20
The phantom portion of the American College of Radiology ACRComputed Tomography CTaccreditation program: Practical tips, artifact examples, and pitfalls to avoid aCynthia H. McCollough b) and Michael R. Bruesewitz Department of Radiology, Mayo Clinic College of Medicine, Rochester, Minnesota 55905 Michael F. McNitt-Gray Department of Radiological Sciences, UCLA School of Medicine, Los Angeles, California 90095 Krista Bush Accreditation and Standards, American College of Radiology, Reston, Virginia 20191 Thomas Ruckdeschel Alliance Medical Physics, Alpharetta, Georgia 30004 J. Thomas Payne Radiation Oncology Department, Abbott Northwestern Hospital, Minneapolis, Minnesota 55407 James A. Brink Yale-New Haven Hospital, New Haven, Connecticut 06520 Robert K. Zeman George Washington University Medical Center, Washington, DC 20037 ~Received 13 March 2004; revised 17 May 2004; accepted for publication 18 May 2004; published 19 August 2004! The ACR CT accreditation program, begun in 2002, requires the submission of approximately 20 images, several completed data sheets and printouts of three Excel worksheets. The procedure manual is very detailed, yet participants unfamiliar with the program or having minimal CT expe- rience have needed to redo aspects of their submission, or in some cases do not receive accredita- tion, due to mistakes made by the physicist. This review of the phantom portion of the ACR CT accreditation program supplements the ACR provided instructions with additional photos of phan- tom setup, region-of-interest ~ROI!, and image placement on the film sheets, and examples of completed portions of actual ~but anonymous! submissions. Common mistakes, as well as uncom- mon but interesting images, are shown and explanations are given as to what could have been done to avoid the problem. Additionally, a review of CT dose measurement techniques and calculations will enable the physicist to better assist sites where typical exam doses are above the ACR reference values. © 2004 American Association of Physicists in Medicine. @DOI: 10.1118/1.1769632# Key words: ACR accreditation, ACR CT phantom, CT image quality, CT radiation dosimetry, CTDI I. INTRODUCTION The ACR CT accreditation program, begun in 2002, requires the submission of approximately 20 images, several com- pleted data sheets, and printouts of three Excel worksheets. As part of the accreditation program materials, a detailed procedure manual is provided to assist the medical physicist in performing the required scanning, measurements, calcula- tions, and documentation ~through submission of films!. Still, some participants have needed to redo aspects of their submission, or in some cases do not receive accreditation, due to mistakes made in the submitted materials In the spirit of the ACR accreditation program, which is meant to be primarily educational in nature, this review supplements the ACR-provided instructions with additional photos of phantom setup, region-of-interest ~ROI! place- ment, and examples from actual ~but anonymous! submis- sions. Common mistakes, as well as uncommon but interest- ing images, are included and explanations given as to the source of the problem. Finally, we review the CT dose measurement techniques and calculations to enable the physicist to better assist sites whose typical exam doses are above the ACR reference val- ues. This review is organized by the various performance tests included in the submission process. For all tests, an example of an acceptable test result is included. II. ACR CT ACCREDITATION PHANTOM The ACR CT accreditation phantom has been designed to examine a broad range of image quality parameters. These include: • positioning accuracy; • CT No. accuracy; • slice width; 2423 2423 Med. Phys. 31 9, September 2004 0094-2405Õ2004Õ319Õ2423Õ20Õ$22.00 © 2004 Am. Assoc. Phys. Med.

The phantom portion of the American College of Radiology ACR … · 2005-05-09 · The phantom portion of the American College of Radiology —ACR– Computed Tomography —CT–accreditation

  • Upload
    others

  • View
    3

  • Download
    0

Embed Size (px)

Citation preview

Page 1: The phantom portion of the American College of Radiology ACR … · 2005-05-09 · The phantom portion of the American College of Radiology —ACR– Computed Tomography —CT–accreditation

The phantom portion of the American College of Radiology „ACR…

Computed Tomography „CT… accreditation program: Practical tips, artifactexamples, and pitfalls to avoid a…

Cynthia H. McColloughb) and Michael R. BruesewitzDepartment of Radiology, Mayo Clinic College of Medicine, Rochester, Minnesota 55905

Michael F. McNitt-GrayDepartment of Radiological Sciences, UCLA School of Medicine, Los Angeles, California 90095

Krista BushAccreditation and Standards, American College of Radiology, Reston, Virginia 20191

Thomas RuckdeschelAlliance Medical Physics, Alpharetta, Georgia 30004

J. Thomas PayneRadiation Oncology Department, Abbott Northwestern Hospital, Minneapolis, Minnesota 55407

James A. BrinkYale-New Haven Hospital, New Haven, Connecticut 06520

Robert K. ZemanGeorge Washington University Medical Center, Washington, DC 20037

~Received 13 March 2004; revised 17 May 2004; accepted for publication 18 May 2004;published 19 August 2004!

The ACR CT accreditation program, begun in 2002, requires the submission of approximately 20images, several completed data sheets and printouts of three Excel worksheets. The proceduremanual is very detailed, yet participants unfamiliar with the program or having minimal CT expe-rience have needed to redo aspects of their submission, or in some cases do not receive accredita-tion, due to mistakes made by the physicist. This review of the phantom portion of the ACR CTaccreditation program supplements the ACR provided instructions with additional photos of phan-tom setup, region-of-interest~ROI!, and image placement on the film sheets, and examples ofcompleted portions of actual~but anonymous! submissions. Common mistakes, as well as uncom-mon but interesting images, are shown and explanations are given as to what could have been doneto avoid the problem. Additionally, a review of CT dose measurement techniques and calculationswill enable the physicist to better assist sites where typical exam doses are above the ACR referencevalues. © 2004 American Association of Physicists in Medicine.@DOI: 10.1118/1.1769632#

Key words: ACR accreditation, ACR CT phantom, CT image quality, CT radiation dosimetry,CTDI

ireme

ileicu

thtio

isw

na

e

the

uesitesval-

stsple

d toese

I. INTRODUCTION

The ACR CT accreditation program, begun in 2002, requthe submission of approximately 20 images, several copleted data sheets, and printouts of three Excel worksheAs part of the accreditation program materials, a detaprocedure manual is provided to assist the medical physin performing the required scanning, measurements, calctions, and documentation~through submission of films!.Still, some participants have needed to redo aspects ofsubmission, or in some cases do not receive accreditadue to mistakes made in the submitted materials

In the spirit of the ACR accreditation program, whichmeant to be primarily educational in nature, this reviesupplements the ACR-provided instructions with additiophotos of phantom setup, region-of-interest~ROI! place-ment, and examples from actual~but anonymous! submis-sions. Common mistakes, as well as uncommon but inter

2423 Med. Phys. 31 „9…, September 2004 0094-2405 Õ2004Õ3

s-

ts.distla-

eirn,

l

st-

ing images, are included and explanations given as tosource of the problem.

Finally, we review the CT dose measurement techniqand calculations to enable the physicist to better assist swhose typical exam doses are above the ACR referenceues.

This review is organized by the various performance teincluded in the submission process. For all tests, an examof an acceptable test result is included.

II. ACR CT ACCREDITATION PHANTOM

The ACR CT accreditation phantom has been designeexamine a broad range of image quality parameters. Thinclude:

• positioning accuracy;

• CT No. accuracy;

• slice width;

24231„9…Õ2423Õ20Õ$22.00 © 2004 Am. Assoc. Phys. Med.

Page 2: The phantom portion of the American College of Radiology ACR … · 2005-05-09 · The phantom portion of the American College of Radiology —ACR– Computed Tomography —CT–accreditation

the central

2424 McCollough et al. : Phantom portion of the ACR CT accreditation program 2424

FIG. 1. ~a! Diagram of the four modules of the ACR CT accreditation phantom.~b! Photograph of a properly aligned phantom.~c! Centering left to right onboth the phantom and the phantom base~optional! substantially simplifies the alignment process.~d! Top view of a properly aligned phantom.~e! Side viewof a properly aligned phantom. The white Teflon rings on the optional phantom base may cause streak artifacts and should be moved away fromportions of each module, which are indicated by the four white lines circumscribing the phantom.

-er

rk-in

up-ent.

• low contrast resolution;

• high contrast~spatial! resolution;

• CT number uniformity; and

• image noise.

The ACR CT accreditation phantom~model 464,Gammex-RMI, Middleton, WI! is a solid phantom containing four modules, and is constructed primarily from a wat

Medical Physics, Vol. 31, No. 9, September 2004

-

equivalent material@Fig. 1~a!#. Each module is 4 cm in depthand 20 cm in diameter. There are external alignment maings on EACH module to allow centering of the phantomthe axial (z axis, cranial/caudal!, coronal (y axis, anterior/posterior!, and sagittal (x axis, left/right! directions. Thereare also ‘‘HEAD,’’ ‘‘FOOT,’’ and ‘‘TOP’’ markings on thephantom to assist with alignment. An optional phantom sport base may be purchased to assist in phantom alignm

Page 3: The phantom portion of the American College of Radiology ACR … · 2005-05-09 · The phantom portion of the American College of Radiology —ACR– Computed Tomography —CT–accreditation

s

2425 McCollough et al. : Phantom portion of the ACR CT accreditation program 2425

TABLE I. Typical image acquisition technical parameters.

Adult head~cerebrum portion!

High resolutionchest Adult abdomen

Pediatricabdomen~5 y.o.!

kVp 120 140 120 120mA 170 120 240 110Time per rotation~s! 2 0.8 0.8 0.8Scan FOV~cm or name!a Head~25 cm! Large ~50 cm! Large ~50 cm! Large ~50 cm!Display FOV~cm!a 25 38 38 30Reconstruction algorithm Std Bone Std StdAxial (A) or helical (H) A A H HZ-axis collimation (T, in mm!b 2.5 1.25 3.75 2.5No. of data channelsusedb (N) 4 1 4 4A: Table increment~mm! orH: Table speed~mm/rot! (I )b

10 10 11.25 15

Pitchc¯ ¯ 0.75 ~HQ! 1.5 ~HS!

Reconstructed scan width~mm! 5 1.25 5 5Reconstructed scan interval~mm! 5 10 5 5Dose reduction technique~s!d

¯ ¯ mA modulation mA modulation

aFOV5field of viewbz-axis collimation(T)5the width of the tomographic section along thez-axis imaged by one data channel. In multidetector row~multislice! CT scanners,several detector elements may be grouped together to form one data channel.Number of data channels(N)5the number of tomographic sections imaged in a single axial scan.Maximum number of data channels(Nmax)5the maximum number of tomographic sections for a single axial scan.Increment(I )5the table increment per axial scan or the table increment per rotation of the x-ray tube in a helical scan.

cApplied to helical scanning. Compute pitch according to the IEC definition: Pitch5table speed (mm/rotation)/N•T (mm)5I /N•T. For some scanners, thicomputed value might differ from the value given by the manufacturer.

dIf selectable, list selection, otherwise leave blank. Example: mA modulation based on patient attenuation.

ulnicurly

heheiof te

anitirt

o-

o

e,

ofthe

an-repa-

--eterctorisur-

hef

l-

ch.sis-’s

The Appendix contains a brief description of each modcontained within the ACR CT accreditation phantom, aincludes a cross-sectional diagram of each module whshows the various test objects contained within each modFigures 1~b!–1~e! show a phantom that has been propealigned with the scanner patient positioning lights.

A. Table I of site scanning data form

A critical element of the submission process is tcompletion of Table I of the site scanning data form—tdescription of the site’s scanning protocols. The acquisitparameters reported in Table I should be representative oscan protocols used for the site’s routine clinical scans. Thparameters will be used for scanning the image qualitydose phantoms and should be consistent with the acquisparameters used for the clinical exams submitted as pathe accreditation process.

Table I refers to the following routine examinations.

~1! Adult head: Head CT~for headaches or to exclude neplasm, brain CT, top of the head!.

~2! High resolution chest: High resolution chest~HRC! CTfor evaluation of diffuse lung disease.

~3! Adult abdomen: Abdomen CT~for detection of possibleliver metastases or lymphoma!.

~4! Pediatric abdomen: Pediatric abdomen CT~for blunttrauma, acute abdominal pain, or infection!. Assume a5-year-old, patient.

The following example shows a correct completionTable I for a four-row multidetector row scanner~Light

Medical Physics, Vol. 31, No. 9, September 2004

edh

le.

nhesed

onof

f

Speed QX/i, General Electric Medical Systems, MilwaukeWI!.

1. Common Table I errors

The most common error observed in the completionTable I occurs for multidetector-row CT systems, whereZ-axis collimation (T, as defined earlier! is often confusedwith the reconstructed scan width. For single-slice CT scners, theZ-axis collimation and reconstructed slice width athe same. However, with multislice CT scanners, theserameters often have different values. TheZ-axis collimation~also referred to as detector collimation! is chosen prospectively ~prior to scan acquisition!. It determines the appropriate x-ray beam collimation and, hence, is a critical paramfor dose measurements. For the example of the four-deterow CT system described in Table I, a 5-mm-thick imagereconstructed from the projection data acquired by the fodetector rows, each one of those rows having aZ-axis widthof 3.75 mm. Hence, many physicists incorrectly report tdetector collimation as 5 mm~instead of the correct value o3.75 mm!.

Similarly, the number of data channels used (N, as de-fined earlier! is an important acquisition parameter. Knowedge of bothT andN ~and table increment per rotation,I ) isrequired to accurately compute the correct value of pit~The value of pitch reported on some systems is not content with the International Electrotechnical Commissiondefinition, which is used in the ACR program.! However, the

Page 4: The phantom portion of the American College of Radiology ACR … · 2005-05-09 · The phantom portion of the American College of Radiology —ACR– Computed Tomography —CT–accreditation

tifddctlic

lu-b

ro

wuler

tha

ysac

le

t i

geC

esn

inlin

nie

i-

No

e

2426 McCollough et al. : Phantom portion of the ACR CT accreditation program 2426

user interfaces for many CT systems do not clearly identhe values of these important acquisition parameters. Ationally, some 16-detector-row scanners use only 12-deterows for certain acquisition modes, yet the user is not expitly notified whetherN512 or 16 for a given clinical proto-col. Hence, serious confusion can exist as to the exact vaof T, N, and I for clinical exam protocols, which are typically prescribed in terms of reconstructed scan width, tatravel speed~mm/rotation!, and for some systems, pitch.

To assist users in determining the correct values ofT, N,and I , information regarding specific scanner models is pvided on the frequently asked questions~FAQ! page of theACR accreditation web site: www.acr.org.

2. Essential criteria for Table I

• Detector configuration must be correctly reported.

• Pitch must be correctly reported.

• Other scan parameters must appear correct.

B. SMPTE test pattern

Prior to filming any images, a SMPTE test pattern~Soci-ety of Motion Picture and Television Engineers! should beprinted using the appropriate window width and windolevel. Users who are unfamiliar with this procedure, shoreview Grayet al.1 and contact their local service enginefor assistance.

A SMPTE test pattern is REQUIRED to demonstratequality of the submitted hard-copy film. Failure to submitSMPTE test pattern results in automatic failure of the phics portion of the accreditation. Alternate patterns arecepted, but not recommended.

Figures 2~a!–2~b! show an example of an acceptabSMPTE image and an image with various deficiencies.

1. Essential criteria for SMPTE test pattern

• SMPTE or other video test pattern must be presenthe first box of each film sheet.

• The 95% square must be visible~whites must not besaturated!.

• The 5% square must be visible~blacks must not besaturated!.

• No aliasing of bar patterns or other artifacts.

C. Phantom and scanner alignment—Modules 1 and 4

The ACR CT accreditation phantom is used for all imaquality scans and must first be positioned carefully on thetable. The HRC protocol from Table I must be used. Eventhe site’s HRC protocol does not use a scan width of lthan 2 mm, it is essential to use a scan width of less thamm for phantom alignment. In the situation where testhas begun and the physicist must yield the scanner for ccal use, the phantom must be realigned~modules 1 and 4!and new alignment images filmed before proceeding.

Figures 3~a!–3~d! show an example of acceptable phatom alignment and several images with various deficienc

Medical Physics, Vol. 31, No. 9, September 2004

yi-or-

es

le

-

d

e

--

n

Tifs2

gi-

-s.

1. Essential criteria for phantom alignment

• Image thickness must be 2 mm or less.

• All four BBs must be seen in one image and have simlar appearance.

• Longest wire must be centrally located (61 wire! onboth top and bottom ramp.

FIG. 2. ~a! Good SMPTE image. Neither blacks nor whites are saturated.artifacts. ~b! Unacceptable SMPTE image. The whites are saturated~notethat the 95% square is not visible! and numerous horizontal line artifacts arpresent.

Page 5: The phantom portion of the American College of Radiology ACR … · 2005-05-09 · The phantom portion of the American College of Radiology —ACR– Computed Tomography —CT–accreditation

on ringsmp buttop

precisiot in v

2427 McCollough et al. : Phantom portion of the ACR CT accreditation program 2427

FIG. 3. ~a! Excellent alignment. Longer central wires are in the center of both the top and bottom ramps. All four BBs are equally bright. Note the Teflfrom the phantom stand.~b! Poor alignment. Text annotation covers some BBs, making them difficult to see. The central wire is visible on the top ranot on the bottom ramp. Also, image orientation~left/right! is wrong.~c! Poor alignment. The central wire is visible on the bottom ramp but not on theramp. Top and left BBs are not as bright as the right BB. Annotation covers the bottom BB. The gray ring partially around the location of the highnsolid water rod is from the phantom manufacturing process and is not a scanner artifact. This ring is not considered to be problematic and is presenaryingdegrees on most of the phantoms produced.~d! FOV too large: Alignment for module 4 is fine, but the reconstruction field of view~FOV! is too large~shouldbe 21–25 cm!.

ixebegeir

haan

TheCTomof

ugh

ng.as

D. Module 1—CT number calibration

CT scanners are quantitative devices where pbrightness should accurately reflect the atomic numand physical density of the material within an imavoxel. By defini-tion, the CT number of water and aare assigned values of 0 and21000, respectively. TheCT numbers for all other materials can vary somewdepending on the system’s x-ray beam spectra

Medical Physics, Vol. 31, No. 9, September 2004

lr

td

other issues such asbeam hardening and scatter.various phantom materials have been assignednumbers consistent with average values obtained frmultiple scanner models. A measured value that is outthe recommended range is not a major deficiency, althothe CT number of water is expected to be 067 HUfor images acquired at any slice thickness or kVp settiThe high-precision solid water rod in module 1 h

Page 6: The phantom portion of the American College of Radiology ACR … · 2005-05-09 · The phantom portion of the American College of Radiology —ACR– Computed Tomography —CT–accreditation

from theers forkVp,

2428 McCollough et al. : Phantom portion of the ACR CT accreditation program 2428

FIG. 4. ~a! Good CT number calibration image. ROIs are well positioned. Note the subtle gray shading around the high-precision water rod. This isphantom manufacturing process and is not a scanner artifact.~b! CT numbers out of recommended range: This image demonstrates mean CT numbwater, acrylic, and polyethylene that are outside the program recommendations.~c! Image lacks the necessary text annotation. Essential details such asmAs, scan thickness, reconstruction algorithm, etc., are all missing. These technical factors must be displayed on all filmed images.

te-haan

io

been shown to give mean CT numbers of approxima062 HU for a wide range of CT scanner models. Hence, mean CT numbers for water tare not within the program tolerances indicate that scner calibration should be repeated. Figures 4~a!–4~c!show an example of acceptable CT number calibratimage and two images with various deficiencies.

Medical Physics, Vol. 31, No. 9, September 2004

ly

t-

n

1. Essential criteria for CT number calibration

• ROIs must be placed within the cylinders.

• Polyethylene mean CT number must be between2107and287 HU.

• Water mean CT number must be between27 and17 HU (65 HU preferred!.

Page 7: The phantom portion of the American College of Radiology ACR … · 2005-05-09 · The phantom portion of the American College of Radiology —ACR– Computed Tomography —CT–accreditation

5 mm.

er

ting up theave

2429 McCollough et al. : Phantom portion of the ACR CT accreditation program 2429

FIG. 5. ~a! Good slice thickness image. In this example, we count 15 wires for both the top and bottom ramp, which indicates a scan thickness of 7.~b!Not centered: The image submitted should have been obtained at the table position where the scan is centered on the center of the ramp~the longer wire shouldbe centered within the other wires!. In this case of a 10 mm scan thickness, the scan width may extend beyond the ramp~where there are no more wires! and,hence, the measurement may not be accurate. Additionally, the reconstructed FOV is too large~400 mm!. Note: The white ring around the high precision watrod is adhesive from the manufacturing process. It is not problematic for the purposes of the accreditation program.~c! Streak artifacts coming off of the wireramp. These are not typical. Also streaks are seen coming in from the bottom left and right from the Teflon roller on the phantom stand. When setphantom, the ring should be placed under the far ends of the phantom to avoid these streaks.~d! Slice thickness too large: The measurement appears to hbeen performed correctly yet the measured scan width is too large. The prescribed width is 1.5 mm, but six wires are counted on the top ramp~3 mm!, andfive wires counted on the bottom ramp~2.5 mm!. Additionally, the reconstructed FOV is too large~400 mm!.

be

• Acrylic mean CT number must be between1110 and1130 HU.

• Bone mean CT number must be between1850 and1970 HU.

Medical Physics, Vol. 31, No. 9, September 2004

• Air mean CT number must be between21005 and2970 HU.

• The adult abdomen protocol from Table I mustused.

Page 8: The phantom portion of the American College of Radiology ACR … · 2005-05-09 · The phantom portion of the American College of Radiology —ACR– Computed Tomography —CT–accreditation

nneretting.theheeyondof.

2430 McCollough et al. : Phantom portion of the ACR CT accreditation program 2430

FIG. 6. ~a! Acceptable image. This 140 kVp image has a mean CT number for water of 0.28 HU. In the 80, 100, and 120 kVp images from this sca~notshown!, the mean CT number of water was 1.46, 0.90, and 0.53 HU, respectively, demonstrating solid scanner calibration independent of kVp s~b!Incorrect mean CT number for water. This 80 kVp image has a mean CT number of26.8 HU, which is at the limit of the program tolerances, suggestingneed for scanner recalibration~many states require water values between25 HU and15 HU). Significant ring artifacts are apparent, indicating that tscanner requires calibration or service.~c! Incorrect mean CT number for water. This 90 kVp image has a mean CT number of 12.6 HU, which is well bprogram limits and indicates the need for scanner recalibration at this kVp setting.~d! Filming artifacts: This 80 kVp image has a mean CT number22.7 HU, which is acceptable. However, horizontal artifacts are well visualized. These are from the hard-copy camera device@same camera as used for Fig2~b! showing the poor quality SMPTE#. The medical physicist should note and address artifacts seen in any of the accreditation images.

ed

roe

am

ing

guldisis

E. Module 1—slice thickness and CT number of waterversus slice thickness

The width of the CT image is most completely describby a section sensitivity profile, particularly for spiral~helical!images. For the purposes of the ACR CT accreditation pgram, only axial slice widths are measured. These are msured using a series of discrete wires positioned on a r

Medical Physics, Vol. 31, No. 9, September 2004

-a-p

inclined with respect to the axial plane such that the spacbetween contiguous wires represent 0.5 mm along thez axis.Thus, an estimate of slice width~in mm! can be obtained bycounting the number of well-visualized wires and dividinthe total number by 2. Very subtle appearing wires shonot be counted. Interobserver variability by this methodabout 0.5 mm. Additionally, the mean CT number of water

Page 9: The phantom portion of the American College of Radiology ACR … · 2005-05-09 · The phantom portion of the American College of Radiology —ACR– Computed Tomography —CT–accreditation

en

rovideadiallyad failed.asr

amete

e been

2431 McCollough et al. : Phantom portion of the ACR CT accreditation program 2431

FIG. 7. ~a! Good low contrast resolution image. Adult abdomen protocol. Minimum resolvable rod size55 mm ~need to see all four rods clearly!. MeasuredCTDIw522.8 mGy, CTDIvol515.2 mGy.~b! Acquisition and analysis errors. Head scan FOV was used~large scan FOV should be used for an abdomprotocol!. ROI positioned radially too far from center. Minimum resolvable rod size56 mm. ~The 5 mm rods are not all well visualized.! ~c! Ring artifact andother deficiencies. The low contrast test object is rotated by 90°. This is a manufacturing error and the site was advised to contact the phantom pr forreplacement or repair. The streaks at the lower left and right of the image are from the Teflon roller on the phantom stand. The ROI is positioned rtoofar from center. Most importantly, there is a dark ring artifact. Finally, some text annotation covers the low contrast test objects. Minimum resolvble rodsize54 mm ~although one is partially hidden by text!. ~d! Failed low-contrast test. The 6 mm diameter low contrast rods are not visible. This site woulthe accreditation. The contrast between the background and the large rod should be 660.5 HU. Site measures 4.3 HU, but the ROI is not properly positionSite was advised to repeat measurement of contrast just to make sure the phantom was not the problem~we have not found any cases were the phantom wthe problem, but it is best to be sure!. The scan protocol likely needs to be revised or the scanner serviced.~e! Failed low-contrast test. The 6 mm diametelow contrast rods are not all clearly visible. This site would fail the accreditation. The likely cause is the use of the detail reconstruction kernel for an abdomenprotocol. This kernel is much too sharp and, hence, the image much too noisy, which severely degrades low contrast resolution. The other scan parrs anddose appear appropriate.~f! Wrong window width and level settings. The images were printed at the wrong window width~39! and level~91!. This large ofdeviation from the prescribed levels~100/100! is not acceptable, as it dramatically alters the appearance of the image and program criteria havestablished relative to specified window width and level settings. Also, essential details such as kVp, mAs, scan thickness, reconstruction algorithm, etc., areall missing. These technical factors must be displayed on all filmed images. The reconstructed FOV is also too large~36 cm instead of 21–25 cm!.

Medical Physics, Vol. 31, No. 9, September 2004

Page 10: The phantom portion of the American College of Radiology ACR … · 2005-05-09 · The phantom portion of the American College of Radiology —ACR– Computed Tomography —CT–accreditation

w

-

achs of

tual-

gs.

yetase

hatanrver

hectsto

ari-ead

e ofe

so-in

e I

ly

m-ith-

2432 McCollough et al. : Phantom portion of the ACR CT accreditation program 2432

evaluated at each slice thickness. Figures 5~a!–5~d! show anacceptable slice thickness image and several imagesvarious deficiencies.

1. Essential criteria for CT slice thickness

• Image data required for HRC,;3, 5, and 7 mm slicethicknesses.

• Water mean CT number must be between27 and17 HU (65 HU preferred!.

• The slice width must be within 1.5 mm of the prescribed width.

FIG. 7. ~Continued.!

Medical Physics, Vol. 31, No. 9, September 2004

ith

F. Module 1—CT number of water versus kVp setting

The mean CT number of water must be evaluated at ekVp setting that can be selected by the operator, regardlesthe frequency of use of those settings in the site’s acclinical practice. Figures 6~a!–6~d! show an acceptable module 1 image at a nonstandard kVp setting~i.e., not the kVpnoted in Table I for the adult abdomen protocol!, as well asseveral images with various deficiencies.

1. Essential criteria for CT number of water versuskVp

• Image data are required for all selectable kVp settin

• Water mean CT number must be between27 and17 HU (65 HU preferred!.

G. Module 2—low contrast resolution

The assessment of low contrast resolution is a difficultclinically relevant task. The scan parameters that increimage noise levels can cause subtle~low contrast! objects tobe poorly visualized. Alternatively, parameters choices tlower image noise and improve low contrast resolution cresult in excessive patient doses. To decrease interobsevariability for this test, the user is instructed to select tsmallest set of low contrast objects for which all four objecan be clearly delineated. It is common for sites to claimbe able to see smaller rods than do the reviewers.

Figures 7~a!–7~f! and 8~a!–8~f! show examples of goodlow contrast resolution images and several images with vous deficiencies, using the adult abdomen and adult hprotocols, respectively. Common problems include the uswrong window and level settings, poor positioning of thROIs ~they should be at the same radial distance from icenter!, use of scan parameters that differ from those listedTable I, and the presence of artifacts.

1. Essential criteria for low contrast resolution

• Adult abdomen and adult head protocols from Tablmust be used.

• Window width5100.

• Window level5100.

• All four cylinders of the 6 mm rods must be clearvisible.

H. Module 3—uniformity

Proper calibration of a CT system should result in an iage of a uniform phantom that appears uniform and is w

Page 11: The phantom portion of the American College of Radiology ACR … · 2005-05-09 · The phantom portion of the American College of Radiology —ACR– Computed Tomography —CT–accreditation

ringand applser

eir

ribed a

ential testis

2433 McCollough et al. : Phantom portion of the ACR CT accreditation program 2433

FIG. 8. ~a! Good low contrast resolution image. Adult head~brain! protocol. Minimum resolvable rod size55 mm. The measured CTDIw~43.0 mGy! is belowthe reference value~60 mGy!. However, the module is rotated~manufacturing error! and the annotation is covering some of the test objects. The brightaround the periphery is not atypical for the head low contrast resolution image, as many scanners assume the presence of skull around the brainy abone correction algorithm. Since the phantom does not have a high-atomic number shell, the bone correction algorithm creates the ring artifact obved. ~b!Shading artifact and somewhat high dose. This image has a minimum resolvable rod size55 mm. However, the measured CTDIw~67.2 mGy! is slightlyabove the reference value~60 mGy!. A shading artifact is present where the right portion of image~left side of phantom! appears brighter than the left.~c! Veryhigh dose. Adult head~brain! protocol. This image has a minimum resolvable rod size56 mm ~all 4 of the 5 mm rods are not clearly present!. However, themeasured CTDIw~114 mGy! is almost twice as high as the reference value~60 mGy!. ROIs are in the correct position. This site needs to work with thservice and applications to optimize the scan parameters such that they can still pass the low contrast resolution~6 mm or less! while hopefully lowering theirdose. The high dose required to achieve the necessary level of image quality is in part due to this system’s use of xenon gas detectors.~d! Wrong scanparameters: This image has a minimum resolvable rod size55 mm. However, the image was acquired using the wrong scan parameters. Table I presc5 mm scan at 160 mA and 2 s~320 mAs! but this scan was taken at 320 mA and 2 s~640 mAs! and is only 2.5 mm thick.~The two errors should theoreticallynegate one another in terms of the appearance of the low contrast image.! The measured CTDIw~87.4 mGy! is well above the reference value~60 mGy!. ROIsare is a good position.~e! Ring artifacts and other problems. This image shows subtle rings artifacts and has the text annotation covering the essobjects. The window and level settings are not correct. The technique is rather high~700 mAs! likely due to the choice of a 2.5 mm scan thickness, whichtypically considered too thin for routine brain evaluations. The measured CTDIw~105 mGy! is almost twice as high as the reference value~60 mGy!. Thebackground ROI is not in the correct position.~f! High dose. Good low contrast resolution image using the adult head~brain! protocol. Minimum resolvablerod size54 mm. However, the measured CTDIw~73.6 mGy! is above the reference value~60 mGy!.

Medical Physics, Vol. 31, No. 9, September 2004

Page 12: The phantom portion of the American College of Radiology ACR … · 2005-05-09 · The phantom portion of the American College of Radiology —ACR– Computed Tomography —CT–accreditation

heg

sfo

it

b

ing

a-

ars

onsingre-

mn

for

for

ionifyage,ure-Ihem

xial

eheialtiontheay

gs.es

2434 McCollough et al. : Phantom portion of the ACR CT accreditation program 2434

out artifacts. Uniformity is quantitated by measuring tmean CT number with a ROI at the center and four edpositions value and calculating the absolute value of@centermean CT No.—edge mean CT No.# for all four edge ROIs.The difference between the edge and center should mea<5 HU. The image should also be carefully examinedartifacts such as rings or streaks. Figures 9~a!–9~f! show anexample of a good uniformity image and several images wvarious deficiencies.

1. Essential criteria for image uniformity

• Edge-to-center mean CT number difference must,5 HU for all four edge positions.

• Correct size and location of ROIs.

FIG. 8. ~Continued.!

Medical Physics, Vol. 31, No. 9, September 2004

e

urer

h

e

• The center CT number must be between27 and17 HU (65 HU preferred!.

• Adult abdomen protocol must be used.

• Window width5100.

• Window level50.

• No image artifacts.

I. Module 4—high contrast resolution

The high contrast resolution image must be viewed usa window width5100 and a window level'1100 ~the win-dow level should be adjusted slightly to optimize visualiztion of the higher frequency bar patterns!. The physicist mustdetermine the highest spatial frequency for which the band spaces are distinctly visualized. Figures 10~a!–10~c! and11~a!–11~c! show examples of good high contrast resolutiimages and several images with various deficiencies, uthe adult abdomen and high resolution chest protocols,spectively.

1. Essential criteria for high contrast resolution

• The adult abdomen and adult HRC protocols froTable I must be used~especially correct reconstructioalgorithm!.

• Window width5100.

• Window level'1100.

• The 5 lp/cm bar pattern must be clearly resolvedadult abdomen.

• The 6 lp/cm bar pattern must be clearly resolvedHRC.

J. CTDI measurements

The ACR CT accreditation program requires submissof images from the CTDI measurements primarily to verthe correct phantom size and position, ion chamber usand correct acquisition parameters during the CTDI measments. Figures 12~a!–12~f! show examples of a good CTDimage and several images having critical deficiencies. Tstandard 16 and 32 cm acrylic CTDI phantoms, 100 mionization chamber, and calculation methods are used.2–10

The CTDI measurement must be performed using an ascan with all other technical parameters~kVp, mA, exposuretime, N, T) the same as the original protocol. For somscanners, this will require paying careful attention to ttechnical parameters when switching from helical to axmode. It is absolutely essential that the detector configuraused for the helical scan prescribed in Table I be used foraxial scan for measuring CTDI. In some scanners this mrequire going into ‘‘service mode’’ to achieve these settinThe FAQ section of the ACR accreditation web site providadditional guidance for this issue: www.acr.org.

1. Essential criteria for CTDI measurements for theadult head protocol

• Adult head protocol must be used.

• An axial scan must be used.

Page 13: The phantom portion of the American College of Radiology ACR … · 2005-05-09 · The phantom portion of the American College of Radiology —ACR– Computed Tomography —CT–accreditation

artifactsber

notttingsre missin

din

2435 McCollough et al. : Phantom portion of the ACR CT accreditation program 2435

FIG. 9. ~a! Good uniformity image. This image obtained with the adult abdomen protocol from Table I, with the exception of kVp value, shows no~cupping, streaks, rings, etc.! ~note that it is a helical scan, as would be expected for body imaging!. The data sheets require measuring the mean CT numat the center and four edge positions, but for many systems only three ROI measurements can be displayed at one time. This is acceptable.~b! Cupping artifact.This image visually demonstrates substantial cupping~darker CT numbers towards the center of the phantom!, even though the measured CT values aresubstantially different from center to edge. Also, the reconstruction FOV is too large and the image was filmed with the wrong window and level se. ~c!ROIs too large. This image shows very good uniformity, however, the sizes of the ROIs are too large. The bright BBs off center and near the edge agfrom the image, most likely due to the site not having filmed the image that is positioned over the center of module 3.~d! Ring artifacts. Concentric ringartifacts are visible in this uniformity image. The four edge ROIs are placed somewhat too near to center.~e! Bright ring artifact around the periphery anhelical ~windmill! artifacts coming off of the bright BBs. The ROI sizes are too large.~f! Dark ring artifact. The ring around the phantom is not reflectedthe acceptable ROI measurements. The ROIs are of good size and position. The BBs appear crisp in spite of a helical pitch factor of 1.5.

cor-can

• Phantom should be in the head holder.

• Phantom must be 16 cm in diameter.

• The nonchamber holes must be filled.

Medical Physics, Vol. 31, No. 9, September 2004

• Complete the dose calculator Excel spreadsheetrectly. ~Parameters must match Table I and actual sparameters.!

Page 14: The phantom portion of the American College of Radiology ACR … · 2005-05-09 · The phantom portion of the American College of Radiology —ACR– Computed Tomography —CT–accreditation

e

coca

eet.

cor-can

eet.

Iw,IngDIe es-am-be-theeettersin

im-ctss ofon

n-anditionbeenEr-ithan.

testringltay-al.lity

esthisgese,

2436 McCollough et al. : Phantom portion of the ACR CT accreditation program 2436

• Print and submit the dose calculator Excel spreadsh

• The CTDIw should not exceed 60 mGy.

2. Essential criteria for CTDI measurements for thepediatric abdomen protocol

• Pediatric abdomen protocol must be used.

• An axial scan must be used.

• Phantom must be on the table top.

• Phantom must be 16 cm in diameter.

• The nonchamber holes must be filled.

• Complete the dose calculator Excel spreadsheetrectly. ~Parameters must match Table I and actual sparameters.!

FIG. 9. ~Continued.!

Medical Physics, Vol. 31, No. 9, September 2004

et.

r-n

• Print and submit the dose calculator Excel spreadsh

• The CTDIw should not exceed 25 mGy.

3. Essential criteria for CTDI measurements for theadult abdomen protocol

• Adult abdomen protocol must be used.

• An axial scan must be used.

• Phantom must be on the table top.

• Phantom must be 32 cm in diameter.

• The nonchamber holes must be filled.

• Complete the dose calculator Excel spreadsheetrectly. ~Parameters must match Table I and actual sparameters.!

• Print and submit the dose calculator Excel spreadsh

• The CTDIw should not exceed 35 mGy.

III. CTDI CALCULATIONS

An Excel spreadsheet is provided to calculate CTDCTDIvol, DLP, and effective dose from the CTDmeasurements.2,3,6–11Clinical dose estimates are made usithe clinical scan parameters from Table I. Although CTmeasurements are made using an axial scan, exam dostimates can be made for spiral acquisitions using the pareter CTDIvol, which takes into account gaps or overlapstween the radiation beams from contiguous rotations ofx-ray source. Table II shows a dose calculation workshthat has been correctly completed. The sample paramecorrespond to the adult abdomen protocol given earlierTable I.

IV. DISCUSSION AND CONCLUSIONS

The provided figures demonstrate acceptable levels ofage quality and dose, as well as a variety of image artifaand scanner performance errors. However, many examplephysicist error have also been shown. The most commoperator errors are listed in Table III.

The goal of the program is to document high quality scaner performance. In order to fairly assess image qualitydose for a wide range of scanner models and scan acquisparameters, a reasonable amount of standardization hasnecessary in the formatting of the submitted image data.rors such as filming the images in the wrong location or wthe wrong window width and level settings can makeapplication difficult or even impossible to review fairlyPhysicist errors such as not including a SMPTE or otherpattern, severe misalignment of the phantom, or measuCTDI with the wrong phantom or in helical mode will resuin the physicist having to repeat the submission. Thus, ping close attention to the provided instructions is essenti

Additionally, the ACR has chosen to assess image quaand dose for four standard imaging applications~adult abdo-men and head, pediatric abdomen, and high resolution ch!.If a site’s clinical image quality is to be assessed by tprogram, it is imperative that the submitted phantom imabe acquired with the relevant clinical protocols. Henc

Page 15: The phantom portion of the American College of Radiology ACR … · 2005-05-09 · The phantom portion of the American College of Radiology —ACR– Computed Tomography —CT–accreditation

e higher

the 7beenbetter

2437 McCollough et al. : Phantom portion of the ACR CT accreditation program 2437

FIG. 10. ~a! Good high contrast resolution image. Adult abdomen protocol. The window setting is 100 and the level setting is 1100, allowing thfrequency bar patterns to well visualized. The limiting spatial resolution~the highest spatial frequency where the bars are still clearly distinguishable! is 7lp/cm. ~b! Wrong window width and level settings. Here the window and level settings have not been properly set (width5832 instead of 100, level51376 instead of 1100!. The limiting spatial resolution is 6 lp/cm, which is passing. A better window width and level setting would likely show thatlp/cm is the limiting spatial resolution value.~c! Wrong window width and level settings. Here the window width setting is correct but the level has notadjusted~from the nominal value of 1100! to allow the higher frequency bar patterns to be visualized. The limiting spatial resolution is 7 lp/cm, but alevel setting might show that the 8 lp/cm is the limiting value.

tlytiam

ete

wl-g athecies

working closely with the site’s lead technologist to correcdocument the clinical scan parameters in Table I is essenThe physicist must then be diligent to use the scan pareters from Table I in the image quality and dose tests.

Through participation in this accreditation program, wbelieve that the medical physicist will become a more in

Medical Physics, Vol. 31, No. 9, September 2004

al.-

-

grated member of the CT imaging team. While we acknoedge the serious time commitment required for completinsuccessful CT accreditation application, we believe thatprogram has helped many to detect and resolve deficienin the their scanner performance~i.e., equipment problems!or scanner utilization~i.e., exam protocols!. The end result,

Page 16: The phantom portion of the American College of Radiology ACR … · 2005-05-09 · The phantom portion of the American College of Radiology —ACR– Computed Tomography —CT–accreditation

lyity to

2438 McCollough et al. : Phantom portion of the ACR CT accreditation program 2438

FIG. 11. ~a! Good high contrast resolution image. High resolution chest protocol. The window width is 108~acceptable! and the window level had beenadjusted to 1156 such that the higher frequency bar patterns are well visualized. The limiting spatial resolution is between 9 and 10 lp/cm.~b! Wrong windowwidth and level settings and wrong reconstruction algorithm. The window width setting is not correct~41 instead of 100! and the window level~1192! settingdoes allow the higher frequency bar patterns to be visualized. Additionally, a high resolution reconstruction algorithm was not used, such that onthe barpatterns between 6 and 7 lp/cm can be visualized.~c! FOV too large. The reconstruction FOV is too large, causing the large pixel size to limit the abilresolve the higher spatial frequency bar patterns that should be visible using the bone algorithm.

ord

her

nyuesin-ton,e

ages

an increase in the overall quality of CT imaging and a mjudicious use of ionizing radiation, is worth the time anattention of the medical physics community.

ACKNOWLEDGMENTS

The authors of this exhibit gratefully acknowledge tassistance of many people in the development of this p

Medical Physics, Vol. 31, No. 9, September 2004

e

o-

gram, including Pamela Wilcox, Lenny Lucey, and PenButler of the ACR, and numerous medical physics colleagwho have provided constructive feedback and invaluableput regarding the program. They additionally would likethank Dr. Frank Zink and Dr. Jim Kofler, Shawna Leitzeand Tim Daly, all of the Mayo Clinic, for their assistancwith numerous phantom scans during the development stof this program.

Page 17: The phantom portion of the American College of Radiology ACR … · 2005-05-09 · The phantom portion of the American College of Radiology —ACR– Computed Tomography —CT–accreditation

e filled.

enered by ttom image.e I exampleed ofmedscanitch of 0.an

or the

2439 McCollough et al. : Phantom portion of the ACR CT accreditation program 2439

FIG. 12. ~a! Good CTDI image. Adult head~brain! protocol. The 16 cm diameter phantom is centered in the head holder and the chamber cavities arThe detector configuration for the scan matches that given in, the respective, Table I~2.5 mm detector collimation!. ~b! Incorrect CTDI scan—helical scan. Ahelical scan was incorrectly used for acquiring the CTDI measurement. The data from these measurements are not valid for CTDI calculations.~c! IncorrectCTDI scan—wrong size phantom for pediatric dose measurement. The large~32 cm diameter! CTDI phantom was incorrectly used for the pediatric abdomCTDI measurements. The data from these measurements are not valid for the pediatric CTDI calculations. The incorrect phantom size was discovheACR reviewer after noting that the measured doses were extremely low and seeing that the 30 cm diameter FOV was completely filled in the phan~d! Correct multislice detector configuration used. This image of an adult abdomen CTDI scan corresponds to the scan parameters given in the Tableabove~with the exception of a different mA value!. The clinical protocol prescribes a helical 5 mm scan thickness using a pitch of 0.75 and a table sp11.25 mm/rotation. These parameters can only be achieved using the 433.75 mm detector configuration. Here the axial CTDI dose scan is correctly perforusing the 433.75 mm detector configuration~3.75 mm/4i!. ~e! Incorrect multislice detector configuration used. This image of an adult abdomen CTDIcorresponds to the scan parameters given in the Table I example above. The clinical protocol prescribes a helical 5 mm scan thickness using a p75and a table speed of 11.25 mm/rotation. These parameters can only be achieved using the 433.75 mm detector configuration. Here the axial CTDI dose scis incorrectly performed using the 435 mm detector configuration~5 mm/4i!. ~f! High dose. This head CTDI image uses the same exam parameters as fhead low contrast resolution image shown in Fig. 8~c!. The technique is high~160 mA, 3 s! and yields a CTDIw of 114 mGy.

Medical Physics, Vol. 31, No. 9, September 2004

Page 18: The phantom portion of the American College of Radiology ACR … · 2005-05-09 · The phantom portion of the American College of Radiology —ACR– Computed Tomography —CT–accreditation

isccatte

mofs of

Tonsist

issists

an-fournddi-

to

2440 McCollough et al. : Phantom portion of the ACR CT accreditation program 2440

APPENDIX: DETAILED DESCRIPTION OF THE ACRCT ACCREDITATION PHANTOM

A. Module 1

Figure 13~a! shows a cross-section of module I, whichused to assess positioning and alignment, CT number aracy, and slice thickness. The background material is wequivalent. For positioning, the module has 1-mm diamesteel BBs embedded at the longitudinal (z axis! center of the

FIG. 12. ~Continued.!

Medical Physics, Vol. 31, No. 9, September 2004

u-err

module, with the outer surface of the BB at the phantosurface at 3, 6, 9, and 12 o’clock positions within the fieldview. To assess CT number accuracy, there are cylinderdifferent materials: bone-mimicking material~‘‘Bone’’ !,polyethylene, water equivalent material, acrylic, and air.assess slice thickness, two ramps are included which coof a series of wires that are visible in 0.5 mmz-axis incre-ments.

B. Module 2

Figure 13~b! shows a cross-section of module 2, whichused to assess low contrast resolution. This module conof a series of cylinders of different diameters, all at 0.6%~6HU! difference from a background material having a meCT number of approximately 90 HU. The cylinder-tobackground contrast is energy independent. There arecylinders for each of the following diameters: 2, 3, 4, 5, a6 mm. The space between each cylinder is equal to theameter of the cylinder. A 25 mm cylinder is includedverify the cylinder-to-background contrast level.

TABLE II. Sample dose calculation worksheet.

CTDI Body Phantom~32-cm diameter PMMA Phantom! Measured Calculated

kVp 120mA 240Exposure time per rotation~s! 0.8Z axis collimationT ~mm! 3.75# data channelsused(N) 4Axial (A): Table Increment (mm)5(I )ORHelical (H): Table Speed (mm/rot)5(I )

11.25

Active Chamber length~mm! 100Chamber correction factor 1CenterMeasurement 1~mR! 152Measurement 2~mR! 155Measurement 3~mR! 154Average of above 3 measurements~mR! 153.7Body CTDI at isocenter in phantom~mGy! 8.912 o’clock positionMeasurement 1~mR! 311Measurement 2~mR! 309Measurement 3~mR! 309Average of above 3 measurements~mR! 309.7Body CTDI at 12 o’clock position inphantom~mGy!

18.0

CTDIw ~mGy! 14.9Clinical exam dose estimates~using measured CTDIw and site’s Adult AbdomenProtocol from Table I!CTDIvol ~mGy! 5CTDIw* N* T/I 19.9DLP ~mGy-cm! 5CTDIvol* 25 498.3Eff Dose ~mSv! 5DLP* 0.015 7.5

Page 19: The phantom portion of the American College of Radiology ACR … · 2005-05-09 · The phantom portion of the American College of Radiology —ACR– Computed Tomography —CT–accreditation

rmetc.

2441 McCollough et al. : Phantom portion of the ACR CT accreditation program 2441

Medical Physics, Vo

TABLE III. Common operator errors.

Scans acquired using parameters that do not match those listed in Table I of the site scanning data fo~kVp, mAs, scan width, detector configuration, reconstruction algorithm, scan and reconstruction FOV,!Incorrect detector configuration~values ofN andT! listed in Table IIncorrect calculation of pitch or table incrementNot submitting a SMPTE pattern or alternate video test patternPoor phantom alignment~central wire not centered in ramp, all 4 BBs not the same brightness!Placing images in the wrong positions~boxes! on the films submitted to the ACRFilming images with the wrong window width and window level settingsWrong size or wrong position ROIsSubmitting images having obvious artifacts or other deficient resultsPerforming CTDI scans using the wrong detector configuration~values ofN andT!Not submitting the printed Excel ‘‘dose calculator’’ spreadsheetNot noting the difference between mA, mAs, and effective mAs (5mAs/pitch) when completing the formsNot displaying sufficient technical parameters on the printed films~all scan parameters must be shown!Positioning ROIs such that the annotation covers important portions of the imageUsing too small or too large a reconstruction FOV

chC

ino

is

12ctlso

1.

C. Module 3

Figure 13~c! shows a cross-section of module 3, whiconsists of a uniform, tissue-equivalent material to assessnumber uniformity. Two very small BBs~0.28 mm each! areincluded for optional use in assessing the accuracy ofplane distance measurements. They may also be used tsess section sensitivity profiles.

l. 31, No. 9, September 2004

T

-as-

D. Module 4

Figure 13~d! shows a cross-section of module 4, whichused to assess high contrast~spatial! resolution. It containseight bar resolution patterns: 4, 5, 6, 7, 8, 9, 10, andlp/cm. The aluminum bar patterns provide very high objecontrast relative to the background material. Module 4 ahas four 1 mm steel beads, as described for module

f

of

-e-

tnd

-

ss

-cy

h

-oe

FIG. 13. ~a! Cross-sectional diagram omodule 1, which contains four cylin-drical rods to assess the CT numberdifferent materials, four BBs to con-firm accurate positioning, and two inclined ramps consisting of discretwires that are spaced in 0.5 mm increments along thez axis. ~b! Cross-sectional diagram of the low-contrasresolution test objects contained imodule 2. Each cylinder is compriseof the material having a nominal CTnumber difference from the background material equal to 6 HU.~c!Cross-sectional diagram used to asseimage uniformity in module 3. Thetwo 0.28 mm beads are used for optional assessment of distance accuraor section-sensitivity profile. ~d!Cross-sectional diagram of the higcontrast ~spatial! resolution test ob-jects contained in module 4. The numerical values shown correspond tthe spatial frequency of the respectivbar pattern, in lp/cm.

Page 20: The phantom portion of the American College of Radiology ACR … · 2005-05-09 · The phantom portion of the American College of Radiology —ACR– Computed Tomography —CT–accreditation

gn

ma

,’’

thed

on

cs:

hy

ip-

in

y,’’

2442 McCollough et al. : Phantom portion of the ACR CT accreditation program 2442

a!Presented at RSNA Annual Meeting, November 2003, Recipient of MaCum Laude Award for Educational Exhibit.

b!Author to whom correspondence should be addressed. [email protected]

1J. E. Grayet al., ‘‘Test pattern for video display and hard-copy cameraRadiology145, 519–527~1985!.

2T. B. Shope, R. M. Gagne, G. C. Johnson, ‘‘A method for describingdoses delivered by transmission x-ray computed tomography,’’ MPhys.8, 488–495~1981!.

3American Association of Physicists in Medicine,Standardized Methodsfor Measuring Diagnostic X-Ray Exposures, Report no. 31~AAPM, NewYork, 1990!.

4American Association of Physicists in Medicine,Specification and Ac-ceptance Testing of Computed Tomography Scanners, Report no. 39~AAPM, New York, 1993!.

5L. N. Rothenberg and K. S. Pentlow, ‘‘CT dosimetry and radiati

Medical Physics, Vol. 31, No. 9, September 2004

a

il:

e.

safety’’ in RSNA Categorical Course in Diagnostic Radiology PhysiCT and US Cross-sectional Imaging. RSNA~2000!.

6European Guidelines for Quality Criteria for Computed Tomograp~European Commission, Luxembourg, 2000!.

7International Electrotechnical Commission,Medical Electrical Equip-ment. Part 2-44: Particular Requirements for the Safety of X-ray Equment for Computed Tomography, IEC publication No. 60601-2-44, 2ndedition, Amendment 1.

8H. D. Nagel, Radiation Exposure in Computed Tomography~COCIR,Frankfurt, 2000!.

9M. Gray-McNitt, ‘‘AAPM/RSNA physics tutorial for residents: Topics inCT radiation dose in CT,’’ Radiographics22, 1541–1553~2002!.

10R. L. Morin, T. C. Gerber, and C. H. McCollough, ‘‘Radiation dosecomputed tomography of the heart,’’ Circulation107, 917–922~2003!.

11C. H. McCollough, ‘‘Patient dose in cardiac computed tomographHerz 28, 1–6 ~2003!.