FLOOR PLAN FOR “LAB 27”
CORRIDOR
+
-
Zone 1Zone 2
Zone 3 - -
--
- -
+
Zone 1-B
-
Zone 1-A
Zone 2
+
Emergency Exit Only
Zone 3Chem.Lab
-
Direction of Flow in designated Zones
SELECTING FUME HOOD REQUIREMENT
Condition when operating at Max. (24" Hood opening) 17,050 CFM
Condition when operating at Min. 3,850 CFM
CORRIDOR
1750 CFM1500 CFM375 CFM
1500 CFM1200 CFM250 CFM
1700 CFM1450 CFM325 CFM 1500 CFM
1200 CFM250 CFM
1450 CFM1150 CFM200 CFM
1750 CFM1500 CFM
1750 CFM
375 CFM
CFH 10
CFH 4
CFH 4
CFH 3CFH 6
CFH 5CFH 1
CFH 2
CFH 11
CFH 12
CFH 13
CFH 11
CFH 12
CFH 13CFH 2
CFH 3
CFH 6
CFH 7
CFH 7
CFH 8
CFH 10
1350 CFM1150 CFM 225 CFM
CFH 5
1500 CFM
Design Condition ( hoods fully open for loading) 20,600 CFM
+
-
1500 CFM1200 CFM250 CFM
Zone 1
Zone 2
Zone 3 - -
-
-
- -
+
375 CFM
1500 CFM1200 CFM250 CFM
1750 CFM1500 CFM375 CFM
Zone 1-B
CFH 1
1500 CFM1200 CFM250 CFM
-
Zone 1-A
Zone 2
CFH 8
CFH 9CFH 9
1000 CFM900 CFM150 CFM
+
Emergency Exit Only
Zone 3
2
Chem.Lab
-
1) VAV Hoods2)100 FPM FV
GENERAL EXHAUST REQUIREMENT
Condition when operating at Max. (24" Hood opening) 17,050 CFM
Condition when operating at Min. 3,850 CFM
CORRIDOR
1750 CFM1500 CFM375 CFM
1500 CFM1200 CFM250 CFM
1700 CFM1450 CFM325 CFM 1500 CFM
1200 CFM250 CFM
1450 CFM1150 CFM200 CFM
1750 CFM1500 CFM
1750 CFM
375 CFM
CFH 10
CFH 4
CFH 4
CFH 3CFH 6
CFH 5CFH 1
CFH 2
CFH 11
CFH 12
CFH 13
CFH 11
CFH 12
CFH 13CFH 2
CFH 3
CFH 6
CFH 7
CFH 7
CFH 8
CFH 10
1350 CFM1150 CFM 225 CFM
CFH 5
1500 CFM
Design Condition ( hoods fully open for loading) 20,600 CFM
+
-
1500 CFM1200 CFM250 CFM
Zone 1
Zone 2
Zone 3 - -
-
-
- -
+
375 CFM
1500 CFM1200 CFM250 CFM
1750 CFM1500 CFM375 CFM
Zone 1-B
CFH 1
1500 CFM1200 CFM250 CFM
-
Zone 1-A
Zone 2
CFH 8
CFH 9CFH 9
1000 CFM900 CFM150 CFM
+
Emergency Exit Only
Zone 3
3
Chem.Lab
-
1700 CFM1450 CFM325 CFM 1500 CFM
1200 CFM250 CFM
1450 CFM1150 CFM
1750 CFM
375 CFM
CFH 4
CFH 4
CFH 13
CFH 13CFH 2
CFH 3
CFH 6 CFH 8
1500 CFM
1500 CFM
- -
300 CFM300 CFM
300 CFM
Offset Zone 3
375 CFM
Lab Exh. AirLab Exh. Air50 CFM50 CFM
500 CFM300 CFM100 CFM
Required for thermal demand atminimum operating condition.
Condition when operating at Max. (24" Hood opening) 17,050 CFM
Condition when operating at Min. 3,850 CFM
CORRIDOR
Design Condition ( hoods fully open for loading) 20,600 CFM
625 CFM625 CFM625 CFM
+
-
30" x 18"
Lab Sup. Air Zone 3
Lab Sup. AirZone 2
4000 CFM3400 CFM 400 CFM
4025 CFM3225 CFM400 CFM
Zone 1
Zone 2
Zone 3 - -
-
-
- -
300 CFM300 CFM
300 CFM
Offset Zone 3
+
Lab Sup. Air
Transfer Grille for offset
14" x 14"
Offset Supply Air 1725 CFM
Zone 1-B
Lab Sup. Air Zone 1
Lab Sup. Air40" x 25"-
Zone 1-A
Zone 2
40" x 18"
10,250 CFM8300 CFM1125 CFM
100 CFM OFFSET
+
600 CFM400 CFM200 CFM
Chem.Lab
Emergency Exit Only
Zone 3
+1,725 CFM Offset
3,850 CFM Minimum Total
Lab Supply 18,875 CFM Labs
20,600 CFM Design Total
2,125 CFM Labs Min.+1,725 CFM Offset
Sup. Air Chem. Lab.
12" x 12"
1000 CFM 1000 CFM1000 CFM
Transfer Grille for offset
4
FRESH AIR REQUIREMENT with OFFSET
OFFSET IS ALWAYS CONSTANT
Condition when operating at Max. (24" Hood opening) 17,050 CFM
Condition when operating at Min. 3,850 CFM
Design Condition ( hoods fully open for loading) 20,600 CFM
Zone 1
Zone 2
Zone 3 - -
-
-
32,828 CFM Total System Flow198 % Dilution Ratio
3,727 CFM Bypass Flow
St r obi c12,051 CFMEntrainment Flow
5
CORRIDOR625 CFM625 CFM625 CFM
+
-
30" x 18"
4000 CFM3400 CFM 400 CFM
- -
300 CFM300 CFM
300 CFM
Offset Zone 3
+
Lab Sup. Air
Transfer Grille for offset
14" x 14"
Offset Supply Air 1725 CFM
Lab Sup. Air
40" x 25"
-
40" x 18"
100 CFM OFFSET
+
600 CFM400 CFM200 CFM
Chem.Lab
Emergency Exit Only
12" x 12"
1000 CFM 1000 CFM1000 CFM
Transfer Grille for offset
1750 CFM1500 CFM375 CFM
1500 CFM1200 CFM250 CFM
1700 CFM1450 CFM325 CFM 1500 CFM
1200 CFM250 CFM
1450 CFM1150 CFM200 CFM
1750 CFM1500 CFM
1750 CFM
375 CFM
CFH 10
CFH 4
CFH 4
CFH 3CFH 6
CFH 5CFH 1
CFH 2
CFH 11
CFH 12
CFH 13
CFH 11
CFH 12
CFH 13CFH 2
CFH 3
CFH 6
CFH 7
CFH 7
CFH 8
CFH 10
1350 CFM1150 CFM 225 CFM
CFH 5
1500 CFM
+
-
1500 CFM1200 CFM250 CFM
- -
300 CFM300 CFM
300 CFM
Offset Zone 3
+
375 CFM
1500 CFM1200 CFM250 CFM
1750 CFM1500 CFM375 CFM
Zone 1-B
CFH 1
1500 CFM1200 CFM250 CFM
-
Zone 1-A
Zone 2
CFH 8
CFH 9
Lab Exh. AirLab Exh. Air50 CFM50 CFM
CFH 9
1000 CFM900 CFM150 CFM
+
500 CFM300 CFM100 CFM
Zone 3
10,250 CFM8300 CFM1125 CFM
4025 CFM3225 CFM
400 CFM
LAB EXHAUST FAN SELECTION
SELECTING MAKE UP AIR UNIT
Condition when operating at Max. (24" Hood opening) 17,050 CFM
Condition when operating at Min. 3,850 CFM
CORRIDOR
Design Condition ( hoods fully open for loading) 20,600 CFM
625 CFM625 CFM625 CFM
+
-
30" x 18"
Lab Sup. Air Zone 3
Lab Sup. AirZone 2
4000 CFM3400 CFM 400 CFM
4025 CFM3225 CFM400 CFM
Zone 1
Zone 2
Zone 3 - -
-
-
- -
300 CFM300 CFM
300 CFM
Offset Zone 3
+
Lab Sup. Air
Transfer Grille for offset
14" x 14"
Offset Supply Air 1725 CFM
Zone 1-B
Lab Sup. Air Zone 1
Lab Sup. Air40" x 25"-
Zone 1-A
Zone 2
40" x 18"
10,250 CFM8300 CFM1125 CFM
100 CFM OFFSET
+
600 CFM400 CFM200 CFM
Chem.Lab
Emergency Exit Only
Zone 3
+1,725 CFM Offset
3,850 CFM Minimum Total
Lab Supply
18,875 CFM Labs
20,600 CFM Design Total
2,125 CFM Labs Min.+1,725 CFM Offset
Sup. Air Chem. Lab.
12" x 12"
1000 CFM 1000 CFM1000 CFM
Transfer Grille for offset
6
St r obi c
3,850 CFM Min
1 2 4 5 873
(GRN)
(ORG)
6
DX
Coi
l
Hea
t Pip
e
Hea
t Pip
e
Filte
r Ban
k
Outdoor Air Damper
Condenser Coil Cond
ense
r Coi
l
Fixe
d Su
pply
Fan
Hea
t Exc
hang
er
Bypass Return Air Plenum
Plen
um S
/A-R
/A
Plen
um F
/A-R
/ A
21,000 CFM MaxMake up Air
CORRIDOR625 CFM625 CFM625 CFM
+
-
30" x 18"
4000 CFM3400 CFM 400 CFM
- -
300 CFM300 CFM
300 CFM
Offset Zone 3
+
Lab Sup. Air
Transfer Grille for offset
14" x 14"
Offset Supply Air 1725 CFM
Lab Sup. Air
40" x 25"
-
40" x 18"
100 CFM OFFSET
+
600 CFM400 CFM200 CFM
Chem.Lab
Emergency Exit Only
12" x 12"
1000 CFM 1000 CFM1000 CFM
Transfer Grille for offset
1750 CFM1500 CFM375 CFM
1500 CFM1200 CFM250 CFM
1700 CFM1450 CFM325 CFM 1500 CFM
1200 CFM250 CFM
1450 CFM1150 CFM200 CFM
1750 CFM1500 CFM
1750 CFM
375 CFM
CFH 10
CFH 4
CFH 4
CFH 3CFH 6
CFH 5CFH 1
CFH 2
CFH 11
CFH 12
CFH 13
CFH 11
CFH 12
CFH 13CFH 2
CFH 3
CFH 6
CFH 7
CFH 7
CFH 8
CFH 10
1350 CFM1150 CFM 225 CFM
CFH 5
1500 CFM
+
-
1500 CFM1200 CFM250 CFM
- -
300 CFM300 CFM
300 CFM
Offset Zone 3
+
375 CFM
1500 CFM1200 CFM250 CFM
1750 CFM1500 CFM375 CFM
Zone 1-B
CFH 1
1500 CFM1200 CFM250 CFM
-
Zone 1-A
Zone 2
CFH 8
CFH 9
Lab Exh. AirLab Exh. Air50 CFM50 CFM
CFH 9
1000 CFM900 CFM150 CFM
+
500 CFM300 CFM100 CFM
Zone 3
10,250 CFM8300 CFM1125 CFM
4025 CFM3225 CFM
400 CFM
ADDING MAU AND EXF CONTROLS
CORRIDOR625 CFM625 CFM625 CFM
+
-
30" x 18"
Lab Sup. Air Zone 3
Lab Sup. AirZone 2
4000 CFM3400 CFM 400 CFM
4025 CFM3225 CFM400 CFM
- -
300 CFM300 CFM
300 CFM
Offset Zone 3
+
Lab Sup. Air
Transfer Grille for offset
14" x 14"
Offset Supply Air 1725 CFM
Zone 1-B
Lab Sup. Air Zone 1
Lab Sup. Air40" x 25"-
Zone 1-A
Zone 2
40" x 18"
10,250 CFM8300 CFM1125 CFM
100 CFM OFFSET
+
600 CFM400 CFM200 CFM
Chem.Lab
Emergency Exit Only
Zone 3
Sup. Air Chem. Lab.
12" x 12"
1000 CFM 1000 CFM1000 CFM
Transfer Grille for offset
7
St r obi c
1 2 4 5 873
(GRN)
(ORG)
6
DX
Coi
l
Hea
t Pip
e
Hea
t Pip
e
Filte
r Ban
k
Outdoor Air Damper
Condenser Coil Cond
ense
r Coi
l
Fixe
d Su
pply
Fan
Hea
t Exc
hang
er
Bypass Return Air Plenum
Plen
um S
/A-R
/A
Plen
um F
/A-R
/ A
CORRIDOR625 CFM625 CFM625 CFM
+
-
30" x 18"
4000 CFM3400 CFM 400 CFM
- -
300 CFM300 CFM
300 CFM
Offset Zone 3
+
Lab Sup. Air
Transfer Grille for offset
14" x 14"
Lab Sup. Air
40" x 25"
-
40" x 18"
100 CFM OFFSET
+
600 CFM400 CFM200 CFM
Chem.Lab
Emergency Exit Only
12" x 12"
1000 CFM 1000 CFM1000 CFM
Transfer Grille for offset
1750 CFM1500 CFM375 CFM
1500 CFM1200 CFM250 CFM
1700 CFM1450 CFM325 CFM 1500 CFM
1200 CFM250 CFM
1450 CFM1150 CFM200 CFM
1750 CFM1500 CFM
1750 CFM
375 CFM
CFH 10
CFH 4
CFH 4
CFH 3CFH 6
CFH 5CFH 1
CFH 2
CFH 11
CFH 12
CFH 13
CFH 11
CFH 12
CFH 13CFH 2
CFH 3
CFH 6
CFH 7
CFH 7
CFH 8
CFH 10
1350 CFM1150 CFM 225 CFM
CFH 5
1500 CFM
+
-
1500 CFM1200 CFM250 CFM
- -
300 CFM300 CFM
300 CFM
Offset Zone 3
+
375 CFM
1500 CFM1200 CFM250 CFM
1750 CFM1500 CFM375 CFM
Zone 1-B
CFH 1
1500 CFM1200 CFM250 CFM
-
Zone 1-A
Zone 2
CFH 8
CFH 9
Lab Exh. AirLab Exh. Air50 CFM50 CFM
CFH 9
1000 CFM900 CFM150 CFM
+
500 CFM300 CFM100 CFM
Zone 3
10,250 CFM8300 CFM1125 CFM
4025 CFM3225 CFM
400 CFM
Supply Duct Pressure
Bypass Damper Control
Supply Air Temp
Outdoor Air TempFilter Status
Supply Start /Stop
Fan Alarm Failure
1 2 4 5 8
Stop / Start Fan Enable
Exhaust Fan Status
Cooling Status
73
Flame Out Alarm
Heating Status
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(GRN)
(PRP)
(PRP)
(PRP)
(ORG)
(BLUE)
(BLUE)
BACnet Control Panel
6
Outdoor Air Damper
Hea
t Exc
hang
er
Plen
um S
/A-R
/A
0-10 vdc Flow Signal
0 -10 vdc Bypass Damper Status(MAU Control Panel & contact points)
10
10
0-10 vdc Temp Signal09
09
(PRP)
Dehumidify Control Dry Contact on/off
11 Fire Alarm Status(PRP)
(PRP)
(PRP)
(PRP)
100 CFM OFFSET
+
Chem.Lab
ADDING SUPPLY AIR DUCT WORK
Supply Duct Pressure
Bypass Damper Control
Supply Air Temp
Outdoor Air TempFilter Status
Supply Start /Stop
Fan Alarm Failure
1 2 4 5 8
Stop / Start Fan Enable
Exhaust Fan Status
Cooling Status
73
Flame Out Alarm
Heating Status
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(GRN)
(PRP)
(PRP)
(PRP)
(ORG)
(BLUE)
(BLUE)
BACnet Control Panel
CORRIDOR
1750 CFM1500 CFM375 CFM
1500 CFM1200 CFM250 CFM
1700 CFM1450 CFM325 CFM 1500 CFM
1200 CFM250 CFM
1450 CFM1150 CFM200 CFM
1750 CFM1500 CFM
1750 CFM
375 CFM
CFH 10
CFH 4
CFH 4
CFH 3CFH 6
CFH 5CFH 1
CFH 2
CFH 11
CFH 12
CFH 13
CFH 11
CFH 12
CFH 13CFH 2
CFH 3
CFH 6
CFH 7
CFH 7
CFH 8
CFH 10
1350 CFM1150 CFM 225 CFM
CFH 5
1500 CFM
625 CFM
1000 CFM 1000 CFM1000 CFM
625 CFM625 CFM
+
-
6
DX
Coi
l
Hea
t Pip
e
Hea
t Pip
e
Filte
r Ban
k
Outdoor Air Damper
Condenser Coil Cond
ense
r Coi
l
Fixe
d Su
pply
Fan
Hea
t Exc
hang
er
Bypass Return Air Plenum
Plen
um S
/A-R
/A
Plen
um F
/A-R
/ A
Total Lab Sup. Air
0-10 vdc Flow Signal
0 -10 vdc Bypass Damper Status(MAU Control Panel & contact points)
10
30" x 18"
1500 CFM1200 CFM250 CFM
Lab Sup. AirZone 3
Lab Sup. AirZone 2
4000 CFM3400 CFM 400 CFM
4025 CFM3225 CFM400 CFM
- -
300 CFM300 CFM
300 CFM
Offset Zone 3
Offset Zone 2
+
Lab Sup. Air
Transfer Grille for offset Transfer Grille
for offset
375 CFM
10
0-10 vdc Temp Signal09
09
(PRP)
Dehumidify Control Dry Contact on/off
14" x 14"
Offset Supply Air 1725 CFM
11 Fire Alarm Status
1500 CFM1200 CFM250 CFM
1750 CFM1500 CFM375 CFM
(PRP)
(PRP)
(PRP)
(PRP)
Zone 1-B
CFH 1
1500 CFM1200 CFM250 CFM
Lab Sup. Air Zone 1
Lab Sup. Air40" x 25"-
Zone 1-A
Zone 2
40" x 18"
CFH 8
CFH 9
Offset Zone 1
10,250 CFM8300 CFM1125 CFM
Lab Exh. Air
Lab Exh. Air
50 CFM
50 CFM
St r obi c
CFH 9
1000 CFM900 CFM150 CFM
100 CFM OFFSET
+
600 CFM400 CFM200 CFM
500 CFM300 CFM100 CFM
Chem.Lab
Emergency Exit Only
Zone 3
9
20,600 CFM Design17,050 CFM Max 24
3,850 CFM Min
ADDING EXHAUST DUCT WORK
Condition when operating at Max. (24" Hood opening) 17,050 CFM
Condition when operating at Min. 3,850 CFM
Supply Duct Pressure
Exhaust Duct Pressure
Bypass Damper Control
Supply Air Temp
Outdoor Air TempFilter Status
Supply Start /Stop
Fan Alarm Failure
1 2 4 5 8
Stop / Start Fan Enable
Exhaust Fan Status
Cooling Status
73
Flame Out Alarm
Heating Status
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(GRN)
(PRP)
(PRP)
(PRP)
(ORG)
(BLUE)
(BLUE)
BACnet Control Panel
CORRIDOR
1750 CFM1500 CFM375 CFM
1500 CFM1200 CFM250 CFM
1700 CFM1450 CFM325 CFM 1500 CFM
1200 CFM250 CFM
1450 CFM1150 CFM200 CFM
1750 CFM1500 CFM
1750 CFM
375 CFM
CFH 10
CFH 4
CFH 4
CFH 3CFH 6
CFH 5CFH 1
CFH 2
CFH 11
CFH 12
CFH 13
CFH 11
CFH 12
CFH 13CFH 2
CFH 3
CFH 6
CFH 7
CFH 7
CFH 8
CFH 10
1350 CFM1150 CFM 225 CFM
CFH 5
1500 CFM
Design Condition ( hoods fully open for loading) 20,600 CFM
625 CFM
1000 CFM 1000 CFM1000 CFM
625 CFM625 CFM
+
-
D : 32,828 CFM Total System Flow198 % Dilution Ratio
3,727 CFM Bypass Flow
6
DX
Coi
l
Hea
t Pip
e
Hea
t Pip
e
Filte
r Ban
k
Outdoor Air Damper
Condenser Coil Cond
ense
r Coi
l
Fixe
d Su
pply
Fan
Hea
t Exc
hang
er
Bypass Return Air Plenum
Plen
um S
/A-R
/A
Plen
um F
/A-R
/ A
Total Lab Sup. Air
0-10 vdc Flow Signal
0 -10 vdc Bypass Damper Status(MAU Control Panel & contact points)
10
30" x 18"
1500 CFM1200 CFM250 CFM
Lab Sup. AirZone 3
Lab Sup. AirZone 2
4000 CFM3400 CFM 400 CFM
4025 CFM3225 CFM400 CFM
- -
300 CFM300 CFM
300 CFM
Offset Zone 3
Offset Zone 2
+
Lab Sup. Air
Transfer Grille for offset Transfer Grille
for offset
375 CFM
10
0-10 vdc Temp Signal09
09
(PRP)
Dehumidify Control Dry Contact on/off
14" x 14"
Offset Supply Air 1725 CFM
11 Fire Alarm Status
1500 CFM1200 CFM250 CFM
1750 CFM1500 CFM375 CFM
(PRP)
(PRP)
(PRP)
(PRP)
Zone 1-B
CFH 1
1500 CFM1200 CFM250 CFM
Lab Sup. Air Zone 1
Lab Sup. Air40" x 25"-
Zone 1-A
Zone 2
40" x 18"
CFH 8
CFH 9
Offset Zone 1
10,250 CFM8300 CFM1125 CFM
Lab Exh. Air
Lab Exh. Air
50 CFM
50 CFM
St r obi cC:12,051 CFMEntrainment
Flow
CFH 9
17,050 CFM
1000 CFM900 CFM150 CFM
100 CFM OFFSET
+
600 CFM400 CFM200 CFM
500 CFM300 CFM100 CFM
Chem.Lab
Emergency Exit Only
Zone 3
9
Condition when operating at Max. (24" Hood opening) 17,050 CFM
Condition when operating at Min. 3,850 CFM
Supply Duct Pressure
Exhaust Duct Pressure
Bypass Damper Control
Supply Air Temp
Outdoor Air TempFilter Status
Supply Start /Stop
Fan Alarm Failure
1 2 4 5 8
Stop / Start Fan Enable
Exhaust Fan Status
Cooling Status
73
Flame Out Alarm
Heating Status
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(GRN)
(PRP)
(PRP)
(PRP)
(ORG)
(BLUE)
(BLUE)
BACnet Control Panel
CORRIDOR
1750 CFM1500 CFM375 CFM
1500 CFM1200 CFM250 CFM
1700 CFM1450 CFM325 CFM 1500 CFM
1200 CFM250 CFM
1450 CFM1150 CFM200 CFM
1750 CFM1500 CFM
1750 CFM
375 CFM
CFH 10
CFH 4
CFH 4
CFH 3CFH 6
CFH 5CFH 1
CFH 2
CFH 11
CFH 12
CFH 13
CFH 11
CFH 12
CFH 13CFH 2
CFH 3
CFH 6
CFH 7
CFH 7
CFH 8
CFH 10
1350 CFM1150 CFM 225 CFM
CFH 5
1500 CFM
Design Condition ( hoods fully open for loading) 20,600 CFM
625 CFM
1000 CFM 1000 CFM1000 CFM
625 CFM625 CFM
+
-
D : 32,828 CFM Total System Flow198 % Dilution Ratio
3,727 CFM Bypass Flow
6
DX
Coi
l
Hea
t Pip
e
Hea
t Pip
e
Filte
r Ban
k
Outdoor Air Damper
Condenser Coil Cond
ense
r Coi
l
Fixe
d Su
pply
Fan
Hea
t Exc
hang
er
Bypass Return Air Plenum
Plen
um S
/A-R
/A
Plen
um F
/A-R
/ A
Total Lab Sup. Air
0-10 vdc Flow Signal
0 -10 vdc Bypass Damper Status(MAU Control Panel & contact points)
10
30" x 18"
1500 CFM1200 CFM250 CFM
Lab Sup. AirZone 3
Lab Sup. AirZone 2
4000 CFM3400 CFM 400 CFM
4025 CFM3225 CFM400 CFM
- -
300 CFM300 CFM
300 CFM
Offset Zone 3
Offset Zone 2
+
Lab Sup. Air
Transfer Grille for offset Transfer Grille
for offset
375 CFM
10
0-10 vdc Temp Signal09
09
(PRP)
Dehumidify Control Dry Contact on/off
14" x 14"
Offset Supply Air 1725 CFM
11 Fire Alarm Status
1500 CFM1200 CFM250 CFM
1750 CFM1500 CFM375 CFM
(PRP)
(PRP)
(PRP)
(PRP)
Zone 1-B
CFH 1
1500 CFM1200 CFM250 CFM
Lab Sup. Air Zone 1
Lab Sup. Air40" x 25"-
Zone 1-A
Zone 2
40" x 18"
CFH 8
CFH 9
Offset Zone 1
10,250 CFM8300 CFM1125 CFM
Lab Exh. Air
Lab Exh. Air
50 CFM
50 CFM
St r obi cC:12,051 CFMEntrainment
Flow
CFH 9
17,050 CFM
1000 CFM900 CFM150 CFM
100 CFM OFFSET
+
600 CFM400 CFM200 CFM
500 CFM300 CFM100 CFM
Chem.Lab
Emergency Exit Only
Zone 3
10
ADDING GENERAL EXHAUST VALVES
Condition when operating at Max. (24" Hood opening) 17,050 CFM
Condition when operating at Min. 3,850 CFM
PHOENIXFume Hood
ExhaustVAV Valve
PHOENIXFume Hood
ExhaustVAV Valve PHOENIX
Fume Hood Exhaust
VAV Valve
Supply Duct Pressure
Exhaust Duct Pressure
Bypass Damper Control
Supply Air Temp
Outdoor Air TempFilter Status
Supply Start /Stop
Fan Alarm Failure
1 2 4 5 8
Stop / Start Fan Enable
Exhaust Fan Status
Cooling Status
73
Flame Out Alarm
Heating Status
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(GRN)
(PRP)
(PRP)
(PRP)
(ORG)
(BLUE)
(BLUE)
BACnet Control Panel
PHOENIXFume Hood
ExhaustVAV Valve
CORRIDOR
1750 CFM1500 CFM375 CFM
1500 CFM1200 CFM250 CFM
1700 CFM1450 CFM325 CFM 1500 CFM
1200 CFM250 CFM
1450 CFM1150 CFM200 CFM
1750 CFM1500 CFM
1750 CFM
375 CFM
CFH 10
CFH 4
CFH 4
CFH 3CFH 6
CFH 5CFH 1
CFH 2
CFH 11
CFH 12
CFH 13
CFH 11
CFH 12
CFH 13CFH 2
CFH 3
CFH 6
CFH 7
CFH 7
CFH 8
CFH 10
1350 CFM1150 CFM 225 CFM
CFH 5
1500 CFM
Design Condition ( hoods fully open for loading) 20,600 CFM
625 CFM
1000 CFM 1000 CFM1000 CFM
625 CFM625 CFM
+
-
D : 32,828 CFM Total System Flow198 % Dilution Ratio
3,727 CFM Bypass Flow
6
DX
Coi
l
Hea
t Pip
e
Hea
t Pip
e
Filte
r Ban
k
Outdoor Air Damper
Condenser Coil Cond
ense
r Coi
l
Fixe
d Su
pply
Fan
Hea
t Exc
hang
er
Bypass Return Air Plenum
Plen
um S
/A-R
/A
Plen
um F
/A-R
/ A
Total Lab Sup. Air
0-10 vdc Flow Signal
0 -10 vdc Bypass Damper Status(MAU Control Panel & contact points)
10
30" x 18"
1500 CFM1200 CFM250 CFM
Lab Sup. AirZone 3
Lab Sup. AirZone 2
4000 CFM3400 CFM 400 CFM
4025 CFM3225 CFM400 CFM
- -
300 CFM300 CFM
300 CFM
Offset Zone 3
Offset Zone 2
+
Lab Sup. Air
Transfer Grille for offset Transfer Grille
for offset
375 CFM
10
0-10 vdc Temp Signal09
09
(PRP)
Dehumidify Control Dry Contact on/off
14" x 14"
Offset Supply Air 1725 CFM
11 Fire Alarm Status
1500 CFM1200 CFM250 CFM
1750 CFM1500 CFM375 CFM
(PRP)
(PRP)
(PRP)
(PRP)
Zone 1-B
CFH 1
1500 CFM1200 CFM250 CFM
Lab Sup. Air Zone 1
Lab Sup. Air40" x 25"-
Zone 1-A
Zone 2
40" x 18"
PHOENIXVAV
SupplyValve
CFH 8
CFH 9
Offset Zone 1
10,250 CFM8300 CFM1125 CFM
Lab Exh. Air
Lab Exh. Air
50 CFM
50 CFM
St r obi cC:12,051 CFMEntrainment
Flow
CFH 9
17,050 CFM
1000 CFM900 CFM150 CFM
100 CFM OFFSET
+
600 CFM400 CFM200 CFM
500 CFM300 CFM100 CFM
Chem.Lab
Emergency Exit Only
Zone 3
11
EXHAUST VALVES FOR FUME HOODS
COATED VALVES
VAV FUME HOOD CONTOL WITH ZPS
(RED)
(RED) (RED)Supply Duct Pressure
Exhaust Duct Pressure
Bypass Damper Control
Supply Air Temp
Outdoor Air TempFilter Status
Supply Start /Stop
Fan Alarm Failure
1 2 4 5 8
Stop / Start Fan Enable
Exhaust Fan Status
Cooling Status
73
Flame Out Alarm
Heating Status
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(GRN)
(PRP)
(PRP)
(PRP)
(ORG)
(BLUE)
(BLUE)
BACnet Control Panel
CORRIDOR
1750 CFM1500 CFM375 CFM
1500 CFM1200 CFM250 CFM
1700 CFM1450 CFM325 CFM 1500 CFM
1200 CFM250 CFM
1450 CFM1150 CFM200 CFM
1750 CFM1500 CFM
1750 CFM
375 CFM
CFH 10
CFH 4
CFH 4
CFH 3CFH 6
CFH 5CFH 1
CFH 2
CFH 11
CFH 12
CFH 13
CFH 11
CFH 12
CFH 13CFH 2
CFH 3
CFH 6
CFH 7
CFH 7
CFH 8
CFH 10
1350 CFM1150 CFM 225 CFM
CFH 5
1500 CFM
625 CFM
1000 CFM 1000 CFM1000 CFM
625 CFM625 CFM
+
-
6
DX
Coi
l
Hea
t Pip
e
Hea
t Pip
e
Filte
r Ban
k
Outdoor Air Damper
Condenser Coil Cond
ense
r Coi
l
Fixe
d Su
pply
Fan
Hea
t Exc
hang
er
Bypass Return Air Plenum
Plen
um S
/A-R
/A
Plen
um F
/A-R
/ A
0-10 vdc Flow Signal
0 -10 vdc Bypass Damper Status(MAU Control Panel & contact points)
10
30" x 18"
1500 CFM1200 CFM250 CFM
Lab Sup. AirZone 3
Lab Sup. AirZone 2
4000 CFM3400 CFM 400 CFM
4025 CFM3225 CFM400 CFM
- -
300 CFM300 CFM
300 CFM
Offset Zone 3
Offset Zone 2
+
Lab Sup. Air
Transfer Grille for offset Transfer Grille
for offset
375 CFM
10
0-10 vdc Temp Signal09
09
(PRP)
Dehumidify Control Dry Contact on/off
14" x 14"
Offset Supply Air 1725 CFM
11 Fire Alarm Status
1500 CFM1200 CFM250 CFM
1750 CFM1500 CFM375 CFM
(PRP)
(PRP)
(PRP)
(PRP)
Zone 1-B
CFH 1
1500 CFM1200 CFM250 CFM
Lab Sup. Air Zone 1
Lab Sup. Air40" x 25"-
Zone 1-A
Zone 2
40" x 18"
CFH 8
CFH 9
Offset Zone 1
10,250 CFM8300 CFM1125 CFM
Lab Exh. Air
Lab Exh. Air
50 CFM
50 CFM
St r obi c
CFH 9
1000 CFM900 CFM150 CFM
100 CFM OFFSET
+
600 CFM400 CFM200 CFM
500 CFM300 CFM100 CFM
Chem.Lab
Emergency Exit Only
Zone 3
12
CFH 81500 CFM1200 CFM250 CFM
VAV SUPPLY VALVES
Condition when operating at Max. (24" Hood opening) 17,050 CFM
Condition when operating at Min. 3,850 CFM
Supply Duct Pressure
Exhaust Duct Pressure
Bypass Damper Control
Supply Air Temp
Outdoor Air TempFilter Status
Supply Start /Stop
Fan Alarm Failure
1 2 4 5 8
Stop / Start Fan Enable
Exhaust Fan Status
Cooling Status
73
Flame Out Alarm
Heating Status
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(GRN)
(PRP)
(PRP)
(PRP)
(ORG)
(BLUE)
(BLUE)
BACnet Control Panel
CORRIDOR
CFH 10
CFH 4
CFH 3CFH 6
CFH 5CFH 1
CFH 2
CFH 11
CFH 12
CFH 13
CFH 7
Design Condition ( hoods fully open for loading) 20,600 CFM
625 CFM
1000 CFM 1000 CFM1000 CFM
625 CFM625 CFM
+
-
6
DX
Coi
l
Hea
t Pip
e
Hea
t Pip
e
Filte
r Ban
k
Outdoor Air Damper
Condenser Coil Cond
ense
r Coi
l
Fixe
d Su
pply
Fan
Hea
t Exc
hang
er
Bypass Return Air Plenum
Plen
um S
/A-R
/A
Plen
um F
/A-R
/ A
20,600 CFM Design17,050 CFM Max 243,850 CFM Min
Total Lab Sup. Air
0-10 vdc Flow Signal
0 -10 vdc Bypass Damper Status(MAU Control Panel & contact points)
10
30" x 18"Lab Sup. Air
Zone 3
Lab Sup. AirZone 2
4000 CFM3400 CFM 400 CFM
4025 CFM3225 CFM400 CFM
- -
300 CFM300 CFM
300 CFM
Offset Zone 3
Offset Zone 2
+
Lab Sup. Air
Transfer Grille for offset Transfer Grille
for offset
10
0-10 vdc Temp Signal09
09
(PRP)
Dehumidify Control Dry Contact on/off
14" x 14"
Offset Supply Air 1725 CFM
11 Fire Alarm Status(PRP)
(PRP)
(PRP)
(PRP)
Zone 1-B
Lab Sup. Air Zone 1
Lab Sup. Air40" x 25"-
Zone 1-A
Zone 2
40" x 18"
CFH 8
CFH 9
Offset Zone 1
10,250 CFM8300 CFM1125 CFM
Lab Exh. Air
Lab Exh. Air
50 CFM
50 CFM
St r obi c
100 CFM OFFSET
+
600 CFM400 CFM200 CFM
500 CFM300 CFM100 CFM
Chem.Lab
Emergency Exit Only
Zone 3
13
REHEAT COILS > For Zone Control
Supply Duct Pressure
Exhaust Duct Pressure
Bypass Damper Control
Supply Air Temp
Outdoor Air TempFilter Status
Supply Start /Stop
Fan Alarm Failure
1 2 4 5 8
Stop / Start Fan Enable
Exhaust Fan Status
Cooling Status
73
Flame Out Alarm
Heating Status
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(GRN)
(PRP)
(PRP)
(PRP)
(ORG)
(BLUE)
(BLUE)
BACnet Control Panel
CORRIDOR
CFH 10
CFH 4
CFH 3CFH 6
CFH 5CFH 1
CFH 2
CFH 11
CFH 12
CFH 13CFH 2
CFH 7
625 CFM
1000 CFM 1000 CFM1000 CFM
625 CFM625 CFM
+
-
6
DX
Coi
l
Hea
t Pip
e
Hea
t Pip
e
Filte
r Ban
k
Outdoor Air Damper
Condenser Coil Cond
ense
r Coi
l
Fixe
d Su
pply
Fan
Hea
t Exc
hang
er
Bypass Return Air Plenum
Plen
um S
/A-R
/A
Plen
um F
/A-R
/ A
0-10 vdc Flow Signal
0 -10 vdc Bypass Damper Status(MAU Control Panel & contact points)
10
30" x 18"Lab Sup. Air
Zone 3
Lab Sup. AirZone 2
4000 CFM3400 CFM 400 CFM
4025 CFM3225 CFM400 CFM
- -
300 CFM300 CFM
300 CFM
Offset Zone 3
Offset Zone 2
+
Lab Sup. Air
Transfer Grille for offset Transfer Grille
for offset
10
0-10 vdc Temp Signal09
09
(PRP)
Dehumidify Control Dry Contact on/off
14" x 14"
Offset Supply Air 1725 CFM
11 Fire Alarm Status(PRP)
(PRP)
(PRP)
(PRP)
Zone 1-B
Lab Sup. Air Zone 1
Lab Sup. Air40" x 25"-
Zone 1-A
Zone 2
40" x 18"
CFH 8
CFH 9
Offset Zone 1
10,250 CFM8300 CFM1125 CFM
Lab Exh. Air
Lab Exh. Air
50 CFM
50 CFM
St r obi c
100 CFM OFFSET
+
600 CFM400 CFM200 CFM
500 CFM300 CFM100 CFM
Chem.Lab
Emergency Exit Only
Zone 3
14
Supply Duct Pressure
Exhaust Duct Pressure
Bypass Damper Control
Supply Air Temp
Outdoor Air TempFilter Status
Supply Start /Stop
Fan Alarm Failure
1 2 4 5 8
Stop / Start Fan Enable
Exhaust Fan Status
Cooling Status
73
Flame Out Alarm
Heating Status
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(GRN)
(PRP)
(PRP)
(PRP)
(ORG)
(BLUE)
(BLUE)
BACnet Control Panel
CORRIDOR
CFH 10
CFH 4
CFH 3CFH 6
CFH 5CFH 1
CFH 2
CFH 11
CFH 12
CFH 13
CFH 7
CFH 8
625 CFM
1000 CFM 1000 CFM1000 CFM
625 CFM625 CFM
+
-
6
DX
Coi
l
Hea
t Pip
e
Hea
t Pip
e
Filte
r Ban
k
Outdoor Air Damper
Condenser Coil Cond
ense
r Coi
l
Fixe
d Su
pply
Fan
Hea
t Exc
hang
er
Bypass Return Air Plenum
Plen
um S
/A-R
/A
Plen
um F
/A-R
/ A
0-10 vdc Flow Signal
0 -10 vdc Bypass Damper Status(MAU Control Panel & contact points)
10
30" x 18"
1500 CFM1200 CFM250 CFM
Lab Sup. AirZone 3
Lab Sup. AirZone 2
4000 CFM3400 CFM 400 CFM
4025 CFM3225 CFM400 CFM
- -
300 CFM300 CFM
300 CFM
Offset Zone 3
Offset Zone 2
+
Lab Sup. Air
Transfer Grille for offset Transfer Grille
for offset
10
0-10 vdc Temp Signal09
09
(PRP)
Dehumidify Control Dry Contact on/off
14" x 14"
Offset Supply Air 1725 CFM
11 Fire Alarm Status(PRP)
(PRP)
(PRP)
(PRP)
Zone 1-B
Lab Sup. Air Zone 1
Lab Sup. Air40" x 25"-
Zone 1-A
Zone 2
40" x 18"
CFH 8
CFH 9
Offset Zone 1
10,250 CFM8300 CFM1125 CFM
Lab Exh. Air
Lab Exh. Air
50 CFM
50 CFM
St r obi c
100 CFM OFFSET
+
600 CFM400 CFM200 CFM
500 CFM300 CFM100 CFM
Chem.Lab
Emergency Exit Only
Zone 3
15
Temp/Hum Zone
Sensor
ADD TEMPERATURE SENSOR
,,
Supply Duct Pressure
Exhaust Duct Pressure
Bypass Damper Control
Supply Air Temp
Outdoor Air TempFilter Status
Supply Start /Stop
Fan Alarm Failure
1 2 4 5 8
Stop / Start Fan Enable
Exhaust Fan Status
Cooling Status
73
Flame Out Alarm
Heating Status
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(GRN)
(PRP)
(PRP)
(PRP)
(ORG)
(BLUE)
(BLUE)
BACnet Control Panel
CORRIDOR
CFH 10
CFH 4
CFH 3CFH 6
CFH 5CFH 1
CFH 2
CFH 11
CFH 12
CFH 13
CFH 7
CFH 8
625 CFM
1000 CFM 1000 CFM1000 CFM
625 CFM625 CFM
+
-
6
DX
Coi
l
Hea
t Pip
e
Hea
t Pip
e
Filte
r Ban
k
Outdoor Air Damper
Condenser Coil Cond
ense
r Coi
l
Fixe
d Su
pply
Fan
Hea
t Exc
hang
er
Bypass Return Air Plenum
Plen
um S
/A-R
/A
Plen
um F
/A-R
/ A
0-10 vdc Flow Signal
0 -10 vdc Bypass Damper Status(MAU Control Panel & contact points)
10
30" x 18"
1500 CFM1200 CFM250 CFM
Lab Sup. AirZone 3
Lab Sup. AirZone 2
4000 CFM3400 CFM 400 CFM
4025 CFM3225 CFM400 CFM
- -
300 CFM300 CFM
300 CFM
Offset Zone 3
Offset Zone 2
+
Lab Sup. Air
Transfer Grille for offset Transfer Grille
for offset
10
0-10 vdc Temp Signal09
09
(PRP)
Dehumidify Control Dry Contact on/off
14" x 14"
Offset Supply Air 1725 CFM
11 Fire Alarm Status(PRP)
(PRP)
(PRP)
(PRP)
Zone 1-B
Lab Sup. Air Zone 1
Lab Sup. Air40" x 25"-
Zone 1-A
Zone 2
40" x 18"
PHOENIXVAV
SupplyValve
CFH 8
CFH 9
Offset Zone 1
10,250 CFM8300 CFM1125 CFM
Lab Exh. Air
Lab Exh. Air
50 CFM
50 CFM
100 CFM OFFSET
+
Chem.Lab
Emergency Exit OnlyOcc. Sensor
Zone 3
16
* With Occupancy Sensor and Low Speed Valves.
NETWORK WIRING of Chem. Lab.
NETWORK WIRING OF ZONE 3
Supply Duct Pressure
Exhaust Duct Pressure
Bypass Damper Control
Supply Air Temp
Outdoor Air TempFilter Status
Supply Start /Stop
Fan Alarm Failure
1 2 4 5 8
Stop / Start Fan Enable
Exhaust Fan Status
Cooling Status
73
Flame Out Alarm
Heating Status
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(GRN)
(PRP)
(PRP)
(PRP)
(ORG)
(BLUE)
(BLUE)
BACnet Control Panel
CORRIDOR
CFH 10
CFH 4
CFH 3CFH 6
CFH 5CFH 1
CFH 2
CFH 11
CFH 12
CFH 13
CFH 7
625 CFM
1000 CFM 1000 CFM1000 CFM
625 CFM625 CFM
+
-
6
DX
Coi
l
Hea
t Pip
e
Hea
t Pip
e
Filte
r Ban
k
Outdoor Air Damper
Condenser Coil Cond
ense
r Coi
l
Fixe
d Su
pply
Fan
Hea
t Exc
hang
er
Bypass Return Air Plenum
Plen
um S
/A-R
/A
Plen
um F
/A-R
/ A
0-10 vdc Flow Signal
0 -10 vdc Bypass Damper Status(MAU Control Panel & contact points)
10
30" x 18"Lab Sup. Air
Zone 3
Lab Sup. AirZone 2
4000 CFM3400 CFM 400 CFM
4025 CFM3225 CFM400 CFM
- -
300 CFM300 CFM
300 CFM
Offset Zone 3
Offset Zone 2
+
Lab Sup. Air
Transfer Grille for offset Transfer Grille
for offset
10
0-10 vdc Temp Signal09
09
(PRP)
Dehumidify Control Dry Contact on/off
14" x 14"
Offset Supply Air 1725 CFM
11 Fire Alarm Status(PRP)
(PRP)
(PRP)
(PRP)
Zone 1-B
Lab Sup. Air Zone 1
Lab Sup. Air40" x 25"-
Zone 1-A
Zone 2
40" x 18"
CFH 8
CFH 9
Offset Zone 1
10,250 CFM8300 CFM1125 CFM
Lab Exh. Air
Lab Exh. Air
50 CFM
50 CFM
St r obi c
100 CFM OFFSET
+
600 CFM400 CFM200 CFM
500 CFM300 CFM100 CFM
Chem.Lab
Emergency Exit Only
Occ. Sensor
Zone 3
17
Local Network
Supply Duct Pressure
Exhaust Duct Pressure
Bypass Damper Control
Supply Air Temp
Outdoor Air TempFilter Status
Supply Start /Stop
Fan Alarm Failure
1 2 4 5 8
Stop / Start Fan Enable
Exhaust Fan Status
Cooling Status
73
Flame Out Alarm
Heating Status
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(GRN)
(PRP)
(PRP)
(PRP)
(ORG)
(BLUE)
(BLUE)
BACnet Control Panel
CORRIDOR
CFH 10
CFH 4
CFH 3CFH 6
CFH 5CFH 1
CFH 2
CFH 11
CFH 12
CFH 13
CFH 7
CFH 8
625 CFM
1000 CFM 1000 CFM1000 CFM
625 CFM625 CFM
+
-
6
DX
Coi
l
Hea
t Pip
e
Hea
t Pip
e
Filte
r Ban
k
Outdoor Air Damper
Condenser Coil Cond
ense
r Coi
l
Fixe
d Su
pply
Fan
Hea
t Exc
hang
er
Bypass Return Air Plenum
Plen
um S
/A-R
/A
Plen
um F
/A-R
/ A
Total Lab Sup. Air
0-10 vdc Flow Signal
0 -10 vdc Bypass Damper Status(MAU Control Panel & contact points)
10
30" x 18"Lab Sup. Air
Zone 3
Lab Sup. AirZone 2
4000 CFM3400 CFM 400 CFM
4025 CFM3225 CFM400 CFM
- -
300 CFM300 CFM
300 CFM
Offset Zone 3
Offset Zone 2
+
Lab Sup. Air
Transfer Grille for offset Transfer Grille
for offset
10
0-10 vdc Temp Signal09
09
(PRP)
Dehumidify Control Dry Contact on/off
14" x 14"
Offset Supply Air 1725 CFM
11 Fire Alarm Status(PRP)
(PRP)
(PRP)
(PRP)
Lab Sup. Air Zone 1
Lab Sup. Air40" x 25"-
Zone 2
40" x 18"
Zone 3
CFH 8
CFH 9
Offset Zone 1
10,250 CFM8300 CFM1125 CFM
Lab Exh. Air
Lab Exh. Air
50 CFM
50 CFM
St r obi c
100 CFM OFFSET
+
600 CFM400 CFM200 CFM
500 CFM300 CFM100 CFM
Chem.Lab
Emergency Exit Only
Occ. Sensor
Zone1-B
Zone1-A
18
NETWORK WIRING OF ZONE 1
Local Network
WIRING OF OFFSET
Supply Duct Pressure
Exhaust Duct Pressure
Bypass Damper Control
Supply Air Temp
Outdoor Air TempFilter Status
Supply Start /Stop
Fan Alarm Failure
1 2 4 5 8
Stop / Start Fan Enable
Exhaust Fan Status
Cooling Status
73
Flame Out Alarm
Heating Status
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(GRN)
(PRP)
(PRP)
(PRP)
(ORG)
(BLUE)
(BLUE)
BACnet Control Panel
CORRIDOR
CFH 10
CFH 4
CFH 3CFH 6
CFH 5CFH 1
CFH 2
CFH 11
CFH 12
CFH 13
CFH 7
CFH 8
625 CFM
1000 CFM 1000 CFM1000 CFM
625 CFM625 CFM
+
-
6
DX
Coi
l
Hea
t Pip
e
Hea
t Pip
e
Filte
r Ban
k
Outdoor Air Damper
Condenser Coil Cond
ense
r Coi
l
Fixe
d Su
pply
Fan
Hea
t Exc
hang
er
Bypass Return Air Plenum
Plen
um S
/A-R
/A
Plen
um F
/A-R
/ A
Total Lab Sup. Air
0-10 vdc Flow Signal
0 -10 vdc Bypass Damper Status(MAU Control Panel & contact points)
10
30" x 18"Lab Sup. Air
Zone 3
Lab Sup. AirZone 2
4000 CFM3400 CFM 400 CFM
4025 CFM3225 CFM400 CFM
- -
300 CFM300 CFM
300 CFM
Offset Zone 3
Offset Zone 2
+
Lab Sup. Air
Transfer Grille for offset Transfer Grille
for offset
10
0-10 vdc Temp Signal09
09
(PRP)
Dehumidify Control Dry Contact on/off
14" x 14"
Offset Supply Air 1725 CFM
11 Fire Alarm Status(PRP)
(PRP)
(PRP)
(PRP)
Lab Sup. Air Zone 1
Lab Sup. Air40" x 25"-
Zone 2
40" x 18"
Zone 3
CFH 8
CFH 9
Offset Zone 1
10,250 CFM8300 CFM1125 CFM
Lab Exh. Air
Lab Exh. Air
50 CFM
50 CFM
St r obi c
100 CFM OFFSET
+
600 CFM400 CFM200 CFM
500 CFM300 CFM100 CFM
Chem.Lab
Emergency Exit Only
Occ. Sensor
Zone1-B
Zone1-A
19
To Local Network
Supply Duct Pressure
Exhaust Duct Pressure
Bypass Damper Control
Supply Air Temp
Outdoor Air TempFilter Status
Supply Start /Stop
Fan Alarm Failure
1 2 4 5 8
Stop / Start Fan Enable
Exhaust Fan Status
Cooling Status
73
Flame Out Alarm
Heating Status
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(GRN)
(PRP)
(PRP)
(PRP)
(ORG)
(BLUE)
(BLUE)
BACnet Control Panel
CORRIDOR
CFH 10
CFH 4
CFH 3CFH 6
CFH 5CFH 1
CFH 2
CFH 11
CFH 12
CFH 13
CFH 7
625 CFM
1000 CFM 1000 CFM1000 CFM
625 CFM625 CFM
+
-
6
DX
Coi
l
Hea
t Pip
e
Hea
t Pip
e
Filte
r Ban
k
Outdoor Air Damper
Condenser Coil Cond
ense
r Coi
l
Fixe
d Su
pply
Fan
Hea
t Exc
hang
er
Bypass Return Air Plenum
Plen
um S
/A-R
/A
Plen
um F
/A-R
/ A
Total Lab Sup. Air
0-10 vdc Flow Signal
0 -10 vdc Bypass Damper Status(MAU Control Panel & contact points)
10
30" x 18"Lab Sup. Air
Zone 3
Lab Sup. AirZone 2
4000 CFM3400 CFM 400 CFM
4025 CFM3225 CFM400 CFM
- -
300 CFM300 CFM
300 CFM
Offset Zone 3
Offset Zone 2
+
Lab Sup. Air
Transfer Grille for offset Transfer Grille
for offset
10
0-10 vdc Temp Signal09
09
(PRP)
Dehumidify Control Dry Contact on/off
14" x 14"
Offset Supply Air 1725 CFM
11 Fire Alarm Status(PRP)
(PRP)
(PRP)
(PRP)
Zone 1-B
Lab Sup. Air Zone 1
Lab Sup. Air40" x 25"-
Zone 1-A
40" x 18"
Zone 3
CFH 8
CFH 9
Offset Zone 1
10,250 CFM8300 CFM1125 CFM
Lab Exh. Air
Lab Exh. Air
50 CFM
50 CFM
St r obi c
100 CFM OFFSET
+
600 CFM400 CFM200 CFM
500 CFM300 CFM100 CFM
Chem.Lab
Emergency Exit Only
Occ. Sensor
Zone 2
20
NETWORK WIRING OF ZONE 2
Local Network
Supply Duct Pressure
Exhaust Duct Pressure
Bypass Damper Control
Supply Air Temp
Outdoor Air TempFilter Status
Supply Start /Stop
Fan Alarm Failure
1 2 4 5 8
Stop / Start Fan Enable
Exhaust Fan Status
Cooling Status
73
Flame Out Alarm
Heating Status
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(GRN)
(PRP)
(PRP)
(PRP)
(ORG)
(BLUE)
(BLUE)
BACnet Control Panel
CORRIDOR
CFH 10
CFH 4
CFH 3CFH 6
CFH 5CFH 1
CFH 2
CFH 11
CFH 12
CFH 13
CFH 7
625 CFM
1000 CFM 1000 CFM1000 CFM
625 CFM625 CFM
+
-
6
DX
Coi
l
Hea
t Pip
e
Hea
t Pip
e
Filte
r Ban
k
Outdoor Air Damper
Condenser Coil Cond
ense
r Coi
l
Fixe
d Su
pply
Fan
Hea
t Exc
hang
er
Bypass Return Air Plenum
Plen
um S
/A-R
/A
Plen
um F
/A-R
/ A
Total Lab Sup. Air
0-10 vdc Flow Signal
0 -10 vdc Bypass Damper Status(MAU Control Panel & contact points)
10
30" x 18"Lab Sup. Air
Zone 3
Lab Sup. AirZone 2
4000 CFM3400 CFM 400 CFM
4025 CFM3225 CFM400 CFM
- -
300 CFM300 CFM
300 CFM
Offset Zone 3
Offset Zone 2
+
Lab Sup. Air
Transfer Grille for offset Transfer Grille
for offset
10
0-10 vdc Temp Signal09
09
(PRP)
Dehumidify Control Dry Contact on/off
14" x 14"
Offset Supply Air 1725 CFM
11 Fire Alarm Status(PRP)
(PRP)
(PRP)
(PRP)
Zone 1-B
Lab Sup. Air Zone 1
Lab Sup. Air40" x 25"-
Zone 1-A
40" x 18"
Zone 3
CFH 8
CFH 9
Offset Zone 1
10,250 CFM8300 CFM1125 CFM
Lab Exh. Air
Lab Exh. Air
50 CFM
50 CFM
St r obi c
100 CFM OFFSET
+
600 CFM400 CFM200 CFM
500 CFM300 CFM100 CFM
Chem.Lab
Emergency Exit Only
Occ. Sensor
Zone 2
21
LDU Local Display Controller
ADDING OPPTIONAL LDU
To Local Network
Supply Duct Pressure
Exhaust Duct Pressure
Bypass Damper Control
Supply Air Temp
Outdoor Air TempFilter Status
Supply Start /Stop
Fan Alarm Failure
1 2 4 5 8
Stop / Start Fan Enable
Exhaust Fan Status
Cooling Status
73
Flame Out Alarm
Heating Status
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(GRN)
(PRP)
(PRP)
(PRP)
(ORG)
(BLUE)
(BLUE)
BACnet Control Panel
CORRIDOR
CFH 10
CFH 4
CFH 3CFH 6
CFH 5CFH 1
CFH 2
CFH 11
CFH 12
CFH 13
CFH 7
+
-
6
DX
Coi
l
Hea
t Pip
e
Hea
t Pip
e
Filte
r Ban
k
Outdoor Air Damper
Condenser Coil Cond
ense
r Coi
l
Fixe
d Su
pply
Fan
Hea
t Exc
hang
er
Bypass Return Air Plenum
Plen
um S
/A-R
/A
Plen
um F
/A-R
/ A
0-10 vdc Flow Signal
0 -10 vdc Bypass Damper Status(MAU Control Panel & contact points)
10
- -
+
Lab Sup. Air
10
0-10 vdc Temp Signal
LabRouter
09
09
(PRP)
Dehumidify Control Dry Contact on/off
To BACnet Lab
Server
11 Fire Alarm Status(PRP)
(PRP)
(PRP)
(PRP)
Zone 1-B
-
Zone 1-A
Zone 2
Zone 3
CFH 8
CFH 9
St r obi c
+
Chem.Lab
22
ADDING ROUTERS
Routing all Local Networks
ADDING BMS Network
Supply Duct Pressure
Exhaust Duct Pressure
Bypass Damper Control
Supply Air Temp
Outdoor Air TempFilter Status
Supply Start /Stop
Fan Alarm Failure
1 2 4 5 8
Stop / Start Fan Enable
Exhaust Fan Status
Cooling Status
73
Flame Out Alarm
Heating Status
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(GRN)
(PRP)
(PRP)
(PRP)
(ORG)
(BLUE)
(BLUE)
BACnet Control Panel
CORRIDOR
CFH 10
CFH 4
CFH 3CFH 6
CFH 5CFH 1
CFH 2
CFH 11
CFH 12
CFH 13
CFH 7
+
-
6
DX
Coi
l
Hea
t Pip
e
Hea
t Pip
e
Filte
r Ban
k
Outdoor Air Damper
Condenser Coil Cond
ense
r Coi
l
Fixe
d Su
pply
Fan
Hea
t Exc
hang
er
Bypass Return Air Plenum
Plen
um S
/A-R
/A
Plen
um F
/A-R
/ A
0-10 vdc Flow Signal
0 -10 vdc Bypass Damper Status(MAU Control Panel & contact points)
10
- -
+
Lab Sup. Air
10
0-10 vdc Temp Signal
PHOENIX
Router
09
09
(PRP)
Dehumidify Control Dry Contact on/off
To BACnet Macro Server
11 Fire Alarm Status(PRP)
(PRP)
(PRP)
(PRP)
Zone 1-B
-
Zone 1-A
Zone 2
Zone 3
CFH 8
CFH 9
St r obi c
+
Chem.Lab
23
Internetor
Intranet
BMS SERVER
Connection MAU , EXF,.... To BMS
Internetor
Intranet
BMSServer
Supply Duct Pressure
Exhaust Duct Pressure
Bypass Damper Control
Supply Air Temp
Outdoor Air TempFilter Status
Supply Start /Stop
Fan Alarm Failure
1 2 4 5 8
Stop / Start Fan Enable
Exhaust Fan Status
Cooling Status
73
Flame Out Alarm
Heating Status
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(GRN)
(PRP)
(PRP)
(PRP)
(ORG)
(BLUE)
(BLUE)
BACnet Control Panel
CORRIDOR
CFH 10
CFH 4
CFH 3CFH 6
CFH 5CFH 1
CFH 2
CFH 11
CFH 12
CFH 13
CFH 7
+
-
6
DX
Coi
l
Hea
t Pip
e
Hea
t Pip
e
Filte
r Ban
k
Outdoor Air Damper
Condenser Coil Cond
ense
r Coi
l
Fixe
d Su
pply
Fan
Hea
t Exc
hang
er
Bypass Return Air Plenum
Plen
um S
/A-R
/A
Plen
um F
/A-R
/ A
0-10 vdc Flow Signal
0 -10 vdc Bypass Damper Status(MAU Control Panel & contact points)
10
- -
+
Lab Sup. Air
10
0-10 vdc Temp Signal
LabRouter
09
09
(PRP)
Dehumidify Control Dry Contact on/off
To BACnet Lab
Server
11 Fire Alarm Status(PRP)
(PRP)
(PRP)
(PRP)
Zone 1-B
-
Zone 1-A
Zone 2
Zone 3
CFH 8
CFH 9
St r obi c
+
Chem.Lab
24
Internetor
Intranet
Internetor
Intranet
Supply Duct Pressure
Exhaust Duct Pressure
Bypass Damper Control
Supply Air Temp
Outdoor Air TempFilter Status
Supply Start /Stop
Fan Alarm Failure
1 2 4 5 8
Stop / Start Fan Enable
Exhaust Fan Status
Cooling Status
73
Flame Out Alarm
Heating Status
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(GRN)
(PRP)
(PRP)
(PRP)
(ORG)
(BLUE)
(BLUE)
BACnet Control Panel
CORRIDOR
CFH 10
CFH 4
CFH 3CFH 6
CFH 5CFH 1
CFH 2
CFH 11
CFH 12
CFH 13
CFH 7
+
-
6
DX
Coi
l
Hea
t Pip
e
Hea
t Pip
e
Filte
r Ban
k
Outdoor Air Damper
Condenser Coil Cond
ense
r Coi
l
Fixe
d Su
pply
Fan
Hea
t Exc
hang
er
Bypass Return Air Plenum
Plen
um S
/A-R
/A
Plen
um F
/A-R
/ A
0-10 vdc Flow Signal
0 -10 vdc Bypass Damper Status(MAU Control Panel & contact points)
10
- -
+
Lab Sup. Air
10
0-10 vdc Temp Signal
LabRouter
09
09
(PRP)
Dehumidify Control Dry Contact on/off
To BACnet Lab
Server
11 Fire Alarm Status(PRP)
(PRP)
(PRP)
(PRP)
Zone 1-B
-
Zone 1-A
Zone 2
Zone 3
CFH 8
CFH 9
St r obi c
+
Chem.Lab
25 BACnet Lab Server “ Macro Server”
ADDING SERVER FOR LAB “INTERFACE”
ROUTING INTERFACE TO BMS
Internetor
Intranet
Supply Duct Pressure
Exhaust Duct Pressure
Bypass Damper Control
Supply Air Temp
Outdoor Air TempFilter Status
Supply Start /Stop
Fan Alarm Failure
1 2 4 5 8
Stop / Start Fan Enable
Exhaust Fan Status
Cooling Status
73
Flame Out Alarm
Heating Status
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(GRN)
(PRP)
(PRP)
(PRP)
(ORG)
(BLUE)
(BLUE)
BACnet Control Panel
CORRIDOR
CFH 10
CFH 4
CFH 3CFH 6
CFH 5CFH 1
CFH 2
CFH 11
CFH 12
CFH 13
CFH 7
+
-
6
DX
Coi
l
Hea
t Pip
e
Hea
t Pip
e
Filte
r Ban
k
Outdoor Air Damper
Condenser Coil Cond
ense
r Coi
l
Fixe
d Su
pply
Fan
Hea
t Exc
hang
er
Bypass Return Air Plenum
Plen
um S
/A-R
/A
Plen
um F
/A-R
/ A
0-10 vdc Flow Signal
0 -10 vdc Bypass Damper Status(MAU Control Panel & contact points)
10
- -
+
Lab Sup. Air
10
0-10 vdc Temp Signal
LabRouter
09
09
(PRP)
Dehumidify Control Dry Contact on/off
To BACnet Lab Server
11 Fire Alarm Status(PRP)
(PRP)
(PRP)
(PRP)
Zone 1-B
-
Zone 1-A
Zone 2
Zone 3
CFH 8
CFH 9
St r obi c
+
Chem.Lab
26 BACnet Macro Server
Sample LAB / BMS Network Interface
From Lab Router
TO FH - 7
To Router
Alm. Help?
Fire Seq. StrbcPrg.H PTTR
Map
Belnor Engineering Inc.
Laboratory Lab 13000 Zone 3 BRoom Supply Air Flow 3800 CFM
Room Supply Set Point 3800 CFM
Static Pressure Alarm Normal
Room Temp St Pt. 21.6°C
Room Temp . 20.8°C
Room Temp Adj. 20.9°C
Fume Hood Exhaust Flow 1200 CFM
Fume Hood Exhaust Set Point 1200 CFM
Static Pressure Alarm Normal
Sash Opening Percent 100%
Hood Flow Command 1200 CFM
Hood Flow Feedback 1200 CFM
Face Velocity 100 FPM
FV Set Point 100 FPM
Sash Cable Alarm Normal
Occ/ UnOcc Occ
ZPS Sensor OFF
Sash Opening Percent 100%
Hood Flow Command 1500 CFM
Hood Flow Feedback 1500 CFM
Face Velocity 100 FPM
FV Set Point 100 FPM
Sash Cable Alarm Normal
Occ/ UnOcc Occ
ZPS Sensor OFF
Fume Hood Exhaust Set Point 1500 CFM
Static Pressure Alarm Normal
Fume Hood Exhaust Flow 1500 CFM
Coated Valve >
VAV Hood
CFH - 10
Exhaust Air< Coated Valve
VAV Hood
CFH - 9
Exhaust Air
Supply Air
Supply Air
Condition when operating at Max. (24" Hood opening) 17,050 CFM
Condition when operating at Min. 3,850 CFM
(RED)
Supply Duct Pressure
Exhaust Duct Pressure
Bypass Damper Control
Supply Air Temp
Outdoor Air TempFilter Status
Supply Start /Stop
3,850 CFM Min
Fan Alarm Failure
1 2 4 5 8
Stop / Start Fan Enable
Exhaust Fan Status
Cooling Status
73
Flame Out Alarm
Heating Status
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(BLUE)
(GRN)
(PRP)
(PRP)
(PRP)
(ORG)
(BLUE)
(BLUE)
BACnet Control Panel
CORRIDOR
1750 CFM1500 CFM375 CFM
1500 CFM1200 CFM250 CFM
1700 CFM1450 CFM325 CFM 1500 CFM
1200 CFM250 CFM
1450 CFM1150 CFM200 CFM
1750 CFM1500 CFM
1750 CFM
375 CFM
CFH 10
CFH 4
CFH 4
CFH 3CFH 6
CFH 5CFH 1
CFH 2
CFH 11
CFH 12
CFH 13
CFH 11
CFH 12
CFH 13CFH 2
CFH 3
CFH 6
CFH 7
CFH 7
CFH 8
CFH 10
1350 CFM1150 CFM 225 CFM
CFH 5
1500 CFM
Design Condition ( hoods fully open for loading) 20,600 CFM
625 CFM
1000 CFM 1000 CFM1000 CFM
625 CFM625 CFM
+
-
Phoenix LDU Local Display Controller
D : 32,828 CFM Total System Flow198 % Dilution Ratio
3,727 CFM Bypass Flow
21,000 CFM Max
6
Sample LAB / BMS Network Interface
Make up Air
DX
Coi
l
Hea
t Pip
e
Hea
t Pip
e
Filte
r Ban
k
Outdoor Air Damper
Condenser Coil Cond
ense
r Coi
l
Fixe
d Su
pply
Fan
Hea
t Exc
hang
er
Bypass Return Air Plenum
Plen
um S
/A-R
/A
Plen
um F
/A-R
/ A
Total Lab Sup. Air
0-10 vdc Flow Signal
0 -10 vdc Bypass Damper Status(MAU Control Panel & contact points)
10
30" x 18"
1500 CFM1200 CFM250 CFM
Lab Sup. AirZone 3
Lab Sup. AirZone 2
4000 CFM3400 CFM 400 CFM
4025 CFM3225 CFM400 CFM
Zone 1
Zone 2
Zone 3 - -
-
-
- -
300 CFM300 CFM
300 CFM
Offset Zone 3
Offset Zone 2
+
Lab Sup. Air
Transfer Grille for offset Transfer Grille
for offset
375 CFM
10
0-10 vdc Temp Signal
LabRouter
Temp/Hum Zone
Sensor
09
09
(PRP)
Dehumidify Control Dry Contact on/off
To BACnet Lab
Server
14" x 14"
Offset Supply Air 1725 CFM
11 Fire Alarm Status
1500 CFM1200 CFM250 CFM
1750 CFM1500 CFM375 CFM
BACnet Macro Server
From LabRouter
Internetor
Intranet
(PRP)
(PRP)
(PRP)
(PRP)
TO FH - 7
To Router
Alm. Help?
Fire Seq. StrbcPrg.H PTTR
Map
Belnor Engineering Inc.
Laboratory Lab 13000 Zone 3 BRoom Supply Air Flow 3800 CFM
Room Supply Set Point 3800 CFM
Static Pressure Alarm Normal
Room Temp St Pt. 21.6°C
Room Temp . 20.8°C
Room Temp Adj. 20.9°C
Fume Hood Exhaust Flow 1200 CFM
Fume Hood Exhaust Set Point 1200 CFM
Static Pressure Alarm Normal
Sash Opening Percent 100%
Hood Flow Command 1200 CFM
Hood Flow Feedback 1200 CFM
Face Velocity 100 FPM
FV Set Point 100 FPM
Sash Cable Alarm Normal
Occ/ UnOcc Occ
ZPS Sensor OFF
Sash Opening Percent 100%
Hood Flow Command 1500 CFM
Hood Flow Feedback 1500 CFM
Face Velocity 100 FPM
FV Set Point 100 FPM
Sash Cable Alarm Normal
Occ/ UnOcc Occ
ZPS Sensor OFF
Fume Hood Exhaust Set Point 1500 CFM
Static Pressure Alarm Normal
Fume Hood Exhaust Flow 1500 CFM
Coated Valve >
VAV Hood
CFH - 10
Exhaust Air< Coated Valve
VAV Hood
CFH - 9
Exhaust Air
Supply Air
Supply Air
Zone 1-B
CFH 1
1500 CFM1200 CFM250 CFM
Lab Sup. Air Zone 1
Lab Sup. Air40" x 25"-
Zone 1-A
Zone 2
40" x 18"
Zone 3
CFH 8
CFH 9
Offset Zone 1
10,250 CFM8300 CFM1125 CFM
Lab Exh. Air
Lab Exh. Air
50 CFM
50 CFM
BMS Server
St r obi cC:12,051 CFMEntrainment
Flow
CFH 9
17,050 CFM
1000 CFM900 CFM150 CFM
100 CFM OFFSET
+
600 CFM400 CFM200 CFM
500 CFM300 CFM100 CFM
Chem.Lab
Emergency Exit Only
Occ. Sensor
+1,725 CFM Offset
3,850 CFM Minimum Total
18,875 CFM Labs
20,600 CFM Design Total 2,125 CFM Labs Min.
+1,725 CFM Offset
27