10
Performance Analysis of Pulse Tube/ 3 He Joule– Thomson Cryocooler for Thermometer Calibration T. Shimazaki National Metrology Institute Japan, National Institute of Advanced Industrial Science and Technology Tsukuba, Ibaraki 305-8563, Japan ABSTRACT Joule–Thomson (JT) circuits enhance the cooling capability of mechanical refrigerators. A pulse tube/ 3 He JT cryocooler for precise thermometer calibration between 0.65 K and 25 K is described in this paper. It comprises a commercial 4 K two–stage pulse tube refrigerator and a JT circuit, which is designed and built at the National Metrology Institute of Japan, National Institute of Advanced Industrial Science and Technology. Owing to the nature of thermometer calibration, the cryocooler requires a wide operational temperature range with a high temperature control stability rather than high cooling power. The cryocooler for the thermometer calibration uses 3 He as the working fluid for the JT circuit to reduce the minimum temperature to 0.5 K. This cryocooler utilizes a needle valve that can vary the flow impedance for JT expansion. An oil rotary pump, a mechanical booster pump, and a compressor were used for 3 He circulation. Using this, the temperature control stability whose standard deviation is less than 0.07 mK can be obtained. The temperature instability observed near 2 K due to the two-phase flow in the JT circuit is also discussed. INTRODUCTION Accurate thermometry is indispensable in many experiments. In principle, the temperature can be measured either using a primary thermometer, which can measure the thermodynamic temperature without calibration or a secondary thermometer which approximates the thermodynamic temperature but requires calibration for the temperature measurement. The primary thermometer performs thermometry based on a well-understood physical system such as a constant volume gas thermometer and a Johnson noise thermometer. 1 The constant volume gas thermometer relies on the relationship between the properties of an ideal gas and thermodynamic temperature. The Johnson noise thermometer relies on the relationship between the thermally induced voltage fluctuations which appear across an electrical resistor and thermodynamic temperature. However, precision primary thermometry usually requires a large and complex measurement system and long measurement time. Applicable temperature ranges for the said methods are limited. The range depends on their physical principles and apparatus settings. These drawbacks make primary thermometry impractical for most experiments except for some experiments relate to temperature standard work. Although secondary thermometers such as Cryocoolers 21, edited by R.G. Ross, Jr., J.R. Raab and S.D. Miller © International Cryocooler Conference, Inc., Boulder, CO, 2021 67

Performance Analysis of Pulse Tube/3He Joule– Thomson

  • Upload
    others

  • View
    8

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Performance Analysis of Pulse Tube/3He Joule– Thomson

P# 10

were taken at the inlet, outlet, and on the heat exchanger using three Lakeshore DT670-CU-1.4L calibratedsilicon diode temperature sensors. Due to the temperature requirement, the operating pressure was chosento be very close to the critical point of helium. The specific heat of helium increases significantly as thecritical point is approached, thus reducing the temperature rise across the magnet heat exchanger.

Figure 14a shows the measured cooling capacity with respect to operating pressure at 5.5K. As thepressure decreases from 30 PSIG to 25 PSIG, the cooling capacity increases from 3.03W to 4.03W.Below 25 PSIG, the temperature became unstable due to cyclic phase changes within the system. Thecooling capacity was then measured at different flow rates while maintaining a constant pressure of 25PSIG, as shown in Figure 14b. The cooling capacity increased from 4.03W to 4.58W at 5.5K as the flowrate decreased from 0.68 g/s to 0.63 g/s. The cooling capacity begins to decrease below 0.63 g/s. It iscritical that the cooling system remain stable in the case of a short period of increased heat load.

A heat impulse test was also run in which a 10W heat load was applied for 60 seconds and thenremoved. The inlet and outlet temperatures increased and returned to the expected operating temperaturein ~45 minutes.

ACKNOWLEDGMENTS

We’d like to thank Lionel Duband (Université Grenoble– Grenoble, France), Gerd Jakob(European Southern Observatory – Garching, Germany), Maria Salatino (Stanford University –California, USA), and Josh West (HPD – a FormFactor company – Colorado, USA), for theirassistance and collaboration on the ground-based telescope projects.

We’d also like to thank Prof. Yoshio Okada (Boston Children’s Hospital and Harvard Medical School)and Dr. Yong-Ho Lee (Korea Research Institute of Standards and Science), for their great support andcontributions on the MEG cryostat projects.

REFERENCES

1. G. Jakob, European Southern Observatory (Private Communication).

2. Wang, C., Sun, L., Lichtenwalter, B., et al., “Compact, ultra-low vibration, closed-cycle helium recyclerfor uninterrupted operation of MEG with SQUID magnetometers,” Cryogenics, vol. 76 (2016), pp. 16-22.

3. Wang, C., Olesh, A., Lee, Y.H., “An Ultra-Low Vibration Helium Reliquefier for Dual-helmetMagnetoencephalography (MEG),” Materials Science and Engineering, vol. 502 (2019).

4. Lee, Y.H., Yu, K.K., Kown, H., et al., “Reduction of Magnetic Noise in Cryocooler Operated Multi-channel,” presented at ASC 2020.

5. https://hyperphysics.phys-astr.gsu.edu/hbase/Solids/bcs.html

6. https://nedm.ornl.gov/

7. W. Wei, The California Institute of Technology (Private Communication).

Figure 14. (a) Cooling capacity vs helium pressure at a constant flow rate of 0.68 g/s and (b) coolingcapacity vs helium flow rate at a constant pressure of 25 PSIG.

(a) (b)

1

Performance Analysis of Pulse Tube/3

He Joule–

Thomson Cryocooler for Thermometer Calibration

T. Shimazaki

National Metrology Institute Japan, National Institute of Advanced Industrial Science and Technology Tsukuba, Ibaraki 305-8563, Japan

ABSTRACT

Joule–Thomson (JT) circuits enhance the cooling capability of mechanical refrigerators. A pulse tube/3He JT cryocooler for precise thermometer calibration between 0.65 K and 25 K is described in this paper. It comprises a commercial 4 K two–stage pulse tube refrigerator and a JT circuit, which is designed and built at the National Metrology Institute of Japan, National Institute of Advanced Industrial Science and Technology.

Owing to the nature of thermometer calibration, the cryocooler requires a wide operational temperature range with a high temperature control stability rather than high cooling power. The cryocooler for the thermometer calibration uses 3He as the working fluid for the JT circuit to reduce the minimum temperature to 0.5 K. This cryocooler utilizes a needle valve that can vary the flow impedance for JT expansion. An oil rotary pump, a mechanical booster pump, and a compressor were used for 3He circulation. Using this, the temperature control stability whose standard deviation is less than 0.07 mK can be obtained. The temperature instability observed near 2 K due to the two-phase flow in the JT circuit is also discussed.

INTRODUCTION Accurate thermometry is indispensable in many experiments. In principle, the temperature

can be measured either using a primary thermometer, which can measure the thermodynamic temperature without calibration or a secondary thermometer which approximates the thermodynamic temperature but requires calibration for the temperature measurement. The primary thermometer performs thermometry based on a well-understood physical system such as a constant volume gas thermometer and a Johnson noise thermometer.1 The constant volume gas thermometer relies on the relationship between the properties of an ideal gas and thermodynamic temperature. The Johnson noise thermometer relies on the relationship between the thermally induced voltage fluctuations which appear across an electrical resistor and thermodynamic temperature. However, precision primary thermometry usually requires a large and complex measurement system and long measurement time. Applicable temperature ranges for the said methods are limited. The range depends on their physical principles and apparatus settings. These drawbacks make primary thermometry impractical for most experiments except for some experiments relate to temperature standard work. Although secondary thermometers such as

���

Cryocoolers 21, edited by R.G. Ross, Jr., J.R. Raab and S.D. Miller© International Cryocooler Conference, Inc., Boulder, CO, 2021 67

Page 2: Performance Analysis of Pulse Tube/3He Joule– Thomson

2

resistance thermometers require calibration, they are compact, stable, easy-to-handle, and widely used as practical thermometry devices for many experiments.1

The International Temperature Scale of 1990 (ITS-90) consists of defining temperature fixed points and interpolation instruments that provide approximated thermodynamic temperature between fixed points.2 ITS-90 defines the temperature scale from 0.65 K to the highest temperature practically measurable in terms of the Plank radiation law using monochromatic radiation. The ITS-90 is designed to be the reference temperature scale for the calibration of secondary thermometers with highest accuracy.

Our institute realized, maintained, and disseminated the abovementioned temperature scale.3 For the cryogenic temperature standard work, a cryocooler that can control the temperature of a thermometer comparison block between 0.65 K and 25 K with sub–millikelvin stability is needed. We have been developing JT cryocoolers for cryogenic thermometer calibration for more than ten years.3-6 The latest version of the cryocooler consists of a 4 K two–stage pulse tube refrigerator and a Joule–Thomson (JT) circuit. The detailed construction of the JT cryocooler and cooling characteristics, including the observed temperature instability are presented in this paper.

The cryocoolers for cryogenic detectors and instruments are usually designed to be operated at a specific target temperature with sufficient cooling power that are determined by the detectors and instruments.7-10 On the other hand, the cryocoolers for the cryogenic temperature standard work are required to be operated at relatively wide temperature range and high temperature control stability. The requirement for the cooling power is usually not very high.

APPARATUS Cryocooler design

The cryocooler for the thermometer calibration consists of a commercial 4 K two–stage pulse tube refrigerator (SHI SPR-052A) and a JT circuit developed at National Metrology Institute of Japan, National Institute of Advanced Industrial Science and Technology. Figure 1 shows the exterior view of the cryocooler. Figure 2 shows a schematic view of the cryocooler. The diameter of the vacuum chamber is 416 mm, and the height of the chamber is 940 mm. The cooling power of the first and second stages of the pulse tube cryocooler is 10 W at 45 K and 0.5W at 4.2 K, respectively. The first stage of the pulse tube refrigerator is thermally connected to the first radiation shield with a copper thermal link and cools the shield. The second stage of the pulse tube refrigerator cools the second radiation shield. The second radiation shield is designed to be an inner vacuum–tight chamber. A heat exchange gas can be introduced through a tube (not shown in Figure 2), which connects the second radiation shield and a heat exchange gas reservoir set outside the vacuum chamber. The heat exchange gas is evacuated by a turbo molecular pumping system at an appropriate timing that is just before the completion of the precooling. The heat exchange gas accelerates the precooling of the components inside the

Figure 1. Exterior view of the pulse tube / 3He JT cryocooler for thermometer calibration.

shield from room temperature to cryogenic temperature, which typically takes 30 h.

3

The JT circuit uses 3He as the working gas. The gas handling system (GHS) supplies room temperature 3He to the high–pressure side of the JT circuit and evacuates the 3He pot through the low–pressure side of the circuit. The details of the GHS are also presented in a separate section. The room temperature 3He supplied to the high–pressure side of the circuit is cooled at the first and second stages of the pulse tube refrigerator when it flows through the copper tubes coiled around the copper blocks attached to the first and second stages. The 3He is further cooled by the counterflow heat exchanger in the second radiation shield. This JT circuit has only one counterflow heat exchanger. This configuration helps to suppress the pressure drop across the low–pressure side of the JT circuit from the 3He pot to the GHS set at room temperature to achieve low vapor pressure at the 3He pot. The drawback of the single counterflow heat exchanger configuration is that precooling of the 3He before the JT expansion takes place only immaturely and a reduction in the cooling power of the JT circuit. Because the required cooling power to the JT circuit for thermometer calibrations is relatively small (order of mW), the reduction in the cooling power is acceptable.

The 3He precooled by the counterflow heat exchanger expands as it flows through a needle valve. This is an isenthalpic expansion. The impedance of the needle valve can be varied by operating a handle set on the top flange of the vacuum chamber. The pressure drop across the needle valve can be varied by changing the impedance of the needle valve. This means that the temperature at the 3He pot can be coarsely set by changing the intensity of the JT expansion effect. Part of the expanded 3He condenses into the 3He pot as liquid under appropriate conditions. The vacuum pumps in the GHS evacuate the 3He pot through the low–pressure side of the counter heat exchanger. The evacuated 3He circulated the JT circuit. The temperatures at the first and second stages of the pulse tube refrigerator in normal operation with the JT circuit operated are approximately 43 K and 3.1 K, respectively. Counterflow heat exchanger

The construction of the counterflow heat exchanger is a coiled capillary in a tube type. This type of counterflow heat exchanges has been successfully applied for dilution refrigerators.11-13 The capillary in the counterflow heat exchanger for the present JT cryocooler is coiled in a twisted spiral shape.4 High–pressure gas flows in the twisted spiral copper–nickel capillary of 0.3 mm inner diameter, 0.5 mm outer diameter, and 4 m length. The capillary of 4 m can be

Figure 2. Schematic representation of the pulse tube / 3He JT cryocooler for thermometer calibration.

COMMERCIAL AND LABORATORY APPLICATIONS 68

Page 3: Performance Analysis of Pulse Tube/3He Joule– Thomson

2

resistance thermometers require calibration, they are compact, stable, easy-to-handle, and widely used as practical thermometry devices for many experiments.1

The International Temperature Scale of 1990 (ITS-90) consists of defining temperature fixed points and interpolation instruments that provide approximated thermodynamic temperature between fixed points.2 ITS-90 defines the temperature scale from 0.65 K to the highest temperature practically measurable in terms of the Plank radiation law using monochromatic radiation. The ITS-90 is designed to be the reference temperature scale for the calibration of secondary thermometers with highest accuracy.

Our institute realized, maintained, and disseminated the abovementioned temperature scale.3 For the cryogenic temperature standard work, a cryocooler that can control the temperature of a thermometer comparison block between 0.65 K and 25 K with sub–millikelvin stability is needed. We have been developing JT cryocoolers for cryogenic thermometer calibration for more than ten years.3-6 The latest version of the cryocooler consists of a 4 K two–stage pulse tube refrigerator and a Joule–Thomson (JT) circuit. The detailed construction of the JT cryocooler and cooling characteristics, including the observed temperature instability are presented in this paper.

The cryocoolers for cryogenic detectors and instruments are usually designed to be operated at a specific target temperature with sufficient cooling power that are determined by the detectors and instruments.7-10 On the other hand, the cryocoolers for the cryogenic temperature standard work are required to be operated at relatively wide temperature range and high temperature control stability. The requirement for the cooling power is usually not very high.

APPARATUS Cryocooler design

The cryocooler for the thermometer calibration consists of a commercial 4 K two–stage pulse tube refrigerator (SHI SPR-052A) and a JT circuit developed at National Metrology Institute of Japan, National Institute of Advanced Industrial Science and Technology. Figure 1 shows the exterior view of the cryocooler. Figure 2 shows a schematic view of the cryocooler. The diameter of the vacuum chamber is 416 mm, and the height of the chamber is 940 mm. The cooling power of the first and second stages of the pulse tube cryocooler is 10 W at 45 K and 0.5W at 4.2 K, respectively. The first stage of the pulse tube refrigerator is thermally connected to the first radiation shield with a copper thermal link and cools the shield. The second stage of the pulse tube refrigerator cools the second radiation shield. The second radiation shield is designed to be an inner vacuum–tight chamber. A heat exchange gas can be introduced through a tube (not shown in Figure 2), which connects the second radiation shield and a heat exchange gas reservoir set outside the vacuum chamber. The heat exchange gas is evacuated by a turbo molecular pumping system at an appropriate timing that is just before the completion of the precooling. The heat exchange gas accelerates the precooling of the components inside the

Figure 1. Exterior view of the pulse tube / 3He JT cryocooler for thermometer calibration.

shield from room temperature to cryogenic temperature, which typically takes 30 h.

3

The JT circuit uses 3He as the working gas. The gas handling system (GHS) supplies room temperature 3He to the high–pressure side of the JT circuit and evacuates the 3He pot through the low–pressure side of the circuit. The details of the GHS are also presented in a separate section. The room temperature 3He supplied to the high–pressure side of the circuit is cooled at the first and second stages of the pulse tube refrigerator when it flows through the copper tubes coiled around the copper blocks attached to the first and second stages. The 3He is further cooled by the counterflow heat exchanger in the second radiation shield. This JT circuit has only one counterflow heat exchanger. This configuration helps to suppress the pressure drop across the low–pressure side of the JT circuit from the 3He pot to the GHS set at room temperature to achieve low vapor pressure at the 3He pot. The drawback of the single counterflow heat exchanger configuration is that precooling of the 3He before the JT expansion takes place only immaturely and a reduction in the cooling power of the JT circuit. Because the required cooling power to the JT circuit for thermometer calibrations is relatively small (order of mW), the reduction in the cooling power is acceptable.

The 3He precooled by the counterflow heat exchanger expands as it flows through a needle valve. This is an isenthalpic expansion. The impedance of the needle valve can be varied by operating a handle set on the top flange of the vacuum chamber. The pressure drop across the needle valve can be varied by changing the impedance of the needle valve. This means that the temperature at the 3He pot can be coarsely set by changing the intensity of the JT expansion effect. Part of the expanded 3He condenses into the 3He pot as liquid under appropriate conditions. The vacuum pumps in the GHS evacuate the 3He pot through the low–pressure side of the counter heat exchanger. The evacuated 3He circulated the JT circuit. The temperatures at the first and second stages of the pulse tube refrigerator in normal operation with the JT circuit operated are approximately 43 K and 3.1 K, respectively. Counterflow heat exchanger

The construction of the counterflow heat exchanger is a coiled capillary in a tube type. This type of counterflow heat exchanges has been successfully applied for dilution refrigerators.11-13 The capillary in the counterflow heat exchanger for the present JT cryocooler is coiled in a twisted spiral shape.4 High–pressure gas flows in the twisted spiral copper–nickel capillary of 0.3 mm inner diameter, 0.5 mm outer diameter, and 4 m length. The capillary of 4 m can be

Figure 2. Schematic representation of the pulse tube / 3He JT cryocooler for thermometer calibration.

ANALYSIS OF PT/ 3HE J -T CRYOCOOLER FOR CALIBRATION 69

Page 4: Performance Analysis of Pulse Tube/3He Joule– Thomson

4

Figure 3. Two types of commercial capsule-type standard resistance thermometers

coiled approximately 100 mm in length and is inserted in a stainless-steel outer tube of 14.4 mm inner diameter, 15 mm outer diameter, and 200 mm length. The low–pressure gas flows in the space between the twisted spiral capillary and the outer tube. For confirmation purposes, the pressure drop across the low–pressure side of the counter flow exchanger is measured before the installation of the JT circuit. It is 0.3 Pa for 300 μmol/s of 4He flow rate at room temperature. The typical 3He flow rate in actual operations ranges from 100 μmol/s to 200 μmol/s.

The temperature distribution around the counterflow heat exchanger is measured using thermometers (Cernox temperature sensor calibrated by manufacture14) numbered from T1 to T5. T1 and T2 are inserted in the copper blocks soldered to the copper–nickel capillary. T3 is inserted in a hole drilled to the 3He pot. T4 and T5 are inserted in the copper blocks clamped to the stainless outer tube. The estimated temperature distribution measurement uncertainties are better than 50 mK. The thermometer number corresponds to the location number on the JT circuit. Thermometer T1 is attached to the location 1, thermometer T2 attached to location 2, and so on. Location 1 is located before the inlet of the high–pressure side of the counterflow heat exchanger. Location 2 is located between the outlet of the high–pressure side of the heat exchanger and the needle valve. Location 3 corresponds the 3He pot. Location 4 and 5 are located at inlet and outlet of the low–pressure side of the counterflow heat exchanger, respectively. The pressure on the high–pressure side of the JT circuit is measured with an absolute pressure gauge labeled PH, and that of the evacuated 3He gas is measured with an absolute pressure gauge labeled PL. The pressure drop across the high–pressure side of the counterflow heat exchanger is measured with a differential pressure gauge. The estimated pressure measurement uncertainties are better than 5 % of readings.

Thermometer comparison block and thermometer calibration

A thermometer comparison block is suspended in the second radiation shield with four strings to improve thermal and mechanical isolation from the second radiation shield. The comparison block and the 3He pot are thermally connected by a flexible copper thermal link. The comparison block is equipped with a resistive heater and thermometer for temperature control. The 3He pot also equips a heater and a thermometer.

The comparison block made of oxygen-free high-conductivity copper has wells for thermometer insertion and can hold up to 10 capsule-type standard resistance thermometers. Capsule-type standard resistance thermometers are specifically designed for precise and high stability cryogenic thermometry of millikelvin order or better measurement uncertainty. They have been used for precision thermometry, such as temperature standard work for several decades worldwide. A coil of wire made of a dilute alloy of iron in rhodium15 (RhFe) or cobalt in platinum16 (PtCo) is used as a sensing element. Both thermometers are used for the present work. Figure 3 shows examples of commercial capsule-type standard resistance thermometers. The coiled sensing element is sealed in a platinum capsule with helium heat exchange gas below atmospheric pressure. The typical size of the capsule is 5 mm in diameter and 60 mm in length. Four platinum lead wires were attached to the sensing element for four–wire resistance

5

measurements. The present cryocooler is primarily designed for the calibration of this type of the thermometer.

In a thermometer calibration procedure, two to three capsule–type standard resistance thermometers which have already been calibrated are installed in the comparison block as reference and confirmation thermometers. These thermometers are used to determined and confirm the temperature of the comparison block. The thermometers to be calibrated are also installed in the same comparison block. First, the temperature of the comparison block is stabilized at a target calibration temperature, and then the resistances of the pre–calibrated and non-calibrated thermometers are measured consecutively. The resistance of the thermometers to be calibrated at the calibration temperature is then obtained. This measurement is repeated at all calibration temperatures.

The heat dissipation from a capsule–type standard resistance thermometer is only several μW or less. The heat dissipations from the resistive heaters attached to the comparison block and the 3He pot for temperature control are less than 0.1 mW in the temperature range at which the JT circuit is running. The cooling power requirement to the JT circuit is not very high for the thermometer evaluation and calibration. The cooling power of the JT circuit depends on the operating condition; it is typically measured to be 1 mW at 0.65 K. On the other hand, requirements for the control temperature stability and operational temperature range are relatively high.

Gas handling system

The GHS circulates 3He through the JT circuit. The GHS supplies room temperature 3He to the high–pressure side of the JT circuit, evacuates 3He from the 3He pot, purifies and, pressurizes the 3He, and then returns 3He to the high–pressure side of the JT circuit. A bellows compressor in the GHS compresses room temperature 3He. Typically, a pressure ratio of 3.6 can be obtained with this compressor. The JT circuit can be operated in various ways. The supply pressure of 3He ranges from approximately 40 kPa to 120 kPa and varies depending on the settings and working conditions.

For example, 3He of up to 120 kPa is temporally supplied to the JT circuit for initial liquification. The supply pressure is reduced to a range of approximately 50 kPa in normal operation. Because the JT circuit is a closed circuit, the supply pressure cannot be set completely independent of other parameters. Practically, it is set by changing several controllable parameters such as openings of valves in the circuit, including the JT needle valve and the amount of circulating gas. The bellows compressor always operates at a constant frequency. It is not regarded as a controllable parameter in the system. A mechanical booster pump and an oil rotary vacuum pump evacuates the 3He pot. According to manufacture’s specifications, the pumping speeds of the pumps are 505 m3/h and 35 m3 /h, respectively. The evacuation speed on the 3He pot can be regulated with a butterfly valve, pressure gauge, and a proportional–integral–derivative (PID) controller. The pressure at the low–pressure side of the JT circuit typically ranges from 8 Pa to 15 kPa. A liquid-nitrogen cooled cold trap and a 25 l gas reservoir tank (not shown in Figure 2) are included in the GHS. The reservoir tank stores the 3He gas when the JT circuit is not in operation and is also used as a buffering volume to regulate the amount of the circulating gas. A cold trap cooled by the second stage of the pulse tube refrigerator is placed in the vacuum chamber.

Temperature control

The cryocooler is used for the thermometer calibrations between 0.65 K and 25 K. The thermometer comparison block needs to be controlled in this temperature range with sub–millikelvin stability. The temperature of the comparison block is set in two steps. First, the temperature of the 3He pot is controlled at a certain temperature, and then the temperature of the comparison block is set and controlled at several mK higher than that of the 3He pot.

COMMERCIAL AND LABORATORY APPLICATIONS 70

Page 5: Performance Analysis of Pulse Tube/3He Joule– Thomson

4

Figure 3. Two types of commercial capsule-type standard resistance thermometers

coiled approximately 100 mm in length and is inserted in a stainless-steel outer tube of 14.4 mm inner diameter, 15 mm outer diameter, and 200 mm length. The low–pressure gas flows in the space between the twisted spiral capillary and the outer tube. For confirmation purposes, the pressure drop across the low–pressure side of the counter flow exchanger is measured before the installation of the JT circuit. It is 0.3 Pa for 300 μmol/s of 4He flow rate at room temperature. The typical 3He flow rate in actual operations ranges from 100 μmol/s to 200 μmol/s.

The temperature distribution around the counterflow heat exchanger is measured using thermometers (Cernox temperature sensor calibrated by manufacture14) numbered from T1 to T5. T1 and T2 are inserted in the copper blocks soldered to the copper–nickel capillary. T3 is inserted in a hole drilled to the 3He pot. T4 and T5 are inserted in the copper blocks clamped to the stainless outer tube. The estimated temperature distribution measurement uncertainties are better than 50 mK. The thermometer number corresponds to the location number on the JT circuit. Thermometer T1 is attached to the location 1, thermometer T2 attached to location 2, and so on. Location 1 is located before the inlet of the high–pressure side of the counterflow heat exchanger. Location 2 is located between the outlet of the high–pressure side of the heat exchanger and the needle valve. Location 3 corresponds the 3He pot. Location 4 and 5 are located at inlet and outlet of the low–pressure side of the counterflow heat exchanger, respectively. The pressure on the high–pressure side of the JT circuit is measured with an absolute pressure gauge labeled PH, and that of the evacuated 3He gas is measured with an absolute pressure gauge labeled PL. The pressure drop across the high–pressure side of the counterflow heat exchanger is measured with a differential pressure gauge. The estimated pressure measurement uncertainties are better than 5 % of readings.

Thermometer comparison block and thermometer calibration

A thermometer comparison block is suspended in the second radiation shield with four strings to improve thermal and mechanical isolation from the second radiation shield. The comparison block and the 3He pot are thermally connected by a flexible copper thermal link. The comparison block is equipped with a resistive heater and thermometer for temperature control. The 3He pot also equips a heater and a thermometer.

The comparison block made of oxygen-free high-conductivity copper has wells for thermometer insertion and can hold up to 10 capsule-type standard resistance thermometers. Capsule-type standard resistance thermometers are specifically designed for precise and high stability cryogenic thermometry of millikelvin order or better measurement uncertainty. They have been used for precision thermometry, such as temperature standard work for several decades worldwide. A coil of wire made of a dilute alloy of iron in rhodium15 (RhFe) or cobalt in platinum16 (PtCo) is used as a sensing element. Both thermometers are used for the present work. Figure 3 shows examples of commercial capsule-type standard resistance thermometers. The coiled sensing element is sealed in a platinum capsule with helium heat exchange gas below atmospheric pressure. The typical size of the capsule is 5 mm in diameter and 60 mm in length. Four platinum lead wires were attached to the sensing element for four–wire resistance

5

measurements. The present cryocooler is primarily designed for the calibration of this type of the thermometer.

In a thermometer calibration procedure, two to three capsule–type standard resistance thermometers which have already been calibrated are installed in the comparison block as reference and confirmation thermometers. These thermometers are used to determined and confirm the temperature of the comparison block. The thermometers to be calibrated are also installed in the same comparison block. First, the temperature of the comparison block is stabilized at a target calibration temperature, and then the resistances of the pre–calibrated and non-calibrated thermometers are measured consecutively. The resistance of the thermometers to be calibrated at the calibration temperature is then obtained. This measurement is repeated at all calibration temperatures.

The heat dissipation from a capsule–type standard resistance thermometer is only several μW or less. The heat dissipations from the resistive heaters attached to the comparison block and the 3He pot for temperature control are less than 0.1 mW in the temperature range at which the JT circuit is running. The cooling power requirement to the JT circuit is not very high for the thermometer evaluation and calibration. The cooling power of the JT circuit depends on the operating condition; it is typically measured to be 1 mW at 0.65 K. On the other hand, requirements for the control temperature stability and operational temperature range are relatively high.

Gas handling system

The GHS circulates 3He through the JT circuit. The GHS supplies room temperature 3He to the high–pressure side of the JT circuit, evacuates 3He from the 3He pot, purifies and, pressurizes the 3He, and then returns 3He to the high–pressure side of the JT circuit. A bellows compressor in the GHS compresses room temperature 3He. Typically, a pressure ratio of 3.6 can be obtained with this compressor. The JT circuit can be operated in various ways. The supply pressure of 3He ranges from approximately 40 kPa to 120 kPa and varies depending on the settings and working conditions.

For example, 3He of up to 120 kPa is temporally supplied to the JT circuit for initial liquification. The supply pressure is reduced to a range of approximately 50 kPa in normal operation. Because the JT circuit is a closed circuit, the supply pressure cannot be set completely independent of other parameters. Practically, it is set by changing several controllable parameters such as openings of valves in the circuit, including the JT needle valve and the amount of circulating gas. The bellows compressor always operates at a constant frequency. It is not regarded as a controllable parameter in the system. A mechanical booster pump and an oil rotary vacuum pump evacuates the 3He pot. According to manufacture’s specifications, the pumping speeds of the pumps are 505 m3/h and 35 m3 /h, respectively. The evacuation speed on the 3He pot can be regulated with a butterfly valve, pressure gauge, and a proportional–integral–derivative (PID) controller. The pressure at the low–pressure side of the JT circuit typically ranges from 8 Pa to 15 kPa. A liquid-nitrogen cooled cold trap and a 25 l gas reservoir tank (not shown in Figure 2) are included in the GHS. The reservoir tank stores the 3He gas when the JT circuit is not in operation and is also used as a buffering volume to regulate the amount of the circulating gas. A cold trap cooled by the second stage of the pulse tube refrigerator is placed in the vacuum chamber.

Temperature control

The cryocooler is used for the thermometer calibrations between 0.65 K and 25 K. The thermometer comparison block needs to be controlled in this temperature range with sub–millikelvin stability. The temperature of the comparison block is set in two steps. First, the temperature of the 3He pot is controlled at a certain temperature, and then the temperature of the comparison block is set and controlled at several mK higher than that of the 3He pot.

ANALYSIS OF PT/ 3HE J -T CRYOCOOLER FOR CALIBRATION 71

Page 6: Performance Analysis of Pulse Tube/3He Joule– Thomson

6

Table 1. 3He pot temperature control methods.

Figure 4. Temperature control stability at the thermometer comparison block between 0.65 K and 25 K. Data A) corresponds to the comparison block temperature at TBlock = 1.50 K. Data B) also corresponds to TBlock = 1.50 K but obtained with different flow setting from that of Data A). Data C) corresponds to TBlock = 1.84 K.

Table 1 shows the three methods employed for the 3He pot temperature control depending on the calibration temperature. The first method is employed in the highest calibration temperature range from 25 K to 5.5 K. The 3He pot is only cooled by the pulse tube refrigerator and the temperature is controlled with a heater attached to the 3He pot. The current to the heater is fed by a commercial temperature controller for cryogenic thermometry and temperature control.

For a middle–temperature range from 5.5 K to 1.8 K, the JT circuit is operated in addition to the pulse tube refrigerator to cool the 3He pot. In this temperature range except near 2 K, liquid 3He does not appear in the 3He pot. The heater attached to the 3He pot is used for temperature control. It is found that, near 2 K, the temperature control turns relatively unstable. This temperature instability may be attributed to the intermittent appearance of the liquid phase at a certain point in the JT circuit. The data are shown in the next section.

For the lowest temperature range from 1.8 K to 0.65 K, the pulse tube refrigerator and JT circuit were used for cooling, but the butterfly valve is used for the 3He pot temperature control. This is because liquid is stable in the 3He pot in this temperature range.

On the other hand, the temperature control of the comparison block is straightforward. It is controlled with a heater attached to the comparison block for the entire temperature range. A bridge for precision thermometry is utilized for temperature control.17 This control method can result in sub–millikelvin temperature stability. The resistance of the thermometer set in the comparison block for temperature control is measured with the bridge, and then a signal related to the deviation from a pre-assigned resistance value that relates to a target calibration temperature can be obtained. The deviation signal is fed to a PID controller to drive the heater of the comparison block to eliminate the deviation.

PPERFORMANCE ANALYSIS

Temperature Control Stability

The temperature control stability at the comparison block measured with a capsule-type standard resistance thermometer is shown in Figure 4. The abscissa represents the comparison block temperature. The ordinate represents the standard deviation of the measured temperature variation at each comparison block temperature, which ranges from 0.65 K to 25 K for a

Calibration temperature Cooling source Control method

High(25 K–5.5 K) Pulse tube refrigerator Heater on 3He potMiddle(5.5 K–1.8 K) Pulse tube refrigerator and JT Heater on 3He potLow(1.8 K–0.65 K) Pulse tube refrigerator and JT Butterfly valve (Vapor pressure in 3He pot)

7

Figure 5. Enthalpy–temperature (h–T) diagrams of the JT circuit for different temperature stability conditions. Estimated states of 3He at numbered locations in the JT circuit as indicated in Figure 2 are shown. The saturated liquid and vapor lines, and isobars corresponded to the pressures at numbered locations are also shown.

measurement duration of 15 min. Before measurement at each block temperature, the temperature of the comparison block is set at a target temperature and wait for stabilization for hours, typically it takes 8 to 10 h. The standard deviation varied depending on the block temperature. However, sub–millikelvin stability (less than 0.07 mK) is achieved. The stability is sufficient for the thermometer calibrations. However, notable instability is observed near 2 K. This or similar instabilities have been pointed out by researchers to relate to the two–phase flow in the JT circuit.3,9

The data labeled A) corresponds to the comparison block temperature at TBlock = 1.50 K. Data B) also corresponds to TBlock = 1.50 K. Data C) corresponds to TBlock = 1.84 K. The difference in the measurement conditions between data A) and B) is the needle valve setting. The needle valve is set at a high impedance (4 x 1012 cm-3) position so that the pressure before the expansion sufficiently increased and liquid stably exists in the 3He pot for data A). On the other hand, the needle valve is set at a low impedance (7 x 109 cm-3) position for B), and therefore, the pressure before the expansion decreased and the liquid does not stably exist in the 3He pot under this condition. For data C), the needle valve is set at the same low impedance position.

Enthalpy–temperature diagram

Figure 5 shows enthalpy–temperature (h–T) diagrams18 of the JT circuit for different temperature stability conditions. Estimated states of 3He at numbered locations in the JT circuit as indicated in Figure 2 are shown in these diagrams. The properties of 3He are obtained by

COMMERCIAL AND LABORATORY APPLICATIONS 72

Page 7: Performance Analysis of Pulse Tube/3He Joule– Thomson

6

Table 1. 3He pot temperature control methods.

Figure 4. Temperature control stability at the thermometer comparison block between 0.65 K and 25 K. Data A) corresponds to the comparison block temperature at TBlock = 1.50 K. Data B) also corresponds to TBlock = 1.50 K but obtained with different flow setting from that of Data A). Data C) corresponds to TBlock = 1.84 K.

Table 1 shows the three methods employed for the 3He pot temperature control depending on the calibration temperature. The first method is employed in the highest calibration temperature range from 25 K to 5.5 K. The 3He pot is only cooled by the pulse tube refrigerator and the temperature is controlled with a heater attached to the 3He pot. The current to the heater is fed by a commercial temperature controller for cryogenic thermometry and temperature control.

For a middle–temperature range from 5.5 K to 1.8 K, the JT circuit is operated in addition to the pulse tube refrigerator to cool the 3He pot. In this temperature range except near 2 K, liquid 3He does not appear in the 3He pot. The heater attached to the 3He pot is used for temperature control. It is found that, near 2 K, the temperature control turns relatively unstable. This temperature instability may be attributed to the intermittent appearance of the liquid phase at a certain point in the JT circuit. The data are shown in the next section.

For the lowest temperature range from 1.8 K to 0.65 K, the pulse tube refrigerator and JT circuit were used for cooling, but the butterfly valve is used for the 3He pot temperature control. This is because liquid is stable in the 3He pot in this temperature range.

On the other hand, the temperature control of the comparison block is straightforward. It is controlled with a heater attached to the comparison block for the entire temperature range. A bridge for precision thermometry is utilized for temperature control.17 This control method can result in sub–millikelvin temperature stability. The resistance of the thermometer set in the comparison block for temperature control is measured with the bridge, and then a signal related to the deviation from a pre-assigned resistance value that relates to a target calibration temperature can be obtained. The deviation signal is fed to a PID controller to drive the heater of the comparison block to eliminate the deviation.

PPERFORMANCE ANALYSIS

Temperature Control Stability

The temperature control stability at the comparison block measured with a capsule-type standard resistance thermometer is shown in Figure 4. The abscissa represents the comparison block temperature. The ordinate represents the standard deviation of the measured temperature variation at each comparison block temperature, which ranges from 0.65 K to 25 K for a

Calibration temperature Cooling source Control method

High(25 K–5.5 K) Pulse tube refrigerator Heater on 3He potMiddle(5.5 K–1.8 K) Pulse tube refrigerator and JT Heater on 3He potLow(1.8 K–0.65 K) Pulse tube refrigerator and JT Butterfly valve (Vapor pressure in 3He pot)

7

Figure 5. Enthalpy–temperature (h–T) diagrams of the JT circuit for different temperature stability conditions. Estimated states of 3He at numbered locations in the JT circuit as indicated in Figure 2 are shown. The saturated liquid and vapor lines, and isobars corresponded to the pressures at numbered locations are also shown.

measurement duration of 15 min. Before measurement at each block temperature, the temperature of the comparison block is set at a target temperature and wait for stabilization for hours, typically it takes 8 to 10 h. The standard deviation varied depending on the block temperature. However, sub–millikelvin stability (less than 0.07 mK) is achieved. The stability is sufficient for the thermometer calibrations. However, notable instability is observed near 2 K. This or similar instabilities have been pointed out by researchers to relate to the two–phase flow in the JT circuit.3,9

The data labeled A) corresponds to the comparison block temperature at TBlock = 1.50 K. Data B) also corresponds to TBlock = 1.50 K. Data C) corresponds to TBlock = 1.84 K. The difference in the measurement conditions between data A) and B) is the needle valve setting. The needle valve is set at a high impedance (4 x 1012 cm-3) position so that the pressure before the expansion sufficiently increased and liquid stably exists in the 3He pot for data A). On the other hand, the needle valve is set at a low impedance (7 x 109 cm-3) position for B), and therefore, the pressure before the expansion decreased and the liquid does not stably exist in the 3He pot under this condition. For data C), the needle valve is set at the same low impedance position.

Enthalpy–temperature diagram

Figure 5 shows enthalpy–temperature (h–T) diagrams18 of the JT circuit for different temperature stability conditions. Estimated states of 3He at numbered locations in the JT circuit as indicated in Figure 2 are shown in these diagrams. The properties of 3He are obtained by

ANALYSIS OF PT/ 3HE J -T CRYOCOOLER FOR CALIBRATION 73

Page 8: Performance Analysis of Pulse Tube/3He Joule– Thomson

8

utilizing the He3Pak Ver. 1.2 Software.19,20 The saturated liquid and vapor lines and isobars corresponding to pressures at numbered locations are also shown in the diagrams. The pressure at location 1 is represented by a dashed isobar line. Location 2 which corresponds the pressure before the JT expansion is represented by a solid isobar line. The pressures at the 3He pot, location 4 and location 5 are represented by the dashed-and-dotted isobar line.

The h–T diagram A) corresponds to data A) of TBlock = 1.50 K shown in the Figure 4. The h–T diagrams B) and C) show the h–T diagrams for the data B) of TBlock = 1.50 K and data C) of TBlock = 1.84 K shown in the Figure 4, respectively. Diagram A) corresponds to the data obtained when the temperature of the comparison block is stably controlled. As already stated in the former subsection, liquid stably existed in the 3He pot in this flow condition. The pressure in the 3He pot is controlled at the saturated vapor pressure. h–T diagrams B) and C) correspond to the data obtained when the comparison block temperature is relatively unstable. Because a liquid phase could not exist in the 3He pot in the flow conditions, the pressure at the 3He pot is evacuated at 0.08 kPa, which is lower than the saturated vapor pressure.

In the two diagrams for unstable temperature control conditions B) and C), the JT expansions start from the wet region or very close to the saturated vapor line to outside the wet region, which is indicated by solid arrows from point 2 to point 3 in the h–T diagrams.

In other words, when JT expansion crosses the saturated vapor line from the wet region, the temperature becomes unstable. In this expansion condition, liquid may appear intermittently. This can lead to temperature instability. In the diagrams for stable condition A), the JT expansion starts and ends within the wet region, as indicated by a dashed arrow from point 2 to point 3 in the h–T diagram.

During the temperature distribution measurements, it is observed that the temperature at point 5 is slightly higher (by a few tens mK) than that at point 1. The temperature difference is almost the same magnitude as the measurement uncertainty. If the counterflow heat exchanger is perfectly thermally isolated from the surroundings and thermally balanced, the temperature at point 5 cannot exceed that at point 1. However, the actual counterflow heat exchanger of the present JT cryocooler is not thermally isolated. There is certain amount of conducting heat flow through the walls of the inner capillary and outer tube that form the counterflow heat exchanger. There are differences in material and diameter between those of the inner capillary and the outer tube, and they lead the difference in the heat flow. These heat flow may affect the temperature distribution measurement. However, further investigation is needed.

It is interesting to note that the flow impedance of the high–pressure side of the present counterflow heat exchange is relatively high because a fine and long capillary is used. If pressure of the high–pressure side increased, frictional pressure drop increase and distributed Joule–Thomson effect18,21-23 may more become noticeable. This also needs further investigation.

CONCLUSION

The Pulse tube/3He JT cryocooler is successfully used for precise thermometer calibration. A temperature control stability of better than 0.07 mK in standard deviation at the thermometer comparison block is obtained. The temperature instability is observed near 2 K. This can be attributed to the two-phase flow in the JT circuit. When JT expansion crosses the saturated vapor line from the wet region, the temperature becomes unstable.

ACKNOWLEDGMENT

The author wishes to thank Dr. H. Nakagawa for variable discussions. This work was partly supported by JSPS KAKENHI Grant Number JP20K04319.

9

REFERENCES

1. Ripple, D. C., Davis, R., Fellmuth, B., Fischer, J., Machin, G., Quinn, T., Steur, P., Tamura, O., White, D. R., “The Roles of the Mise en Pratique for the Definition of the Kelvin,” Int J Thermophys., vol. 31, (2010), pp. 1795-1808.

2. Preston-Thomas, H., “The International Temperature Scale of 1990,” Metrologia, vol. 27, (1990), pp. 3-10, 107.

3. Shimazaki, T., Toyoda, K., and Tamura, O., “Realization of the 3He Vapor-Pressure Temperature Scale and Development of a Liquid-He-Free Calibration Apparatus,” Int. J. Thermophys., vol. 32 (2011), pp. 2171-2182.

4. Shimazaki, T., Toyoda, K., and Tamura, O., “Gifford-McMahon/Joule-Thomson cryocooler with high-flow-conductance counterflow heat exchanger for use in resistance thermometer calibration,” Rev. Sci. Instrum, vol. 77, (2006), 034902.

5. Shimazaki, T., Toyoda, K., Oota. A., Nozato, H., Usuda, T., and Tamura, O., “Closed-Cycle Joule-Thomson Cryocooler for Resistance Thermometer Calibration down to 0.65 K,” Int. J. Thermophys., vol. 29 (2008), pp. 42-50.

6. Shimazaki, T., “Cooling Characteristics of 3He Joule-Thomson cryocooler,” Proceedings of ICEC 24-ICMC 2012, Cryogenics and Superconductivity Society of Japan, Tokyo (2013), pp. 435-438.

7. Crook, M., Bradshaw, T., Gilley, G., Hills, M., Watson, S., Green, B., Pulker, C., Rawlings, T., “Development of a 2 K Joule-Thomson Closed-Cycle Cryocooler,” Cryocoolers 19, International Cryocooler Conference, Inc., Boulder (2016), pp. 9-18.

8. Narasaki, K., Tsunematsu, S., Otsuka, K., Kanao, K., Okabayashi, A., Yoshida, S., Sugita, H., Sato, Y., Mitsuda, K., Nakagawa, T., Nishibori, T., “Lifetime Test and Heritage On-Orbit of SHI Coolers for Space Use,” Cryocoolers 19, International Cryocooler Conference, Inc., Boulder (2016), pp. 613-622.

9. Kotsubo, V., Ullom, J. N., and Nam, S. W., “Compact 1.7 K Cryocooler for Superconducting Nanowire Single-Photon Detectors,” Cryocoolers 20, International Cryocooler Conference, Inc., Boulder (2018), pp. 295-304.

10. Wu, Y., Zalewski, D. R., Vermeer, C. H., Doornink, J., Benthem, B., Boom, E., Ter Brake, H. J. M., “Sorption–Based Vibration–Free Cooler for the METIS Instrument on E–ELT,” Cryocoolers 17, International Cryocooler Conference, Inc., Boulder (2012), pp. 377-386.

11. Maeda, M., Shigematsu, T., Li, Z., Shigi, T., Fujii, Y., Yamaguchi, M., and Nakamura, M., “Experimental study of the dilution refrigerator without 1 K pot,” Proceedings of the Sixteenth International Cryogenic Engineering Conference / International Cryogenic Material Conference, Kitakyusyu, Elsevier Science, Oxford (1997), pp. 461-464.

12. Uhlig, K., “Cryogen-Free Dilution Refrigerator with 1 K-Stage,” Cryocoolers 17, International Cryocooler Conference, Inc., Boulder (2012), pp. 471-477.

13. Uhlig, K., “3He/4He Dilution Refrigerator without a Pumping 4He Stage,” Cryogenics, vol. 27 (1987), pp. 454-457.

14. LakeShore Cryotronics, https://www.Lakeshore.com/

15. Rusby, R. L., “Temperature: Its Measurement and Control in Science and Industry,” vol. 5, part. 2, AIP, New York (1982), pp. 829-834.

16. Shiratori, T., Mitsui, K., “Platinum-Cobalt Resistance Thermometer for Low Temperature Use,” Jpn. J. Appl. Phys., vol. 17 (1978), pp. 1289-1290.

17. White, G. K., and Meeson, P. J. Experimental Techniques in Low-Temperature Physics, Fourth Edition, Clarendon Press, Oxford (2002), pp. 130-133.

18. De Waele, A. T. A. M., “Basics of Joule-Thomson Liquefaction and JT Cooling,” J. Low. Temp. Phys., vol. 186 (2017), pp. 385-403.

19. He3Pak v.1.2 software, Horizon Technologies, Littleton, CO, USA, http://www.htess.com.

20. Huang, Y., Chen, G., Lai, B., Wang, S., “p–H and T–S diagrams of 3He from 0.2 K to 20 K,” Cryogenics, vol. 45 (2005), pp. 687-693.

COMMERCIAL AND LABORATORY APPLICATIONS 74

Page 9: Performance Analysis of Pulse Tube/3He Joule– Thomson

8

utilizing the He3Pak Ver. 1.2 Software.19,20 The saturated liquid and vapor lines and isobars corresponding to pressures at numbered locations are also shown in the diagrams. The pressure at location 1 is represented by a dashed isobar line. Location 2 which corresponds the pressure before the JT expansion is represented by a solid isobar line. The pressures at the 3He pot, location 4 and location 5 are represented by the dashed-and-dotted isobar line.

The h–T diagram A) corresponds to data A) of TBlock = 1.50 K shown in the Figure 4. The h–T diagrams B) and C) show the h–T diagrams for the data B) of TBlock = 1.50 K and data C) of TBlock = 1.84 K shown in the Figure 4, respectively. Diagram A) corresponds to the data obtained when the temperature of the comparison block is stably controlled. As already stated in the former subsection, liquid stably existed in the 3He pot in this flow condition. The pressure in the 3He pot is controlled at the saturated vapor pressure. h–T diagrams B) and C) correspond to the data obtained when the comparison block temperature is relatively unstable. Because a liquid phase could not exist in the 3He pot in the flow conditions, the pressure at the 3He pot is evacuated at 0.08 kPa, which is lower than the saturated vapor pressure.

In the two diagrams for unstable temperature control conditions B) and C), the JT expansions start from the wet region or very close to the saturated vapor line to outside the wet region, which is indicated by solid arrows from point 2 to point 3 in the h–T diagrams.

In other words, when JT expansion crosses the saturated vapor line from the wet region, the temperature becomes unstable. In this expansion condition, liquid may appear intermittently. This can lead to temperature instability. In the diagrams for stable condition A), the JT expansion starts and ends within the wet region, as indicated by a dashed arrow from point 2 to point 3 in the h–T diagram.

During the temperature distribution measurements, it is observed that the temperature at point 5 is slightly higher (by a few tens mK) than that at point 1. The temperature difference is almost the same magnitude as the measurement uncertainty. If the counterflow heat exchanger is perfectly thermally isolated from the surroundings and thermally balanced, the temperature at point 5 cannot exceed that at point 1. However, the actual counterflow heat exchanger of the present JT cryocooler is not thermally isolated. There is certain amount of conducting heat flow through the walls of the inner capillary and outer tube that form the counterflow heat exchanger. There are differences in material and diameter between those of the inner capillary and the outer tube, and they lead the difference in the heat flow. These heat flow may affect the temperature distribution measurement. However, further investigation is needed.

It is interesting to note that the flow impedance of the high–pressure side of the present counterflow heat exchange is relatively high because a fine and long capillary is used. If pressure of the high–pressure side increased, frictional pressure drop increase and distributed Joule–Thomson effect18,21-23 may more become noticeable. This also needs further investigation.

CONCLUSION

The Pulse tube/3He JT cryocooler is successfully used for precise thermometer calibration. A temperature control stability of better than 0.07 mK in standard deviation at the thermometer comparison block is obtained. The temperature instability is observed near 2 K. This can be attributed to the two-phase flow in the JT circuit. When JT expansion crosses the saturated vapor line from the wet region, the temperature becomes unstable.

ACKNOWLEDGMENT

The author wishes to thank Dr. H. Nakagawa for variable discussions. This work was partly supported by JSPS KAKENHI Grant Number JP20K04319.

9

REFERENCES

1. Ripple, D. C., Davis, R., Fellmuth, B., Fischer, J., Machin, G., Quinn, T., Steur, P., Tamura, O., White, D. R., “The Roles of the Mise en Pratique for the Definition of the Kelvin,” Int J Thermophys., vol. 31, (2010), pp. 1795-1808.

2. Preston-Thomas, H., “The International Temperature Scale of 1990,” Metrologia, vol. 27, (1990), pp. 3-10, 107.

3. Shimazaki, T., Toyoda, K., and Tamura, O., “Realization of the 3He Vapor-Pressure Temperature Scale and Development of a Liquid-He-Free Calibration Apparatus,” Int. J. Thermophys., vol. 32 (2011), pp. 2171-2182.

4. Shimazaki, T., Toyoda, K., and Tamura, O., “Gifford-McMahon/Joule-Thomson cryocooler with high-flow-conductance counterflow heat exchanger for use in resistance thermometer calibration,” Rev. Sci. Instrum, vol. 77, (2006), 034902.

5. Shimazaki, T., Toyoda, K., Oota. A., Nozato, H., Usuda, T., and Tamura, O., “Closed-Cycle Joule-Thomson Cryocooler for Resistance Thermometer Calibration down to 0.65 K,” Int. J. Thermophys., vol. 29 (2008), pp. 42-50.

6. Shimazaki, T., “Cooling Characteristics of 3He Joule-Thomson cryocooler,” Proceedings of ICEC 24-ICMC 2012, Cryogenics and Superconductivity Society of Japan, Tokyo (2013), pp. 435-438.

7. Crook, M., Bradshaw, T., Gilley, G., Hills, M., Watson, S., Green, B., Pulker, C., Rawlings, T., “Development of a 2 K Joule-Thomson Closed-Cycle Cryocooler,” Cryocoolers 19, International Cryocooler Conference, Inc., Boulder (2016), pp. 9-18.

8. Narasaki, K., Tsunematsu, S., Otsuka, K., Kanao, K., Okabayashi, A., Yoshida, S., Sugita, H., Sato, Y., Mitsuda, K., Nakagawa, T., Nishibori, T., “Lifetime Test and Heritage On-Orbit of SHI Coolers for Space Use,” Cryocoolers 19, International Cryocooler Conference, Inc., Boulder (2016), pp. 613-622.

9. Kotsubo, V., Ullom, J. N., and Nam, S. W., “Compact 1.7 K Cryocooler for Superconducting Nanowire Single-Photon Detectors,” Cryocoolers 20, International Cryocooler Conference, Inc., Boulder (2018), pp. 295-304.

10. Wu, Y., Zalewski, D. R., Vermeer, C. H., Doornink, J., Benthem, B., Boom, E., Ter Brake, H. J. M., “Sorption–Based Vibration–Free Cooler for the METIS Instrument on E–ELT,” Cryocoolers 17, International Cryocooler Conference, Inc., Boulder (2012), pp. 377-386.

11. Maeda, M., Shigematsu, T., Li, Z., Shigi, T., Fujii, Y., Yamaguchi, M., and Nakamura, M., “Experimental study of the dilution refrigerator without 1 K pot,” Proceedings of the Sixteenth International Cryogenic Engineering Conference / International Cryogenic Material Conference, Kitakyusyu, Elsevier Science, Oxford (1997), pp. 461-464.

12. Uhlig, K., “Cryogen-Free Dilution Refrigerator with 1 K-Stage,” Cryocoolers 17, International Cryocooler Conference, Inc., Boulder (2012), pp. 471-477.

13. Uhlig, K., “3He/4He Dilution Refrigerator without a Pumping 4He Stage,” Cryogenics, vol. 27 (1987), pp. 454-457.

14. LakeShore Cryotronics, https://www.Lakeshore.com/

15. Rusby, R. L., “Temperature: Its Measurement and Control in Science and Industry,” vol. 5, part. 2, AIP, New York (1982), pp. 829-834.

16. Shiratori, T., Mitsui, K., “Platinum-Cobalt Resistance Thermometer for Low Temperature Use,” Jpn. J. Appl. Phys., vol. 17 (1978), pp. 1289-1290.

17. White, G. K., and Meeson, P. J. Experimental Techniques in Low-Temperature Physics, Fourth Edition, Clarendon Press, Oxford (2002), pp. 130-133.

18. De Waele, A. T. A. M., “Basics of Joule-Thomson Liquefaction and JT Cooling,” J. Low. Temp. Phys., vol. 186 (2017), pp. 385-403.

19. He3Pak v.1.2 software, Horizon Technologies, Littleton, CO, USA, http://www.htess.com.

20. Huang, Y., Chen, G., Lai, B., Wang, S., “p–H and T–S diagrams of 3He from 0.2 K to 20 K,” Cryogenics, vol. 45 (2005), pp. 687-693.

ANALYSIS OF PT/ 3HE J -T CRYOCOOLER FOR CALIBRATION 75

Page 10: Performance Analysis of Pulse Tube/3He Joule– Thomson

21. Hwang, G., Jeong, S., “Pressure Loss Effect on Recuperative Heat Exchanger and Its Thermal Performance,” Cryogenics, vol. 50 (2010), pp. 13-17.

22. Maytal, B-Z., “Hampson’s Type Cryocooler with Distributed Joule–Thomson Effect for Mixed Refrigerants Closed Cycle,” Cryogenics, vol. 61 (2014), pp. 92-96.

23. Jeong, S., Park, C., Kim, K., “Design of Distributed JT (Joule–Thomson) Effect Heat Exchanger for Superfluid 2 K Cooling Device,” J. Phys.: Conf. Ser., vol. 969 (2018), 012084.

�� C21_09 9 1

Cryocooler Technology for ElectronParticle Accelerators

G. E. Lawler, N. Majernik, A. Fukasawa, Y. Sakai, and J.B. Rosenzweig

UCL A Department of Physics & AstronomyL os Angeles, CA 9 009 5

ABSTRACTParticle accelerator technology is ubiq uitous in modern discovery science. Accelerators are

no longer limited to the more well-known high energy physics applications. They are of vital im-portance to many diverse fields as far reaching as biology, chemistry, condensed matter physics, medicine, and industrial manufacturing procedures. It has then of the utmost importance to develop better understanding of the physics and engineering principles necessary for advanced particle ac-celerator design.

The Particle B eam Physics L aboratory (PB PL ) at the University of California, L os Angeles (UCL A) is at the forefront of this research. O ne major thrust in accelerator physics in general and PB PL in particular is the application of cryogenic technology to accelerators. The presentation will focus on the use of cryogenic technology in general accelerator R& D research as well as particular developments taking place at UCLA. Discussions will introduce general improvements of figures of merit such as beam brightness ex pected to come from cryogenic operations. In particular, three ex periments utilizing cryocoolers will be presented: two radiofreq uency (RF) resonant cavity tests of normal conducting copper at cryogenic temperatures, at low and high RF power, and one inte-grated test bed for studies of photoemission at cryogenic temperatures.

INTRODUCTIONMotivations for continued advancement of the field of particle accelerator physics often canoni-

cally make note of the essential role they have played in discovery science over the past century. Notable highlights include the successes in elementary particle physics at the Fermilab and CERN accelerator complex es, nuclear physics studies at B rookhaven National L aboratory and CERN, and high brilliance photon sources at synchrotron storage rings and free electron lasers such the Advanced Photon Source at Argonne National L ab and the L inac Coherent L ight Source (L CL S) at SLAC [1-4]. In many respects light sources are the most compelling as they benefit fields beyond basic physics. These machines, often electron accelerators, offer extensive capabilities for scientific users, making available large numbers of high energy photons allowing breakthroughs in material science, chemistry, biology and others. Medical and various industrial applications are also at the forefront of development motivation. Many if not all of these facilities are enormous in scale, with kilometers of space needed and costs in the billions of dollars for construction and maintenance. We thus find enormous incentive to reduce the size and cost of these devices.

COMMERCIAL AND LABORATORY APPLICATIONS 76