Transcript
Page 1: Implant Dentar Personalizat BioMicron Transilvania-Study 1-1

Simulation of Implant dentar personalizat BioMicron TransilvaniaDate: 7 iulie 2014Designer: SolidworksStudy name: Study 1Analysis type: Static

Table of ContentsDescription...............................................1

Assumptions.............................................2

Model Information.....................................2

Study Properties.......................................4

Units.........................................................4

Material Properties...................................5

Loads and Fixtures...................................6

Connector Definitions...............................7

Contact Information..................................8

Mesh Information....................................10

Sensor Details.........................................12

Resultant Forces.....................................12

Beams....................................................12

Study Results..........................................13

Conclusion..............................................17

DescriptionNo Data

Analyzed with SolidWorks SimulationSimulation of Implant dentar personalizat

BioMicron Transilvania 1

Page 2: Implant Dentar Personalizat BioMicron Transilvania-Study 1-1

Assumptions

Analyzed with SolidWorks SimulationSimulation of Implant dentar personalizat

BioMicron Transilvania 2

Page 3: Implant Dentar Personalizat BioMicron Transilvania-Study 1-1

Analyzed with SolidWorks SimulationSimulation of Implant dentar personalizat

BioMicron Transilvania 3

Page 4: Implant Dentar Personalizat BioMicron Transilvania-Study 1-1

Model Information

Analyzed with SolidWorks SimulationSimulation of Implant dentar personalizat

BioMicron Transilvania 4

Page 5: Implant Dentar Personalizat BioMicron Transilvania-Study 1-1

Model name: Implant dentar personalizat BioMicron Transilvania

Current Configuration: Default

Solid BodiesDocument Name and

ReferenceTreated As Volumetric Properties Document Path/Date

ModifiedCut-Revolve1

Solid Body

Mass:0.000311209 kgVolume:6.90041e-008

m^3Density:4510 kg/m^3Weight:0.00304984 N

C:\Users\A&C\Desktop\DIPLOMA

POGACEAN_DENTAL IMPLANT\SORIN

MATEI_ARHIVA 2\AU EI 4,1- 12.SLDPRT

Jul 06 19:15:27 2014

Cut-Extrude2

Solid Body

Mass:6.5389e-005 kgVolume:1.44987e-008

m^3Density:4510 kg/m^3

Weight:0.000640812 N

C:\Users\A&C\Desktop\DIPLOMA

POGACEAN_DENTAL IMPLANT\SORIN

MATEI_ARHIVA 2\AU EI 4,1.SLDPRT

Jul 06 15:36:32 2014

Cut-Revolve1

Solid Body

Mass:8.36985e-005 kgVolume:1.88988e-008

m^3Density:4428.78 kg/m^3Weight:0.000820246 N

C:\Users\A&C\Desktop\DIPLOMA

POGACEAN_DENTAL IMPLANT\SORIN

MATEI_ARHIVA 2\BM-

Analyzed with SolidWorks SimulationSimulation of Implant dentar personalizat

BioMicron Transilvania 5

Page 6: Implant Dentar Personalizat BioMicron Transilvania-Study 1-1

Study PropertiesStudy name Study 1

Analysis type Static

Mesh type Solid Mesh

Thermal Effect: On

Thermal option Include temperature loads

Zero strain temperature 298 Kelvin

Include fluid pressure effects from SolidWorks Flow Simulation

Off

Solver type FFEPlus

Inplane Effect: Off

Soft Spring: Off

Inertial Relief: Off

Incompatible bonding options Automatic

Large displacement Off

Compute free body forces On

Friction Off

Use Adaptive Method: Off

Result folder SolidWorks document (C:\Users\A&C\Desktop\DIPLOMA POGACEAN_DENTAL IMPLANT\SORIN MATEI_ARHIVA 2)

UnitsUnit system: SI (MKS)

Length/Displacement mm

Temperature Kelvin

Angular velocity Rad/sec

Pressure/Stress N/m^2

Analyzed with SolidWorks SimulationSimulation of Implant dentar personalizat

BioMicron Transilvania 6

Page 7: Implant Dentar Personalizat BioMicron Transilvania-Study 1-1

Material PropertiesModel Reference Properties Components

Name: Commercially Pure CP-Ti UNS R50700 Grade 4 (SS)

Model type: Linear Elastic Isotropic

Default failure criterion:

Max von Mises Stress

Yield strength: 5e+008 N/m^2Tensile strength: 5.5e+008 N/m^2Elastic modulus: 1.05e+011 N/m^2Poisson's ratio: 0.37

Mass density: 4510 kg/m^3Shear modulus: 4.5e+010 N/m^2

Thermal expansion

coefficient:

9e-006 /Kelvin

SolidBody 1(Cut-Revolve1)(AU EI 4,1- 12-1),SolidBody 1(Cut-Extrude2)(AU EI 4,1-1)

Curve Data:N/A

Name: Ti-6Al-4VSolution treated and aged (SS)

Model type: Linear Elastic Isotropic

Default failure criterion:

Max von Mises Stress

Yield strength: 8.27371e+008 N/m^2

Tensile strength: 1.05e+009 N/m^2Elastic modulus: 1.048e+011 N/m^2Poisson's ratio: 0.31

Mass density: 4428.78 kg/m^3Shear modulus: 4.10238e+010

N/m^2Thermal

expansion coefficient:

9e-006 /Kelvin

SolidBody 1(Cut-Revolve1)(BM-0056-1)

Curve Data:N/A

Name: Ti-6Al-4VSolution treated and aged (SS)

Model type: Linear Elastic Isotropic

Default failure criterion:

Unknown

Yield strength: 8.27371e+008 N/m^2

SolidBody 1(Chamfer1)(Bont 15 AU-VL 1-1)

Analyzed with SolidWorks SimulationSimulation of Implant dentar personalizat

BioMicron Transilvania 7

Page 8: Implant Dentar Personalizat BioMicron Transilvania-Study 1-1

Tensile strength: 1.05e+009 N/m^2Elastic modulus: 1.048e+011 N/m^2Poisson's ratio: 0.31

Mass density: 4428.78 kg/m^3Shear modulus: 4.10238e+010

N/m^2Thermal

expansion coefficient:

9e-006 /Kelvin

Curve Data:N/A

Analyzed with SolidWorks SimulationSimulation of Implant dentar personalizat

BioMicron Transilvania 8

Page 9: Implant Dentar Personalizat BioMicron Transilvania-Study 1-1

Loads and FixturesFixture name

Fixture Image Fixture Details

Fixed-1

Entities: 1 edge(s), 23 face(s)Type: Fixed Geometry

Resultant ForcesComponents X Y Z Resultant

Reaction force(N) 0.0163343 0.0214571 -349.872 349.872Reaction

Moment(N·m)0 0 0 0

Fixed-2

Entities: 11 face(s)Type: Fixed Geometry

Resultant ForcesComponents X Y Z Resultant

Reaction force(N) 0.000126843 0.000350551 -0.000837069 0.000916329Reaction

Moment(N·m)0 0 0 0

Load name Load Image Load Details

Gravity-1

Reference: Top PlaneValues: 0 0 -9.81

Units: SI

Analyzed with SolidWorks SimulationSimulation of Implant dentar personalizat

BioMicron Transilvania 9

Page 10: Implant Dentar Personalizat BioMicron Transilvania-Study 1-1

Load name Load Image Load Details

Force-1

Entities: 1 face(s), 1 plane(s)Type: Apply normal force

Value: 350 N

Connector DefinitionsNo Data

Analyzed with SolidWorks SimulationSimulation of Implant dentar personalizat

BioMicron Transilvania 10

Page 11: Implant Dentar Personalizat BioMicron Transilvania-Study 1-1

Contact Information

Contact Contact Image Contact Properties

Contact Set-1

Type: Bonded contact pair

Entites: 2 face(s)

Contact Set-2

Type: Bonded contact pair

Entites: 2 face(s)

Contact Set-4

Type: Bonded contact pair

Entites: 2 face(s)

Contact Set-5

Type: Bonded contact pair

Entites: 2 face(s)

Analyzed with SolidWorks SimulationSimulation of Implant dentar personalizat

BioMicron Transilvania 11

Page 12: Implant Dentar Personalizat BioMicron Transilvania-Study 1-1

Contact Contact Image Contact Properties

Contact Set-6

Type: Bonded contact pair

Entites: 2 face(s)

Contact Set-7

Type: Bonded contact pair

Entites: 2 face(s)

Contact Set-8

Type: Bonded contact pair

Entites: 2 face(s)

Global Contact

Type: BondedComponents: 1

component(s)Options: Compatible

mesh

Analyzed with SolidWorks SimulationSimulation of Implant dentar personalizat

BioMicron Transilvania 12

Page 13: Implant Dentar Personalizat BioMicron Transilvania-Study 1-1

Mesh InformationMesh type Solid Mesh

Mesher Used: Curvature based mesh

Jacobian points 4 Points

Maximum element size 0.592353 mm

Minimum element size 0.118471 mm

Mesh Quality High

Remesh failed parts with incompatible mesh

Off

Mesh Information - DetailsTotal Nodes 1018738

Total Elements 703194

Maximum Aspect Ratio 193.36

% of elements with Aspect Ratio < 3 99

% of elements with Aspect Ratio > 10 0.244

% of distorted elements(Jacobian) 0.00356

Time to complete mesh(hh;mm;ss): 00:00:51

Computer name: AC

Analyzed with SolidWorks SimulationSimulation of Implant dentar personalizat

BioMicron Transilvania 13

Page 14: Implant Dentar Personalizat BioMicron Transilvania-Study 1-1

Mesh Control Information:

Mesh Control Name Mesh Control Image Mesh Control Details

Control-1

Entities: 1 Solid Body (s)Units: mmSize: 0.141211

Ratio: 1.5

Control-2

Entities: 1 Solid Body (s)Units: mmSize: 0.151543

Ratio: 1.5

Control-3

Entities: 1 Solid Body (s)Units: mmSize: 0.151543

Ratio: 1.5

Control-4

Entities: 1 Solid Body (s)Units: mmSize: 0.10543

Ratio: 1.5

Analyzed with SolidWorks SimulationSimulation of Implant dentar personalizat

BioMicron Transilvania 14

Page 15: Implant Dentar Personalizat BioMicron Transilvania-Study 1-1

Sensor DetailsNo Data

Resultant Forces

Reaction ForcesSelection set

Units Sum XSum Y Sum Z Resultant

Entire Model N 0.0164535 0.0216554 -349.872 349.872

Reaction MomentsSelection set

Units Sum XSum Y Sum Z Resultant

Entire Model N·m 0 0 0 0

BeamsNo Data

Analyzed with SolidWorks SimulationSimulation of Implant dentar personalizat

BioMicron Transilvania 15

Page 16: Implant Dentar Personalizat BioMicron Transilvania-Study 1-1

Study Results

Name Type Min MaxStress1 VON: von Mises Stress 0.00151846 N/m^2

Node: 433032.40191e+009 N/m^2Node: 516497

Implant dentar personalizat BioMicron Transilvania-Study 1-Stress-Stress1

Name Type Min MaxDisplacement1 URES: Resultant Displacement 0 mm

Node: 250.0937302 mmNode: 438126

Analyzed with SolidWorks SimulationSimulation of Implant dentar personalizat

BioMicron Transilvania 16

Page 17: Implant Dentar Personalizat BioMicron Transilvania-Study 1-1

Name Type Min Max

Implant dentar personalizat BioMicron Transilvania-Study 1-Displacement-Displacement1

Name Type Min MaxStrain1 ESTRN: Equivalent Strain 6.62202e-012

Element: 1374420.0163425Element: 295982

Analyzed with SolidWorks SimulationSimulation of Implant dentar personalizat

BioMicron Transilvania 17

Page 18: Implant Dentar Personalizat BioMicron Transilvania-Study 1-1

Name Type Min Max

Implant dentar personalizat BioMicron Transilvania-Study 1-Strain-Strain1

Name TypeDisplacement1{1} Deformed Shape

Analyzed with SolidWorks SimulationSimulation of Implant dentar personalizat

BioMicron Transilvania 18

Page 19: Implant Dentar Personalizat BioMicron Transilvania-Study 1-1

Implant dentar personalizat BioMicron Transilvania-Study 1-Displacement-Displacement1{1}

Name Type Min MaxStress2 VON: von Mises Stress 0.00151846 N/m^2

Node: 433032.40191e+009 N/m^2Node: 516497

Implant dentar personalizat BioMicron Transilvania-Study 1-Stress-Stress2

Name Type Min MaxStress3 VON: von Mises Stress 0.00151846 N/m^2

Node: 433032.40191e+009 N/m^2Node: 516497

Analyzed with SolidWorks SimulationSimulation of Implant dentar personalizat

BioMicron Transilvania 19

Page 20: Implant Dentar Personalizat BioMicron Transilvania-Study 1-1

Name Type Min Max

Implant dentar personalizat BioMicron Transilvania-Study 1-Stress-Stress3

Analyzed with SolidWorks SimulationSimulation of Implant dentar personalizat

BioMicron Transilvania 20

Page 21: Implant Dentar Personalizat BioMicron Transilvania-Study 1-1

Image-1

Conclusion

Analyzed with SolidWorks SimulationSimulation of Implant dentar personalizat

BioMicron Transilvania 21