DANIEL ULTRASONIC METER T-2 TO T-11 TRANSDUCER UPGRADE INSTRUCTIONS __________________________________________

DANIEL MEASUREMENT AND CONTROL, INC.AN EMERSON PROCESS MANAGEMENT COMPANY

HOUSTON, TEXAS

Part Number 3-9000-751Revision A

DECEMBER 2005

Page 2: Daniel Ultrasonic Meter T-2 to T-11 Transducer Upgrade Instructions
Page 3: Daniel Ultrasonic Meter T-2 to T-11 Transducer Upgrade Instructions

IMPORTANT INSTRUCTIONS

Because these measurement instruments are sophisticated technical products, you must install, useand maintain them in accordance with Daniel’s guidelines to ensure that they operate within therange specified on the equipment nameplate. The following instructions must be adhered to andintegrated into your safety program when installing, using and maintaining Daniel Products.

• Read all instructions prior to installing, operating and servicing the product. If thisinstruction manual is not the correct manual, call 1-713-467-6000 (24-hour response numberfor both Service and Sales Support) and the requested manual will be provided. Save thisinstruction manual for future reference.

• If you do not understand any of the instructions, contact your Daniel representative forclarification.

• Follow all warnings, cautions and instructions marked on and supplied with the product.• Inform and educate your personnel in the proper installation, operation and maintenance of

the product.• Install your equipment as specified in the installation instructions of the appropriate

instruction manual and per applicable local and national codes. Connect all products to theproper electrical and pressure sources.

• To ensure proper performance, use qualified personnel to install, operate, update, programand maintain the product.

• When replacement parts are required, ensure that qualified people use replacement partsspecified by the manufacturer. Unauthorized parts and procedures can affect the product'sperformance and place the safe operation of your process at risk. Look-alike substitutionsmay result in fire, electrical hazards or improper operation.

• Ensure that all equipment doors are closed and protective covers are in place, except whenmaintenance is being performed by qualified persons, to prevent personal injury.

• ALWAYS READ AND FOLLOW THE DANIEL ULTRASONIC METER T-2 TO T-11 TRANSDUCER UPGRADE INSTRUCTIONS AND ALL PRODUCT WARNINGSAND INSTRUCTIONS.

• Use of this equipment for any purpose other than its intended purpose may result in propertydamage and/or serious personal injury or death.

• Before opening the flameproof enclosure in a flammable atmosphere, the electrical circuitsmust be interrupted.

Page 4: Daniel Ultrasonic Meter T-2 to T-11 Transducer Upgrade Instructions

This page intentionally left blank.

Page 5: Daniel Ultrasonic Meter T-2 to T-11 Transducer Upgrade Instructions

DANIEL T-2 TO T-11 TRANSDUCER UPGRADE INSTRUCTIONS DEC 2005

PREFACE i

DANIEL MEASUREMENT AND CONTROL, INC.DANIEL ULTRASONIC METER T-2 TO T-11TRANSDUCER UPGRADE INSTRUCTIONS

NOTICE

DANIEL MEASUREMENT AND CONTROL, INC. ("DANIEL") SHALL NOT BE LIABLE FOR TECHNICAL OR

EDITORIAL ERRORS IN THIS MANUAL OR OMISSIONS FROM THIS MANUAL. DANIEL MAKES NO

WARRANTIES, EXPRESSED OR IM PLIED, INCLUDING THE IMPLIED WARRANTIES OF

MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE WITH RESPECT TO THIS

MANUAL AND, IN NO EVENT, SHALL DANIEL BE LIABLE FOR ANY SPECIAL OR CONSEQUENTIAL

DAMAGES INCLUDING, BUT NOT LIMITED TO, LOSS OF PRODUCTION, LOSS OF PROFITS, ETC.

PRODUCT NAMES USED HEREIN ARE FOR MANUFACTURER OR SUPPLIER IDENTIFICATION ONLY

AND MAY BE TRADEM ARKS/REGISTERED TRADEM ARKS OF THESE COMPANIES.

THE CONTENTS OF THIS PUBLICATION ARE PRESENTED FOR INFORMATIONAL PURPOSES ONLY, AND

WHILE EVERY EFFORT HAS BEEN MADE TO ENSURE THEIR ACCURACY, THEY ARE NOT TO BE

CONSTRUED AS WARRANTIES OR GUARANTEES, EXPRESSED OR IMPLIED, REGARDING THE

PRODUCTS OR SERVICES DESCRIBED HEREIN OR THEIR USE OR APPLICABILITY. WE RESERVE THE

RIGHT TO MODIFY OR IMPROVE THE DESIGNS OR SPECIFICATIONS OF SUCH PRODUCTS AT ANY TIME.

DANIEL DOES NOT ASSUME RESPONSIBILITY FOR THE SELECTION, USE OR MAINTENANCE OF ANY

PRODUCT. RESPONSIBILITY FOR PROPER SELECTION, USE AND MAINTENANCE OF ANY DANIEL

PRODUCT REMAINS SOLELY WITH THE PURCHASER AND END-USER.

DANIEL AND THE DANIEL LOGO ARE REGISTERED TRADEMARKS OF DANIEL INDUSTRIES, INC. THE

EMERSON LOGO IS A TRADEMARK AND SERVICE MARK OF EMERSON ELECTRIC CO.

THE DANIEL ULTRASONIC GAS FLOWMETER INCLUDING THE MARK III ELECTRONICS IS COVERED

BY UNITED STATES PATENTS 5,983,730 AND 4,646,575 AND PATENTS PENDING.

COPYRIGHT © 2005BY DANIEL MEASUREMENT AND CONTROL, INC.

HOUSTON, TEXAS, U.S.A.

All rights reserved. No part of this work may be reproduced orcopied in any form or by any means - graphic, electronic ormechanical - without first receiving the written permission of

Daniel Measurement and Control, Inc., Houston, Texas, U.S.A.

Page 6: Daniel Ultrasonic Meter T-2 to T-11 Transducer Upgrade Instructions

DEC 2005 DANIEL T-2 TO T-11 TRANSDUCER UPGRADE INSTRUCTIONS

PREFACEii

WARRANTY

1. LIMITED WARRANTY: Subject to the limitations contained in Section 2 herein and except as otherwise expressly

provided herein, Daniel Measurement and Control, Inc. ("Daniel") warrants that the firmware will execute the

programming instructions provided by Daniel, and that the Goods manufactured or Services provided by Daniel will be

free from defects in materials or workmanship under normal use and care until the expiration of the applicable warranty

period. Goods are warranted for twelve (12) months from the date of initial installation or eighteen (18) months from

the date of shipment by Daniel, whichever period expires first. Consumables and Services are warranted for a period

of 90 days from the date of shipment or completion of the Services. Products purchased by Daniel from a third party for

resale to Buyer ("Resale Products") shall carry only the warranty extended by the original manufacturer. Buyer agrees

that Daniel has no liability for Resale Products beyond making a reasonable commercial effort to arrange for procurement

and shipping of the Resale Products. If Buyer discovers any warranty defects and notifies Daniel thereof in writing during

the applicable warranty period, Daniel shall, at its option, promptly correct any errors that are found by Daniel in the

firmware or Services, or repair or replace F.O.B. point of manufacture that portion of the Goods or firmware found by

Daniel to be defective, or refund the purchase price of the defective portion of the Goods/Services. All replacements or

repairs necessitated by inadequate maintenance, normal wear and usage, unsuitable power sources, unsuitable

environmental conditions, accident, misuse, improper installation, modification, repair, storage or handling, or any other

cause not the fault of Daniel are not covered by this limited warranty, and shall be at Buyer's expense. Daniel shall not

be obligated to pay any costs or charges incurred by Buyer or any other party except as may be agreed upon in writing

in advance by an authorized Daniel representative. All costs of dismantling, reinstallation and freight and the time and

expenses of Daniel's personnel for site travel and diagnosis under this warranty clause shall be borne by Buyer unless

accepted in writing by Daniel. Goods repaired and parts replaced during the warranty period shall be in warranty for the

remainder of the original warranty period or ninety (90) days, whichever is longer. This limited warranty is the only

warranty made by Daniel and can be amended only in a writing signed by an authorized representative of Daniel. Except

as otherwise expressly provided in the Agreement, THERE ARE NO REPRESENTATIONS OR WARRANTIES OF

ANY KIND, EXPRESSED OR IMPLIED, AS TO MERCHANTABILITY, FITNESS FOR PARTICULAR PURPOSE,

OR ANY OTHER MATTER WITH RESPECT TO ANY OF THE GOODS OR SERVICES. It is understood thatcorrosion or erosion of materials is not covered by our guarantee.

2. LIMITATION OF REMEDY AND LIABILITY: DANIEL SHALL NOT BE LIABLE FOR DAMAGES

CAUSED BY DELAY IN PERFORMANCE. THE SOLE AND EXCLUSIVE REMEDY FOR BREACH OF

WARRANTY HEREUNDER SHALL BE LIMITED TO REPAIR, CORRECTION, REPLACEMENT OR REFUND

OF PURCHASE PRICE UNDER THE LIMITED WARRANTY CLAUSE IN SECTION 1 HEREIN. IN NO EVENT,

REGARDLESS OF THE FORM OF THE CLAIM OR CAUSE OF ACTION (WHETHER BASED IN CONTRACT,

INFRINGEMENT, NEGLIGENCE, STRICT LIABILITY, OTHER TORT OR OTHERWISE), SHALL DANIEL'S

LIABILITY TO BUYER AND/OR ITS CUSTOMERS EXCEED THE PRICE TO BUYER OF THE SPECIFIC

GOODS MANUFACTURED OR SERVICES PROVIDED BY DANIEL GIVING RISE TO THE CLAIM OR CAUSE

OF ACTION. BUYER AGREES THAT IN NO EVENT SHALL DANIEL'S LIABILITY TO BUYER AND/OR ITS

CUSTOMERS EXTEND TO INCLUDE INCIDENTAL, CONSEQUENTIAL OR PUNITIVE DAMAGES. THE

TERM "CONSEQUENTIAL DAMAGES" SHALL INCLUDE, BUT NOT BE LIMITED TO, LOSS OF

ANTICIPATED PROFITS, LOSS OF USE, LOSS OF REVENUE AND COST OF CAPITAL.

Page 7: Daniel Ultrasonic Meter T-2 to T-11 Transducer Upgrade Instructions

DANIEL T-2 TO T-11 TRANSDUCER UPGRADE INSTRUCTIONS DEC 2005

TABLE OF CONTENTS iii

TABLE OF CONTENTS

1.0 INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-11.1 General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1

2.0 BEFORE YOU BEGIN! . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-12.1 Tools and Materials Required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1

3.0 KEY POINTS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1

4.0 TRANSDUCER UPGRADE PROCEDURE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1

APPENDICES

APPENDIX A REMOVAL AND RE-INSTALLATION OF WRITE-PROTECTION JUMPER ON MARK II ELECTRONICS . . . . . A-1

Page 8: Daniel Ultrasonic Meter T-2 to T-11 Transducer Upgrade Instructions

DEC 2005 DANIEL T-2 TO T-11 TRANSDUCER UPGRADE INSTRUCTIONS

TABLE OF CONTENTSiv

This page intentionally left blank.

Page 9: Daniel Ultrasonic Meter T-2 to T-11 Transducer Upgrade Instructions

DANIEL T-2 TO T-11 TRANSDUCER UPGRADE INSTRUCTIONS DEC 2005

INTRODUCTION 1-1

1.0 INTRODUCTION

1.1 General

Welcome to the Daniel Ultrasonic Meter T-2 to T-11 Transducer Upgrade Instructions Manual.

This manual has been designed to provide you with a step-by-step set of instructions for upgradingDaniel JuniorSonic™ and SeniorSonic™ meters from T-2 to T-11 transducers.

Please read the ‘Before You Begin’ section to make sure you have all the componentsnecessary to perform the upgrade before taking the JuniorSonic™ and SeniorSonic™ metersout of service.

Page 10: Daniel Ultrasonic Meter T-2 to T-11 Transducer Upgrade Instructions

DEC 2005 DANIEL T-2 TO T-11 TRANSDUCER UPGRADE INSTRUCTIONS

INTRODUCTION1-2

This page intentionally left blank.

Page 11: Daniel Ultrasonic Meter T-2 to T-11 Transducer Upgrade Instructions

DANIEL T-2 TO T-11 TRANSDUCER UPGRADE INSTRUCTIONS DEC 2005

BEFORE YOU BEGIN 2-1

2.0 BEFORE YOU BEGIN

This section lists all of the parts and tools required to perform the upgrade. Ensure that all of theseitems are available before taking the JuniorSonic™ and SeniorSonic™ meters out of service.

SERIOUS PERSONAL INJURY OR DEATH POSSIBLE

If the meter cannot be depressurized, use the ultrasonic extraction tool to remove and re-install the transducer assemblies.

Failure to follow the Ultrasonic Extractor Tool Operation Manual (3-9000-729) proceduresmay result in serious personal injury or death and may cause equipment damage.

If the meter cannot be depressurized, then the Daniel Ultrasonic Extractor Tool will berequired to remove the transducer assembly while the meter and associated piping remainsunder pressure. If you do NOT have this tool, it can be purchased or rented from DanielMeasurement Services.

2.1 Required Tools and Materials

1. 1¼" wrench or ratchet wrench c/w 1¼" socket2. 1/4" flat-blade screw driver3. 1/16" Allen wrench (2 or more recommended)4. Heat gun (type used for heat shrink tubing)5. Dow Corning® III silicone grease or equivalent6. Nickel anti-seize compound (P/N 3-9960-134)7. Sharpie® (or equivalent) permanent ink marker pen8. Laptop computer complete with Daniel CUI 3.10 or later installed 9. Zero Flow Calibration Data Sheet (transducer assembly dimensions)

Page 12: Daniel Ultrasonic Meter T-2 to T-11 Transducer Upgrade Instructions

DEC 2005 DANIEL T-2 TO T-11 TRANSDUCER UPGRADE INSTRUCTIONS

BEFORE YOU BEGIN2-2

This page intentionally left blank.

Page 13: Daniel Ultrasonic Meter T-2 to T-11 Transducer Upgrade Instructions

DANIEL T-2 TO T-11 TRANSDUCER UPGRADE INSTRUCTIONS DEC 2005

KEY POINTS 3-1

3.0 KEY POINTS

EQUIPMENT DAMAGE

DO NOT proceed with the transducer upgrade if the original component lengths are notavailable. Obtain the Zero Flow Calibration Data Sheet from Daniel. The originalcomponent lengths of all equipment being replaced must be known in order to calculatethe new path length for each chord.

Exchanging any component in the transducer assembly may alter the length of the chord.Document each component exchanged to ensure that the changes are correctly entered into themeter configuration. Failure to do so, or entry of erroneous data, may impact meter accuracy.

• It is recommended that Daniel CUI 3.10 or later be used for this procedure. Although olderversions will work, the latest version is always recommended.

All the screen images in this document are from Daniel CUI 3.10. Daniel CUI updates areavailable at the following website: http://www.emersonprocess.com/daniel/.

• Transducers are always replaced in pairs, and it is recommended they be installed one pairat a time to avoid accidental interchange of components, leading to erroneous dimensions.When replacing the transducers, it is important to note their position (upstream ordownstream) and replace them in the correct location to ensure proper low-velocityperformance. Each transducer pair is marked with either a number 1 or 2 that relates to themeter location (i.e. A1, A2, B1, B2, etc). The number “1” transducers must be installed ina location such as A1, B1, C1 or D1 and the “2” transducer must be installed in thecorresponding A2, B2, C2 or D2.

• The transducer kit, P/N 2-9-3400-847, includes four pairs of transducers and is designed forthe SeniorSonic™ meter. If a JuniorSonic™ meter is being upgraded, this kit can be used,and the remaining transducers can be kept as spares. If the spares are not needed, contactDaniel Measurement Services for a list of components and part numbers.

Page 14: Daniel Ultrasonic Meter T-2 to T-11 Transducer Upgrade Instructions

DEC 2005 DANIEL T-2 TO T-11 TRANSDUCER UPGRADE INSTRUCTIONS

KEY POINTS3-2

• Older transducer assemblies (transducer, stalk, and holder) did not have the lengths etchedon the individual component. New ones do. These lengths are required for the Daniel CUI3.10 software to calculate the new chord length.

To ensure that the required information is available, it is highly recommended that the originalZero Flow Calibration Data Sheet be obtained from Daniel prior to replacing the transducerpair.

The data sheet, shown in Figure 4-13, provides all of the serial numbers (S/N) and length datapertaining to the meter’s original transducer assembly components. This data sheet can be obtainedby contacting Daniel Measurement Services at 713-827-6314.

• Prior to replacing the transducers, it is imperative that each of the transducer pairs beinstalled in the correct location (i.e. A1, A2, etc) since changing the transducers, or anycomponents of the assembly, will impact the chord length. Mark the location on theTransducer Pair Data Sheet (see Figure 4-13, the data sheet included in the upgrade kit fornew transducers), and also mark the location on the side of the transducer with the markerpen (see Figure 4-5 for these components). This will prevent any confusion when thetransducers are installed and the meter’s configuration parameters are updated. The valueson the Transducer Pair Data Sheet relate to a unique pair of transducers. Do not mixtransducers from different pairs.

• It is recommended that only one pair (i.e. A1 & A2 or B1 & B2 etc.) of transducer assembliesbe removed at a time to avoid installing the transducer assemblies in the wrong position.When this isn’t practical, before removing the transducer assemblies from the mount, ensurethat the transducer location (i.e. A1, A2, etc) is marked on one of the flats of the 1¼" holderhead (see Figure 4-4). Ensure the marking will not be accidentally erased during removalof the old transducer.

• When the transducers are replaced, the applicable chord parameters must be updated. Usethe Transducer Swap-out Wizard in Daniel CUI (see Figures 4-6 through 4-9 for screen shotsof this wizard). This procedure requires writing changes to the meter. If the write-protectionjumper (JP2 on the DFI Board) is installed, it will need to be removed.

Page 15: Daniel Ultrasonic Meter T-2 to T-11 Transducer Upgrade Instructions

DANIEL T-2 TO T-11 TRANSDUCER UPGRADE INSTRUCTIONS DEC 2005

TRANSDUCER UPGRADE PROCEDURE 4-1

4.0 TRANSDUCER UPGRADE PROCEDURE

1. Prior to starting the transducer replacements, a minimum of a two-minute maintenance logfile should be collected.

2. Use the Edit/Compare Configuration Wizard to read the configuration from the meter.

3. Name the configuration file clearly to indicate it is before the transducer upgrade (by editingthe Daniel CUI automated file name shown in the Edit/Compare Configuration Wizardscreen (e.g. "Meter Name, Meter config, T-2 Transducers, 3-3-2005 4-48-58 PM.cfg").

4. When the upgrade procedure is complete and all of transducers are replaced and the dataupdated, save the configuration file named "Meter Name, Meter config, T-11 Transducers,3-3-2005 5-10-07 PM.cfg".

The chart in Figure 4-1 shows a typical SOS (Speed of Sound) comparison chart for a meter at zeroflow. It can be seen on the Charts sheet in the maintenance log file.

Figure 4-1. Typical SOS Comparison Chart

Page 16: Daniel Ultrasonic Meter T-2 to T-11 Transducer Upgrade Instructions

DEC 2005 DANIEL T-2 TO T-11 TRANSDUCER UPGRADE INSTRUCTIONS

TRANSDUCER UPGRADE PROCEDURE4-2

SERIOUS PERSONAL INJURY OR DEATH POSSIBLE

If the meter cannot be depressurized, use the Ultrasonic Extraction tool to remove andre-install the transducer assemblies.

Failure to follow the extractor tool manual procedures may result in serious personal injury ordeath and may cause equipment damage.

If the meter cannot be depressurized, refer to the Daniel Ultrasonic Extractor Tool OperationManual (P/N 3-9000-729). The most recent copy of this manual may be downloaded from the Danielwebsite (http://www.emersonprocess.com/daniel/). Follow the steps in Section 4 of the UltrasonicExtractor Tool Operation Manual for removing the transducers under pressure, instead of steps 5through 8 below.

5. Depressurize the meter and isolate the meter run piping, and vent the gas to atmosphere.Verify that the pressure is 0 psig before proceeding.

6. Prior to removing the holder cover, put an alignment mark on the edges of both the holdercover and the mount cover using the marker pen. This will help facilitate rotationalalignment during the re-assembly. Remove the holder cover (see Figure 4-2) by completelyloosening the two captive screws. Gently pull back the cover and cable, unplug the cable andexpose the mount cover assembly.

Figure 4-2. Side and End View of M-Mount Transducer Assembly

Page 17: Daniel Ultrasonic Meter T-2 to T-11 Transducer Upgrade Instructions

DANIEL T-2 TO T-11 TRANSDUCER UPGRADE INSTRUCTIONS DEC 2005

TRANSDUCER UPGRADE PROCEDURE 4-3

7. After labeling the mount cover location (A1, A2, etc.), loosen the two screws in the mountcover as shown in Figure 4-3, below. Note that these two screws may or may not be captive.If they are not captive, remove them carefully. With the two screws fully unscrewed, pull themount cover off of the hex head of the holder exposing the mount and holder assembly.

Figure 4-3. Mount Cover

8. Loosen the T-Slot transducer assembly holder with the 1¼" wrench or socket (see Figure 4-4). Once it has been loosened, it should be possible to unscrew it by hand.

Unless the meter/piping has a vent open to the atmosphere, pressure may build up in themeter/piping due to minor valve leakage.

Ensure there is no pressure buildup before completely unscrewing the holder. Carefully pullout the transducer assembly (Figure 4-5).

If the hex head of the holder does not clearly indicate the location of the transducer assembly(A1, A2, etc. as shown in Figure 4-4), mark it now with a Sharpie® marker pen (or equivalent)to ensure the assembly is re-installed in the correct location.

Page 18: Daniel Ultrasonic Meter T-2 to T-11 Transducer Upgrade Instructions

DEC 2005 DANIEL T-2 TO T-11 TRANSDUCER UPGRADE INSTRUCTIONS

TRANSDUCER UPGRADE PROCEDURE4-4

Figure 4-4. Mount and T-Slot Transducer Assembly

9. Loosen the three 6-32 set screws that hold the transducer in place. The screws do not needto be completely removed. On some of the older assemblies, Loc-Tite™ is used on the setscrews. If the set screw is very tight and does not appear to come out without damage to thewrench or the set screw, try heating the area around the set screw with a heat gun. Often thiswill help in releasing the bond and allows the screw to be removed without damage. Usecaution when heating the stalk area around the set screw. Overheating may cause damageto the stalk.

If the inside of the set screw is stripped due to Allen wrench slippage and the screws cannot beextracted, it may be necessary to loosen the set screws by holding the stalk in place. If this isthe case, a new stalk must be obtained and installed. The length of the replacement stalk willbe required to update the meter configuration.

Page 19: Daniel Ultrasonic Meter T-2 to T-11 Transducer Upgrade Instructions

DANIEL T-2 TO T-11 TRANSDUCER UPGRADE INSTRUCTIONS DEC 2005

TRANSDUCER UPGRADE PROCEDURE 4-5

Figure 4-5. Transducer Assembly

10. After all three set screws have been loosened, as shown in Figure 4-5, remove the transducerfrom the holder assembly.

11. Replace the O-rings and back-up rings on the transducer holder. It is recommended that theybe replaced while the transducer is removed from the holder/stalk to prevent damage to thetransducer.

12. Install the new transducer designated for that location. The parts are keyed so that they onlyfit together one way. Tighten the three set screws equally to lock the new transducer intoplace. DO NOT OVER TIGHTEN. Confirm the serial number of the transducer, andinstallation location (i.e. A1, A2, etc.) previously marked on the Transducer Pair Data Sheet.See annotated data sheet in Figure 4-13.

13. Clean the transducer ports, mounts, and T-Slot transducer before re-installing the newtransducer pair.

Page 20: Daniel Ultrasonic Meter T-2 to T-11 Transducer Upgrade Instructions

DEC 2005 DANIEL T-2 TO T-11 TRANSDUCER UPGRADE INSTRUCTIONS

TRANSDUCER UPGRADE PROCEDURE4-6

14. Lubricate the O-rings/back-up rings with a small amount of Dow Corning III silicon grease®

(or equivalent).

15. Apply a small amount of Nickel anti-seize compound (P/N 3-9960-134 or equivalent) to thethreads on the transducer holder.

SERIOUS PERSONAL INJURY OR DEATH POSSIBLE

If the extractor tool was used to remove the transducer assemblies while under pressure,use the tool to re-install the transducers. Follow the procedures in Section 5 of theUltrasonic Extractor Tool manual, instead of step 16 in this procedure. Once thetransducer has been installed, continue with step 17.

Failure to follow the extractor tool manual procedures may result in serious personal injury ordeath and may cause equipment damage.

16. Carefully insert the assembly into the assigned transducer port. With the 1¼" wrench orsocket, tighten the holder in the mount, until it is snug and securely seated (specificationtorque = 35 lb-ft). DO NOT OVER TIGHTEN.

17. Repeat the entire procedure for each of the remaining transducer pairs being replaced.

18. With the system sealed, apply minimal pressure to the meter. Using soapy water, or otherleak check method, check for leakage around the transducer mounts and holder. Fix any leakbefore proceeding. If no leaks are detected, fully pressurize the meter to line conditions (orat least 200 psi).

19. Replace the mount covers in their original locations. Position them so the cover slips overthe 1 ¼” hex head screw of the transducer holder and the screw slots on the side allow accessto the screw-hole in the transducer mount underneath. Observe the alignment marks appliedin step 6. The metal locking plate on the surface of the cover can be rotated approximately¼" to engage the plate on the 1¼" hex head of the holder. While holding the body of thecover, rotate the locking plate by pushing on the tabs with the thumb and finger whilepushing down (see Figure 4-3). Once the locking plate is engaged and the cover is fullyseated on the mount, install the two screws that attach the cover to the mount.

Page 21: Daniel Ultrasonic Meter T-2 to T-11 Transducer Upgrade Instructions

DANIEL T-2 TO T-11 TRANSDUCER UPGRADE INSTRUCTIONS DEC 2005

TRANSDUCER UPGRADE PROCEDURE 4-7

20. Align the cable connector plug with the transducer connector pins in the holder, and plug inthe cable. The internal connector is keyed to align with the holder. Observe the alignmentmarks applied in step 6 to align the screws in the holder cover with the holes in the mountcover. Also, note that there are small “V” shaped marks on the sides of the mount cover toalign it with the holder cover.

21. Once the transducers have been replaced, the meter configuration must be updated for thenew chord length, Delay times and Delta Delay times. This is most easily done using theTransducer Swap-out Wizard in Daniel CUI 3.10.

22. Prior to making any changes to the meter’s configuration, open the communicationsconnection to the meter and ensure that all the transducers are operating correctly. The SOSdifference graph will show different values, but all chords should be operational. If any chordhas failed, troubleshoot and repair before proceeding.

The following steps discuss configuration changes required to maintain the meter’s accuracy. If thejumpers (Mark II electronics) are configured for write-protection, they must be removed (seeAppendix A for detailed procedures).

23. If the configuration has not yet been read from the meter and saved, do so now. Figure 4-6shows an example of the Edit/Compare Configuration wizard.

Page 22: Daniel Ultrasonic Meter T-2 to T-11 Transducer Upgrade Instructions

DEC 2005 DANIEL T-2 TO T-11 TRANSDUCER UPGRADE INSTRUCTIONS

TRANSDUCER UPGRADE PROCEDURE4-8

Figure 4-6. Edit/Compare Configuration Screen

24. Use the Daniel CUI Tools menu path and select Transducer Swap-out. The Test Meter dialogbox shown in Figure 4-7 displays. Select all the chords by clicking the check box beside thechord name in the lower left-hand corner. A check mark appears in the box as shown in thefigure.

25. Click the Next button to continue the configuration process (see Figure 4-8).

Page 23: Daniel Ultrasonic Meter T-2 to T-11 Transducer Upgrade Instructions

DANIEL T-2 TO T-11 TRANSDUCER UPGRADE INSTRUCTIONS DEC 2005

TRANSDUCER UPGRADE PROCEDURE 4-9

Figure 4-7. Chord Parameters Adjustment Dialog

26. From this dialog (Figure 4-8) select the components that have been replaced by clicking theradio buttons beside the applicable component. Note that only one chord will be entered ata time. The example in Figure 4-8 shows that only the transducer on Chord A was replaced(chord indicated at the top of the Select Components dialog).

Page 24: Daniel Ultrasonic Meter T-2 to T-11 Transducer Upgrade Instructions

DEC 2005 DANIEL T-2 TO T-11 TRANSDUCER UPGRADE INSTRUCTIONS

TRANSDUCER UPGRADE PROCEDURE4-10

Figure 4-8. Select Components for Chord A Dialog

27. Click the Next button to display the Select Components for Chord A data entry dialog (see

Figure 4-9). Enter the lengths of the “Components Removed” (from original Zero FlowCalibration Data Sheet), and the “Components Added”. The Daniel CUI softwareautomatically calculates the new chord length based on the data input.

28. Next enter the Delay and Delta times for the new transducer pair (from data sheet sent witheach transducer pair).

Page 25: Daniel Ultrasonic Meter T-2 to T-11 Transducer Upgrade Instructions

DANIEL T-2 TO T-11 TRANSDUCER UPGRADE INSTRUCTIONS DEC 2005

TRANSDUCER UPGRADE PROCEDURE 4-11

Figure 4-9. Update Chord A Data Entry Dialog

29. Click the Write to Meter button to apply the new values to the meter configuration.

30. Repeat this process for each of the remaining chords. Again, this process only permitschanges to the component information one chord at a time.

31. Read the configuration from the meter. As suggested in step 1, give this file a name whichclearly indicates that it is the configuration after the transducers have been replaced (i.e."Meter Name, Meter config, T-11 Transducers, 3-3-2005 5-10-07 PM.cfg").

Page 26: Daniel Ultrasonic Meter T-2 to T-11 Transducer Upgrade Instructions

DEC 2005 DANIEL T-2 TO T-11 TRANSDUCER UPGRADE INSTRUCTIONS

TRANSDUCER UPGRADE PROCEDURE4-12

32. Perform a file comparison to verify the changes to the meter’s configuration. To perform thiscomparison, use the Tools>Edit/Compare Configuration menu path.

(a) In the File section at the bottom of the screen click the Open button. Select the

directory and file folder in which the original configuration (created in step 1) was saved.

(b) Open the configuration file. The Daniel CUI configuration window displays theconfiguration.

(c) Next, click the Compare button at the bottom of the screen. Select the directory and file

folder in which the new configuration (created in step 31) was saved.

(d) Open the configuration file. The Compare Configurations window displays. An exampleof this comparison is shown in Figure 4-10. Only the "L" dimensions (Lx, Registers 312- 318), Delay (AvgDlyx, Registers 332 - 338) and Delta times (Dltdlyx, Registers 340 -346) should be different.

Figure 4-10. Compare Configurations Window

(e) Collect a maintenance log file for at least two minutes while the meter is flowing fasterthan 3 ft./sec. Verify that the SOS of each chord is within 2 ft./sec. of the others. If a SOSwas calculated using AGA10 and a known gas composition, verify that the meter averageSOS is within 2 ft./sec. of the AGA10 SOS.

Page 27: Daniel Ultrasonic Meter T-2 to T-11 Transducer Upgrade Instructions

DANIEL T-2 TO T-11 TRANSDUCER UPGRADE INSTRUCTIONS DEC 2005

TRANSDUCER UPGRADE PROCEDURE 4-13

Figure 4-11. SOS Comparison after Transducer Replacement

(f) If the security jumpers were removed, power down the meter and re-install the jumpersto implement the write-protection. When complete, restore power to the meter and checkthe meter to ensure proper operation using Daniel CUI.

(g) Disconnect the communication to the meter via Daniel CUI and install the end covers onthe meter head.

Page 28: Daniel Ultrasonic Meter T-2 to T-11 Transducer Upgrade Instructions

DEC 2005 DANIEL T-2 TO T-11 TRANSDUCER UPGRADE INSTRUCTIONS

TRANSDUCER UPGRADE PROCEDURE4-14

Figure 12. Ultrasonic Flow Meter Zero Flow Calibration Report

Page 29: Daniel Ultrasonic Meter T-2 to T-11 Transducer Upgrade Instructions

DANIEL T-2 TO T-11 TRANSDUCER UPGRADE INSTRUCTIONS DEC 2005

TRANSDUCER UPGRADE PROCEDURE 4-15

Figure 4-13. Transducer Pair Data Sheet

Page 30: Daniel Ultrasonic Meter T-2 to T-11 Transducer Upgrade Instructions

DEC 2005 DANIEL T-2 TO T-11 TRANSDUCER UPGRADE INSTRUCTIONS

TRANSDUCER UPGRADE PROCEDURE4-16

This page intentionally left blank.

Page 31: Daniel Ultrasonic Meter T-2 to T-11 Transducer Upgrade Instructions

DANIEL T-2 TO T-11 TRANSDUCER UPGRADE INSTRUCTIONS DEC 2005

APPENDIX A A-1

A.1 REMOVAL AND RE-INSTALLATION OF WRITE-PROTECTION JUMPER ONMARK II ELECTRONICS

If the jumper has been installed on JP2 on the DFI Board, then the meter configuration will be write-protected. The effect of this jumper is to prevent changes to the values in the meter’s configuration.When the transducers or any component of the transducer assembly are replaced, the values relatedto the chord length “L”, the delay and the delta delay times will need to be revised in order tomaintain the meter’s accuracy. It is then necessary to remove the jumper, change the applicableconfiguration values and re-install the jumper.

Page 32: Daniel Ultrasonic Meter T-2 to T-11 Transducer Upgrade Instructions

DEC 2005 DANIEL T-2 TO T-11 TRANSDUCER UPGRADE INSTRUCTIONS

APPENDIX AA-2

Figure A-1. Mark II Electronics

Page 33: Daniel Ultrasonic Meter T-2 to T-11 Transducer Upgrade Instructions

DANIEL T-2 TO T-11 TRANSDUCER UPGRADE INSTRUCTIONS DEC 2005

APPENDIX A A-3

Figure A-2. View of Mark II Electronics Board Stack from End of Enclosure

1. From Daniel CUI use the Meter pull-down menu and select Meter Information. The MeterInformation dialog box appears (see Figure A-2). The S/W Ver. shows the firmware versioninstalled in the meter electronics. If the version is 3.34 or earlier, then jumpers JP2 on theDFI Board and JP9 on the CPU must be removed in order to modify the meter’sconfiguration. If the firmware version is 3.62 or later, only jumper JP2 must be removed toapply the changes. Once the firmware version has been determined, click the Close button

and exit Daniel CUI..

Page 34: Daniel Ultrasonic Meter T-2 to T-11 Transducer Upgrade Instructions

DEC 2005 DANIEL T-2 TO T-11 TRANSDUCER UPGRADE INSTRUCTIONS

APPENDIX AA-4

Figure A-3. Daniel CUI Meter Information Dialog

2. Turn off power to the electronics.

3. Disconnect the cable originating from the Field Connection Board by unplugging theconnector on the Power Supply Board (see Figure A-3). Loosen the screws completely tounplug the connector.

4. Completely loosen the four captive screws that hold the Electronics Board stack in theenclosure. With the cable held to the side, grasp the screws and pull the assembly straight outof the enclosure. The three-board assembly consisting of the Power Supply Board, the DFIBoard and the CPU Board are removed as a single unit.

Page 35: Daniel Ultrasonic Meter T-2 to T-11 Transducer Upgrade Instructions

DANIEL T-2 TO T-11 TRANSDUCER UPGRADE INSTRUCTIONS DEC 2005

APPENDIX A A-5

5. As shown in Figure A-4, jumper JP2 is located near the outer edge of the DFI Board. Accessto the jumper is possible without disassembling the three boards. Using needle nose pliers,carefully lift the jumper from the contact pins. One side of the jumper can be installed onone of the pins so that the jumper is not lost. If the firmware version is 3.62 or later, then theelectronics can be re-installed so that the configuration can be changed. If the firmwareversion is 3.34 or earlier, then jumper JP9 on the CPU Board (see Figure A-5) will also needto be removed. As with jumper JP2 on the DFI Board, jumper JP9 on the CPU Board canalso be carefully removed by using a pair of needle nose pliers.

Figure A-4. DFI Board Showing Jumper JP2

Page 36: Daniel Ultrasonic Meter T-2 to T-11 Transducer Upgrade Instructions

DEC 2005 DANIEL T-2 TO T-11 TRANSDUCER UPGRADE INSTRUCTIONS

APPENDIX AA-6

Figure A-5. CPU Board Showing Jumper JP9

6. Align the electronics assembly and re-install it in the enclosure, ensuring that the connectionbetween the electronics and the enclosure is correctly engaged. Tighten the four captivescrews and reconnect the cable from the Field Connection Board.

7. Power up the electronics. With the write-protection jumper removed, it is now possible tochange values, which influence the measurement function, in the meter configuration.

8. After the values in the configuration have been changed, and the meter is operating withinthe expected tolerances, re-install jumper JP2 on the DFI Board, and jumper JP9 on the CPUBoard (if removed) using the same procedure as outlined above, except that the jumper needsto be re-installed to the position it was originally found, bridging both jumper contact pins.

Page 37: Daniel Ultrasonic Meter T-2 to T-11 Transducer Upgrade Instructions

NOTES

Page 38: Daniel Ultrasonic Meter T-2 to T-11 Transducer Upgrade Instructions

NOTES

Page 39: Daniel Ultrasonic Meter T-2 to T-11 Transducer Upgrade Instructions

WARRANTY CLAIM PROCEDURES To make a warranty claim, you, the Purchaser, must: 1. Provide Daniel with proof of the Date of Purchase and proof of the Date of Shipment of the

product in question.

2. Return the product to Daniel within twelve (12) months of the date of original shipment ofthe product, or within eighteen (18) months of the date of original shipment of the productto destinations outside of the United States. The Purchaser must prepay any shipping charges.In addition, the Purchaser is responsible for insuring any product shipped for return, andassumes the risk of loss of the product during shipment.

3. To obtain Warranty service or to locate the nearest Daniel office, sales, or service center call

(713) 827-6314, Fax (713) 827-6312 write to:

Daniel Measurement Services11100 Brittmoore Park DriveHouston, Texas 77041

Or contact Daniel via the following site:

www.emersonprocess.com/daniel

4. When contacting Daniel for product service, the Purchaser is asked to provide informationas indicated on the following page entitled "Customer Repair Report".

5. For product returns from locations outside the United States, it will be necessary for you toobtain the import consignment address so that Daniel’s customs broker can handle theimportation with the U.S. Customs Service.

6. Daniel Measurement Services offers both on call and contract maintenance service designedto afford single source responsibility for all its products.

7. Daniel reserves the right to make changes at any time to any product to improve its design

and to insure the best available product.

Page 40: Daniel Ultrasonic Meter T-2 to T-11 Transducer Upgrade Instructions
Page 41: Daniel Ultrasonic Meter T-2 to T-11 Transducer Upgrade Instructions

DANIEL MEASUREMENT AND CONTROL, INC.CUSTOMER REPAIR REPORT

FOR SERVICE, COMPLETE THIS FORM, AND RETURN IT ALONG WITH THE AFFECTED EQUIPMENT TO CUSTOMERSERVICE AT THE ADDRESS INDICATED BELOW.

COMPANY NAME:

TECHNICAL CONTACT: PHONE:

REPAIR P. O. #: IF WARRANTY, UNIT S/N:

INVOICE ADDRESS:

SHIPPING ADDRESS:

RETURN SHIPPING METHOD:

EQUIPMENT MODEL #: S/N: FAILURE DATE:

DESCRIPTION OF PROBLEM:

WHAT WAS HAPPENING AT TIME OF FAILURE?

ADDITIONAL COMMENTS:

REPORT PREPARED BY: TITLE:

IF YOU REQUIRE TECHNICAL ASSISTANCE, PLEASE FAX OR WRITE THE CUSTOMER SERVICE DEPARTMENT AT:

DANIEL FIELD SERVICES PHONE: (713) 827-6314ATTN: CUSTOMER SERVICE FAX: (713) 827-631211100 BRITTMOORE PARK DRIVEHOUSTON, TEXAS 77041

FOR FASTEST SERVICE CONTACT DANIEL VIA OUR WEBSITE:www.emersonprocess.com/daniel

Page 42: Daniel Ultrasonic Meter T-2 to T-11 Transducer Upgrade Instructions
Page 43: Daniel Ultrasonic Meter T-2 to T-11 Transducer Upgrade Instructions
Page 44: Daniel Ultrasonic Meter T-2 to T-11 Transducer Upgrade Instructions

The sales and service offices of Daniel Measurement and Control are locatedthroughout the United States and in major countries overseas.

Please contact Daniel Measurement Services at11100 Brittmoore Park Drive, Houston, Texas 77041, or phone (713) 467-6000

for the location of the sales or service office nearest you. Daniel Measurement Services offers both on-call and contract

maintenance service designed to provide single-source responsibility for all Daniel products.

Daniel Measurement and Control, Inc., and Daniel Measurement Services, Inc.Divisions of Emerson Process Management reserves the right to make changes to any of its products or services

at any time without prior notification in order to improve that product or service and to supplythe best product or service possible.www.emersonprocess.com/daniel