Upload
auth
View
0
Download
0
Embed Size (px)
Citation preview
A general approach to the derivation of peak area flow dependence inFIA and HPLC amperometric detection
P. Agrafiotou, C. Maliakas, A. Pappa-Louisi 1, S. Sotiropoulos *
Physical Chemistry Laboratory, Department of Chemistry, Aristotle University of Thessaloniki, Thessaloniki 54124, Greece
Received 18 March 2003; received in revised form 30 March 2003; accepted 30 March 2003
Abstract
The dependence of the peak area, Q , on the analyte volumetric flow rate, U , in flow injection analysis (FIA) and high
performance liquid chromatography (HPLC) with amperometric detection, was studied for typical electroactive species and in a
wide flow rate range. Based on the hydrodynamics of thin channel flow cells (as established by steady state experiments) and simple
dispersion theory considerations, a linear relationship between log Q and log U with a �/2/3 slope has been derived in a general
manner (irrespective to the type of dispersion) for amperometric detectors operated in the limiting current potential region. In the
case of mixed mass transfer and kinetic control the variation of Q with U is more complicated but the peak area is still smoothly
decreasing with the flow rate. These predictions were found to be in reasonable agreement with experiment for a few indicative
systems both in FIA and HPLC experiments. On the contrary, the non-steady state current corresponding to the peak maximum of
FIA and HPLC exhibited local maxima and the former could not be described by any of the equations proposed for the dispersion
in FIA experiments. The practical implications of the form of integrated signal, Q , dependence on flow rate for FIA and HPLC
amperometric detection are also discussed.
# 2003 Elsevier Science Ltd. All rights reserved.
Keywords: Electrochemical detection; HPLC; FIA; Flow rate
1. Introduction
Electrochemical detection has long been established
as a useful detection method in flow injection analysis
(FIA) and high performance liquid chromatography
(HPLC) due to a number of advantages it offers namely,
high sensitivity and selectivity, low cost and detector
miniaturisation [1�/4]. The latter is of great importance
in chromatographic and flowing analysis since mini-
misation of detector dead volume limits additional
sample dispersion and chromatographic peak distortion.
This is particularly useful for the current trend to
employ microbore columns and low dispersion chroma-
tography in order to achieve faster analysis with
minimal sample volumes [5].
Among the various types of amperometric and
coulometric electrochemical sensors for FIA and
HPLC [4�/6], perhaps the most widely used is that of
the thin channel flow cell, made of two blocks and a
spacer. The indicator electrode is usually glassy carbon
(GC), carbon paste electrode (CPE), Au or interdigi-
tated electrodes (IDE), embedded in an insulating block
and separated by the opposite block (which can be the
counter electrode) by an insulating spacer which defines
the channel thickness [7,8].
Analytical expressions for the steady state, mass
transport limited current at rectangular electrodes in
thin channel flow cells have been derived by Weber and
Purdy [9,10], Roosendaal and Poppe [11], Moldoveanu
and Anderson [12] and Compton and Daly [13].
Numerical solutions to the equations describing the
cell hydrodynamics are also due to Anderson and co-
workers both for macro-electrodes [14] and microarray
electrodes [15], to Compton and co-workers [16] for
microband electrodes and to Bond and co-workers for
microdisc electrodes [17]. In these expressions, the
* Corresponding author. Tel.: �/30-2310-99-7742; fax: �/30-2310-
44-3922.
E-mail addresses: [email protected], [email protected]
(S. Sotiropoulos).1 ISE member.
Electrochimica Acta 48 (2003) 2447�/2462
www.elsevier.com/locate/electacta
0013-4686/03/$ - see front matter # 2003 Elsevier Science Ltd. All rights reserved.
doi:10.1016/S0013-4686(03)00271-8
steady state current is related to the cell geometric
characteristics, the solute and solvent properties and,
more important, to the solution flow rate. In cases that
radial diffusion to the electrode can be neglected (as it isthe norm except for very small electrodes and wide
channels), the steady state limiting current at an
amperometric detector has been found both theoreti-
cally and experimentally to vary with the cube root of
analyte flow velocity. However, there is little work on
the prediction of the non-steady state peak current in
HPLC (see for example Ref. [18] where a complicated
expression for the peak current as a function of flowrate, based on the van Deemter model for the plate
height, has been fitted to experimental data). At the
same time, we are only aware of a single paper
discussing the effect of flow on peak current in FIA,
based on experiments limited in a narrow flow rate
range [19] and proposing a U�1/6 dependence of the
peak current. There has not either been any attempt to
transform existing theoretical expressions for the flowdependence of dispersion coefficient, D [20�/22] to
corresponding peak current expressions in FIA with
electrochemical detection. Note that the former propose
conflicting dependencies of D on either U1/2 [20] or
U�0.206 [21,22].
Apart from the peak height, a very useful parameter
in quantitative FIA and HPLC analysis is the area under
the peak which, in the case of electrochemical detection,corresponds to the charge passing through the electrode.
The peak area is independent of the type of dispersion
(hence, of system geometric characteristics and arrange-
ment) in FIA and of peak asymmetry (hence of ideal
column behaviour) in HPLC, since it is directly related
to the amount of analyte eluted [23,24]. In the case of
electrochemical detection, the peak area corresponds to
the amount of analyte having reacted as it passesthrough the detector. Despite the usefulness of this
parameter, there is again a single paper discussing the
variation of amperometric detection peak area with flow
rate and testing it in FIA experiments [25]; that
theoretical derivation, based on the analytical solution
of concentration profile equations and predicting a
U�2/3 dependence of Q on U , was limited to the cases
of extended analyte dispersion (Taylor dispersion regime[26,27], i.e. small samples, long reactors, low-to-moder-
ate flow rates [28]). Its validity was experimentally tested
only for a few indicative velocity values and only for
FIA experiments with a single species (not for HPLC
detection or a variety of analytes).
The main aim of this paper is to propose a general
approach to the prediction of the effect of flow rate on
peak area both in FIA and HPLC with amperometricdetection. In more detail, its main objectives in order of
discussion are (i) to test expressions for maximum
current in FIA with electrochemical detection based on
the most popular dispersion equations, (ii) to derive in a
simple and general manner the peak area-flow rate
dependence in FIA and HPLC and test its validity for a
number of systems and experimental conditions and,
(iii) to compare the effect of flow rate on peak area inelectrochemical detection under mass transfer control
with that under mixed or kinetic control (as well as with
that in UV-photometric detection) and discuss the
practical implications for FIA and HPLC analysis.
2. Experimental
2.1. FIA and HPLC system
A Schimatzu model LC-9A pump and a Rheodyne
model 7125 syringe-loading sample injector, fitted with a
20 ml injector loop, have been used in all FIA and HPLC
experiments. The volumetric flow rate range studied was
0.01�/3 ml min�1.
Teflon tubing of 0.01 in. (0.025 cm) ID from Alltech
was employed; its length up to the inlet of the detector inFIA was 24.5 cm (in one set of experiments this length
was 100 cm). This corresponds to a tubing volume
(‘reactor volume’), VR, of 0.012 ml for the 24.5 cm long
tubing and 0.049 ml for the 100 cm one. Since the
sample volume is VS�/0.020 ml, it follows that the ratio
VS/VR is 1.66 and 0.44, respectively, i.e. relatively large
samples and small volume tubing have been used and
hence not very high dispersion is expected in our FIAexperiments, at least at moderate and high flow rates.
Note that Valcarcel et al. (Ref. [28], p. 91) give as an
empirical criterion for limited dispersion in FIA, at flow
rates higher than 2 ml min�1, that of VS/VR�/0.2. Thus,
for the samples and tubing used in this study, the Taylor
dispersion regime (large dispersion, pure diffusional
transport) is only expected for the lowest flow rates
used (for a more quantitative argument see Section 3below).
A 250�/4.6 mm i.d. Kromasil C18 reverse phase
column of 5 mm particle size from MZ-Analyzentechnik
was employed in HPLC experiments with electrochemi-
cal detection, whereas an Alltech 100�/4.6 mm Adsor-
bosphere Catecholamine 3U reverse phase column was
used in the few UV detection experiments carried out.
Again, 24.5 cm long Teflon tubing of 0.01 in. (0.025 cm)ID from Alltech were used both for the column inlet and
outlet.
2.2. Thin layer flow cell and electrochemical
instrumentation
A commercial two-block flow cell was the electro-
chemical detector (Gillson model 141). It consisted of aTeflon† block where a 3 mm diameter GC indicator
electrode is embedded, a stainless steel block serving as
the counter electrode, and 100 mm thick Teflon† spacers
P. Agrafiotou et al. / Electrochimica Acta 48 (2003) 2447�/24622448
(used individually or in stacks of two or three) to
separate the two blocks. The reference electrode (Ag/
AgCl in 3 M NaCl) was screwed into a port drilled
through the stainless steel block; the solution inlet andoutlet ports were on the stainless steel and Teflon†
blocks, respectively. This arrangement was similar to
that of many flow cells that appear in the literature and
first introduced by Poppe and co-workers [29] and
Wightman and co-workers [30]. The opening in the
spacer was such that the channel length was 1.2 cm and
its width 0.5 cm; these, together with the spacer
thickness of 100 mm or 0.01 cm, result in a total detectorvolume of 0.006 ml and a dead volume of 0.00225 ml
from the cell input up to the leading electrode edge (for a
single spacer). In some experiments a modified cell was
used whereby the working electrodes were four carbon
fibres 30 mm thick (WPI Inc.), attached to the Teflon
block and placed perpendicular to the flow.
A laboratory-made potentiostat and a computer-
based data acquisition system were used for electro-chemical detection.
2.3. Chemicals
Ascorbic acid (99.9%) and hydroquinone (99%) from
Aldrich were used in steady state experiments asreversible electrochemical probes. They were also used
together with m -dopa (L-a-methyl-dopa or 2-methyl-3(-
3,4,-dihydroxy-phenyl-L-alanine) from Sigma Chemicals
(99.5%), 5-hydroxy-tryptophan (5-htp; Biochemica, 99%
NT) from Fluka and potassium ferrocyanide from
Aldrich (99.9%) in FIA and HPLC experiments. In
HPLC with UV detection, vanillylmandelic acid (99.5%)
from Sigma Chemicals Co. was used. The supportingelectrolytes in the carrier solution (FIA), the mobile
phase (HPLC) and the sample solutions were phosphate
buffers of required pH and ionic strength values [31].
Potassium phosphate, potassium dihydrogen phosphate,
potassium hydrogen phosphate and phosphoric acid for
buffer preparation were analytical reagent grade. In
HPLC experiments with m -dopa and 5-htp, the mobile
phase included 10% v/v of methanol from Sigma (HPLCgrade).
3. Results and discussion
3.1. Theoretical considerations
3.1.1. Steady state current at a disc electrode in a thin
channel flow cell
The dimensionless number correlation describing theflow in a parallel plate cell with short rectangular
electrodes and for fully developed laminar flow is of
the general form [32]:
Sh�1:85FRe1=3Sc1=3
�de
Lel
�1=3
(1)
where Sh�kmde=D is the Sherwood number,//
/Re�(vde=n)�((U=S)de=n) is the Reynolds number,
Sc�n=D is the Schmidt number, km the electroactivespecies mass transport coefficient, D its diffusion
coefficient, n the solution kinematic viscosity, v the
solution mean linear velocity, U the solution volumetric
flow rate, S the channel cross-section, Lel the electrode
length and de the cell hydraulic diameter. The latter is
equal to four times the area S of the cross-section
divided by its perimeter P and in the case of a
rectangular channel of width wch and thickness dch isgiven by:
de�4S
P�
4(wchdch)
2(wch � dch)�
2(wchdch)
(wch � dch)(2)
F is a parameter in the 0.788�/1 range, depending on cell
geometry (Ref. [32], p. 24). For the equivalent disc
electrode length (Lel)eq�/0.2658 cm (i.e. the length of a
square electrode having the same area as the disc
electrode of del�/0.3 cm diameter; (Lel)eq2 �/pdel
2 /4),
channel width wch�/0.5 cm and thickness dch�/0.01
cm of our cell, F approaches 1 [32]. Eq. (1) (with F�/1)has also been proposed by Picket et al. [33] to hold for
[33,34] ReB/2000, wch�/dch and L /deB/35. Based on the
cell geometric characteristics presented above, it readily
follows that the second condition holds whereas from
Eq. (2) we have de�/0.0196 cm and (Lel)eq/de�/13.5B/
35. Furthermore, for the U�/0.01�/3 ml min�1 range
used, the cell dimensions, and a typical value of 0.011
cm2 s�1 for n , very low Reynolds numbers result (in theRe�/0.059�/17.82 range) and hence laminar flow is
expected in our case. Finally, the distance between the
solution inlet and the electrode is 0.5 cm in the cell used,
i.e. much higher than the entry length of le�/0.01deRe�/
35 mm (calculated for the maximum value of Re)
required according to Ref. [32], p. 18 for fully developed
laminar flow. Summarising, all conditions are met for
Eq. (1) (with F�/1) to describe our steady stateexperiments.
If we take into account that the steady state mass
transport limiting current is given by:
ISS�nFAkmC (3)
where A is the electrode area and km the analyte mass
transfer coefficient, then, substituting the expressions ofthe dimensionless numbers in (1), solving for km and
substituting in (3), it follows:
ISS�1:85nFCAD2=3 1
L1=3
el
1
d1=3e
�U
dchwch
�1=3
(4)
For dch�/wch (as is in our case), the hydraulic diameter
can be approximated according to (2) by de�/2dch. Then
P. Agrafiotou et al. / Electrochimica Acta 48 (2003) 2447�/2462 2449
Eq. (4) becomes:
ISS�1:47nFCAD2=3 1
L1=3el
1
d2=3ch
�U
wch
�1=3
(5)
For rectangular electrodes, where A�/welLel, Eq. (5)
becomes:
ISS�1:47nFCwelD2=3L
2=3el
1
d2=3
ch
�U
wch
�1=3
(6)
which is the equation derived by Weber and Purdy [9,10]
and Roosendaal and Poppe [11] and widely used for
rectangular electrodes in thin channel cells. However,
although there exists an analytical solution for a flow
perpendicular to a disc electrode [6] and a numerical
solution for a flow parallel to microdisc electrodes [17],we are not aware of an analytical solution for the
current at a disc electrode parallel to the flow in a
parallel plate cell. We can attempt an approximation by
modifying Eq. (5) and replacing the electrode length
parallel to the flow, Lel, by an equivalent electrode
length (Lel)eq that corresponds to the length of a square
electrode having the same area as the disc electrode of del
diameter. It then holds by definition (Lel)eq2 �/pdel
2 /4 andsubstituting (Lel)eq from this expression and A�/pdel
2 /4
in (5), we get:
ISS�1:20nFCd5=3el
�D
dch
�2=3� U
wch
�1=3
(7)
It should be noted that if one used as Lel simply the discdiameter del then this would result to a small (ca. 4%)
underestimate of the current. According to Eqs. (5)�/(7),
both for rectangular and disc electrodes, the steady state
current in amperometric detection mode is expected to
be proportional to the cubic root of the flow rate U .
3.1.2. Peak current in FIA experiments with an
amperometric detector
A useful parameter that characterises the extent of
dilution of the injected sample up to the point of
detection (and hence the shape of the FIA peak too) is
the dispersion coefficient D introduced by Ruzicka et al.
[20]. It is defined as the ratio of the injected sample
concentration Cs to the average (in the radial tubing
direction z , see Fig. 1(A) and (B)) concentration
C(xD; t)/of the dispersed analyte over the detector (atan axial coordinate position same as the axial coordi-
nate of the detector, x�/xD, see Fig. 1(B)) at a particular
elution time:
D(t)�CS
C(xD; t)(8)
This means that each point in the concentration
versus time profile (hence in the FIA peak too)
corresponds to a different value of D. At the peak
maximum the concentration of the dispersed analyte
over the detector takes its maximum value and hence the
dispersion coefficient takes its minimum value Dmin (the
subscript is often dropped in the literature):
Dmin�D�CS
Cmax(xD)(9)
When detectors responding to the average radial
analyte concentration are used (such as for example
UV spectrometers) then, introducing the steady state,
non-steady state and maximum signals Sss�/KCs,/S�KC(xD; t); Smax�KCmax(xD) into (8) and (9) is
straightforward (where K is a proportionality constant).
However, in the case of an amperometric detector, as
pointed out by Meschi and Johnson [19], the signal
(mass transport limited current) is proportional to the
analyte concentration at a distance from the electrode in
the radial direction equal to the length of the diffusion
layer d , i.e. at z�/d (see Fig. 1(B)):
I �KelC(xD; d; t) (10)
and for the peak current:
Imax�KelCmax(xD; d) (11)
However, the concentration profile of the dispersed
analyte is not in general homogeneous in the z direction.
This is depicted in Fig. 1 which presents the parabolic,
highly non-homogeneous profile (A) for dispersion in
the convective regime (very short tubing, very highvelocities) and a more homogenised one (B) for disper-
sion in the diffusive or in the mixed convective�/diffusive
regimes [19,28]. It follows that, in general, the average
concentration in the z direction above the detector
C(xD; t)/does not coincide with C (xD, d , t). This
difference is expected to be significant in the case of
purely convective transport (very high flow rates, very
small tubing lengths) but it is expected to graduallydiminish as diffusive transport gets into play at moder-
ate flow rates and tubing lengths (mixed transport,
Taylor dispersion, Taylor�/Aris dispersion) and radial
diffusion tends to homogenise the concentration profile.
Indeed, Meschi and Johnson (Ref. [19], Fig. 2 therein)
have shown that the FIA currents calculated based on
Eq. (10) differ very little from those calculated based on
the average concentration in cases of Taylor-typedispersion. Furthermore, they proved that in these cases
the maximum of the concentration variation with time
at z�/d from the electrode coincides with the maximum
average concentration at the same location, i.e. Cmax(xD,
d )�//Cmax(xD): All these facts point out that, in all cases
but those of pure convective dispersion, Eq. (11) can be
replaced by Imax�KelCmax(xD): On the other hand, the
steady state (limiting) current can be written accordingto (7) as ISS�/KelCS. Hence, for an amperometric
detector in FIA experiments of mixed or diffusion-type
dispersion, Eq. (9) becomes:
P. Agrafiotou et al. / Electrochimica Acta 48 (2003) 2447�/24622450
D�ISS
Imax
(12)
It follows that, knowledge of the flow rate dependence
of the dispersion coefficient D , in connection with the
cubic root dependence of Iss on the flow rate (Eq. (7),
would provide via (12) the dependence of the peak
current Imax on U . There have been two main equations
suggested in the literature for the dependence of D onU .
The first one is due to Ruzicka et al. ([20] and Ref.
[28], p. 76) and has been derived analytically for Taylor
dispersion conditions:
D�KDU1=2 (13)
where KD is a constant. Given that according to (7) Iss�/
KssU1/3 where Kss a constant, this means that Eq. (12)
results in:
Imax�(KSS=KD)U�1=6�KImaxU�1=6 (14)
where KImax�/KSS/KD.
It should be noted that Meschi and Johnson [19] have
derived the same �/1/6 dependence solving directly the
problem of peak current rather than using FIA theory.
The second equation is empirical and was derived by
Valcarcel and co-workers [22,28] who fitted a large
number of experimental results, in the mixed convectiveand diffusive dispersion regimes, into equations with
adjustable parameters:
D�K ?DU�0:206 (15)
Again, according to (7) Iss�/KssU1/3, and Eq. (12) results
in:
Imax�(KSS=K ?D)U0:54�K ?ImaxU0:54 (16)
At this point, it would be useful to introduce the
parameters whose values define in a quantitative mannerthe type of dispersion in FIA experiments. According to
References [35,28] (pp. 61�/63), a convenient parameter
for dispersion classification is the ratio of reactor
(tubing) length, L , to the flow rate, U . They give the
following indicative limits for a typical analyte diffusion
coefficient of 1�/10�5 cm2 s�1: for L /U B/2.1 cm min
ml�1, the dispersion is governed by pure convection
(very short times, very large sample volumes, very smallreactor volumes, very high flow rates); for 53.1B/L /
U B/159.1 cm min ml�1, mixed convective and diffu-
sional transport prevails (the usual situation under
Fig. 1. (A) Schematic representation of the entire injected analyte dispersion profile in a tubular FIA reactor for convective mass transport
conditions; v denotes the mean linear flow velocity. (B) Schematic representation of injected analyte dispersion in the vicinity of the electrochemical
detector (depicted with its centre positioned at x�/xD) in a tubular FIA reactor for diffusive or mixed mass transport conditions; dVx is the
elementary part of reactor volume above the detector and C(xD)/the mean analyte concentration in that volume; v denotes the mean linear flow
velocity and d the thickness of the electrochemical diffusion layer.
P. Agrafiotou et al. / Electrochimica Acta 48 (2003) 2447�/2462 2451
practical FIA conditions); for L /U �/425 cm min ml�1,
diffusional transport prevails (Taylor regime, long
times, small samples, long reactors, very low flow rates).
3.1.3. Peak area in FIA and HPLC experiments with an
amperometric detector
Let Cs and Vs be the concentration and volume of the
injected sample in a FIA or HPLC experiment. If C is
the analyte concentration of an elementary volume dV
in the dispersion zone then, the mass balance require-
ment between the injected sample and the dispersedsample in the total system volume, Vtotal, can be written
in terms of analyte mol at all times as:
CSVS� gVtotal
0
CdV (17)
Let us also consider an elementary volume dVx at an
axial coordinate x in the direction parallel to the flow
and extending over the entire channel diameter in a
direction perpendicular to the flow, as shown in Fig.
1(B). We can then replace in (17) the sum of all
elementary mol quantities CdV within the elementaryvolume dVx by Cx/dVx , where Cx/is the average analyte
concentration in the radial direction within the elemen-
tary volume dVx .
Thus, Eq. (17) can be rewritten as:
CSVS� gVtotal
0
CxdVx (18)
Eqs. (17) and (18) express the mass balance at all times
in the space domain.
In a flowing system the entire quantity of the
dispersed analyte has to pass from any specified point
or elementary region of the system over the period of its
residence in the system, tr, i.e. from the time of injectiont�/0 until the time t�/tr when all the analyte has passed
by this point or elementary region.
We can choose this point or elementary region to be
above the detector (of very small dimensions), i.e. close
to a location of an axial coordinate x�/xD. If we take
into account that the volumetric flow rate in the system
in general, and over the detector specifically, is U �dVxD
=dt then, it follows that the number of analyte molwithin the elementary volume dVxD
at any time is:
C(xD; t)dVxD�UC(xD; t)dt (19)
Then the mass balance of (18) can be expressed in the
time domain as:
Fig. 2. Steady-state current, Iss, recorded at �/0.8 V vs. Ag/AgCl on a GC disc electrode in a thin channel cell as a function of volumetric flow rate,
U , from flowing solutions of 5�/10�6 M ascorbic acid �/0.05 M phosphate buffer (pH 5) and for interelectrode gaps of 100, 200 and 300 mm as
indicated on the graph. Inset; log�/log plots of Iss vs. U .
P. Agrafiotou et al. / Electrochimica Acta 48 (2003) 2447�/24622452
CSVS�U gtr
0
C(xD; t)dt (20)
As discussed in Section 3.1.2 above, the signal-current
of an amperometric detector is at any time proportionalto the analyte concentration at a radial coordinate
above the electrode equal to the diffusion layer, C (xD,
d , t). Due to the non-homogeneous profile of concen-
tration in general, the latter expression for concentration
does not coincide with the average radial concentration
of Eq. (20) but for the case of purely diffusive mass
transport (Taylor regime), where the concentration in
the radial direction is fully homogenised.However, the left-hand-side of Eq. (20), CsVs, is equal
to the number of injected analyte mol and hence
independent of the type of dispersion and associated
concentration profile. Therefore, the integral of the
right-hand-side of (20) should be independent of these
parameters too. Hence, in calculating this integral for
any type of experimental conditions and dispersion we
can use (with no loss of generality) instead of the realconcentration profile an equivalent profile with a
constant concentration along the radial direction (per-
pendicular to the flow). For this equivalent profile the
above-mentioned concentrations coincide, i.e. C (xD, d ,
t)�//C(xD; t); and hence we can use Eq. (10) to
introduce the current in Eq. (20):
CSVS�U
Kelgtr
0
Idt (21)
The integral in Eq. (21) is the total charge, Q , passing
through the amperometric detector as the eluted analyte
is oxidised or reduced and corresponds to the area under
the FIA or chromatographic peak. Taking also intoaccount (as discussed in Section 3.1.2) that the steady
state current is Iss�/KelCs, and substituting for Kel in
(21), we have:
Q�ISSVS
U(22)
From Eq. (7) we have for the flow rate dependence ofthe current that Iss�/KssU
1/3 (where Kss sums up all
analyte and cell constants of (7)) and, therefore, the
peak area dependence for amperometric detection
becomes:
Q�KQU�2=3 (23)
where KQ�/KssVs, a constant.
The simple form of peak area flow rate dependence
predicted by Eq. (23) is only valid if the potential of
electrochemical detection of the analyte is in the masstransfer control potential range, since it is only then that
the steady state current is given by Eq. (7). However, in
many cases of electrochemical detection the electrode
reaction is under mixed kinetic and mass transfer
control. In that case, the steady state current is made
up of a kinetic current contribution, (Iss)k, which is
independent of flow rate and depends only on reactionkinetics and electrode potential, and a mass transfer
controlled current contribution, (Iss)m, which is given by
Eq. (7). The three currents are related by:
1
ISS
�1
(ISS)k
�1
(ISS)m
(24)
which, if we replace the mass transfer current by
(Iss)m�/KssU1/3 and solve for Iss, gives:
ISS�(ISS)kKSSU1=3
(ISS)k � KSSU1=3(25)
Putting Eq. (25) into (22) we obtain for the peak area of
amperometric detection under mixed kinetic and masstransfer control:
Q�VSKSS(ISS)k
(ISS)kU2=3 � KSSU�
VSKSS
U2=3 �KSS
(ISS)k
U
(26)
It can be seen that the dependence of Q on U is also
affected in this case by the standard rate constant of the
electrochemical reaction of the particular analyte and
the electrode potential, as expressed by the kinetic
current (Iss)k.
In the limit of pure mass transfer control (Iss)k�/Kss
and Eq. (26) is diminished to the form of Eq. (23)
(taking into account that VsKss�/KQ). In the limit ofpure kinetic control (Iss)k�/Kss and (26) becomes:
Q�VS(ISS)kU�1 (27)
At this point we would like to comment briefly on the
very popular technique of pulsed amperometric detec-tion (PAD) which is also being used extensively in the
chromatographic detection of important classes of
organic molecules (see for example Ref. [2], pp. 145�/
148 and Ref. [4], pp. 836�/840). Since the pulse duration
is very short (a few tens of ms) and the diffusion layer is
smaller than the convective shear layer, convection has
no effect on the current and hence the steady state pulse
current, (Iss)p, is at all potentials independent of flowrate. Therefore, by putting (Iss)p in (22) it follows that, in
PAD too, the peak area is given by an equation similar
to (27), i.e. it is proportional to U�1.
It can be seen that in the case of simple amperometric
detection the log Q versus log U plots are only predicted
to be linear in the limits of pure mass transfer or kinetic
control, with a slope of �/2/3 or �/1, respectively. In all
other cases, the dependence of Q on U takes the morecomplicated form of (26) and the slope of the empirical
linear approximation of the log Q versus log U curve is
expected to have values between �/2/3 and �/1.
P. Agrafiotou et al. / Electrochimica Acta 48 (2003) 2447�/2462 2453
It is interesting to note that the case of amperometric
detection under pure kinetic control is formally the
same, with respect to peak area flow dependence, to the
case of spectrophotometric (e.g. UV) detection, with adetector signal S independent of flow rate and a peak
area A. Then Eqs. (22) and (23) become:
A�SSSVS
U�KAU�1 (28)
3.2. Steady state experiments for flow cell
characterisation
Fig. 2 shows the variation with volumetric flow rate of
the steady-state limiting current for the two-electron
oxidation of ascorbic acid, recorded at �/0.8 V versus
Ag/AgCl on a GC disc electrode in the thin channel cell
of this work. Flowing solutions of 5�/10�6 M ascorbicacid and phosphate buffer of pH 5 and 0.05 M ionic
strength were used and three inter-electrode distances
were tested (100, 200 and 300 mm depending on whether
a single spacer or a stack of two or three spacers were
employed). The experiments were carried out to check
the validity of the steady state current flow rate
dependence predicted by Eq. (7) for a disc electrode in
a thin channel cell. The Inset in Fig. 2 shows log Iss
versus log U plots exhibiting very good linearity and
slopes close to the theoretically expected slope of 0.33
according to (7).
Furthermore, if we substitute in the same equation
n�/2, D�/6.9�/10�6 cm2 s�1 [7] and C�/5�/10�6
M�/5�/10�9 mol cm�3 for ascorbic acid, del�/3
mm�/0.3 cm, dch�/100 mm�/0.01 cm (one spacer) and
wch�/0.5 cm for our thin channel cell, and U�/1 mlmin�1�/1/60 cm3 s�1 (as an indicative flow rate), we
predict a theoretical current of (Iss)th�/391.2 nA which
is very close to the experimentally observed value of
(Iss)exp�/400 nA. The same tests on the validity of Eq.
(7), were performed using the two-electron oxidation of
hydroquinone (5�/10�6 M in 0.02 M phosphate buffer
of pH 4.68; n�/2, C�/5�/10�6 M, D�/8.5�/10�6 cm2
s�1 [9]) as a model reaction and a ratio of (Iss)exp/(Iss)th�/0.97 was obtained. A summary of results
obtained both for ascorbic acid and hydroquinone
oxidation steady-state experiments is presented in Table
1.
3.3. Variation of peak current with flow rate in FIA
experiments
Fig. 3(A) shows the amperometric responses at twoindicative flow rates, during FIA experiments with a
0.05 M phosphate buffer (pH 5) carrier solution and a
5�/10�6 M ascorbic acid�/0.05 M phosphate buffer
(pH 5) injected sample of 20 ml (reactor length of 24.5
cm). Fig. 3(B) shows similar curves for an experiment
with a 0.05 M phosphate buffer (pH 3) carrier solution
and a 5�/10�6 M m -dopa �/0.05 M phosphate buffer
(pH 3) injected sample of 20 ml (reactor length of 24.5
cm). The potential of the GC electrode was in both cases
�/0.8 V versus Ag/AgCl, i.e. in the limiting current
regime for both compounds (as preliminary variation of
the electrode potential showed) and the current was
measured from the baseline current. It can be seen that
in both cases, as the flow rate increases from 0.5 to 1 ml
min�1 the FIA peak becomes narrower and more
asymmetric, as simple dispersion theory predicts [28]
for the shift from conditions of mixed convective�/
diffusive transport to those of convective transport,
and the peak current increases as well.
A more detailed study of the flow rate effect on FIA
amperometric peak height for the two systems men-
tioned above (reactor length of 24.5 cm), as well as for a
5�/10�6 M hydroquinone �/0.02 M phosphate (pH
4.68) injected sample of 20 ml (reactor length of 100 cm),
is presented in Fig. 4 where the maximum current Imax is
plotted against the flow rate. A maximum is observed in
all three cases and this was also the situation for a
number of other analytes tested (e.g. ferrocyanide and
tryptophan derivatives). The inset shows the variation of
the dispersion coefficient D with flow rate for the
ascorbic acid and hydroquinone systems; D was calcu-
lated from Imax at each flow rate and the steady-state
current Iss for that flow rate, as obtained from the
steady-state experiments described in Section 3.2 above.
It can be seen that the dispersion coefficient shows an
increase for flow rates higher than 1 ml min�1 (the flow
rate range of the maximum in the Imax vs. U curves).
None of Eqs. (14) or (16) could be fitted to an
appreciable range of the results. This is not surprising,
for two reasons. First of all, both equations were
proposed for very limited parts of the dispersion regime,
whereas our results cover a wide range of flow rates. Eq.
(14) strictly holds for Taylor dispersion and when tested
by Meschi and Johnson [19] for a single experimental
system, with L /U values in the 298�/73.5 cm min ml�1
range, it was found that it could only predict the results
obtained at the lowest flow rate. Eq. (16) is based on the
empirical equation for the dispersion derived in Ref. [22]
and tested for L /U values in the 200�/20 cm min ml�1.
It should also be noted that the two equations proposed
by these authors are contradicting each other since Eq.
(14) predicts a decrease of the peak current with
increasing flow rate whereas Eq. (16) predicts the
opposite trend. Secondly, both the theoretical treatment
and the experimental set up of the above two studies
involved a flowing system of a linear and uniform
geometry whereas our system (as well as similar systems
based on electrochemical detection with a thin channel
P. Agrafiotou et al. / Electrochimica Acta 48 (2003) 2447�/24622454
cell) is characterised by different channel thickness for
the reactor-tubing (internal diameter, dr�/0.025 cm)
and the detector (channel thickness, dch�/100 mm�/0.01
cm) as well as a 908 change in flow direction upon entry
to the detector cell. Hence, flow conditions are expected
to change as the analyte passes from the reactor to the
detector.
Table 2 gives the Reynolds numbers both within the
reactor-tubing of the FIA system, (Re)r, and within the
thin channel detector (Re)d for the flow rates used. The
values of the dispersion type characteristic parameter L /
U and of dispersion coefficient D are also tabulated for
the ascorbic acid and hydroquinone experiments dis-
cussed above. The Reynolds number within the tubing-
reactor was calculated based on:
(Re)r�vdr
n�
(U ?=S)dr
n�
Udr
60Sn�
Udr
60(pd2r =4)n
(29)
where dr�/0.025 cm the tubing diameter, U? and U the
volumetric flow rates per second and minute, respec-
tively, and n�/0.011 cm2 s�1 the typical kinematic
viscosity of aqueous solutions. The Reynolds number
within the thin channel cell-detector was calculated (as
in Section 3.1.1 above) based on:
(Re)r�vde
n�
(U ?=S)de
n�
Udr
60Sn�
Ude
60(dchwch)n(30)
Table 1
Slopes and linear correlation coefficients of log�/log plots of the steady state current Iss vs. flow rate U (for three values of channel thickness dch), for
ascorbic acid and hydroquinone oxidation in a thin channel flow cell
Ascorbic acid Hydroquinone
1 spacer 2 spacers 3 spacers 1 spacer 2 spacers 3 spacers
Slope of log Iss vs. log U
plot
0.3170 (R2�/
0.9995)
0.3123 (R2�/
0.9995)
0.3234 (R2�/
0.9981)
0.342 (R2�/
0.9999)
0.3331 (R2�/
0.9998)
0.3291 (R2�/
0.9994)
Fig. 3. (A) Electrochemical detection FIA peaks recorded at �/0.8 V vs. Ag/AgCl on a GC disc electrode in a thin channel cell for a 20 ml injected
sample of 5�/10�6 M ascorbic acid in a 0.05 M phosphate buffer (pH 5) carrier solution and for flow rates as indicated on the graph. The arrow
indicates the point of injection. (B) Electrochemical detection FIA peaks recorded at �/0.8 V vs. Ag/AgCl on a GC disc electrode in a thin channel
cell for a 20 ml injected sample of 5�/10�6 M m -dopa in a 0.05 M phosphate buffer (pH 3) carrier solution and for flow rates as indicated on the
graph. The arrow indicates the point of injection.
P. Agrafiotou et al. / Electrochimica Acta 48 (2003) 2447�/2462 2455
where de�/0.0196 cm the hydraulic cell diameter (as
defined and calculated in Section 3.1.1), whereas dch�/
100 mm�/0.01 cm and wch�/0.5 cm the channel thick-
ness and width, respectively. From the values presented
in this table it can be seen that our results cover an
extended range of the diffusional and mixed transport
dispersion regimes unlike the results of the works of Eqs.
(14) and (16) which cover only parts of them. The
significant change in the Reynolds numbers as the
analyte passes from the reactor to the detector can be
readily seen. Although the volume of the latter is too
small when compared with that of the former (see
Fig. 4. Variation of peak current, Imax, with flow rate, U , in electrochemical FIA experiments under the conditions of Fig. 3 and for the same
ascorbic acid and m -dopa systems, as well as for a 20 ml injected sample of 5�/10�6 M hydroquinone in a 0.02 M phosphate buffer (pH 4.68) carrier
solution. Inset; log�/log plots of dispersion coefficient, D , and volumetric flow rate.
Table 2
L /U (reactor length/flow rate) parameter, reactor and detector Reynolds numbers, (Re)r and (Re)d, and dispersion coefficient, D , for various
volumetric flow rates, U , for ascorbic acid (L�/24.5 cm) and hydroquinone (L�/100 cm) FIA experiments
U (ml min�1) L /U (cm ml min�1) (Re)r (Re)d D
Ascorbic acid
(L�/24.5 cm)
Hydroquinone
(L�/100 cm)
Ascorbic acid
(L�/24.5 cm)
Hydroquinone
(L�/100 cm)
0.01 2450 0.78 0.06 2.9
0.03 817 2.3 0.18 2.2
0.05 490 3.9 0.30 2.4
425 lower limit for purely diffusional transport [28,35]
0.07 350 5.4 0.42 2.5
0.1 245 7.7 0.60 2.5
159.1 upper limit for mixed transport [28,35]
0.3 81.7 333.3 23.2 1.78 3.1 4.45
53.1 lower limit for mixed transport [28,35]
0.5 49 200 38.6 2.97 3.4 4.55
0.7 35 143 54 4.16 3.6 4.7
1 24.5 100 77.2 5.94 3.7 5.2
2 12.25 50 154.4 11.88 5.8 7.25
3 8.2 3.3 231.6 17.82 9.1 9.2
2.1 upper limit for purely convective transport [28,35]
P. Agrafiotou et al. / Electrochimica Acta 48 (2003) 2447�/24622456
Section 2) to affect to a considerable extent analyte
dispersion, such a drastic change in flow conditions
(combined with the abrupt change in flow direction) is
nevertheless expected to affect to some extend the Imax
versus U curve characteristics. At this point, it should
also be mentioned that the variation of D with U has
often be reported in the experimental literature to pass
through a minimum (see for example Ref. [28], p. 78) in
accordance with our results of the Inset to Fig. 4.
3.4. Variation of peak area with flow rate in FIA and
HPLC experiments
Fig. 5 shows the variation of the peak area-charge, Q ,
of FIA experiments with electrochemical detection for
the m -dopa system for which results have been pre-
sented in Fig. 3(B) and Fig. 4 above. The Inset gives the
log�/log plot of the peak area versus the flow rate and
excellent linearity is observed, with a slope equal to�/0.64, i.e. very close to the expected slope of �/0.66
according to Eq. (23) and our arguments in Section 3.1.3
above. Also, the applicability of (23) extends over a wide
flow rate range (0.03�/3 ml min�1 in this case) in
accordance with our prediction that the peak area-flow
rate relationship is independent of the type of dispersion
in FIA. A test of the validity of Eq. (22) (Q�/IssVs/U , U
in ml s�1 or Q�/IssVs/(U /60), U in ml min�1) was alsoperformed, based on the experimentally found Q values
for various flow rates and the Iss values found from
steady state experiments for the same flow rates, for the
ascorbic acid and hydroquinone systems. The Vs values
thus calculated were 19.59/2 and 21.39/1.7 ml, in
reasonable agreement with the nominal 20 ml of injected
sample.
Fig. 6 presents two typical electrochemical detection
HPLC chromatograms (at two indicative flow rates of
0.5 and 1 ml min�1 flow rates) for a 5�/10�6 M
m -dopa �/0.05 M phosphate buffer (pH 3) injected
sample of 20 ml, through the HPLC column described in
Section 2 and in a mobile phase of a 0.05 M phosphate
buffer (pH 3) aqueous solution �/10% v/v methanol.
Inset (A) shows the variation of the maximum peak
current Imax with the flow rate. Again, similar to our
FIA results, an unclear dependence is observed, not
markedly different to that presented for electrochemical
HPLC experiments in Ref. [18]. Of course, in HPLC
experiments the peak shape and maximum depend
mainly on the thermodynamics and kinetics of analyte
distribution between the mobile and stationary phases
and an attempt to fit the peak current to empirical
equations containing the capacity factor and the height
equivalent of a theoretical plate, based on the Snyder
[36] or van Deemter [37] approximations is also given in
[18]. However, a smooth variation of the peak area, Q ,
with flow rate was observed and, more important, the
log Q versus log U plot given in Inset (B) of the same
figure shows good linearity and a slope of �/0.62 again.
This indicates that our arguments of Section 3.1.3.
regarding the peak area flow dependence hold equally
well both for FIA and HPLC experiments. Table 3
presents the slopes of the log Q versus log U curves for a
number of FIA or HPLC experiments with ampero-
Fig. 5. Variation of peak area, Q , with flow rate, U , in the electrochemical FIA experiment of the m -dopa sample of Fig. 4. Inset; log�/log plot of Q
vs. U .
P. Agrafiotou et al. / Electrochimica Acta 48 (2003) 2447�/2462 2457
metric detection, for four different analytes, detected
under conditions of mass transfer control. It can be seen
that, within experimental error, all slopes are close to the
theoretical value of �/0.66.
In order to explore the variation of Q with U in the
case of amperometric detection under mixed control we
have performed experiments for the detection of m -dopa
at different constant potentials so that the electrode
reaction is either at mass transfer or mixed control,
depending on the choice of potential. The Inset in Fig. 7
shows the variation of peak area Q for a similar HPLC
experiment to that of Fig. 6 (at a flow rate of 0.5 ml
min�1) but at different values of electrode potential. In
this curve (also known as hydrodynamic diagram in the
electrochemical HPLC and FIA literature) the mixed
control (ascending part of the curve) and the mass
transfer limiting (constant charge part of the curve)
regions are clearly seen. Fig. 7 itself presents the log Q
versus log U plots for different detection potentials. It is
shown that, as one moves from potentials in the mass
transfer control region (�/0.8 V vs. Ag/AgCl) to
potentials in the mixed control region (�/0.7, �/0.6
and �/0.5 V vs. Ag/AgCl), the linearity of the plots
deteriorates and the slope of the linear approximations
of the curves (its magnitude) increases from 0.6121 to
0.7426, 0.8206 and 0.9034, respectively, i.e. in qualitative
agreement with the prediction of Section 3.1.3 that the
slope for mixed control cases should lie between the
theoretically values of �/0.66 and �/1, predicted for pure
mass transfer or kinetic control, respectively. Fig. 8
presents a similar log Q versus log U plot for the HPLC
amperometric detection of a 20 ml injected sample of 1
mg ml�1 5-htp in a pH 3 phosphate buffer of 0.05 M
ionic strength, on four carbon fibre electrodes at �/1.2 V
Fig. 6. Electrochemical detection reverse phase HPLC peaks recorded at �/0.8 V vs. Ag/AgCl on a GC disc electrode in a thin channel cell, for a 20 ml
injected sample of 5�/10�6 M m -dopa in a 0.05 M phosphate buffer (pH 3) �/10% methanol mobile phase and for flow rates as indicated on the
graph (injection at t�/0). Inset (A); variation of peak current, Imax, with flow rate, U . Inset (B); log�/log plot of peak area, Q , and volumetric flow
rate, U .
Table 3
Values of the slope of log�/log plots of the peak area-charge, Q vs. flow rate U , for ascorbic acid, m -dopa, potassium ferrocyanide and hydroquinone
electrochemical detection FIA and HPLC experiments
Ascorbic acida m -dopab Ferrocyanidec Hydroquinoned
Slope of log Q vs. log U plot 0.669/0.08 0.659/0.05 0.679/0.09 0.639/0.01
a 5 FIA experiments.b 4 FIA and 3 HPLC experiments.c 5 FIA experiments.d 2 FIA and 1 HPLC experiments.
P. Agrafiotou et al. / Electrochimica Acta 48 (2003) 2447�/24622458
Fig. 7. Log�/log plots of chromatographic peak area, Q , and volumetric flow rate, U , from electrochemical detection reverse phase HPLC
experiments, using a 3 mm diameter GC disc electrode in a thin channel cell, for a 20 ml injected sample of 5�/10�6 M m -dopa in a 0.05 M phosphate
buffer (pH 3) �/10% methanol mobile phase and at various electrode potentials vs. Ag/AgCl, as indicated on the graph. Inset; hydrodynamic (charge)
voltammogram for the m -dopa sample during reverse phase HPLC electrochemical detection experiments under conditions described above, at a 0.5
ml min�1 flow rate.
Fig. 8. Log�/log plot of chromatographic peak area, Q , with volumetric flow rate, U , from electrochemical detection reverse phase HPLC
experiments, using four carbon fibre electrodes of 30 mm diameter at �/1.2 V vs. Ag/AgCl in a thin channel cell, for a 20 ml injected sample of 1 mg
ml�1 5-htp in a pH 3 phosphate buffer of 0.05 M ionic strength; the mobile phase was a 90% v/v pH 5 phosphate buffer of 0.05 M ionic strength �/
10% v/v of methanol. Inset; hydrodynamic (charge) voltammogram for the 5-htp sample during reverse phase HPLC electrochemical detection
experiments under conditions described above, at a 1 ml min�1 flow rate.
P. Agrafiotou et al. / Electrochimica Acta 48 (2003) 2447�/2462 2459
versus Ag/AgCl. As it can be seen from the Inset to the
Figure which presents the hydrodynamic voltammo-
gram at 1 ml min�1, this potential is not in the pure
mass transfer control region of the system hence theslope of the linear approximation of the log Q versus
log U plot is �/0.8724.
Finally, Fig. 9 presents a log�/log plot of the HPLC
chromatographic peak area with UV detection (at 254
nm), A, versus the flow rate, U , for the case of 10�3 M
vanillylmandelic acid injected in a mobile phase of a
0.005 M total ionic strength and pH 7. Excellent
linearity is observed and a slope of �/1 is found, incomplete agreement with the theoretical arguments of
Section 3.1.3 and Eq. (28).
4. Conclusions
Steady-state experiments of constant analyte concen-
tration in the flowing solution at various flow rates haveproven that the current at a disc electrode of a thin
channel flow cell is given by an expression similar to that
established for rectangular electrodes. Most important,
the steady state limiting current, Iss, is proportional to
the volumetric flow rate to the power of 1/3, U1/3.
No simple dependence of maximum peak current,
Imax, on flow rate, U , has been found neither in FIA or
HPLC experiments with a common amperometric thinchannel flow cell; instead, Imax versus U curves exhibited
local maxima and, in the case of FIA, none of the main
dispersion equations proposed in the literature could be
fitted to the results over any appreciable flow rate range.
This was attributed to the limited theoretical and
experimental range that these equations have been
derived for or tested in (as opposed to the wide range
of our experiments) and, mainly, to the 908 change in
flow direction of the analyte as it passes from the FIA
tubing-reactor to the detector as well as to the differ-
ences in flow conditions between the reactor and the
detector due to different channel thickness.
On the contrary, in all electrochemical FIA and
HPLC experiments the peak area-charge, Q , varied
smoothly with flow rate, U , being a decreasing function
of the latter. The log Q versus log U curves for analytes
detected under mass transfer control had a slope close to
�/0.66. This value has been derived by simple dispersion
theory and mass balance considerations that hold for
any type of dispersion in FIA (hence irrespective of flow
rate conditions and system geometric characteristics)
and any type of analyte-column-mobile phase interac-
tions in HPLC. Based on similar considerations, the
peak area of FIA and HPLC with amperometric
detection under pure kinetic control as well as with
UV detection have both been found to be proportional
to U�1. In cases of amperometric detection under mixed
kinetic and mass transfer control, Q is a more compli-
cated function of U and the linear approximations of
the log Q versus log U plots have slopes between the �/
0.66 and �/1 values of the mass transfer and kinetic
control limits, respectively.
The use of peak area as the analytical signal instead of
the peak maximum has been known to be advantageous
because the former is independent of the type of
dispersion (hence, of system geometric characteristics
and arrangement) in FIA and of peak asymmetry (hence
of ideal column behaviour) in HPLC [23,24]. Based on
the flow rate dependence of the peak area in electro-
chemical detection FIA and HPLC, we can comment on
Fig. 9. Log�/log plot of peak area, A, vs. flow rate, U , in the UV detection (at 254 nm) reverse phase HPLC experiment of a 20 ml injected sample of
10�3 M vanillylmandelic acid in a mobile phase of a 0.005 M total ionic strength phosphate buffer (pH 7).
P. Agrafiotou et al. / Electrochimica Acta 48 (2003) 2447�/24622460
a number of special features associated with the choice
of the peak area, Q , as the analytical signal, S , in the
case of amperometric detection.
First of all, the establishment of a universal lineardependence of peak area on U�2/3 in the case of pure
mass transfer control or of a smooth descending Q
versus U working curve in the case of mixed control (as
opposed to the maxima of the Imax versus U curves),
permits for comparisons of results obtained in the same
system but at different flow rates as well as limits the
number of calibrations needed.
Second, if one chooses to work at relatively low flowrates (of course, the lowest practical flow rate will be
determined by elution time, peak broadening and peak
overlapping considerations) then the signal Q for a
given analyte concentration will increase resulting in an
increase in sensitivity. This effect will be enhanced in the
case of analytes with very slow electron transfer kinetics
where an increase in flow rate-mass transport conditions
would shift the voltammetric curve to higher over-potentials and hence decrease the current at a given
potential; on the contrary, working at low flow rates
with such systems and choosing Q as the signal, not only
will the signal increase due to the dependence of Q on U
discussed above but also due to the shift of the
voltammetric wave to lower overpotentials.
Finally, particular attention has to be given on the
definition and assessment of the signal-to-noise ratio (S /N ratio) when using the peak area as the signal. The
baseline of the FIA signal or HPLC chromatogram is
due to the steady-state oxidation (or reduction) of
impurities dissolved in the carrier solution/mobile phase
or of the solvent itself. If at the detection potential the
oxidation (or reduction) of the impurities is under pure
kinetic control (as that of the solvent is), then the
baseline current, Ib, will remain unchanged if the flowrate changes and equal to a kinetic current (Ib)k. If,
however, the reaction of impurities is under mixed or
mass transport limited control, then the steady-state
baseline current will change with flow rate. In the case of
pure mass transport control it will increase linearly with
U1/3 as described in Section 3.1.1, i.e.:
Ib�KbU1=3 (31)
whereas in the case of mixed control and in accordance
with (25) it will hold:
Ib�(Ib)kKbU1=3
(Ib)k � KbU1=3(32)
It can easily follow from (31) and (32) that an increase of
flow rate results in an increase of this type of contribu-
tion to the baseline current.Variations and fluctuations of the background cur-
rent result in signal drifts or peaks that give rise to the
noise, N, of the analytical determination. These current
fluctuations and the corresponding noise can be due
either to voltage fluctuations or changes in the electrode
surface (e.g. by changing adsorption or wetting extent)
or flow fluctuations. The first two sources of noise areindependent of flow rate (dIb is constant). The last
source is only operative when impurities oxidised (or
reduced) under mixed or mass transport control are
present, as discussed above. Therefore, in the latter case,
a fluctuation of flow rate by dU will result, according to
(31) in a baseline current fluctuation-noise of:
dIb
dU�1=3KbU�2=3[dIb�1=3KbU�2=3dU (33)
if the impurities react under pure mass transfer control
and, according to (32) in:
dIb
dU�
1=3[(Ib)k]2KbU�2=3
[(Ib)2k � KbU1=3]2
[
dIb�1=3[(Ib)k]2KbU�2=3
[(Ib)2k � KbU1=3]2
dU (34)
if the impurities are transformed under mixed kinetic
and mass transfer control.
The noise N is defined as the maximum value of the
peak current, dIb (above the baseline current Ib), from a
number of observed noise peaks, when the peak current
is used as the signal. However, when the peak area is thesignal, then the corresponding noise can be defined
either as the area under the maximum noise peak or, as
the product of the maximum noise peak current multi-
plied by the maximum duration-width of the recorded
noise peaks, Dtnoise (i.e. as N�/dIbDtnoise), or (in a
modification of the definition proposed in Ref. [29]) as
the product of the maximum noise peak current multi-
plied by the width of the analyte chromatographic peak,Dtsignal (i.e. as N�/dIb Dtsignal). The dependence of the
S /N ratio (which becomes Q /(dIbDtnoise) or Q /(dIbDt-
signal) in this case) on flow rate, can be predicted by all
possible combinations of Eqs. (23), (26), and (27) with
Equations (34), (35) and that of a flow independent dIb,
for the analyte and the impurities being transformed
under either pure mass transfer, or mixed, or kinetic
control. The situation becomes even more complicated ifone takes into account that Dtsignal and, in general,
Dtnoise too, both depend on flow rate. Regarding the
former, the width of the FIA peak is suggested to be
proportional to U�1 (Ref. [28], p. 90) whereas for the
width of the HPLC peak, empirical expressions based on
the dependence of chromatographic plate height on flow
which describe peak broadening have been proposed
[38]. On the other hand, the width of a noise peak is onlyexpected to depend on flow in cases that the main
contribution to baseline current is from impurities
transformed under mixed or mass transfer control.
P. Agrafiotou et al. / Electrochimica Acta 48 (2003) 2447�/2462 2461
A simplification of the situation arises by considering
that the background current is often mainly due to
solvent oxidation (always under kinetic control, hence
associated with a flow independent noise dIb and noisepeak width Dtnoise) and that Q always increases with
decreasing flow rate (Eqs. (23), (26), and (27)). Then, the
S /N ratio (when defined as Q /(dIbDtnoise) is expected to
be higher at low flow rates.
Summarising, it can be seen that the choice of peak
area as the signal in amperometric FIA and HPLC
detection, due to its specific dependence on flow rate,
has a number of implications for electroanalysis inflowing systems.
References
[1] K. Stulik, V. Pacakova, Electroanalytical Measurements in
Flowing Liquids, Ellis Horwood, Chichester, 1987.
[2] H. Parvez, M. Bastart-Malsot, S. Parvez, T. Nagatsu, G.
Carpenter (Eds.), Electrochemical Detection in Medicine and
Chemistry, Progress in HPLC, vol. 2, VNU Science Press,
Ultrecht, 1987.
[3] J. Wang, in: G. Patonay (Ed.), Electrochemical Detection for
Liquid Chromatography, in HPLC Detection*/Newer Methods,
VCH, NY, 1992, pp. 91�/107.
[4] S.M. Lunte, C.E. Lunte, P.T. Kissinger, Electrochemical Detec-
tion in Liquid Chromatography and Capillary Electrophoresis, in
Laboratory Techniques in Electroanalytical Chemistry, second ed
(Chapter 27), Marcel Dekker, New York, 1996, pp. 813�/853.
[5] K. .Slais, Electrochemical detectors for low-dispersion liquid
chromatography, in: H. Parvez, M. Bastart-Malsot, S. Parvez,
T. Nagatsu, G. Carpenter (Eds.), Electrochemical Detection in
Medicine and Chemistry, Progress in HPLC, vol. 2, VNU Science
Press, Ultrecht, 1987, pp. 53�/94.
[6] H.B. Hanekamp, H.J. Van Nieuwkerk, Anal. Chim. Acta 121
(1980) 13.
[7] J. Lankelma, H. Poppe, J. Chromatogr. 125 (1976) 375.
[8] W.L. Caudill, J.O. Howell, R.M. Wightman, Anal. Chem. 54
(1982) 2532.
[9] S.G. Weber, W. Purdy, Anal. Chim. Acta 100 (1978) 531.
[10] S.G. Weber, J. Electroanal. Chem. 145 (1983) 1.
[11] E.M.M. Roosendaal, H. Poppe, Anal. Chim. Acta 158 (1984) 323.
[12] S. Moldoveanu, J.L. Anderson, J. Electroanal. Chem. 175 (1984)
67.
[13] R.G. Compton, P.J. Daly, J. Electroanal. Chem. 179 (1984) 107.
[14] J.L. Anderson, S. Moldoveanu, J. Electroanal. Chem. 179 (1984)
107.
[15] S. Moldoveanu, J.L. Anderson, J. Electroanal. Chem. 185 (1985)
239.
[16] J.A. Alden, R.G. Compton, J. Electroanal. Chem. 404 (1996) 27.
[17] R.J. Tait, P.C. Bury, B.C. Finnin, B.L. Reed, A.M. Bond, J.
Electroanal. Chem. 356 (1996) 25.
[18] S. Prabhu, J.L. Anderson, Anal. Chem. 59 (1987) 157.
[19] P. Meschi, D.C. Johnson, Anal. Chim. Acta 124 (1981) 305.
[20] J. Ruzicka, E.H. Hansen, Anal. Chim. Acta 99 (1978) 37.
[21] J.T. Vanderslice, K.K. Stewart, A.G. Rosenfeld, D. Higgs,
Talanta 28 (1981) 11.
[22] M.A. Gomez-Nieto, M.D.L. De Castro, A. Martin, M. Valcarcel,
Talanta 32 (4) (1985) 319.
[23] R.P.W. Scott, Chromatographic Detectors*/Design, Function
and operation (Chapter 18), Marcel Dekker, New York, 1985, pp.
482�/487.
[24] J.M. Harris, Anal. Chem. 54 (1982) 2337.
[25] P. Meschi, D.C. Johnson, Anal. Chim. Acta 124 (1981) 315.
[26] G. Taylor, Proc. R. Soc. London Ser. A 219 (1953) 186.
[27] G. Taylor, Proc. R. Soc. London Ser. A 219 (1954) 473.
[28] M. Valcarcel, M.D.L. De Castro, in: A. Losada (Transl.), Flow-
Injection Analysis*/Principles and Applications, Ellis Horwood,
Chichester, 1986.
[29] J. Lankelma, H. Poppe, J. Chromatogr. 125 (1976) 375.
[30] W.L. Caudill, J.O. Howell, R.M. Wightman, Anal. Chem. 54
(1982) 2532.
[31] A.K. Convington, R.G. Bates, R.A. Durst, Pure Appl. Chem 57
(3) (1985) 531.
[32] I. Rousar, K. Micka, A. Kimla, Electrochemical Engineering II
(Parts D-F), Chemical Engineering Monographs (Chapter 22),
Elsevier, Amsterdam, 1986, pp. 18�/24.
[33] D.J. Pickett, K.L. Ong, Electrochim. Acta 19 (1974) 875.
[34] D. Pletcher, F.C. Walsh, Industrial Electrochemistry, second ed,
Blackie Academic and Professional, London, 1993, p. 112.
[35] K.K. Stewart, Anal. Chem. 55 (1983) 931A.
[36] L.R. Snyder, J.J. Kirkland, Introduction to Modern Liquid
Chromatography, second ed, Wiley, New York, 1979, p. 235.
[37] J.J. van Deemter, F.J. Zuiderweg, A. Klinkenberg, Chem. Eng.
Sci. 5 (1956) 271.
[38] J.A. Jonsson, in: J.A. Jonsson (Ed.), Chromatographic Theory
and Basic Principles (Chapter 3), Marcel Dekker, New York,
1987.
P. Agrafiotou et al. / Electrochimica Acta 48 (2003) 2447�/24622462