Spatially Resolved High Resolution X-Ray Crystal Spectrometry and Radiation Tolerant 2D X-Ray Imaging Schemes for ICF Plasmas including NIF Ignition Experiments Continuously spatially resolved high resolution x-ray spectra Simple – 1 spherically bent crystal and 1 2D detector Has revolutionized physics of self-generated flows and thermal transport in tokamaks Mature diagnostic – deployed on 7 spectrometers in 4 countries Primary diagnostic for ion temperature and rotation profiles on ITER We are doing the ITER conceptual design Will work for HED plasmas Would allow extensive shielding against neutrons for NIF ignition experiments Two dimensional x-ray imaging Matched pair of spherical crystals 10,000 times larger solid angle than pinhole imager Magnification possible Will allow extensive shielding of line-of-sight neutrons and detector for NIF ignition High efficiency of x-ray detection – no conversion losses 1

Spatially Resolved High Resolution X-Ray Crystal Spectrometry and Radiation Tolerant 2D X-Ray Imaging Schemes for ICF Plasmas including NIF Ignition Experiments

Embed Size (px)

Citation preview

Spatially Resolved High Resolution X-Ray Crystal Spectrometry and Radiation Tolerant 2D X-Ray Imaging Schemes for ICF Plasmas including NIF Ignition Experiments

• Continuously spatially resolved high resolution x-ray spectra

• Simple – 1 spherically bent crystal and 1 2D detector

• Has revolutionized physics of self-generated flows and thermal transport in tokamaks

• Mature diagnostic – deployed on 7 spectrometers in 4 countries

• Primary diagnostic for ion temperature and rotation profiles on ITER

• We are doing the ITER conceptual design

• Will work for HED plasmas

• Would allow extensive shielding against neutrons for NIF ignition experiments

• Two dimensional x-ray imaging

• Matched pair of spherical crystals

• 10,000 times larger solid angle than pinhole imager

• Magnification possible

• Will allow extensive shielding of line-of-sight neutrons and detector for NIF ignition

• High efficiency of x-ray detection – no conversion losses

1

Spherically bent crystal wavelength-disperses x rays horizontally and images them vertically

Bragg relation =2d sin)

Spherically bent crystal

Rotational axis of symmetry

Sagittal focalline

C-Mod spectral image

Verticalfocus

fm=R sin

Astigmatism can be eliminated by using a matched pair of spherically bent crystals

objectimage

Concave crystal

convexcrystal convex-concave pair

provides magnification

Schemes tested in visible light.

To be tested with x rays.

1D imaging x-ray spectrometer and 2D x-ray imager with spherical crystals can be well shielded against NIF neutrons

K. W. Hill OLUG 4/2012 4