Simulation of Implant dentar personalizat BioMicron Transilvania Date: 7 iulie 2014 Designer: Solidworks Study name: Study 1 Analysis type: Static Table of Contents Description........................1 Assumptions........................2 Model Information..................2 Study Properties...................4 Units..............................4 Material Properties................5 Loads and Fixtures.................6 Connector Definitions..............7 Contact Information................8 Mesh Information..................10 Sensor Details....................12 Resultant Forces..................12 Beams.............................12 Study Results.....................13 Conclusion........................17 Description No Data Analyzed with SolidWorks Simulation Simulation of Implant dentar personalizat BioMicron Transilvania 1

Implant Dentar Personalizat BioMicron Transilvania-Study 1-1

Embed Size (px)

Citation preview

Page 1: Implant Dentar Personalizat BioMicron Transilvania-Study 1-1

Simulation of Implant dentar personalizat BioMicron TransilvaniaDate: 7 iulie 2014Designer: SolidworksStudy name: Study 1Analysis type: Static

Table of ContentsDescription...............................................1

Assumptions.............................................2

Model Information.....................................2

Study Properties.......................................4

Units.........................................................4

Material Properties...................................5

Loads and Fixtures...................................6

Connector Definitions...............................7

Contact Information..................................8

Mesh Information....................................10

Sensor Details.........................................12

Resultant Forces.....................................12

Beams....................................................12

Study Results..........................................13

Conclusion..............................................17

DescriptionNo Data

Analyzed with SolidWorks SimulationSimulation of Implant dentar personalizat

BioMicron Transilvania 1

Page 2: Implant Dentar Personalizat BioMicron Transilvania-Study 1-1

Assumptions

Analyzed with SolidWorks SimulationSimulation of Implant dentar personalizat

BioMicron Transilvania 2

Page 3: Implant Dentar Personalizat BioMicron Transilvania-Study 1-1

Analyzed with SolidWorks SimulationSimulation of Implant dentar personalizat

BioMicron Transilvania 3

Page 4: Implant Dentar Personalizat BioMicron Transilvania-Study 1-1

Model Information

Analyzed with SolidWorks SimulationSimulation of Implant dentar personalizat

BioMicron Transilvania 4

Page 5: Implant Dentar Personalizat BioMicron Transilvania-Study 1-1

Model name: Implant dentar personalizat BioMicron Transilvania

Current Configuration: Default

Solid BodiesDocument Name and

ReferenceTreated As Volumetric Properties Document Path/Date

ModifiedCut-Revolve1

Solid Body

Mass:0.000311209 kgVolume:6.90041e-008

m^3Density:4510 kg/m^3Weight:0.00304984 N

C:\Users\A&C\Desktop\DIPLOMA

POGACEAN_DENTAL IMPLANT\SORIN

MATEI_ARHIVA 2\AU EI 4,1- 12.SLDPRT

Jul 06 19:15:27 2014

Cut-Extrude2

Solid Body

Mass:6.5389e-005 kgVolume:1.44987e-008

m^3Density:4510 kg/m^3

Weight:0.000640812 N

C:\Users\A&C\Desktop\DIPLOMA

POGACEAN_DENTAL IMPLANT\SORIN

MATEI_ARHIVA 2\AU EI 4,1.SLDPRT

Jul 06 15:36:32 2014

Cut-Revolve1

Solid Body

Mass:8.36985e-005 kgVolume:1.88988e-008

m^3Density:4428.78 kg/m^3Weight:0.000820246 N

C:\Users\A&C\Desktop\DIPLOMA

POGACEAN_DENTAL IMPLANT\SORIN

MATEI_ARHIVA 2\BM-

Analyzed with SolidWorks SimulationSimulation of Implant dentar personalizat

BioMicron Transilvania 5

Page 6: Implant Dentar Personalizat BioMicron Transilvania-Study 1-1

Study PropertiesStudy name Study 1

Analysis type Static

Mesh type Solid Mesh

Thermal Effect: On

Thermal option Include temperature loads

Zero strain temperature 298 Kelvin

Include fluid pressure effects from SolidWorks Flow Simulation

Off

Solver type FFEPlus

Inplane Effect: Off

Soft Spring: Off

Inertial Relief: Off

Incompatible bonding options Automatic

Large displacement Off

Compute free body forces On

Friction Off

Use Adaptive Method: Off

Result folder SolidWorks document (C:\Users\A&C\Desktop\DIPLOMA POGACEAN_DENTAL IMPLANT\SORIN MATEI_ARHIVA 2)

UnitsUnit system: SI (MKS)

Length/Displacement mm

Temperature Kelvin

Angular velocity Rad/sec

Pressure/Stress N/m^2

Analyzed with SolidWorks SimulationSimulation of Implant dentar personalizat

BioMicron Transilvania 6

Page 7: Implant Dentar Personalizat BioMicron Transilvania-Study 1-1

Material PropertiesModel Reference Properties Components

Name: Commercially Pure CP-Ti UNS R50700 Grade 4 (SS)

Model type: Linear Elastic Isotropic

Default failure criterion:

Max von Mises Stress

Yield strength: 5e+008 N/m^2Tensile strength: 5.5e+008 N/m^2Elastic modulus: 1.05e+011 N/m^2Poisson's ratio: 0.37

Mass density: 4510 kg/m^3Shear modulus: 4.5e+010 N/m^2

Thermal expansion

coefficient:

9e-006 /Kelvin

SolidBody 1(Cut-Revolve1)(AU EI 4,1- 12-1),SolidBody 1(Cut-Extrude2)(AU EI 4,1-1)

Curve Data:N/A

Name: Ti-6Al-4VSolution treated and aged (SS)

Model type: Linear Elastic Isotropic

Default failure criterion:

Max von Mises Stress

Yield strength: 8.27371e+008 N/m^2

Tensile strength: 1.05e+009 N/m^2Elastic modulus: 1.048e+011 N/m^2Poisson's ratio: 0.31

Mass density: 4428.78 kg/m^3Shear modulus: 4.10238e+010

N/m^2Thermal

expansion coefficient:

9e-006 /Kelvin

SolidBody 1(Cut-Revolve1)(BM-0056-1)

Curve Data:N/A

Name: Ti-6Al-4VSolution treated and aged (SS)

Model type: Linear Elastic Isotropic

Default failure criterion:

Unknown

Yield strength: 8.27371e+008 N/m^2

SolidBody 1(Chamfer1)(Bont 15 AU-VL 1-1)

Analyzed with SolidWorks SimulationSimulation of Implant dentar personalizat

BioMicron Transilvania 7

Page 8: Implant Dentar Personalizat BioMicron Transilvania-Study 1-1

Tensile strength: 1.05e+009 N/m^2Elastic modulus: 1.048e+011 N/m^2Poisson's ratio: 0.31

Mass density: 4428.78 kg/m^3Shear modulus: 4.10238e+010

N/m^2Thermal

expansion coefficient:

9e-006 /Kelvin

Curve Data:N/A

Analyzed with SolidWorks SimulationSimulation of Implant dentar personalizat

BioMicron Transilvania 8

Page 9: Implant Dentar Personalizat BioMicron Transilvania-Study 1-1

Loads and FixturesFixture name

Fixture Image Fixture Details

Fixed-1

Entities: 1 edge(s), 23 face(s)Type: Fixed Geometry

Resultant ForcesComponents X Y Z Resultant

Reaction force(N) 0.0163343 0.0214571 -349.872 349.872Reaction

Moment(N·m)0 0 0 0

Fixed-2

Entities: 11 face(s)Type: Fixed Geometry

Resultant ForcesComponents X Y Z Resultant

Reaction force(N) 0.000126843 0.000350551 -0.000837069 0.000916329Reaction

Moment(N·m)0 0 0 0

Load name Load Image Load Details

Gravity-1

Reference: Top PlaneValues: 0 0 -9.81

Units: SI

Analyzed with SolidWorks SimulationSimulation of Implant dentar personalizat

BioMicron Transilvania 9

Page 10: Implant Dentar Personalizat BioMicron Transilvania-Study 1-1

Load name Load Image Load Details

Force-1

Entities: 1 face(s), 1 plane(s)Type: Apply normal force

Value: 350 N

Connector DefinitionsNo Data

Analyzed with SolidWorks SimulationSimulation of Implant dentar personalizat

BioMicron Transilvania 10

Page 11: Implant Dentar Personalizat BioMicron Transilvania-Study 1-1

Contact Information

Contact Contact Image Contact Properties

Contact Set-1

Type: Bonded contact pair

Entites: 2 face(s)

Contact Set-2

Type: Bonded contact pair

Entites: 2 face(s)

Contact Set-4

Type: Bonded contact pair

Entites: 2 face(s)

Contact Set-5

Type: Bonded contact pair

Entites: 2 face(s)

Analyzed with SolidWorks SimulationSimulation of Implant dentar personalizat

BioMicron Transilvania 11

Page 12: Implant Dentar Personalizat BioMicron Transilvania-Study 1-1

Contact Contact Image Contact Properties

Contact Set-6

Type: Bonded contact pair

Entites: 2 face(s)

Contact Set-7

Type: Bonded contact pair

Entites: 2 face(s)

Contact Set-8

Type: Bonded contact pair

Entites: 2 face(s)

Global Contact

Type: BondedComponents: 1

component(s)Options: Compatible

mesh

Analyzed with SolidWorks SimulationSimulation of Implant dentar personalizat

BioMicron Transilvania 12

Page 13: Implant Dentar Personalizat BioMicron Transilvania-Study 1-1

Mesh InformationMesh type Solid Mesh

Mesher Used: Curvature based mesh

Jacobian points 4 Points

Maximum element size 0.592353 mm

Minimum element size 0.118471 mm

Mesh Quality High

Remesh failed parts with incompatible mesh

Off

Mesh Information - DetailsTotal Nodes 1018738

Total Elements 703194

Maximum Aspect Ratio 193.36

% of elements with Aspect Ratio < 3 99

% of elements with Aspect Ratio > 10 0.244

% of distorted elements(Jacobian) 0.00356

Time to complete mesh(hh;mm;ss): 00:00:51

Computer name: AC

Analyzed with SolidWorks SimulationSimulation of Implant dentar personalizat

BioMicron Transilvania 13

Page 14: Implant Dentar Personalizat BioMicron Transilvania-Study 1-1

Mesh Control Information:

Mesh Control Name Mesh Control Image Mesh Control Details

Control-1

Entities: 1 Solid Body (s)Units: mmSize: 0.141211

Ratio: 1.5

Control-2

Entities: 1 Solid Body (s)Units: mmSize: 0.151543

Ratio: 1.5

Control-3

Entities: 1 Solid Body (s)Units: mmSize: 0.151543

Ratio: 1.5

Control-4

Entities: 1 Solid Body (s)Units: mmSize: 0.10543

Ratio: 1.5

Analyzed with SolidWorks SimulationSimulation of Implant dentar personalizat

BioMicron Transilvania 14

Page 15: Implant Dentar Personalizat BioMicron Transilvania-Study 1-1

Sensor DetailsNo Data

Resultant Forces

Reaction ForcesSelection set

Units Sum XSum Y Sum Z Resultant

Entire Model N 0.0164535 0.0216554 -349.872 349.872

Reaction MomentsSelection set

Units Sum XSum Y Sum Z Resultant

Entire Model N·m 0 0 0 0

BeamsNo Data

Analyzed with SolidWorks SimulationSimulation of Implant dentar personalizat

BioMicron Transilvania 15

Page 16: Implant Dentar Personalizat BioMicron Transilvania-Study 1-1

Study Results

Name Type Min MaxStress1 VON: von Mises Stress 0.00151846 N/m^2

Node: 433032.40191e+009 N/m^2Node: 516497

Implant dentar personalizat BioMicron Transilvania-Study 1-Stress-Stress1

Name Type Min MaxDisplacement1 URES: Resultant Displacement 0 mm

Node: 250.0937302 mmNode: 438126

Analyzed with SolidWorks SimulationSimulation of Implant dentar personalizat

BioMicron Transilvania 16

Page 17: Implant Dentar Personalizat BioMicron Transilvania-Study 1-1

Name Type Min Max

Implant dentar personalizat BioMicron Transilvania-Study 1-Displacement-Displacement1

Name Type Min MaxStrain1 ESTRN: Equivalent Strain 6.62202e-012

Element: 1374420.0163425Element: 295982

Analyzed with SolidWorks SimulationSimulation of Implant dentar personalizat

BioMicron Transilvania 17

Page 18: Implant Dentar Personalizat BioMicron Transilvania-Study 1-1

Name Type Min Max

Implant dentar personalizat BioMicron Transilvania-Study 1-Strain-Strain1

Name TypeDisplacement1{1} Deformed Shape

Analyzed with SolidWorks SimulationSimulation of Implant dentar personalizat

BioMicron Transilvania 18

Page 19: Implant Dentar Personalizat BioMicron Transilvania-Study 1-1

Implant dentar personalizat BioMicron Transilvania-Study 1-Displacement-Displacement1{1}

Name Type Min MaxStress2 VON: von Mises Stress 0.00151846 N/m^2

Node: 433032.40191e+009 N/m^2Node: 516497

Implant dentar personalizat BioMicron Transilvania-Study 1-Stress-Stress2

Name Type Min MaxStress3 VON: von Mises Stress 0.00151846 N/m^2

Node: 433032.40191e+009 N/m^2Node: 516497

Analyzed with SolidWorks SimulationSimulation of Implant dentar personalizat

BioMicron Transilvania 19

Page 20: Implant Dentar Personalizat BioMicron Transilvania-Study 1-1

Name Type Min Max

Implant dentar personalizat BioMicron Transilvania-Study 1-Stress-Stress3

Analyzed with SolidWorks SimulationSimulation of Implant dentar personalizat

BioMicron Transilvania 20

Page 21: Implant Dentar Personalizat BioMicron Transilvania-Study 1-1

Image-1

Conclusion