Upload
justina-johnston
View
215
Download
0
Embed Size (px)
Citation preview
TEM-investigation of CrSiON sample of Lorenzo Castaldi (EMPA)report 3
EMEZ-Proposal 0190
Elisabeth Müller 26.5.2008
Time used:
Microscopy I used your sample as test sample for EDX-mapping for 80h and 50h respectivelyData-analysis/reporting
If any of the TEM-data is used for a publication, I would certainly like to read the publication before submission
I made a test with EDX-mapping and used the CrSiON sample for this. Two of the maps were at lower, one at higher resolution. The former ones for the chemical distibution around two of the big „balls“ and one for the composition modulation.
The times used for the mappings were up to 80 hours measuring time at the Tecnai F30.
Energy (keV)
Co
un
ts
6.04.02.00.0
100
80
60
40
20
0
O
Si
Cr
Cr
Cr
Cr
N
EDX HAADF Detector Area 2
Energy (keV)
Co
un
ts
0.700.600.500.400.30
50
40
30
20
10
0
O
Cr
Cr
CrCr
N
EDX HAADF Detector Area 2
Problem: EDX-mapping on Tecnai F30Windows as defined for the mapping
Problems:
N, Si, O ↔ signal to noise ratio
O ↔ Cr-L lines extremely near
Energy windows as chosen for the EDX-mapping:
N-K: 0.344 - 0.427 eV
O-K: 0.5182 - 0.5358 eV
Si-K: 1.638 – 2.863 eV
Cr-K: 5.257 – 5.578 eV
CrSiON:faults in the layer with big balls (low magnification) and concentration modulation (higher magnification)
Cr-K map N-K map O-K map Si-K map
• small window: used for drift correction• large window: area chosen for the map• ball = Cr-rich area with possibly some Si• area below the ball = Si-rich area
Si-rich
Cr-rich
Very low-magnification EDX-map
Cr-K map N-K map O-K map Si-K map
• small window: used for drift correction• large window: area chosen for the map• size: 256 x 256 pixels = 1532 x 1532 nm• measuring time per pixel: 4“, total time 80h• area per pixel: 6 x 6 nm2
• ball = Cr-rich area with possibly some Si• area below the ball = Si-rich area
Si-rich
Cr-rich
Low-magnification EDX-map
Cr-K map N-K map O-K map Si-K map
• black and white: windows chosen• colored: automatic routine(190 x 190 pixels, 50h)
Energy (keV)
Co
un
ts
0.700.600.500.400.30
50
40
30
20
10
0
O
Cr
Cr
CrCr
N
EDX HAADF Detector Area 2
High-magnification EDX-map• small window: used for drift correction• large window: area chosen for the map• size: 190 x 190 pixels = 236 x 236 nm• measuring time per pixel: 4“, total time 50h• area per pixel: 1.24 x 1.24 nm2
O-blue_N-yellow
O-green_Si-rose
Si-turquoise_N-red
• O and N (resp. Cr): maxima and minima just exchanged. Gray areas besides blue lines and between yellow lines contain all three elements: Cr,N,OThe O-N EDX-map nicely confirms the EFTEM result. The main difference is, that the EFTEM result shows a better resolution than the EDX-map, as a slight splitting of the weak blue maximum appears to be split, which is not visble in the EDX-map.
• Si and N (resp. Cr): maxima and minima of the Si is shifted slightly towards sample surface as compared to N and Cr.
Sample surface
• Si and O: maxima of the Si is shifted slightly towards sample substrate as compared to O.
EFTEM
EDX
• Cr and N: maxima and minima at the same positions. Therefore only one of these elements is compared to the others.
elemental distribution
50
70
90
110
130
150
170
190
210
230
250
0 10 20 30 40 50 60 70 80
pixel
inte
nsit
y
Si-K Cr-K
N-K O-K Summary:Concentration maxima • N = Cr (double
maximum with a small minimum in between and separated by a strong minimum)
• O = -N (always one strong maximum followed by a weak one)
• Si: maxima always between the top side of N/Cr-maximum and the bottom side of the O-maximum
top bottom
STEM image acquired during EDX-mapping
HAADF STEM image acquired with small spot size
Cr Cr Cr CrSi Si Si Si