PI-25197-00-DTTS (CD3CD4CD45)

Embed Size (px)

Citation preview

  • 8/14/2019 PI-25197-00-DTTS (CD3CD4CD45)

    1/35

    1 Test per tube

    Product Catalog No: 25197-00

    Dry-Tri T-STAT

    CD3/CD4/CD45Reagent

    Manufactured by

  • 8/14/2019 PI-25197-00-DTTS (CD3CD4CD45)

    2/35

    2

  • 8/14/2019 PI-25197-00-DTTS (CD3CD4CD45)

    3/35

    3

  • 8/14/2019 PI-25197-00-DTTS (CD3CD4CD45)

    4/35

    Dry-Tri T-STAT CD3/CD4/CD45

    Reagent1. INTENDED USE

    The Dry-Tri T-STAT CD3/CD4/CD45 reagent is a three color

    immunofluorescence stain for the labeling, identification, and enumeration of

    mature human total T-lymphocytes (CD3+), and helper/inducer T-

    lymphocytes (CD3+CD4+) combined with a precise number of fluorescentcounting beads for absolute CD4+ and CD3+ T-Cell counts and

    percentages. This reagent is intended to be used in a lyse-no wash

    protocol for flow cytometry analysis of erythrocyte-lysed human whole blood.

    2. SUMMARY AND EXPLANATION

    The Dry-Tri T-STAT CD3/CD4/CD45 reagent contains fluorescently labeled

    antibodies that bind to CD45, CD3, and CD4 antigens found on the surface

    4

  • 8/14/2019 PI-25197-00-DTTS (CD3CD4CD45)

    5/35

    of circulating leukocytes. The CD3 antigen is a complex of at least six

    proteins known collectively as the T-cell receptor (TCR) complex. The

    antibody used in this reagent binds to the 20kDa chain of this complex

    (1).

    The CD4 antigen is a 59kDa protein. It interacts with class II molecules of

    the major histocompatibility complex and is the primary receptor for the

    Human Immunodeficiency Virus (HIV) (2, 3).

    The CD45 antigen is a complex 180-220kDa transmembrane glycoprotein

    which is expressed in high levels on leukocyte cell surfaces and its presencedistinguishes leukocytes from non-haematopoietic cells (1).

    Clinical relevance

    Identification of abnormal levels of CD4+ T-cells and corresponding % CD4

    have been associated with certain types of immunodeficiency disease. For

    example infection with human immunodeficiency virus (HIV), results in

    profound immunosuppression due predominantly to a selective depletion of

    the CD4+ lymphocytes that express the receptor for the virus. Progressive

    5

  • 8/14/2019 PI-25197-00-DTTS (CD3CD4CD45)

    6/35

    clinical and immunologic deterioration generally correlates with a decreasing

    CD4+ lymphocyte count. (5-8).

    3. PRINCIPLE OF THE PROCEDURE

    The Dry Tri T-STAT CD3/CD4/CD45 reagent consists of murine monoclonal

    antibodies that specifically recognize the human leukocyte surface antigens,

    CD3, CD4, and CD45. Each of the monoclonal antibodies is labeled with a

    unique fluorochrome. Specific cell subsets are stained when blood is

    combined with the reagent and each monoclonal antibody binds to the binds

    to the cell determinant molecules on the cell surface. Specific cell subsetsare identified when passed through a flow cytometer laser beam.

    The Dry Tri T-STAT CD3/CD4/CD45 reagent also contains a precise number

    of fluorescent beads. When the reagent is combined with a known volume

    of blood the reagent provides for the single platform determination of the

    absolute cell concentrations of the stained subsets. The volume of sampleanalyzed can be determined by multiplication of the total sample volume by

    the fraction of total beads that were detected during the analysis.

    6

  • 8/14/2019 PI-25197-00-DTTS (CD3CD4CD45)

    7/35

    4. REAGENT

    The Dry-Tri T-STAT CD3/CD4/CD45 reagent is formulated in buffered saline

    with sodium azide and stabilizers. It contains Atto488 labeled CD4

    monoclonal antibody, clone RPA-T4; Phycoerythrin (PE) labeled CD45monoclonal antibody, clone HI30; and PEDyomics649 labeled CD3

    monoclonal antibody, clone UCHT1. The monoclonal antibodies used in the

    Dry-Tri T-STAT CD3/CD4/CD45 were assigned these specificities at the 8th

    International Workshop on Human Leukocyte Differentiation Antigens (9). A

    precise number of fluorescent counting beads are included in the Dry-Tri T-

    STAT CD3/CD4/CD45 reagent to allow single-platform determination ofabsolute CD4+ and CD3+ T-cell counts. The Dry-Tri T-STAT

    CD3/CD4/CD45 reagent is provided in dried form and dispensed in flow

    cytometer compatible sample tubes with each tube containing one ready-to-

    use test. Fifty (50) tests are supplied in each Dry Tri T-STAT

    CD3/CD4/CD45 package.

    Precautions

    1. Warning: The Dry-Tri T-STAT CD3/CD4/CD45 reagent

    7

  • 8/14/2019 PI-25197-00-DTTS (CD3CD4CD45)

    8/35

    contains Sodium azide. Sodium azide is harmful if swallowed.

    Wear suitable protective clothing. If swallowed, seek medical

    advice immediately. Contact with acids liberates toxic gas.

    Azides should be flushed with large amounts of water during

    disposal to avoid deposits in lead or copper plumbing.2. Warning: All blood specimens are considered biohazards.

    Handle them as if they are capable of transmitting infection

    and dispose off with proper precautions and accordance with

    governmental regulations.

    3. The addition of the precise volume of blood is critical to obtain

    correct results. Use a calibrated pipette and operate accordingto the manufacturers instructions.

    Storage and Handling

    1. Store the reagent at room temperature in a dry place. Do not

    use the reagent after the expiry date on the label.2. Do not freeze or refrigerate Dry-Tri T-STAT CD3/CD4/CD45

    reagent.

    3. The Dry-Tri T-STAT CD3/CD4/CD45 reagent is light sensitive.

    8

  • 8/14/2019 PI-25197-00-DTTS (CD3CD4CD45)

    9/35

    Do not expose to direct light either during storage or when

    mixed with blood.

    4. Use the ziplock to reseal the reagent pouch after you have

    removed the desired number of tests.

    Indication of Instability

    The Dry Tri T-STAT CD3/CD4/CD45 reagent is a dry glassy coating at the

    bottom of the reaction tube. Do not use if the reagents appears to be moist

    or it has become dislodged from the bottom of the reaction tube.

    5. INSTRUMENT

    The Dry-Tri T-STAT CD3/CD4/CD45 reagent is designed to be used on flow

    cytometers equipped with a 488nm laser, that are capable of detecting

    forward and side scatter, and detecting three color fluorescence in threeranges: 515-545nm, 562-607nm, and >650nm. Examples of flow cytometers

    with the above features are the BD FACScan, BD FACSCalibur, Coulter

    EPICS-XL and Accuri flow cytometer systems. Calibrate the instrument

    9

  • 8/14/2019 PI-25197-00-DTTS (CD3CD4CD45)

    10/35

    for setting photomultiplier tube voltages, fluorescence compensation, and

    checking instrument sensitivity according to the manufacturers guidelines.

    Refer to the manufacturers instructions for setting up three color

    immunophenotyping, principles of operation, operating instructions, andoperating precautions, limitations, and hazards.

    6. SPECIMEN COLLECTION

    The blood sample should be collected in a sterile blood collection tube

    containing K3EDTA. Follow the collection tube manufacturers guidelines forthe minimum volume of blood to be collected.

    The anti-coagulated blood must be stored at room temperature (20C -

    25C) and should be stained and analyzed ideally within 24 hours of draw.

    Interfering conditionsRefrigerated, hemolyzed, and previously fixed blood specimens can yield

    erroneous results and should be rejected. Aged or partially clotted blood

    specimens can affect flow cytometer performance.

    10

  • 8/14/2019 PI-25197-00-DTTS (CD3CD4CD45)

    11/35

    7. PROCEDURE

    Reagent Provided: Dry-Tri T-STAT CD3/CD4/CD45 reagent in a dried

    format.

    Reagents and Materials required but not provided

    1. Blood collection tube containing K3EDTA

    2. Calibrated pipettes

    3. Vortex mixer

    4. Lysing SolutionA or similar blood fix/lyse solution

    5. Sheath fluidB

    6. 0.2m filtered, deionized distilled water or Milli-Q waterC

    7. Flow cytometer Calibration BeadsD

    A. ReaLyse Lysing solution: ReaMetrix Catalog No: 25237-00

    B. Sheath fluid: BD Catalog No.342003/ BC Catalog No. 8546859 or equivalent

    C. Accuri Flow cytometer uses Milli-Q water/Distilled deionized water instead of

    11

  • 8/14/2019 PI-25197-00-DTTS (CD3CD4CD45)

    12/35

    Sheath fluid

    D. Calibration Beads: BD Catalog No. 340486/ BC Catalog No. 6605359 or

    equivalent

    Assay Protocol

    1. Start flow cytometer according to manufacturers instructions.(In the case of FACScan or FACSCalibur, the instrument should

    be setup and calibrated using CaliBRITE beads and FACSComp

    software; For Beckman Coulter EPICS instruments, the instrument

    should be set up using the Flow-Check fluorospheres provided by

    Beckman Coulter. For Accuri flow cytometer instruments, theperformance of the instrument should be checked by Validation

    beads provided by Accuri.)

    3. Mix blood sample (invert blood tube at least 10 times) and

    reverse pipette 50L of blood into the correctly labeled tube

    that contains the dry down reagent.

    3. Gently vortex each tube for 30 seconds. Incubate for 30minutes at room temperature. Protect the tube from direct

    light.

    4. Add 450L of 1X ReaLyse lysing solution to each tube and

    12

  • 8/14/2019 PI-25197-00-DTTS (CD3CD4CD45)

    13/35

  • 8/14/2019 PI-25197-00-DTTS (CD3CD4CD45)

    14/35

    View 3: CD4 Atto488 (FL1) versus CD45 PE fluorescence

    (FL2) of all events from first view. View-3 will have a gate (R2)

    to capture the fluorescent reference beads. The beads have a

    high FL1 and FL2 Mean Fluorescence Intensity.

    2. Set the instrument Threshold on FL2. Before acquisition adjust theFL-2 threshold to minimize debris.

    3. Set the number of events to capture at 3000 for R-2 gate bead

    events (Figure 1; View 3) and start data acquisition. When 3000

    beads are acquired in the gate, acquisition stops.

    4. Gate the brightest CD45+ cell cluster in View-1(Figure 1) and label

    the population as R-1. This CD45+ cell cluster is that population

    which has the lower SSC intensity and higher FL-2 mean

    fluorescence intensity. It represents the lymphocyte population. The

    Gate R-1 is applied on View-2.

    5. View-2 (Figure 1) cell clusters are separated using quadrant gating.

    The CD4+CD3+ cell populations can be identified in View-2 (Figure

    1) in the upper right quadrant as double positives. Total CD3+

    events can be identified by adding up the cell events from both the

    upper quadrants (Upper Left and Upper Right quadrants).

    14

  • 8/14/2019 PI-25197-00-DTTS (CD3CD4CD45)

    15/35

    6. The absolute count for each cell type can be calculated using the

    following equation, where the Gated Cell count and Bead count are

    obtained from the Flow Cytometer analysis and the Bead count per

    test from the test kit provided.

    inVolume(TestxCount)Bead(Gated

    tperCount(BeadxCount)Cell(GatedL/CountCellAbsolute =

    Figure 1. Dot plot views: View 1: CD45-PE fluorescence (FL2 log scale) versus

    side scatter (SSC Linear scale) View 2: CD4-Atto488 (FL1 log scale) versus

    15

  • 8/14/2019 PI-25197-00-DTTS (CD3CD4CD45)

    16/35

    CD3-PE-Dyomics649 fluorescence (FL3 log scale) of events gated as CD45+

    R1 from the first view. View 3: CD4-Atto488 (FL1) versus CD45-PE

    fluorescence (FL2) of all events from first view. View-3 will have a gate R2 to

    capture the fluorescent reference beads.

    For Beckman-Coulter Flow Cytometer instruments (EPICS series)

    1. Draw five dot-plot views Refer Figure 2.

    View 1: Display CD45 PE fluorescence (FL2 - log scale)

    versus side scatter (SSC Linear scale)

    View 2: side scatter (SSC Linear scale) versus CD3 PE-

    Dyomics649 (FL4 - log scale)

    View 3: CD4 Atto488 (FL1 log scale) versus CD3 PE-

    Dyomics649 fluorescence (FL4 log scale) of events gated as

    A (CD45+) from the first view.

    View 4: CD4 Atto488 (FL1) versus CD45 PE fluorescence

    (FL2) of all events from first view. View-4 will have a gate ( D) tocapture the fluorescent reference beads. The beads have a

    high FL1 and FL2 Mean Fluorescence Intensity.

    16

  • 8/14/2019 PI-25197-00-DTTS (CD3CD4CD45)

    17/35

    View 5: CD4 Atto488 (FL1 log scale) versus CD3 PE-

    Dyomics649 fluorescence (FL4 log scale) of events gated as

    D (Beads) from the fourth view. View-5 will have a gate (E) to

    capture the fluorescent reference beads

    2. Set the instrument Discriminator on FL2. Before acquisition adjustthe FL-2 Discriminator to minimize debris.

    3. Set the number of events to capture at 3000 for E gate bead events

    (Figure 2; View 5) and start data acquisition. When 3000 beads are

    acquired in the gate E, acquisition stops.

    4. Gate the brightest CD45+ cell cluster in View-1(Figure 2) and label

    the population as A. This CD45+ cell cluster is that population

    which has the lower SSC intensity and higher FL-2 mean

    fluorescence intensity. It represents the lymphocyte population. The

    Gate A is applied on View-2 and 3.

    5. View-2 (Figure 2); CD3+ cell clusters are separated using gate B.

    The CD3+ cell population has the lower SSC intensity and higher

    FL-4 mean fluorescence intensity.

    6. View-3 (Figure 2) cell clusters are separated using quadrant gating

    C. The CD4+CD3+ cell populations can be identified in View-3

    17

  • 8/14/2019 PI-25197-00-DTTS (CD3CD4CD45)

    18/35

    (Figure 2) in the upper right quadrant (C2) as double positives.

    7. View-4 (Figure 2); fluorescent reference beads are separated using

    gate D. The beads have a high FL1 and FL2 Mean Fluorescence

    Intensity. The gate D is applied on View-5.

    8. View-5 (Figure 2); fluorescent reference beads are separated usinggate E. Set the number of events to capture at 3000 for E gate

    bead events.

    9. Enter the Calibrator (CAL) value in the Region E on the template to

    generate absolute cell counts. The Calibrator value is calculated

    using bead count per test from the test kit provided.

    (L)VolumeTest

    TestperCountBead(CAL)Calibrator =

    18

  • 8/14/2019 PI-25197-00-DTTS (CD3CD4CD45)

    19/35

    Figure 2. Dot plot views: View 1: CD45-PE fluorescence (FL2 log scale) versus sidescatter (SSC Linear scale) View 2: CD3-PE-Dyomics649 fluorescence (FL4 log

    scale) versus side scatter (SSC Linear scale) View 3: CD4-Atto488 (FL1 log scale)

    versus CD3-PE-Dyomics649 fluorescence (FL4 log scale) of events gated as A

    19

  • 8/14/2019 PI-25197-00-DTTS (CD3CD4CD45)

    20/35

    (CD45+) from the first view. View 4: CD4-Atto488 (FL1) versus CD45-PE

    fluorescence (FL2) of all events from first view. View-4 will have a gate D to capture

    the fluorescent beads View 5: CD4-Atto488 (FL1 log scale) versus CD45-PE

    fluorescence (FL2) of events gated as D (Beads) from the fourth view. View-5 will

    have a gate E to capture the fluorescent reference beads

    For Accuri Flow Cytometer Instruments

    1. Draw four dot-plot views Refer Figure 3.

    View 1: Display CD45 PE fluorescence (FL2 - log scale)

    versus side scatter (SSC Linear scale) View 2: Side scatter (SSC Linear scale) versus CD3 PE-

    Dyomics649 (FL3 - log scale) of events gated as P1 (CD45+)

    from the first view.

    View 3: CD4 Atto488 (FL1 log scale) versus CD3 PE-

    Dyomics649 fluorescence (FL3 log scale) of events gated as

    P1 (CD45+) from the first view.

    View 4: CD4 Atto488 (FL1) versus CD45 PE fluorescence

    (FL2) of all events from first view. View-4 will have a gate P2 to

    20

  • 8/14/2019 PI-25197-00-DTTS (CD3CD4CD45)

    21/35

    capture the fluorescent reference beads. The beads have a

    high FL1 and FL2 Mean Fluorescence Intensity.

    Following table (Table-1) represents the Plot Specification Value to

    set the scales:

    Table-1: Dot Plot View X & Y-Axis Scale Specifications

    Dot Plot

    ViewsAxis Scale Min Value Max value

    View 1X-Axis FL2-Log scale 800 1,000,000

    Y-Axis SSC-Linear scale 0 180,000

    View 2

    X-Axis FL3-Log scale 300 600,000

    Y-Axis SSC-Linear scale 0 150,000

    View 3X-Axis FL1-Log scale 1000 200,000

    Y-Axis FL3-Log scale 100 5,000,000

    View 4X-Axis FL1-Log scale 1000 10,000,000

    Y-Axis FL2-Log scale 500 2,000,000

    2. In the Run Limit section of the Instrument template panel, set theinstrument Threshold on FL2 and set the value to 2000.

    Before/After acquisition adjust the FL-2 Threshold value to

    minimize debris.

    21

  • 8/14/2019 PI-25197-00-DTTS (CD3CD4CD45)

    22/35

    3. In the Run Limit section of the Instrument template panel, set the

    fluidics flow rate on MEDIUM. Set the number of events to capture

    at 3000 for P2 gate bead events (Figure 3; View 4).

    4. Click on the Set Colour Compensation to open the compensation

    matrix and set the compensation as per Figure 4.5. Vortex sample tube thoroughly at low speed and load onto flow

    cytometer for acquisition. When 3000 bead events are acquired in

    the gate P2, acquisition stops.

    6. Gate the brightest CD45+ cell cluster in View-1(Figure 3) and label

    the population as P1. This CD45+ cell cluster is that population

    which has the lower SSC intensity and higher FL2 meanfluorescence intensity. It represents the lymphocyte population. The

    Gate P1 is applied on View-2 and 3.

    7. CD3+ cell clusters are separated using gate R1 (View-2, Figure 3).

    The CD3+ cell population has the lower SSC intensity and higher

    FL3 mean fluorescence intensity.

    8. In View-3 (Figure 3) cell clusters are separated using quadrant gate

    Q1. The CD4+CD3+ cell populations can be identified in View-3

    (Figure 3) in the upper right quadrant Q1-UR as double positives.

    22

  • 8/14/2019 PI-25197-00-DTTS (CD3CD4CD45)

    23/35

    9. View-4 (Figure 3); fluorescent reference beads are separated using

    gate P2. The beads have a high FL1 and FL2 Mean Fluorescence

    Intensity.

    10. The absolute count for each cell type can be calculated using the

    following equation, where the Gated Cell count and Bead count areobtained from the Flow Cytometer analysis and the Bead count per

    test from the test kit provided.

    )LinVolume(TestxCount)Bead(Gated

    test)perCount(BeadxCount)Cell(GatedL/CountCellAbsolute =

    23

    View 1 View 2

  • 8/14/2019 PI-25197-00-DTTS (CD3CD4CD45)

    24/35

    View 3 View 4

    Figure 3. Dot plot views: View 1: CD45-PE fluorescence (FL2 log scale) versus side

    scatter (SSC Linear scale) View 2: CD3-PE-Dyomics649 fluorescence (FL3 log

    scale) versus side scatter (SSC Linear scale) View 3: CD4-Atto488 (FL1 log scale)

    versus CD3-PE-Dyomics649 fluorescence (FL3 log scale) View 4: CD4-Atto488 (FL1

    log scale) versus CD45-PE fluorescence (FL2 log scale)

    24

  • 8/14/2019 PI-25197-00-DTTS (CD3CD4CD45)

    25/35

    Figure 4: Accuri Flow cytometer Compensation settings for Dry-Tri T-STAT

    25

  • 8/14/2019 PI-25197-00-DTTS (CD3CD4CD45)

    26/35

    CD3CD4CD45 Reagent

    9. LIMITATIONS1. The Dry-Tri T-STAT CD3/CD4/CD45 reagent has only been

    validated with K3EDTA treated whole blood.

    2. Laboratories should establish their own reference ranges for

    the absolute counts obtained using the Dry-Tri T-STAT

    CD3/CD4/CD45 reagent.

    10. PERFORMANCE CHARACTERISTICS

    Accuracy

    The absolute counts for CD4+CD3+ and CD3+ T-Cells determined usingDry-Tri T-STAT CD3/CD4/CD45 reagent were compared with a

    commercially available ReaPan 34845 reagent. Aliquots of the same blood

    samples were analyzed with the two reagents at the same time. Regression

    26

  • 8/14/2019 PI-25197-00-DTTS (CD3CD4CD45)

    27/35

    analysis reported in the Table-2 below and the plots in Figures 5 - 8 indicate

    that the results are substantially equivalent.

    Table-2. Regression analysis

    T-Cell Subset n Slope Intercept R Range

    CD3+CD4+ T-Cells (cells/L) 53 1.09 -19 0.99 9 - 1786

    Total T lymphocytes - CD3+ (cells/L) 53 1.00 71 0.96 243 - 3577

    CD3+CD4+ T-Cells (%) 53 1.00 -1.1 0.99 3.3 55.6

    Absolute lymphocyte count (cells/L) 53 1.03 65 0.96 280 - 4069

    27

  • 8/14/2019 PI-25197-00-DTTS (CD3CD4CD45)

    28/35

    Figure 5. Correlation of absolute CD4+CD3+

    T-Cell counts determined using the Dry-Tri T-

    STAT CD3/CD4/CD45 reagent and comparedto ReaPan 34845 reagent

    Figure 6. Correlation of % CD4 determined

    using the Dry-Tri T-STAT CD3/CD4/CD45

    reagent and compared to ReaPan 34845reagent

    28

  • 8/14/2019 PI-25197-00-DTTS (CD3CD4CD45)

    29/35

    Figure 7. Correlation of absolute CD3+ T-Cell

    counts determined using the Dry-Tri T-STAT

    CD3/CD4/CD45 reagent and compared toReaPan 34845 reagent

    Figure 8. Correlation of Absolute

    Lymphocyte counts (ALC) determined using

    the Dry-Tri T-STAT CD3/CD4/CD45 reagentand compared to ReaPan 34845 reagent

    29

  • 8/14/2019 PI-25197-00-DTTS (CD3CD4CD45)

    30/35

    Precision

    The repeatability of measurement was determined for three samples

    representing high, medium and low CD3+CD4+ and CD3+ T-cells counts

    were analyzed by performing triplicates per level. These results are

    summarized in Table-3.

    Table-3. Repeatability of CD4+CD3+ and CD3+ T-Cell enumeration

    Cell Type Level Mean SD CV (%)

    CD4+CD3+ T-Cells

    High 917 67.5 7.3

    Medium 378 9.5 2.5

    Low 147 4.7 3.1

    CD3+ T-Cells

    High 1812 104.0 5.7

    Medium 891 29.9 3.2

    Low 361 31.7 3.1

    Specificity

    30

  • 8/14/2019 PI-25197-00-DTTS (CD3CD4CD45)

    31/35

    The antigen specificities for the CD3, CD4, and CD45 monoclonal antibodies

    used in the Dry Tri T-STAT CD3/CD4/CD45 reagent were confirmed by the

    Human Leukocyte Differentiation Antigens (HLDA) Workshops. After

    conjugation of the antibodies no additional cross-reactivity was observed.

    11. BIBLIOGRAPHY

    1. Knapp W, Dorken B, Gilks WR, Rieber EP, Schmidt RE, Stein H, &

    von dem BorneAEGKr, eds. Leukocyte typing IV. White cell

    differentiation antigens. Oxford university press 1989

    2.Reinherz EL, Meuer SC, and Schlossman SF. The delineation ofantigen receptors on human T lymphocytes. Immunology Today,

    1983, 4:5-8.

    3. Fauci AS: The human deficiency virus: Infectivity and mechanism of

    pathogenesis. Science, 1988, 239:617-622.

    4. Reinherz EL and Schlossman SF. The differentiation and function of

    human T lymphocytes. Cell, 1980, 19:821-827.5. Phillips A, Lee C, Elford J, et al. Serial CD4 lymphocyte counts and

    the development of AIDS. Lancet 1991, 337:389-392.

    6. Nicholson J. Use of flow cytometry in the evaluation and diagnosis

    31

  • 8/14/2019 PI-25197-00-DTTS (CD3CD4CD45)

    32/35

    of primary and secondary immunodeficiency diseases. Arch. Pathol.

    Lab. Med. 1989, 113:598-605.

    7. Yarchoan R, Venzon D, Pluda J, et al. CD4 count and the risk of

    death in patients infected with HIV receiving antiretroviral therapy.

    1991, 115:184-189.8. Giorgi J & Detels R. T-cell subset alterations in HIV-infected

    homosexual men: NIAID multicenter AIDS cohort study. 1989,

    52:10-18.

    9. Heddy Z, Swart B, Nicholson I, & Voss E, eds. Leukocyte and

    stromal cell molecules; the CD markers. John Wiley & Sons 2007

    12. WARRANTY

    This product is warranted only to conform to the quantity and contents

    stated on the label at the time of delivery to the customer. There are no

    warranties, expressed or implied, that extend beyond the description on

    the label of the product. ReaMetrixs sole liability is limited to

    replacement of the product. ReaMetrix is not liable for property damage,

    personal injury, or economic loss caused by the product.

    32

  • 8/14/2019 PI-25197-00-DTTS (CD3CD4CD45)

    33/35

    Note: FACScan, FACSCalibur, CaliBRITE, FACSComp are all registered trade

    names of Becton-Dickinson; EPICS, Flow-Checkare all registered trade names of

    Beckman Coulter, Accuri C6 flow cytometer is the trade name of Accuri.

    33

  • 8/14/2019 PI-25197-00-DTTS (CD3CD4CD45)

    34/35

    Manufactured by ReaMetrix India Pvt. Ltd.

    Manufacturing License Number: KTK/25/519/2006

    50-B, II Phase, Peenya Industrial Area

    Peenya, Bangalore 560058, IndiaPh: +91-80-28378693/5,

    Fax: +91-80-41172451

    E-mail: [email protected],

    34

  • 8/14/2019 PI-25197-00-DTTS (CD3CD4CD45)

    35/35

    www.reametrix.com

    Rev No. 4.0, 22-Sep-09

    35