
Must be backed: scroll pump, diaphragm pump or rotary oil pump.
Turbo pump performance
Turbo pumps can start even at atmospheric pressure (although they labor) and they can go down to 10-8 T
It is best to pre-pump the system with a clean backing system
TMP do not pump all gases with same efficiency - large molecules are pumped faster than smaller molecules
1 milliT = 0.13Pa = 1 μmHg
Turbo pump performance
Ion Pumps Ionized molecules
spiral in magnetic field and get buried in Ti wall coating
A large number of these structures are run in parallel to improve the pumping speed
Diode pumps only handle gases that are easily ionized (no noble gases)
The triode pump If noble or unusual
gases are expected to be found in the SEM (nitrogen, helium, counter gases from a WDS system etc.) then a triode pump must be used.
The additional electrode then makes it possible to ionize these gases
Ion pump performance
“The” UHV pump - goes to 10-9 Pa (10-11 Torr) and below in a properly designed vacuum system
Requires no backing…more than a little misleading… in fact it works best in a sealed system. Entrainment pump!
The IP requires a periodic bake-out into rough pumped system to clean the buried gas from the pump. This is done during the gun bake procedure
Check for electrical instability by slapping the pump with an open hand. Instability indicates need for a bakeIon Pump Performance
Cryogenic pump Cryo-pumps use liquid
helium and activated charcoal absorbers to pump to 10-12 T
Very high pumping speeds No vibration or magnetic
fields But they need periodic
bake-outs into a rough pump to clean the absorbers
They are expensive to run unless used with a closed-circuit (Stirling engine) liquid He pump
Vacuum Gauges Vacuum systems must be
monitored constantly to ensure satisfactory performance, but manufacturers seem to be reluctant to provide gauges which allow this to be done
Many different types of gauges are available because each only covers a limited range of pressures
Never trust a gauge unless you can check it independently
Range of gauge utility
Pirani gauge The Pirani is a
dedicated low vacuum gauge device
The resistance of the hot wire changes with the rate of heat loss (conduction) to the gas
The Wheatstone bridge then measures the change in resistance of the hot wire
Pirani’s are rugged and generally reliable and rarely need attention
Schematic Circuit for a Pirani (hot wire) gauge
Pirani calibration The calibration of a
Pirani depends on thermal conductivity and so on the actual gas in the system
Beware when using a crystal spectrometer as gases leaking from the counter tubes will degrade the accuracy of the Pirani gauge
Correction Curve for Pirani Gauges
Penning (Cold cathode) Gauge
A Penning gauge measures the ion current flowing from the cathode to the anode
The magnetic field increases sensitivity by making the ions spiral as they travel to cause secondary ionization
Beware - a Penning gauge reads zero current when the pressure is both very low and very high. The gauge must ‘strike’ to be operational
Check with a Pirani gauge if in doubt
Penning gauges require routine cleaning and testing
Capacitance Manometer
Gauge head on chamber Controller and digital read-out
Capacitance Manometer
A = Annular electrode D = Disk electrode S = Substrate G = Getter (in vacuum
space) Differential capacitance
between annulus and disk depends on pressure difference between Test Chamber and “Getter”. (Earlier reference to “unbacked getter pump”)
Ion gaugesPressures lower than 10-5 Torr can be
measured with ion gauges (which are miniature ion pumps) or (more usually) directly from the actual ion pump
Mass spectrometer gauges (residual gas analyzers) are a desirable extra. These can measure partial pressures of e.g helium (for leak testing) or of water vapor.
O-ring seals O-rings (from the 1950s) made
it possible to build demountable vacuum systems
The rings are now made of high tech polymers such as VITON
Two kinds are in common use... Black/shiny - has filler. Low
vacuum only. Lubricate with finger grease to prevent cracking
Brown/dull - high vacuum, and bakeable. Do not grease
Do not crush or cut the ring - ensure that it is in the groove designed for it
o-ring compressed to fill groove
remove dust,hairs andlubricate withfinger
UHV metal to metal seals
Knife edges
copper gasket
bolt
Tighten boltsin sequence1,4,2,5,3,6
1
2
3
4
5
6
First used in the 1960s Knife edges on the flanges
cut into OFHC (oxygen free high conductivity) copper rings about 5mm thick to make an impermeable metal to metal seal
Good down to pressures as low as 10-10 Pa
Bakeable, clean, long lasting and (with care) reusable
Expensive - an 8 inch gasket costs ~ $100
Don’t touch !
Vacuum Hygiene Always keep vacuum systems
running 24/7/365 Use LN2 cooled maze traps, and
fore-line traps, to reduce backstreaming in older machines
Do not overpump the specimen exchange chamber (SEC) as this can result in backstreaming (unique to Hitachi FEG systems)
Keep your fingers away from samples and from the specimen chamber area - wear gloves
If column contamination occurs try nitrogen purging (laminar flow) over a weekend
Maze trap fitted to a rotary pump
(mfp discussion)
Cleaning samples Do not use organic solvents as
these are always contaminated, even the fresh ‘electronic grade’ material in brown glass bottles
Never, never, use squeeze or spray bottles as the TEFLON filler goes into solution
Use detergents instead e.g. Alconox ‘Detergent 8’ which are bio-degradeable and leaves no residue
Carbon Dioxide ‘snow’ cleaning -no residue and good solvent action but expensive to set up. www.co2clean.com
CO2 “snow gun” for sample cleaning
Clean is not for ever ...
As soon as a specimen is prepared for observation it begins to get dirty again (CCW rule: “one monolayer/sec at 10-6 Torr)
Even storing the sample in a vacuum dessicator will not prevent the growth of bacterial or microbial surface contaminant films because the source of the problem is carried in by the specimen itself
Remedial action is required
As prepared
After one week
Plasma cleaning
Plasma cleaning is a rapid and easy way of removing the build-up of surface contaminants
Fast and non-destructiveSame sample after plasma cleaning
The Dark Side of SEM
The interaction of electrons with solids results in a variety of interactions which give us uniquely valuable information about the sample
But these same interactions can also result in either temporary or permanent damage to the sample
Know your enemy!
Unwanted Beam Interactions
Intrinsic to electron beam irradiation
Radiation Damage
IonizationDisplacementHeating
ContaminationEtching
Results fromvacuum problems
Both are usually important
Unwanted beam interactions
Electron beams have bad effects on organic, polymeric, and ionic materials
This is ‘radiolysis’
Effect of 0.01C/cm on
protein protoxin
500nmShrinkage of ArF resist 1mC/cm2
Radiolysis is….Radiolysis is the breaking
of bonds as the result of ionization by the electron.
Electrons are the most intense source of ionizing radiation available - the typical dose in an SEM is equivalent to standing 6 foot from a 10 megaton H-bomb
Compare SEM to Sun and SPEAR
Radiolysis damage in Polymers
In polymers radiolysis produces swelling or shrinking in the material and the actual loss of the sample
Despite appearances this damage is not due to heating in the sample
The effect may be reduced by coating with metal or a thin carbon layer
Courtesy Dale Newbury NIST
Dose does matter A typical SEM dose for
a photo-record is about 0.1 C/cm2 or 100 el/Å2
Typically at 1 -10el/Å2 we see a
loss of crystallinity at 10-100 el/Å2 mass loss and above 100 el/Å2 limiting mass loss
Dose from a single photo scan
Is a high beam energy bad?
It is often said that low beam energies minimize or eliminates beam induced damage
From casual observation this statement may appear to be true, but physics and measurements show that the truth is just the opposite
And note - even a very low energy electron (1eV) has an equivalent temperature of 10,000oK, which is hotter than the surface of the sun
‘Mythbuster’ fact All electrons damage At low energies the
damage is high but limited by the range
Damage is a maximum when range & feature sizes are similar
At higher energies damage falls - energy deposition occurs outside the feature
Adapted from Egerton (2004)
Range<<size damage limited
Range~size damage is maximum
Range>>size damage limited
Damage in semiconductors
e- beam damage of devices shifts the threshold voltage
damage is localized in gate oxides and is usually reversible
Damage depends on the beam energy…. and generally appears to get worse as the energy is increased
Thermal damage?Not usually a serious problem as the
energy deposited is quite smallFor a typical material of medium density
and thermal diffusivity the temperature rise with energy and beam dose is minimal
Magnification 5keV 15keV 30keV
400x 0.1C/nA 0.24C/nA 0.56C/nA
4000x 0.15C/nA 0.34C/nA 0.79C/nA
Other beam induced damage
In addition to Radiolysis the beam can produce ‘knock-on’ damage
In this the incident electron strikes an atom head on and knocks it out of position generating vacancies, e.g. Frenkel defects
This requires a minimum beam energy before it can occur, the value varying with the atomic number of the sample
For Carbon (Z=6) the knock-on threshold energy is about 80keV, for Silicon (Z=14) the knock-on threshold is 220keV
Not currently a problem with SEMs
Contamination and Etching
Contamination is beam induced polymerization of the hydrocarbons present on the sample surface
Etching is the removal of surface layer by impact of ions (C + H2O 2- --> CO + H2 )
Both phenomena are affected by surface charging and often occur together
Both are temperature dependentYour microscope is not to blame!
Modern SEMs are very clean
RGA of S4300 chamber just before the
specimen is inserted
H2O Nhydrocarbon
but samples are not..
S4300 chamber vacuum just after sample insertion
hydrocarbons from the SEC pump
Contamination and Etching
Electrons break down the hydrocarbon film by radiolysis.
The residue charges +ve and the field pulls in fresh material for radiolysis. If water vapor is present then negative drift to + ve
charged regions and can etch that area away
Low magnification At low magnifications the
hydrocarbon film is polymerized into a thin sheet.
This will charge positive (and so look black in the SE image) but is not a serious problem
Minimize by pre-exposing the sample at the lowest possible magnification prior to examination
Schematic of contamination build-up at low magnification
scans
Black squares... The black squares are visible
evidence of the charging that occurs
Post facto in situ removal of contamination is possible using plasma sources in the chamber although the process is slow
Use plasma cleaning before observation for best results
Example shown is by courtesy of Dr. Bryan Tracy, Spansion Inc.
High magnification
At high magnification the contamination grows a cone which scatters the beam
Avoid spot mode - always keep the beam scanning the sample
Pre-pump samples before use Keep your hands off the sample Avoid the use of dirty solvents Plasma clean before use if
possible
~ 0.03
Ant-hill contamination
Virtue of necessity..
Contamination cones can grow to a height of tens of nanometers and are so tough they are used for high resolution AFM tips
Measured diameters of carbon nanotubes can be high by half an order of magnitude!
Can prevent this growth by pre-irradiating the area at low magnification before going up to a high magnification
30nm high cones grown on a silicon wafer in spot mode - 1min
Temperature effects Altering both the
temperature of the sample and its surroundings will switch contamination to etching as the temperature falls
This is because water vapor condenses out on the sample surface and etches the contaminant
But the situation is unstable and leads to sample erosion
Temperature Effects II
Holding the sample at room temperature, but placing a cold surface close to it, can dramatically reduce the contamination rate
At a low enough temperature the situation becomes stable and non-contaminating
Such a device is called a “Cold Finger”
It is actually a disc placed just above the sample surface
operate here
The Cold Finger Standard fitting on
S4700, and beyond and available as an option for the S4500
The finger is held at LN2 temperatures, a few mm from the specimen surface
After allow the sample enough time to reach thermal equilibrium before starting to image
Without a cold finger
This high resolution image of gold on carbon disappears in just a few seconds of observation because of the contamination build-up that occurs
With a cold finger in use...
The equivalent area stays clean and high in contrast for an extended period of time.
Remember to give the sample time enough to reach thermal equilibrium before trying to achieve high resolution
Controlling contamination “Cold fingers” are a good start Beam blanking during flyback and
settling periods reduces LHS edge contamination in the S5500
Anything that reduces charging also reduces contamination and/or etching - so coating samples, pre-cleaning them, heating them prior to observation etc. all help
Keep beam currents and magnifications low, use minimum dose procedures, work fast.
50 nm
The combination of a cold finger and maintaining the sample at a low temperature (-90C) eliminates contamination