Topographic Changes Indu ced During Flap Producti on by Epi-LASIK surgery Department of Ophthalmology, University of Ulsan Colle ge of Medicine, Asan Medical Center, Seoul, Korea. Kyung Hoon Kim, Jooeun Lee, Chuljin Shin, Hee Kyung Lee, Sung Yong Kang, Jae Yong Kim, Hungwon T chah Authors have no financial interest in this study.

Topographic Changes Induced During Flap Production by Epi-LASIK surgery

  • Upload
    aletha

  • View
    82

  • Download
    4

Embed Size (px)

DESCRIPTION

Topographic Changes Induced During Flap Production by Epi-LASIK surgery. Department of Ophthalmology, University of Ulsan College of Medicine, Asan Medical Center, Seoul, Korea. Kyung Hoon Kim, Jooeun Lee, Chuljin Shin, Hee Kyung Lee, Sung Yong Kang, Jae Yong Kim, Hungwon Tchah. - PowerPoint PPT Presentation

Citation preview

Page 1: Topographic Changes Induced During Flap Production by  Epi-LASIK surgery

Topographic Changes Induced During Flap Production by

Epi-LASIK surgery

Department of Ophthalmology, University of Ulsan College of Medicine,Asan Medical Center, Seoul, Korea.

Kyung Hoon Kim, Jooeun Lee, Chuljin Shin, Hee Kyung Lee, Sung Yong Kang, Jae Yong Kim, Hungwon Tchah

Authors have no financial interest in this study.

Page 2: Topographic Changes Induced During Flap Production by  Epi-LASIK surgery

Introduction• Changes in corneal topography and ocular aberration are highly evalu

ated as a measurement of visual quality after refractive surgery.

• Currently, customized surface ablation techniques are more frequently used to improve patients quality of vision.

• However, these refractive procedures are conducted under the assumption that removed corneal epithelium does not significantly affect outcomes of refractive surface ablation and that they will be regenerated into their original distribution.

• Thus, the purpose of this study was to investigate induced changes of corneal topographic data before and after epithelial flap removal during Epi-LASIK.

Page 3: Topographic Changes Induced During Flap Production by  Epi-LASIK surgery

Patients and Methods• Patients

– Eighteen eyes of ten subjects undergoing Epi-LASIK for correction of myopia or myopic astigmatism were enrolled.

• Surgical Technique– Microkeratome (Amadeus II®, AMO, Irvine, CA) for epithelial flap production. – Flap diameter : 9.0mm– Flap thickness : Approximately 55~60µm

• Topography – Corneal topography with ray tracing aberrometry, i-Trace® (Technologies, Housto

n, Texas), was used to measure corneal topography before and after epithelium removal.

– Corneal aberrations are calculated using the program integrated in the i-Trace system based on the topographic data.

Fig.1 i-Trace®

Fig.2 Amadeus microkeratome®

Page 4: Topographic Changes Induced During Flap Production by  Epi-LASIK surgery

Results (1)• Simulated K showed statistically meaningful change between the ep

ithelium on and off state. – Tendency of Increasing refractive power after epithelium removal.

• However, other topographic parameters (central power, astigmatism on simulated K, asphericity and best fit sphere) did not significantly change.

Fig.3 Topography with epithelium Fig.4 Topography without epithelium

Page 5: Topographic Changes Induced During Flap Production by  Epi-LASIK surgery

Results (2)• Total Corneal higher order aberrations (HOAs) significantly increase

d at 3.0, 5.0, and 7.0mm diameters after epithelial removal.

3mm 5mm 7mm

WithEpithelium

WithoutEpithelium

Total Corneal Higher Order Aberrations

Page 6: Topographic Changes Induced During Flap Production by  Epi-LASIK surgery

i-Trace Data WithEpithelium

Without Epithelium

P value (Wilcoxon Test)

Simulated K (D) <0.001*

Mean±SD 42.52±1.67 43.21±1.75

Range 39.59~45.61 39.95~46.57

Astigmatism on simulated K (D) 1.000

Mean±SD 1.47±0.52 1.56±0.72

Range 0.59~2.59 0.61~3.12

Best-Fit Sphere, central 3.0mm(mm) 0.913

Mean±SD 7.95±0.32 7.93±0.33

Range 7.43 ~ 8.52 7.22~8.63

Central Power (D) 0.913

Mean±SD 42.53±1.69 42.63±1.76

Range 39.60~45.44 39.09~46.73

Asphericity (Q value) 0.248

Mean±SD -0.10±0.14 0.08±0.84

Range -0.33~0.15 -1.82~1.52

Total Corneal Higher Order Aberration, 3.0mm(µm) <0.001*

Mean±SD 0.057±0.034 0.170±0.066

Range 0.021~0.161 0.095~0.342

Total Corneal Higher Order Aberration, 5.0mm(µm) <0.001*

Mean±SD 0.246±0.097 0.639±0.265

Range 0.122~0.482 0.357~1.352

Total Corneal Higher Order Aberration, 7.0mm(µm) <0.001*

Mean±SD 0.669±0.190 1.563±0.763

Range 0.437~1.139 0.626~3.411 *: p<0.05

Page 7: Topographic Changes Induced During Flap Production by  Epi-LASIK surgery

Discussion (1)• According to Damien et al, epithelial distribution on the corneal surfa

ce before and after its removal during PRK is closely related to astigmatism and changes in corneal irregularity. They reported a trend for increased anterior prolateness after epithelial removal, which agrees with results of previous reports.

• In this study, however, neither astigmatism nor anterior prolateness were significantly altered. Differing methods of epithelial removal may underlie such a result.

• Increases of simulated K may be due to thicker epithelium in the periphery than in the center.

Page 8: Topographic Changes Induced During Flap Production by  Epi-LASIK surgery

Discussion (2)• Increases of total HOAs after the removal of epithelium may indicate its

role to smoothe the corneal surface over the Bowman’s layer and to decrease the corneal total HOAs.

• However, current customized surface ablation is conducted in the epithelium off state whereas pre-operative wavefront data are measured from epithelium on normal cornea. This assumption that the epithelium does not affect significantly on refractive surface ablation may be erroneous.

• For this reason, customized techniques in surface ablation may cause unpredictable results. Therefore, more concern should be given to changes of HOAs attributable to the corneal epithelium, and further study is needed to analyze causes in increase of HOAs.

Page 9: Topographic Changes Induced During Flap Production by  Epi-LASIK surgery

Conclusions• Astigmatism of simulated K in corneal topography showed a tendency to inc

rease after removing the epithelial flap during Epi-LASIK.

• Total corneal HOAs at 3.0, 5.0, and 7.0mm increased after epithelial removal.

• This suggests that the corneal epithelium could have a significant effect on corneal topography and the results of corneal surface ablation.

• Future surface ablation refractive surgery may require more concern and investigation into the role of the corneal epithelium.

Page 10: Topographic Changes Induced During Flap Production by  Epi-LASIK surgery

References1. Damien G, Louis R, Thanh H-X. Contribution of the corneal epithelium to anterior corne

al topography in patients having myopic photorefractive keratectomy. J Cataract Refract Surg 2007; 33:1860-1865

2. Serrao S, Lombardo M. Corneal epithelial healing after photorefractive keratectomy: analytical study. J Cataract Refract. Surg 2005; 31: 930-937

3. Simon G, Ren Q, Kervick GN, Parel J-M. Optics of the corneal epithelium. Refract Corneal Surg 1993; 9:42-50

4. Patel S, Marshall J, Fitzke FW III. Refractive index of the human corneal epithelium and stroma. J Refract Surg 1995; 11:100-105

5. Corneal Topography in the Wavefront EraWang M, Swartz T. Corneal Topography in the Wavefront Era; a Guide for Clinical Applic

ation. Thorofare, SLACK Incorporated, 2006