Team PANDA, Zurich, Switzerland - [email protected] - TEAM PANDA TEST PROTOCOL INLINE PEEP VALVE COMPARISON Project PANDApeep Date/Place 4.5.2020 / Winterthur, Switzerland Authors Dominik Textor, Stefan Stahl Introduction This test series shall compare the PANDApeep Inline valve with the commercial PEEP valve Ambu PEEP 20. This is not a medical test. Components PEEP valves: PANDApeep Gen2 inline: https://www.thingiverse.com/thing:4316582 [Configuration: Bike hub spring mounted on retrofit kit / Material: Profill PLA (case)] o Rubber seal (1 mm thickness) o 3D-printed TPU seal Ambu PEEP 20 disposable: https://www.ambu.com/emergency-care-and-training/resus- citators/product/peep-valves Flow restrictor (Tests without PEEP valve only): PANDAflow: https://www.thingiverse.com/thing:4289425 Test rig: Compressor: Revell Airbrush Gamma Check valve SMC AKH Magnetic fill valve Bürkert 6012 Test lung Dräger MP02400 1000 ml Manual ball valve RB Patient Tube: length 400 mm, connectors 22 mm Sensors & Controls: Pressure Sensor: Bosch BMP280 Control electronics: Arduino Uno with 2x relay bank 24 VDC Power supply Manual ball valve Check valve Adjustable PEEP valve Pressure sensor Test lung Compressor Magnetic 3/2 valve Patient tube 400 mm

TEAM PANDA TEST PROTOCOL INLINE PEEP VALVE COMPARISON€¦ · 04/05/2020  · Team PANDA, Zurich, Switzerland - [email protected] - TEAM PANDA TEST PROTOCOL INLINE PEEP VALVE

  • Upload
    others

  • View
    3

  • Download
    1

Embed Size (px)

Citation preview

Page 1: TEAM PANDA TEST PROTOCOL INLINE PEEP VALVE COMPARISON€¦ · 04/05/2020  · Team PANDA, Zurich, Switzerland - teampandacovid19@gmail.com - TEAM PANDA TEST PROTOCOL INLINE PEEP VALVE

Team PANDA, Zurich, Switzerland - [email protected] -

TEAM PANDA TEST PROTOCOL

INLINE PEEP VALVE COMPARISON

Project PANDApeep

Date/Place 4.5.2020 / Winterthur, Switzerland

Authors Dominik Textor, Stefan Stahl

Introduction

This test series shall compare the PANDApeep Inline valve with the commercial PEEP valve Ambu PEEP 20. This is not a medical test.

Components

PEEP valves:

PANDApeep Gen2 inline: https://www.thingiverse.com/thing:4316582 [Configuration: Bike hub spring mounted on retrofit kit / Material: Profill PLA (case)]

o Rubber seal (1 mm thickness) o 3D-printed TPU seal

Ambu PEEP 20 disposable: https://www.ambu.com/emergency-care-and-training/resus-citators/product/peep-valves

Flow restrictor (Tests without PEEP valve only):

PANDAflow: https://www.thingiverse.com/thing:4289425 Test rig:

Compressor: Revell Airbrush Gamma

Check valve SMC AKH

Magnetic fill valve Bürkert 6012

Test lung Dräger MP02400 1000 ml

Manual ball valve RB

Patient Tube: length 400 mm, connectors 22 mm Sensors & Controls:

Pressure Sensor: Bosch BMP280

Control electronics: Arduino Uno with 2x relay bank

24 VDC Power supply

Manual ball valveCheck valve Adjustable PEEP valve

Pressure sensor

Test lung

Compressor Magnetic 3/2 valve

Patient tube 400 mm

Page 2: TEAM PANDA TEST PROTOCOL INLINE PEEP VALVE COMPARISON€¦ · 04/05/2020  · Team PANDA, Zurich, Switzerland - teampandacovid19@gmail.com - TEAM PANDA TEST PROTOCOL INLINE PEEP VALVE

Page 2

Team PANDA, Zurich, Switzerland - [email protected]

Page 3: TEAM PANDA TEST PROTOCOL INLINE PEEP VALVE COMPARISON€¦ · 04/05/2020  · Team PANDA, Zurich, Switzerland - teampandacovid19@gmail.com - TEAM PANDA TEST PROTOCOL INLINE PEEP VALVE

Page 3

Team PANDA, Zurich, Switzerland - [email protected]

Method Sensitivity testing: Check response to test lung expiration at different PEEP pressure values

Fully open the PEEP valve

Close the PEEP valve to start position

Fill test lung up to 30 cm H2O

Start pressure measurement

Wait 1 s

Open ball valve

Expiration for 5 s

Stop pressure measurement

Close ball valve

Close the PEEP valve to the next position

Iterate until last position is measured

Measure frequency of pressure sensor: 10 Hz Repeatability testing: Check variation of response to test lung expiration

Same procedure as sensitivity testing with a constant PEEP valve position (ca. 15 cm H2O) Overview PEEP valve test positions:

PEEP valve Test series Protocol Pos. 1 Pos. 2 Pos. 3 Pos. 4 Pos. 5

PANDApeep Gen2 inline TPU seal

Series 1 Sensitivity Start +1/4 turn +1/2 turn +3/4 turn +1 turn

Series 2 Repeatability +3/4 turn +3/4 turn +3/4 turn +3/4 turn +3/4 turn

Series 2a Repeatability Same position as Series 2 but top down (minimum PEEP direction)

+3/4 turn +3/4 turn +3/4 turn +3/4 turn +3/4 turn

Series 3 Sensitivity Same positions as Series 1 but starting from 40 cm H2O

Start +1/4 turn +1/2 turn +3/4 turn +1 turn

PANDApeep Gen2 inline Rubber seal

Series 1 Sensitivity Start +1/4 turn +1/2 turn +3/4 turn +1 turn

Series 2 Repeatability +1/2 turn +1/2 turn +1/2 turn +1/2 turn +1/2 turn

Ambu PEEP 20 Series 1 Sensitivity 5 10 15 20 Max

Series 2 Repeatability 15 15 15 15 15

System testing: The test systems expiration curve without PEEP valve was tested qualitatively with a flow restrictor in the same place.

Page 4: TEAM PANDA TEST PROTOCOL INLINE PEEP VALVE COMPARISON€¦ · 04/05/2020  · Team PANDA, Zurich, Switzerland - teampandacovid19@gmail.com - TEAM PANDA TEST PROTOCOL INLINE PEEP VALVE

Page 4

Team PANDA, Zurich, Switzerland - [email protected]

Results The following results were achieved:

Figure 1: Results PANDApeep Gen2 Inline with TPU seal

Page 5: TEAM PANDA TEST PROTOCOL INLINE PEEP VALVE COMPARISON€¦ · 04/05/2020  · Team PANDA, Zurich, Switzerland - teampandacovid19@gmail.com - TEAM PANDA TEST PROTOCOL INLINE PEEP VALVE

Page 5

Team PANDA, Zurich, Switzerland - [email protected]

Figure 2: Results PANDApeep Gen2 Inline with Rubber seal

Figure 3: Results Ambu PEEP 20

Figure 4: Results System testing without PEEP valve and restricted flow (PANDAflow)

Page 6: TEAM PANDA TEST PROTOCOL INLINE PEEP VALVE COMPARISON€¦ · 04/05/2020  · Team PANDA, Zurich, Switzerland - teampandacovid19@gmail.com - TEAM PANDA TEST PROTOCOL INLINE PEEP VALVE

Page 6

Team PANDA, Zurich, Switzerland - [email protected]

Discussion Sensitivity:

All tested valve versions offer enough sensitivity to adjust the PEEP to < 1 cm H2O. The Ambu PEEP 20 has much more revolutions per pressure difference.

All valves need pressure measurement for exact adjustment. Whereas the PANDApeep has no absolute indication, the Ambu PEEP 20 can be adjusted to ±1-2 cm H2O (See Figure 3).

Expiration behavior:

The PANDApeep Gen2 Inline with Rubber seal and the Ambu PEEP 20 have a very similar behavior (Figure 5) The PANDApeep Gen2 Inline with TPU seal shows slower discharge and inferior closing behavior (Figure 6).

Figure 5: Expiration curves PANDApeep Gen2 Inline with rubber seal (Series 1) and Ambu PEEP 20 (Series 1)

Page 7: TEAM PANDA TEST PROTOCOL INLINE PEEP VALVE COMPARISON€¦ · 04/05/2020  · Team PANDA, Zurich, Switzerland - teampandacovid19@gmail.com - TEAM PANDA TEST PROTOCOL INLINE PEEP VALVE

Page 7

Team PANDA, Zurich, Switzerland - [email protected]

Figure 6: Expiraton curves PANDApeep Gen2 inline with TPU and rubber seal

Repeatability

All valves show excellent repeatability over several expiration cycles (see Figure 1, Figure 2 and Figure 3).

Hysteresis

The PANDApeep Gen2 Inline with TPU seal was tested for dependency on turning direc-tion. A hysteresis of ca. 2 cm H2O can be seen (Figure 7).

Figure 7: Hysteresis in Expiration curves depending on turning direction

Page 8: TEAM PANDA TEST PROTOCOL INLINE PEEP VALVE COMPARISON€¦ · 04/05/2020  · Team PANDA, Zurich, Switzerland - teampandacovid19@gmail.com - TEAM PANDA TEST PROTOCOL INLINE PEEP VALVE

Page 8

Team PANDA, Zurich, Switzerland - [email protected]

System expiration behavior: As there were ripples in the expiration curves with all valves at low PEEP levels, a qualitative test with a constant flow restriction was conducted. According to this test, the ripples are result-ing of the test system (test lung, connectors and patient tubes) and not caused by the valves (Figure 4). The test setup tries to mimic the setup used in the clinical application. Measuring without the ad-ditional patient tube results in cleaner opening behavior for the commercial as well as for the 3D printed valves. This shows that the valve behavior is dependent on the tube setup.

Conclusion

In our test setup, the PANDApeep Gen2 Inline with Rubber seal and bike hub spring achieves very promising results: Expiration behavior and repeatability are comparable to the commercial Ambu PEEP 20. The version with the 3D-printed TPU seal is working as well, but with much slower discharge and therefore higher dependency on inspiration pressure and frequency.