Upload
others
View
2
Download
0
Embed Size (px)
Citation preview
Safety aspects of a new Hybrid
Angio-MR: P3-Stroke-Horizon 2020 project of EIT Health
Michiel Dehairs1, Nicholas Marshall1,2, Hilde Bosmans1,2
1 Medical Imaging Research Center, Medical Physics and Quality Assessment , Katholieke Universiteit Leuven, 3000 Leuven, Belgium2 Department of Radiology, UZ Gasthuisberg, Herestraat 49, 3000 Leuven, Belgium
The concepts and information presented in this presentation are based on research and are not commercially available
Angio-MR system
• Hybrid system combining an MRI with an Angio x-ray source and
detector
• One combined system not two systems next to each other
Siemens nexaris Angio-CT
© Siemens Healthcare GmbH, 2018
Angio-MR system
Mock-up of the system
18426 P3-Stroke - Predictive Prevention and Personalized Interventional Stroke
Therapy
• Project partners
• UK Erlangen, Prof Arnd Dörfler
• FAU Erlangen, Prof Andreas Maier
• KU Leuven, Prof Hilde Bosmans
• Univ Bordeaux *, Prof Pierre Jais
• ESTESC Coimbra * Prof Graciano Paulo
• Siemens Healthcare GmbH, Dr Martino Leghissa
• * External project partners
Purpose• Increasing the efficiency of Stroke treatment
o Stroke remains the second leading cause of death worldwide
o Rapid treatment may minimize the long-term effects of a stroke and prevent death
o By combining both imaging modalities into one hybrid system it is possible to reduce treatment
time significantly no more time delay between diagnosis and treatment
o Immediate verification whether treatment is successful
• Electro Physiology (EP)
• Interventional oncology
• Percutaneous RF ablation
• Trans Arterial Chemo-Embolization (TACE)
• Selective Internal Radio Therapy (SIRT)
Take advantage of the Hybrid System
• Make use of the MRI system when possible
• Find ways to reduce the use of X-rays
immediate reduction of dose
• Task for the physicians to help with this
• MRI should/could replace CT scan when possible
Safety
• It’s a hybrid system look at the safety measures for:
o MRI
o Angio
• MRI safety: deal with magnetic fields
• Angio safety deal with the X-rays
Magnetic Safety
• Testing of MR compatibility and safety of neuro-interventional devices
• Metallic coating of catheter artifacts in MRI (similar for guidewires)
Various stent retrievers Guidewires Catheters
X-ray Safety
• Under couch vs Over couch configuration
patient eye lens becomes a risk factor
Eye lens dose
• Threshold for developing lens opacities: 0.5 - 2.0 Gy
• Threshold for developing cataract: 5.0 Gy
The Peak Skin Dose (PSD),
averaged over 60 patients
undergoing a thrombectomy
procedure in our hospital was
0.26 Gy with a maximum value of
1.97 Gy. In total four patients had
a PSD above 0.5 Gy.
Eye lens dose
Simulation TLD Measurements
Eye lens dose Results• Simulation ratio Anterior-Posterior/Posterior-Anterior:
o Small FOV: ~4
o Large FOV: ~40
• TLD Measurement ratio AP/PA:
o Small FOV: ~4
o Large FOV: ~35
• Bismuth eye shield lowered eye lens dose by factor ~3
Eye lens dose Conclusion• Patient eye lens dose is significantly increased in the Angio-MR system
meaning necessary precautions need to be considered:
good collimation, eye shields, head tilt, larger primary angulations,…
Secondary
angulation is not
possible with the
Angio-MRI
Tilting the head
when possible
removes the eyes
from the primary
FOV
Physician Dose
• Comparative study of the scattered radiation towards the primary
physician was performed
Bore exit
Safe position for theinterventionalist
Patient table
3D Dose Distribution(eV/g) per history
Patient table
No lead shields in thissimulation
Head
Head
Feet
Feet
3D Dose Distribution(eV/g) per history
Physician Dose Results:• The scattered radiation towards the physician reduced by a factor: 2 - 4
• When taking a step back the scatter reduced a further factor of 15 for the
Angio-MRI setup
• After adding both a ceiling mounted and table mounted radiation shield in the
conventional setup, the MRI setup had factor 2 lower dose
• The head of the physician is also protected when standing close enough to the
MRI bore exit
possible reduction in eye lens dose
Physician Dose Conclusion
• The presence of the MRI can serve as a radiation shield
• When possible, the medical personnel should step outside the
scattered x-ray field during irradiation
• Adding a ceiling mounted shield could reduce the scatter even more
• Physician eye lens dose could potentially be lowered
Thank You!