Heriot-Watt University Research Gateway Heriot-Watt University A compressed-sensing approach for ultrasound imaging Besson, Adrien; Carrillo, Rafael E.; Perdios, Dimitris; Arditi, Marcel; Wiaux, Yves; Thiran, Jean-Philippe Published in: Proceedings of SPARS 2017 Publication date: 2017 Document Version Peer reviewed version Link to publication in Heriot-Watt University Research Gateway Citation for published version (APA): Besson, A., Carrillo, R. E., Perdios, D., Arditi, M., Wiaux, Y., & Thiran, J-P. (2017). A compressed-sensing approach for ultrasound imaging. In Proceedings of SPARS 2017 (pp. 1-2) General rights Copyright and moral rights for the publications made accessible in the public portal are retained by the authors and/or other copyright owners and it is a condition of accessing publications that users recognise and abide by the legal requirements associated with these rights. If you believe that this document breaches copyright please contact us providing details, and we will remove access to the work immediately and investigate your claim.

A compressed-sensing approach for ultrasound imaging · PDF fileA compressed-sensing approach for ultrasound imaging Adrien Besson, Rafael E. Carrilloy, Dimitris Perdios, Marcel Arditi,

  • Upload
    dangdat

  • View
    216

  • Download
    1

Embed Size (px)

Citation preview

Page 1: A compressed-sensing approach for ultrasound imaging · PDF fileA compressed-sensing approach for ultrasound imaging Adrien Besson, Rafael E. Carrilloy, Dimitris Perdios, Marcel Arditi,

Heriot-Watt University Research Gateway

Heriot-Watt University

A compressed-sensing approach for ultrasound imagingBesson, Adrien; Carrillo, Rafael E.; Perdios, Dimitris; Arditi, Marcel; Wiaux, Yves; Thiran,Jean-PhilippePublished in:Proceedings of SPARS 2017

Publication date:2017

Document VersionPeer reviewed version

Link to publication in Heriot-Watt University Research Gateway

Citation for published version (APA):Besson, A., Carrillo, R. E., Perdios, D., Arditi, M., Wiaux, Y., & Thiran, J-P. (2017). A compressed-sensingapproach for ultrasound imaging. In Proceedings of SPARS 2017 (pp. 1-2)

General rightsCopyright and moral rights for the publications made accessible in the public portal are retained by the authors and/or other copyright ownersand it is a condition of accessing publications that users recognise and abide by the legal requirements associated with these rights.

If you believe that this document breaches copyright please contact us providing details, and we will remove access to the work immediatelyand investigate your claim.

Page 2: A compressed-sensing approach for ultrasound imaging · PDF fileA compressed-sensing approach for ultrasound imaging Adrien Besson, Rafael E. Carrilloy, Dimitris Perdios, Marcel Arditi,

Download date: 21. Apr. 2018

Page 3: A compressed-sensing approach for ultrasound imaging · PDF fileA compressed-sensing approach for ultrasound imaging Adrien Besson, Rafael E. Carrilloy, Dimitris Perdios, Marcel Arditi,

A compressed-sensing approach for ultrasound imagingAdrien Besson∗, Rafael E. Carrillo†, Dimitris Perdios∗, Marcel Arditi∗, Yves Wiaux‡, and Jean-Philippe Thiran∗§

∗Signal Processing Laboratory (LTS5), Ecole Polytechnique Federale de Lausanne, Lausanne, Switzerland†Centre Suisse d’Electronique et de Microtechnique (CSEM), Neuchatel, Switzerland

‡Institute of Sensors, Signals and Systems, Heriot-Watt University, Edinburgh, United-Kingdom§Department of Radiology, University Hospital Center (CHUV) and University of Lausanne (UNIL), Lausanne, Switzerland

Abstract—Ultrasonography uses multiple piezo-electric element probesto image tissues. Current time-domain beamforming techniques requirethe signal at each transducer-element to be sampled at a rate higherthan the Nyquist criterion, resulting in an extensive amount of data tobe received, stored and processed. In this work, we propose to exploitsparsity of the signal received at each transducer-element. The proposedapproach uses multiple compressive multiplexers for signal encodingand solves an `1-minimization in the decoding step, resulting in thereduction of 75 % of the amount of data, the number of cables and thenumber of analog-to-digital converters required to perform high qualityreconstruction.

Medical ultrasonography is a widely used modality nowadaysdue to its non-invasiveness and real-time capability. In many ul-trasound (US) systems, an array of transducer-elements is used totransmit acoustic pulses, which, when reflected back by the mediuminhomogeneities, are sensed by the same array. According to theShannon-Nyquist theorem, the sampling rate at each element must beat least twice the bandwidth of the received signal. In practice, time-domain beamforming techniques require sampling rates between 3and 10 times the center frequency to minimize the delay-quantizationerrors [1]. The large number of transducer-elements and the highcentral frequency required in medical ultrasonography motivate theresearch towards sampling rate reduction.

Let us consider an US probe made of Nel transducer-elements andcall ri (t) with i ∈ {1, ..., Nel} the corresponding echo signals re-ceived by each transducer-element at time t. If we consider a mediummade of K inhomogeneities, then ri (t) =

∑Kk=1 aikψ (t− tk),

with (aik, tk) amplitudes and times-of-arrival of the K echo-pulses to the ith transducer-element and ψ (t) the elementarywaveform. Assuming a linear propagation, we can state thatψ (t) = (e ∗ hTx ∗ hRx) (t) where e (t) denotes the excita-tion and hTx (t) and hRx (t) are the transmit and receive im-pulse responses of the transducer-elements, respectively. Giventhe model described above, it can be stated that the vectorri = [ri (t1) , ..., ri (tNt)]

T ∈ RNt , where Nt denotes thenumber of time samples, obeys a K-sparse synthesis model in anovercomplete dictionary Ψ ∈ RNt×Nt made of all the shifted replicasof the pulse [2]. Thus, ri = Ψai with ||ai||0 = K, which canbe exploited under the compressed sensing (CS) framework. Manyefforts have been made in order to provide sub-Nyquist acquisitionsystems which result in different CS architectures such as the RandomDemodulator [3] and the Random-Modulator Pre-Integrator [4] forsingle-channel signals, and the compressive multiplexer for multi-channel signals [5], [6]. In US imaging, Chernyakova et al. haverecently proposed a hardware architecture based on finite rate ofinnovation and Xampling ideas [7].

In this work, we propose a proof of concept for a compressivemultiplexer (CMUX) applied to US imaging. The architecture, de-noted as US-CMUX is derived from the work of Kim et al. [6]. Theidea is to split the Nel channels of the US probe into L groups of Mchannels and to mix each group according to the CMUX framework.

In the decoding step, a convex problem is solved.The CMUX, described on Figure 1, is based on modulating each

channel ri (t) of a given group by a chipping sequence pi (t) sampledfrom a Rademacher distribution. The modulated channels are thensummed to y (t) =

∑Mi=1 pi (t) ri (t) and sampled at fs.

The US-CMUX, described on Figure 2, uses L CMUX sharing thesame chipping sequences to perform the signal encoding, giving riseto the matrix Y = [y1, ...,yL] ∈ RNt×L. In the decoding step, thefollowing convex problem is solved:

minA∈RMNt×L

||A||11 subject to ‖Y −ΨPA‖F ≤ ε, (1)

where ||.||11 accounts for the `11-norm, ||.||F is the Frobe-nius norm, ΨP = [Ψp1, ...,ΨpM] ∈ RNt×MNt in whichΨpi = [pi ⊗Ψ1, ...,pi ⊗ΨNt ] ∈ RNt×Nt , where ⊗ denotes theHadamard product, and

A =

a1 aM+1 · · · aNel−M+1

......

...aM a2M · · · aNel

∈ RMNt×L,

with ai ∈ RNt the representation coefficients of ri.Problem (1) is solved using the primal-dual forward backward

algorithm [8] and each channel ri is recovered from A as: ri = Ψai.In order to validate the proposed method, we present numerical

results on an in-vitro hyperechoic inclusion phantom (Model 54GS,Computerized Imaging Reference Systems Inc., Norfolk, USA) andan in-vivo carotid acquired with a Verasonics research scanner (V1-128, Verasonics Inc., Redmond, WA). The US probe used for thedifferent experiments is a L12-5 50 mm probe, with 128 activetransducer-elements, working at 5 MHz with 100% bandwidth. Thesampling frequency is 31.2 MHz, corresponding to 4 times thebandwidth. The architecture is simulated on MATLAB thus theexperiments are carried out on a digital setting. The hardwareimplementation of the proposed scheme will be investigated in futurework. The architecture is tested for L = 2, 4 which means areduction of 50 % and 75 % of the sampling rate, respectively. Inthe decoding process, ε is set to 10−6 ||Y||F and 1500 iterations ofthe algorithm are run. Radio-frequency images are computed with aclassical delay-and-sum (DAS) algorithm, with a linear interpolationfor the delay calculation and without apodization, and the B-modeimage is obtained by Hilbert demodulation, normalization and log-compression with a dynamic range of 40 dB.

The performance of the proposed method is quantified by thesignal-to-noise ratio (SNR) and the structural similarity index (SSIM)against the computed image using 100 % of the data, calculated onthe B-mode image without log-compression.

The results, displayed on Table I, as well as a visual evaluation onFigure 3 and Figure 4 show that the proposed architecture leads tohigh quality reconstruction with 25 % of data only since anechoic,hyperechoic and speckle regions are preserved.

Page 4: A compressed-sensing approach for ultrasound imaging · PDF fileA compressed-sensing approach for ultrasound imaging Adrien Besson, Rafael E. Carrilloy, Dimitris Perdios, Marcel Arditi,

𝑟1(𝑡)

𝑟2(𝑡)

𝑟𝑀(𝑡)

𝑝1(𝑡)

𝑝2(𝑡)

𝑝𝑀(𝑡)

𝑇𝑟𝑎𝑛𝑠𝑑𝑢𝑐𝑒𝑟 𝑒𝑙𝑒𝑚𝑒𝑛𝑡𝑠

𝑦(𝑡)ADC

𝑓𝑠

𝑦[𝑛]

.

.

.

Figure 1 Compressive multiplexer (CMUX) architecture for M

transducer-elements.

𝑟1(𝑡)

𝑟𝑀+1(𝑡)

.

.

.

𝑟𝑖(𝑡)

𝑟𝑖+𝑀(𝑡)

.

.

.

𝐶𝑀𝑈𝑋

𝐶𝑀𝑈𝑋

𝑌 = 𝑦1, … , 𝑦𝐿

𝑦1 𝑛

𝑦𝐿 𝑛

𝐺𝑟𝑜𝑢𝑝 1

𝐺𝑟𝑜𝑢𝑝 𝐿

𝑈𝑙𝑡𝑟𝑎𝑠𝑜𝑢𝑛𝑑 𝑝𝑟𝑜𝑏𝑒

.

.

.

Figure 2 Ultrasound compressive multiplexer architecture using L

CMUX.

Hyperechoic inclusion In-vivo carotidSNR - L = 2 39 36SSIM - L = 2 0.94 0.87SNR - L = 4 32 29SSIM - L = 4 0.81 0.72

Table I Average values of the SNR (dB) and SSIM over 10 draws forthe different images, for L = 2 and L = 4.

REFERENCES

[1] T. L. Szabo, Diagnostic Ultrasound Imaging: Inside Out, second edition.Elsevier, 2014.

[2] F. M. Naini, R. Gribonval, L. Jacques, and P. Vandergheynst, “Compres-sive sampling of pulse trains: Spread the spectrum!” in 2009 IEEE Int.Conf. Acoust. Speech Signal Process., 2009, pp. 2877–2880.

[3] J. A. Tropp, J. N. Laska, M. F. Duarte, J. K. Romberg, and R. G. Baraniuk,“Beyond Nyquist: Efficient sampling of sparse bandlimited signals,” IEEETrans. Inf. Theory, vol. 56, no. 1, pp. 520–544, 2010.

[4] S. R. Becker, “Practical compressed sensing: modern data acquisition andsignal processing,” Ph.D. dissertation, California Institute of Technology,2011.

[5] J. P. Slavinsky, J. N. Laska, M. A. Davenport, and R. G. Baraniuk, “Thecompressive multiplexer for multi-channel compressive sensing,” in 2011IEEE Int. Conf. Acoust. Speech Signal Process., 2011, pp. 3980–3983.

-10 0 10

Lateral position [mm]

5

10

15

20

25

30

35

40

45

De

pth

[m

m]

-40 dB

-30 dB

-20 dB

-10 dB

0 dB

(a) Reference

-10 0 10

Lateral position [mm]

5

10

15

20

25

30

35

40

45

Dep

th [

mm

]

-40 dB

-30 dB

-20 dB

-10 dB

0 dB

(b) CS reconstruction

Figure 3 B-mode image of the hyperechoic inclusion reconstructed with(a) 100 % of the data and (b) 25 % of the data acquired with the US-CMUX architecture.

-10 0 10

Lateral position [mm]

5

10

15

20

25

30

35

40

De

pth

[m

m]

-40 dB

-30 dB

-20 dB

-10 dB

0 dB

(a) Reference

-10 0 10

Lateral position [mm]

5

10

15

20

25

30

35

40

De

pth

[m

m]

-40 dB

-30 dB

-20 dB

-10 dB

0 dB

(b) CS reconstruction

Figure 4 B-mode image of the carotid reconstructed with (a) 100 % of thedata and (b) 25 % of the data acquired with the US-CMUX architecture.