2014 http://informahealthcare.com/lpr ISSN: 0898-2104 (print), 1532-2394 (electronic) J Liposome Res, 2014; 24(1): 53–58 ! 2014 Informa Healthcare USA, Inc. DOI: 10.3109/08982104.2013.833224 RESEARCH ARTICLE Density gradient separation of carborane-containing liposome from low density lipoprotein and detection by inductively coupled plasma spectrometry Fars K. Alanazi 1,2,3 , D. Robert Lu 3 , Faiyaz Shakeel 1,4 , and Nazrul Haq 1,4 1 Department of Pharmaceutics, College of Pharmacy, King Saud University, Riyadh, Saudi Arabia, 2 Kayyali Chair for Pharmaceutical Industry, Department of Pharmaceutics, College of Pharmacy, King Saud University, Riyadh, Saudi Arabia, 3 Department of Pharmaceutical and Biomedical Sciences, College of Pharmacy, The University of Georgia, Athens, GA, USA, and 4 Center of Excellence in Biotechnology Research (CEBR), King Saud University, Riyadh, Saudi Arabia Abstract Inductively coupled plasma atomic emission spectroscopy (ICP-AES) was used for analyzing the new cholesterol-based compounds (BCH, BCH-Da, BCH-Db and BCH-Dc) in liposomal formulations. Not only the boron compounds but also the phospholipid compositions of the liposome formulation were quantitatively analyzed. Reasonable limit of detection for boron (0.5 mg/ml) and phosphorous (0.09 mg/ml), respectively, was observed. ICP-MS was also utilized for analyzing BCH in a brain distribution study. The detection limit of boron analysis by ICP-MS is at least three orders of magnitude lower than of that of ICP-AES (1 ng B/ml). The method was linear in the range of 500-1 ng B/ml and the linearity correlation coefficient was 1. In addition, an ultracentrifugation method was developed to separate liposomes from low-density lipoprotein (LDL). Factors such as density gradient and size of liposomes were adjusted to optimize separation and it was observed that in conjunction to time, speed and density gradient, size of the liposome also had impact on the separation using centrifugation method. These findings show the importance of ICP-AES as an analytical method for the analysis of element-based compounds encapsulated in phospholipid vesicles. Keywords Boron compound, inductively coupled plasma, liposomes, low-density lipoprotein, ultracentrifugation History Received 30 May 2013 Revised 27 July 2013 Accepted 6 August 2013 Published online 5 September 2013 Introduction Because of high potential of boronated compounds in cancer therapy, they are currently of great interest for pharmacist and drug discovery scientists (Barth, 2003; Barth et al., 2005; Bregadze et al., 2006; Valliant et al., 2002). With such great importance comes the need to find a good analytical method for analyzing these compounds in various matrices such as plasma, tissues and other biological matrices such as liposomal formulations. Inductively coupled plasma spec- trometry (ICP) can be used in the quantitative analysis in several research areas such as natural materials (rocks, minerals, soil, sediment air, water and plant) and animal tissues. Furthermore, it has been used widely for quantitative analysis of compound-containing elements, for example, boron compounds (Hiratsuka et al., 2000; Morrison et al., 2010; Svantesson et al., 2002), selenium compounds (Li et al., 1999; Zheng et al., 2002) and platinum compounds (Liu et al., 2002). An advantage of using ICP as a bio-analytical technique is its ability to identify and quantify most of the elements. Since many wavelengths of varied sensitivity are available for determination of any one element, ICP is suitable for all concentrations from ultra-trace levels to major components. Detection limits are generally low for most elements with a typical range of 1 mg/l (lower limit) to 1 g/l (upper limit). The main advantage of employing an ICP when performing quantitative analysis is the fact that multi-elemental analysis can be accomplished rapidly. A complete multi-element analysis can be undertaken in a period as short as 30 seconds, consuming less than 0.5 ml of sample solution. In the present study, new cholesterol-based carborane compounds have been synthesized as potential anticancer targeting molecules (Alanazi et al., 2003; Ji et al., 2002). These compounds are designed to mimic the native choles- terol esters and to be delivered through the elevated LDL receptor for cancer treatment. In order to enhance the efficacy of boron neutron capture therapy, it is of utmost importance to deliver the boron-containing compounds deep into the malignant colonies in the vicinity of nucleus and increase the chances of selective uptake of boronated compounds. Liposomes have been chosen to facilitate the delivery of targeted molecules to an endogenous cholesterol carrier Address for correspondence: Dr. Nazrul Haq, Assistant Professor, Center of Excellence in Biotechnology Research (CEBR), King Saud Uni- versity, P.O. Box 2460, Riyadh-11451, Saudi Arabia. Tel: +966- 531148715. E-mail: [email protected] Journal of Liposome Research Downloaded from informahealthcare.com by University of Melbourne on 09/15/14 For personal use only.

Density gradient separation of carborane-containing liposome from low density lipoprotein and detection by inductively coupled plasma spectrometry

  • Upload
    nazrul

  • View
    216

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Density gradient separation of carborane-containing liposome from low density lipoprotein and detection by inductively coupled plasma spectrometry

2014

http://informahealthcare.com/lprISSN: 0898-2104 (print), 1532-2394 (electronic)

J Liposome Res, 2014; 24(1): 53–58! 2014 Informa Healthcare USA, Inc. DOI: 10.3109/08982104.2013.833224

RESEARCH ARTICLE

Density gradient separation of carborane-containing liposome from lowdensity lipoprotein and detection by inductively coupled plasmaspectrometry

Fars K. Alanazi1,2,3, D. Robert Lu3, Faiyaz Shakeel1,4, and Nazrul Haq1,4

1Department of Pharmaceutics, College of Pharmacy, King Saud University, Riyadh, Saudi Arabia, 2Kayyali Chair for Pharmaceutical Industry,

Department of Pharmaceutics, College of Pharmacy, King Saud University, Riyadh, Saudi Arabia, 3Department of Pharmaceutical and Biomedical

Sciences, College of Pharmacy, The University of Georgia, Athens, GA, USA, and 4Center of Excellence in Biotechnology Research (CEBR), King Saud

University, Riyadh, Saudi Arabia

Abstract

Inductively coupled plasma atomic emission spectroscopy (ICP-AES) was used for analyzing thenew cholesterol-based compounds (BCH, BCH-Da, BCH-Db and BCH-Dc) in liposomalformulations. Not only the boron compounds but also the phospholipid compositions of theliposome formulation were quantitatively analyzed. Reasonable limit of detection for boron(0.5mg/ml) and phosphorous (0.09 mg/ml), respectively, was observed. ICP-MS was also utilizedfor analyzing BCH in a brain distribution study. The detection limit of boron analysis by ICP-MSis at least three orders of magnitude lower than of that of ICP-AES (1 ng B/ml). The method waslinear in the range of 500-1 ng B/ml and the linearity correlation coefficient was 1. In addition,an ultracentrifugation method was developed to separate liposomes from low-densitylipoprotein (LDL). Factors such as density gradient and size of liposomes were adjusted tooptimize separation and it was observed that in conjunction to time, speed and densitygradient, size of the liposome also had impact on the separation using centrifugation method.These findings show the importance of ICP-AES as an analytical method for the analysis ofelement-based compounds encapsulated in phospholipid vesicles.

Keywords

Boron compound, inductively coupledplasma, liposomes, low-densitylipoprotein, ultracentrifugation

History

Received 30 May 2013Revised 27 July 2013Accepted 6 August 2013Published online 5 September 2013

Introduction

Because of high potential of boronated compounds in cancer

therapy, they are currently of great interest for pharmacist and

drug discovery scientists (Barth, 2003; Barth et al., 2005;

Bregadze et al., 2006; Valliant et al., 2002). With such great

importance comes the need to find a good analytical method

for analyzing these compounds in various matrices such as

plasma, tissues and other biological matrices such as

liposomal formulations. Inductively coupled plasma spec-

trometry (ICP) can be used in the quantitative analysis in

several research areas such as natural materials (rocks,

minerals, soil, sediment air, water and plant) and animal

tissues. Furthermore, it has been used widely for quantitative

analysis of compound-containing elements, for example,

boron compounds (Hiratsuka et al., 2000; Morrison et al.,

2010; Svantesson et al., 2002), selenium compounds (Li et al.,

1999; Zheng et al., 2002) and platinum compounds (Liu et al.,

2002).

An advantage of using ICP as a bio-analytical technique is

its ability to identify and quantify most of the elements. Since

many wavelengths of varied sensitivity are available for

determination of any one element, ICP is suitable for all

concentrations from ultra-trace levels to major components.

Detection limits are generally low for most elements with a

typical range of 1 mg/l (lower limit) to 1 g/l (upper limit). The

main advantage of employing an ICP when performing

quantitative analysis is the fact that multi-elemental analysis

can be accomplished rapidly. A complete multi-element

analysis can be undertaken in a period as short as 30 seconds,

consuming less than 0.5 ml of sample solution.

In the present study, new cholesterol-based carborane

compounds have been synthesized as potential anticancer

targeting molecules (Alanazi et al., 2003; Ji et al., 2002).

These compounds are designed to mimic the native choles-

terol esters and to be delivered through the elevated LDL

receptor for cancer treatment. In order to enhance the efficacy

of boron neutron capture therapy, it is of utmost importance to

deliver the boron-containing compounds deep into the

malignant colonies in the vicinity of nucleus and increase

the chances of selective uptake of boronated compounds.

Liposomes have been chosen to facilitate the delivery of

targeted molecules to an endogenous cholesterol carrier

Address for correspondence: Dr. Nazrul Haq, Assistant Professor, Centerof Excellence in Biotechnology Research (CEBR), King Saud Uni-versity, P.O. Box 2460, Riyadh-11451, Saudi Arabia. Tel: +966-531148715. E-mail: [email protected]

Jour

nal o

f L

ipos

ome

Res

earc

h D

ownl

oade

d fr

om in

form

ahea

lthca

re.c

om b

y U

nive

rsity

of

Mel

bour

ne o

n 09

/15/

14Fo

r pe

rson

al u

se o

nly.

Page 2: Density gradient separation of carborane-containing liposome from low density lipoprotein and detection by inductively coupled plasma spectrometry

(LDL) and to tumor tissue accordingly (Heyes et al., 2002;

Kader et al., 1998; Meidan et al., 2000).

In the present study, ICP-AES (inductively coupled plasma

atomic emission spectroscopy) was used for analyzing the

new cholesterol-based compounds (BCH, BCH-Da, BCH-Db

and BCH-Dc) in liposomal formulations. These compounds

are boron derivatives. Each drug molecule contains 10 boron

atoms. Detection of the boron in the samples was used as

indirect method for analyzing the whole drug molecule. The

determination of phospholipids in liposome suspension is

necessary to estimate the loss or dilution of phospholipids.

Many methods, such as the spectrophotometric method based

on digestion of phospholipid molecules with acid, have been

developed for the determination of phospholipids (Hidalgo

et al., 2004). However, these methods are laborious and time-

consuming. Notably, only few reports on the determination of

phospholipids in liposomal suspension by ICP-AES were

found (Kamidate et al., 1996; Suita & Kamidate, 1999; Wasan

et al., 1997). Multi-element analysis as provided by ICP-AES

allows measurement of boron concentration as well as

phosphorus concentration at the same time. Consequently,

phosphorus concentration was used to calculate phospholipid

concentration to determine incorporation efficiency.

Additionally, inductively coupled plasma-mass spectrometry

(ICP-MS) was used for analyzing BCH in brain tissue samples

in a brain distribution study.

One of the objectives of the study after synthesizing the

cholesterol-based compounds and incorporating them in a

suitable drug delivery system (liposomes) is to examine the

transfer of these compounds from the liposomal formulation

to low density lipoprotein (LDL) in vitro with different

incubation media. To conduct this study, it was imperative to

develop a method to separate the liposomal formulation from

LDL after finishing the incubation period. Looking into the

literature reveals that the technique most commonly used for

the aforesaid purpose in clinical research laboratories is ultra-

centrifugation. Ultra-centrifugation was used for separation

based on the fact that at solvent density higher than the

liposomes or lipoproteins, they float at rates based on their

densities. This technique has been used to separate liposomes

from lipoproteins (Kader et al., 1998; Wasan et al., 1993,

1997). Liposomal nystatin (particle size 321� 192 nm) was

recovered from solution at density values range of 1.21 to

1.25 g/ml (on the bottom) and LDL was recovered at a density

of 1.063 g/ml [on the top] (Wasan et al., 1997). Also, it has

been reported that doxorubicin liposomal preparation floated

on the top of solution with density of 1.006 g/ml and LDL

remained at the bottom (Kader et al., 1998). The size of the

liposomes was not clarified (&200 nm). Not only did

liposomes float differently according to these reports but

also the processes were time consuming (20 h). In the present

study, details of how to separate liposomes from LDL by

ultra-centrifugation are presented.

Materials and methods

Materials

BCH (cholesteryl 1,12-dicarba-closo-dodecacaborane 1-

carboxylate), BCH-Da (cholesteryl-3b-12-octyl-1,

12-dicarbo-closo-dodecacaborane-1-carboxylate), BCH-Db

(cholesteryl-3b-12-(1-octenyl)-1,12-dicarbo-closo-dodecaca-

borane-1-carboxylate), BCHDc (cholesteryl-3b-12-octadecyl-

1,12-dicarbo-closo-dodecacaborane-1-carboxylate) have been

synthesized and purified according to a method published by

Alanazi et al. (2003) and Ji et al. (2002). Cholesterol, nile red

dye and dipalmitoyl DL-a- phosphotidylcholine (DPPC) were

purchased from Sigma Aldrich (St. Louis, MO). Liposomal

formulations of these compounds have been prepared as

described earlier (Alanazi et al., 2003). Low-density lipopro-

tein from human plasma (5–7 mg/ml protein solution in

0.15 M NaCl; 0.01% EDTA; pH 7.4) was obtained from

Sigma Aldrich (St. Louis, MO). All other chemicals were of

analytical grade. ICP instruments are VG inductively coupled

plasma mass spectrometer (ICP-MS) and the thermo Jarrell-

Ash 965 inductively coupled argon plasma (ICP-AES).

Densities of the prepared solution were measured with a

specific gravity balance (Henery Troemner, Philadephia, PA).

Analysis of boron compounds in liposomes usingICP-AES

The concentration of boron compounds in the liposomal

formulations was measured by ICP-AES. About 0.5–1 ml

samples were diluted to sufficient volume (about 2 ml).

Deionized water was used to formulate liposomes and for

dilutions. The steps involved in the analysis of compounds in

liposomal samples by ICP-AES were as follows: the

liposomal formulation was introduced to the system by

using an electronic controlled pump followed by the

nebulization step in which the sample was converted from

liquid to aerosol via nebuliser. In the next step, the

desolvation/volatization step, water was driven off and

remaining solid and liquid portions were converted to gases.

In the subsequent step called atomization step, gas phase

bonds were broken, and only atoms were present. In the

penultimate step, i.e. excitation/emission step, atoms gain

energy from collisions and emit light of a characteristic

wavelength (Table 1). In the final step, the separation/

detection step, a grating disperses light that was quantitatively

measured. Two control steps were used to justify the analysis:

Running a standard matrix containing most of the elements

between the unknown samples to exclude any fluctuation of

the data due to instrumental changes during the analysis and

running boric acid (boron reference solution, Fisher

Chemicals, Fair Lawn, NJ) as an external standard for boron

analysis.

Elements are often measured by ICP-AES in units of ppm

metal (part-per-million) based on mass:

ppm metal ¼ g metal

106 g sample¼ mg metal

gm of sample:

Table 1. Properties and ICP-AES limit of detections of boron andphosphorus.

Element Mol. wt. Detection limits (ppm) Wavelength (nm)

Boron (B) 10.81 0.50–600 249.68Phosphorus (P) 30.97 0.09–900 178.29

54 F. K. Alanazi et al. J Liposome Res, 2014; 24(1): 53–58

Jour

nal o

f L

ipos

ome

Res

earc

h D

ownl

oade

d fr

om in

form

ahea

lthca

re.c

om b

y U

nive

rsity

of

Mel

bour

ne o

n 09

/15/

14Fo

r pe

rson

al u

se o

nly.

Page 3: Density gradient separation of carborane-containing liposome from low density lipoprotein and detection by inductively coupled plasma spectrometry

An ICP requires that the elements, which are to be

analyzed, be in solution. The samples are mostly aqueous

samples (water as solvent or suspending media) with density

of 1 g/ml. Therefore, it could be presented as follows:

ppm metal ¼ mg metal

gm of sample¼ mg metal

ml of sample:

Elements are often measured by ICP-MS in units of ppp

metal (part-per-pillion):

ppp metal ¼ mg metal

kg of sample¼ mg metal

Liter of sample:

Analysis of tissue samples by ICP-MS

Brain tissue samples were collected from normal tissue and

tumor originated tissue. The concentration of boron com-

pounds was measured using ICP-MS. Tissue samples were

weighed and placed in a teflon tube. Five hundred ml of nitric

acid was added. The teflon tube was capped and seated in the

metallic jacket (teflon bomb). It was, then, placed in an oven

at about 140 �C for more than 60 min. After leaving it to cool,

digested solution was transferred to a plastic tube. The

remaining traces of digested solution in the teflon tube was

washed with 500 ml of deionized water and transferred to the

plastic tube to complete the sample (to be analyzed) to 1 ml.

The steps involved in the analysis of BCH in tissue samples

by ICP-MS were the same of ICP-AES, except the analyte was

analyzed with a mass spectrometer. Because ICP-MS was

used for single element analysis (i.e. boron), only boric acid

as an external control was used.

Ultra-centrifugation for separating LDL fromliposomal formulation

Liposomal formulation of BCH as a representative model of

the other compounds was used in this study. Preparation of

BCH liposomal formulation was described previously

(Alanazi et al., 2003). Size of the liposome was reduced

using extrusion methods (LipoFast extrusion device, Avestin,

Ottawa, Canada). Nicomp submicron particle sizer was used

to measure particle size (Model 380, Nicomp, San Francisco,

CA). For determining the density of BCH liposomes, 200 ml

of BCH liposomal formulation was added to 1.8 ml of

phosphate buffer saline (PBS) [tube A]. About fifty ml of

nile red dye solution (100mg/ml nile red dye in acetone) was

placed in a glass tube (tube B) and dried using nitrogen gas.

Tube A was added to tube B and the dye re-suspended through

the mixture via shaking. The density of the mixture was then

adjusted to 1.21 g/ml with KBr. The mixtures were placed in

centrifuge tubes (13 ml in size, n¼ 6). The rest of the tubes

were filled with additional KBr solutions (densities; 1.03 g/

ml, 1.04 g/ml, 1.05 g/ml, 1.063 g/ml, 1.07 g/ml and 1.08 g/ml

for tube 6, 5, 4, 3, 2 and 1, respectively) (Table 2). The tubes

were centrifuged at 40,000 rpm at 10 �C for 1 h, and the rotor

was allowed to stop without using the brake (SW 40 rotor).

Also, the effect of liposome size on the floating behavior was

studied. Large liposomes were prepared with the same

procedure as small liposomes except their size was not

reduced by the extrusion method. Two hundred ml of large

liposomes (MLV) and small liposomes went through the same

conditions described for tube 3 (the bottom density was

1.21 g/ml and the top density was 1.063 g/ml). In addition,

large liposomes and LDL were compared in terms of floating

behavior when they were treated similarly. Two hundred ml of

BCH liposomal formulation and 100 ml of LDL were

centrifuged at similar conditions to that of tube 3. All tubes

were exposed to UV light to make the visualization of the nile

red dye complex with either liposomes or LDL possible.

Results and discussion

Cholesteryl carborane conjugates were synthesized in our

laboratory to mimic native cholesterol esters as novel

anticancer compounds (Alanazi et al., 2003; Ji et al., 2002).

As these compounds share similar chemical and physical

characteristics with native cholesteryl esters, they may be

effectively transferred into LDL in biological fluids and thus,

follow the LDL pathway for targeted delivery utilizing the

elevated LDL receptor expression on tumor cells (Alanazi

et al., 2003). Inductively coupled plasma spectrometry (ICP)

was used to analyze these compounds based on indirect boron

analysis. In the present study, the usefulness of ICP as

analytical techniques for such compounds in liposomal

formulation and brain tissue is presented.

Results from analyzing boron compounds (for boron) and

phospholipids (for phosphorus) in liposomal formulation are

presented in Figure 1. When ICP-AES was used to measure

boron compound concentrations in liposomal formulation, the

method was simple and easy since it does not require drug

extraction from liposomes (i.e. no pre-analysis treatment). In

addition, it was very fast and it had a reasonable limit of

detection for boron (0.5 mg/ml) and phosphorous (0.09 mg/ml).

The method was linear in the range of 50–0.5 mg of B/ml and

the linearity correlation coefficient was 0.9994 (Figure 2A).

In the present study, ICP-AES was used to obtain boron

compound concentrations as well as phospholipid concentra-

tion at the same time and on the same run. This finding

provides a feasible method for analyzing element-based

compounds encapsulated in phospholipid vesicles.

Results of the determination of boron compound concen-

trations in brain tissues using ICP-MS are presented in

Table 3. The detection limit of boron analysis by ICP-MS is at

least three orders of magnitude lower than of that of ICP-AES

(1 ng B/ml). The method was linear in range of 500-1 ng B/ml

and the linearity correlation coefficient was 1 (Figure 2B).

While the method has the same advantage of ICP-AES, it

needs the tissue samples to be digested which increases time

of analysis (sample preparation). Dilution of the samples is

Table 2. Density gradient used for determination of density of BCHliposomal pellets.

Tube Top r (g/ml) Bottom r (g/ml) Liposome r (g/ml)

1 1.08 1.21 51.082 1.07 1.21 51.073 1.063 1.21 ¼1.0634 1.05 1.21 41.055 1.04 1.21 41.046 1.03 1.21 41.03

DOI: 10.3109/08982104.2013.833224 Density gradient separation of carborane-containing liposome 55

Jour

nal o

f L

ipos

ome

Res

earc

h D

ownl

oade

d fr

om in

form

ahea

lthca

re.c

om b

y U

nive

rsity

of

Mel

bour

ne o

n 09

/15/

14Fo

r pe

rson

al u

se o

nly.

Page 4: Density gradient separation of carborane-containing liposome from low density lipoprotein and detection by inductively coupled plasma spectrometry

required since the viscous liquid may result in clogging of the

instrumentation.

In the present study, a description of how to separate

liposomes from LDL by an ultra-centrifugation technique is

presented. The result of determining the density of BCH

liposomes (small liposome) is shown in Figure 3(A). Nile

red complex with liposomes gave a fluorescent pink color

(see online version) under UV light. Liposomes were

recovered on the top of tube 1 and 2. This indicated that

the liposomes had a density less than 1.07 g/ml; but in

tubes 4, 5 and 6, liposomes were retained at the interface

between the two densities (the top densities: 1.05–1.03 g/ml

and the bottom density: 1.21 g/ml). This indicated that the

liposomes had a density greater than 1.05 g/ml. Results

from tube 4 indicated that liposomes were distributed

throughout the 1.063 layer. This showed that the liposomes

had a density of 1.063 g/ml. In general, LDL has a density

of 1.019–1.063 g/ml. Therefore; the same experimental

conditions of tube 3 (top density 1.063 and bottom density

is 1.21) were used to find the density of LDL. The results

showed that LDL was retained at the interface between the

two densities. Thus, the separation of LDL from liposomes

is not complete because they were overlapped on the

bottom fraction of the 1.063 layer. To separate them

efficiently liposomes should float on the top of the 1.063

density solution and LDL should be retained between the

two solutions. The only factor that can be modified to

adjust the floatation is the size of liposome. Also, no report

has been found in the literature on the effect of liposome

size on improving the LDL–liposome separation. In this

study, the effect of liposome size on floating behaviors was

Figure 1. Results from analyzing boron compounds in liposomal formulation by ICP-AES.

Figure 2. Linearity plots of ICP-AES (A) and ICP-MS (B).

Table 3. Results from analyzing BCH in brain tissue sample by ICP-MS.

Timeb

Tissue type 2 h 6 h

Tumora 27.7� 5.54 12.5� 2.5Normala 5.07� 1.014 7.99� 1.598

aConcentrations are in mg/g of tissue.bTime after injection of BCH intracerebrally.

56 F. K. Alanazi et al. J Liposome Res, 2014; 24(1): 53–58

Jour

nal o

f L

ipos

ome

Res

earc

h D

ownl

oade

d fr

om in

form

ahea

lthca

re.c

om b

y U

nive

rsity

of

Mel

bour

ne o

n 09

/15/

14Fo

r pe

rson

al u

se o

nly.

Page 5: Density gradient separation of carborane-containing liposome from low density lipoprotein and detection by inductively coupled plasma spectrometry

studied. The result is shown in Figure 3(B). Large

liposomes (tube a, 1.63 mm) were recovered on the top of

the 1.063 layer. On the other hand, small liposomes (tube

b, 34 nm) were distributed throughout the 1.063 layer. This

can be explained based on isopycnic theory (also termed

density gradient centrifugation and sedimentation equilib-

rium centrifugation). The separation is based on particle

density. When samples were underlaid at the bottom of the

tube, liposomal particles floated to their corrected densities

during centrifugation (i.e. molecules move to the position

in the gradient where their density is the same as the

gradient material). Small liposomes had a greater phospho-

lipid to water ratio (large volume of liposomal phospho-

lipid) than large liposomes (Versluis et al., 1999).

Consequently, the small liposomes are denser than the

large liposomes. Shawer et al. reported similar phenomena

with submicron emulsions (SE); the largest particle size SE

were recovered first and then the smallest ones (Shawer

et al., 2002). To confirm the success of this method to

separate LDL from liposomes, large liposomes and LDL

were ultracentrifuged in separate tube under the same

conditions. Liposomes were recovered on the top of the

1.063 layer and LDL was retained at the interface between

the two density layers (Figure 3C).

Conclusion

ICP spectrometry was used to analyze these compounds

indirectly based on boron analysis. ICP-AES was used to

obtain boron compound concentrations as well as phospho-

lipid concentration at the same time and on the same run.

This finding provides a feasible method for analyzing

element-based compounds encapsulated in phospholipid

vesicles. ICP-MS was used for analyzing BCH in brain

tissue samples in brain distribution study. Details of how to

separate liposomes from LDL by ultra-centrifugation tech-

nique are presented. Not only time, speed and density

gradient but also the size of the liposome has effects on the

separation using centrifugation method. This method gave a

good separation between LDL and liposome and could be

used in future studies.

Acknowledgements

The authors are grateful to Department of Pharmaceutical and

Biomedical Sciences, College of Pharmacy, The University of

Georgia, Athens, GA, USA for facilitating the equipments and

laboratories used.

Declaration of interest

The authors report no declaration of interest. The authors

alone are responsible for the content and writing of the article.

The authors are also thankful to National Plan for Science

& Technology and Innovation for generous financial support

(Grant No. 11 NAN 1286-02).

References

Alanazi F, Li H, Halpern DS, et al. (2003). Synthesis, preformulation andliposomal formulation of cholesteryl carborane esters with variousfatty chains. Int J Pharm 255:189–97.

Barth RF. (2003). A critical assessment of boron neutron capturetherapy: an overview. J Neurooncol 62:1–5.

Barth RF, Coderre JA, Vicente MGH, Blue TE. (2005). Boron neutroncapture therapy of cancer: current status and future prospects. ClinCancer Res 11:3987–4002.

Bregadze VI, Sivaev IB, Glazun SA. (2006). Polyhedral boroncompounds as potential diagnostic and therapeutic antitumor agents.Anticancer Agents Med Chem 6:75–109.

Heyes JA, Niculescu-Duvaz D, Cooper RG, Springer CJ. (2002).Synthesis of novel cationic lipids: effect of structural modificationon the efficiency of gene transfer. J Med Chem 45:99–114.

Hidalgo FJ, Nogales F, Zamora R. (2004). Determination of pyrrolizedphospholipids in oxidized phospholipid vesicles and lipoproteins.Anal Biochem 334:155–63.

Hiratsuka J, Yoshino K, Kondoh H, et al. (2000). Biodistribution ofboron concentration on melanoma-bearing hamsters after administra-tion of p-, m-, o-boronophenylalanine. Jpn J Cancer Res 91:446–50.

Ji B, Peacock G, Lu DR. (2002). Synthesis of cholesterol–carboraneconjugate for targeted drug delivery. Bioorg Med Chem Lett 12:2455–8.

Kader A, Davis PJ, Kara M, Liu H. (1998). Drug targeting using low-density lipoprotein (LDL): physicochemical factors affecting drugloading into LDL particles. J Control Rel 55:231–43.

Kamidate T, Suita T, Watanabe H. (1996). Assay of phosphatidylcholinein liposome suspensions by inductively coupled plasma atomicemission spectrometry with ultrasonic nebulizer. Anal Biochem 241:264–6.

Figure 3. Results of determining density of BCH liposomal pellets (A), comparison between large [a] and small liposome [b] (B) and comparisonbetween LDL [a] and large liposome [b] (C).

DOI: 10.3109/08982104.2013.833224 Density gradient separation of carborane-containing liposome 57

Jour

nal o

f L

ipos

ome

Res

earc

h D

ownl

oade

d fr

om in

form

ahea

lthca

re.c

om b

y U

nive

rsity

of

Mel

bour

ne o

n 09

/15/

14Fo

r pe

rson

al u

se o

nly.

Page 6: Density gradient separation of carborane-containing liposome from low density lipoprotein and detection by inductively coupled plasma spectrometry

Li F, Goessler W, Irgolic KJ. (1999). Determination of seleniumcompounds by HPLC with ICP-MS or FAAS as selenium-specificdetector. Se Pu 17:240–4.

Liu J, Kraut E, Bender J, et al. (2002). Pharmacokinetics of oxaliplatin(NSC 266046) alone and in combination with paclitaxel in cancerpatients. Cancer Chemother Pharmacol 49:367–74.

Meidan VM, Cohen JS, Amariglio N, et al. (2000). Interaction ofoligonucleotides with cationic lipids: the relationship between electro-statics, hydration and state of aggregation. Biochim Biophys Acta 1464:251–61.

Morrison DE, Issa F, Bhadbhade M, et al. (2010). Boronatedphosphonium salts containing arylboronic acid, closo-carborane, ornido-carborane: synthesis, X-ray diffraction, in vitro cytotoxicity andcellular uptake. J Biol Inorg Chem 15:1305–18.

Shawer M, Greenspan P, Oie S, Lu DR. (2002). VLDL-resemblingphospholipid-submicron emulsion for cholesterol-based drug target-ing. J Pharm Sci 91:1405–13.

Suita T, Kamidate T. (1999). Preparation of antibody-coupled liposomescontaining horseradish peroxidase as a marker molecule. Anal Sci 15:349–52.

Svantesson E, Capala J, Markides KE, Pettersson J. (2002).Determination of boron-containing compounds in urine and bloodplasma from boron neutron capture therapy patients: the importanceof using coupled techniques. Anal Chem 74:5358–63.

Valliant JF, Guenther KJ, King AS, et al. (2002). The medicinalchemistry of carboranes. Coord Chem Rev 232:173–230.

Versluis AJ, Rump ET, Rensen PC, et al. (1999). Stable incorporation of alipophilic daunorubicin prodrug into apolipoprotein E-exposing lipo-somes induces uptake of prodrug via low-density lipoprotein receptorin vivo. J Pharmacol Exp Ther 289:1–7.

Wasan KM, Brazeau GA, Keyhani A, et al. (1993). Roles of liposomecomposition and temperature in distribution of amphotericin B inserum lipoproteins. Antimicrob Agents Chemother 37:246–50.

Wasan KM, Ramasway M, Cassidy SM, et al. (1997). Physicalcharacteristics and lipoprotein distribution of liposomal nystatin inhuman plasma. Antimicrob Agents Chemother 41:1871–5.

Zheng J, Shibata Y, Tanaka A. (2002). Study of the stability of seleniumcompounds in human urine and determination by mixed ion-pairreversed-phase chromatography with ICPMS detection. Anal BioanalChem 374:348–53.

58 F. K. Alanazi et al. J Liposome Res, 2014; 24(1): 53–58

Jour

nal o

f L

ipos

ome

Res

earc

h D

ownl

oade

d fr

om in

form

ahea

lthca

re.c

om b

y U

nive

rsity

of

Mel

bour

ne o

n 09

/15/

14Fo

r pe

rson

al u

se o

nly.