User Manual Release 1.2 Copyright2017 Seongsan Lab, Seoul, South Korea This manual is protected under international copyright laws. No part of this manual may be reproduced, distributed, translated, or transmitted in any form or by any means, electronic, magnetic, optical, chemical, manual or otherwise, without the prior written permission of Seongsan Lab. All rights reserved DEEPSOUND M3 www.seongsanlab.com DEEPSOUND M3 - Easy, Precise, Fast Phased Array UT device ®

Release 1.2 Sdf sdf sdf sdf sdf DEEPSOUND M3

  • Upload
    others

  • View
    43

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Release 1.2 Sdf sdf sdf sdf sdf DEEPSOUND M3

User Manual

Release 1.2

Copyrightⓒ 2017 Seongsan Lab, Seoul, South Korea This manual is protected under international copyright laws. No part of this manual may be reproduced, distributed, translated, or transmitted in any form or by any means, electronic, magnetic, optical, chemical, manual or otherwise, without the prior written permission of Seongsan Lab.

All rights reserved

DEEPSOUND M3

www.seongsanlab.com

Sdf sdf sdf sdf sdf

DEEPSOUND M3 - Easy, Precise, Fast Phased Array UT device

®

Page 2: Release 1.2 Sdf sdf sdf sdf sdf DEEPSOUND M3

Release 1.2, 2017

Copyright⒞ Seongsan Lab. All rights Reserved 1

About this user manual Manual release 1.2 (rev 1.0)

This instruction manual supports the usage and methodology of the DEEPSOUND M3 device.

The user of this document is assumed to have knowledge of ultrasound techniques, testing,

imaging and hardware components such as probe etc.

Audience

This document and the information contained in its strictly reserved for current Seongsan Lab

personnel, Seongsan Lab licensed representatives, and Seongsan Lab customers who have

purchased a valid service agreement for use by the customer’s designated in-house service

employee on equipment located at the customer’s designated site.

General Precautions

Before using and turning on the device, please carefully read the instructions in this

user manual.

Keep this manual as priority for future reference.

Please follow the installation and operation procedures carefully.

As software reference, see the user manual, DSVision Software User Manual, Rel.1.2

User manual format

This manual is in Portable Document Format (PDF), for viewing on a laptop computer or any

other electronic device compatible with Acrobat Reader® . A list of bookmarks functions as an

additional table of contents.

Conventions of the user manual

This user manual is designed to be easy to understand and user-friendly, but to take full

benefit of the information provided, conventions are useful to know easily.

Various typographical conventions explained below were defined to standardize and simplify

the look and feel of this documentation. :

Hypertext links are bold.

All procedures are numbered, and sub-procedures are

lettered.

Control names and menu items or titles are spelled as

they are in the device, and they appear in bold text.

Symbols, menu bars and images appear as they appear

in the software part

An italic typeface is used to indicate on a specific word

or phrase (for example: should always).

Small capitals are generally used when reference is

made to inscriptions found on the device (indicator

lights, buttons, etc.)

Page 3: Release 1.2 Sdf sdf sdf sdf sdf DEEPSOUND M3

Release 1.2, 2017

Copyright⒞ Seongsan Lab. All rights Reserved 2

Safety and Other Indications

The purpose of the various safety indications is to ensure operator safety and device

integrity.

This sign indicates a hazardous action. It denotes that if not correctly performed or adhered to, could result in material damage of loss of data.

Important: The Important sign denotes to a note that provides important information, or

information necessary to the completion of a task with device.

Note: The Note sign indicates an operating procedure, practice that requires special

attention. It also denotes related information, but not imperative ones.

Questions or comments about this manual

If you have questions about this manual, please contact Seongsan Lab,

[email protected]

Research and Development team, at the address

below

ATTENTION

Seongsan Lab, 2017. Rev 1.0. All rights reserved. All the information herein is subject to change without prior notification.

Contact us:

Seongsan Lab

Seung-il venture tower 802, Gasan Digital Road 2-gil 136 (Gasan-dong 371-42) Geumcheon-gu, Seoul, Korea Seoul, South Korea P:02-2039-5725 F:02-2039-5726

www.seongsanlab.com

Quality

All work is done in accordance with Seongsan Lab Quality standard, which complies with ISO 5577:2000, ISO 2400:1972 and ISO 10375:1997

Page 4: Release 1.2 Sdf sdf sdf sdf sdf DEEPSOUND M3

Release 1.2, 2017

Copyright⒞ Seongsan Lab. All rights Reserved 3

Table of Contents

About this User Manual ……………………………………………….……………………………1

Introduction of the DEEPSOUND M3..…………………………………………………………...4

1 Description of the M3 Body Panel………………………………………………………..6

2 Detailed specifactions of the M3 ................................................................................................ 7

3 Preparing M3 for in-use ............................................................................................................... 9

3.1 Proper standing spot for the M3 .............................................................................................. 9

3.2 Connecting M3 to the Host ....................................................................................................... 9

3.3 Probe Connections ................................................................................................................... 12

3.4 Encoder Connections ............................................................................................................... 13

3.5 Turning ON/OFF the M3 ......................................................................................................... 13

4 Maintanence of M3 ...................................................................................................................... 14

4.1 Cleaning the M3 ....................................................................................................................... 14

4.2 Light indicator status ............................................................................................................... 14

5 Connector Reference ................................................................................................................... 15

5.1 Pulser/Receiver Connectors ................................................................................................... 15

5.2 Gigabit Ethernet Connector .................................................................................................... 16

5.3 Encoder Connector .................................................................................................................. 16

6 Basic Troubleshooting ................................................................................................................ 18

SUPPORT … ………………………………………………………………………………………. 20

Page 5: Release 1.2 Sdf sdf sdf sdf sdf DEEPSOUND M3

Release 1.2, 2017

Copyright⒞ Seongsan Lab. All rights Reserved 4

Introduction of the DEEPSOUND M3

Seongsan Lab’s DEEPSOUND M3 device is a pioneer in its class for South Korean domestic

market that targeted for providing a high quality, portable and highly efficient, user friendly

ultrasonic inspection using PA-UT (phased array UT technology).

Phased Array Ultrasonic Testing (PA-UT) is used for a flaw detection, sizing, and imaging.

PA-UT technology has the ability to modify electronically the acoustic probe characteristics.

The device offers following features & dimension of spacing with high-performance and

efficiency with portability.

DEEPSOUND M3 with PC – The device is developed to enable control and data collection

and analysis by using built-in PC, unlike regular PAUT devices does not include. Since it can

store and manage equipment. It has various advantages such as reducing the risk of losing

data and uniformity of management.

32 channel /256 element system PR – Innovative phased array system, up to 32 simultaneous

active channels and 256 in total. 32 indicates the maximum number of simultaneously active

beam-forming channels, whereas 256 indicates the total number of available probe elements.

With the same processing power, the system can adapt focal laws to conduct inspection on

complex surfaces. In addition, you can connect two Phased array probe at the same time

through two Mini DLP connector without extension connector.

Bipolar Pulse wave type – While other PAUT device uses Unipolar Pulse wave method, but

DEPPSOUND M3 adopts Bipolar Pulse wave method which is recognized in medical

ultrasonic technology, therefore device will generate clear signal with less noise.

Dynamic Sample Focusing – In the receiving phase of ultrasonic signal, the signal

components collected from several oscillators are separated into the noise component and

signal component by the calculated Focus value at the specific position. Conventional PAUT

device uses Dynamic Depth Focusing method to acquire signals with different focus values

with every 10mm, but DEEPSOUND M3 uses Dynamic Sample Focusing to calculate focus

every 0.06mm by varying the value to obtain a better signals.

DEEPSOUND M3 with software - The M3 device properly works with remote PC running

Windows Xp, Windows 7 or Windows 10 and Seongsan Lab’s DSVision software. However

M3 includes computer inside, PC is remotely used for send/receive data, as PC will be

referred as “Host” in this document. Since DEEPSOUND M3 is driven by Seongsan Lab’s

powerful DSVision software platform, M3 offers advanced data acquisition and analysis.

Additionally, it can dynamically adapt focal laws to conduct inspections on complex surfaces.

In terms of performance point, M3 device is capable of handling data files in high speed, which

gives extra feature to users to increase scanning speed without losing its quality

For more information about DSVision software, please see the software user manual,

Release1.2.

Page 6: Release 1.2 Sdf sdf sdf sdf sdf DEEPSOUND M3

Release 1.2, 2017

Copyright⒞ Seongsan Lab. All rights Reserved 5

1. Description of DEEPSOUND M3 body panels

Following image shows entire view of M3 surface with 3-Dimension.

Figure 1-1 M3 device – 3 Dimensional view

1. Top Panel

2. Side Panel (including power , Air-intake, encoder connector, fan, switch)

3. Front Panel (including power button, probe connector, handler)

1. Following image describes a quick overview of M3 front panel:

Figure 1-2 M3 Front panel

Important: The front panel of the M3 is not splash-proof.

1. Pulser/ Receiver connectors (Fig 1-2)

2. On/Off button (Fig 1-2)

2. The top panel of M3:

1 Sdf sdf sdf sdf sdf

2 Sdf sdf sdf sdf sdf

1

Sdf sdf sdf sdf sdf

1

2

Sdf sdf sdf sdf sdf

1

3

Sdf sdf sdf sdf sdf

1

Page 7: Release 1.2 Sdf sdf sdf sdf sdf DEEPSOUND M3

Release 1.2, 2017

Copyright⒞ Seongsan Lab. All rights Reserved 6

The top panel has four side-bumper for avoid damage.

3. The following image demonstrates side body panel of M3 device.

Figure 1-3 M3 Side panel

Important: The side panel of the M3 is not splash-proof.

1. Cooling fan

2. ENCODER connector

3. Ethernet connector

4. USB connector x 4 (USB version 2.0)

5. HDMI connector

1 Sdf sdf sdf sdf sdf

2 Sdf sdf sdf sdf sdf

3 Sdf sdf sdf sdf sdf

4 Sdf sdf sdf sdf sdf

5 Sdf sdf sdf sdf sdf

6 Sdf sdf sdf sdf sdf

7 Sdf sdf sdf sdf sdf

8 Sdf sdf sdf sdf sdf

9 Sdf sdf sdf sdf sdf

10 Sdf sdf sdf sdf sdf

Page 8: Release 1.2 Sdf sdf sdf sdf sdf DEEPSOUND M3

Release 1.2, 2017

Copyright⒞ Seongsan Lab. All rights Reserved 7

6. VGA connector

7. SPK connector

8. Power connector

9. ON/OFF switch

10. Air intake in/out

2. Detailed Specifications of the DEEPSOUND M3

This chapter will provide more information about specifications of the M3.

Main body of the DEEPSOUND M3

Size (H x W x D): 37cm x 23cm x 25cm (14.56in. x 9.05in. x 9.84 in.) Net weight: Approximately 12.5kg (27.55 lb.)

Environment

Operating temperature: 0C to 45C (32F to 113F) Operating humidity: up to 80% at 31C

General specifications:

Air intake: Yes

Phased Array Connector: 2 Mini DLP

Pulse wave type: Bipolar

Phased Array Channel: up to 32/256 PR

Digitizing frequency: up to 50MHZ

Amplitude Resolution Phased Array Signal: 16 bit

Network interface: High speed (1Gb/s)

Maximum PRF: 33 KHz

Maximum pulser voltage: 220V

Bandwidth: from 0.25 to 20 MHz

Real-time data compression: Yes

Rectification: Digital

Filtering: Analog/Digital

Smoothing (Video filter): Yes

Dynamic Depth Focusing(DDF) : Yes

Encoder Interface: 8 quadrature type/ 4 differential type

PC Software control: Yes

Page 9: Release 1.2 Sdf sdf sdf sdf sdf DEEPSOUND M3

Release 1.2, 2017

Copyright⒞ Seongsan Lab. All rights Reserved 8

External trigger: Yes

Note: M3 has specially designed to operate in extreme conditions, those

environmental operating temperature may fixed when M3 using Air intake

feature.

Interface

Ethernet Interface: Gigabit Ethernet (High-speed (1 Gb/s)) Ethernet cable length: up to 100m, max. Global data throughput: up to 20 MB/s USB: High speed 2.0 Visual connection interface: HDMI/VGA

Power supply

Voltage: 0C to 45C (32F to 113F) Frequency: up to 80% at 31C

System/Pc software control

Data acquisition Data analysis Software: DSVision software

Features

Bipolar pulse wave type Dynamic sample focusing

Page 10: Release 1.2 Sdf sdf sdf sdf sdf DEEPSOUND M3

Release 1.2, 2017

Copyright⒞ Seongsan Lab. All rights Reserved 9

3. Preparing the DEEPSOUND M3 for in-use

This section will explain following steps required to setting up before in-use:

Proper standing spot for the M3

Connecting M3 to the host computer

Probe connections

Encoder connections

Turning ON/OFF the M3

Furthermore, appendix section will provide more information about probes, which will lead

you to obtaining even better acquisition scanning result.

We strongly recommend user to read all this manual thoroughly before real time

application or unintended hazardous action, since this document provide all

information you may need.

3.1 Proper standing spot for DEEPSOUND M3

M3 device is relatively lighter (appx. 12.5 kg) than other market products, therefore device

can be used on the desk, floor or other flat spaces.

Since M3 met the safe and light performance in hardware matter, but safety is still essential.

Therefore we recommend users to use M3 device always makes sure positioned away from

heat sources, and leave a distance of at least 5-10 cm in front of the fan outlets, unless Air

Intake feature not in access.

Important: It is recommended to that device should positioned close to a power outlet that

could avoid disconnect unexpectedly in some manner.

Also, we encourage you to not block cooling fan for while intentionally or not.

3.2 Connecting DEEPSOUND M3 to the Host

Inspection with the M3 are remotely controlled via Ethernet connection to Host PC. Following

steps will guide you to connect M3 to the Host PC and configure setting with Seongsan Lab’s

DSVision software.

Important: It is recommended to plug off device from wall before accessing side panel

connectors.

Page 11: Release 1.2 Sdf sdf sdf sdf sdf DEEPSOUND M3

Release 1.2, 2017

Copyright⒞ Seongsan Lab. All rights Reserved 10

The M3 is equipped with one Gigabit ethernet adapters and network card and network

dedicated to the M3. Once the device and Host PC connected via LAN (should not exceed 100

meters, you need longer, you may need repeaters or boosters), further steps below should

followed.

Seongsan Lab’s DSVision software has network module with make connection between

hardware device (DEEPSOUND M3) and host computer by 1:1 bonding via network.

■ Settings

Desktop computer network information

Network equipment information on DSVision

To enter Network Mode:

DSVision – Tool – Setting – Network

Figure 3-1 Network Setting mode

Local Network setting can be handled by Desktop computer – Network settings, and

insert your network credentials into Network window, click Set-OK to save setting.

For example as:

Receive Port : 9900

IP Address : 192.168.0.99

Subnet Mask : 255.255.255.0

Gateway : 192.168.0.1

Device Network in Fig.3.2 stands for the DSVision network information of

hardware equipment. Click ( ) icon to edit DSVision device IP Address and

click Check – Network check.

Page 12: Release 1.2 Sdf sdf sdf sdf sdf DEEPSOUND M3

Release 1.2, 2017

Copyright⒞ Seongsan Lab. All rights Reserved 11

Figure 3-2 Network Setting – Network check

After checking network credentials, Click OK button on main setting window

.

Figure 3-3 Full Network Settings

Page 13: Release 1.2 Sdf sdf sdf sdf sdf DEEPSOUND M3

Release 1.2, 2017

Copyright⒞ Seongsan Lab. All rights Reserved 12

3.3 Probe connections

This section will demonstrate how to connect probes to the M3 device. Since M3 has 2 phased

array connections slots for probe connectivity, next few pages will demonstrate the main

configurations for DeepSound M3 PAUT device.

DEEPSOUND M3 32/256PR – Innovative phased array system, up to 32 simultaneous active

channels and 256 in total.

* M3 device has up to the 32/256 PR model phased array channel.

Figure 3-4 32/256 phased array channel configurations

Installing the connector:

To install connector:

1. Remove the adapter from the storage.

2. Align the adapter on Pulser/Receiver, and gently push it in. The adapter should slide

easily.

On the 32/256 PR configurations, connector connects Pulser/Receiver.

3. Turn Open/Close with tightening screws clockwise, and adapter is ready to receive

conventional probes.

Removing the connector

To remove connector:

1. Turn tightening screws counterclockwise until they are completely out of the

sockets.

2. Gently pull out to remove connector from the socket.

We strongly recommend users to pull out the adapter in straight direction.

Otherwise connector pins could bend and render the adapter unusable.

Up to 256 element, each for 128 elements Sdf sdf sdf sdf sdf

Page 14: Release 1.2 Sdf sdf sdf sdf sdf DEEPSOUND M3

Release 1.2, 2017

Copyright⒞ Seongsan Lab. All rights Reserved 13

3.4 Encoder connections

The Encoder connector is fully compatible with the scanning mechanism. Following image

shows that encoder connection port of the M3, which compatible with manually driven

scanning mechanisms. M3 encoder interface has 8 quadrature type and 4 differential type.

Figure 3-5 128 Encoder connection in side panel

Important: Encoder adapter cable is necessary for when connecting basic manual scanner

to the M3 via Encoder connector, depending on the pin-out of manual scanner

type.

3.5 Turning ON/OFF the M3

We strongly recommend users to do not turn on the M3 device until it has reach

room temperature, if M3 has carried under extreme low temperature for long

time, immediate turning on may lead M3 to the damage.

To start up the M3:

1. Plug the power cord into power outlet or wall.

2. Press ON/OFF switch, where located on the side panel.

3. On your HOST PC, start the Seongsan Lab’s DSVision software.

4. Press Tool-Settings – Network to connect device.

5. Click refresh if the device address is not properly recognized. This will ensure that the

M3 is ready for connection.

Manual encoder Sdf sdf sdf sdf sdf

Air intake Sdf sdf sdf sdf sdf

Page 15: Release 1.2 Sdf sdf sdf sdf sdf DEEPSOUND M3

Release 1.2, 2017

Copyright⒞ Seongsan Lab. All rights Reserved 14

4. Maintenance of DEEPSOUND M3

Regarding good working conditions of M3 device, following section will provide basic

maintenance instructions. M3 has designed to require minimal maintenance.

4.1 Cleaning the DEEPSOUND M3

The M3 external cases (front and side panel) can be cleaned under some circumstances.

To clean the device:

Always make sure the device power cord is disconnected and turned off completely.

Keep in mind that device body surface is better to cleaned with a soft clothing

material.

We strongly recommend users to not clean the device with liquid, especially when

device is turned on. Liquid could make the device partially or wholly inoperable,

or when connectors stay wet without immediate drying process, it may produce a

short circuit when plugging cables.

Wait until the device dries completely before plugging into the power cord or any

other cables.

4.2 Light indicator status

The light indicator let you know the information on current status of M3 device.

Figure 4-1 Status light indicator

Power status light indicator

The Power status light (Fig.4.1) indicator appears white blue when device is properly turned

On and operating smoothly. When turned Off, the light indicator will appear white.

Page 16: Release 1.2 Sdf sdf sdf sdf sdf DEEPSOUND M3

Release 1.2, 2017

Copyright⒞ Seongsan Lab. All rights Reserved 15

5 Connector references

Following section will demonstrate all the relevant connector information of the M3 device.

5.1 Pulser/Receiver Connectors

Specification: Pitch 0.8mm, 4 Row, 260P, Receptacle, DIP

Manufacturer: TE Connectivity, model number: TC-ZIF REC ASSY 260P

*Note that M3 has 2 connector on device body, those are identical to each other.

Figure 5-1 a) Connector model b) Pin arrangement

Important: Number of positions and number of signals are equal, which 260.

Installing the pulser/receiver connector

To install connector:

4. Remove the adapter from the storage.

5. Align the adapter on Pulser/Receiver, and gently push it in. The adapter should slide

easily.

On the 32/256 PR configurations, connector connects Pulser/Receiver.

6. Turn Open/Close with tightening screws clockwise, and adapter is ready to receive

conventional probes.

Removing the connector

To remove connector:

3. Turn tightening screws counterclockwise until they are completely out of the

sockets.

4. Gently pull out to remove connector from the socket.

We strongly recommend users to pull out the adapter in straight direction.

Otherwise connector pins could bend and render the adapter unusable.

Page 17: Release 1.2 Sdf sdf sdf sdf sdf DEEPSOUND M3

Release 1.2, 2017

Copyright⒞ Seongsan Lab. All rights Reserved 16

5.2 Gigabit Ethernet Connector

M3 uses auto-configurable Gigabit Ethernet port, user may use any type of Ethernet cable to

connect M3 device to the HOST-PC.

Description: RJ-45, female connector

Figure 5-2 Gigabit Ethernet connector

Parameters Specifications Description RJ-45 , four pair

Maximum length 100

5.3 Encoder Connector (Receptacle)

Description: Amphenol 58-570124-18S, Circular MIL Spec connector

Suggested cable connector: ITT Cannon KPT 061319-18P

Figure 5-3 Encoder connector

Table 5-1: Pinout for Encoder connector (M3 device)

Pin I/O Signal Description Type

F Output +5V +5Vb -

G Input PHA3 Encoder1, phase A TTL

H Input PHB3 Encoder1, phase B TTL

J - GND Ground -

K Input PHA4 Encoder 2 TTL

L Input PHB4 Encoder 2 TTL

M - GND Ground -

Page 18: Release 1.2 Sdf sdf sdf sdf sdf DEEPSOUND M3

Release 1.2, 2017

Copyright⒞ Seongsan Lab. All rights Reserved 17

The M3 is designed for phased array inspections, and Seongsan Lab provides connector that

allows you work with conventional UT probes or with a combination of conventional and

phased array UT probes.

Important thing that working with probe is make sure that the connector pins are

properly aligned and that the adapter is not positioned in wrong side. Also when

manipulating the adapter, never use excessive force, that may bend the pins and

render the adapter unusable.

Installing the encoder connector

To install the encoder connector:

1. Remove the adapter from the storage.

2. Align the adapter on encoder, and gently push it in. The adapter should slide easily.

Removing the connector

To remove connector:

5. Turn tightening screws counterclockwise until they are completely out of the

sockets.

6. Gently pull out to remove it from the connector.

We strongly recommend users to pull out the adapter in straight direction.

Otherwise connector pins could bend and render the adapter unusable.

Page 19: Release 1.2 Sdf sdf sdf sdf sdf DEEPSOUND M3

Release 1.2, 2017

Copyright⒞ Seongsan Lab. All rights Reserved 18

6 Basic Troubleshooting

This section will provide basic troubleshooting if device malfunctioned or problem arises.

Following solutions has suggested in Table 6-1.

Table 6-1: Basic troubleshooting

Problem Cause Recommended solution

M3 does not turn ON

Power cord disconnection

Check the power cord and replace

it with proper size.

Communication problems occur

between the HOST-PC and the M3

device

HOST and M3 are not connected

with the proper type of Ethernet

cable

Check the Ethernet cable that meet

the proper length and quality.

DSVision cannot accessed

The license key is not valid on your

HOST PC.

Insert valid license key provided

from manufacturer.

The M3 shuts down unexpectedly

The device has been used for

extreme temperatures for an

extended time

Wait few minutes before turning

on device.

Change the working location or

reduce the ambient temperature.

Start using Air intake feature

Encoder signal are not detected

Wrong setup configured on

DSVision software

Verify pin-out and refer to

DSVision software manual.

Page 20: Release 1.2 Sdf sdf sdf sdf sdf DEEPSOUND M3

Release 1.2, 2017

Copyright⒞ Seongsan Lab. All rights Reserved 19

DEEPSOUND M3 PAUT Device User Manual Release 1.2

Important NOTICE:

Seongsan Lab reserve the right to make corrections, modifications, enhancements,

improvements, and other changes to its products and services at any time and to discontinue

any product or service without notice. Customers should obtain the latest relevant

information before placing orders and should verify that such information is current and

complete. All products (hardware and software) are sold subject to Seongsan Lab’s terms and

conditions of sale supplied at the time of order acknowledgment.

Problem with DEEPSOUND M3 device

Here at Seongsan Lab, we understand that sometime there may some problem may arise. In

that case, we strongly recommend that:

1. Read this document deeply, since this document provide all the information you

will need.

2. If some solution could not found, please refer “basic troubleshooting” on

Chapter 6.

3. If you do not find any proper solution from this document, feel free to send us

e-mail or mail to addressing below, since testing and other quality control

techniques are used to the extent Seongsan Lab deems necessary to support this

user manual.

Mailing Address:

Seung-il venture tower 802, Gasan Digital Road 2-gil 136, Geumcheon-gu, seoul, Korea

E-mail: [email protected]

Copyright © 2017, Seongsan Lab

Page 21: Release 1.2 Sdf sdf sdf sdf sdf DEEPSOUND M3

Release 1.2, 2017

Copyright⒞ Seongsan Lab. All rights Reserved 20

SUPPORT

Product type (check the box)

DeepSound M3 device DSVision software

Problem Report Sheet

Company

Reported by

Telephone/Fax

Report date

Product version

Problem identification (draw underline)

Electric part/Probe & cable connections

Power interruption/Estimation/Inspection times/Other issues

Problem description (optional)

……………………………………………………………………………………………………………

……………………………………………………………………………………………………………

……………………………………………………………………………………………………………

……………………………………………………………………………………………………………

……………………………………………………………………………………………………………

……………………………………………………………………………………………………………

……………………………………………………………………………………………………………

……………………………………………………………………………………………………………

……………………………………………………………………………………………………………

……………………………………………………………………………………………………………

……………………………………………………………………………………………………………

Page 22: Release 1.2 Sdf sdf sdf sdf sdf DEEPSOUND M3

Release 1.2, 2017

Copyright⒞ Seongsan Lab. All rights Reserved 21

NOTE: Mailing or e-mail address has written in Page 1.

User Manual

Release 1.2

Jan, 2017

Copyrightⓒ 2017 Seongsan Lab, Seoul, South Korea This manual is protected under international copyright laws. No part of this manual may be reproduced, distributed, translated, or transmitted in any form or by any means, electronic, magnetic, optical, chemical, manual or otherwise, without the prior written permission of Seongsan Lab.