User Manual
Release 1.2
Copyrightⓒ 2017 Seongsan Lab, Seoul, South Korea This manual is protected under international copyright laws. No part of this manual may be reproduced, distributed, translated, or transmitted in any form or by any means, electronic, magnetic, optical, chemical, manual or otherwise, without the prior written permission of Seongsan Lab.
All rights reserved
DEEPSOUND M3
www.seongsanlab.com
Sdf sdf sdf sdf sdf
DEEPSOUND M3 - Easy, Precise, Fast Phased Array UT device
®
Release 1.2, 2017
Copyright⒞ Seongsan Lab. All rights Reserved 1
About this user manual Manual release 1.2 (rev 1.0)
This instruction manual supports the usage and methodology of the DEEPSOUND M3 device.
The user of this document is assumed to have knowledge of ultrasound techniques, testing,
imaging and hardware components such as probe etc.
Audience
This document and the information contained in its strictly reserved for current Seongsan Lab
personnel, Seongsan Lab licensed representatives, and Seongsan Lab customers who have
purchased a valid service agreement for use by the customer’s designated in-house service
employee on equipment located at the customer’s designated site.
General Precautions
Before using and turning on the device, please carefully read the instructions in this
user manual.
Keep this manual as priority for future reference.
Please follow the installation and operation procedures carefully.
As software reference, see the user manual, DSVision Software User Manual, Rel.1.2
User manual format
This manual is in Portable Document Format (PDF), for viewing on a laptop computer or any
other electronic device compatible with Acrobat Reader® . A list of bookmarks functions as an
additional table of contents.
Conventions of the user manual
This user manual is designed to be easy to understand and user-friendly, but to take full
benefit of the information provided, conventions are useful to know easily.
Various typographical conventions explained below were defined to standardize and simplify
the look and feel of this documentation. :
Hypertext links are bold.
All procedures are numbered, and sub-procedures are
lettered.
Control names and menu items or titles are spelled as
they are in the device, and they appear in bold text.
Symbols, menu bars and images appear as they appear
in the software part
An italic typeface is used to indicate on a specific word
or phrase (for example: should always).
Small capitals are generally used when reference is
made to inscriptions found on the device (indicator
lights, buttons, etc.)
Release 1.2, 2017
Copyright⒞ Seongsan Lab. All rights Reserved 2
Safety and Other Indications
The purpose of the various safety indications is to ensure operator safety and device
integrity.
This sign indicates a hazardous action. It denotes that if not correctly performed or adhered to, could result in material damage of loss of data.
Important: The Important sign denotes to a note that provides important information, or
information necessary to the completion of a task with device.
Note: The Note sign indicates an operating procedure, practice that requires special
attention. It also denotes related information, but not imperative ones.
Questions or comments about this manual
If you have questions about this manual, please contact Seongsan Lab,
Research and Development team, at the address
below
ATTENTION
Seongsan Lab, 2017. Rev 1.0. All rights reserved. All the information herein is subject to change without prior notification.
Contact us:
Seongsan Lab
Seung-il venture tower 802, Gasan Digital Road 2-gil 136 (Gasan-dong 371-42) Geumcheon-gu, Seoul, Korea Seoul, South Korea P:02-2039-5725 F:02-2039-5726
www.seongsanlab.com
Quality
All work is done in accordance with Seongsan Lab Quality standard, which complies with ISO 5577:2000, ISO 2400:1972 and ISO 10375:1997
Release 1.2, 2017
Copyright⒞ Seongsan Lab. All rights Reserved 3
Table of Contents
About this User Manual ……………………………………………….……………………………1
Introduction of the DEEPSOUND M3..…………………………………………………………...4
1 Description of the M3 Body Panel………………………………………………………..6
2 Detailed specifactions of the M3 ................................................................................................ 7
3 Preparing M3 for in-use ............................................................................................................... 9
3.1 Proper standing spot for the M3 .............................................................................................. 9
3.2 Connecting M3 to the Host ....................................................................................................... 9
3.3 Probe Connections ................................................................................................................... 12
3.4 Encoder Connections ............................................................................................................... 13
3.5 Turning ON/OFF the M3 ......................................................................................................... 13
4 Maintanence of M3 ...................................................................................................................... 14
4.1 Cleaning the M3 ....................................................................................................................... 14
4.2 Light indicator status ............................................................................................................... 14
5 Connector Reference ................................................................................................................... 15
5.1 Pulser/Receiver Connectors ................................................................................................... 15
5.2 Gigabit Ethernet Connector .................................................................................................... 16
5.3 Encoder Connector .................................................................................................................. 16
6 Basic Troubleshooting ................................................................................................................ 18
SUPPORT … ………………………………………………………………………………………. 20
Release 1.2, 2017
Copyright⒞ Seongsan Lab. All rights Reserved 4
Introduction of the DEEPSOUND M3
Seongsan Lab’s DEEPSOUND M3 device is a pioneer in its class for South Korean domestic
market that targeted for providing a high quality, portable and highly efficient, user friendly
ultrasonic inspection using PA-UT (phased array UT technology).
Phased Array Ultrasonic Testing (PA-UT) is used for a flaw detection, sizing, and imaging.
PA-UT technology has the ability to modify electronically the acoustic probe characteristics.
The device offers following features & dimension of spacing with high-performance and
efficiency with portability.
DEEPSOUND M3 with PC – The device is developed to enable control and data collection
and analysis by using built-in PC, unlike regular PAUT devices does not include. Since it can
store and manage equipment. It has various advantages such as reducing the risk of losing
data and uniformity of management.
32 channel /256 element system PR – Innovative phased array system, up to 32 simultaneous
active channels and 256 in total. 32 indicates the maximum number of simultaneously active
beam-forming channels, whereas 256 indicates the total number of available probe elements.
With the same processing power, the system can adapt focal laws to conduct inspection on
complex surfaces. In addition, you can connect two Phased array probe at the same time
through two Mini DLP connector without extension connector.
Bipolar Pulse wave type – While other PAUT device uses Unipolar Pulse wave method, but
DEPPSOUND M3 adopts Bipolar Pulse wave method which is recognized in medical
ultrasonic technology, therefore device will generate clear signal with less noise.
Dynamic Sample Focusing – In the receiving phase of ultrasonic signal, the signal
components collected from several oscillators are separated into the noise component and
signal component by the calculated Focus value at the specific position. Conventional PAUT
device uses Dynamic Depth Focusing method to acquire signals with different focus values
with every 10mm, but DEEPSOUND M3 uses Dynamic Sample Focusing to calculate focus
every 0.06mm by varying the value to obtain a better signals.
DEEPSOUND M3 with software - The M3 device properly works with remote PC running
Windows Xp, Windows 7 or Windows 10 and Seongsan Lab’s DSVision software. However
M3 includes computer inside, PC is remotely used for send/receive data, as PC will be
referred as “Host” in this document. Since DEEPSOUND M3 is driven by Seongsan Lab’s
powerful DSVision software platform, M3 offers advanced data acquisition and analysis.
Additionally, it can dynamically adapt focal laws to conduct inspections on complex surfaces.
In terms of performance point, M3 device is capable of handling data files in high speed, which
gives extra feature to users to increase scanning speed without losing its quality
For more information about DSVision software, please see the software user manual,
Release1.2.
Release 1.2, 2017
Copyright⒞ Seongsan Lab. All rights Reserved 5
1. Description of DEEPSOUND M3 body panels
Following image shows entire view of M3 surface with 3-Dimension.
Figure 1-1 M3 device – 3 Dimensional view
1. Top Panel
2. Side Panel (including power , Air-intake, encoder connector, fan, switch)
3. Front Panel (including power button, probe connector, handler)
1. Following image describes a quick overview of M3 front panel:
Figure 1-2 M3 Front panel
Important: The front panel of the M3 is not splash-proof.
1. Pulser/ Receiver connectors (Fig 1-2)
2. On/Off button (Fig 1-2)
2. The top panel of M3:
1 Sdf sdf sdf sdf sdf
2 Sdf sdf sdf sdf sdf
1
Sdf sdf sdf sdf sdf
1
2
Sdf sdf sdf sdf sdf
1
3
Sdf sdf sdf sdf sdf
1
Release 1.2, 2017
Copyright⒞ Seongsan Lab. All rights Reserved 6
The top panel has four side-bumper for avoid damage.
3. The following image demonstrates side body panel of M3 device.
Figure 1-3 M3 Side panel
Important: The side panel of the M3 is not splash-proof.
1. Cooling fan
2. ENCODER connector
3. Ethernet connector
4. USB connector x 4 (USB version 2.0)
5. HDMI connector
1 Sdf sdf sdf sdf sdf
2 Sdf sdf sdf sdf sdf
3 Sdf sdf sdf sdf sdf
4 Sdf sdf sdf sdf sdf
5 Sdf sdf sdf sdf sdf
6 Sdf sdf sdf sdf sdf
7 Sdf sdf sdf sdf sdf
8 Sdf sdf sdf sdf sdf
9 Sdf sdf sdf sdf sdf
10 Sdf sdf sdf sdf sdf
Release 1.2, 2017
Copyright⒞ Seongsan Lab. All rights Reserved 7
6. VGA connector
7. SPK connector
8. Power connector
9. ON/OFF switch
10. Air intake in/out
2. Detailed Specifications of the DEEPSOUND M3
This chapter will provide more information about specifications of the M3.
Main body of the DEEPSOUND M3
Size (H x W x D): 37cm x 23cm x 25cm (14.56in. x 9.05in. x 9.84 in.) Net weight: Approximately 12.5kg (27.55 lb.)
Environment
Operating temperature: 0C to 45C (32F to 113F) Operating humidity: up to 80% at 31C
General specifications:
Air intake: Yes
Phased Array Connector: 2 Mini DLP
Pulse wave type: Bipolar
Phased Array Channel: up to 32/256 PR
Digitizing frequency: up to 50MHZ
Amplitude Resolution Phased Array Signal: 16 bit
Network interface: High speed (1Gb/s)
Maximum PRF: 33 KHz
Maximum pulser voltage: 220V
Bandwidth: from 0.25 to 20 MHz
Real-time data compression: Yes
Rectification: Digital
Filtering: Analog/Digital
Smoothing (Video filter): Yes
Dynamic Depth Focusing(DDF) : Yes
Encoder Interface: 8 quadrature type/ 4 differential type
PC Software control: Yes
Release 1.2, 2017
Copyright⒞ Seongsan Lab. All rights Reserved 8
External trigger: Yes
Note: M3 has specially designed to operate in extreme conditions, those
environmental operating temperature may fixed when M3 using Air intake
feature.
Interface
Ethernet Interface: Gigabit Ethernet (High-speed (1 Gb/s)) Ethernet cable length: up to 100m, max. Global data throughput: up to 20 MB/s USB: High speed 2.0 Visual connection interface: HDMI/VGA
Power supply
Voltage: 0C to 45C (32F to 113F) Frequency: up to 80% at 31C
System/Pc software control
Data acquisition Data analysis Software: DSVision software
Features
Bipolar pulse wave type Dynamic sample focusing
Release 1.2, 2017
Copyright⒞ Seongsan Lab. All rights Reserved 9
3. Preparing the DEEPSOUND M3 for in-use
This section will explain following steps required to setting up before in-use:
Proper standing spot for the M3
Connecting M3 to the host computer
Probe connections
Encoder connections
Turning ON/OFF the M3
Furthermore, appendix section will provide more information about probes, which will lead
you to obtaining even better acquisition scanning result.
We strongly recommend user to read all this manual thoroughly before real time
application or unintended hazardous action, since this document provide all
information you may need.
3.1 Proper standing spot for DEEPSOUND M3
M3 device is relatively lighter (appx. 12.5 kg) than other market products, therefore device
can be used on the desk, floor or other flat spaces.
Since M3 met the safe and light performance in hardware matter, but safety is still essential.
Therefore we recommend users to use M3 device always makes sure positioned away from
heat sources, and leave a distance of at least 5-10 cm in front of the fan outlets, unless Air
Intake feature not in access.
Important: It is recommended to that device should positioned close to a power outlet that
could avoid disconnect unexpectedly in some manner.
Also, we encourage you to not block cooling fan for while intentionally or not.
3.2 Connecting DEEPSOUND M3 to the Host
Inspection with the M3 are remotely controlled via Ethernet connection to Host PC. Following
steps will guide you to connect M3 to the Host PC and configure setting with Seongsan Lab’s
DSVision software.
Important: It is recommended to plug off device from wall before accessing side panel
connectors.
Release 1.2, 2017
Copyright⒞ Seongsan Lab. All rights Reserved 10
The M3 is equipped with one Gigabit ethernet adapters and network card and network
dedicated to the M3. Once the device and Host PC connected via LAN (should not exceed 100
meters, you need longer, you may need repeaters or boosters), further steps below should
followed.
Seongsan Lab’s DSVision software has network module with make connection between
hardware device (DEEPSOUND M3) and host computer by 1:1 bonding via network.
■ Settings
Desktop computer network information
Network equipment information on DSVision
To enter Network Mode:
DSVision – Tool – Setting – Network
Figure 3-1 Network Setting mode
Local Network setting can be handled by Desktop computer – Network settings, and
insert your network credentials into Network window, click Set-OK to save setting.
For example as:
Receive Port : 9900
IP Address : 192.168.0.99
Subnet Mask : 255.255.255.0
Gateway : 192.168.0.1
Device Network in Fig.3.2 stands for the DSVision network information of
hardware equipment. Click ( ) icon to edit DSVision device IP Address and
click Check – Network check.
Release 1.2, 2017
Copyright⒞ Seongsan Lab. All rights Reserved 11
Figure 3-2 Network Setting – Network check
After checking network credentials, Click OK button on main setting window
.
Figure 3-3 Full Network Settings
Release 1.2, 2017
Copyright⒞ Seongsan Lab. All rights Reserved 12
3.3 Probe connections
This section will demonstrate how to connect probes to the M3 device. Since M3 has 2 phased
array connections slots for probe connectivity, next few pages will demonstrate the main
configurations for DeepSound M3 PAUT device.
DEEPSOUND M3 32/256PR – Innovative phased array system, up to 32 simultaneous active
channels and 256 in total.
* M3 device has up to the 32/256 PR model phased array channel.
Figure 3-4 32/256 phased array channel configurations
Installing the connector:
To install connector:
1. Remove the adapter from the storage.
2. Align the adapter on Pulser/Receiver, and gently push it in. The adapter should slide
easily.
On the 32/256 PR configurations, connector connects Pulser/Receiver.
3. Turn Open/Close with tightening screws clockwise, and adapter is ready to receive
conventional probes.
Removing the connector
To remove connector:
1. Turn tightening screws counterclockwise until they are completely out of the
sockets.
2. Gently pull out to remove connector from the socket.
We strongly recommend users to pull out the adapter in straight direction.
Otherwise connector pins could bend and render the adapter unusable.
Up to 256 element, each for 128 elements Sdf sdf sdf sdf sdf
Release 1.2, 2017
Copyright⒞ Seongsan Lab. All rights Reserved 13
3.4 Encoder connections
The Encoder connector is fully compatible with the scanning mechanism. Following image
shows that encoder connection port of the M3, which compatible with manually driven
scanning mechanisms. M3 encoder interface has 8 quadrature type and 4 differential type.
Figure 3-5 128 Encoder connection in side panel
Important: Encoder adapter cable is necessary for when connecting basic manual scanner
to the M3 via Encoder connector, depending on the pin-out of manual scanner
type.
3.5 Turning ON/OFF the M3
We strongly recommend users to do not turn on the M3 device until it has reach
room temperature, if M3 has carried under extreme low temperature for long
time, immediate turning on may lead M3 to the damage.
To start up the M3:
1. Plug the power cord into power outlet or wall.
2. Press ON/OFF switch, where located on the side panel.
3. On your HOST PC, start the Seongsan Lab’s DSVision software.
4. Press Tool-Settings – Network to connect device.
5. Click refresh if the device address is not properly recognized. This will ensure that the
M3 is ready for connection.
Manual encoder Sdf sdf sdf sdf sdf
Air intake Sdf sdf sdf sdf sdf
Release 1.2, 2017
Copyright⒞ Seongsan Lab. All rights Reserved 14
4. Maintenance of DEEPSOUND M3
Regarding good working conditions of M3 device, following section will provide basic
maintenance instructions. M3 has designed to require minimal maintenance.
4.1 Cleaning the DEEPSOUND M3
The M3 external cases (front and side panel) can be cleaned under some circumstances.
To clean the device:
Always make sure the device power cord is disconnected and turned off completely.
Keep in mind that device body surface is better to cleaned with a soft clothing
material.
We strongly recommend users to not clean the device with liquid, especially when
device is turned on. Liquid could make the device partially or wholly inoperable,
or when connectors stay wet without immediate drying process, it may produce a
short circuit when plugging cables.
Wait until the device dries completely before plugging into the power cord or any
other cables.
4.2 Light indicator status
The light indicator let you know the information on current status of M3 device.
Figure 4-1 Status light indicator
Power status light indicator
The Power status light (Fig.4.1) indicator appears white blue when device is properly turned
On and operating smoothly. When turned Off, the light indicator will appear white.
Release 1.2, 2017
Copyright⒞ Seongsan Lab. All rights Reserved 15
5 Connector references
Following section will demonstrate all the relevant connector information of the M3 device.
5.1 Pulser/Receiver Connectors
Specification: Pitch 0.8mm, 4 Row, 260P, Receptacle, DIP
Manufacturer: TE Connectivity, model number: TC-ZIF REC ASSY 260P
*Note that M3 has 2 connector on device body, those are identical to each other.
Figure 5-1 a) Connector model b) Pin arrangement
Important: Number of positions and number of signals are equal, which 260.
Installing the pulser/receiver connector
To install connector:
4. Remove the adapter from the storage.
5. Align the adapter on Pulser/Receiver, and gently push it in. The adapter should slide
easily.
On the 32/256 PR configurations, connector connects Pulser/Receiver.
6. Turn Open/Close with tightening screws clockwise, and adapter is ready to receive
conventional probes.
Removing the connector
To remove connector:
3. Turn tightening screws counterclockwise until they are completely out of the
sockets.
4. Gently pull out to remove connector from the socket.
We strongly recommend users to pull out the adapter in straight direction.
Otherwise connector pins could bend and render the adapter unusable.
Release 1.2, 2017
Copyright⒞ Seongsan Lab. All rights Reserved 16
5.2 Gigabit Ethernet Connector
M3 uses auto-configurable Gigabit Ethernet port, user may use any type of Ethernet cable to
connect M3 device to the HOST-PC.
Description: RJ-45, female connector
Figure 5-2 Gigabit Ethernet connector
Parameters Specifications Description RJ-45 , four pair
Maximum length 100
5.3 Encoder Connector (Receptacle)
Description: Amphenol 58-570124-18S, Circular MIL Spec connector
Suggested cable connector: ITT Cannon KPT 061319-18P
Figure 5-3 Encoder connector
Table 5-1: Pinout for Encoder connector (M3 device)
Pin I/O Signal Description Type
F Output +5V +5Vb -
G Input PHA3 Encoder1, phase A TTL
H Input PHB3 Encoder1, phase B TTL
J - GND Ground -
K Input PHA4 Encoder 2 TTL
L Input PHB4 Encoder 2 TTL
M - GND Ground -
Release 1.2, 2017
Copyright⒞ Seongsan Lab. All rights Reserved 17
The M3 is designed for phased array inspections, and Seongsan Lab provides connector that
allows you work with conventional UT probes or with a combination of conventional and
phased array UT probes.
Important thing that working with probe is make sure that the connector pins are
properly aligned and that the adapter is not positioned in wrong side. Also when
manipulating the adapter, never use excessive force, that may bend the pins and
render the adapter unusable.
Installing the encoder connector
To install the encoder connector:
1. Remove the adapter from the storage.
2. Align the adapter on encoder, and gently push it in. The adapter should slide easily.
Removing the connector
To remove connector:
5. Turn tightening screws counterclockwise until they are completely out of the
sockets.
6. Gently pull out to remove it from the connector.
We strongly recommend users to pull out the adapter in straight direction.
Otherwise connector pins could bend and render the adapter unusable.
Release 1.2, 2017
Copyright⒞ Seongsan Lab. All rights Reserved 18
6 Basic Troubleshooting
This section will provide basic troubleshooting if device malfunctioned or problem arises.
Following solutions has suggested in Table 6-1.
Table 6-1: Basic troubleshooting
Problem Cause Recommended solution
M3 does not turn ON
Power cord disconnection
Check the power cord and replace
it with proper size.
Communication problems occur
between the HOST-PC and the M3
device
HOST and M3 are not connected
with the proper type of Ethernet
cable
Check the Ethernet cable that meet
the proper length and quality.
DSVision cannot accessed
The license key is not valid on your
HOST PC.
Insert valid license key provided
from manufacturer.
The M3 shuts down unexpectedly
The device has been used for
extreme temperatures for an
extended time
Wait few minutes before turning
on device.
Change the working location or
reduce the ambient temperature.
Start using Air intake feature
Encoder signal are not detected
Wrong setup configured on
DSVision software
Verify pin-out and refer to
DSVision software manual.
Release 1.2, 2017
Copyright⒞ Seongsan Lab. All rights Reserved 19
DEEPSOUND M3 PAUT Device User Manual Release 1.2
Important NOTICE:
Seongsan Lab reserve the right to make corrections, modifications, enhancements,
improvements, and other changes to its products and services at any time and to discontinue
any product or service without notice. Customers should obtain the latest relevant
information before placing orders and should verify that such information is current and
complete. All products (hardware and software) are sold subject to Seongsan Lab’s terms and
conditions of sale supplied at the time of order acknowledgment.
Problem with DEEPSOUND M3 device
Here at Seongsan Lab, we understand that sometime there may some problem may arise. In
that case, we strongly recommend that:
1. Read this document deeply, since this document provide all the information you
will need.
2. If some solution could not found, please refer “basic troubleshooting” on
Chapter 6.
3. If you do not find any proper solution from this document, feel free to send us
e-mail or mail to addressing below, since testing and other quality control
techniques are used to the extent Seongsan Lab deems necessary to support this
user manual.
Mailing Address:
Seung-il venture tower 802, Gasan Digital Road 2-gil 136, Geumcheon-gu, seoul, Korea
E-mail: [email protected]
Copyright © 2017, Seongsan Lab
Release 1.2, 2017
Copyright⒞ Seongsan Lab. All rights Reserved 20
SUPPORT
Product type (check the box)
DeepSound M3 device DSVision software
Problem Report Sheet
Company
Reported by
Telephone/Fax
Report date
Product version
Problem identification (draw underline)
Electric part/Probe & cable connections
Power interruption/Estimation/Inspection times/Other issues
Problem description (optional)
……………………………………………………………………………………………………………
……………………………………………………………………………………………………………
……………………………………………………………………………………………………………
……………………………………………………………………………………………………………
……………………………………………………………………………………………………………
……………………………………………………………………………………………………………
……………………………………………………………………………………………………………
……………………………………………………………………………………………………………
……………………………………………………………………………………………………………
……………………………………………………………………………………………………………
……………………………………………………………………………………………………………
Release 1.2, 2017
Copyright⒞ Seongsan Lab. All rights Reserved 21
NOTE: Mailing or e-mail address has written in Page 1.
User Manual
Release 1.2
Jan, 2017
Copyrightⓒ 2017 Seongsan Lab, Seoul, South Korea This manual is protected under international copyright laws. No part of this manual may be reproduced, distributed, translated, or transmitted in any form or by any means, electronic, magnetic, optical, chemical, manual or otherwise, without the prior written permission of Seongsan Lab.
www.seongsanlab.com Sdf sdf sdf sdf sdf
DEEPSOUND M3
®
Release 1.2, 2017
Copyright⒞ Seongsan Lab. All rights Reserved 22
All rights reserved