Ⅱ-8. Accelerator - 153 - RIKEN Accel. Prog. Rep. 47 (2014) Reinforcement of magnetic shield for HTc SQUID beam current monitor at the RIBF T. Watanabe, *1 N. Fukunishi, *1 M. Kase, *1 O. Kamigaito, *1 S. Inamori, *2 and K. Kon *2 To measure the DC current of high-energy heavy-ion beams nondestructively at a high resolution, a high- critical-temperature (HTc) superconducting quantum interference device (SQUID) beam current monitor henceforth referred as HTc SQUID monitor has been developed for use in the radioactive isotope beam fac- tory (RIBF) at RIKEN. 1) Beginning this year, the magnetic shielding system has been greatly reinforced. Since the measurement resolution is determined by the signal to noise ratio, this resolution is improved by at- tenuating the external magnetic noise and RF back ground noise. These noises are mainly produced by the distribution and transmission lines from the high- current power supplies and high-power RF cavities of the cyclotrons. To reinforce the existing magnetic shield, we devel- oped a hybrid magnetic shielding method based on the properties of perfect diamagnetic materials and fer- romagnetic materials; we were able to realize a high shielding effect despite the compact system. This sys- tem consists of two shielding parts: one for the HTc current sensor and the other for ferromagnetic shield- ing materials. The HTc current sensor used to pro- duce a shielding current produced by the beam 1) also works as the superconducting shield via the Meiss- ner effect (perfect diamagnetism). The ferromagnetic shielding materials are composed of high permeability alloys (Permalloy, Mu-metal, etc.). The HTc SQUID is installed inside the frame and onto the HTc current sensor, and the frame is covered by the cap. Con- sequently, the HTc SQUID is almost completely sur- rounded by the hybrid magnetic shielding system. A photograph of the completed hybrid magnetic system is shown in Fig. 1. In the acceleration facility, since there exist AC mag- netic noises of 50 Hz and higher order and which are much stronger than terrestrial magnetism, an active magnetic field canceller system (JEOL Ltd.) was de- signed and introduced to the HTc SQUID monitor. This system is comprised of a magnetic field control unit, combined AC/DC magnetic field sensors, and compensation coils. The compensation coils consist of three pairs of coils that are arranged perpendicular to each other. Each of these pairs forms a so-called “Helmholtz-Coil-Pair,” able to produce a homogenous magnetic field in between the pairs; each pair controls one direction (along x-, y-, or z- axis). A photograph of Condensed from the proceedings in European Conference on Applied Superconductivity (EUCAS 2013) *1 RIKEN Nishina Center *2 TEP Corporation Fig. 1. Photograph of the completed hybrid magnetic sys- tem. 1: HTc current sensor with ferromagnetic shield- ing materials, 2: frame, 3: cap, and 4: band. Fig. 2. HTc SQUID monitor with active magnetic field can- celler system. the active magnetic canceller system is shown in Fig. 2. To evaluate the performance of the hybrid magnetic shielding system and the active magnetic canceller sys- tem, the output signals of the HTc SQUID were ana- lyzed in the time and frequency domains. The signal was measured in the room next to where the power supplies for RIBF were located, where the leakage mag- netic field of the 50 Hz component was measured by a Gauss meter as 4.5 ×10 -4 T. On the other hand, the output signal of the 50 Hz component of the HTc SQUID monitor was 6 ×10 -14 T. Based on these find- ings, we consider that the combination of the hybrid magnetic shielding system and the active magnetic canceller system can attenuate the external magnetic noise to 10 -10 . References 1) T. Watanabe et al., Proc. 2010 Beam Instrumentation Workshop (BIW10), Santa Fe, U.S.A (2010) p. 523. 1 HTc Current Sensor 2 Flame (inside SQUID) 3 Cap 4 Band Bundling Band HTc SQUID Monitor Helmholtz-Coil-Cage Mgnetic Field Sensors

Reinforcement of magnetic shield for HTc SQUID beam ... · The HTc current sensor used to pro-duce a shielding current produced by the beam1) also works as the superconducting shield

  • Upload
    others

  • View
    3

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Reinforcement of magnetic shield for HTc SQUID beam ... · The HTc current sensor used to pro-duce a shielding current produced by the beam1) also works as the superconducting shield

Ⅱ-8. Accelerator

- 153 -

RIKEN Accel. Prog. Rep. 47 (2014)

Reinforcement of magnetic shield for HTc SQUID beam currentmonitor at the RIBF†

T. Watanabe,∗1 N. Fukunishi,∗1 M. Kase,∗1 O. Kamigaito,∗1 S. Inamori,∗2 and K. Kon∗2

To measure the DC current of high-energy heavy-ionbeams nondestructively at a high resolution, a high-critical-temperature (HTc) superconducting quantuminterference device (SQUID) beam current monitorhenceforth referred as HTc SQUID monitor has beendeveloped for use in the radioactive isotope beam fac-tory (RIBF) at RIKEN.1) Beginning this year, themagnetic shielding system has been greatly reinforced.Since the measurement resolution is determined by thesignal to noise ratio, this resolution is improved by at-tenuating the external magnetic noise and RF background noise. These noises are mainly produced bythe distribution and transmission lines from the high-current power supplies and high-power RF cavities ofthe cyclotrons.

To reinforce the existing magnetic shield, we devel-oped a hybrid magnetic shielding method based on theproperties of perfect diamagnetic materials and fer-romagnetic materials; we were able to realize a highshielding effect despite the compact system. This sys-tem consists of two shielding parts: one for the HTccurrent sensor and the other for ferromagnetic shield-ing materials. The HTc current sensor used to pro-duce a shielding current produced by the beam1) alsoworks as the superconducting shield via the Meiss-ner effect (perfect diamagnetism). The ferromagneticshielding materials are composed of high permeabilityalloys (Permalloy, Mu-metal, etc.). The HTc SQUIDis installed inside the frame and onto the HTc currentsensor, and the frame is covered by the cap. Con-sequently, the HTc SQUID is almost completely sur-rounded by the hybrid magnetic shielding system. Aphotograph of the completed hybrid magnetic systemis shown in Fig. 1.

In the acceleration facility, since there exist AC mag-netic noises of 50 Hz and higher order and which aremuch stronger than terrestrial magnetism, an activemagnetic field canceller system (JEOL Ltd.) was de-signed and introduced to the HTc SQUID monitor.This system is comprised of a magnetic field controlunit, combined AC/DC magnetic field sensors, andcompensation coils. The compensation coils consistof three pairs of coils that are arranged perpendicularto each other. Each of these pairs forms a so-called“Helmholtz-Coil-Pair,” able to produce a homogenousmagnetic field in between the pairs; each pair controlsone direction (along x-, y-, or z- axis). A photograph of

† Condensed from the proceedings in European Conference onApplied Superconductivity (EUCAS 2013)

∗1 RIKEN Nishina Center∗2 TEP Corporation

Fig. 1. Photograph of the completed hybrid magnetic sys-

tem. 1: HTc current sensor with ferromagnetic shield-

ing materials, 2: frame, 3: cap, and 4: band.

Fig. 2. HTc SQUID monitor with active magnetic field can-

celler system.

the active magnetic canceller system is shown in Fig. 2.To evaluate the performance of the hybrid magnetic

shielding system and the active magnetic canceller sys-tem, the output signals of the HTc SQUID were ana-lyzed in the time and frequency domains. The signalwas measured in the room next to where the powersupplies for RIBF were located, where the leakage mag-netic field of the 50 Hz component was measured bya Gauss meter as 4.5 ×10−4 T. On the other hand,the output signal of the 50 Hz component of the HTcSQUID monitor was 6 ×10−14 T. Based on these find-ings, we consider that the combination of the hybridmagnetic shielding system and the active magneticcanceller system can attenuate the external magneticnoise to 10−10.

References1) T. Watanabe et al., Proc. 2010 Beam Instrumentation

Workshop (BIW10), Santa Fe, U.S.A (2010) p. 523.

1 HTc Current Sensor

2 Flame(inside SQUID)3 Cap

4 Band

Bundling Band

HTc SQUIDMonitor

Helmholtz-Coil-Cage

Mgnetic FieldSensors