49

P igh erformance L iquid C hromatography HPLC is a form of liquid chromatography used to separate compounds that are dissolved in solution. HPLC instruments

Embed Size (px)

Citation preview

Page 1: P igh erformance L iquid C hromatography  HPLC is a form of liquid chromatography used to separate compounds that are dissolved in solution. HPLC instruments

Pigh

erformance

L iquid

C hromatography

HPLC is a form of liquid chromatography used to separate compounds that are dissolved in solution HPLC instruments consist of a reservoir of mobile phase a pump an injector a separation column and a detector

Compounds are separated by injecting a sample mixture onto the column The different component in the mixture pass through the column at differentiates due to differences in their partition behavior between the mobile phase and the stationary phase The mobile phase must be degassed to eliminate the formation of air bubbles

This unit is equipped

with two pump units

and a UVVis

detector

The gradient is

controlled via the

pump controllers

LC

methods are not

as sensitive to

temperature

Columns are

commonly

mounted outside

the instrument

Solvents

bull All solvents should be lsquoHPLCrsquo grade

ndash This is a type of reagent grade material

ndash It has been filtered using a 02 μm filter

bull Filtered solvent helps extend pump life by preventing scoring It also reduces the chances of a column plugging

Solvent degassing

All solvents should be degassed prior to use

This reduces the chances of bubbles being formed in the column or detector Oxygen present at high pressure can also cause a problem

Methods that can be used

Displacement with a less soluble gas

Applying a vacuum

Heating the solvent

HPLC Injectors

In order to introduce a sample onto the column for analysis a special valve called the injector must be used to transfer the sample into the pressurized system

Injectors may look different from the outside but internally most are 6-port rotary valves These valves consist of a fixed body (the stator) plus an internal seal that rotates (the rotor)

Three internal passages connect alternate pairs of external ports The valves can switch between two positions referred to as the inject and load positions respectively In the load position the pump is connected to the column and the sample inlet is connected to one end of a piece of tubing called the sample loop

The other end of the sample loop is connected to the waste port Rotation results in reconnecting the various lines that enter the valve so that a sample volume can be inserted into the mobile phase that flows from the pump to the column inlet

The column

HPLC has seen significant improvement over

the last 20 years primarily due to improved

column technology

Packings are more uniform and smaller

Phases are commonly chemically bound to

the packing

Packing methods have improved

Pump types

bullIsocratic pump - delivers constant mobile phase composition

bullsolvent must be pre-mixed

bulllowest cost pump

bullGradient pump -delivers variable mobile phase composition

bullcan be used to mix and deliver an isocratic mobile phase or a gradient mobile phase

1048729 Reciprocation pumps

--- advantages small internal volume high output pressures

ready adaptability to gradient elution and constant flow rates

--- disadvantages produce a pulsed flow cause baseline noise

1048729 Displacement pumps

--- advantages output is pulse free

--- disadvantages limited solvent capacity (lt250 mL)

inconvenience when change solvents

1048729 Pneumatic pumps

--- advantages inexpensive output is pulse free

--- disadvantages limited solvent capacity not amenable to

gradient elution and limit pressures lt2000 psi

Packings

Originally these were irregular silica and

alumina A range of synthetic regularly

shaped packings are now available

Porous - channels through packing

Superficially porous - rough surface

Smooth - bead like

Packing size

As packing size is decreased efficiency and

pressure requirements are increased

Common diameters for analytical work

diameter plates

10 μm 5000

5 μm 9000

3 μm 15000

All are for a 15 cm x 46 mm column

Column body

Typically consist of stainless steel with a high

precision internal bore

Some manufacturers offer column inserts

- donrsquot need to repurchase the column

fittings

Others offer columns where the external

body can be compressed to improve packing

efficiency

HPLC column examples

Column stationary phases

Today most packing fall into four classes

Silica or alumina

Bound phases on either alumina or silica

Gels

Controlled-pore glass or silica

Absorption phases - alumina

common mobile phases

hexane chloroform 2-propanol

example application - amines

silica

common mobile phases

hexane chloroform 2-propanol

example applications - ethers esters

porphyrins fat-soluble vitamins

Partition phases

Can be broken down into

Normal phase - polar materials bound to the support

Reverse phase - non-polar materials bound to the support

Mixed phase - may have some of each

Partition phases

Normal

Amino (-NH2)

Cyano (-CN)

Diol (glycidoxy-ethylmethoxysilane)

Reverse

C-2 or RP-2 (-Si-CH2CH3)

C-8 or RP-8 (-Si-(CH2)7CH3)

C-18 or RP-18 (-Si-(CH2)17CH3)

Increasing the C number results in a thicker more retentive phase

Ion exchange phases

Strong cation - sulfonic acid group

Strong anionic - quarternary amine

Weak anion - primary amine

Weak cation - COOH

Size exclusion phases

Gels - organic or aqueous based

Controlled-pore - silica or glass

Must be selected based on pressure

requirements and size range

required for your application

Capillary and Microbore columns

Several companies have begun offering

columns with smaller ID

Microbore column - 1 mm ID packed

column

Capillary column - lt 1 mm ID internal

bound phase

These columns require smaller solvent flows

reduced sample size and improved detector

design

Capillary and Microbore columns Capillary and Microbore columns

Aromatic Compounds

mobile phase2 ethylacetate in

hexane

flow rate 4 μlmin

column Fusica II 300μm ID x

25 cm silica

sample

1 toluene

2 nitrobenzene

3 acetophenone

4 26-dinitrobenzene

injection 60 nl

detection UV 254 nm

Silica based columns

Silica is the ideal support for HPLC columns It offers a large mechanical stability excellent

physicochemical surface properties a wide range of bonding chemistries and is compatible

with a broad range of organic solvents

pH stability

In general HPLC columns are stable within a pH range of 2 to 8 If you are measuring a pH value the measurement must be done in the aqueous media before mixing the eluent with organic solvents

Modern HPLC columns can be used outside that pH range The new bonding chemistries allow use down to pH 1 for some stationary phases However please check vendorrsquos product information before using silica based column outside the pH range of 2 to 8 However best lifetimes are obtained between pH 20 and pH 68

Mechanical stability

Stationary phases based on silica are mechanically very stable The packed columns show no pressure limit and can be used at more than 40 MPa (6000 psi) without any problem However please avoid pressure shocks on the column Pressure shocks lead to channelling in the column which results in peak splitting in the corresponding chromatogram

Mobile phases (Eluents)

Silica based stationary phases are compatible with all organic solvents in the above mentioned pH range

bull Filter all prepared buffer through a 05 μm filter before using them in your HPLC system

The use of non pure solvents in HPLC causes irreversible adsorption of impurities on the column head These impurities block adsorption sites change the selectivity of the column and lead to peak splitting in the chromatogram In gradient elution impurities cause so called ldquoGhost Peaksrdquo

Ghost peaks are peaks that always appear in the same position on the chromatogram Their origin is not the sample but the impurities from the solvents or solvent additives Therefore it is highly recommended to run a gradientwithout injection in the beginning of each method to determine the ghost peaks

To avoid irreversible adsorption at the column head you should always use a pre-column

The use of a pre-column increases the life time of a column dramatically In addition a pre-column can filter solid parts stemming from pump seals or injection rotors An alternative to a pre-column is an in-line filter These filters are attached directly to the column These filters get rid of solid parts in the eluent but will not avoid irreversible adsorption of organic impurities

Detector Systems

Virtually every chemical and physical

property that can be measured in

solution has been look at

Detectors fall roughly into two classes

Bulk property - measures an overall change in the mobile phase

Solute property - measures a solute specific property

Properties of a good detector

A detector must provide

bull high sensitivity low detection limits

bull linearity

bullReproducibility

This is true for any detector

Each detector will have specific advantages and will vary as to peak shape and spread noise and flowtemperature dependence they have

UVVis detector

A solute property detector

Sample must exhibit absorption in UVVis range Solvent must not absorb significantly at the measured wavelength

Types Filter photometer - single

bull Variable wavelength

bull Multiwavelength

Refractive index detector

Bulk property detector - general purpose

Based on refraction of light as it passes from

one media to another Presence of a solute

changes the refractive index of the solvent

Heat of absorption detector

A small amount of heat is released when a sample absorbs on a suitable surface

This detector can measure this

Electrochemical detectors

A number of properties have been evaluated

Detector types

Dielectic constant

Amperometric

Conductometric

Polarographic

Potentiometric

Dielectric constant detector

Bulk property detector

Measures changes in polarity of the liquid phase passing through the cell

Conductometric detector

Measures conductivity of the solvent Useful for

solutions of ions

Amperometric detectors

Most frequently applied type of electrochemical

detector

A known potential is applied across a set of

electrodes - typically a glassy carbon type

Ability to oxidize or reduce a species can be

measured

Typically limited to working with a specific class

of materials per analysis

Separation is based on the analytersquos relative solubility between two liquid phases

Stationary PhaseMobile Phase

Solvent Bonded Phase

Normal Phase- Polar stationary phase and non-polar solvent

bull Reverse Phase- Non-polar stationary phase and a polar solvent

Methan0l CH3OH

bull Acetonitrile CH3CN

bull Tetrahydrofuran

bull Water H2O

Solid Support - Backbone for bonded phases Usually 10micro 5micro or 3micro silica or polymeric

particles Bonded Phases - Functional groups firmly linked

(chemically bound) to the solid support Extremely stable Reproducible

Guard - Protects the analytical column Particles Interferences Prolongs the life of the analytical column

bull Analytical - Performs the separation

C-2 Ethyl Silyl -Si-CH2-CH3

bullCN Cyanopropyl Silyl -Si-(CH2)3-CN

bull C-18C-18 Octadecyl SilylOctadecyl Silyl -Si-(CH -Si-(CH22))1717--

CHCH33

bullC-8 Octyl Silyl -Si-(CH2)7-CH3

Pump

Injector

ColumnDetector

Mobile Phases

Gradient Controller

bull

UV Single wavelength (filter) [610 8330] Variable wavelength (monochromator)

[8316 8325] Multiple wavelengths (PDA) [555]

Fluorescence [610]Electrochemical [605] Mass Spectrometric [8325]

Restekreg ULTRA C-18 and CN Columns (250mm x 46mm 5micro)

Mobile Phase (11 MethanolWater) 15 mLmin

Supelcosil LC-PAH Columns A BConditions A 150mm x 46mm 5micro

Flow Rate 15 mLmin

Conditions B 50mm x 46mm 3micro

Flow Rate 30 mLmin

  • Slide 5
  • Slide 6
  • Slide 7
  • Slide 8
  • Slide 9
  • Slide 10
  • Slide 11
  • Slide 12
  • Slide 13
  • Slide 14
  • Slide 15
  • Slide 16
  • Slide 17
  • Slide 18
  • Slide 19
  • Slide 20
  • Slide 21
  • Slide 22
  • Slide 23
  • Slide 24
  • Slide 25
  • Slide 26
  • Slide 27
  • Slide 28
  • Slide 29
  • Slide 30
  • Slide 31
  • Slide 32
  • Slide 33
  • Slide 34
  • Slide 35
  • Slide 36
  • Slide 37
  • Slide 38
  • Slide 39
  • Slide 40
Page 2: P igh erformance L iquid C hromatography  HPLC is a form of liquid chromatography used to separate compounds that are dissolved in solution. HPLC instruments

HPLC is a form of liquid chromatography used to separate compounds that are dissolved in solution HPLC instruments consist of a reservoir of mobile phase a pump an injector a separation column and a detector

Compounds are separated by injecting a sample mixture onto the column The different component in the mixture pass through the column at differentiates due to differences in their partition behavior between the mobile phase and the stationary phase The mobile phase must be degassed to eliminate the formation of air bubbles

This unit is equipped

with two pump units

and a UVVis

detector

The gradient is

controlled via the

pump controllers

LC

methods are not

as sensitive to

temperature

Columns are

commonly

mounted outside

the instrument

Solvents

bull All solvents should be lsquoHPLCrsquo grade

ndash This is a type of reagent grade material

ndash It has been filtered using a 02 μm filter

bull Filtered solvent helps extend pump life by preventing scoring It also reduces the chances of a column plugging

Solvent degassing

All solvents should be degassed prior to use

This reduces the chances of bubbles being formed in the column or detector Oxygen present at high pressure can also cause a problem

Methods that can be used

Displacement with a less soluble gas

Applying a vacuum

Heating the solvent

HPLC Injectors

In order to introduce a sample onto the column for analysis a special valve called the injector must be used to transfer the sample into the pressurized system

Injectors may look different from the outside but internally most are 6-port rotary valves These valves consist of a fixed body (the stator) plus an internal seal that rotates (the rotor)

Three internal passages connect alternate pairs of external ports The valves can switch between two positions referred to as the inject and load positions respectively In the load position the pump is connected to the column and the sample inlet is connected to one end of a piece of tubing called the sample loop

The other end of the sample loop is connected to the waste port Rotation results in reconnecting the various lines that enter the valve so that a sample volume can be inserted into the mobile phase that flows from the pump to the column inlet

The column

HPLC has seen significant improvement over

the last 20 years primarily due to improved

column technology

Packings are more uniform and smaller

Phases are commonly chemically bound to

the packing

Packing methods have improved

Pump types

bullIsocratic pump - delivers constant mobile phase composition

bullsolvent must be pre-mixed

bulllowest cost pump

bullGradient pump -delivers variable mobile phase composition

bullcan be used to mix and deliver an isocratic mobile phase or a gradient mobile phase

1048729 Reciprocation pumps

--- advantages small internal volume high output pressures

ready adaptability to gradient elution and constant flow rates

--- disadvantages produce a pulsed flow cause baseline noise

1048729 Displacement pumps

--- advantages output is pulse free

--- disadvantages limited solvent capacity (lt250 mL)

inconvenience when change solvents

1048729 Pneumatic pumps

--- advantages inexpensive output is pulse free

--- disadvantages limited solvent capacity not amenable to

gradient elution and limit pressures lt2000 psi

Packings

Originally these were irregular silica and

alumina A range of synthetic regularly

shaped packings are now available

Porous - channels through packing

Superficially porous - rough surface

Smooth - bead like

Packing size

As packing size is decreased efficiency and

pressure requirements are increased

Common diameters for analytical work

diameter plates

10 μm 5000

5 μm 9000

3 μm 15000

All are for a 15 cm x 46 mm column

Column body

Typically consist of stainless steel with a high

precision internal bore

Some manufacturers offer column inserts

- donrsquot need to repurchase the column

fittings

Others offer columns where the external

body can be compressed to improve packing

efficiency

HPLC column examples

Column stationary phases

Today most packing fall into four classes

Silica or alumina

Bound phases on either alumina or silica

Gels

Controlled-pore glass or silica

Absorption phases - alumina

common mobile phases

hexane chloroform 2-propanol

example application - amines

silica

common mobile phases

hexane chloroform 2-propanol

example applications - ethers esters

porphyrins fat-soluble vitamins

Partition phases

Can be broken down into

Normal phase - polar materials bound to the support

Reverse phase - non-polar materials bound to the support

Mixed phase - may have some of each

Partition phases

Normal

Amino (-NH2)

Cyano (-CN)

Diol (glycidoxy-ethylmethoxysilane)

Reverse

C-2 or RP-2 (-Si-CH2CH3)

C-8 or RP-8 (-Si-(CH2)7CH3)

C-18 or RP-18 (-Si-(CH2)17CH3)

Increasing the C number results in a thicker more retentive phase

Ion exchange phases

Strong cation - sulfonic acid group

Strong anionic - quarternary amine

Weak anion - primary amine

Weak cation - COOH

Size exclusion phases

Gels - organic or aqueous based

Controlled-pore - silica or glass

Must be selected based on pressure

requirements and size range

required for your application

Capillary and Microbore columns

Several companies have begun offering

columns with smaller ID

Microbore column - 1 mm ID packed

column

Capillary column - lt 1 mm ID internal

bound phase

These columns require smaller solvent flows

reduced sample size and improved detector

design

Capillary and Microbore columns Capillary and Microbore columns

Aromatic Compounds

mobile phase2 ethylacetate in

hexane

flow rate 4 μlmin

column Fusica II 300μm ID x

25 cm silica

sample

1 toluene

2 nitrobenzene

3 acetophenone

4 26-dinitrobenzene

injection 60 nl

detection UV 254 nm

Silica based columns

Silica is the ideal support for HPLC columns It offers a large mechanical stability excellent

physicochemical surface properties a wide range of bonding chemistries and is compatible

with a broad range of organic solvents

pH stability

In general HPLC columns are stable within a pH range of 2 to 8 If you are measuring a pH value the measurement must be done in the aqueous media before mixing the eluent with organic solvents

Modern HPLC columns can be used outside that pH range The new bonding chemistries allow use down to pH 1 for some stationary phases However please check vendorrsquos product information before using silica based column outside the pH range of 2 to 8 However best lifetimes are obtained between pH 20 and pH 68

Mechanical stability

Stationary phases based on silica are mechanically very stable The packed columns show no pressure limit and can be used at more than 40 MPa (6000 psi) without any problem However please avoid pressure shocks on the column Pressure shocks lead to channelling in the column which results in peak splitting in the corresponding chromatogram

Mobile phases (Eluents)

Silica based stationary phases are compatible with all organic solvents in the above mentioned pH range

bull Filter all prepared buffer through a 05 μm filter before using them in your HPLC system

The use of non pure solvents in HPLC causes irreversible adsorption of impurities on the column head These impurities block adsorption sites change the selectivity of the column and lead to peak splitting in the chromatogram In gradient elution impurities cause so called ldquoGhost Peaksrdquo

Ghost peaks are peaks that always appear in the same position on the chromatogram Their origin is not the sample but the impurities from the solvents or solvent additives Therefore it is highly recommended to run a gradientwithout injection in the beginning of each method to determine the ghost peaks

To avoid irreversible adsorption at the column head you should always use a pre-column

The use of a pre-column increases the life time of a column dramatically In addition a pre-column can filter solid parts stemming from pump seals or injection rotors An alternative to a pre-column is an in-line filter These filters are attached directly to the column These filters get rid of solid parts in the eluent but will not avoid irreversible adsorption of organic impurities

Detector Systems

Virtually every chemical and physical

property that can be measured in

solution has been look at

Detectors fall roughly into two classes

Bulk property - measures an overall change in the mobile phase

Solute property - measures a solute specific property

Properties of a good detector

A detector must provide

bull high sensitivity low detection limits

bull linearity

bullReproducibility

This is true for any detector

Each detector will have specific advantages and will vary as to peak shape and spread noise and flowtemperature dependence they have

UVVis detector

A solute property detector

Sample must exhibit absorption in UVVis range Solvent must not absorb significantly at the measured wavelength

Types Filter photometer - single

bull Variable wavelength

bull Multiwavelength

Refractive index detector

Bulk property detector - general purpose

Based on refraction of light as it passes from

one media to another Presence of a solute

changes the refractive index of the solvent

Heat of absorption detector

A small amount of heat is released when a sample absorbs on a suitable surface

This detector can measure this

Electrochemical detectors

A number of properties have been evaluated

Detector types

Dielectic constant

Amperometric

Conductometric

Polarographic

Potentiometric

Dielectric constant detector

Bulk property detector

Measures changes in polarity of the liquid phase passing through the cell

Conductometric detector

Measures conductivity of the solvent Useful for

solutions of ions

Amperometric detectors

Most frequently applied type of electrochemical

detector

A known potential is applied across a set of

electrodes - typically a glassy carbon type

Ability to oxidize or reduce a species can be

measured

Typically limited to working with a specific class

of materials per analysis

Separation is based on the analytersquos relative solubility between two liquid phases

Stationary PhaseMobile Phase

Solvent Bonded Phase

Normal Phase- Polar stationary phase and non-polar solvent

bull Reverse Phase- Non-polar stationary phase and a polar solvent

Methan0l CH3OH

bull Acetonitrile CH3CN

bull Tetrahydrofuran

bull Water H2O

Solid Support - Backbone for bonded phases Usually 10micro 5micro or 3micro silica or polymeric

particles Bonded Phases - Functional groups firmly linked

(chemically bound) to the solid support Extremely stable Reproducible

Guard - Protects the analytical column Particles Interferences Prolongs the life of the analytical column

bull Analytical - Performs the separation

C-2 Ethyl Silyl -Si-CH2-CH3

bullCN Cyanopropyl Silyl -Si-(CH2)3-CN

bull C-18C-18 Octadecyl SilylOctadecyl Silyl -Si-(CH -Si-(CH22))1717--

CHCH33

bullC-8 Octyl Silyl -Si-(CH2)7-CH3

Pump

Injector

ColumnDetector

Mobile Phases

Gradient Controller

bull

UV Single wavelength (filter) [610 8330] Variable wavelength (monochromator)

[8316 8325] Multiple wavelengths (PDA) [555]

Fluorescence [610]Electrochemical [605] Mass Spectrometric [8325]

Restekreg ULTRA C-18 and CN Columns (250mm x 46mm 5micro)

Mobile Phase (11 MethanolWater) 15 mLmin

Supelcosil LC-PAH Columns A BConditions A 150mm x 46mm 5micro

Flow Rate 15 mLmin

Conditions B 50mm x 46mm 3micro

Flow Rate 30 mLmin

  • Slide 5
  • Slide 6
  • Slide 7
  • Slide 8
  • Slide 9
  • Slide 10
  • Slide 11
  • Slide 12
  • Slide 13
  • Slide 14
  • Slide 15
  • Slide 16
  • Slide 17
  • Slide 18
  • Slide 19
  • Slide 20
  • Slide 21
  • Slide 22
  • Slide 23
  • Slide 24
  • Slide 25
  • Slide 26
  • Slide 27
  • Slide 28
  • Slide 29
  • Slide 30
  • Slide 31
  • Slide 32
  • Slide 33
  • Slide 34
  • Slide 35
  • Slide 36
  • Slide 37
  • Slide 38
  • Slide 39
  • Slide 40
Page 3: P igh erformance L iquid C hromatography  HPLC is a form of liquid chromatography used to separate compounds that are dissolved in solution. HPLC instruments

This unit is equipped

with two pump units

and a UVVis

detector

The gradient is

controlled via the

pump controllers

LC

methods are not

as sensitive to

temperature

Columns are

commonly

mounted outside

the instrument

Solvents

bull All solvents should be lsquoHPLCrsquo grade

ndash This is a type of reagent grade material

ndash It has been filtered using a 02 μm filter

bull Filtered solvent helps extend pump life by preventing scoring It also reduces the chances of a column plugging

Solvent degassing

All solvents should be degassed prior to use

This reduces the chances of bubbles being formed in the column or detector Oxygen present at high pressure can also cause a problem

Methods that can be used

Displacement with a less soluble gas

Applying a vacuum

Heating the solvent

HPLC Injectors

In order to introduce a sample onto the column for analysis a special valve called the injector must be used to transfer the sample into the pressurized system

Injectors may look different from the outside but internally most are 6-port rotary valves These valves consist of a fixed body (the stator) plus an internal seal that rotates (the rotor)

Three internal passages connect alternate pairs of external ports The valves can switch between two positions referred to as the inject and load positions respectively In the load position the pump is connected to the column and the sample inlet is connected to one end of a piece of tubing called the sample loop

The other end of the sample loop is connected to the waste port Rotation results in reconnecting the various lines that enter the valve so that a sample volume can be inserted into the mobile phase that flows from the pump to the column inlet

The column

HPLC has seen significant improvement over

the last 20 years primarily due to improved

column technology

Packings are more uniform and smaller

Phases are commonly chemically bound to

the packing

Packing methods have improved

Pump types

bullIsocratic pump - delivers constant mobile phase composition

bullsolvent must be pre-mixed

bulllowest cost pump

bullGradient pump -delivers variable mobile phase composition

bullcan be used to mix and deliver an isocratic mobile phase or a gradient mobile phase

1048729 Reciprocation pumps

--- advantages small internal volume high output pressures

ready adaptability to gradient elution and constant flow rates

--- disadvantages produce a pulsed flow cause baseline noise

1048729 Displacement pumps

--- advantages output is pulse free

--- disadvantages limited solvent capacity (lt250 mL)

inconvenience when change solvents

1048729 Pneumatic pumps

--- advantages inexpensive output is pulse free

--- disadvantages limited solvent capacity not amenable to

gradient elution and limit pressures lt2000 psi

Packings

Originally these were irregular silica and

alumina A range of synthetic regularly

shaped packings are now available

Porous - channels through packing

Superficially porous - rough surface

Smooth - bead like

Packing size

As packing size is decreased efficiency and

pressure requirements are increased

Common diameters for analytical work

diameter plates

10 μm 5000

5 μm 9000

3 μm 15000

All are for a 15 cm x 46 mm column

Column body

Typically consist of stainless steel with a high

precision internal bore

Some manufacturers offer column inserts

- donrsquot need to repurchase the column

fittings

Others offer columns where the external

body can be compressed to improve packing

efficiency

HPLC column examples

Column stationary phases

Today most packing fall into four classes

Silica or alumina

Bound phases on either alumina or silica

Gels

Controlled-pore glass or silica

Absorption phases - alumina

common mobile phases

hexane chloroform 2-propanol

example application - amines

silica

common mobile phases

hexane chloroform 2-propanol

example applications - ethers esters

porphyrins fat-soluble vitamins

Partition phases

Can be broken down into

Normal phase - polar materials bound to the support

Reverse phase - non-polar materials bound to the support

Mixed phase - may have some of each

Partition phases

Normal

Amino (-NH2)

Cyano (-CN)

Diol (glycidoxy-ethylmethoxysilane)

Reverse

C-2 or RP-2 (-Si-CH2CH3)

C-8 or RP-8 (-Si-(CH2)7CH3)

C-18 or RP-18 (-Si-(CH2)17CH3)

Increasing the C number results in a thicker more retentive phase

Ion exchange phases

Strong cation - sulfonic acid group

Strong anionic - quarternary amine

Weak anion - primary amine

Weak cation - COOH

Size exclusion phases

Gels - organic or aqueous based

Controlled-pore - silica or glass

Must be selected based on pressure

requirements and size range

required for your application

Capillary and Microbore columns

Several companies have begun offering

columns with smaller ID

Microbore column - 1 mm ID packed

column

Capillary column - lt 1 mm ID internal

bound phase

These columns require smaller solvent flows

reduced sample size and improved detector

design

Capillary and Microbore columns Capillary and Microbore columns

Aromatic Compounds

mobile phase2 ethylacetate in

hexane

flow rate 4 μlmin

column Fusica II 300μm ID x

25 cm silica

sample

1 toluene

2 nitrobenzene

3 acetophenone

4 26-dinitrobenzene

injection 60 nl

detection UV 254 nm

Silica based columns

Silica is the ideal support for HPLC columns It offers a large mechanical stability excellent

physicochemical surface properties a wide range of bonding chemistries and is compatible

with a broad range of organic solvents

pH stability

In general HPLC columns are stable within a pH range of 2 to 8 If you are measuring a pH value the measurement must be done in the aqueous media before mixing the eluent with organic solvents

Modern HPLC columns can be used outside that pH range The new bonding chemistries allow use down to pH 1 for some stationary phases However please check vendorrsquos product information before using silica based column outside the pH range of 2 to 8 However best lifetimes are obtained between pH 20 and pH 68

Mechanical stability

Stationary phases based on silica are mechanically very stable The packed columns show no pressure limit and can be used at more than 40 MPa (6000 psi) without any problem However please avoid pressure shocks on the column Pressure shocks lead to channelling in the column which results in peak splitting in the corresponding chromatogram

Mobile phases (Eluents)

Silica based stationary phases are compatible with all organic solvents in the above mentioned pH range

bull Filter all prepared buffer through a 05 μm filter before using them in your HPLC system

The use of non pure solvents in HPLC causes irreversible adsorption of impurities on the column head These impurities block adsorption sites change the selectivity of the column and lead to peak splitting in the chromatogram In gradient elution impurities cause so called ldquoGhost Peaksrdquo

Ghost peaks are peaks that always appear in the same position on the chromatogram Their origin is not the sample but the impurities from the solvents or solvent additives Therefore it is highly recommended to run a gradientwithout injection in the beginning of each method to determine the ghost peaks

To avoid irreversible adsorption at the column head you should always use a pre-column

The use of a pre-column increases the life time of a column dramatically In addition a pre-column can filter solid parts stemming from pump seals or injection rotors An alternative to a pre-column is an in-line filter These filters are attached directly to the column These filters get rid of solid parts in the eluent but will not avoid irreversible adsorption of organic impurities

Detector Systems

Virtually every chemical and physical

property that can be measured in

solution has been look at

Detectors fall roughly into two classes

Bulk property - measures an overall change in the mobile phase

Solute property - measures a solute specific property

Properties of a good detector

A detector must provide

bull high sensitivity low detection limits

bull linearity

bullReproducibility

This is true for any detector

Each detector will have specific advantages and will vary as to peak shape and spread noise and flowtemperature dependence they have

UVVis detector

A solute property detector

Sample must exhibit absorption in UVVis range Solvent must not absorb significantly at the measured wavelength

Types Filter photometer - single

bull Variable wavelength

bull Multiwavelength

Refractive index detector

Bulk property detector - general purpose

Based on refraction of light as it passes from

one media to another Presence of a solute

changes the refractive index of the solvent

Heat of absorption detector

A small amount of heat is released when a sample absorbs on a suitable surface

This detector can measure this

Electrochemical detectors

A number of properties have been evaluated

Detector types

Dielectic constant

Amperometric

Conductometric

Polarographic

Potentiometric

Dielectric constant detector

Bulk property detector

Measures changes in polarity of the liquid phase passing through the cell

Conductometric detector

Measures conductivity of the solvent Useful for

solutions of ions

Amperometric detectors

Most frequently applied type of electrochemical

detector

A known potential is applied across a set of

electrodes - typically a glassy carbon type

Ability to oxidize or reduce a species can be

measured

Typically limited to working with a specific class

of materials per analysis

Separation is based on the analytersquos relative solubility between two liquid phases

Stationary PhaseMobile Phase

Solvent Bonded Phase

Normal Phase- Polar stationary phase and non-polar solvent

bull Reverse Phase- Non-polar stationary phase and a polar solvent

Methan0l CH3OH

bull Acetonitrile CH3CN

bull Tetrahydrofuran

bull Water H2O

Solid Support - Backbone for bonded phases Usually 10micro 5micro or 3micro silica or polymeric

particles Bonded Phases - Functional groups firmly linked

(chemically bound) to the solid support Extremely stable Reproducible

Guard - Protects the analytical column Particles Interferences Prolongs the life of the analytical column

bull Analytical - Performs the separation

C-2 Ethyl Silyl -Si-CH2-CH3

bullCN Cyanopropyl Silyl -Si-(CH2)3-CN

bull C-18C-18 Octadecyl SilylOctadecyl Silyl -Si-(CH -Si-(CH22))1717--

CHCH33

bullC-8 Octyl Silyl -Si-(CH2)7-CH3

Pump

Injector

ColumnDetector

Mobile Phases

Gradient Controller

bull

UV Single wavelength (filter) [610 8330] Variable wavelength (monochromator)

[8316 8325] Multiple wavelengths (PDA) [555]

Fluorescence [610]Electrochemical [605] Mass Spectrometric [8325]

Restekreg ULTRA C-18 and CN Columns (250mm x 46mm 5micro)

Mobile Phase (11 MethanolWater) 15 mLmin

Supelcosil LC-PAH Columns A BConditions A 150mm x 46mm 5micro

Flow Rate 15 mLmin

Conditions B 50mm x 46mm 3micro

Flow Rate 30 mLmin

  • Slide 5
  • Slide 6
  • Slide 7
  • Slide 8
  • Slide 9
  • Slide 10
  • Slide 11
  • Slide 12
  • Slide 13
  • Slide 14
  • Slide 15
  • Slide 16
  • Slide 17
  • Slide 18
  • Slide 19
  • Slide 20
  • Slide 21
  • Slide 22
  • Slide 23
  • Slide 24
  • Slide 25
  • Slide 26
  • Slide 27
  • Slide 28
  • Slide 29
  • Slide 30
  • Slide 31
  • Slide 32
  • Slide 33
  • Slide 34
  • Slide 35
  • Slide 36
  • Slide 37
  • Slide 38
  • Slide 39
  • Slide 40
Page 4: P igh erformance L iquid C hromatography  HPLC is a form of liquid chromatography used to separate compounds that are dissolved in solution. HPLC instruments

LC

methods are not

as sensitive to

temperature

Columns are

commonly

mounted outside

the instrument

Solvents

bull All solvents should be lsquoHPLCrsquo grade

ndash This is a type of reagent grade material

ndash It has been filtered using a 02 μm filter

bull Filtered solvent helps extend pump life by preventing scoring It also reduces the chances of a column plugging

Solvent degassing

All solvents should be degassed prior to use

This reduces the chances of bubbles being formed in the column or detector Oxygen present at high pressure can also cause a problem

Methods that can be used

Displacement with a less soluble gas

Applying a vacuum

Heating the solvent

HPLC Injectors

In order to introduce a sample onto the column for analysis a special valve called the injector must be used to transfer the sample into the pressurized system

Injectors may look different from the outside but internally most are 6-port rotary valves These valves consist of a fixed body (the stator) plus an internal seal that rotates (the rotor)

Three internal passages connect alternate pairs of external ports The valves can switch between two positions referred to as the inject and load positions respectively In the load position the pump is connected to the column and the sample inlet is connected to one end of a piece of tubing called the sample loop

The other end of the sample loop is connected to the waste port Rotation results in reconnecting the various lines that enter the valve so that a sample volume can be inserted into the mobile phase that flows from the pump to the column inlet

The column

HPLC has seen significant improvement over

the last 20 years primarily due to improved

column technology

Packings are more uniform and smaller

Phases are commonly chemically bound to

the packing

Packing methods have improved

Pump types

bullIsocratic pump - delivers constant mobile phase composition

bullsolvent must be pre-mixed

bulllowest cost pump

bullGradient pump -delivers variable mobile phase composition

bullcan be used to mix and deliver an isocratic mobile phase or a gradient mobile phase

1048729 Reciprocation pumps

--- advantages small internal volume high output pressures

ready adaptability to gradient elution and constant flow rates

--- disadvantages produce a pulsed flow cause baseline noise

1048729 Displacement pumps

--- advantages output is pulse free

--- disadvantages limited solvent capacity (lt250 mL)

inconvenience when change solvents

1048729 Pneumatic pumps

--- advantages inexpensive output is pulse free

--- disadvantages limited solvent capacity not amenable to

gradient elution and limit pressures lt2000 psi

Packings

Originally these were irregular silica and

alumina A range of synthetic regularly

shaped packings are now available

Porous - channels through packing

Superficially porous - rough surface

Smooth - bead like

Packing size

As packing size is decreased efficiency and

pressure requirements are increased

Common diameters for analytical work

diameter plates

10 μm 5000

5 μm 9000

3 μm 15000

All are for a 15 cm x 46 mm column

Column body

Typically consist of stainless steel with a high

precision internal bore

Some manufacturers offer column inserts

- donrsquot need to repurchase the column

fittings

Others offer columns where the external

body can be compressed to improve packing

efficiency

HPLC column examples

Column stationary phases

Today most packing fall into four classes

Silica or alumina

Bound phases on either alumina or silica

Gels

Controlled-pore glass or silica

Absorption phases - alumina

common mobile phases

hexane chloroform 2-propanol

example application - amines

silica

common mobile phases

hexane chloroform 2-propanol

example applications - ethers esters

porphyrins fat-soluble vitamins

Partition phases

Can be broken down into

Normal phase - polar materials bound to the support

Reverse phase - non-polar materials bound to the support

Mixed phase - may have some of each

Partition phases

Normal

Amino (-NH2)

Cyano (-CN)

Diol (glycidoxy-ethylmethoxysilane)

Reverse

C-2 or RP-2 (-Si-CH2CH3)

C-8 or RP-8 (-Si-(CH2)7CH3)

C-18 or RP-18 (-Si-(CH2)17CH3)

Increasing the C number results in a thicker more retentive phase

Ion exchange phases

Strong cation - sulfonic acid group

Strong anionic - quarternary amine

Weak anion - primary amine

Weak cation - COOH

Size exclusion phases

Gels - organic or aqueous based

Controlled-pore - silica or glass

Must be selected based on pressure

requirements and size range

required for your application

Capillary and Microbore columns

Several companies have begun offering

columns with smaller ID

Microbore column - 1 mm ID packed

column

Capillary column - lt 1 mm ID internal

bound phase

These columns require smaller solvent flows

reduced sample size and improved detector

design

Capillary and Microbore columns Capillary and Microbore columns

Aromatic Compounds

mobile phase2 ethylacetate in

hexane

flow rate 4 μlmin

column Fusica II 300μm ID x

25 cm silica

sample

1 toluene

2 nitrobenzene

3 acetophenone

4 26-dinitrobenzene

injection 60 nl

detection UV 254 nm

Silica based columns

Silica is the ideal support for HPLC columns It offers a large mechanical stability excellent

physicochemical surface properties a wide range of bonding chemistries and is compatible

with a broad range of organic solvents

pH stability

In general HPLC columns are stable within a pH range of 2 to 8 If you are measuring a pH value the measurement must be done in the aqueous media before mixing the eluent with organic solvents

Modern HPLC columns can be used outside that pH range The new bonding chemistries allow use down to pH 1 for some stationary phases However please check vendorrsquos product information before using silica based column outside the pH range of 2 to 8 However best lifetimes are obtained between pH 20 and pH 68

Mechanical stability

Stationary phases based on silica are mechanically very stable The packed columns show no pressure limit and can be used at more than 40 MPa (6000 psi) without any problem However please avoid pressure shocks on the column Pressure shocks lead to channelling in the column which results in peak splitting in the corresponding chromatogram

Mobile phases (Eluents)

Silica based stationary phases are compatible with all organic solvents in the above mentioned pH range

bull Filter all prepared buffer through a 05 μm filter before using them in your HPLC system

The use of non pure solvents in HPLC causes irreversible adsorption of impurities on the column head These impurities block adsorption sites change the selectivity of the column and lead to peak splitting in the chromatogram In gradient elution impurities cause so called ldquoGhost Peaksrdquo

Ghost peaks are peaks that always appear in the same position on the chromatogram Their origin is not the sample but the impurities from the solvents or solvent additives Therefore it is highly recommended to run a gradientwithout injection in the beginning of each method to determine the ghost peaks

To avoid irreversible adsorption at the column head you should always use a pre-column

The use of a pre-column increases the life time of a column dramatically In addition a pre-column can filter solid parts stemming from pump seals or injection rotors An alternative to a pre-column is an in-line filter These filters are attached directly to the column These filters get rid of solid parts in the eluent but will not avoid irreversible adsorption of organic impurities

Detector Systems

Virtually every chemical and physical

property that can be measured in

solution has been look at

Detectors fall roughly into two classes

Bulk property - measures an overall change in the mobile phase

Solute property - measures a solute specific property

Properties of a good detector

A detector must provide

bull high sensitivity low detection limits

bull linearity

bullReproducibility

This is true for any detector

Each detector will have specific advantages and will vary as to peak shape and spread noise and flowtemperature dependence they have

UVVis detector

A solute property detector

Sample must exhibit absorption in UVVis range Solvent must not absorb significantly at the measured wavelength

Types Filter photometer - single

bull Variable wavelength

bull Multiwavelength

Refractive index detector

Bulk property detector - general purpose

Based on refraction of light as it passes from

one media to another Presence of a solute

changes the refractive index of the solvent

Heat of absorption detector

A small amount of heat is released when a sample absorbs on a suitable surface

This detector can measure this

Electrochemical detectors

A number of properties have been evaluated

Detector types

Dielectic constant

Amperometric

Conductometric

Polarographic

Potentiometric

Dielectric constant detector

Bulk property detector

Measures changes in polarity of the liquid phase passing through the cell

Conductometric detector

Measures conductivity of the solvent Useful for

solutions of ions

Amperometric detectors

Most frequently applied type of electrochemical

detector

A known potential is applied across a set of

electrodes - typically a glassy carbon type

Ability to oxidize or reduce a species can be

measured

Typically limited to working with a specific class

of materials per analysis

Separation is based on the analytersquos relative solubility between two liquid phases

Stationary PhaseMobile Phase

Solvent Bonded Phase

Normal Phase- Polar stationary phase and non-polar solvent

bull Reverse Phase- Non-polar stationary phase and a polar solvent

Methan0l CH3OH

bull Acetonitrile CH3CN

bull Tetrahydrofuran

bull Water H2O

Solid Support - Backbone for bonded phases Usually 10micro 5micro or 3micro silica or polymeric

particles Bonded Phases - Functional groups firmly linked

(chemically bound) to the solid support Extremely stable Reproducible

Guard - Protects the analytical column Particles Interferences Prolongs the life of the analytical column

bull Analytical - Performs the separation

C-2 Ethyl Silyl -Si-CH2-CH3

bullCN Cyanopropyl Silyl -Si-(CH2)3-CN

bull C-18C-18 Octadecyl SilylOctadecyl Silyl -Si-(CH -Si-(CH22))1717--

CHCH33

bullC-8 Octyl Silyl -Si-(CH2)7-CH3

Pump

Injector

ColumnDetector

Mobile Phases

Gradient Controller

bull

UV Single wavelength (filter) [610 8330] Variable wavelength (monochromator)

[8316 8325] Multiple wavelengths (PDA) [555]

Fluorescence [610]Electrochemical [605] Mass Spectrometric [8325]

Restekreg ULTRA C-18 and CN Columns (250mm x 46mm 5micro)

Mobile Phase (11 MethanolWater) 15 mLmin

Supelcosil LC-PAH Columns A BConditions A 150mm x 46mm 5micro

Flow Rate 15 mLmin

Conditions B 50mm x 46mm 3micro

Flow Rate 30 mLmin

  • Slide 5
  • Slide 6
  • Slide 7
  • Slide 8
  • Slide 9
  • Slide 10
  • Slide 11
  • Slide 12
  • Slide 13
  • Slide 14
  • Slide 15
  • Slide 16
  • Slide 17
  • Slide 18
  • Slide 19
  • Slide 20
  • Slide 21
  • Slide 22
  • Slide 23
  • Slide 24
  • Slide 25
  • Slide 26
  • Slide 27
  • Slide 28
  • Slide 29
  • Slide 30
  • Slide 31
  • Slide 32
  • Slide 33
  • Slide 34
  • Slide 35
  • Slide 36
  • Slide 37
  • Slide 38
  • Slide 39
  • Slide 40
Page 5: P igh erformance L iquid C hromatography  HPLC is a form of liquid chromatography used to separate compounds that are dissolved in solution. HPLC instruments

Solvents

bull All solvents should be lsquoHPLCrsquo grade

ndash This is a type of reagent grade material

ndash It has been filtered using a 02 μm filter

bull Filtered solvent helps extend pump life by preventing scoring It also reduces the chances of a column plugging

Solvent degassing

All solvents should be degassed prior to use

This reduces the chances of bubbles being formed in the column or detector Oxygen present at high pressure can also cause a problem

Methods that can be used

Displacement with a less soluble gas

Applying a vacuum

Heating the solvent

HPLC Injectors

In order to introduce a sample onto the column for analysis a special valve called the injector must be used to transfer the sample into the pressurized system

Injectors may look different from the outside but internally most are 6-port rotary valves These valves consist of a fixed body (the stator) plus an internal seal that rotates (the rotor)

Three internal passages connect alternate pairs of external ports The valves can switch between two positions referred to as the inject and load positions respectively In the load position the pump is connected to the column and the sample inlet is connected to one end of a piece of tubing called the sample loop

The other end of the sample loop is connected to the waste port Rotation results in reconnecting the various lines that enter the valve so that a sample volume can be inserted into the mobile phase that flows from the pump to the column inlet

The column

HPLC has seen significant improvement over

the last 20 years primarily due to improved

column technology

Packings are more uniform and smaller

Phases are commonly chemically bound to

the packing

Packing methods have improved

Pump types

bullIsocratic pump - delivers constant mobile phase composition

bullsolvent must be pre-mixed

bulllowest cost pump

bullGradient pump -delivers variable mobile phase composition

bullcan be used to mix and deliver an isocratic mobile phase or a gradient mobile phase

1048729 Reciprocation pumps

--- advantages small internal volume high output pressures

ready adaptability to gradient elution and constant flow rates

--- disadvantages produce a pulsed flow cause baseline noise

1048729 Displacement pumps

--- advantages output is pulse free

--- disadvantages limited solvent capacity (lt250 mL)

inconvenience when change solvents

1048729 Pneumatic pumps

--- advantages inexpensive output is pulse free

--- disadvantages limited solvent capacity not amenable to

gradient elution and limit pressures lt2000 psi

Packings

Originally these were irregular silica and

alumina A range of synthetic regularly

shaped packings are now available

Porous - channels through packing

Superficially porous - rough surface

Smooth - bead like

Packing size

As packing size is decreased efficiency and

pressure requirements are increased

Common diameters for analytical work

diameter plates

10 μm 5000

5 μm 9000

3 μm 15000

All are for a 15 cm x 46 mm column

Column body

Typically consist of stainless steel with a high

precision internal bore

Some manufacturers offer column inserts

- donrsquot need to repurchase the column

fittings

Others offer columns where the external

body can be compressed to improve packing

efficiency

HPLC column examples

Column stationary phases

Today most packing fall into four classes

Silica or alumina

Bound phases on either alumina or silica

Gels

Controlled-pore glass or silica

Absorption phases - alumina

common mobile phases

hexane chloroform 2-propanol

example application - amines

silica

common mobile phases

hexane chloroform 2-propanol

example applications - ethers esters

porphyrins fat-soluble vitamins

Partition phases

Can be broken down into

Normal phase - polar materials bound to the support

Reverse phase - non-polar materials bound to the support

Mixed phase - may have some of each

Partition phases

Normal

Amino (-NH2)

Cyano (-CN)

Diol (glycidoxy-ethylmethoxysilane)

Reverse

C-2 or RP-2 (-Si-CH2CH3)

C-8 or RP-8 (-Si-(CH2)7CH3)

C-18 or RP-18 (-Si-(CH2)17CH3)

Increasing the C number results in a thicker more retentive phase

Ion exchange phases

Strong cation - sulfonic acid group

Strong anionic - quarternary amine

Weak anion - primary amine

Weak cation - COOH

Size exclusion phases

Gels - organic or aqueous based

Controlled-pore - silica or glass

Must be selected based on pressure

requirements and size range

required for your application

Capillary and Microbore columns

Several companies have begun offering

columns with smaller ID

Microbore column - 1 mm ID packed

column

Capillary column - lt 1 mm ID internal

bound phase

These columns require smaller solvent flows

reduced sample size and improved detector

design

Capillary and Microbore columns Capillary and Microbore columns

Aromatic Compounds

mobile phase2 ethylacetate in

hexane

flow rate 4 μlmin

column Fusica II 300μm ID x

25 cm silica

sample

1 toluene

2 nitrobenzene

3 acetophenone

4 26-dinitrobenzene

injection 60 nl

detection UV 254 nm

Silica based columns

Silica is the ideal support for HPLC columns It offers a large mechanical stability excellent

physicochemical surface properties a wide range of bonding chemistries and is compatible

with a broad range of organic solvents

pH stability

In general HPLC columns are stable within a pH range of 2 to 8 If you are measuring a pH value the measurement must be done in the aqueous media before mixing the eluent with organic solvents

Modern HPLC columns can be used outside that pH range The new bonding chemistries allow use down to pH 1 for some stationary phases However please check vendorrsquos product information before using silica based column outside the pH range of 2 to 8 However best lifetimes are obtained between pH 20 and pH 68

Mechanical stability

Stationary phases based on silica are mechanically very stable The packed columns show no pressure limit and can be used at more than 40 MPa (6000 psi) without any problem However please avoid pressure shocks on the column Pressure shocks lead to channelling in the column which results in peak splitting in the corresponding chromatogram

Mobile phases (Eluents)

Silica based stationary phases are compatible with all organic solvents in the above mentioned pH range

bull Filter all prepared buffer through a 05 μm filter before using them in your HPLC system

The use of non pure solvents in HPLC causes irreversible adsorption of impurities on the column head These impurities block adsorption sites change the selectivity of the column and lead to peak splitting in the chromatogram In gradient elution impurities cause so called ldquoGhost Peaksrdquo

Ghost peaks are peaks that always appear in the same position on the chromatogram Their origin is not the sample but the impurities from the solvents or solvent additives Therefore it is highly recommended to run a gradientwithout injection in the beginning of each method to determine the ghost peaks

To avoid irreversible adsorption at the column head you should always use a pre-column

The use of a pre-column increases the life time of a column dramatically In addition a pre-column can filter solid parts stemming from pump seals or injection rotors An alternative to a pre-column is an in-line filter These filters are attached directly to the column These filters get rid of solid parts in the eluent but will not avoid irreversible adsorption of organic impurities

Detector Systems

Virtually every chemical and physical

property that can be measured in

solution has been look at

Detectors fall roughly into two classes

Bulk property - measures an overall change in the mobile phase

Solute property - measures a solute specific property

Properties of a good detector

A detector must provide

bull high sensitivity low detection limits

bull linearity

bullReproducibility

This is true for any detector

Each detector will have specific advantages and will vary as to peak shape and spread noise and flowtemperature dependence they have

UVVis detector

A solute property detector

Sample must exhibit absorption in UVVis range Solvent must not absorb significantly at the measured wavelength

Types Filter photometer - single

bull Variable wavelength

bull Multiwavelength

Refractive index detector

Bulk property detector - general purpose

Based on refraction of light as it passes from

one media to another Presence of a solute

changes the refractive index of the solvent

Heat of absorption detector

A small amount of heat is released when a sample absorbs on a suitable surface

This detector can measure this

Electrochemical detectors

A number of properties have been evaluated

Detector types

Dielectic constant

Amperometric

Conductometric

Polarographic

Potentiometric

Dielectric constant detector

Bulk property detector

Measures changes in polarity of the liquid phase passing through the cell

Conductometric detector

Measures conductivity of the solvent Useful for

solutions of ions

Amperometric detectors

Most frequently applied type of electrochemical

detector

A known potential is applied across a set of

electrodes - typically a glassy carbon type

Ability to oxidize or reduce a species can be

measured

Typically limited to working with a specific class

of materials per analysis

Separation is based on the analytersquos relative solubility between two liquid phases

Stationary PhaseMobile Phase

Solvent Bonded Phase

Normal Phase- Polar stationary phase and non-polar solvent

bull Reverse Phase- Non-polar stationary phase and a polar solvent

Methan0l CH3OH

bull Acetonitrile CH3CN

bull Tetrahydrofuran

bull Water H2O

Solid Support - Backbone for bonded phases Usually 10micro 5micro or 3micro silica or polymeric

particles Bonded Phases - Functional groups firmly linked

(chemically bound) to the solid support Extremely stable Reproducible

Guard - Protects the analytical column Particles Interferences Prolongs the life of the analytical column

bull Analytical - Performs the separation

C-2 Ethyl Silyl -Si-CH2-CH3

bullCN Cyanopropyl Silyl -Si-(CH2)3-CN

bull C-18C-18 Octadecyl SilylOctadecyl Silyl -Si-(CH -Si-(CH22))1717--

CHCH33

bullC-8 Octyl Silyl -Si-(CH2)7-CH3

Pump

Injector

ColumnDetector

Mobile Phases

Gradient Controller

bull

UV Single wavelength (filter) [610 8330] Variable wavelength (monochromator)

[8316 8325] Multiple wavelengths (PDA) [555]

Fluorescence [610]Electrochemical [605] Mass Spectrometric [8325]

Restekreg ULTRA C-18 and CN Columns (250mm x 46mm 5micro)

Mobile Phase (11 MethanolWater) 15 mLmin

Supelcosil LC-PAH Columns A BConditions A 150mm x 46mm 5micro

Flow Rate 15 mLmin

Conditions B 50mm x 46mm 3micro

Flow Rate 30 mLmin

  • Slide 5
  • Slide 6
  • Slide 7
  • Slide 8
  • Slide 9
  • Slide 10
  • Slide 11
  • Slide 12
  • Slide 13
  • Slide 14
  • Slide 15
  • Slide 16
  • Slide 17
  • Slide 18
  • Slide 19
  • Slide 20
  • Slide 21
  • Slide 22
  • Slide 23
  • Slide 24
  • Slide 25
  • Slide 26
  • Slide 27
  • Slide 28
  • Slide 29
  • Slide 30
  • Slide 31
  • Slide 32
  • Slide 33
  • Slide 34
  • Slide 35
  • Slide 36
  • Slide 37
  • Slide 38
  • Slide 39
  • Slide 40
Page 6: P igh erformance L iquid C hromatography  HPLC is a form of liquid chromatography used to separate compounds that are dissolved in solution. HPLC instruments

Solvent degassing

All solvents should be degassed prior to use

This reduces the chances of bubbles being formed in the column or detector Oxygen present at high pressure can also cause a problem

Methods that can be used

Displacement with a less soluble gas

Applying a vacuum

Heating the solvent

HPLC Injectors

In order to introduce a sample onto the column for analysis a special valve called the injector must be used to transfer the sample into the pressurized system

Injectors may look different from the outside but internally most are 6-port rotary valves These valves consist of a fixed body (the stator) plus an internal seal that rotates (the rotor)

Three internal passages connect alternate pairs of external ports The valves can switch between two positions referred to as the inject and load positions respectively In the load position the pump is connected to the column and the sample inlet is connected to one end of a piece of tubing called the sample loop

The other end of the sample loop is connected to the waste port Rotation results in reconnecting the various lines that enter the valve so that a sample volume can be inserted into the mobile phase that flows from the pump to the column inlet

The column

HPLC has seen significant improvement over

the last 20 years primarily due to improved

column technology

Packings are more uniform and smaller

Phases are commonly chemically bound to

the packing

Packing methods have improved

Pump types

bullIsocratic pump - delivers constant mobile phase composition

bullsolvent must be pre-mixed

bulllowest cost pump

bullGradient pump -delivers variable mobile phase composition

bullcan be used to mix and deliver an isocratic mobile phase or a gradient mobile phase

1048729 Reciprocation pumps

--- advantages small internal volume high output pressures

ready adaptability to gradient elution and constant flow rates

--- disadvantages produce a pulsed flow cause baseline noise

1048729 Displacement pumps

--- advantages output is pulse free

--- disadvantages limited solvent capacity (lt250 mL)

inconvenience when change solvents

1048729 Pneumatic pumps

--- advantages inexpensive output is pulse free

--- disadvantages limited solvent capacity not amenable to

gradient elution and limit pressures lt2000 psi

Packings

Originally these were irregular silica and

alumina A range of synthetic regularly

shaped packings are now available

Porous - channels through packing

Superficially porous - rough surface

Smooth - bead like

Packing size

As packing size is decreased efficiency and

pressure requirements are increased

Common diameters for analytical work

diameter plates

10 μm 5000

5 μm 9000

3 μm 15000

All are for a 15 cm x 46 mm column

Column body

Typically consist of stainless steel with a high

precision internal bore

Some manufacturers offer column inserts

- donrsquot need to repurchase the column

fittings

Others offer columns where the external

body can be compressed to improve packing

efficiency

HPLC column examples

Column stationary phases

Today most packing fall into four classes

Silica or alumina

Bound phases on either alumina or silica

Gels

Controlled-pore glass or silica

Absorption phases - alumina

common mobile phases

hexane chloroform 2-propanol

example application - amines

silica

common mobile phases

hexane chloroform 2-propanol

example applications - ethers esters

porphyrins fat-soluble vitamins

Partition phases

Can be broken down into

Normal phase - polar materials bound to the support

Reverse phase - non-polar materials bound to the support

Mixed phase - may have some of each

Partition phases

Normal

Amino (-NH2)

Cyano (-CN)

Diol (glycidoxy-ethylmethoxysilane)

Reverse

C-2 or RP-2 (-Si-CH2CH3)

C-8 or RP-8 (-Si-(CH2)7CH3)

C-18 or RP-18 (-Si-(CH2)17CH3)

Increasing the C number results in a thicker more retentive phase

Ion exchange phases

Strong cation - sulfonic acid group

Strong anionic - quarternary amine

Weak anion - primary amine

Weak cation - COOH

Size exclusion phases

Gels - organic or aqueous based

Controlled-pore - silica or glass

Must be selected based on pressure

requirements and size range

required for your application

Capillary and Microbore columns

Several companies have begun offering

columns with smaller ID

Microbore column - 1 mm ID packed

column

Capillary column - lt 1 mm ID internal

bound phase

These columns require smaller solvent flows

reduced sample size and improved detector

design

Capillary and Microbore columns Capillary and Microbore columns

Aromatic Compounds

mobile phase2 ethylacetate in

hexane

flow rate 4 μlmin

column Fusica II 300μm ID x

25 cm silica

sample

1 toluene

2 nitrobenzene

3 acetophenone

4 26-dinitrobenzene

injection 60 nl

detection UV 254 nm

Silica based columns

Silica is the ideal support for HPLC columns It offers a large mechanical stability excellent

physicochemical surface properties a wide range of bonding chemistries and is compatible

with a broad range of organic solvents

pH stability

In general HPLC columns are stable within a pH range of 2 to 8 If you are measuring a pH value the measurement must be done in the aqueous media before mixing the eluent with organic solvents

Modern HPLC columns can be used outside that pH range The new bonding chemistries allow use down to pH 1 for some stationary phases However please check vendorrsquos product information before using silica based column outside the pH range of 2 to 8 However best lifetimes are obtained between pH 20 and pH 68

Mechanical stability

Stationary phases based on silica are mechanically very stable The packed columns show no pressure limit and can be used at more than 40 MPa (6000 psi) without any problem However please avoid pressure shocks on the column Pressure shocks lead to channelling in the column which results in peak splitting in the corresponding chromatogram

Mobile phases (Eluents)

Silica based stationary phases are compatible with all organic solvents in the above mentioned pH range

bull Filter all prepared buffer through a 05 μm filter before using them in your HPLC system

The use of non pure solvents in HPLC causes irreversible adsorption of impurities on the column head These impurities block adsorption sites change the selectivity of the column and lead to peak splitting in the chromatogram In gradient elution impurities cause so called ldquoGhost Peaksrdquo

Ghost peaks are peaks that always appear in the same position on the chromatogram Their origin is not the sample but the impurities from the solvents or solvent additives Therefore it is highly recommended to run a gradientwithout injection in the beginning of each method to determine the ghost peaks

To avoid irreversible adsorption at the column head you should always use a pre-column

The use of a pre-column increases the life time of a column dramatically In addition a pre-column can filter solid parts stemming from pump seals or injection rotors An alternative to a pre-column is an in-line filter These filters are attached directly to the column These filters get rid of solid parts in the eluent but will not avoid irreversible adsorption of organic impurities

Detector Systems

Virtually every chemical and physical

property that can be measured in

solution has been look at

Detectors fall roughly into two classes

Bulk property - measures an overall change in the mobile phase

Solute property - measures a solute specific property

Properties of a good detector

A detector must provide

bull high sensitivity low detection limits

bull linearity

bullReproducibility

This is true for any detector

Each detector will have specific advantages and will vary as to peak shape and spread noise and flowtemperature dependence they have

UVVis detector

A solute property detector

Sample must exhibit absorption in UVVis range Solvent must not absorb significantly at the measured wavelength

Types Filter photometer - single

bull Variable wavelength

bull Multiwavelength

Refractive index detector

Bulk property detector - general purpose

Based on refraction of light as it passes from

one media to another Presence of a solute

changes the refractive index of the solvent

Heat of absorption detector

A small amount of heat is released when a sample absorbs on a suitable surface

This detector can measure this

Electrochemical detectors

A number of properties have been evaluated

Detector types

Dielectic constant

Amperometric

Conductometric

Polarographic

Potentiometric

Dielectric constant detector

Bulk property detector

Measures changes in polarity of the liquid phase passing through the cell

Conductometric detector

Measures conductivity of the solvent Useful for

solutions of ions

Amperometric detectors

Most frequently applied type of electrochemical

detector

A known potential is applied across a set of

electrodes - typically a glassy carbon type

Ability to oxidize or reduce a species can be

measured

Typically limited to working with a specific class

of materials per analysis

Separation is based on the analytersquos relative solubility between two liquid phases

Stationary PhaseMobile Phase

Solvent Bonded Phase

Normal Phase- Polar stationary phase and non-polar solvent

bull Reverse Phase- Non-polar stationary phase and a polar solvent

Methan0l CH3OH

bull Acetonitrile CH3CN

bull Tetrahydrofuran

bull Water H2O

Solid Support - Backbone for bonded phases Usually 10micro 5micro or 3micro silica or polymeric

particles Bonded Phases - Functional groups firmly linked

(chemically bound) to the solid support Extremely stable Reproducible

Guard - Protects the analytical column Particles Interferences Prolongs the life of the analytical column

bull Analytical - Performs the separation

C-2 Ethyl Silyl -Si-CH2-CH3

bullCN Cyanopropyl Silyl -Si-(CH2)3-CN

bull C-18C-18 Octadecyl SilylOctadecyl Silyl -Si-(CH -Si-(CH22))1717--

CHCH33

bullC-8 Octyl Silyl -Si-(CH2)7-CH3

Pump

Injector

ColumnDetector

Mobile Phases

Gradient Controller

bull

UV Single wavelength (filter) [610 8330] Variable wavelength (monochromator)

[8316 8325] Multiple wavelengths (PDA) [555]

Fluorescence [610]Electrochemical [605] Mass Spectrometric [8325]

Restekreg ULTRA C-18 and CN Columns (250mm x 46mm 5micro)

Mobile Phase (11 MethanolWater) 15 mLmin

Supelcosil LC-PAH Columns A BConditions A 150mm x 46mm 5micro

Flow Rate 15 mLmin

Conditions B 50mm x 46mm 3micro

Flow Rate 30 mLmin

  • Slide 5
  • Slide 6
  • Slide 7
  • Slide 8
  • Slide 9
  • Slide 10
  • Slide 11
  • Slide 12
  • Slide 13
  • Slide 14
  • Slide 15
  • Slide 16
  • Slide 17
  • Slide 18
  • Slide 19
  • Slide 20
  • Slide 21
  • Slide 22
  • Slide 23
  • Slide 24
  • Slide 25
  • Slide 26
  • Slide 27
  • Slide 28
  • Slide 29
  • Slide 30
  • Slide 31
  • Slide 32
  • Slide 33
  • Slide 34
  • Slide 35
  • Slide 36
  • Slide 37
  • Slide 38
  • Slide 39
  • Slide 40
Page 7: P igh erformance L iquid C hromatography  HPLC is a form of liquid chromatography used to separate compounds that are dissolved in solution. HPLC instruments

HPLC Injectors

In order to introduce a sample onto the column for analysis a special valve called the injector must be used to transfer the sample into the pressurized system

Injectors may look different from the outside but internally most are 6-port rotary valves These valves consist of a fixed body (the stator) plus an internal seal that rotates (the rotor)

Three internal passages connect alternate pairs of external ports The valves can switch between two positions referred to as the inject and load positions respectively In the load position the pump is connected to the column and the sample inlet is connected to one end of a piece of tubing called the sample loop

The other end of the sample loop is connected to the waste port Rotation results in reconnecting the various lines that enter the valve so that a sample volume can be inserted into the mobile phase that flows from the pump to the column inlet

The column

HPLC has seen significant improvement over

the last 20 years primarily due to improved

column technology

Packings are more uniform and smaller

Phases are commonly chemically bound to

the packing

Packing methods have improved

Pump types

bullIsocratic pump - delivers constant mobile phase composition

bullsolvent must be pre-mixed

bulllowest cost pump

bullGradient pump -delivers variable mobile phase composition

bullcan be used to mix and deliver an isocratic mobile phase or a gradient mobile phase

1048729 Reciprocation pumps

--- advantages small internal volume high output pressures

ready adaptability to gradient elution and constant flow rates

--- disadvantages produce a pulsed flow cause baseline noise

1048729 Displacement pumps

--- advantages output is pulse free

--- disadvantages limited solvent capacity (lt250 mL)

inconvenience when change solvents

1048729 Pneumatic pumps

--- advantages inexpensive output is pulse free

--- disadvantages limited solvent capacity not amenable to

gradient elution and limit pressures lt2000 psi

Packings

Originally these were irregular silica and

alumina A range of synthetic regularly

shaped packings are now available

Porous - channels through packing

Superficially porous - rough surface

Smooth - bead like

Packing size

As packing size is decreased efficiency and

pressure requirements are increased

Common diameters for analytical work

diameter plates

10 μm 5000

5 μm 9000

3 μm 15000

All are for a 15 cm x 46 mm column

Column body

Typically consist of stainless steel with a high

precision internal bore

Some manufacturers offer column inserts

- donrsquot need to repurchase the column

fittings

Others offer columns where the external

body can be compressed to improve packing

efficiency

HPLC column examples

Column stationary phases

Today most packing fall into four classes

Silica or alumina

Bound phases on either alumina or silica

Gels

Controlled-pore glass or silica

Absorption phases - alumina

common mobile phases

hexane chloroform 2-propanol

example application - amines

silica

common mobile phases

hexane chloroform 2-propanol

example applications - ethers esters

porphyrins fat-soluble vitamins

Partition phases

Can be broken down into

Normal phase - polar materials bound to the support

Reverse phase - non-polar materials bound to the support

Mixed phase - may have some of each

Partition phases

Normal

Amino (-NH2)

Cyano (-CN)

Diol (glycidoxy-ethylmethoxysilane)

Reverse

C-2 or RP-2 (-Si-CH2CH3)

C-8 or RP-8 (-Si-(CH2)7CH3)

C-18 or RP-18 (-Si-(CH2)17CH3)

Increasing the C number results in a thicker more retentive phase

Ion exchange phases

Strong cation - sulfonic acid group

Strong anionic - quarternary amine

Weak anion - primary amine

Weak cation - COOH

Size exclusion phases

Gels - organic or aqueous based

Controlled-pore - silica or glass

Must be selected based on pressure

requirements and size range

required for your application

Capillary and Microbore columns

Several companies have begun offering

columns with smaller ID

Microbore column - 1 mm ID packed

column

Capillary column - lt 1 mm ID internal

bound phase

These columns require smaller solvent flows

reduced sample size and improved detector

design

Capillary and Microbore columns Capillary and Microbore columns

Aromatic Compounds

mobile phase2 ethylacetate in

hexane

flow rate 4 μlmin

column Fusica II 300μm ID x

25 cm silica

sample

1 toluene

2 nitrobenzene

3 acetophenone

4 26-dinitrobenzene

injection 60 nl

detection UV 254 nm

Silica based columns

Silica is the ideal support for HPLC columns It offers a large mechanical stability excellent

physicochemical surface properties a wide range of bonding chemistries and is compatible

with a broad range of organic solvents

pH stability

In general HPLC columns are stable within a pH range of 2 to 8 If you are measuring a pH value the measurement must be done in the aqueous media before mixing the eluent with organic solvents

Modern HPLC columns can be used outside that pH range The new bonding chemistries allow use down to pH 1 for some stationary phases However please check vendorrsquos product information before using silica based column outside the pH range of 2 to 8 However best lifetimes are obtained between pH 20 and pH 68

Mechanical stability

Stationary phases based on silica are mechanically very stable The packed columns show no pressure limit and can be used at more than 40 MPa (6000 psi) without any problem However please avoid pressure shocks on the column Pressure shocks lead to channelling in the column which results in peak splitting in the corresponding chromatogram

Mobile phases (Eluents)

Silica based stationary phases are compatible with all organic solvents in the above mentioned pH range

bull Filter all prepared buffer through a 05 μm filter before using them in your HPLC system

The use of non pure solvents in HPLC causes irreversible adsorption of impurities on the column head These impurities block adsorption sites change the selectivity of the column and lead to peak splitting in the chromatogram In gradient elution impurities cause so called ldquoGhost Peaksrdquo

Ghost peaks are peaks that always appear in the same position on the chromatogram Their origin is not the sample but the impurities from the solvents or solvent additives Therefore it is highly recommended to run a gradientwithout injection in the beginning of each method to determine the ghost peaks

To avoid irreversible adsorption at the column head you should always use a pre-column

The use of a pre-column increases the life time of a column dramatically In addition a pre-column can filter solid parts stemming from pump seals or injection rotors An alternative to a pre-column is an in-line filter These filters are attached directly to the column These filters get rid of solid parts in the eluent but will not avoid irreversible adsorption of organic impurities

Detector Systems

Virtually every chemical and physical

property that can be measured in

solution has been look at

Detectors fall roughly into two classes

Bulk property - measures an overall change in the mobile phase

Solute property - measures a solute specific property

Properties of a good detector

A detector must provide

bull high sensitivity low detection limits

bull linearity

bullReproducibility

This is true for any detector

Each detector will have specific advantages and will vary as to peak shape and spread noise and flowtemperature dependence they have

UVVis detector

A solute property detector

Sample must exhibit absorption in UVVis range Solvent must not absorb significantly at the measured wavelength

Types Filter photometer - single

bull Variable wavelength

bull Multiwavelength

Refractive index detector

Bulk property detector - general purpose

Based on refraction of light as it passes from

one media to another Presence of a solute

changes the refractive index of the solvent

Heat of absorption detector

A small amount of heat is released when a sample absorbs on a suitable surface

This detector can measure this

Electrochemical detectors

A number of properties have been evaluated

Detector types

Dielectic constant

Amperometric

Conductometric

Polarographic

Potentiometric

Dielectric constant detector

Bulk property detector

Measures changes in polarity of the liquid phase passing through the cell

Conductometric detector

Measures conductivity of the solvent Useful for

solutions of ions

Amperometric detectors

Most frequently applied type of electrochemical

detector

A known potential is applied across a set of

electrodes - typically a glassy carbon type

Ability to oxidize or reduce a species can be

measured

Typically limited to working with a specific class

of materials per analysis

Separation is based on the analytersquos relative solubility between two liquid phases

Stationary PhaseMobile Phase

Solvent Bonded Phase

Normal Phase- Polar stationary phase and non-polar solvent

bull Reverse Phase- Non-polar stationary phase and a polar solvent

Methan0l CH3OH

bull Acetonitrile CH3CN

bull Tetrahydrofuran

bull Water H2O

Solid Support - Backbone for bonded phases Usually 10micro 5micro or 3micro silica or polymeric

particles Bonded Phases - Functional groups firmly linked

(chemically bound) to the solid support Extremely stable Reproducible

Guard - Protects the analytical column Particles Interferences Prolongs the life of the analytical column

bull Analytical - Performs the separation

C-2 Ethyl Silyl -Si-CH2-CH3

bullCN Cyanopropyl Silyl -Si-(CH2)3-CN

bull C-18C-18 Octadecyl SilylOctadecyl Silyl -Si-(CH -Si-(CH22))1717--

CHCH33

bullC-8 Octyl Silyl -Si-(CH2)7-CH3

Pump

Injector

ColumnDetector

Mobile Phases

Gradient Controller

bull

UV Single wavelength (filter) [610 8330] Variable wavelength (monochromator)

[8316 8325] Multiple wavelengths (PDA) [555]

Fluorescence [610]Electrochemical [605] Mass Spectrometric [8325]

Restekreg ULTRA C-18 and CN Columns (250mm x 46mm 5micro)

Mobile Phase (11 MethanolWater) 15 mLmin

Supelcosil LC-PAH Columns A BConditions A 150mm x 46mm 5micro

Flow Rate 15 mLmin

Conditions B 50mm x 46mm 3micro

Flow Rate 30 mLmin

  • Slide 5
  • Slide 6
  • Slide 7
  • Slide 8
  • Slide 9
  • Slide 10
  • Slide 11
  • Slide 12
  • Slide 13
  • Slide 14
  • Slide 15
  • Slide 16
  • Slide 17
  • Slide 18
  • Slide 19
  • Slide 20
  • Slide 21
  • Slide 22
  • Slide 23
  • Slide 24
  • Slide 25
  • Slide 26
  • Slide 27
  • Slide 28
  • Slide 29
  • Slide 30
  • Slide 31
  • Slide 32
  • Slide 33
  • Slide 34
  • Slide 35
  • Slide 36
  • Slide 37
  • Slide 38
  • Slide 39
  • Slide 40
Page 8: P igh erformance L iquid C hromatography  HPLC is a form of liquid chromatography used to separate compounds that are dissolved in solution. HPLC instruments

Injectors may look different from the outside but internally most are 6-port rotary valves These valves consist of a fixed body (the stator) plus an internal seal that rotates (the rotor)

Three internal passages connect alternate pairs of external ports The valves can switch between two positions referred to as the inject and load positions respectively In the load position the pump is connected to the column and the sample inlet is connected to one end of a piece of tubing called the sample loop

The other end of the sample loop is connected to the waste port Rotation results in reconnecting the various lines that enter the valve so that a sample volume can be inserted into the mobile phase that flows from the pump to the column inlet

The column

HPLC has seen significant improvement over

the last 20 years primarily due to improved

column technology

Packings are more uniform and smaller

Phases are commonly chemically bound to

the packing

Packing methods have improved

Pump types

bullIsocratic pump - delivers constant mobile phase composition

bullsolvent must be pre-mixed

bulllowest cost pump

bullGradient pump -delivers variable mobile phase composition

bullcan be used to mix and deliver an isocratic mobile phase or a gradient mobile phase

1048729 Reciprocation pumps

--- advantages small internal volume high output pressures

ready adaptability to gradient elution and constant flow rates

--- disadvantages produce a pulsed flow cause baseline noise

1048729 Displacement pumps

--- advantages output is pulse free

--- disadvantages limited solvent capacity (lt250 mL)

inconvenience when change solvents

1048729 Pneumatic pumps

--- advantages inexpensive output is pulse free

--- disadvantages limited solvent capacity not amenable to

gradient elution and limit pressures lt2000 psi

Packings

Originally these were irregular silica and

alumina A range of synthetic regularly

shaped packings are now available

Porous - channels through packing

Superficially porous - rough surface

Smooth - bead like

Packing size

As packing size is decreased efficiency and

pressure requirements are increased

Common diameters for analytical work

diameter plates

10 μm 5000

5 μm 9000

3 μm 15000

All are for a 15 cm x 46 mm column

Column body

Typically consist of stainless steel with a high

precision internal bore

Some manufacturers offer column inserts

- donrsquot need to repurchase the column

fittings

Others offer columns where the external

body can be compressed to improve packing

efficiency

HPLC column examples

Column stationary phases

Today most packing fall into four classes

Silica or alumina

Bound phases on either alumina or silica

Gels

Controlled-pore glass or silica

Absorption phases - alumina

common mobile phases

hexane chloroform 2-propanol

example application - amines

silica

common mobile phases

hexane chloroform 2-propanol

example applications - ethers esters

porphyrins fat-soluble vitamins

Partition phases

Can be broken down into

Normal phase - polar materials bound to the support

Reverse phase - non-polar materials bound to the support

Mixed phase - may have some of each

Partition phases

Normal

Amino (-NH2)

Cyano (-CN)

Diol (glycidoxy-ethylmethoxysilane)

Reverse

C-2 or RP-2 (-Si-CH2CH3)

C-8 or RP-8 (-Si-(CH2)7CH3)

C-18 or RP-18 (-Si-(CH2)17CH3)

Increasing the C number results in a thicker more retentive phase

Ion exchange phases

Strong cation - sulfonic acid group

Strong anionic - quarternary amine

Weak anion - primary amine

Weak cation - COOH

Size exclusion phases

Gels - organic or aqueous based

Controlled-pore - silica or glass

Must be selected based on pressure

requirements and size range

required for your application

Capillary and Microbore columns

Several companies have begun offering

columns with smaller ID

Microbore column - 1 mm ID packed

column

Capillary column - lt 1 mm ID internal

bound phase

These columns require smaller solvent flows

reduced sample size and improved detector

design

Capillary and Microbore columns Capillary and Microbore columns

Aromatic Compounds

mobile phase2 ethylacetate in

hexane

flow rate 4 μlmin

column Fusica II 300μm ID x

25 cm silica

sample

1 toluene

2 nitrobenzene

3 acetophenone

4 26-dinitrobenzene

injection 60 nl

detection UV 254 nm

Silica based columns

Silica is the ideal support for HPLC columns It offers a large mechanical stability excellent

physicochemical surface properties a wide range of bonding chemistries and is compatible

with a broad range of organic solvents

pH stability

In general HPLC columns are stable within a pH range of 2 to 8 If you are measuring a pH value the measurement must be done in the aqueous media before mixing the eluent with organic solvents

Modern HPLC columns can be used outside that pH range The new bonding chemistries allow use down to pH 1 for some stationary phases However please check vendorrsquos product information before using silica based column outside the pH range of 2 to 8 However best lifetimes are obtained between pH 20 and pH 68

Mechanical stability

Stationary phases based on silica are mechanically very stable The packed columns show no pressure limit and can be used at more than 40 MPa (6000 psi) without any problem However please avoid pressure shocks on the column Pressure shocks lead to channelling in the column which results in peak splitting in the corresponding chromatogram

Mobile phases (Eluents)

Silica based stationary phases are compatible with all organic solvents in the above mentioned pH range

bull Filter all prepared buffer through a 05 μm filter before using them in your HPLC system

The use of non pure solvents in HPLC causes irreversible adsorption of impurities on the column head These impurities block adsorption sites change the selectivity of the column and lead to peak splitting in the chromatogram In gradient elution impurities cause so called ldquoGhost Peaksrdquo

Ghost peaks are peaks that always appear in the same position on the chromatogram Their origin is not the sample but the impurities from the solvents or solvent additives Therefore it is highly recommended to run a gradientwithout injection in the beginning of each method to determine the ghost peaks

To avoid irreversible adsorption at the column head you should always use a pre-column

The use of a pre-column increases the life time of a column dramatically In addition a pre-column can filter solid parts stemming from pump seals or injection rotors An alternative to a pre-column is an in-line filter These filters are attached directly to the column These filters get rid of solid parts in the eluent but will not avoid irreversible adsorption of organic impurities

Detector Systems

Virtually every chemical and physical

property that can be measured in

solution has been look at

Detectors fall roughly into two classes

Bulk property - measures an overall change in the mobile phase

Solute property - measures a solute specific property

Properties of a good detector

A detector must provide

bull high sensitivity low detection limits

bull linearity

bullReproducibility

This is true for any detector

Each detector will have specific advantages and will vary as to peak shape and spread noise and flowtemperature dependence they have

UVVis detector

A solute property detector

Sample must exhibit absorption in UVVis range Solvent must not absorb significantly at the measured wavelength

Types Filter photometer - single

bull Variable wavelength

bull Multiwavelength

Refractive index detector

Bulk property detector - general purpose

Based on refraction of light as it passes from

one media to another Presence of a solute

changes the refractive index of the solvent

Heat of absorption detector

A small amount of heat is released when a sample absorbs on a suitable surface

This detector can measure this

Electrochemical detectors

A number of properties have been evaluated

Detector types

Dielectic constant

Amperometric

Conductometric

Polarographic

Potentiometric

Dielectric constant detector

Bulk property detector

Measures changes in polarity of the liquid phase passing through the cell

Conductometric detector

Measures conductivity of the solvent Useful for

solutions of ions

Amperometric detectors

Most frequently applied type of electrochemical

detector

A known potential is applied across a set of

electrodes - typically a glassy carbon type

Ability to oxidize or reduce a species can be

measured

Typically limited to working with a specific class

of materials per analysis

Separation is based on the analytersquos relative solubility between two liquid phases

Stationary PhaseMobile Phase

Solvent Bonded Phase

Normal Phase- Polar stationary phase and non-polar solvent

bull Reverse Phase- Non-polar stationary phase and a polar solvent

Methan0l CH3OH

bull Acetonitrile CH3CN

bull Tetrahydrofuran

bull Water H2O

Solid Support - Backbone for bonded phases Usually 10micro 5micro or 3micro silica or polymeric

particles Bonded Phases - Functional groups firmly linked

(chemically bound) to the solid support Extremely stable Reproducible

Guard - Protects the analytical column Particles Interferences Prolongs the life of the analytical column

bull Analytical - Performs the separation

C-2 Ethyl Silyl -Si-CH2-CH3

bullCN Cyanopropyl Silyl -Si-(CH2)3-CN

bull C-18C-18 Octadecyl SilylOctadecyl Silyl -Si-(CH -Si-(CH22))1717--

CHCH33

bullC-8 Octyl Silyl -Si-(CH2)7-CH3

Pump

Injector

ColumnDetector

Mobile Phases

Gradient Controller

bull

UV Single wavelength (filter) [610 8330] Variable wavelength (monochromator)

[8316 8325] Multiple wavelengths (PDA) [555]

Fluorescence [610]Electrochemical [605] Mass Spectrometric [8325]

Restekreg ULTRA C-18 and CN Columns (250mm x 46mm 5micro)

Mobile Phase (11 MethanolWater) 15 mLmin

Supelcosil LC-PAH Columns A BConditions A 150mm x 46mm 5micro

Flow Rate 15 mLmin

Conditions B 50mm x 46mm 3micro

Flow Rate 30 mLmin

  • Slide 5
  • Slide 6
  • Slide 7
  • Slide 8
  • Slide 9
  • Slide 10
  • Slide 11
  • Slide 12
  • Slide 13
  • Slide 14
  • Slide 15
  • Slide 16
  • Slide 17
  • Slide 18
  • Slide 19
  • Slide 20
  • Slide 21
  • Slide 22
  • Slide 23
  • Slide 24
  • Slide 25
  • Slide 26
  • Slide 27
  • Slide 28
  • Slide 29
  • Slide 30
  • Slide 31
  • Slide 32
  • Slide 33
  • Slide 34
  • Slide 35
  • Slide 36
  • Slide 37
  • Slide 38
  • Slide 39
  • Slide 40
Page 9: P igh erformance L iquid C hromatography  HPLC is a form of liquid chromatography used to separate compounds that are dissolved in solution. HPLC instruments

The column

HPLC has seen significant improvement over

the last 20 years primarily due to improved

column technology

Packings are more uniform and smaller

Phases are commonly chemically bound to

the packing

Packing methods have improved

Pump types

bullIsocratic pump - delivers constant mobile phase composition

bullsolvent must be pre-mixed

bulllowest cost pump

bullGradient pump -delivers variable mobile phase composition

bullcan be used to mix and deliver an isocratic mobile phase or a gradient mobile phase

1048729 Reciprocation pumps

--- advantages small internal volume high output pressures

ready adaptability to gradient elution and constant flow rates

--- disadvantages produce a pulsed flow cause baseline noise

1048729 Displacement pumps

--- advantages output is pulse free

--- disadvantages limited solvent capacity (lt250 mL)

inconvenience when change solvents

1048729 Pneumatic pumps

--- advantages inexpensive output is pulse free

--- disadvantages limited solvent capacity not amenable to

gradient elution and limit pressures lt2000 psi

Packings

Originally these were irregular silica and

alumina A range of synthetic regularly

shaped packings are now available

Porous - channels through packing

Superficially porous - rough surface

Smooth - bead like

Packing size

As packing size is decreased efficiency and

pressure requirements are increased

Common diameters for analytical work

diameter plates

10 μm 5000

5 μm 9000

3 μm 15000

All are for a 15 cm x 46 mm column

Column body

Typically consist of stainless steel with a high

precision internal bore

Some manufacturers offer column inserts

- donrsquot need to repurchase the column

fittings

Others offer columns where the external

body can be compressed to improve packing

efficiency

HPLC column examples

Column stationary phases

Today most packing fall into four classes

Silica or alumina

Bound phases on either alumina or silica

Gels

Controlled-pore glass or silica

Absorption phases - alumina

common mobile phases

hexane chloroform 2-propanol

example application - amines

silica

common mobile phases

hexane chloroform 2-propanol

example applications - ethers esters

porphyrins fat-soluble vitamins

Partition phases

Can be broken down into

Normal phase - polar materials bound to the support

Reverse phase - non-polar materials bound to the support

Mixed phase - may have some of each

Partition phases

Normal

Amino (-NH2)

Cyano (-CN)

Diol (glycidoxy-ethylmethoxysilane)

Reverse

C-2 or RP-2 (-Si-CH2CH3)

C-8 or RP-8 (-Si-(CH2)7CH3)

C-18 or RP-18 (-Si-(CH2)17CH3)

Increasing the C number results in a thicker more retentive phase

Ion exchange phases

Strong cation - sulfonic acid group

Strong anionic - quarternary amine

Weak anion - primary amine

Weak cation - COOH

Size exclusion phases

Gels - organic or aqueous based

Controlled-pore - silica or glass

Must be selected based on pressure

requirements and size range

required for your application

Capillary and Microbore columns

Several companies have begun offering

columns with smaller ID

Microbore column - 1 mm ID packed

column

Capillary column - lt 1 mm ID internal

bound phase

These columns require smaller solvent flows

reduced sample size and improved detector

design

Capillary and Microbore columns Capillary and Microbore columns

Aromatic Compounds

mobile phase2 ethylacetate in

hexane

flow rate 4 μlmin

column Fusica II 300μm ID x

25 cm silica

sample

1 toluene

2 nitrobenzene

3 acetophenone

4 26-dinitrobenzene

injection 60 nl

detection UV 254 nm

Silica based columns

Silica is the ideal support for HPLC columns It offers a large mechanical stability excellent

physicochemical surface properties a wide range of bonding chemistries and is compatible

with a broad range of organic solvents

pH stability

In general HPLC columns are stable within a pH range of 2 to 8 If you are measuring a pH value the measurement must be done in the aqueous media before mixing the eluent with organic solvents

Modern HPLC columns can be used outside that pH range The new bonding chemistries allow use down to pH 1 for some stationary phases However please check vendorrsquos product information before using silica based column outside the pH range of 2 to 8 However best lifetimes are obtained between pH 20 and pH 68

Mechanical stability

Stationary phases based on silica are mechanically very stable The packed columns show no pressure limit and can be used at more than 40 MPa (6000 psi) without any problem However please avoid pressure shocks on the column Pressure shocks lead to channelling in the column which results in peak splitting in the corresponding chromatogram

Mobile phases (Eluents)

Silica based stationary phases are compatible with all organic solvents in the above mentioned pH range

bull Filter all prepared buffer through a 05 μm filter before using them in your HPLC system

The use of non pure solvents in HPLC causes irreversible adsorption of impurities on the column head These impurities block adsorption sites change the selectivity of the column and lead to peak splitting in the chromatogram In gradient elution impurities cause so called ldquoGhost Peaksrdquo

Ghost peaks are peaks that always appear in the same position on the chromatogram Their origin is not the sample but the impurities from the solvents or solvent additives Therefore it is highly recommended to run a gradientwithout injection in the beginning of each method to determine the ghost peaks

To avoid irreversible adsorption at the column head you should always use a pre-column

The use of a pre-column increases the life time of a column dramatically In addition a pre-column can filter solid parts stemming from pump seals or injection rotors An alternative to a pre-column is an in-line filter These filters are attached directly to the column These filters get rid of solid parts in the eluent but will not avoid irreversible adsorption of organic impurities

Detector Systems

Virtually every chemical and physical

property that can be measured in

solution has been look at

Detectors fall roughly into two classes

Bulk property - measures an overall change in the mobile phase

Solute property - measures a solute specific property

Properties of a good detector

A detector must provide

bull high sensitivity low detection limits

bull linearity

bullReproducibility

This is true for any detector

Each detector will have specific advantages and will vary as to peak shape and spread noise and flowtemperature dependence they have

UVVis detector

A solute property detector

Sample must exhibit absorption in UVVis range Solvent must not absorb significantly at the measured wavelength

Types Filter photometer - single

bull Variable wavelength

bull Multiwavelength

Refractive index detector

Bulk property detector - general purpose

Based on refraction of light as it passes from

one media to another Presence of a solute

changes the refractive index of the solvent

Heat of absorption detector

A small amount of heat is released when a sample absorbs on a suitable surface

This detector can measure this

Electrochemical detectors

A number of properties have been evaluated

Detector types

Dielectic constant

Amperometric

Conductometric

Polarographic

Potentiometric

Dielectric constant detector

Bulk property detector

Measures changes in polarity of the liquid phase passing through the cell

Conductometric detector

Measures conductivity of the solvent Useful for

solutions of ions

Amperometric detectors

Most frequently applied type of electrochemical

detector

A known potential is applied across a set of

electrodes - typically a glassy carbon type

Ability to oxidize or reduce a species can be

measured

Typically limited to working with a specific class

of materials per analysis

Separation is based on the analytersquos relative solubility between two liquid phases

Stationary PhaseMobile Phase

Solvent Bonded Phase

Normal Phase- Polar stationary phase and non-polar solvent

bull Reverse Phase- Non-polar stationary phase and a polar solvent

Methan0l CH3OH

bull Acetonitrile CH3CN

bull Tetrahydrofuran

bull Water H2O

Solid Support - Backbone for bonded phases Usually 10micro 5micro or 3micro silica or polymeric

particles Bonded Phases - Functional groups firmly linked

(chemically bound) to the solid support Extremely stable Reproducible

Guard - Protects the analytical column Particles Interferences Prolongs the life of the analytical column

bull Analytical - Performs the separation

C-2 Ethyl Silyl -Si-CH2-CH3

bullCN Cyanopropyl Silyl -Si-(CH2)3-CN

bull C-18C-18 Octadecyl SilylOctadecyl Silyl -Si-(CH -Si-(CH22))1717--

CHCH33

bullC-8 Octyl Silyl -Si-(CH2)7-CH3

Pump

Injector

ColumnDetector

Mobile Phases

Gradient Controller

bull

UV Single wavelength (filter) [610 8330] Variable wavelength (monochromator)

[8316 8325] Multiple wavelengths (PDA) [555]

Fluorescence [610]Electrochemical [605] Mass Spectrometric [8325]

Restekreg ULTRA C-18 and CN Columns (250mm x 46mm 5micro)

Mobile Phase (11 MethanolWater) 15 mLmin

Supelcosil LC-PAH Columns A BConditions A 150mm x 46mm 5micro

Flow Rate 15 mLmin

Conditions B 50mm x 46mm 3micro

Flow Rate 30 mLmin

  • Slide 5
  • Slide 6
  • Slide 7
  • Slide 8
  • Slide 9
  • Slide 10
  • Slide 11
  • Slide 12
  • Slide 13
  • Slide 14
  • Slide 15
  • Slide 16
  • Slide 17
  • Slide 18
  • Slide 19
  • Slide 20
  • Slide 21
  • Slide 22
  • Slide 23
  • Slide 24
  • Slide 25
  • Slide 26
  • Slide 27
  • Slide 28
  • Slide 29
  • Slide 30
  • Slide 31
  • Slide 32
  • Slide 33
  • Slide 34
  • Slide 35
  • Slide 36
  • Slide 37
  • Slide 38
  • Slide 39
  • Slide 40
Page 10: P igh erformance L iquid C hromatography  HPLC is a form of liquid chromatography used to separate compounds that are dissolved in solution. HPLC instruments

Pump types

bullIsocratic pump - delivers constant mobile phase composition

bullsolvent must be pre-mixed

bulllowest cost pump

bullGradient pump -delivers variable mobile phase composition

bullcan be used to mix and deliver an isocratic mobile phase or a gradient mobile phase

1048729 Reciprocation pumps

--- advantages small internal volume high output pressures

ready adaptability to gradient elution and constant flow rates

--- disadvantages produce a pulsed flow cause baseline noise

1048729 Displacement pumps

--- advantages output is pulse free

--- disadvantages limited solvent capacity (lt250 mL)

inconvenience when change solvents

1048729 Pneumatic pumps

--- advantages inexpensive output is pulse free

--- disadvantages limited solvent capacity not amenable to

gradient elution and limit pressures lt2000 psi

Packings

Originally these were irregular silica and

alumina A range of synthetic regularly

shaped packings are now available

Porous - channels through packing

Superficially porous - rough surface

Smooth - bead like

Packing size

As packing size is decreased efficiency and

pressure requirements are increased

Common diameters for analytical work

diameter plates

10 μm 5000

5 μm 9000

3 μm 15000

All are for a 15 cm x 46 mm column

Column body

Typically consist of stainless steel with a high

precision internal bore

Some manufacturers offer column inserts

- donrsquot need to repurchase the column

fittings

Others offer columns where the external

body can be compressed to improve packing

efficiency

HPLC column examples

Column stationary phases

Today most packing fall into four classes

Silica or alumina

Bound phases on either alumina or silica

Gels

Controlled-pore glass or silica

Absorption phases - alumina

common mobile phases

hexane chloroform 2-propanol

example application - amines

silica

common mobile phases

hexane chloroform 2-propanol

example applications - ethers esters

porphyrins fat-soluble vitamins

Partition phases

Can be broken down into

Normal phase - polar materials bound to the support

Reverse phase - non-polar materials bound to the support

Mixed phase - may have some of each

Partition phases

Normal

Amino (-NH2)

Cyano (-CN)

Diol (glycidoxy-ethylmethoxysilane)

Reverse

C-2 or RP-2 (-Si-CH2CH3)

C-8 or RP-8 (-Si-(CH2)7CH3)

C-18 or RP-18 (-Si-(CH2)17CH3)

Increasing the C number results in a thicker more retentive phase

Ion exchange phases

Strong cation - sulfonic acid group

Strong anionic - quarternary amine

Weak anion - primary amine

Weak cation - COOH

Size exclusion phases

Gels - organic or aqueous based

Controlled-pore - silica or glass

Must be selected based on pressure

requirements and size range

required for your application

Capillary and Microbore columns

Several companies have begun offering

columns with smaller ID

Microbore column - 1 mm ID packed

column

Capillary column - lt 1 mm ID internal

bound phase

These columns require smaller solvent flows

reduced sample size and improved detector

design

Capillary and Microbore columns Capillary and Microbore columns

Aromatic Compounds

mobile phase2 ethylacetate in

hexane

flow rate 4 μlmin

column Fusica II 300μm ID x

25 cm silica

sample

1 toluene

2 nitrobenzene

3 acetophenone

4 26-dinitrobenzene

injection 60 nl

detection UV 254 nm

Silica based columns

Silica is the ideal support for HPLC columns It offers a large mechanical stability excellent

physicochemical surface properties a wide range of bonding chemistries and is compatible

with a broad range of organic solvents

pH stability

In general HPLC columns are stable within a pH range of 2 to 8 If you are measuring a pH value the measurement must be done in the aqueous media before mixing the eluent with organic solvents

Modern HPLC columns can be used outside that pH range The new bonding chemistries allow use down to pH 1 for some stationary phases However please check vendorrsquos product information before using silica based column outside the pH range of 2 to 8 However best lifetimes are obtained between pH 20 and pH 68

Mechanical stability

Stationary phases based on silica are mechanically very stable The packed columns show no pressure limit and can be used at more than 40 MPa (6000 psi) without any problem However please avoid pressure shocks on the column Pressure shocks lead to channelling in the column which results in peak splitting in the corresponding chromatogram

Mobile phases (Eluents)

Silica based stationary phases are compatible with all organic solvents in the above mentioned pH range

bull Filter all prepared buffer through a 05 μm filter before using them in your HPLC system

The use of non pure solvents in HPLC causes irreversible adsorption of impurities on the column head These impurities block adsorption sites change the selectivity of the column and lead to peak splitting in the chromatogram In gradient elution impurities cause so called ldquoGhost Peaksrdquo

Ghost peaks are peaks that always appear in the same position on the chromatogram Their origin is not the sample but the impurities from the solvents or solvent additives Therefore it is highly recommended to run a gradientwithout injection in the beginning of each method to determine the ghost peaks

To avoid irreversible adsorption at the column head you should always use a pre-column

The use of a pre-column increases the life time of a column dramatically In addition a pre-column can filter solid parts stemming from pump seals or injection rotors An alternative to a pre-column is an in-line filter These filters are attached directly to the column These filters get rid of solid parts in the eluent but will not avoid irreversible adsorption of organic impurities

Detector Systems

Virtually every chemical and physical

property that can be measured in

solution has been look at

Detectors fall roughly into two classes

Bulk property - measures an overall change in the mobile phase

Solute property - measures a solute specific property

Properties of a good detector

A detector must provide

bull high sensitivity low detection limits

bull linearity

bullReproducibility

This is true for any detector

Each detector will have specific advantages and will vary as to peak shape and spread noise and flowtemperature dependence they have

UVVis detector

A solute property detector

Sample must exhibit absorption in UVVis range Solvent must not absorb significantly at the measured wavelength

Types Filter photometer - single

bull Variable wavelength

bull Multiwavelength

Refractive index detector

Bulk property detector - general purpose

Based on refraction of light as it passes from

one media to another Presence of a solute

changes the refractive index of the solvent

Heat of absorption detector

A small amount of heat is released when a sample absorbs on a suitable surface

This detector can measure this

Electrochemical detectors

A number of properties have been evaluated

Detector types

Dielectic constant

Amperometric

Conductometric

Polarographic

Potentiometric

Dielectric constant detector

Bulk property detector

Measures changes in polarity of the liquid phase passing through the cell

Conductometric detector

Measures conductivity of the solvent Useful for

solutions of ions

Amperometric detectors

Most frequently applied type of electrochemical

detector

A known potential is applied across a set of

electrodes - typically a glassy carbon type

Ability to oxidize or reduce a species can be

measured

Typically limited to working with a specific class

of materials per analysis

Separation is based on the analytersquos relative solubility between two liquid phases

Stationary PhaseMobile Phase

Solvent Bonded Phase

Normal Phase- Polar stationary phase and non-polar solvent

bull Reverse Phase- Non-polar stationary phase and a polar solvent

Methan0l CH3OH

bull Acetonitrile CH3CN

bull Tetrahydrofuran

bull Water H2O

Solid Support - Backbone for bonded phases Usually 10micro 5micro or 3micro silica or polymeric

particles Bonded Phases - Functional groups firmly linked

(chemically bound) to the solid support Extremely stable Reproducible

Guard - Protects the analytical column Particles Interferences Prolongs the life of the analytical column

bull Analytical - Performs the separation

C-2 Ethyl Silyl -Si-CH2-CH3

bullCN Cyanopropyl Silyl -Si-(CH2)3-CN

bull C-18C-18 Octadecyl SilylOctadecyl Silyl -Si-(CH -Si-(CH22))1717--

CHCH33

bullC-8 Octyl Silyl -Si-(CH2)7-CH3

Pump

Injector

ColumnDetector

Mobile Phases

Gradient Controller

bull

UV Single wavelength (filter) [610 8330] Variable wavelength (monochromator)

[8316 8325] Multiple wavelengths (PDA) [555]

Fluorescence [610]Electrochemical [605] Mass Spectrometric [8325]

Restekreg ULTRA C-18 and CN Columns (250mm x 46mm 5micro)

Mobile Phase (11 MethanolWater) 15 mLmin

Supelcosil LC-PAH Columns A BConditions A 150mm x 46mm 5micro

Flow Rate 15 mLmin

Conditions B 50mm x 46mm 3micro

Flow Rate 30 mLmin

  • Slide 5
  • Slide 6
  • Slide 7
  • Slide 8
  • Slide 9
  • Slide 10
  • Slide 11
  • Slide 12
  • Slide 13
  • Slide 14
  • Slide 15
  • Slide 16
  • Slide 17
  • Slide 18
  • Slide 19
  • Slide 20
  • Slide 21
  • Slide 22
  • Slide 23
  • Slide 24
  • Slide 25
  • Slide 26
  • Slide 27
  • Slide 28
  • Slide 29
  • Slide 30
  • Slide 31
  • Slide 32
  • Slide 33
  • Slide 34
  • Slide 35
  • Slide 36
  • Slide 37
  • Slide 38
  • Slide 39
  • Slide 40
Page 11: P igh erformance L iquid C hromatography  HPLC is a form of liquid chromatography used to separate compounds that are dissolved in solution. HPLC instruments

1048729 Reciprocation pumps

--- advantages small internal volume high output pressures

ready adaptability to gradient elution and constant flow rates

--- disadvantages produce a pulsed flow cause baseline noise

1048729 Displacement pumps

--- advantages output is pulse free

--- disadvantages limited solvent capacity (lt250 mL)

inconvenience when change solvents

1048729 Pneumatic pumps

--- advantages inexpensive output is pulse free

--- disadvantages limited solvent capacity not amenable to

gradient elution and limit pressures lt2000 psi

Packings

Originally these were irregular silica and

alumina A range of synthetic regularly

shaped packings are now available

Porous - channels through packing

Superficially porous - rough surface

Smooth - bead like

Packing size

As packing size is decreased efficiency and

pressure requirements are increased

Common diameters for analytical work

diameter plates

10 μm 5000

5 μm 9000

3 μm 15000

All are for a 15 cm x 46 mm column

Column body

Typically consist of stainless steel with a high

precision internal bore

Some manufacturers offer column inserts

- donrsquot need to repurchase the column

fittings

Others offer columns where the external

body can be compressed to improve packing

efficiency

HPLC column examples

Column stationary phases

Today most packing fall into four classes

Silica or alumina

Bound phases on either alumina or silica

Gels

Controlled-pore glass or silica

Absorption phases - alumina

common mobile phases

hexane chloroform 2-propanol

example application - amines

silica

common mobile phases

hexane chloroform 2-propanol

example applications - ethers esters

porphyrins fat-soluble vitamins

Partition phases

Can be broken down into

Normal phase - polar materials bound to the support

Reverse phase - non-polar materials bound to the support

Mixed phase - may have some of each

Partition phases

Normal

Amino (-NH2)

Cyano (-CN)

Diol (glycidoxy-ethylmethoxysilane)

Reverse

C-2 or RP-2 (-Si-CH2CH3)

C-8 or RP-8 (-Si-(CH2)7CH3)

C-18 or RP-18 (-Si-(CH2)17CH3)

Increasing the C number results in a thicker more retentive phase

Ion exchange phases

Strong cation - sulfonic acid group

Strong anionic - quarternary amine

Weak anion - primary amine

Weak cation - COOH

Size exclusion phases

Gels - organic or aqueous based

Controlled-pore - silica or glass

Must be selected based on pressure

requirements and size range

required for your application

Capillary and Microbore columns

Several companies have begun offering

columns with smaller ID

Microbore column - 1 mm ID packed

column

Capillary column - lt 1 mm ID internal

bound phase

These columns require smaller solvent flows

reduced sample size and improved detector

design

Capillary and Microbore columns Capillary and Microbore columns

Aromatic Compounds

mobile phase2 ethylacetate in

hexane

flow rate 4 μlmin

column Fusica II 300μm ID x

25 cm silica

sample

1 toluene

2 nitrobenzene

3 acetophenone

4 26-dinitrobenzene

injection 60 nl

detection UV 254 nm

Silica based columns

Silica is the ideal support for HPLC columns It offers a large mechanical stability excellent

physicochemical surface properties a wide range of bonding chemistries and is compatible

with a broad range of organic solvents

pH stability

In general HPLC columns are stable within a pH range of 2 to 8 If you are measuring a pH value the measurement must be done in the aqueous media before mixing the eluent with organic solvents

Modern HPLC columns can be used outside that pH range The new bonding chemistries allow use down to pH 1 for some stationary phases However please check vendorrsquos product information before using silica based column outside the pH range of 2 to 8 However best lifetimes are obtained between pH 20 and pH 68

Mechanical stability

Stationary phases based on silica are mechanically very stable The packed columns show no pressure limit and can be used at more than 40 MPa (6000 psi) without any problem However please avoid pressure shocks on the column Pressure shocks lead to channelling in the column which results in peak splitting in the corresponding chromatogram

Mobile phases (Eluents)

Silica based stationary phases are compatible with all organic solvents in the above mentioned pH range

bull Filter all prepared buffer through a 05 μm filter before using them in your HPLC system

The use of non pure solvents in HPLC causes irreversible adsorption of impurities on the column head These impurities block adsorption sites change the selectivity of the column and lead to peak splitting in the chromatogram In gradient elution impurities cause so called ldquoGhost Peaksrdquo

Ghost peaks are peaks that always appear in the same position on the chromatogram Their origin is not the sample but the impurities from the solvents or solvent additives Therefore it is highly recommended to run a gradientwithout injection in the beginning of each method to determine the ghost peaks

To avoid irreversible adsorption at the column head you should always use a pre-column

The use of a pre-column increases the life time of a column dramatically In addition a pre-column can filter solid parts stemming from pump seals or injection rotors An alternative to a pre-column is an in-line filter These filters are attached directly to the column These filters get rid of solid parts in the eluent but will not avoid irreversible adsorption of organic impurities

Detector Systems

Virtually every chemical and physical

property that can be measured in

solution has been look at

Detectors fall roughly into two classes

Bulk property - measures an overall change in the mobile phase

Solute property - measures a solute specific property

Properties of a good detector

A detector must provide

bull high sensitivity low detection limits

bull linearity

bullReproducibility

This is true for any detector

Each detector will have specific advantages and will vary as to peak shape and spread noise and flowtemperature dependence they have

UVVis detector

A solute property detector

Sample must exhibit absorption in UVVis range Solvent must not absorb significantly at the measured wavelength

Types Filter photometer - single

bull Variable wavelength

bull Multiwavelength

Refractive index detector

Bulk property detector - general purpose

Based on refraction of light as it passes from

one media to another Presence of a solute

changes the refractive index of the solvent

Heat of absorption detector

A small amount of heat is released when a sample absorbs on a suitable surface

This detector can measure this

Electrochemical detectors

A number of properties have been evaluated

Detector types

Dielectic constant

Amperometric

Conductometric

Polarographic

Potentiometric

Dielectric constant detector

Bulk property detector

Measures changes in polarity of the liquid phase passing through the cell

Conductometric detector

Measures conductivity of the solvent Useful for

solutions of ions

Amperometric detectors

Most frequently applied type of electrochemical

detector

A known potential is applied across a set of

electrodes - typically a glassy carbon type

Ability to oxidize or reduce a species can be

measured

Typically limited to working with a specific class

of materials per analysis

Separation is based on the analytersquos relative solubility between two liquid phases

Stationary PhaseMobile Phase

Solvent Bonded Phase

Normal Phase- Polar stationary phase and non-polar solvent

bull Reverse Phase- Non-polar stationary phase and a polar solvent

Methan0l CH3OH

bull Acetonitrile CH3CN

bull Tetrahydrofuran

bull Water H2O

Solid Support - Backbone for bonded phases Usually 10micro 5micro or 3micro silica or polymeric

particles Bonded Phases - Functional groups firmly linked

(chemically bound) to the solid support Extremely stable Reproducible

Guard - Protects the analytical column Particles Interferences Prolongs the life of the analytical column

bull Analytical - Performs the separation

C-2 Ethyl Silyl -Si-CH2-CH3

bullCN Cyanopropyl Silyl -Si-(CH2)3-CN

bull C-18C-18 Octadecyl SilylOctadecyl Silyl -Si-(CH -Si-(CH22))1717--

CHCH33

bullC-8 Octyl Silyl -Si-(CH2)7-CH3

Pump

Injector

ColumnDetector

Mobile Phases

Gradient Controller

bull

UV Single wavelength (filter) [610 8330] Variable wavelength (monochromator)

[8316 8325] Multiple wavelengths (PDA) [555]

Fluorescence [610]Electrochemical [605] Mass Spectrometric [8325]

Restekreg ULTRA C-18 and CN Columns (250mm x 46mm 5micro)

Mobile Phase (11 MethanolWater) 15 mLmin

Supelcosil LC-PAH Columns A BConditions A 150mm x 46mm 5micro

Flow Rate 15 mLmin

Conditions B 50mm x 46mm 3micro

Flow Rate 30 mLmin

  • Slide 5
  • Slide 6
  • Slide 7
  • Slide 8
  • Slide 9
  • Slide 10
  • Slide 11
  • Slide 12
  • Slide 13
  • Slide 14
  • Slide 15
  • Slide 16
  • Slide 17
  • Slide 18
  • Slide 19
  • Slide 20
  • Slide 21
  • Slide 22
  • Slide 23
  • Slide 24
  • Slide 25
  • Slide 26
  • Slide 27
  • Slide 28
  • Slide 29
  • Slide 30
  • Slide 31
  • Slide 32
  • Slide 33
  • Slide 34
  • Slide 35
  • Slide 36
  • Slide 37
  • Slide 38
  • Slide 39
  • Slide 40
Page 12: P igh erformance L iquid C hromatography  HPLC is a form of liquid chromatography used to separate compounds that are dissolved in solution. HPLC instruments

Packings

Originally these were irregular silica and

alumina A range of synthetic regularly

shaped packings are now available

Porous - channels through packing

Superficially porous - rough surface

Smooth - bead like

Packing size

As packing size is decreased efficiency and

pressure requirements are increased

Common diameters for analytical work

diameter plates

10 μm 5000

5 μm 9000

3 μm 15000

All are for a 15 cm x 46 mm column

Column body

Typically consist of stainless steel with a high

precision internal bore

Some manufacturers offer column inserts

- donrsquot need to repurchase the column

fittings

Others offer columns where the external

body can be compressed to improve packing

efficiency

HPLC column examples

Column stationary phases

Today most packing fall into four classes

Silica or alumina

Bound phases on either alumina or silica

Gels

Controlled-pore glass or silica

Absorption phases - alumina

common mobile phases

hexane chloroform 2-propanol

example application - amines

silica

common mobile phases

hexane chloroform 2-propanol

example applications - ethers esters

porphyrins fat-soluble vitamins

Partition phases

Can be broken down into

Normal phase - polar materials bound to the support

Reverse phase - non-polar materials bound to the support

Mixed phase - may have some of each

Partition phases

Normal

Amino (-NH2)

Cyano (-CN)

Diol (glycidoxy-ethylmethoxysilane)

Reverse

C-2 or RP-2 (-Si-CH2CH3)

C-8 or RP-8 (-Si-(CH2)7CH3)

C-18 or RP-18 (-Si-(CH2)17CH3)

Increasing the C number results in a thicker more retentive phase

Ion exchange phases

Strong cation - sulfonic acid group

Strong anionic - quarternary amine

Weak anion - primary amine

Weak cation - COOH

Size exclusion phases

Gels - organic or aqueous based

Controlled-pore - silica or glass

Must be selected based on pressure

requirements and size range

required for your application

Capillary and Microbore columns

Several companies have begun offering

columns with smaller ID

Microbore column - 1 mm ID packed

column

Capillary column - lt 1 mm ID internal

bound phase

These columns require smaller solvent flows

reduced sample size and improved detector

design

Capillary and Microbore columns Capillary and Microbore columns

Aromatic Compounds

mobile phase2 ethylacetate in

hexane

flow rate 4 μlmin

column Fusica II 300μm ID x

25 cm silica

sample

1 toluene

2 nitrobenzene

3 acetophenone

4 26-dinitrobenzene

injection 60 nl

detection UV 254 nm

Silica based columns

Silica is the ideal support for HPLC columns It offers a large mechanical stability excellent

physicochemical surface properties a wide range of bonding chemistries and is compatible

with a broad range of organic solvents

pH stability

In general HPLC columns are stable within a pH range of 2 to 8 If you are measuring a pH value the measurement must be done in the aqueous media before mixing the eluent with organic solvents

Modern HPLC columns can be used outside that pH range The new bonding chemistries allow use down to pH 1 for some stationary phases However please check vendorrsquos product information before using silica based column outside the pH range of 2 to 8 However best lifetimes are obtained between pH 20 and pH 68

Mechanical stability

Stationary phases based on silica are mechanically very stable The packed columns show no pressure limit and can be used at more than 40 MPa (6000 psi) without any problem However please avoid pressure shocks on the column Pressure shocks lead to channelling in the column which results in peak splitting in the corresponding chromatogram

Mobile phases (Eluents)

Silica based stationary phases are compatible with all organic solvents in the above mentioned pH range

bull Filter all prepared buffer through a 05 μm filter before using them in your HPLC system

The use of non pure solvents in HPLC causes irreversible adsorption of impurities on the column head These impurities block adsorption sites change the selectivity of the column and lead to peak splitting in the chromatogram In gradient elution impurities cause so called ldquoGhost Peaksrdquo

Ghost peaks are peaks that always appear in the same position on the chromatogram Their origin is not the sample but the impurities from the solvents or solvent additives Therefore it is highly recommended to run a gradientwithout injection in the beginning of each method to determine the ghost peaks

To avoid irreversible adsorption at the column head you should always use a pre-column

The use of a pre-column increases the life time of a column dramatically In addition a pre-column can filter solid parts stemming from pump seals or injection rotors An alternative to a pre-column is an in-line filter These filters are attached directly to the column These filters get rid of solid parts in the eluent but will not avoid irreversible adsorption of organic impurities

Detector Systems

Virtually every chemical and physical

property that can be measured in

solution has been look at

Detectors fall roughly into two classes

Bulk property - measures an overall change in the mobile phase

Solute property - measures a solute specific property

Properties of a good detector

A detector must provide

bull high sensitivity low detection limits

bull linearity

bullReproducibility

This is true for any detector

Each detector will have specific advantages and will vary as to peak shape and spread noise and flowtemperature dependence they have

UVVis detector

A solute property detector

Sample must exhibit absorption in UVVis range Solvent must not absorb significantly at the measured wavelength

Types Filter photometer - single

bull Variable wavelength

bull Multiwavelength

Refractive index detector

Bulk property detector - general purpose

Based on refraction of light as it passes from

one media to another Presence of a solute

changes the refractive index of the solvent

Heat of absorption detector

A small amount of heat is released when a sample absorbs on a suitable surface

This detector can measure this

Electrochemical detectors

A number of properties have been evaluated

Detector types

Dielectic constant

Amperometric

Conductometric

Polarographic

Potentiometric

Dielectric constant detector

Bulk property detector

Measures changes in polarity of the liquid phase passing through the cell

Conductometric detector

Measures conductivity of the solvent Useful for

solutions of ions

Amperometric detectors

Most frequently applied type of electrochemical

detector

A known potential is applied across a set of

electrodes - typically a glassy carbon type

Ability to oxidize or reduce a species can be

measured

Typically limited to working with a specific class

of materials per analysis

Separation is based on the analytersquos relative solubility between two liquid phases

Stationary PhaseMobile Phase

Solvent Bonded Phase

Normal Phase- Polar stationary phase and non-polar solvent

bull Reverse Phase- Non-polar stationary phase and a polar solvent

Methan0l CH3OH

bull Acetonitrile CH3CN

bull Tetrahydrofuran

bull Water H2O

Solid Support - Backbone for bonded phases Usually 10micro 5micro or 3micro silica or polymeric

particles Bonded Phases - Functional groups firmly linked

(chemically bound) to the solid support Extremely stable Reproducible

Guard - Protects the analytical column Particles Interferences Prolongs the life of the analytical column

bull Analytical - Performs the separation

C-2 Ethyl Silyl -Si-CH2-CH3

bullCN Cyanopropyl Silyl -Si-(CH2)3-CN

bull C-18C-18 Octadecyl SilylOctadecyl Silyl -Si-(CH -Si-(CH22))1717--

CHCH33

bullC-8 Octyl Silyl -Si-(CH2)7-CH3

Pump

Injector

ColumnDetector

Mobile Phases

Gradient Controller

bull

UV Single wavelength (filter) [610 8330] Variable wavelength (monochromator)

[8316 8325] Multiple wavelengths (PDA) [555]

Fluorescence [610]Electrochemical [605] Mass Spectrometric [8325]

Restekreg ULTRA C-18 and CN Columns (250mm x 46mm 5micro)

Mobile Phase (11 MethanolWater) 15 mLmin

Supelcosil LC-PAH Columns A BConditions A 150mm x 46mm 5micro

Flow Rate 15 mLmin

Conditions B 50mm x 46mm 3micro

Flow Rate 30 mLmin

  • Slide 5
  • Slide 6
  • Slide 7
  • Slide 8
  • Slide 9
  • Slide 10
  • Slide 11
  • Slide 12
  • Slide 13
  • Slide 14
  • Slide 15
  • Slide 16
  • Slide 17
  • Slide 18
  • Slide 19
  • Slide 20
  • Slide 21
  • Slide 22
  • Slide 23
  • Slide 24
  • Slide 25
  • Slide 26
  • Slide 27
  • Slide 28
  • Slide 29
  • Slide 30
  • Slide 31
  • Slide 32
  • Slide 33
  • Slide 34
  • Slide 35
  • Slide 36
  • Slide 37
  • Slide 38
  • Slide 39
  • Slide 40
Page 13: P igh erformance L iquid C hromatography  HPLC is a form of liquid chromatography used to separate compounds that are dissolved in solution. HPLC instruments

Packing size

As packing size is decreased efficiency and

pressure requirements are increased

Common diameters for analytical work

diameter plates

10 μm 5000

5 μm 9000

3 μm 15000

All are for a 15 cm x 46 mm column

Column body

Typically consist of stainless steel with a high

precision internal bore

Some manufacturers offer column inserts

- donrsquot need to repurchase the column

fittings

Others offer columns where the external

body can be compressed to improve packing

efficiency

HPLC column examples

Column stationary phases

Today most packing fall into four classes

Silica or alumina

Bound phases on either alumina or silica

Gels

Controlled-pore glass or silica

Absorption phases - alumina

common mobile phases

hexane chloroform 2-propanol

example application - amines

silica

common mobile phases

hexane chloroform 2-propanol

example applications - ethers esters

porphyrins fat-soluble vitamins

Partition phases

Can be broken down into

Normal phase - polar materials bound to the support

Reverse phase - non-polar materials bound to the support

Mixed phase - may have some of each

Partition phases

Normal

Amino (-NH2)

Cyano (-CN)

Diol (glycidoxy-ethylmethoxysilane)

Reverse

C-2 or RP-2 (-Si-CH2CH3)

C-8 or RP-8 (-Si-(CH2)7CH3)

C-18 or RP-18 (-Si-(CH2)17CH3)

Increasing the C number results in a thicker more retentive phase

Ion exchange phases

Strong cation - sulfonic acid group

Strong anionic - quarternary amine

Weak anion - primary amine

Weak cation - COOH

Size exclusion phases

Gels - organic or aqueous based

Controlled-pore - silica or glass

Must be selected based on pressure

requirements and size range

required for your application

Capillary and Microbore columns

Several companies have begun offering

columns with smaller ID

Microbore column - 1 mm ID packed

column

Capillary column - lt 1 mm ID internal

bound phase

These columns require smaller solvent flows

reduced sample size and improved detector

design

Capillary and Microbore columns Capillary and Microbore columns

Aromatic Compounds

mobile phase2 ethylacetate in

hexane

flow rate 4 μlmin

column Fusica II 300μm ID x

25 cm silica

sample

1 toluene

2 nitrobenzene

3 acetophenone

4 26-dinitrobenzene

injection 60 nl

detection UV 254 nm

Silica based columns

Silica is the ideal support for HPLC columns It offers a large mechanical stability excellent

physicochemical surface properties a wide range of bonding chemistries and is compatible

with a broad range of organic solvents

pH stability

In general HPLC columns are stable within a pH range of 2 to 8 If you are measuring a pH value the measurement must be done in the aqueous media before mixing the eluent with organic solvents

Modern HPLC columns can be used outside that pH range The new bonding chemistries allow use down to pH 1 for some stationary phases However please check vendorrsquos product information before using silica based column outside the pH range of 2 to 8 However best lifetimes are obtained between pH 20 and pH 68

Mechanical stability

Stationary phases based on silica are mechanically very stable The packed columns show no pressure limit and can be used at more than 40 MPa (6000 psi) without any problem However please avoid pressure shocks on the column Pressure shocks lead to channelling in the column which results in peak splitting in the corresponding chromatogram

Mobile phases (Eluents)

Silica based stationary phases are compatible with all organic solvents in the above mentioned pH range

bull Filter all prepared buffer through a 05 μm filter before using them in your HPLC system

The use of non pure solvents in HPLC causes irreversible adsorption of impurities on the column head These impurities block adsorption sites change the selectivity of the column and lead to peak splitting in the chromatogram In gradient elution impurities cause so called ldquoGhost Peaksrdquo

Ghost peaks are peaks that always appear in the same position on the chromatogram Their origin is not the sample but the impurities from the solvents or solvent additives Therefore it is highly recommended to run a gradientwithout injection in the beginning of each method to determine the ghost peaks

To avoid irreversible adsorption at the column head you should always use a pre-column

The use of a pre-column increases the life time of a column dramatically In addition a pre-column can filter solid parts stemming from pump seals or injection rotors An alternative to a pre-column is an in-line filter These filters are attached directly to the column These filters get rid of solid parts in the eluent but will not avoid irreversible adsorption of organic impurities

Detector Systems

Virtually every chemical and physical

property that can be measured in

solution has been look at

Detectors fall roughly into two classes

Bulk property - measures an overall change in the mobile phase

Solute property - measures a solute specific property

Properties of a good detector

A detector must provide

bull high sensitivity low detection limits

bull linearity

bullReproducibility

This is true for any detector

Each detector will have specific advantages and will vary as to peak shape and spread noise and flowtemperature dependence they have

UVVis detector

A solute property detector

Sample must exhibit absorption in UVVis range Solvent must not absorb significantly at the measured wavelength

Types Filter photometer - single

bull Variable wavelength

bull Multiwavelength

Refractive index detector

Bulk property detector - general purpose

Based on refraction of light as it passes from

one media to another Presence of a solute

changes the refractive index of the solvent

Heat of absorption detector

A small amount of heat is released when a sample absorbs on a suitable surface

This detector can measure this

Electrochemical detectors

A number of properties have been evaluated

Detector types

Dielectic constant

Amperometric

Conductometric

Polarographic

Potentiometric

Dielectric constant detector

Bulk property detector

Measures changes in polarity of the liquid phase passing through the cell

Conductometric detector

Measures conductivity of the solvent Useful for

solutions of ions

Amperometric detectors

Most frequently applied type of electrochemical

detector

A known potential is applied across a set of

electrodes - typically a glassy carbon type

Ability to oxidize or reduce a species can be

measured

Typically limited to working with a specific class

of materials per analysis

Separation is based on the analytersquos relative solubility between two liquid phases

Stationary PhaseMobile Phase

Solvent Bonded Phase

Normal Phase- Polar stationary phase and non-polar solvent

bull Reverse Phase- Non-polar stationary phase and a polar solvent

Methan0l CH3OH

bull Acetonitrile CH3CN

bull Tetrahydrofuran

bull Water H2O

Solid Support - Backbone for bonded phases Usually 10micro 5micro or 3micro silica or polymeric

particles Bonded Phases - Functional groups firmly linked

(chemically bound) to the solid support Extremely stable Reproducible

Guard - Protects the analytical column Particles Interferences Prolongs the life of the analytical column

bull Analytical - Performs the separation

C-2 Ethyl Silyl -Si-CH2-CH3

bullCN Cyanopropyl Silyl -Si-(CH2)3-CN

bull C-18C-18 Octadecyl SilylOctadecyl Silyl -Si-(CH -Si-(CH22))1717--

CHCH33

bullC-8 Octyl Silyl -Si-(CH2)7-CH3

Pump

Injector

ColumnDetector

Mobile Phases

Gradient Controller

bull

UV Single wavelength (filter) [610 8330] Variable wavelength (monochromator)

[8316 8325] Multiple wavelengths (PDA) [555]

Fluorescence [610]Electrochemical [605] Mass Spectrometric [8325]

Restekreg ULTRA C-18 and CN Columns (250mm x 46mm 5micro)

Mobile Phase (11 MethanolWater) 15 mLmin

Supelcosil LC-PAH Columns A BConditions A 150mm x 46mm 5micro

Flow Rate 15 mLmin

Conditions B 50mm x 46mm 3micro

Flow Rate 30 mLmin

  • Slide 5
  • Slide 6
  • Slide 7
  • Slide 8
  • Slide 9
  • Slide 10
  • Slide 11
  • Slide 12
  • Slide 13
  • Slide 14
  • Slide 15
  • Slide 16
  • Slide 17
  • Slide 18
  • Slide 19
  • Slide 20
  • Slide 21
  • Slide 22
  • Slide 23
  • Slide 24
  • Slide 25
  • Slide 26
  • Slide 27
  • Slide 28
  • Slide 29
  • Slide 30
  • Slide 31
  • Slide 32
  • Slide 33
  • Slide 34
  • Slide 35
  • Slide 36
  • Slide 37
  • Slide 38
  • Slide 39
  • Slide 40
Page 14: P igh erformance L iquid C hromatography  HPLC is a form of liquid chromatography used to separate compounds that are dissolved in solution. HPLC instruments

Column body

Typically consist of stainless steel with a high

precision internal bore

Some manufacturers offer column inserts

- donrsquot need to repurchase the column

fittings

Others offer columns where the external

body can be compressed to improve packing

efficiency

HPLC column examples

Column stationary phases

Today most packing fall into four classes

Silica or alumina

Bound phases on either alumina or silica

Gels

Controlled-pore glass or silica

Absorption phases - alumina

common mobile phases

hexane chloroform 2-propanol

example application - amines

silica

common mobile phases

hexane chloroform 2-propanol

example applications - ethers esters

porphyrins fat-soluble vitamins

Partition phases

Can be broken down into

Normal phase - polar materials bound to the support

Reverse phase - non-polar materials bound to the support

Mixed phase - may have some of each

Partition phases

Normal

Amino (-NH2)

Cyano (-CN)

Diol (glycidoxy-ethylmethoxysilane)

Reverse

C-2 or RP-2 (-Si-CH2CH3)

C-8 or RP-8 (-Si-(CH2)7CH3)

C-18 or RP-18 (-Si-(CH2)17CH3)

Increasing the C number results in a thicker more retentive phase

Ion exchange phases

Strong cation - sulfonic acid group

Strong anionic - quarternary amine

Weak anion - primary amine

Weak cation - COOH

Size exclusion phases

Gels - organic or aqueous based

Controlled-pore - silica or glass

Must be selected based on pressure

requirements and size range

required for your application

Capillary and Microbore columns

Several companies have begun offering

columns with smaller ID

Microbore column - 1 mm ID packed

column

Capillary column - lt 1 mm ID internal

bound phase

These columns require smaller solvent flows

reduced sample size and improved detector

design

Capillary and Microbore columns Capillary and Microbore columns

Aromatic Compounds

mobile phase2 ethylacetate in

hexane

flow rate 4 μlmin

column Fusica II 300μm ID x

25 cm silica

sample

1 toluene

2 nitrobenzene

3 acetophenone

4 26-dinitrobenzene

injection 60 nl

detection UV 254 nm

Silica based columns

Silica is the ideal support for HPLC columns It offers a large mechanical stability excellent

physicochemical surface properties a wide range of bonding chemistries and is compatible

with a broad range of organic solvents

pH stability

In general HPLC columns are stable within a pH range of 2 to 8 If you are measuring a pH value the measurement must be done in the aqueous media before mixing the eluent with organic solvents

Modern HPLC columns can be used outside that pH range The new bonding chemistries allow use down to pH 1 for some stationary phases However please check vendorrsquos product information before using silica based column outside the pH range of 2 to 8 However best lifetimes are obtained between pH 20 and pH 68

Mechanical stability

Stationary phases based on silica are mechanically very stable The packed columns show no pressure limit and can be used at more than 40 MPa (6000 psi) without any problem However please avoid pressure shocks on the column Pressure shocks lead to channelling in the column which results in peak splitting in the corresponding chromatogram

Mobile phases (Eluents)

Silica based stationary phases are compatible with all organic solvents in the above mentioned pH range

bull Filter all prepared buffer through a 05 μm filter before using them in your HPLC system

The use of non pure solvents in HPLC causes irreversible adsorption of impurities on the column head These impurities block adsorption sites change the selectivity of the column and lead to peak splitting in the chromatogram In gradient elution impurities cause so called ldquoGhost Peaksrdquo

Ghost peaks are peaks that always appear in the same position on the chromatogram Their origin is not the sample but the impurities from the solvents or solvent additives Therefore it is highly recommended to run a gradientwithout injection in the beginning of each method to determine the ghost peaks

To avoid irreversible adsorption at the column head you should always use a pre-column

The use of a pre-column increases the life time of a column dramatically In addition a pre-column can filter solid parts stemming from pump seals or injection rotors An alternative to a pre-column is an in-line filter These filters are attached directly to the column These filters get rid of solid parts in the eluent but will not avoid irreversible adsorption of organic impurities

Detector Systems

Virtually every chemical and physical

property that can be measured in

solution has been look at

Detectors fall roughly into two classes

Bulk property - measures an overall change in the mobile phase

Solute property - measures a solute specific property

Properties of a good detector

A detector must provide

bull high sensitivity low detection limits

bull linearity

bullReproducibility

This is true for any detector

Each detector will have specific advantages and will vary as to peak shape and spread noise and flowtemperature dependence they have

UVVis detector

A solute property detector

Sample must exhibit absorption in UVVis range Solvent must not absorb significantly at the measured wavelength

Types Filter photometer - single

bull Variable wavelength

bull Multiwavelength

Refractive index detector

Bulk property detector - general purpose

Based on refraction of light as it passes from

one media to another Presence of a solute

changes the refractive index of the solvent

Heat of absorption detector

A small amount of heat is released when a sample absorbs on a suitable surface

This detector can measure this

Electrochemical detectors

A number of properties have been evaluated

Detector types

Dielectic constant

Amperometric

Conductometric

Polarographic

Potentiometric

Dielectric constant detector

Bulk property detector

Measures changes in polarity of the liquid phase passing through the cell

Conductometric detector

Measures conductivity of the solvent Useful for

solutions of ions

Amperometric detectors

Most frequently applied type of electrochemical

detector

A known potential is applied across a set of

electrodes - typically a glassy carbon type

Ability to oxidize or reduce a species can be

measured

Typically limited to working with a specific class

of materials per analysis

Separation is based on the analytersquos relative solubility between two liquid phases

Stationary PhaseMobile Phase

Solvent Bonded Phase

Normal Phase- Polar stationary phase and non-polar solvent

bull Reverse Phase- Non-polar stationary phase and a polar solvent

Methan0l CH3OH

bull Acetonitrile CH3CN

bull Tetrahydrofuran

bull Water H2O

Solid Support - Backbone for bonded phases Usually 10micro 5micro or 3micro silica or polymeric

particles Bonded Phases - Functional groups firmly linked

(chemically bound) to the solid support Extremely stable Reproducible

Guard - Protects the analytical column Particles Interferences Prolongs the life of the analytical column

bull Analytical - Performs the separation

C-2 Ethyl Silyl -Si-CH2-CH3

bullCN Cyanopropyl Silyl -Si-(CH2)3-CN

bull C-18C-18 Octadecyl SilylOctadecyl Silyl -Si-(CH -Si-(CH22))1717--

CHCH33

bullC-8 Octyl Silyl -Si-(CH2)7-CH3

Pump

Injector

ColumnDetector

Mobile Phases

Gradient Controller

bull

UV Single wavelength (filter) [610 8330] Variable wavelength (monochromator)

[8316 8325] Multiple wavelengths (PDA) [555]

Fluorescence [610]Electrochemical [605] Mass Spectrometric [8325]

Restekreg ULTRA C-18 and CN Columns (250mm x 46mm 5micro)

Mobile Phase (11 MethanolWater) 15 mLmin

Supelcosil LC-PAH Columns A BConditions A 150mm x 46mm 5micro

Flow Rate 15 mLmin

Conditions B 50mm x 46mm 3micro

Flow Rate 30 mLmin

  • Slide 5
  • Slide 6
  • Slide 7
  • Slide 8
  • Slide 9
  • Slide 10
  • Slide 11
  • Slide 12
  • Slide 13
  • Slide 14
  • Slide 15
  • Slide 16
  • Slide 17
  • Slide 18
  • Slide 19
  • Slide 20
  • Slide 21
  • Slide 22
  • Slide 23
  • Slide 24
  • Slide 25
  • Slide 26
  • Slide 27
  • Slide 28
  • Slide 29
  • Slide 30
  • Slide 31
  • Slide 32
  • Slide 33
  • Slide 34
  • Slide 35
  • Slide 36
  • Slide 37
  • Slide 38
  • Slide 39
  • Slide 40
Page 15: P igh erformance L iquid C hromatography  HPLC is a form of liquid chromatography used to separate compounds that are dissolved in solution. HPLC instruments

HPLC column examples

Column stationary phases

Today most packing fall into four classes

Silica or alumina

Bound phases on either alumina or silica

Gels

Controlled-pore glass or silica

Absorption phases - alumina

common mobile phases

hexane chloroform 2-propanol

example application - amines

silica

common mobile phases

hexane chloroform 2-propanol

example applications - ethers esters

porphyrins fat-soluble vitamins

Partition phases

Can be broken down into

Normal phase - polar materials bound to the support

Reverse phase - non-polar materials bound to the support

Mixed phase - may have some of each

Partition phases

Normal

Amino (-NH2)

Cyano (-CN)

Diol (glycidoxy-ethylmethoxysilane)

Reverse

C-2 or RP-2 (-Si-CH2CH3)

C-8 or RP-8 (-Si-(CH2)7CH3)

C-18 or RP-18 (-Si-(CH2)17CH3)

Increasing the C number results in a thicker more retentive phase

Ion exchange phases

Strong cation - sulfonic acid group

Strong anionic - quarternary amine

Weak anion - primary amine

Weak cation - COOH

Size exclusion phases

Gels - organic or aqueous based

Controlled-pore - silica or glass

Must be selected based on pressure

requirements and size range

required for your application

Capillary and Microbore columns

Several companies have begun offering

columns with smaller ID

Microbore column - 1 mm ID packed

column

Capillary column - lt 1 mm ID internal

bound phase

These columns require smaller solvent flows

reduced sample size and improved detector

design

Capillary and Microbore columns Capillary and Microbore columns

Aromatic Compounds

mobile phase2 ethylacetate in

hexane

flow rate 4 μlmin

column Fusica II 300μm ID x

25 cm silica

sample

1 toluene

2 nitrobenzene

3 acetophenone

4 26-dinitrobenzene

injection 60 nl

detection UV 254 nm

Silica based columns

Silica is the ideal support for HPLC columns It offers a large mechanical stability excellent

physicochemical surface properties a wide range of bonding chemistries and is compatible

with a broad range of organic solvents

pH stability

In general HPLC columns are stable within a pH range of 2 to 8 If you are measuring a pH value the measurement must be done in the aqueous media before mixing the eluent with organic solvents

Modern HPLC columns can be used outside that pH range The new bonding chemistries allow use down to pH 1 for some stationary phases However please check vendorrsquos product information before using silica based column outside the pH range of 2 to 8 However best lifetimes are obtained between pH 20 and pH 68

Mechanical stability

Stationary phases based on silica are mechanically very stable The packed columns show no pressure limit and can be used at more than 40 MPa (6000 psi) without any problem However please avoid pressure shocks on the column Pressure shocks lead to channelling in the column which results in peak splitting in the corresponding chromatogram

Mobile phases (Eluents)

Silica based stationary phases are compatible with all organic solvents in the above mentioned pH range

bull Filter all prepared buffer through a 05 μm filter before using them in your HPLC system

The use of non pure solvents in HPLC causes irreversible adsorption of impurities on the column head These impurities block adsorption sites change the selectivity of the column and lead to peak splitting in the chromatogram In gradient elution impurities cause so called ldquoGhost Peaksrdquo

Ghost peaks are peaks that always appear in the same position on the chromatogram Their origin is not the sample but the impurities from the solvents or solvent additives Therefore it is highly recommended to run a gradientwithout injection in the beginning of each method to determine the ghost peaks

To avoid irreversible adsorption at the column head you should always use a pre-column

The use of a pre-column increases the life time of a column dramatically In addition a pre-column can filter solid parts stemming from pump seals or injection rotors An alternative to a pre-column is an in-line filter These filters are attached directly to the column These filters get rid of solid parts in the eluent but will not avoid irreversible adsorption of organic impurities

Detector Systems

Virtually every chemical and physical

property that can be measured in

solution has been look at

Detectors fall roughly into two classes

Bulk property - measures an overall change in the mobile phase

Solute property - measures a solute specific property

Properties of a good detector

A detector must provide

bull high sensitivity low detection limits

bull linearity

bullReproducibility

This is true for any detector

Each detector will have specific advantages and will vary as to peak shape and spread noise and flowtemperature dependence they have

UVVis detector

A solute property detector

Sample must exhibit absorption in UVVis range Solvent must not absorb significantly at the measured wavelength

Types Filter photometer - single

bull Variable wavelength

bull Multiwavelength

Refractive index detector

Bulk property detector - general purpose

Based on refraction of light as it passes from

one media to another Presence of a solute

changes the refractive index of the solvent

Heat of absorption detector

A small amount of heat is released when a sample absorbs on a suitable surface

This detector can measure this

Electrochemical detectors

A number of properties have been evaluated

Detector types

Dielectic constant

Amperometric

Conductometric

Polarographic

Potentiometric

Dielectric constant detector

Bulk property detector

Measures changes in polarity of the liquid phase passing through the cell

Conductometric detector

Measures conductivity of the solvent Useful for

solutions of ions

Amperometric detectors

Most frequently applied type of electrochemical

detector

A known potential is applied across a set of

electrodes - typically a glassy carbon type

Ability to oxidize or reduce a species can be

measured

Typically limited to working with a specific class

of materials per analysis

Separation is based on the analytersquos relative solubility between two liquid phases

Stationary PhaseMobile Phase

Solvent Bonded Phase

Normal Phase- Polar stationary phase and non-polar solvent

bull Reverse Phase- Non-polar stationary phase and a polar solvent

Methan0l CH3OH

bull Acetonitrile CH3CN

bull Tetrahydrofuran

bull Water H2O

Solid Support - Backbone for bonded phases Usually 10micro 5micro or 3micro silica or polymeric

particles Bonded Phases - Functional groups firmly linked

(chemically bound) to the solid support Extremely stable Reproducible

Guard - Protects the analytical column Particles Interferences Prolongs the life of the analytical column

bull Analytical - Performs the separation

C-2 Ethyl Silyl -Si-CH2-CH3

bullCN Cyanopropyl Silyl -Si-(CH2)3-CN

bull C-18C-18 Octadecyl SilylOctadecyl Silyl -Si-(CH -Si-(CH22))1717--

CHCH33

bullC-8 Octyl Silyl -Si-(CH2)7-CH3

Pump

Injector

ColumnDetector

Mobile Phases

Gradient Controller

bull

UV Single wavelength (filter) [610 8330] Variable wavelength (monochromator)

[8316 8325] Multiple wavelengths (PDA) [555]

Fluorescence [610]Electrochemical [605] Mass Spectrometric [8325]

Restekreg ULTRA C-18 and CN Columns (250mm x 46mm 5micro)

Mobile Phase (11 MethanolWater) 15 mLmin

Supelcosil LC-PAH Columns A BConditions A 150mm x 46mm 5micro

Flow Rate 15 mLmin

Conditions B 50mm x 46mm 3micro

Flow Rate 30 mLmin

  • Slide 5
  • Slide 6
  • Slide 7
  • Slide 8
  • Slide 9
  • Slide 10
  • Slide 11
  • Slide 12
  • Slide 13
  • Slide 14
  • Slide 15
  • Slide 16
  • Slide 17
  • Slide 18
  • Slide 19
  • Slide 20
  • Slide 21
  • Slide 22
  • Slide 23
  • Slide 24
  • Slide 25
  • Slide 26
  • Slide 27
  • Slide 28
  • Slide 29
  • Slide 30
  • Slide 31
  • Slide 32
  • Slide 33
  • Slide 34
  • Slide 35
  • Slide 36
  • Slide 37
  • Slide 38
  • Slide 39
  • Slide 40
Page 16: P igh erformance L iquid C hromatography  HPLC is a form of liquid chromatography used to separate compounds that are dissolved in solution. HPLC instruments

Column stationary phases

Today most packing fall into four classes

Silica or alumina

Bound phases on either alumina or silica

Gels

Controlled-pore glass or silica

Absorption phases - alumina

common mobile phases

hexane chloroform 2-propanol

example application - amines

silica

common mobile phases

hexane chloroform 2-propanol

example applications - ethers esters

porphyrins fat-soluble vitamins

Partition phases

Can be broken down into

Normal phase - polar materials bound to the support

Reverse phase - non-polar materials bound to the support

Mixed phase - may have some of each

Partition phases

Normal

Amino (-NH2)

Cyano (-CN)

Diol (glycidoxy-ethylmethoxysilane)

Reverse

C-2 or RP-2 (-Si-CH2CH3)

C-8 or RP-8 (-Si-(CH2)7CH3)

C-18 or RP-18 (-Si-(CH2)17CH3)

Increasing the C number results in a thicker more retentive phase

Ion exchange phases

Strong cation - sulfonic acid group

Strong anionic - quarternary amine

Weak anion - primary amine

Weak cation - COOH

Size exclusion phases

Gels - organic or aqueous based

Controlled-pore - silica or glass

Must be selected based on pressure

requirements and size range

required for your application

Capillary and Microbore columns

Several companies have begun offering

columns with smaller ID

Microbore column - 1 mm ID packed

column

Capillary column - lt 1 mm ID internal

bound phase

These columns require smaller solvent flows

reduced sample size and improved detector

design

Capillary and Microbore columns Capillary and Microbore columns

Aromatic Compounds

mobile phase2 ethylacetate in

hexane

flow rate 4 μlmin

column Fusica II 300μm ID x

25 cm silica

sample

1 toluene

2 nitrobenzene

3 acetophenone

4 26-dinitrobenzene

injection 60 nl

detection UV 254 nm

Silica based columns

Silica is the ideal support for HPLC columns It offers a large mechanical stability excellent

physicochemical surface properties a wide range of bonding chemistries and is compatible

with a broad range of organic solvents

pH stability

In general HPLC columns are stable within a pH range of 2 to 8 If you are measuring a pH value the measurement must be done in the aqueous media before mixing the eluent with organic solvents

Modern HPLC columns can be used outside that pH range The new bonding chemistries allow use down to pH 1 for some stationary phases However please check vendorrsquos product information before using silica based column outside the pH range of 2 to 8 However best lifetimes are obtained between pH 20 and pH 68

Mechanical stability

Stationary phases based on silica are mechanically very stable The packed columns show no pressure limit and can be used at more than 40 MPa (6000 psi) without any problem However please avoid pressure shocks on the column Pressure shocks lead to channelling in the column which results in peak splitting in the corresponding chromatogram

Mobile phases (Eluents)

Silica based stationary phases are compatible with all organic solvents in the above mentioned pH range

bull Filter all prepared buffer through a 05 μm filter before using them in your HPLC system

The use of non pure solvents in HPLC causes irreversible adsorption of impurities on the column head These impurities block adsorption sites change the selectivity of the column and lead to peak splitting in the chromatogram In gradient elution impurities cause so called ldquoGhost Peaksrdquo

Ghost peaks are peaks that always appear in the same position on the chromatogram Their origin is not the sample but the impurities from the solvents or solvent additives Therefore it is highly recommended to run a gradientwithout injection in the beginning of each method to determine the ghost peaks

To avoid irreversible adsorption at the column head you should always use a pre-column

The use of a pre-column increases the life time of a column dramatically In addition a pre-column can filter solid parts stemming from pump seals or injection rotors An alternative to a pre-column is an in-line filter These filters are attached directly to the column These filters get rid of solid parts in the eluent but will not avoid irreversible adsorption of organic impurities

Detector Systems

Virtually every chemical and physical

property that can be measured in

solution has been look at

Detectors fall roughly into two classes

Bulk property - measures an overall change in the mobile phase

Solute property - measures a solute specific property

Properties of a good detector

A detector must provide

bull high sensitivity low detection limits

bull linearity

bullReproducibility

This is true for any detector

Each detector will have specific advantages and will vary as to peak shape and spread noise and flowtemperature dependence they have

UVVis detector

A solute property detector

Sample must exhibit absorption in UVVis range Solvent must not absorb significantly at the measured wavelength

Types Filter photometer - single

bull Variable wavelength

bull Multiwavelength

Refractive index detector

Bulk property detector - general purpose

Based on refraction of light as it passes from

one media to another Presence of a solute

changes the refractive index of the solvent

Heat of absorption detector

A small amount of heat is released when a sample absorbs on a suitable surface

This detector can measure this

Electrochemical detectors

A number of properties have been evaluated

Detector types

Dielectic constant

Amperometric

Conductometric

Polarographic

Potentiometric

Dielectric constant detector

Bulk property detector

Measures changes in polarity of the liquid phase passing through the cell

Conductometric detector

Measures conductivity of the solvent Useful for

solutions of ions

Amperometric detectors

Most frequently applied type of electrochemical

detector

A known potential is applied across a set of

electrodes - typically a glassy carbon type

Ability to oxidize or reduce a species can be

measured

Typically limited to working with a specific class

of materials per analysis

Separation is based on the analytersquos relative solubility between two liquid phases

Stationary PhaseMobile Phase

Solvent Bonded Phase

Normal Phase- Polar stationary phase and non-polar solvent

bull Reverse Phase- Non-polar stationary phase and a polar solvent

Methan0l CH3OH

bull Acetonitrile CH3CN

bull Tetrahydrofuran

bull Water H2O

Solid Support - Backbone for bonded phases Usually 10micro 5micro or 3micro silica or polymeric

particles Bonded Phases - Functional groups firmly linked

(chemically bound) to the solid support Extremely stable Reproducible

Guard - Protects the analytical column Particles Interferences Prolongs the life of the analytical column

bull Analytical - Performs the separation

C-2 Ethyl Silyl -Si-CH2-CH3

bullCN Cyanopropyl Silyl -Si-(CH2)3-CN

bull C-18C-18 Octadecyl SilylOctadecyl Silyl -Si-(CH -Si-(CH22))1717--

CHCH33

bullC-8 Octyl Silyl -Si-(CH2)7-CH3

Pump

Injector

ColumnDetector

Mobile Phases

Gradient Controller

bull

UV Single wavelength (filter) [610 8330] Variable wavelength (monochromator)

[8316 8325] Multiple wavelengths (PDA) [555]

Fluorescence [610]Electrochemical [605] Mass Spectrometric [8325]

Restekreg ULTRA C-18 and CN Columns (250mm x 46mm 5micro)

Mobile Phase (11 MethanolWater) 15 mLmin

Supelcosil LC-PAH Columns A BConditions A 150mm x 46mm 5micro

Flow Rate 15 mLmin

Conditions B 50mm x 46mm 3micro

Flow Rate 30 mLmin

  • Slide 5
  • Slide 6
  • Slide 7
  • Slide 8
  • Slide 9
  • Slide 10
  • Slide 11
  • Slide 12
  • Slide 13
  • Slide 14
  • Slide 15
  • Slide 16
  • Slide 17
  • Slide 18
  • Slide 19
  • Slide 20
  • Slide 21
  • Slide 22
  • Slide 23
  • Slide 24
  • Slide 25
  • Slide 26
  • Slide 27
  • Slide 28
  • Slide 29
  • Slide 30
  • Slide 31
  • Slide 32
  • Slide 33
  • Slide 34
  • Slide 35
  • Slide 36
  • Slide 37
  • Slide 38
  • Slide 39
  • Slide 40
Page 17: P igh erformance L iquid C hromatography  HPLC is a form of liquid chromatography used to separate compounds that are dissolved in solution. HPLC instruments

Absorption phases - alumina

common mobile phases

hexane chloroform 2-propanol

example application - amines

silica

common mobile phases

hexane chloroform 2-propanol

example applications - ethers esters

porphyrins fat-soluble vitamins

Partition phases

Can be broken down into

Normal phase - polar materials bound to the support

Reverse phase - non-polar materials bound to the support

Mixed phase - may have some of each

Partition phases

Normal

Amino (-NH2)

Cyano (-CN)

Diol (glycidoxy-ethylmethoxysilane)

Reverse

C-2 or RP-2 (-Si-CH2CH3)

C-8 or RP-8 (-Si-(CH2)7CH3)

C-18 or RP-18 (-Si-(CH2)17CH3)

Increasing the C number results in a thicker more retentive phase

Ion exchange phases

Strong cation - sulfonic acid group

Strong anionic - quarternary amine

Weak anion - primary amine

Weak cation - COOH

Size exclusion phases

Gels - organic or aqueous based

Controlled-pore - silica or glass

Must be selected based on pressure

requirements and size range

required for your application

Capillary and Microbore columns

Several companies have begun offering

columns with smaller ID

Microbore column - 1 mm ID packed

column

Capillary column - lt 1 mm ID internal

bound phase

These columns require smaller solvent flows

reduced sample size and improved detector

design

Capillary and Microbore columns Capillary and Microbore columns

Aromatic Compounds

mobile phase2 ethylacetate in

hexane

flow rate 4 μlmin

column Fusica II 300μm ID x

25 cm silica

sample

1 toluene

2 nitrobenzene

3 acetophenone

4 26-dinitrobenzene

injection 60 nl

detection UV 254 nm

Silica based columns

Silica is the ideal support for HPLC columns It offers a large mechanical stability excellent

physicochemical surface properties a wide range of bonding chemistries and is compatible

with a broad range of organic solvents

pH stability

In general HPLC columns are stable within a pH range of 2 to 8 If you are measuring a pH value the measurement must be done in the aqueous media before mixing the eluent with organic solvents

Modern HPLC columns can be used outside that pH range The new bonding chemistries allow use down to pH 1 for some stationary phases However please check vendorrsquos product information before using silica based column outside the pH range of 2 to 8 However best lifetimes are obtained between pH 20 and pH 68

Mechanical stability

Stationary phases based on silica are mechanically very stable The packed columns show no pressure limit and can be used at more than 40 MPa (6000 psi) without any problem However please avoid pressure shocks on the column Pressure shocks lead to channelling in the column which results in peak splitting in the corresponding chromatogram

Mobile phases (Eluents)

Silica based stationary phases are compatible with all organic solvents in the above mentioned pH range

bull Filter all prepared buffer through a 05 μm filter before using them in your HPLC system

The use of non pure solvents in HPLC causes irreversible adsorption of impurities on the column head These impurities block adsorption sites change the selectivity of the column and lead to peak splitting in the chromatogram In gradient elution impurities cause so called ldquoGhost Peaksrdquo

Ghost peaks are peaks that always appear in the same position on the chromatogram Their origin is not the sample but the impurities from the solvents or solvent additives Therefore it is highly recommended to run a gradientwithout injection in the beginning of each method to determine the ghost peaks

To avoid irreversible adsorption at the column head you should always use a pre-column

The use of a pre-column increases the life time of a column dramatically In addition a pre-column can filter solid parts stemming from pump seals or injection rotors An alternative to a pre-column is an in-line filter These filters are attached directly to the column These filters get rid of solid parts in the eluent but will not avoid irreversible adsorption of organic impurities

Detector Systems

Virtually every chemical and physical

property that can be measured in

solution has been look at

Detectors fall roughly into two classes

Bulk property - measures an overall change in the mobile phase

Solute property - measures a solute specific property

Properties of a good detector

A detector must provide

bull high sensitivity low detection limits

bull linearity

bullReproducibility

This is true for any detector

Each detector will have specific advantages and will vary as to peak shape and spread noise and flowtemperature dependence they have

UVVis detector

A solute property detector

Sample must exhibit absorption in UVVis range Solvent must not absorb significantly at the measured wavelength

Types Filter photometer - single

bull Variable wavelength

bull Multiwavelength

Refractive index detector

Bulk property detector - general purpose

Based on refraction of light as it passes from

one media to another Presence of a solute

changes the refractive index of the solvent

Heat of absorption detector

A small amount of heat is released when a sample absorbs on a suitable surface

This detector can measure this

Electrochemical detectors

A number of properties have been evaluated

Detector types

Dielectic constant

Amperometric

Conductometric

Polarographic

Potentiometric

Dielectric constant detector

Bulk property detector

Measures changes in polarity of the liquid phase passing through the cell

Conductometric detector

Measures conductivity of the solvent Useful for

solutions of ions

Amperometric detectors

Most frequently applied type of electrochemical

detector

A known potential is applied across a set of

electrodes - typically a glassy carbon type

Ability to oxidize or reduce a species can be

measured

Typically limited to working with a specific class

of materials per analysis

Separation is based on the analytersquos relative solubility between two liquid phases

Stationary PhaseMobile Phase

Solvent Bonded Phase

Normal Phase- Polar stationary phase and non-polar solvent

bull Reverse Phase- Non-polar stationary phase and a polar solvent

Methan0l CH3OH

bull Acetonitrile CH3CN

bull Tetrahydrofuran

bull Water H2O

Solid Support - Backbone for bonded phases Usually 10micro 5micro or 3micro silica or polymeric

particles Bonded Phases - Functional groups firmly linked

(chemically bound) to the solid support Extremely stable Reproducible

Guard - Protects the analytical column Particles Interferences Prolongs the life of the analytical column

bull Analytical - Performs the separation

C-2 Ethyl Silyl -Si-CH2-CH3

bullCN Cyanopropyl Silyl -Si-(CH2)3-CN

bull C-18C-18 Octadecyl SilylOctadecyl Silyl -Si-(CH -Si-(CH22))1717--

CHCH33

bullC-8 Octyl Silyl -Si-(CH2)7-CH3

Pump

Injector

ColumnDetector

Mobile Phases

Gradient Controller

bull

UV Single wavelength (filter) [610 8330] Variable wavelength (monochromator)

[8316 8325] Multiple wavelengths (PDA) [555]

Fluorescence [610]Electrochemical [605] Mass Spectrometric [8325]

Restekreg ULTRA C-18 and CN Columns (250mm x 46mm 5micro)

Mobile Phase (11 MethanolWater) 15 mLmin

Supelcosil LC-PAH Columns A BConditions A 150mm x 46mm 5micro

Flow Rate 15 mLmin

Conditions B 50mm x 46mm 3micro

Flow Rate 30 mLmin

  • Slide 5
  • Slide 6
  • Slide 7
  • Slide 8
  • Slide 9
  • Slide 10
  • Slide 11
  • Slide 12
  • Slide 13
  • Slide 14
  • Slide 15
  • Slide 16
  • Slide 17
  • Slide 18
  • Slide 19
  • Slide 20
  • Slide 21
  • Slide 22
  • Slide 23
  • Slide 24
  • Slide 25
  • Slide 26
  • Slide 27
  • Slide 28
  • Slide 29
  • Slide 30
  • Slide 31
  • Slide 32
  • Slide 33
  • Slide 34
  • Slide 35
  • Slide 36
  • Slide 37
  • Slide 38
  • Slide 39
  • Slide 40
Page 18: P igh erformance L iquid C hromatography  HPLC is a form of liquid chromatography used to separate compounds that are dissolved in solution. HPLC instruments

Partition phases

Can be broken down into

Normal phase - polar materials bound to the support

Reverse phase - non-polar materials bound to the support

Mixed phase - may have some of each

Partition phases

Normal

Amino (-NH2)

Cyano (-CN)

Diol (glycidoxy-ethylmethoxysilane)

Reverse

C-2 or RP-2 (-Si-CH2CH3)

C-8 or RP-8 (-Si-(CH2)7CH3)

C-18 or RP-18 (-Si-(CH2)17CH3)

Increasing the C number results in a thicker more retentive phase

Ion exchange phases

Strong cation - sulfonic acid group

Strong anionic - quarternary amine

Weak anion - primary amine

Weak cation - COOH

Size exclusion phases

Gels - organic or aqueous based

Controlled-pore - silica or glass

Must be selected based on pressure

requirements and size range

required for your application

Capillary and Microbore columns

Several companies have begun offering

columns with smaller ID

Microbore column - 1 mm ID packed

column

Capillary column - lt 1 mm ID internal

bound phase

These columns require smaller solvent flows

reduced sample size and improved detector

design

Capillary and Microbore columns Capillary and Microbore columns

Aromatic Compounds

mobile phase2 ethylacetate in

hexane

flow rate 4 μlmin

column Fusica II 300μm ID x

25 cm silica

sample

1 toluene

2 nitrobenzene

3 acetophenone

4 26-dinitrobenzene

injection 60 nl

detection UV 254 nm

Silica based columns

Silica is the ideal support for HPLC columns It offers a large mechanical stability excellent

physicochemical surface properties a wide range of bonding chemistries and is compatible

with a broad range of organic solvents

pH stability

In general HPLC columns are stable within a pH range of 2 to 8 If you are measuring a pH value the measurement must be done in the aqueous media before mixing the eluent with organic solvents

Modern HPLC columns can be used outside that pH range The new bonding chemistries allow use down to pH 1 for some stationary phases However please check vendorrsquos product information before using silica based column outside the pH range of 2 to 8 However best lifetimes are obtained between pH 20 and pH 68

Mechanical stability

Stationary phases based on silica are mechanically very stable The packed columns show no pressure limit and can be used at more than 40 MPa (6000 psi) without any problem However please avoid pressure shocks on the column Pressure shocks lead to channelling in the column which results in peak splitting in the corresponding chromatogram

Mobile phases (Eluents)

Silica based stationary phases are compatible with all organic solvents in the above mentioned pH range

bull Filter all prepared buffer through a 05 μm filter before using them in your HPLC system

The use of non pure solvents in HPLC causes irreversible adsorption of impurities on the column head These impurities block adsorption sites change the selectivity of the column and lead to peak splitting in the chromatogram In gradient elution impurities cause so called ldquoGhost Peaksrdquo

Ghost peaks are peaks that always appear in the same position on the chromatogram Their origin is not the sample but the impurities from the solvents or solvent additives Therefore it is highly recommended to run a gradientwithout injection in the beginning of each method to determine the ghost peaks

To avoid irreversible adsorption at the column head you should always use a pre-column

The use of a pre-column increases the life time of a column dramatically In addition a pre-column can filter solid parts stemming from pump seals or injection rotors An alternative to a pre-column is an in-line filter These filters are attached directly to the column These filters get rid of solid parts in the eluent but will not avoid irreversible adsorption of organic impurities

Detector Systems

Virtually every chemical and physical

property that can be measured in

solution has been look at

Detectors fall roughly into two classes

Bulk property - measures an overall change in the mobile phase

Solute property - measures a solute specific property

Properties of a good detector

A detector must provide

bull high sensitivity low detection limits

bull linearity

bullReproducibility

This is true for any detector

Each detector will have specific advantages and will vary as to peak shape and spread noise and flowtemperature dependence they have

UVVis detector

A solute property detector

Sample must exhibit absorption in UVVis range Solvent must not absorb significantly at the measured wavelength

Types Filter photometer - single

bull Variable wavelength

bull Multiwavelength

Refractive index detector

Bulk property detector - general purpose

Based on refraction of light as it passes from

one media to another Presence of a solute

changes the refractive index of the solvent

Heat of absorption detector

A small amount of heat is released when a sample absorbs on a suitable surface

This detector can measure this

Electrochemical detectors

A number of properties have been evaluated

Detector types

Dielectic constant

Amperometric

Conductometric

Polarographic

Potentiometric

Dielectric constant detector

Bulk property detector

Measures changes in polarity of the liquid phase passing through the cell

Conductometric detector

Measures conductivity of the solvent Useful for

solutions of ions

Amperometric detectors

Most frequently applied type of electrochemical

detector

A known potential is applied across a set of

electrodes - typically a glassy carbon type

Ability to oxidize or reduce a species can be

measured

Typically limited to working with a specific class

of materials per analysis

Separation is based on the analytersquos relative solubility between two liquid phases

Stationary PhaseMobile Phase

Solvent Bonded Phase

Normal Phase- Polar stationary phase and non-polar solvent

bull Reverse Phase- Non-polar stationary phase and a polar solvent

Methan0l CH3OH

bull Acetonitrile CH3CN

bull Tetrahydrofuran

bull Water H2O

Solid Support - Backbone for bonded phases Usually 10micro 5micro or 3micro silica or polymeric

particles Bonded Phases - Functional groups firmly linked

(chemically bound) to the solid support Extremely stable Reproducible

Guard - Protects the analytical column Particles Interferences Prolongs the life of the analytical column

bull Analytical - Performs the separation

C-2 Ethyl Silyl -Si-CH2-CH3

bullCN Cyanopropyl Silyl -Si-(CH2)3-CN

bull C-18C-18 Octadecyl SilylOctadecyl Silyl -Si-(CH -Si-(CH22))1717--

CHCH33

bullC-8 Octyl Silyl -Si-(CH2)7-CH3

Pump

Injector

ColumnDetector

Mobile Phases

Gradient Controller

bull

UV Single wavelength (filter) [610 8330] Variable wavelength (monochromator)

[8316 8325] Multiple wavelengths (PDA) [555]

Fluorescence [610]Electrochemical [605] Mass Spectrometric [8325]

Restekreg ULTRA C-18 and CN Columns (250mm x 46mm 5micro)

Mobile Phase (11 MethanolWater) 15 mLmin

Supelcosil LC-PAH Columns A BConditions A 150mm x 46mm 5micro

Flow Rate 15 mLmin

Conditions B 50mm x 46mm 3micro

Flow Rate 30 mLmin

  • Slide 5
  • Slide 6
  • Slide 7
  • Slide 8
  • Slide 9
  • Slide 10
  • Slide 11
  • Slide 12
  • Slide 13
  • Slide 14
  • Slide 15
  • Slide 16
  • Slide 17
  • Slide 18
  • Slide 19
  • Slide 20
  • Slide 21
  • Slide 22
  • Slide 23
  • Slide 24
  • Slide 25
  • Slide 26
  • Slide 27
  • Slide 28
  • Slide 29
  • Slide 30
  • Slide 31
  • Slide 32
  • Slide 33
  • Slide 34
  • Slide 35
  • Slide 36
  • Slide 37
  • Slide 38
  • Slide 39
  • Slide 40
Page 19: P igh erformance L iquid C hromatography  HPLC is a form of liquid chromatography used to separate compounds that are dissolved in solution. HPLC instruments

Partition phases

Normal

Amino (-NH2)

Cyano (-CN)

Diol (glycidoxy-ethylmethoxysilane)

Reverse

C-2 or RP-2 (-Si-CH2CH3)

C-8 or RP-8 (-Si-(CH2)7CH3)

C-18 or RP-18 (-Si-(CH2)17CH3)

Increasing the C number results in a thicker more retentive phase

Ion exchange phases

Strong cation - sulfonic acid group

Strong anionic - quarternary amine

Weak anion - primary amine

Weak cation - COOH

Size exclusion phases

Gels - organic or aqueous based

Controlled-pore - silica or glass

Must be selected based on pressure

requirements and size range

required for your application

Capillary and Microbore columns

Several companies have begun offering

columns with smaller ID

Microbore column - 1 mm ID packed

column

Capillary column - lt 1 mm ID internal

bound phase

These columns require smaller solvent flows

reduced sample size and improved detector

design

Capillary and Microbore columns Capillary and Microbore columns

Aromatic Compounds

mobile phase2 ethylacetate in

hexane

flow rate 4 μlmin

column Fusica II 300μm ID x

25 cm silica

sample

1 toluene

2 nitrobenzene

3 acetophenone

4 26-dinitrobenzene

injection 60 nl

detection UV 254 nm

Silica based columns

Silica is the ideal support for HPLC columns It offers a large mechanical stability excellent

physicochemical surface properties a wide range of bonding chemistries and is compatible

with a broad range of organic solvents

pH stability

In general HPLC columns are stable within a pH range of 2 to 8 If you are measuring a pH value the measurement must be done in the aqueous media before mixing the eluent with organic solvents

Modern HPLC columns can be used outside that pH range The new bonding chemistries allow use down to pH 1 for some stationary phases However please check vendorrsquos product information before using silica based column outside the pH range of 2 to 8 However best lifetimes are obtained between pH 20 and pH 68

Mechanical stability

Stationary phases based on silica are mechanically very stable The packed columns show no pressure limit and can be used at more than 40 MPa (6000 psi) without any problem However please avoid pressure shocks on the column Pressure shocks lead to channelling in the column which results in peak splitting in the corresponding chromatogram

Mobile phases (Eluents)

Silica based stationary phases are compatible with all organic solvents in the above mentioned pH range

bull Filter all prepared buffer through a 05 μm filter before using them in your HPLC system

The use of non pure solvents in HPLC causes irreversible adsorption of impurities on the column head These impurities block adsorption sites change the selectivity of the column and lead to peak splitting in the chromatogram In gradient elution impurities cause so called ldquoGhost Peaksrdquo

Ghost peaks are peaks that always appear in the same position on the chromatogram Their origin is not the sample but the impurities from the solvents or solvent additives Therefore it is highly recommended to run a gradientwithout injection in the beginning of each method to determine the ghost peaks

To avoid irreversible adsorption at the column head you should always use a pre-column

The use of a pre-column increases the life time of a column dramatically In addition a pre-column can filter solid parts stemming from pump seals or injection rotors An alternative to a pre-column is an in-line filter These filters are attached directly to the column These filters get rid of solid parts in the eluent but will not avoid irreversible adsorption of organic impurities

Detector Systems

Virtually every chemical and physical

property that can be measured in

solution has been look at

Detectors fall roughly into two classes

Bulk property - measures an overall change in the mobile phase

Solute property - measures a solute specific property

Properties of a good detector

A detector must provide

bull high sensitivity low detection limits

bull linearity

bullReproducibility

This is true for any detector

Each detector will have specific advantages and will vary as to peak shape and spread noise and flowtemperature dependence they have

UVVis detector

A solute property detector

Sample must exhibit absorption in UVVis range Solvent must not absorb significantly at the measured wavelength

Types Filter photometer - single

bull Variable wavelength

bull Multiwavelength

Refractive index detector

Bulk property detector - general purpose

Based on refraction of light as it passes from

one media to another Presence of a solute

changes the refractive index of the solvent

Heat of absorption detector

A small amount of heat is released when a sample absorbs on a suitable surface

This detector can measure this

Electrochemical detectors

A number of properties have been evaluated

Detector types

Dielectic constant

Amperometric

Conductometric

Polarographic

Potentiometric

Dielectric constant detector

Bulk property detector

Measures changes in polarity of the liquid phase passing through the cell

Conductometric detector

Measures conductivity of the solvent Useful for

solutions of ions

Amperometric detectors

Most frequently applied type of electrochemical

detector

A known potential is applied across a set of

electrodes - typically a glassy carbon type

Ability to oxidize or reduce a species can be

measured

Typically limited to working with a specific class

of materials per analysis

Separation is based on the analytersquos relative solubility between two liquid phases

Stationary PhaseMobile Phase

Solvent Bonded Phase

Normal Phase- Polar stationary phase and non-polar solvent

bull Reverse Phase- Non-polar stationary phase and a polar solvent

Methan0l CH3OH

bull Acetonitrile CH3CN

bull Tetrahydrofuran

bull Water H2O

Solid Support - Backbone for bonded phases Usually 10micro 5micro or 3micro silica or polymeric

particles Bonded Phases - Functional groups firmly linked

(chemically bound) to the solid support Extremely stable Reproducible

Guard - Protects the analytical column Particles Interferences Prolongs the life of the analytical column

bull Analytical - Performs the separation

C-2 Ethyl Silyl -Si-CH2-CH3

bullCN Cyanopropyl Silyl -Si-(CH2)3-CN

bull C-18C-18 Octadecyl SilylOctadecyl Silyl -Si-(CH -Si-(CH22))1717--

CHCH33

bullC-8 Octyl Silyl -Si-(CH2)7-CH3

Pump

Injector

ColumnDetector

Mobile Phases

Gradient Controller

bull

UV Single wavelength (filter) [610 8330] Variable wavelength (monochromator)

[8316 8325] Multiple wavelengths (PDA) [555]

Fluorescence [610]Electrochemical [605] Mass Spectrometric [8325]

Restekreg ULTRA C-18 and CN Columns (250mm x 46mm 5micro)

Mobile Phase (11 MethanolWater) 15 mLmin

Supelcosil LC-PAH Columns A BConditions A 150mm x 46mm 5micro

Flow Rate 15 mLmin

Conditions B 50mm x 46mm 3micro

Flow Rate 30 mLmin

  • Slide 5
  • Slide 6
  • Slide 7
  • Slide 8
  • Slide 9
  • Slide 10
  • Slide 11
  • Slide 12
  • Slide 13
  • Slide 14
  • Slide 15
  • Slide 16
  • Slide 17
  • Slide 18
  • Slide 19
  • Slide 20
  • Slide 21
  • Slide 22
  • Slide 23
  • Slide 24
  • Slide 25
  • Slide 26
  • Slide 27
  • Slide 28
  • Slide 29
  • Slide 30
  • Slide 31
  • Slide 32
  • Slide 33
  • Slide 34
  • Slide 35
  • Slide 36
  • Slide 37
  • Slide 38
  • Slide 39
  • Slide 40
Page 20: P igh erformance L iquid C hromatography  HPLC is a form of liquid chromatography used to separate compounds that are dissolved in solution. HPLC instruments

Ion exchange phases

Strong cation - sulfonic acid group

Strong anionic - quarternary amine

Weak anion - primary amine

Weak cation - COOH

Size exclusion phases

Gels - organic or aqueous based

Controlled-pore - silica or glass

Must be selected based on pressure

requirements and size range

required for your application

Capillary and Microbore columns

Several companies have begun offering

columns with smaller ID

Microbore column - 1 mm ID packed

column

Capillary column - lt 1 mm ID internal

bound phase

These columns require smaller solvent flows

reduced sample size and improved detector

design

Capillary and Microbore columns Capillary and Microbore columns

Aromatic Compounds

mobile phase2 ethylacetate in

hexane

flow rate 4 μlmin

column Fusica II 300μm ID x

25 cm silica

sample

1 toluene

2 nitrobenzene

3 acetophenone

4 26-dinitrobenzene

injection 60 nl

detection UV 254 nm

Silica based columns

Silica is the ideal support for HPLC columns It offers a large mechanical stability excellent

physicochemical surface properties a wide range of bonding chemistries and is compatible

with a broad range of organic solvents

pH stability

In general HPLC columns are stable within a pH range of 2 to 8 If you are measuring a pH value the measurement must be done in the aqueous media before mixing the eluent with organic solvents

Modern HPLC columns can be used outside that pH range The new bonding chemistries allow use down to pH 1 for some stationary phases However please check vendorrsquos product information before using silica based column outside the pH range of 2 to 8 However best lifetimes are obtained between pH 20 and pH 68

Mechanical stability

Stationary phases based on silica are mechanically very stable The packed columns show no pressure limit and can be used at more than 40 MPa (6000 psi) without any problem However please avoid pressure shocks on the column Pressure shocks lead to channelling in the column which results in peak splitting in the corresponding chromatogram

Mobile phases (Eluents)

Silica based stationary phases are compatible with all organic solvents in the above mentioned pH range

bull Filter all prepared buffer through a 05 μm filter before using them in your HPLC system

The use of non pure solvents in HPLC causes irreversible adsorption of impurities on the column head These impurities block adsorption sites change the selectivity of the column and lead to peak splitting in the chromatogram In gradient elution impurities cause so called ldquoGhost Peaksrdquo

Ghost peaks are peaks that always appear in the same position on the chromatogram Their origin is not the sample but the impurities from the solvents or solvent additives Therefore it is highly recommended to run a gradientwithout injection in the beginning of each method to determine the ghost peaks

To avoid irreversible adsorption at the column head you should always use a pre-column

The use of a pre-column increases the life time of a column dramatically In addition a pre-column can filter solid parts stemming from pump seals or injection rotors An alternative to a pre-column is an in-line filter These filters are attached directly to the column These filters get rid of solid parts in the eluent but will not avoid irreversible adsorption of organic impurities

Detector Systems

Virtually every chemical and physical

property that can be measured in

solution has been look at

Detectors fall roughly into two classes

Bulk property - measures an overall change in the mobile phase

Solute property - measures a solute specific property

Properties of a good detector

A detector must provide

bull high sensitivity low detection limits

bull linearity

bullReproducibility

This is true for any detector

Each detector will have specific advantages and will vary as to peak shape and spread noise and flowtemperature dependence they have

UVVis detector

A solute property detector

Sample must exhibit absorption in UVVis range Solvent must not absorb significantly at the measured wavelength

Types Filter photometer - single

bull Variable wavelength

bull Multiwavelength

Refractive index detector

Bulk property detector - general purpose

Based on refraction of light as it passes from

one media to another Presence of a solute

changes the refractive index of the solvent

Heat of absorption detector

A small amount of heat is released when a sample absorbs on a suitable surface

This detector can measure this

Electrochemical detectors

A number of properties have been evaluated

Detector types

Dielectic constant

Amperometric

Conductometric

Polarographic

Potentiometric

Dielectric constant detector

Bulk property detector

Measures changes in polarity of the liquid phase passing through the cell

Conductometric detector

Measures conductivity of the solvent Useful for

solutions of ions

Amperometric detectors

Most frequently applied type of electrochemical

detector

A known potential is applied across a set of

electrodes - typically a glassy carbon type

Ability to oxidize or reduce a species can be

measured

Typically limited to working with a specific class

of materials per analysis

Separation is based on the analytersquos relative solubility between two liquid phases

Stationary PhaseMobile Phase

Solvent Bonded Phase

Normal Phase- Polar stationary phase and non-polar solvent

bull Reverse Phase- Non-polar stationary phase and a polar solvent

Methan0l CH3OH

bull Acetonitrile CH3CN

bull Tetrahydrofuran

bull Water H2O

Solid Support - Backbone for bonded phases Usually 10micro 5micro or 3micro silica or polymeric

particles Bonded Phases - Functional groups firmly linked

(chemically bound) to the solid support Extremely stable Reproducible

Guard - Protects the analytical column Particles Interferences Prolongs the life of the analytical column

bull Analytical - Performs the separation

C-2 Ethyl Silyl -Si-CH2-CH3

bullCN Cyanopropyl Silyl -Si-(CH2)3-CN

bull C-18C-18 Octadecyl SilylOctadecyl Silyl -Si-(CH -Si-(CH22))1717--

CHCH33

bullC-8 Octyl Silyl -Si-(CH2)7-CH3

Pump

Injector

ColumnDetector

Mobile Phases

Gradient Controller

bull

UV Single wavelength (filter) [610 8330] Variable wavelength (monochromator)

[8316 8325] Multiple wavelengths (PDA) [555]

Fluorescence [610]Electrochemical [605] Mass Spectrometric [8325]

Restekreg ULTRA C-18 and CN Columns (250mm x 46mm 5micro)

Mobile Phase (11 MethanolWater) 15 mLmin

Supelcosil LC-PAH Columns A BConditions A 150mm x 46mm 5micro

Flow Rate 15 mLmin

Conditions B 50mm x 46mm 3micro

Flow Rate 30 mLmin

  • Slide 5
  • Slide 6
  • Slide 7
  • Slide 8
  • Slide 9
  • Slide 10
  • Slide 11
  • Slide 12
  • Slide 13
  • Slide 14
  • Slide 15
  • Slide 16
  • Slide 17
  • Slide 18
  • Slide 19
  • Slide 20
  • Slide 21
  • Slide 22
  • Slide 23
  • Slide 24
  • Slide 25
  • Slide 26
  • Slide 27
  • Slide 28
  • Slide 29
  • Slide 30
  • Slide 31
  • Slide 32
  • Slide 33
  • Slide 34
  • Slide 35
  • Slide 36
  • Slide 37
  • Slide 38
  • Slide 39
  • Slide 40
Page 21: P igh erformance L iquid C hromatography  HPLC is a form of liquid chromatography used to separate compounds that are dissolved in solution. HPLC instruments

Size exclusion phases

Gels - organic or aqueous based

Controlled-pore - silica or glass

Must be selected based on pressure

requirements and size range

required for your application

Capillary and Microbore columns

Several companies have begun offering

columns with smaller ID

Microbore column - 1 mm ID packed

column

Capillary column - lt 1 mm ID internal

bound phase

These columns require smaller solvent flows

reduced sample size and improved detector

design

Capillary and Microbore columns Capillary and Microbore columns

Aromatic Compounds

mobile phase2 ethylacetate in

hexane

flow rate 4 μlmin

column Fusica II 300μm ID x

25 cm silica

sample

1 toluene

2 nitrobenzene

3 acetophenone

4 26-dinitrobenzene

injection 60 nl

detection UV 254 nm

Silica based columns

Silica is the ideal support for HPLC columns It offers a large mechanical stability excellent

physicochemical surface properties a wide range of bonding chemistries and is compatible

with a broad range of organic solvents

pH stability

In general HPLC columns are stable within a pH range of 2 to 8 If you are measuring a pH value the measurement must be done in the aqueous media before mixing the eluent with organic solvents

Modern HPLC columns can be used outside that pH range The new bonding chemistries allow use down to pH 1 for some stationary phases However please check vendorrsquos product information before using silica based column outside the pH range of 2 to 8 However best lifetimes are obtained between pH 20 and pH 68

Mechanical stability

Stationary phases based on silica are mechanically very stable The packed columns show no pressure limit and can be used at more than 40 MPa (6000 psi) without any problem However please avoid pressure shocks on the column Pressure shocks lead to channelling in the column which results in peak splitting in the corresponding chromatogram

Mobile phases (Eluents)

Silica based stationary phases are compatible with all organic solvents in the above mentioned pH range

bull Filter all prepared buffer through a 05 μm filter before using them in your HPLC system

The use of non pure solvents in HPLC causes irreversible adsorption of impurities on the column head These impurities block adsorption sites change the selectivity of the column and lead to peak splitting in the chromatogram In gradient elution impurities cause so called ldquoGhost Peaksrdquo

Ghost peaks are peaks that always appear in the same position on the chromatogram Their origin is not the sample but the impurities from the solvents or solvent additives Therefore it is highly recommended to run a gradientwithout injection in the beginning of each method to determine the ghost peaks

To avoid irreversible adsorption at the column head you should always use a pre-column

The use of a pre-column increases the life time of a column dramatically In addition a pre-column can filter solid parts stemming from pump seals or injection rotors An alternative to a pre-column is an in-line filter These filters are attached directly to the column These filters get rid of solid parts in the eluent but will not avoid irreversible adsorption of organic impurities

Detector Systems

Virtually every chemical and physical

property that can be measured in

solution has been look at

Detectors fall roughly into two classes

Bulk property - measures an overall change in the mobile phase

Solute property - measures a solute specific property

Properties of a good detector

A detector must provide

bull high sensitivity low detection limits

bull linearity

bullReproducibility

This is true for any detector

Each detector will have specific advantages and will vary as to peak shape and spread noise and flowtemperature dependence they have

UVVis detector

A solute property detector

Sample must exhibit absorption in UVVis range Solvent must not absorb significantly at the measured wavelength

Types Filter photometer - single

bull Variable wavelength

bull Multiwavelength

Refractive index detector

Bulk property detector - general purpose

Based on refraction of light as it passes from

one media to another Presence of a solute

changes the refractive index of the solvent

Heat of absorption detector

A small amount of heat is released when a sample absorbs on a suitable surface

This detector can measure this

Electrochemical detectors

A number of properties have been evaluated

Detector types

Dielectic constant

Amperometric

Conductometric

Polarographic

Potentiometric

Dielectric constant detector

Bulk property detector

Measures changes in polarity of the liquid phase passing through the cell

Conductometric detector

Measures conductivity of the solvent Useful for

solutions of ions

Amperometric detectors

Most frequently applied type of electrochemical

detector

A known potential is applied across a set of

electrodes - typically a glassy carbon type

Ability to oxidize or reduce a species can be

measured

Typically limited to working with a specific class

of materials per analysis

Separation is based on the analytersquos relative solubility between two liquid phases

Stationary PhaseMobile Phase

Solvent Bonded Phase

Normal Phase- Polar stationary phase and non-polar solvent

bull Reverse Phase- Non-polar stationary phase and a polar solvent

Methan0l CH3OH

bull Acetonitrile CH3CN

bull Tetrahydrofuran

bull Water H2O

Solid Support - Backbone for bonded phases Usually 10micro 5micro or 3micro silica or polymeric

particles Bonded Phases - Functional groups firmly linked

(chemically bound) to the solid support Extremely stable Reproducible

Guard - Protects the analytical column Particles Interferences Prolongs the life of the analytical column

bull Analytical - Performs the separation

C-2 Ethyl Silyl -Si-CH2-CH3

bullCN Cyanopropyl Silyl -Si-(CH2)3-CN

bull C-18C-18 Octadecyl SilylOctadecyl Silyl -Si-(CH -Si-(CH22))1717--

CHCH33

bullC-8 Octyl Silyl -Si-(CH2)7-CH3

Pump

Injector

ColumnDetector

Mobile Phases

Gradient Controller

bull

UV Single wavelength (filter) [610 8330] Variable wavelength (monochromator)

[8316 8325] Multiple wavelengths (PDA) [555]

Fluorescence [610]Electrochemical [605] Mass Spectrometric [8325]

Restekreg ULTRA C-18 and CN Columns (250mm x 46mm 5micro)

Mobile Phase (11 MethanolWater) 15 mLmin

Supelcosil LC-PAH Columns A BConditions A 150mm x 46mm 5micro

Flow Rate 15 mLmin

Conditions B 50mm x 46mm 3micro

Flow Rate 30 mLmin

  • Slide 5
  • Slide 6
  • Slide 7
  • Slide 8
  • Slide 9
  • Slide 10
  • Slide 11
  • Slide 12
  • Slide 13
  • Slide 14
  • Slide 15
  • Slide 16
  • Slide 17
  • Slide 18
  • Slide 19
  • Slide 20
  • Slide 21
  • Slide 22
  • Slide 23
  • Slide 24
  • Slide 25
  • Slide 26
  • Slide 27
  • Slide 28
  • Slide 29
  • Slide 30
  • Slide 31
  • Slide 32
  • Slide 33
  • Slide 34
  • Slide 35
  • Slide 36
  • Slide 37
  • Slide 38
  • Slide 39
  • Slide 40
Page 22: P igh erformance L iquid C hromatography  HPLC is a form of liquid chromatography used to separate compounds that are dissolved in solution. HPLC instruments

Capillary and Microbore columns

Several companies have begun offering

columns with smaller ID

Microbore column - 1 mm ID packed

column

Capillary column - lt 1 mm ID internal

bound phase

These columns require smaller solvent flows

reduced sample size and improved detector

design

Capillary and Microbore columns Capillary and Microbore columns

Aromatic Compounds

mobile phase2 ethylacetate in

hexane

flow rate 4 μlmin

column Fusica II 300μm ID x

25 cm silica

sample

1 toluene

2 nitrobenzene

3 acetophenone

4 26-dinitrobenzene

injection 60 nl

detection UV 254 nm

Silica based columns

Silica is the ideal support for HPLC columns It offers a large mechanical stability excellent

physicochemical surface properties a wide range of bonding chemistries and is compatible

with a broad range of organic solvents

pH stability

In general HPLC columns are stable within a pH range of 2 to 8 If you are measuring a pH value the measurement must be done in the aqueous media before mixing the eluent with organic solvents

Modern HPLC columns can be used outside that pH range The new bonding chemistries allow use down to pH 1 for some stationary phases However please check vendorrsquos product information before using silica based column outside the pH range of 2 to 8 However best lifetimes are obtained between pH 20 and pH 68

Mechanical stability

Stationary phases based on silica are mechanically very stable The packed columns show no pressure limit and can be used at more than 40 MPa (6000 psi) without any problem However please avoid pressure shocks on the column Pressure shocks lead to channelling in the column which results in peak splitting in the corresponding chromatogram

Mobile phases (Eluents)

Silica based stationary phases are compatible with all organic solvents in the above mentioned pH range

bull Filter all prepared buffer through a 05 μm filter before using them in your HPLC system

The use of non pure solvents in HPLC causes irreversible adsorption of impurities on the column head These impurities block adsorption sites change the selectivity of the column and lead to peak splitting in the chromatogram In gradient elution impurities cause so called ldquoGhost Peaksrdquo

Ghost peaks are peaks that always appear in the same position on the chromatogram Their origin is not the sample but the impurities from the solvents or solvent additives Therefore it is highly recommended to run a gradientwithout injection in the beginning of each method to determine the ghost peaks

To avoid irreversible adsorption at the column head you should always use a pre-column

The use of a pre-column increases the life time of a column dramatically In addition a pre-column can filter solid parts stemming from pump seals or injection rotors An alternative to a pre-column is an in-line filter These filters are attached directly to the column These filters get rid of solid parts in the eluent but will not avoid irreversible adsorption of organic impurities

Detector Systems

Virtually every chemical and physical

property that can be measured in

solution has been look at

Detectors fall roughly into two classes

Bulk property - measures an overall change in the mobile phase

Solute property - measures a solute specific property

Properties of a good detector

A detector must provide

bull high sensitivity low detection limits

bull linearity

bullReproducibility

This is true for any detector

Each detector will have specific advantages and will vary as to peak shape and spread noise and flowtemperature dependence they have

UVVis detector

A solute property detector

Sample must exhibit absorption in UVVis range Solvent must not absorb significantly at the measured wavelength

Types Filter photometer - single

bull Variable wavelength

bull Multiwavelength

Refractive index detector

Bulk property detector - general purpose

Based on refraction of light as it passes from

one media to another Presence of a solute

changes the refractive index of the solvent

Heat of absorption detector

A small amount of heat is released when a sample absorbs on a suitable surface

This detector can measure this

Electrochemical detectors

A number of properties have been evaluated

Detector types

Dielectic constant

Amperometric

Conductometric

Polarographic

Potentiometric

Dielectric constant detector

Bulk property detector

Measures changes in polarity of the liquid phase passing through the cell

Conductometric detector

Measures conductivity of the solvent Useful for

solutions of ions

Amperometric detectors

Most frequently applied type of electrochemical

detector

A known potential is applied across a set of

electrodes - typically a glassy carbon type

Ability to oxidize or reduce a species can be

measured

Typically limited to working with a specific class

of materials per analysis

Separation is based on the analytersquos relative solubility between two liquid phases

Stationary PhaseMobile Phase

Solvent Bonded Phase

Normal Phase- Polar stationary phase and non-polar solvent

bull Reverse Phase- Non-polar stationary phase and a polar solvent

Methan0l CH3OH

bull Acetonitrile CH3CN

bull Tetrahydrofuran

bull Water H2O

Solid Support - Backbone for bonded phases Usually 10micro 5micro or 3micro silica or polymeric

particles Bonded Phases - Functional groups firmly linked

(chemically bound) to the solid support Extremely stable Reproducible

Guard - Protects the analytical column Particles Interferences Prolongs the life of the analytical column

bull Analytical - Performs the separation

C-2 Ethyl Silyl -Si-CH2-CH3

bullCN Cyanopropyl Silyl -Si-(CH2)3-CN

bull C-18C-18 Octadecyl SilylOctadecyl Silyl -Si-(CH -Si-(CH22))1717--

CHCH33

bullC-8 Octyl Silyl -Si-(CH2)7-CH3

Pump

Injector

ColumnDetector

Mobile Phases

Gradient Controller

bull

UV Single wavelength (filter) [610 8330] Variable wavelength (monochromator)

[8316 8325] Multiple wavelengths (PDA) [555]

Fluorescence [610]Electrochemical [605] Mass Spectrometric [8325]

Restekreg ULTRA C-18 and CN Columns (250mm x 46mm 5micro)

Mobile Phase (11 MethanolWater) 15 mLmin

Supelcosil LC-PAH Columns A BConditions A 150mm x 46mm 5micro

Flow Rate 15 mLmin

Conditions B 50mm x 46mm 3micro

Flow Rate 30 mLmin

  • Slide 5
  • Slide 6
  • Slide 7
  • Slide 8
  • Slide 9
  • Slide 10
  • Slide 11
  • Slide 12
  • Slide 13
  • Slide 14
  • Slide 15
  • Slide 16
  • Slide 17
  • Slide 18
  • Slide 19
  • Slide 20
  • Slide 21
  • Slide 22
  • Slide 23
  • Slide 24
  • Slide 25
  • Slide 26
  • Slide 27
  • Slide 28
  • Slide 29
  • Slide 30
  • Slide 31
  • Slide 32
  • Slide 33
  • Slide 34
  • Slide 35
  • Slide 36
  • Slide 37
  • Slide 38
  • Slide 39
  • Slide 40
Page 23: P igh erformance L iquid C hromatography  HPLC is a form of liquid chromatography used to separate compounds that are dissolved in solution. HPLC instruments

Capillary and Microbore columns Capillary and Microbore columns

Aromatic Compounds

mobile phase2 ethylacetate in

hexane

flow rate 4 μlmin

column Fusica II 300μm ID x

25 cm silica

sample

1 toluene

2 nitrobenzene

3 acetophenone

4 26-dinitrobenzene

injection 60 nl

detection UV 254 nm

Silica based columns

Silica is the ideal support for HPLC columns It offers a large mechanical stability excellent

physicochemical surface properties a wide range of bonding chemistries and is compatible

with a broad range of organic solvents

pH stability

In general HPLC columns are stable within a pH range of 2 to 8 If you are measuring a pH value the measurement must be done in the aqueous media before mixing the eluent with organic solvents

Modern HPLC columns can be used outside that pH range The new bonding chemistries allow use down to pH 1 for some stationary phases However please check vendorrsquos product information before using silica based column outside the pH range of 2 to 8 However best lifetimes are obtained between pH 20 and pH 68

Mechanical stability

Stationary phases based on silica are mechanically very stable The packed columns show no pressure limit and can be used at more than 40 MPa (6000 psi) without any problem However please avoid pressure shocks on the column Pressure shocks lead to channelling in the column which results in peak splitting in the corresponding chromatogram

Mobile phases (Eluents)

Silica based stationary phases are compatible with all organic solvents in the above mentioned pH range

bull Filter all prepared buffer through a 05 μm filter before using them in your HPLC system

The use of non pure solvents in HPLC causes irreversible adsorption of impurities on the column head These impurities block adsorption sites change the selectivity of the column and lead to peak splitting in the chromatogram In gradient elution impurities cause so called ldquoGhost Peaksrdquo

Ghost peaks are peaks that always appear in the same position on the chromatogram Their origin is not the sample but the impurities from the solvents or solvent additives Therefore it is highly recommended to run a gradientwithout injection in the beginning of each method to determine the ghost peaks

To avoid irreversible adsorption at the column head you should always use a pre-column

The use of a pre-column increases the life time of a column dramatically In addition a pre-column can filter solid parts stemming from pump seals or injection rotors An alternative to a pre-column is an in-line filter These filters are attached directly to the column These filters get rid of solid parts in the eluent but will not avoid irreversible adsorption of organic impurities

Detector Systems

Virtually every chemical and physical

property that can be measured in

solution has been look at

Detectors fall roughly into two classes

Bulk property - measures an overall change in the mobile phase

Solute property - measures a solute specific property

Properties of a good detector

A detector must provide

bull high sensitivity low detection limits

bull linearity

bullReproducibility

This is true for any detector

Each detector will have specific advantages and will vary as to peak shape and spread noise and flowtemperature dependence they have

UVVis detector

A solute property detector

Sample must exhibit absorption in UVVis range Solvent must not absorb significantly at the measured wavelength

Types Filter photometer - single

bull Variable wavelength

bull Multiwavelength

Refractive index detector

Bulk property detector - general purpose

Based on refraction of light as it passes from

one media to another Presence of a solute

changes the refractive index of the solvent

Heat of absorption detector

A small amount of heat is released when a sample absorbs on a suitable surface

This detector can measure this

Electrochemical detectors

A number of properties have been evaluated

Detector types

Dielectic constant

Amperometric

Conductometric

Polarographic

Potentiometric

Dielectric constant detector

Bulk property detector

Measures changes in polarity of the liquid phase passing through the cell

Conductometric detector

Measures conductivity of the solvent Useful for

solutions of ions

Amperometric detectors

Most frequently applied type of electrochemical

detector

A known potential is applied across a set of

electrodes - typically a glassy carbon type

Ability to oxidize or reduce a species can be

measured

Typically limited to working with a specific class

of materials per analysis

Separation is based on the analytersquos relative solubility between two liquid phases

Stationary PhaseMobile Phase

Solvent Bonded Phase

Normal Phase- Polar stationary phase and non-polar solvent

bull Reverse Phase- Non-polar stationary phase and a polar solvent

Methan0l CH3OH

bull Acetonitrile CH3CN

bull Tetrahydrofuran

bull Water H2O

Solid Support - Backbone for bonded phases Usually 10micro 5micro or 3micro silica or polymeric

particles Bonded Phases - Functional groups firmly linked

(chemically bound) to the solid support Extremely stable Reproducible

Guard - Protects the analytical column Particles Interferences Prolongs the life of the analytical column

bull Analytical - Performs the separation

C-2 Ethyl Silyl -Si-CH2-CH3

bullCN Cyanopropyl Silyl -Si-(CH2)3-CN

bull C-18C-18 Octadecyl SilylOctadecyl Silyl -Si-(CH -Si-(CH22))1717--

CHCH33

bullC-8 Octyl Silyl -Si-(CH2)7-CH3

Pump

Injector

ColumnDetector

Mobile Phases

Gradient Controller

bull

UV Single wavelength (filter) [610 8330] Variable wavelength (monochromator)

[8316 8325] Multiple wavelengths (PDA) [555]

Fluorescence [610]Electrochemical [605] Mass Spectrometric [8325]

Restekreg ULTRA C-18 and CN Columns (250mm x 46mm 5micro)

Mobile Phase (11 MethanolWater) 15 mLmin

Supelcosil LC-PAH Columns A BConditions A 150mm x 46mm 5micro

Flow Rate 15 mLmin

Conditions B 50mm x 46mm 3micro

Flow Rate 30 mLmin

  • Slide 5
  • Slide 6
  • Slide 7
  • Slide 8
  • Slide 9
  • Slide 10
  • Slide 11
  • Slide 12
  • Slide 13
  • Slide 14
  • Slide 15
  • Slide 16
  • Slide 17
  • Slide 18
  • Slide 19
  • Slide 20
  • Slide 21
  • Slide 22
  • Slide 23
  • Slide 24
  • Slide 25
  • Slide 26
  • Slide 27
  • Slide 28
  • Slide 29
  • Slide 30
  • Slide 31
  • Slide 32
  • Slide 33
  • Slide 34
  • Slide 35
  • Slide 36
  • Slide 37
  • Slide 38
  • Slide 39
  • Slide 40
Page 24: P igh erformance L iquid C hromatography  HPLC is a form of liquid chromatography used to separate compounds that are dissolved in solution. HPLC instruments

Silica based columns

Silica is the ideal support for HPLC columns It offers a large mechanical stability excellent

physicochemical surface properties a wide range of bonding chemistries and is compatible

with a broad range of organic solvents

pH stability

In general HPLC columns are stable within a pH range of 2 to 8 If you are measuring a pH value the measurement must be done in the aqueous media before mixing the eluent with organic solvents

Modern HPLC columns can be used outside that pH range The new bonding chemistries allow use down to pH 1 for some stationary phases However please check vendorrsquos product information before using silica based column outside the pH range of 2 to 8 However best lifetimes are obtained between pH 20 and pH 68

Mechanical stability

Stationary phases based on silica are mechanically very stable The packed columns show no pressure limit and can be used at more than 40 MPa (6000 psi) without any problem However please avoid pressure shocks on the column Pressure shocks lead to channelling in the column which results in peak splitting in the corresponding chromatogram

Mobile phases (Eluents)

Silica based stationary phases are compatible with all organic solvents in the above mentioned pH range

bull Filter all prepared buffer through a 05 μm filter before using them in your HPLC system

The use of non pure solvents in HPLC causes irreversible adsorption of impurities on the column head These impurities block adsorption sites change the selectivity of the column and lead to peak splitting in the chromatogram In gradient elution impurities cause so called ldquoGhost Peaksrdquo

Ghost peaks are peaks that always appear in the same position on the chromatogram Their origin is not the sample but the impurities from the solvents or solvent additives Therefore it is highly recommended to run a gradientwithout injection in the beginning of each method to determine the ghost peaks

To avoid irreversible adsorption at the column head you should always use a pre-column

The use of a pre-column increases the life time of a column dramatically In addition a pre-column can filter solid parts stemming from pump seals or injection rotors An alternative to a pre-column is an in-line filter These filters are attached directly to the column These filters get rid of solid parts in the eluent but will not avoid irreversible adsorption of organic impurities

Detector Systems

Virtually every chemical and physical

property that can be measured in

solution has been look at

Detectors fall roughly into two classes

Bulk property - measures an overall change in the mobile phase

Solute property - measures a solute specific property

Properties of a good detector

A detector must provide

bull high sensitivity low detection limits

bull linearity

bullReproducibility

This is true for any detector

Each detector will have specific advantages and will vary as to peak shape and spread noise and flowtemperature dependence they have

UVVis detector

A solute property detector

Sample must exhibit absorption in UVVis range Solvent must not absorb significantly at the measured wavelength

Types Filter photometer - single

bull Variable wavelength

bull Multiwavelength

Refractive index detector

Bulk property detector - general purpose

Based on refraction of light as it passes from

one media to another Presence of a solute

changes the refractive index of the solvent

Heat of absorption detector

A small amount of heat is released when a sample absorbs on a suitable surface

This detector can measure this

Electrochemical detectors

A number of properties have been evaluated

Detector types

Dielectic constant

Amperometric

Conductometric

Polarographic

Potentiometric

Dielectric constant detector

Bulk property detector

Measures changes in polarity of the liquid phase passing through the cell

Conductometric detector

Measures conductivity of the solvent Useful for

solutions of ions

Amperometric detectors

Most frequently applied type of electrochemical

detector

A known potential is applied across a set of

electrodes - typically a glassy carbon type

Ability to oxidize or reduce a species can be

measured

Typically limited to working with a specific class

of materials per analysis

Separation is based on the analytersquos relative solubility between two liquid phases

Stationary PhaseMobile Phase

Solvent Bonded Phase

Normal Phase- Polar stationary phase and non-polar solvent

bull Reverse Phase- Non-polar stationary phase and a polar solvent

Methan0l CH3OH

bull Acetonitrile CH3CN

bull Tetrahydrofuran

bull Water H2O

Solid Support - Backbone for bonded phases Usually 10micro 5micro or 3micro silica or polymeric

particles Bonded Phases - Functional groups firmly linked

(chemically bound) to the solid support Extremely stable Reproducible

Guard - Protects the analytical column Particles Interferences Prolongs the life of the analytical column

bull Analytical - Performs the separation

C-2 Ethyl Silyl -Si-CH2-CH3

bullCN Cyanopropyl Silyl -Si-(CH2)3-CN

bull C-18C-18 Octadecyl SilylOctadecyl Silyl -Si-(CH -Si-(CH22))1717--

CHCH33

bullC-8 Octyl Silyl -Si-(CH2)7-CH3

Pump

Injector

ColumnDetector

Mobile Phases

Gradient Controller

bull

UV Single wavelength (filter) [610 8330] Variable wavelength (monochromator)

[8316 8325] Multiple wavelengths (PDA) [555]

Fluorescence [610]Electrochemical [605] Mass Spectrometric [8325]

Restekreg ULTRA C-18 and CN Columns (250mm x 46mm 5micro)

Mobile Phase (11 MethanolWater) 15 mLmin

Supelcosil LC-PAH Columns A BConditions A 150mm x 46mm 5micro

Flow Rate 15 mLmin

Conditions B 50mm x 46mm 3micro

Flow Rate 30 mLmin

  • Slide 5
  • Slide 6
  • Slide 7
  • Slide 8
  • Slide 9
  • Slide 10
  • Slide 11
  • Slide 12
  • Slide 13
  • Slide 14
  • Slide 15
  • Slide 16
  • Slide 17
  • Slide 18
  • Slide 19
  • Slide 20
  • Slide 21
  • Slide 22
  • Slide 23
  • Slide 24
  • Slide 25
  • Slide 26
  • Slide 27
  • Slide 28
  • Slide 29
  • Slide 30
  • Slide 31
  • Slide 32
  • Slide 33
  • Slide 34
  • Slide 35
  • Slide 36
  • Slide 37
  • Slide 38
  • Slide 39
  • Slide 40
Page 25: P igh erformance L iquid C hromatography  HPLC is a form of liquid chromatography used to separate compounds that are dissolved in solution. HPLC instruments

pH stability

In general HPLC columns are stable within a pH range of 2 to 8 If you are measuring a pH value the measurement must be done in the aqueous media before mixing the eluent with organic solvents

Modern HPLC columns can be used outside that pH range The new bonding chemistries allow use down to pH 1 for some stationary phases However please check vendorrsquos product information before using silica based column outside the pH range of 2 to 8 However best lifetimes are obtained between pH 20 and pH 68

Mechanical stability

Stationary phases based on silica are mechanically very stable The packed columns show no pressure limit and can be used at more than 40 MPa (6000 psi) without any problem However please avoid pressure shocks on the column Pressure shocks lead to channelling in the column which results in peak splitting in the corresponding chromatogram

Mobile phases (Eluents)

Silica based stationary phases are compatible with all organic solvents in the above mentioned pH range

bull Filter all prepared buffer through a 05 μm filter before using them in your HPLC system

The use of non pure solvents in HPLC causes irreversible adsorption of impurities on the column head These impurities block adsorption sites change the selectivity of the column and lead to peak splitting in the chromatogram In gradient elution impurities cause so called ldquoGhost Peaksrdquo

Ghost peaks are peaks that always appear in the same position on the chromatogram Their origin is not the sample but the impurities from the solvents or solvent additives Therefore it is highly recommended to run a gradientwithout injection in the beginning of each method to determine the ghost peaks

To avoid irreversible adsorption at the column head you should always use a pre-column

The use of a pre-column increases the life time of a column dramatically In addition a pre-column can filter solid parts stemming from pump seals or injection rotors An alternative to a pre-column is an in-line filter These filters are attached directly to the column These filters get rid of solid parts in the eluent but will not avoid irreversible adsorption of organic impurities

Detector Systems

Virtually every chemical and physical

property that can be measured in

solution has been look at

Detectors fall roughly into two classes

Bulk property - measures an overall change in the mobile phase

Solute property - measures a solute specific property

Properties of a good detector

A detector must provide

bull high sensitivity low detection limits

bull linearity

bullReproducibility

This is true for any detector

Each detector will have specific advantages and will vary as to peak shape and spread noise and flowtemperature dependence they have

UVVis detector

A solute property detector

Sample must exhibit absorption in UVVis range Solvent must not absorb significantly at the measured wavelength

Types Filter photometer - single

bull Variable wavelength

bull Multiwavelength

Refractive index detector

Bulk property detector - general purpose

Based on refraction of light as it passes from

one media to another Presence of a solute

changes the refractive index of the solvent

Heat of absorption detector

A small amount of heat is released when a sample absorbs on a suitable surface

This detector can measure this

Electrochemical detectors

A number of properties have been evaluated

Detector types

Dielectic constant

Amperometric

Conductometric

Polarographic

Potentiometric

Dielectric constant detector

Bulk property detector

Measures changes in polarity of the liquid phase passing through the cell

Conductometric detector

Measures conductivity of the solvent Useful for

solutions of ions

Amperometric detectors

Most frequently applied type of electrochemical

detector

A known potential is applied across a set of

electrodes - typically a glassy carbon type

Ability to oxidize or reduce a species can be

measured

Typically limited to working with a specific class

of materials per analysis

Separation is based on the analytersquos relative solubility between two liquid phases

Stationary PhaseMobile Phase

Solvent Bonded Phase

Normal Phase- Polar stationary phase and non-polar solvent

bull Reverse Phase- Non-polar stationary phase and a polar solvent

Methan0l CH3OH

bull Acetonitrile CH3CN

bull Tetrahydrofuran

bull Water H2O

Solid Support - Backbone for bonded phases Usually 10micro 5micro or 3micro silica or polymeric

particles Bonded Phases - Functional groups firmly linked

(chemically bound) to the solid support Extremely stable Reproducible

Guard - Protects the analytical column Particles Interferences Prolongs the life of the analytical column

bull Analytical - Performs the separation

C-2 Ethyl Silyl -Si-CH2-CH3

bullCN Cyanopropyl Silyl -Si-(CH2)3-CN

bull C-18C-18 Octadecyl SilylOctadecyl Silyl -Si-(CH -Si-(CH22))1717--

CHCH33

bullC-8 Octyl Silyl -Si-(CH2)7-CH3

Pump

Injector

ColumnDetector

Mobile Phases

Gradient Controller

bull

UV Single wavelength (filter) [610 8330] Variable wavelength (monochromator)

[8316 8325] Multiple wavelengths (PDA) [555]

Fluorescence [610]Electrochemical [605] Mass Spectrometric [8325]

Restekreg ULTRA C-18 and CN Columns (250mm x 46mm 5micro)

Mobile Phase (11 MethanolWater) 15 mLmin

Supelcosil LC-PAH Columns A BConditions A 150mm x 46mm 5micro

Flow Rate 15 mLmin

Conditions B 50mm x 46mm 3micro

Flow Rate 30 mLmin

  • Slide 5
  • Slide 6
  • Slide 7
  • Slide 8
  • Slide 9
  • Slide 10
  • Slide 11
  • Slide 12
  • Slide 13
  • Slide 14
  • Slide 15
  • Slide 16
  • Slide 17
  • Slide 18
  • Slide 19
  • Slide 20
  • Slide 21
  • Slide 22
  • Slide 23
  • Slide 24
  • Slide 25
  • Slide 26
  • Slide 27
  • Slide 28
  • Slide 29
  • Slide 30
  • Slide 31
  • Slide 32
  • Slide 33
  • Slide 34
  • Slide 35
  • Slide 36
  • Slide 37
  • Slide 38
  • Slide 39
  • Slide 40
Page 26: P igh erformance L iquid C hromatography  HPLC is a form of liquid chromatography used to separate compounds that are dissolved in solution. HPLC instruments

Mechanical stability

Stationary phases based on silica are mechanically very stable The packed columns show no pressure limit and can be used at more than 40 MPa (6000 psi) without any problem However please avoid pressure shocks on the column Pressure shocks lead to channelling in the column which results in peak splitting in the corresponding chromatogram

Mobile phases (Eluents)

Silica based stationary phases are compatible with all organic solvents in the above mentioned pH range

bull Filter all prepared buffer through a 05 μm filter before using them in your HPLC system

The use of non pure solvents in HPLC causes irreversible adsorption of impurities on the column head These impurities block adsorption sites change the selectivity of the column and lead to peak splitting in the chromatogram In gradient elution impurities cause so called ldquoGhost Peaksrdquo

Ghost peaks are peaks that always appear in the same position on the chromatogram Their origin is not the sample but the impurities from the solvents or solvent additives Therefore it is highly recommended to run a gradientwithout injection in the beginning of each method to determine the ghost peaks

To avoid irreversible adsorption at the column head you should always use a pre-column

The use of a pre-column increases the life time of a column dramatically In addition a pre-column can filter solid parts stemming from pump seals or injection rotors An alternative to a pre-column is an in-line filter These filters are attached directly to the column These filters get rid of solid parts in the eluent but will not avoid irreversible adsorption of organic impurities

Detector Systems

Virtually every chemical and physical

property that can be measured in

solution has been look at

Detectors fall roughly into two classes

Bulk property - measures an overall change in the mobile phase

Solute property - measures a solute specific property

Properties of a good detector

A detector must provide

bull high sensitivity low detection limits

bull linearity

bullReproducibility

This is true for any detector

Each detector will have specific advantages and will vary as to peak shape and spread noise and flowtemperature dependence they have

UVVis detector

A solute property detector

Sample must exhibit absorption in UVVis range Solvent must not absorb significantly at the measured wavelength

Types Filter photometer - single

bull Variable wavelength

bull Multiwavelength

Refractive index detector

Bulk property detector - general purpose

Based on refraction of light as it passes from

one media to another Presence of a solute

changes the refractive index of the solvent

Heat of absorption detector

A small amount of heat is released when a sample absorbs on a suitable surface

This detector can measure this

Electrochemical detectors

A number of properties have been evaluated

Detector types

Dielectic constant

Amperometric

Conductometric

Polarographic

Potentiometric

Dielectric constant detector

Bulk property detector

Measures changes in polarity of the liquid phase passing through the cell

Conductometric detector

Measures conductivity of the solvent Useful for

solutions of ions

Amperometric detectors

Most frequently applied type of electrochemical

detector

A known potential is applied across a set of

electrodes - typically a glassy carbon type

Ability to oxidize or reduce a species can be

measured

Typically limited to working with a specific class

of materials per analysis

Separation is based on the analytersquos relative solubility between two liquid phases

Stationary PhaseMobile Phase

Solvent Bonded Phase

Normal Phase- Polar stationary phase and non-polar solvent

bull Reverse Phase- Non-polar stationary phase and a polar solvent

Methan0l CH3OH

bull Acetonitrile CH3CN

bull Tetrahydrofuran

bull Water H2O

Solid Support - Backbone for bonded phases Usually 10micro 5micro or 3micro silica or polymeric

particles Bonded Phases - Functional groups firmly linked

(chemically bound) to the solid support Extremely stable Reproducible

Guard - Protects the analytical column Particles Interferences Prolongs the life of the analytical column

bull Analytical - Performs the separation

C-2 Ethyl Silyl -Si-CH2-CH3

bullCN Cyanopropyl Silyl -Si-(CH2)3-CN

bull C-18C-18 Octadecyl SilylOctadecyl Silyl -Si-(CH -Si-(CH22))1717--

CHCH33

bullC-8 Octyl Silyl -Si-(CH2)7-CH3

Pump

Injector

ColumnDetector

Mobile Phases

Gradient Controller

bull

UV Single wavelength (filter) [610 8330] Variable wavelength (monochromator)

[8316 8325] Multiple wavelengths (PDA) [555]

Fluorescence [610]Electrochemical [605] Mass Spectrometric [8325]

Restekreg ULTRA C-18 and CN Columns (250mm x 46mm 5micro)

Mobile Phase (11 MethanolWater) 15 mLmin

Supelcosil LC-PAH Columns A BConditions A 150mm x 46mm 5micro

Flow Rate 15 mLmin

Conditions B 50mm x 46mm 3micro

Flow Rate 30 mLmin

  • Slide 5
  • Slide 6
  • Slide 7
  • Slide 8
  • Slide 9
  • Slide 10
  • Slide 11
  • Slide 12
  • Slide 13
  • Slide 14
  • Slide 15
  • Slide 16
  • Slide 17
  • Slide 18
  • Slide 19
  • Slide 20
  • Slide 21
  • Slide 22
  • Slide 23
  • Slide 24
  • Slide 25
  • Slide 26
  • Slide 27
  • Slide 28
  • Slide 29
  • Slide 30
  • Slide 31
  • Slide 32
  • Slide 33
  • Slide 34
  • Slide 35
  • Slide 36
  • Slide 37
  • Slide 38
  • Slide 39
  • Slide 40
Page 27: P igh erformance L iquid C hromatography  HPLC is a form of liquid chromatography used to separate compounds that are dissolved in solution. HPLC instruments

Mobile phases (Eluents)

Silica based stationary phases are compatible with all organic solvents in the above mentioned pH range

bull Filter all prepared buffer through a 05 μm filter before using them in your HPLC system

The use of non pure solvents in HPLC causes irreversible adsorption of impurities on the column head These impurities block adsorption sites change the selectivity of the column and lead to peak splitting in the chromatogram In gradient elution impurities cause so called ldquoGhost Peaksrdquo

Ghost peaks are peaks that always appear in the same position on the chromatogram Their origin is not the sample but the impurities from the solvents or solvent additives Therefore it is highly recommended to run a gradientwithout injection in the beginning of each method to determine the ghost peaks

To avoid irreversible adsorption at the column head you should always use a pre-column

The use of a pre-column increases the life time of a column dramatically In addition a pre-column can filter solid parts stemming from pump seals or injection rotors An alternative to a pre-column is an in-line filter These filters are attached directly to the column These filters get rid of solid parts in the eluent but will not avoid irreversible adsorption of organic impurities

Detector Systems

Virtually every chemical and physical

property that can be measured in

solution has been look at

Detectors fall roughly into two classes

Bulk property - measures an overall change in the mobile phase

Solute property - measures a solute specific property

Properties of a good detector

A detector must provide

bull high sensitivity low detection limits

bull linearity

bullReproducibility

This is true for any detector

Each detector will have specific advantages and will vary as to peak shape and spread noise and flowtemperature dependence they have

UVVis detector

A solute property detector

Sample must exhibit absorption in UVVis range Solvent must not absorb significantly at the measured wavelength

Types Filter photometer - single

bull Variable wavelength

bull Multiwavelength

Refractive index detector

Bulk property detector - general purpose

Based on refraction of light as it passes from

one media to another Presence of a solute

changes the refractive index of the solvent

Heat of absorption detector

A small amount of heat is released when a sample absorbs on a suitable surface

This detector can measure this

Electrochemical detectors

A number of properties have been evaluated

Detector types

Dielectic constant

Amperometric

Conductometric

Polarographic

Potentiometric

Dielectric constant detector

Bulk property detector

Measures changes in polarity of the liquid phase passing through the cell

Conductometric detector

Measures conductivity of the solvent Useful for

solutions of ions

Amperometric detectors

Most frequently applied type of electrochemical

detector

A known potential is applied across a set of

electrodes - typically a glassy carbon type

Ability to oxidize or reduce a species can be

measured

Typically limited to working with a specific class

of materials per analysis

Separation is based on the analytersquos relative solubility between two liquid phases

Stationary PhaseMobile Phase

Solvent Bonded Phase

Normal Phase- Polar stationary phase and non-polar solvent

bull Reverse Phase- Non-polar stationary phase and a polar solvent

Methan0l CH3OH

bull Acetonitrile CH3CN

bull Tetrahydrofuran

bull Water H2O

Solid Support - Backbone for bonded phases Usually 10micro 5micro or 3micro silica or polymeric

particles Bonded Phases - Functional groups firmly linked

(chemically bound) to the solid support Extremely stable Reproducible

Guard - Protects the analytical column Particles Interferences Prolongs the life of the analytical column

bull Analytical - Performs the separation

C-2 Ethyl Silyl -Si-CH2-CH3

bullCN Cyanopropyl Silyl -Si-(CH2)3-CN

bull C-18C-18 Octadecyl SilylOctadecyl Silyl -Si-(CH -Si-(CH22))1717--

CHCH33

bullC-8 Octyl Silyl -Si-(CH2)7-CH3

Pump

Injector

ColumnDetector

Mobile Phases

Gradient Controller

bull

UV Single wavelength (filter) [610 8330] Variable wavelength (monochromator)

[8316 8325] Multiple wavelengths (PDA) [555]

Fluorescence [610]Electrochemical [605] Mass Spectrometric [8325]

Restekreg ULTRA C-18 and CN Columns (250mm x 46mm 5micro)

Mobile Phase (11 MethanolWater) 15 mLmin

Supelcosil LC-PAH Columns A BConditions A 150mm x 46mm 5micro

Flow Rate 15 mLmin

Conditions B 50mm x 46mm 3micro

Flow Rate 30 mLmin

  • Slide 5
  • Slide 6
  • Slide 7
  • Slide 8
  • Slide 9
  • Slide 10
  • Slide 11
  • Slide 12
  • Slide 13
  • Slide 14
  • Slide 15
  • Slide 16
  • Slide 17
  • Slide 18
  • Slide 19
  • Slide 20
  • Slide 21
  • Slide 22
  • Slide 23
  • Slide 24
  • Slide 25
  • Slide 26
  • Slide 27
  • Slide 28
  • Slide 29
  • Slide 30
  • Slide 31
  • Slide 32
  • Slide 33
  • Slide 34
  • Slide 35
  • Slide 36
  • Slide 37
  • Slide 38
  • Slide 39
  • Slide 40
Page 28: P igh erformance L iquid C hromatography  HPLC is a form of liquid chromatography used to separate compounds that are dissolved in solution. HPLC instruments

The use of non pure solvents in HPLC causes irreversible adsorption of impurities on the column head These impurities block adsorption sites change the selectivity of the column and lead to peak splitting in the chromatogram In gradient elution impurities cause so called ldquoGhost Peaksrdquo

Ghost peaks are peaks that always appear in the same position on the chromatogram Their origin is not the sample but the impurities from the solvents or solvent additives Therefore it is highly recommended to run a gradientwithout injection in the beginning of each method to determine the ghost peaks

To avoid irreversible adsorption at the column head you should always use a pre-column

The use of a pre-column increases the life time of a column dramatically In addition a pre-column can filter solid parts stemming from pump seals or injection rotors An alternative to a pre-column is an in-line filter These filters are attached directly to the column These filters get rid of solid parts in the eluent but will not avoid irreversible adsorption of organic impurities

Detector Systems

Virtually every chemical and physical

property that can be measured in

solution has been look at

Detectors fall roughly into two classes

Bulk property - measures an overall change in the mobile phase

Solute property - measures a solute specific property

Properties of a good detector

A detector must provide

bull high sensitivity low detection limits

bull linearity

bullReproducibility

This is true for any detector

Each detector will have specific advantages and will vary as to peak shape and spread noise and flowtemperature dependence they have

UVVis detector

A solute property detector

Sample must exhibit absorption in UVVis range Solvent must not absorb significantly at the measured wavelength

Types Filter photometer - single

bull Variable wavelength

bull Multiwavelength

Refractive index detector

Bulk property detector - general purpose

Based on refraction of light as it passes from

one media to another Presence of a solute

changes the refractive index of the solvent

Heat of absorption detector

A small amount of heat is released when a sample absorbs on a suitable surface

This detector can measure this

Electrochemical detectors

A number of properties have been evaluated

Detector types

Dielectic constant

Amperometric

Conductometric

Polarographic

Potentiometric

Dielectric constant detector

Bulk property detector

Measures changes in polarity of the liquid phase passing through the cell

Conductometric detector

Measures conductivity of the solvent Useful for

solutions of ions

Amperometric detectors

Most frequently applied type of electrochemical

detector

A known potential is applied across a set of

electrodes - typically a glassy carbon type

Ability to oxidize or reduce a species can be

measured

Typically limited to working with a specific class

of materials per analysis

Separation is based on the analytersquos relative solubility between two liquid phases

Stationary PhaseMobile Phase

Solvent Bonded Phase

Normal Phase- Polar stationary phase and non-polar solvent

bull Reverse Phase- Non-polar stationary phase and a polar solvent

Methan0l CH3OH

bull Acetonitrile CH3CN

bull Tetrahydrofuran

bull Water H2O

Solid Support - Backbone for bonded phases Usually 10micro 5micro or 3micro silica or polymeric

particles Bonded Phases - Functional groups firmly linked

(chemically bound) to the solid support Extremely stable Reproducible

Guard - Protects the analytical column Particles Interferences Prolongs the life of the analytical column

bull Analytical - Performs the separation

C-2 Ethyl Silyl -Si-CH2-CH3

bullCN Cyanopropyl Silyl -Si-(CH2)3-CN

bull C-18C-18 Octadecyl SilylOctadecyl Silyl -Si-(CH -Si-(CH22))1717--

CHCH33

bullC-8 Octyl Silyl -Si-(CH2)7-CH3

Pump

Injector

ColumnDetector

Mobile Phases

Gradient Controller

bull

UV Single wavelength (filter) [610 8330] Variable wavelength (monochromator)

[8316 8325] Multiple wavelengths (PDA) [555]

Fluorescence [610]Electrochemical [605] Mass Spectrometric [8325]

Restekreg ULTRA C-18 and CN Columns (250mm x 46mm 5micro)

Mobile Phase (11 MethanolWater) 15 mLmin

Supelcosil LC-PAH Columns A BConditions A 150mm x 46mm 5micro

Flow Rate 15 mLmin

Conditions B 50mm x 46mm 3micro

Flow Rate 30 mLmin

  • Slide 5
  • Slide 6
  • Slide 7
  • Slide 8
  • Slide 9
  • Slide 10
  • Slide 11
  • Slide 12
  • Slide 13
  • Slide 14
  • Slide 15
  • Slide 16
  • Slide 17
  • Slide 18
  • Slide 19
  • Slide 20
  • Slide 21
  • Slide 22
  • Slide 23
  • Slide 24
  • Slide 25
  • Slide 26
  • Slide 27
  • Slide 28
  • Slide 29
  • Slide 30
  • Slide 31
  • Slide 32
  • Slide 33
  • Slide 34
  • Slide 35
  • Slide 36
  • Slide 37
  • Slide 38
  • Slide 39
  • Slide 40
Page 29: P igh erformance L iquid C hromatography  HPLC is a form of liquid chromatography used to separate compounds that are dissolved in solution. HPLC instruments

To avoid irreversible adsorption at the column head you should always use a pre-column

The use of a pre-column increases the life time of a column dramatically In addition a pre-column can filter solid parts stemming from pump seals or injection rotors An alternative to a pre-column is an in-line filter These filters are attached directly to the column These filters get rid of solid parts in the eluent but will not avoid irreversible adsorption of organic impurities

Detector Systems

Virtually every chemical and physical

property that can be measured in

solution has been look at

Detectors fall roughly into two classes

Bulk property - measures an overall change in the mobile phase

Solute property - measures a solute specific property

Properties of a good detector

A detector must provide

bull high sensitivity low detection limits

bull linearity

bullReproducibility

This is true for any detector

Each detector will have specific advantages and will vary as to peak shape and spread noise and flowtemperature dependence they have

UVVis detector

A solute property detector

Sample must exhibit absorption in UVVis range Solvent must not absorb significantly at the measured wavelength

Types Filter photometer - single

bull Variable wavelength

bull Multiwavelength

Refractive index detector

Bulk property detector - general purpose

Based on refraction of light as it passes from

one media to another Presence of a solute

changes the refractive index of the solvent

Heat of absorption detector

A small amount of heat is released when a sample absorbs on a suitable surface

This detector can measure this

Electrochemical detectors

A number of properties have been evaluated

Detector types

Dielectic constant

Amperometric

Conductometric

Polarographic

Potentiometric

Dielectric constant detector

Bulk property detector

Measures changes in polarity of the liquid phase passing through the cell

Conductometric detector

Measures conductivity of the solvent Useful for

solutions of ions

Amperometric detectors

Most frequently applied type of electrochemical

detector

A known potential is applied across a set of

electrodes - typically a glassy carbon type

Ability to oxidize or reduce a species can be

measured

Typically limited to working with a specific class

of materials per analysis

Separation is based on the analytersquos relative solubility between two liquid phases

Stationary PhaseMobile Phase

Solvent Bonded Phase

Normal Phase- Polar stationary phase and non-polar solvent

bull Reverse Phase- Non-polar stationary phase and a polar solvent

Methan0l CH3OH

bull Acetonitrile CH3CN

bull Tetrahydrofuran

bull Water H2O

Solid Support - Backbone for bonded phases Usually 10micro 5micro or 3micro silica or polymeric

particles Bonded Phases - Functional groups firmly linked

(chemically bound) to the solid support Extremely stable Reproducible

Guard - Protects the analytical column Particles Interferences Prolongs the life of the analytical column

bull Analytical - Performs the separation

C-2 Ethyl Silyl -Si-CH2-CH3

bullCN Cyanopropyl Silyl -Si-(CH2)3-CN

bull C-18C-18 Octadecyl SilylOctadecyl Silyl -Si-(CH -Si-(CH22))1717--

CHCH33

bullC-8 Octyl Silyl -Si-(CH2)7-CH3

Pump

Injector

ColumnDetector

Mobile Phases

Gradient Controller

bull

UV Single wavelength (filter) [610 8330] Variable wavelength (monochromator)

[8316 8325] Multiple wavelengths (PDA) [555]

Fluorescence [610]Electrochemical [605] Mass Spectrometric [8325]

Restekreg ULTRA C-18 and CN Columns (250mm x 46mm 5micro)

Mobile Phase (11 MethanolWater) 15 mLmin

Supelcosil LC-PAH Columns A BConditions A 150mm x 46mm 5micro

Flow Rate 15 mLmin

Conditions B 50mm x 46mm 3micro

Flow Rate 30 mLmin

  • Slide 5
  • Slide 6
  • Slide 7
  • Slide 8
  • Slide 9
  • Slide 10
  • Slide 11
  • Slide 12
  • Slide 13
  • Slide 14
  • Slide 15
  • Slide 16
  • Slide 17
  • Slide 18
  • Slide 19
  • Slide 20
  • Slide 21
  • Slide 22
  • Slide 23
  • Slide 24
  • Slide 25
  • Slide 26
  • Slide 27
  • Slide 28
  • Slide 29
  • Slide 30
  • Slide 31
  • Slide 32
  • Slide 33
  • Slide 34
  • Slide 35
  • Slide 36
  • Slide 37
  • Slide 38
  • Slide 39
  • Slide 40
Page 30: P igh erformance L iquid C hromatography  HPLC is a form of liquid chromatography used to separate compounds that are dissolved in solution. HPLC instruments

Detector Systems

Virtually every chemical and physical

property that can be measured in

solution has been look at

Detectors fall roughly into two classes

Bulk property - measures an overall change in the mobile phase

Solute property - measures a solute specific property

Properties of a good detector

A detector must provide

bull high sensitivity low detection limits

bull linearity

bullReproducibility

This is true for any detector

Each detector will have specific advantages and will vary as to peak shape and spread noise and flowtemperature dependence they have

UVVis detector

A solute property detector

Sample must exhibit absorption in UVVis range Solvent must not absorb significantly at the measured wavelength

Types Filter photometer - single

bull Variable wavelength

bull Multiwavelength

Refractive index detector

Bulk property detector - general purpose

Based on refraction of light as it passes from

one media to another Presence of a solute

changes the refractive index of the solvent

Heat of absorption detector

A small amount of heat is released when a sample absorbs on a suitable surface

This detector can measure this

Electrochemical detectors

A number of properties have been evaluated

Detector types

Dielectic constant

Amperometric

Conductometric

Polarographic

Potentiometric

Dielectric constant detector

Bulk property detector

Measures changes in polarity of the liquid phase passing through the cell

Conductometric detector

Measures conductivity of the solvent Useful for

solutions of ions

Amperometric detectors

Most frequently applied type of electrochemical

detector

A known potential is applied across a set of

electrodes - typically a glassy carbon type

Ability to oxidize or reduce a species can be

measured

Typically limited to working with a specific class

of materials per analysis

Separation is based on the analytersquos relative solubility between two liquid phases

Stationary PhaseMobile Phase

Solvent Bonded Phase

Normal Phase- Polar stationary phase and non-polar solvent

bull Reverse Phase- Non-polar stationary phase and a polar solvent

Methan0l CH3OH

bull Acetonitrile CH3CN

bull Tetrahydrofuran

bull Water H2O

Solid Support - Backbone for bonded phases Usually 10micro 5micro or 3micro silica or polymeric

particles Bonded Phases - Functional groups firmly linked

(chemically bound) to the solid support Extremely stable Reproducible

Guard - Protects the analytical column Particles Interferences Prolongs the life of the analytical column

bull Analytical - Performs the separation

C-2 Ethyl Silyl -Si-CH2-CH3

bullCN Cyanopropyl Silyl -Si-(CH2)3-CN

bull C-18C-18 Octadecyl SilylOctadecyl Silyl -Si-(CH -Si-(CH22))1717--

CHCH33

bullC-8 Octyl Silyl -Si-(CH2)7-CH3

Pump

Injector

ColumnDetector

Mobile Phases

Gradient Controller

bull

UV Single wavelength (filter) [610 8330] Variable wavelength (monochromator)

[8316 8325] Multiple wavelengths (PDA) [555]

Fluorescence [610]Electrochemical [605] Mass Spectrometric [8325]

Restekreg ULTRA C-18 and CN Columns (250mm x 46mm 5micro)

Mobile Phase (11 MethanolWater) 15 mLmin

Supelcosil LC-PAH Columns A BConditions A 150mm x 46mm 5micro

Flow Rate 15 mLmin

Conditions B 50mm x 46mm 3micro

Flow Rate 30 mLmin

  • Slide 5
  • Slide 6
  • Slide 7
  • Slide 8
  • Slide 9
  • Slide 10
  • Slide 11
  • Slide 12
  • Slide 13
  • Slide 14
  • Slide 15
  • Slide 16
  • Slide 17
  • Slide 18
  • Slide 19
  • Slide 20
  • Slide 21
  • Slide 22
  • Slide 23
  • Slide 24
  • Slide 25
  • Slide 26
  • Slide 27
  • Slide 28
  • Slide 29
  • Slide 30
  • Slide 31
  • Slide 32
  • Slide 33
  • Slide 34
  • Slide 35
  • Slide 36
  • Slide 37
  • Slide 38
  • Slide 39
  • Slide 40
Page 31: P igh erformance L iquid C hromatography  HPLC is a form of liquid chromatography used to separate compounds that are dissolved in solution. HPLC instruments

Properties of a good detector

A detector must provide

bull high sensitivity low detection limits

bull linearity

bullReproducibility

This is true for any detector

Each detector will have specific advantages and will vary as to peak shape and spread noise and flowtemperature dependence they have

UVVis detector

A solute property detector

Sample must exhibit absorption in UVVis range Solvent must not absorb significantly at the measured wavelength

Types Filter photometer - single

bull Variable wavelength

bull Multiwavelength

Refractive index detector

Bulk property detector - general purpose

Based on refraction of light as it passes from

one media to another Presence of a solute

changes the refractive index of the solvent

Heat of absorption detector

A small amount of heat is released when a sample absorbs on a suitable surface

This detector can measure this

Electrochemical detectors

A number of properties have been evaluated

Detector types

Dielectic constant

Amperometric

Conductometric

Polarographic

Potentiometric

Dielectric constant detector

Bulk property detector

Measures changes in polarity of the liquid phase passing through the cell

Conductometric detector

Measures conductivity of the solvent Useful for

solutions of ions

Amperometric detectors

Most frequently applied type of electrochemical

detector

A known potential is applied across a set of

electrodes - typically a glassy carbon type

Ability to oxidize or reduce a species can be

measured

Typically limited to working with a specific class

of materials per analysis

Separation is based on the analytersquos relative solubility between two liquid phases

Stationary PhaseMobile Phase

Solvent Bonded Phase

Normal Phase- Polar stationary phase and non-polar solvent

bull Reverse Phase- Non-polar stationary phase and a polar solvent

Methan0l CH3OH

bull Acetonitrile CH3CN

bull Tetrahydrofuran

bull Water H2O

Solid Support - Backbone for bonded phases Usually 10micro 5micro or 3micro silica or polymeric

particles Bonded Phases - Functional groups firmly linked

(chemically bound) to the solid support Extremely stable Reproducible

Guard - Protects the analytical column Particles Interferences Prolongs the life of the analytical column

bull Analytical - Performs the separation

C-2 Ethyl Silyl -Si-CH2-CH3

bullCN Cyanopropyl Silyl -Si-(CH2)3-CN

bull C-18C-18 Octadecyl SilylOctadecyl Silyl -Si-(CH -Si-(CH22))1717--

CHCH33

bullC-8 Octyl Silyl -Si-(CH2)7-CH3

Pump

Injector

ColumnDetector

Mobile Phases

Gradient Controller

bull

UV Single wavelength (filter) [610 8330] Variable wavelength (monochromator)

[8316 8325] Multiple wavelengths (PDA) [555]

Fluorescence [610]Electrochemical [605] Mass Spectrometric [8325]

Restekreg ULTRA C-18 and CN Columns (250mm x 46mm 5micro)

Mobile Phase (11 MethanolWater) 15 mLmin

Supelcosil LC-PAH Columns A BConditions A 150mm x 46mm 5micro

Flow Rate 15 mLmin

Conditions B 50mm x 46mm 3micro

Flow Rate 30 mLmin

  • Slide 5
  • Slide 6
  • Slide 7
  • Slide 8
  • Slide 9
  • Slide 10
  • Slide 11
  • Slide 12
  • Slide 13
  • Slide 14
  • Slide 15
  • Slide 16
  • Slide 17
  • Slide 18
  • Slide 19
  • Slide 20
  • Slide 21
  • Slide 22
  • Slide 23
  • Slide 24
  • Slide 25
  • Slide 26
  • Slide 27
  • Slide 28
  • Slide 29
  • Slide 30
  • Slide 31
  • Slide 32
  • Slide 33
  • Slide 34
  • Slide 35
  • Slide 36
  • Slide 37
  • Slide 38
  • Slide 39
  • Slide 40
Page 32: P igh erformance L iquid C hromatography  HPLC is a form of liquid chromatography used to separate compounds that are dissolved in solution. HPLC instruments

UVVis detector

A solute property detector

Sample must exhibit absorption in UVVis range Solvent must not absorb significantly at the measured wavelength

Types Filter photometer - single

bull Variable wavelength

bull Multiwavelength

Refractive index detector

Bulk property detector - general purpose

Based on refraction of light as it passes from

one media to another Presence of a solute

changes the refractive index of the solvent

Heat of absorption detector

A small amount of heat is released when a sample absorbs on a suitable surface

This detector can measure this

Electrochemical detectors

A number of properties have been evaluated

Detector types

Dielectic constant

Amperometric

Conductometric

Polarographic

Potentiometric

Dielectric constant detector

Bulk property detector

Measures changes in polarity of the liquid phase passing through the cell

Conductometric detector

Measures conductivity of the solvent Useful for

solutions of ions

Amperometric detectors

Most frequently applied type of electrochemical

detector

A known potential is applied across a set of

electrodes - typically a glassy carbon type

Ability to oxidize or reduce a species can be

measured

Typically limited to working with a specific class

of materials per analysis

Separation is based on the analytersquos relative solubility between two liquid phases

Stationary PhaseMobile Phase

Solvent Bonded Phase

Normal Phase- Polar stationary phase and non-polar solvent

bull Reverse Phase- Non-polar stationary phase and a polar solvent

Methan0l CH3OH

bull Acetonitrile CH3CN

bull Tetrahydrofuran

bull Water H2O

Solid Support - Backbone for bonded phases Usually 10micro 5micro or 3micro silica or polymeric

particles Bonded Phases - Functional groups firmly linked

(chemically bound) to the solid support Extremely stable Reproducible

Guard - Protects the analytical column Particles Interferences Prolongs the life of the analytical column

bull Analytical - Performs the separation

C-2 Ethyl Silyl -Si-CH2-CH3

bullCN Cyanopropyl Silyl -Si-(CH2)3-CN

bull C-18C-18 Octadecyl SilylOctadecyl Silyl -Si-(CH -Si-(CH22))1717--

CHCH33

bullC-8 Octyl Silyl -Si-(CH2)7-CH3

Pump

Injector

ColumnDetector

Mobile Phases

Gradient Controller

bull

UV Single wavelength (filter) [610 8330] Variable wavelength (monochromator)

[8316 8325] Multiple wavelengths (PDA) [555]

Fluorescence [610]Electrochemical [605] Mass Spectrometric [8325]

Restekreg ULTRA C-18 and CN Columns (250mm x 46mm 5micro)

Mobile Phase (11 MethanolWater) 15 mLmin

Supelcosil LC-PAH Columns A BConditions A 150mm x 46mm 5micro

Flow Rate 15 mLmin

Conditions B 50mm x 46mm 3micro

Flow Rate 30 mLmin

  • Slide 5
  • Slide 6
  • Slide 7
  • Slide 8
  • Slide 9
  • Slide 10
  • Slide 11
  • Slide 12
  • Slide 13
  • Slide 14
  • Slide 15
  • Slide 16
  • Slide 17
  • Slide 18
  • Slide 19
  • Slide 20
  • Slide 21
  • Slide 22
  • Slide 23
  • Slide 24
  • Slide 25
  • Slide 26
  • Slide 27
  • Slide 28
  • Slide 29
  • Slide 30
  • Slide 31
  • Slide 32
  • Slide 33
  • Slide 34
  • Slide 35
  • Slide 36
  • Slide 37
  • Slide 38
  • Slide 39
  • Slide 40
Page 33: P igh erformance L iquid C hromatography  HPLC is a form of liquid chromatography used to separate compounds that are dissolved in solution. HPLC instruments

Refractive index detector

Bulk property detector - general purpose

Based on refraction of light as it passes from

one media to another Presence of a solute

changes the refractive index of the solvent

Heat of absorption detector

A small amount of heat is released when a sample absorbs on a suitable surface

This detector can measure this

Electrochemical detectors

A number of properties have been evaluated

Detector types

Dielectic constant

Amperometric

Conductometric

Polarographic

Potentiometric

Dielectric constant detector

Bulk property detector

Measures changes in polarity of the liquid phase passing through the cell

Conductometric detector

Measures conductivity of the solvent Useful for

solutions of ions

Amperometric detectors

Most frequently applied type of electrochemical

detector

A known potential is applied across a set of

electrodes - typically a glassy carbon type

Ability to oxidize or reduce a species can be

measured

Typically limited to working with a specific class

of materials per analysis

Separation is based on the analytersquos relative solubility between two liquid phases

Stationary PhaseMobile Phase

Solvent Bonded Phase

Normal Phase- Polar stationary phase and non-polar solvent

bull Reverse Phase- Non-polar stationary phase and a polar solvent

Methan0l CH3OH

bull Acetonitrile CH3CN

bull Tetrahydrofuran

bull Water H2O

Solid Support - Backbone for bonded phases Usually 10micro 5micro or 3micro silica or polymeric

particles Bonded Phases - Functional groups firmly linked

(chemically bound) to the solid support Extremely stable Reproducible

Guard - Protects the analytical column Particles Interferences Prolongs the life of the analytical column

bull Analytical - Performs the separation

C-2 Ethyl Silyl -Si-CH2-CH3

bullCN Cyanopropyl Silyl -Si-(CH2)3-CN

bull C-18C-18 Octadecyl SilylOctadecyl Silyl -Si-(CH -Si-(CH22))1717--

CHCH33

bullC-8 Octyl Silyl -Si-(CH2)7-CH3

Pump

Injector

ColumnDetector

Mobile Phases

Gradient Controller

bull

UV Single wavelength (filter) [610 8330] Variable wavelength (monochromator)

[8316 8325] Multiple wavelengths (PDA) [555]

Fluorescence [610]Electrochemical [605] Mass Spectrometric [8325]

Restekreg ULTRA C-18 and CN Columns (250mm x 46mm 5micro)

Mobile Phase (11 MethanolWater) 15 mLmin

Supelcosil LC-PAH Columns A BConditions A 150mm x 46mm 5micro

Flow Rate 15 mLmin

Conditions B 50mm x 46mm 3micro

Flow Rate 30 mLmin

  • Slide 5
  • Slide 6
  • Slide 7
  • Slide 8
  • Slide 9
  • Slide 10
  • Slide 11
  • Slide 12
  • Slide 13
  • Slide 14
  • Slide 15
  • Slide 16
  • Slide 17
  • Slide 18
  • Slide 19
  • Slide 20
  • Slide 21
  • Slide 22
  • Slide 23
  • Slide 24
  • Slide 25
  • Slide 26
  • Slide 27
  • Slide 28
  • Slide 29
  • Slide 30
  • Slide 31
  • Slide 32
  • Slide 33
  • Slide 34
  • Slide 35
  • Slide 36
  • Slide 37
  • Slide 38
  • Slide 39
  • Slide 40
Page 34: P igh erformance L iquid C hromatography  HPLC is a form of liquid chromatography used to separate compounds that are dissolved in solution. HPLC instruments

Heat of absorption detector

A small amount of heat is released when a sample absorbs on a suitable surface

This detector can measure this

Electrochemical detectors

A number of properties have been evaluated

Detector types

Dielectic constant

Amperometric

Conductometric

Polarographic

Potentiometric

Dielectric constant detector

Bulk property detector

Measures changes in polarity of the liquid phase passing through the cell

Conductometric detector

Measures conductivity of the solvent Useful for

solutions of ions

Amperometric detectors

Most frequently applied type of electrochemical

detector

A known potential is applied across a set of

electrodes - typically a glassy carbon type

Ability to oxidize or reduce a species can be

measured

Typically limited to working with a specific class

of materials per analysis

Separation is based on the analytersquos relative solubility between two liquid phases

Stationary PhaseMobile Phase

Solvent Bonded Phase

Normal Phase- Polar stationary phase and non-polar solvent

bull Reverse Phase- Non-polar stationary phase and a polar solvent

Methan0l CH3OH

bull Acetonitrile CH3CN

bull Tetrahydrofuran

bull Water H2O

Solid Support - Backbone for bonded phases Usually 10micro 5micro or 3micro silica or polymeric

particles Bonded Phases - Functional groups firmly linked

(chemically bound) to the solid support Extremely stable Reproducible

Guard - Protects the analytical column Particles Interferences Prolongs the life of the analytical column

bull Analytical - Performs the separation

C-2 Ethyl Silyl -Si-CH2-CH3

bullCN Cyanopropyl Silyl -Si-(CH2)3-CN

bull C-18C-18 Octadecyl SilylOctadecyl Silyl -Si-(CH -Si-(CH22))1717--

CHCH33

bullC-8 Octyl Silyl -Si-(CH2)7-CH3

Pump

Injector

ColumnDetector

Mobile Phases

Gradient Controller

bull

UV Single wavelength (filter) [610 8330] Variable wavelength (monochromator)

[8316 8325] Multiple wavelengths (PDA) [555]

Fluorescence [610]Electrochemical [605] Mass Spectrometric [8325]

Restekreg ULTRA C-18 and CN Columns (250mm x 46mm 5micro)

Mobile Phase (11 MethanolWater) 15 mLmin

Supelcosil LC-PAH Columns A BConditions A 150mm x 46mm 5micro

Flow Rate 15 mLmin

Conditions B 50mm x 46mm 3micro

Flow Rate 30 mLmin

  • Slide 5
  • Slide 6
  • Slide 7
  • Slide 8
  • Slide 9
  • Slide 10
  • Slide 11
  • Slide 12
  • Slide 13
  • Slide 14
  • Slide 15
  • Slide 16
  • Slide 17
  • Slide 18
  • Slide 19
  • Slide 20
  • Slide 21
  • Slide 22
  • Slide 23
  • Slide 24
  • Slide 25
  • Slide 26
  • Slide 27
  • Slide 28
  • Slide 29
  • Slide 30
  • Slide 31
  • Slide 32
  • Slide 33
  • Slide 34
  • Slide 35
  • Slide 36
  • Slide 37
  • Slide 38
  • Slide 39
  • Slide 40
Page 35: P igh erformance L iquid C hromatography  HPLC is a form of liquid chromatography used to separate compounds that are dissolved in solution. HPLC instruments

Electrochemical detectors

A number of properties have been evaluated

Detector types

Dielectic constant

Amperometric

Conductometric

Polarographic

Potentiometric

Dielectric constant detector

Bulk property detector

Measures changes in polarity of the liquid phase passing through the cell

Conductometric detector

Measures conductivity of the solvent Useful for

solutions of ions

Amperometric detectors

Most frequently applied type of electrochemical

detector

A known potential is applied across a set of

electrodes - typically a glassy carbon type

Ability to oxidize or reduce a species can be

measured

Typically limited to working with a specific class

of materials per analysis

Separation is based on the analytersquos relative solubility between two liquid phases

Stationary PhaseMobile Phase

Solvent Bonded Phase

Normal Phase- Polar stationary phase and non-polar solvent

bull Reverse Phase- Non-polar stationary phase and a polar solvent

Methan0l CH3OH

bull Acetonitrile CH3CN

bull Tetrahydrofuran

bull Water H2O

Solid Support - Backbone for bonded phases Usually 10micro 5micro or 3micro silica or polymeric

particles Bonded Phases - Functional groups firmly linked

(chemically bound) to the solid support Extremely stable Reproducible

Guard - Protects the analytical column Particles Interferences Prolongs the life of the analytical column

bull Analytical - Performs the separation

C-2 Ethyl Silyl -Si-CH2-CH3

bullCN Cyanopropyl Silyl -Si-(CH2)3-CN

bull C-18C-18 Octadecyl SilylOctadecyl Silyl -Si-(CH -Si-(CH22))1717--

CHCH33

bullC-8 Octyl Silyl -Si-(CH2)7-CH3

Pump

Injector

ColumnDetector

Mobile Phases

Gradient Controller

bull

UV Single wavelength (filter) [610 8330] Variable wavelength (monochromator)

[8316 8325] Multiple wavelengths (PDA) [555]

Fluorescence [610]Electrochemical [605] Mass Spectrometric [8325]

Restekreg ULTRA C-18 and CN Columns (250mm x 46mm 5micro)

Mobile Phase (11 MethanolWater) 15 mLmin

Supelcosil LC-PAH Columns A BConditions A 150mm x 46mm 5micro

Flow Rate 15 mLmin

Conditions B 50mm x 46mm 3micro

Flow Rate 30 mLmin

  • Slide 5
  • Slide 6
  • Slide 7
  • Slide 8
  • Slide 9
  • Slide 10
  • Slide 11
  • Slide 12
  • Slide 13
  • Slide 14
  • Slide 15
  • Slide 16
  • Slide 17
  • Slide 18
  • Slide 19
  • Slide 20
  • Slide 21
  • Slide 22
  • Slide 23
  • Slide 24
  • Slide 25
  • Slide 26
  • Slide 27
  • Slide 28
  • Slide 29
  • Slide 30
  • Slide 31
  • Slide 32
  • Slide 33
  • Slide 34
  • Slide 35
  • Slide 36
  • Slide 37
  • Slide 38
  • Slide 39
  • Slide 40
Page 36: P igh erformance L iquid C hromatography  HPLC is a form of liquid chromatography used to separate compounds that are dissolved in solution. HPLC instruments

Dielectric constant detector

Bulk property detector

Measures changes in polarity of the liquid phase passing through the cell

Conductometric detector

Measures conductivity of the solvent Useful for

solutions of ions

Amperometric detectors

Most frequently applied type of electrochemical

detector

A known potential is applied across a set of

electrodes - typically a glassy carbon type

Ability to oxidize or reduce a species can be

measured

Typically limited to working with a specific class

of materials per analysis

Separation is based on the analytersquos relative solubility between two liquid phases

Stationary PhaseMobile Phase

Solvent Bonded Phase

Normal Phase- Polar stationary phase and non-polar solvent

bull Reverse Phase- Non-polar stationary phase and a polar solvent

Methan0l CH3OH

bull Acetonitrile CH3CN

bull Tetrahydrofuran

bull Water H2O

Solid Support - Backbone for bonded phases Usually 10micro 5micro or 3micro silica or polymeric

particles Bonded Phases - Functional groups firmly linked

(chemically bound) to the solid support Extremely stable Reproducible

Guard - Protects the analytical column Particles Interferences Prolongs the life of the analytical column

bull Analytical - Performs the separation

C-2 Ethyl Silyl -Si-CH2-CH3

bullCN Cyanopropyl Silyl -Si-(CH2)3-CN

bull C-18C-18 Octadecyl SilylOctadecyl Silyl -Si-(CH -Si-(CH22))1717--

CHCH33

bullC-8 Octyl Silyl -Si-(CH2)7-CH3

Pump

Injector

ColumnDetector

Mobile Phases

Gradient Controller

bull

UV Single wavelength (filter) [610 8330] Variable wavelength (monochromator)

[8316 8325] Multiple wavelengths (PDA) [555]

Fluorescence [610]Electrochemical [605] Mass Spectrometric [8325]

Restekreg ULTRA C-18 and CN Columns (250mm x 46mm 5micro)

Mobile Phase (11 MethanolWater) 15 mLmin

Supelcosil LC-PAH Columns A BConditions A 150mm x 46mm 5micro

Flow Rate 15 mLmin

Conditions B 50mm x 46mm 3micro

Flow Rate 30 mLmin

  • Slide 5
  • Slide 6
  • Slide 7
  • Slide 8
  • Slide 9
  • Slide 10
  • Slide 11
  • Slide 12
  • Slide 13
  • Slide 14
  • Slide 15
  • Slide 16
  • Slide 17
  • Slide 18
  • Slide 19
  • Slide 20
  • Slide 21
  • Slide 22
  • Slide 23
  • Slide 24
  • Slide 25
  • Slide 26
  • Slide 27
  • Slide 28
  • Slide 29
  • Slide 30
  • Slide 31
  • Slide 32
  • Slide 33
  • Slide 34
  • Slide 35
  • Slide 36
  • Slide 37
  • Slide 38
  • Slide 39
  • Slide 40
Page 37: P igh erformance L iquid C hromatography  HPLC is a form of liquid chromatography used to separate compounds that are dissolved in solution. HPLC instruments

Amperometric detectors

Most frequently applied type of electrochemical

detector

A known potential is applied across a set of

electrodes - typically a glassy carbon type

Ability to oxidize or reduce a species can be

measured

Typically limited to working with a specific class

of materials per analysis

Separation is based on the analytersquos relative solubility between two liquid phases

Stationary PhaseMobile Phase

Solvent Bonded Phase

Normal Phase- Polar stationary phase and non-polar solvent

bull Reverse Phase- Non-polar stationary phase and a polar solvent

Methan0l CH3OH

bull Acetonitrile CH3CN

bull Tetrahydrofuran

bull Water H2O

Solid Support - Backbone for bonded phases Usually 10micro 5micro or 3micro silica or polymeric

particles Bonded Phases - Functional groups firmly linked

(chemically bound) to the solid support Extremely stable Reproducible

Guard - Protects the analytical column Particles Interferences Prolongs the life of the analytical column

bull Analytical - Performs the separation

C-2 Ethyl Silyl -Si-CH2-CH3

bullCN Cyanopropyl Silyl -Si-(CH2)3-CN

bull C-18C-18 Octadecyl SilylOctadecyl Silyl -Si-(CH -Si-(CH22))1717--

CHCH33

bullC-8 Octyl Silyl -Si-(CH2)7-CH3

Pump

Injector

ColumnDetector

Mobile Phases

Gradient Controller

bull

UV Single wavelength (filter) [610 8330] Variable wavelength (monochromator)

[8316 8325] Multiple wavelengths (PDA) [555]

Fluorescence [610]Electrochemical [605] Mass Spectrometric [8325]

Restekreg ULTRA C-18 and CN Columns (250mm x 46mm 5micro)

Mobile Phase (11 MethanolWater) 15 mLmin

Supelcosil LC-PAH Columns A BConditions A 150mm x 46mm 5micro

Flow Rate 15 mLmin

Conditions B 50mm x 46mm 3micro

Flow Rate 30 mLmin

  • Slide 5
  • Slide 6
  • Slide 7
  • Slide 8
  • Slide 9
  • Slide 10
  • Slide 11
  • Slide 12
  • Slide 13
  • Slide 14
  • Slide 15
  • Slide 16
  • Slide 17
  • Slide 18
  • Slide 19
  • Slide 20
  • Slide 21
  • Slide 22
  • Slide 23
  • Slide 24
  • Slide 25
  • Slide 26
  • Slide 27
  • Slide 28
  • Slide 29
  • Slide 30
  • Slide 31
  • Slide 32
  • Slide 33
  • Slide 34
  • Slide 35
  • Slide 36
  • Slide 37
  • Slide 38
  • Slide 39
  • Slide 40
Page 38: P igh erformance L iquid C hromatography  HPLC is a form of liquid chromatography used to separate compounds that are dissolved in solution. HPLC instruments

Separation is based on the analytersquos relative solubility between two liquid phases

Stationary PhaseMobile Phase

Solvent Bonded Phase

Normal Phase- Polar stationary phase and non-polar solvent

bull Reverse Phase- Non-polar stationary phase and a polar solvent

Methan0l CH3OH

bull Acetonitrile CH3CN

bull Tetrahydrofuran

bull Water H2O

Solid Support - Backbone for bonded phases Usually 10micro 5micro or 3micro silica or polymeric

particles Bonded Phases - Functional groups firmly linked

(chemically bound) to the solid support Extremely stable Reproducible

Guard - Protects the analytical column Particles Interferences Prolongs the life of the analytical column

bull Analytical - Performs the separation

C-2 Ethyl Silyl -Si-CH2-CH3

bullCN Cyanopropyl Silyl -Si-(CH2)3-CN

bull C-18C-18 Octadecyl SilylOctadecyl Silyl -Si-(CH -Si-(CH22))1717--

CHCH33

bullC-8 Octyl Silyl -Si-(CH2)7-CH3

Pump

Injector

ColumnDetector

Mobile Phases

Gradient Controller

bull

UV Single wavelength (filter) [610 8330] Variable wavelength (monochromator)

[8316 8325] Multiple wavelengths (PDA) [555]

Fluorescence [610]Electrochemical [605] Mass Spectrometric [8325]

Restekreg ULTRA C-18 and CN Columns (250mm x 46mm 5micro)

Mobile Phase (11 MethanolWater) 15 mLmin

Supelcosil LC-PAH Columns A BConditions A 150mm x 46mm 5micro

Flow Rate 15 mLmin

Conditions B 50mm x 46mm 3micro

Flow Rate 30 mLmin

  • Slide 5
  • Slide 6
  • Slide 7
  • Slide 8
  • Slide 9
  • Slide 10
  • Slide 11
  • Slide 12
  • Slide 13
  • Slide 14
  • Slide 15
  • Slide 16
  • Slide 17
  • Slide 18
  • Slide 19
  • Slide 20
  • Slide 21
  • Slide 22
  • Slide 23
  • Slide 24
  • Slide 25
  • Slide 26
  • Slide 27
  • Slide 28
  • Slide 29
  • Slide 30
  • Slide 31
  • Slide 32
  • Slide 33
  • Slide 34
  • Slide 35
  • Slide 36
  • Slide 37
  • Slide 38
  • Slide 39
  • Slide 40
Page 39: P igh erformance L iquid C hromatography  HPLC is a form of liquid chromatography used to separate compounds that are dissolved in solution. HPLC instruments

Normal Phase- Polar stationary phase and non-polar solvent

bull Reverse Phase- Non-polar stationary phase and a polar solvent

Methan0l CH3OH

bull Acetonitrile CH3CN

bull Tetrahydrofuran

bull Water H2O

Solid Support - Backbone for bonded phases Usually 10micro 5micro or 3micro silica or polymeric

particles Bonded Phases - Functional groups firmly linked

(chemically bound) to the solid support Extremely stable Reproducible

Guard - Protects the analytical column Particles Interferences Prolongs the life of the analytical column

bull Analytical - Performs the separation

C-2 Ethyl Silyl -Si-CH2-CH3

bullCN Cyanopropyl Silyl -Si-(CH2)3-CN

bull C-18C-18 Octadecyl SilylOctadecyl Silyl -Si-(CH -Si-(CH22))1717--

CHCH33

bullC-8 Octyl Silyl -Si-(CH2)7-CH3

Pump

Injector

ColumnDetector

Mobile Phases

Gradient Controller

bull

UV Single wavelength (filter) [610 8330] Variable wavelength (monochromator)

[8316 8325] Multiple wavelengths (PDA) [555]

Fluorescence [610]Electrochemical [605] Mass Spectrometric [8325]

Restekreg ULTRA C-18 and CN Columns (250mm x 46mm 5micro)

Mobile Phase (11 MethanolWater) 15 mLmin

Supelcosil LC-PAH Columns A BConditions A 150mm x 46mm 5micro

Flow Rate 15 mLmin

Conditions B 50mm x 46mm 3micro

Flow Rate 30 mLmin

  • Slide 5
  • Slide 6
  • Slide 7
  • Slide 8
  • Slide 9
  • Slide 10
  • Slide 11
  • Slide 12
  • Slide 13
  • Slide 14
  • Slide 15
  • Slide 16
  • Slide 17
  • Slide 18
  • Slide 19
  • Slide 20
  • Slide 21
  • Slide 22
  • Slide 23
  • Slide 24
  • Slide 25
  • Slide 26
  • Slide 27
  • Slide 28
  • Slide 29
  • Slide 30
  • Slide 31
  • Slide 32
  • Slide 33
  • Slide 34
  • Slide 35
  • Slide 36
  • Slide 37
  • Slide 38
  • Slide 39
  • Slide 40
Page 40: P igh erformance L iquid C hromatography  HPLC is a form of liquid chromatography used to separate compounds that are dissolved in solution. HPLC instruments

Methan0l CH3OH

bull Acetonitrile CH3CN

bull Tetrahydrofuran

bull Water H2O

Solid Support - Backbone for bonded phases Usually 10micro 5micro or 3micro silica or polymeric

particles Bonded Phases - Functional groups firmly linked

(chemically bound) to the solid support Extremely stable Reproducible

Guard - Protects the analytical column Particles Interferences Prolongs the life of the analytical column

bull Analytical - Performs the separation

C-2 Ethyl Silyl -Si-CH2-CH3

bullCN Cyanopropyl Silyl -Si-(CH2)3-CN

bull C-18C-18 Octadecyl SilylOctadecyl Silyl -Si-(CH -Si-(CH22))1717--

CHCH33

bullC-8 Octyl Silyl -Si-(CH2)7-CH3

Pump

Injector

ColumnDetector

Mobile Phases

Gradient Controller

bull

UV Single wavelength (filter) [610 8330] Variable wavelength (monochromator)

[8316 8325] Multiple wavelengths (PDA) [555]

Fluorescence [610]Electrochemical [605] Mass Spectrometric [8325]

Restekreg ULTRA C-18 and CN Columns (250mm x 46mm 5micro)

Mobile Phase (11 MethanolWater) 15 mLmin

Supelcosil LC-PAH Columns A BConditions A 150mm x 46mm 5micro

Flow Rate 15 mLmin

Conditions B 50mm x 46mm 3micro

Flow Rate 30 mLmin

  • Slide 5
  • Slide 6
  • Slide 7
  • Slide 8
  • Slide 9
  • Slide 10
  • Slide 11
  • Slide 12
  • Slide 13
  • Slide 14
  • Slide 15
  • Slide 16
  • Slide 17
  • Slide 18
  • Slide 19
  • Slide 20
  • Slide 21
  • Slide 22
  • Slide 23
  • Slide 24
  • Slide 25
  • Slide 26
  • Slide 27
  • Slide 28
  • Slide 29
  • Slide 30
  • Slide 31
  • Slide 32
  • Slide 33
  • Slide 34
  • Slide 35
  • Slide 36
  • Slide 37
  • Slide 38
  • Slide 39
  • Slide 40
Page 41: P igh erformance L iquid C hromatography  HPLC is a form of liquid chromatography used to separate compounds that are dissolved in solution. HPLC instruments

Solid Support - Backbone for bonded phases Usually 10micro 5micro or 3micro silica or polymeric

particles Bonded Phases - Functional groups firmly linked

(chemically bound) to the solid support Extremely stable Reproducible

Guard - Protects the analytical column Particles Interferences Prolongs the life of the analytical column

bull Analytical - Performs the separation

C-2 Ethyl Silyl -Si-CH2-CH3

bullCN Cyanopropyl Silyl -Si-(CH2)3-CN

bull C-18C-18 Octadecyl SilylOctadecyl Silyl -Si-(CH -Si-(CH22))1717--

CHCH33

bullC-8 Octyl Silyl -Si-(CH2)7-CH3

Pump

Injector

ColumnDetector

Mobile Phases

Gradient Controller

bull

UV Single wavelength (filter) [610 8330] Variable wavelength (monochromator)

[8316 8325] Multiple wavelengths (PDA) [555]

Fluorescence [610]Electrochemical [605] Mass Spectrometric [8325]

Restekreg ULTRA C-18 and CN Columns (250mm x 46mm 5micro)

Mobile Phase (11 MethanolWater) 15 mLmin

Supelcosil LC-PAH Columns A BConditions A 150mm x 46mm 5micro

Flow Rate 15 mLmin

Conditions B 50mm x 46mm 3micro

Flow Rate 30 mLmin

  • Slide 5
  • Slide 6
  • Slide 7
  • Slide 8
  • Slide 9
  • Slide 10
  • Slide 11
  • Slide 12
  • Slide 13
  • Slide 14
  • Slide 15
  • Slide 16
  • Slide 17
  • Slide 18
  • Slide 19
  • Slide 20
  • Slide 21
  • Slide 22
  • Slide 23
  • Slide 24
  • Slide 25
  • Slide 26
  • Slide 27
  • Slide 28
  • Slide 29
  • Slide 30
  • Slide 31
  • Slide 32
  • Slide 33
  • Slide 34
  • Slide 35
  • Slide 36
  • Slide 37
  • Slide 38
  • Slide 39
  • Slide 40
Page 42: P igh erformance L iquid C hromatography  HPLC is a form of liquid chromatography used to separate compounds that are dissolved in solution. HPLC instruments

C-2 Ethyl Silyl -Si-CH2-CH3

bullCN Cyanopropyl Silyl -Si-(CH2)3-CN

bull C-18C-18 Octadecyl SilylOctadecyl Silyl -Si-(CH -Si-(CH22))1717--

CHCH33

bullC-8 Octyl Silyl -Si-(CH2)7-CH3

Pump

Injector

ColumnDetector

Mobile Phases

Gradient Controller

bull

UV Single wavelength (filter) [610 8330] Variable wavelength (monochromator)

[8316 8325] Multiple wavelengths (PDA) [555]

Fluorescence [610]Electrochemical [605] Mass Spectrometric [8325]

Restekreg ULTRA C-18 and CN Columns (250mm x 46mm 5micro)

Mobile Phase (11 MethanolWater) 15 mLmin

Supelcosil LC-PAH Columns A BConditions A 150mm x 46mm 5micro

Flow Rate 15 mLmin

Conditions B 50mm x 46mm 3micro

Flow Rate 30 mLmin

  • Slide 5
  • Slide 6
  • Slide 7
  • Slide 8
  • Slide 9
  • Slide 10
  • Slide 11
  • Slide 12
  • Slide 13
  • Slide 14
  • Slide 15
  • Slide 16
  • Slide 17
  • Slide 18
  • Slide 19
  • Slide 20
  • Slide 21
  • Slide 22
  • Slide 23
  • Slide 24
  • Slide 25
  • Slide 26
  • Slide 27
  • Slide 28
  • Slide 29
  • Slide 30
  • Slide 31
  • Slide 32
  • Slide 33
  • Slide 34
  • Slide 35
  • Slide 36
  • Slide 37
  • Slide 38
  • Slide 39
  • Slide 40
Page 43: P igh erformance L iquid C hromatography  HPLC is a form of liquid chromatography used to separate compounds that are dissolved in solution. HPLC instruments

Pump

Injector

ColumnDetector

Mobile Phases

Gradient Controller

bull

UV Single wavelength (filter) [610 8330] Variable wavelength (monochromator)

[8316 8325] Multiple wavelengths (PDA) [555]

Fluorescence [610]Electrochemical [605] Mass Spectrometric [8325]

Restekreg ULTRA C-18 and CN Columns (250mm x 46mm 5micro)

Mobile Phase (11 MethanolWater) 15 mLmin

Supelcosil LC-PAH Columns A BConditions A 150mm x 46mm 5micro

Flow Rate 15 mLmin

Conditions B 50mm x 46mm 3micro

Flow Rate 30 mLmin

  • Slide 5
  • Slide 6
  • Slide 7
  • Slide 8
  • Slide 9
  • Slide 10
  • Slide 11
  • Slide 12
  • Slide 13
  • Slide 14
  • Slide 15
  • Slide 16
  • Slide 17
  • Slide 18
  • Slide 19
  • Slide 20
  • Slide 21
  • Slide 22
  • Slide 23
  • Slide 24
  • Slide 25
  • Slide 26
  • Slide 27
  • Slide 28
  • Slide 29
  • Slide 30
  • Slide 31
  • Slide 32
  • Slide 33
  • Slide 34
  • Slide 35
  • Slide 36
  • Slide 37
  • Slide 38
  • Slide 39
  • Slide 40
Page 44: P igh erformance L iquid C hromatography  HPLC is a form of liquid chromatography used to separate compounds that are dissolved in solution. HPLC instruments

UV Single wavelength (filter) [610 8330] Variable wavelength (monochromator)

[8316 8325] Multiple wavelengths (PDA) [555]

Fluorescence [610]Electrochemical [605] Mass Spectrometric [8325]

Restekreg ULTRA C-18 and CN Columns (250mm x 46mm 5micro)

Mobile Phase (11 MethanolWater) 15 mLmin

Supelcosil LC-PAH Columns A BConditions A 150mm x 46mm 5micro

Flow Rate 15 mLmin

Conditions B 50mm x 46mm 3micro

Flow Rate 30 mLmin

  • Slide 5
  • Slide 6
  • Slide 7
  • Slide 8
  • Slide 9
  • Slide 10
  • Slide 11
  • Slide 12
  • Slide 13
  • Slide 14
  • Slide 15
  • Slide 16
  • Slide 17
  • Slide 18
  • Slide 19
  • Slide 20
  • Slide 21
  • Slide 22
  • Slide 23
  • Slide 24
  • Slide 25
  • Slide 26
  • Slide 27
  • Slide 28
  • Slide 29
  • Slide 30
  • Slide 31
  • Slide 32
  • Slide 33
  • Slide 34
  • Slide 35
  • Slide 36
  • Slide 37
  • Slide 38
  • Slide 39
  • Slide 40
Page 45: P igh erformance L iquid C hromatography  HPLC is a form of liquid chromatography used to separate compounds that are dissolved in solution. HPLC instruments

Restekreg ULTRA C-18 and CN Columns (250mm x 46mm 5micro)

Mobile Phase (11 MethanolWater) 15 mLmin

Supelcosil LC-PAH Columns A BConditions A 150mm x 46mm 5micro

Flow Rate 15 mLmin

Conditions B 50mm x 46mm 3micro

Flow Rate 30 mLmin

  • Slide 5
  • Slide 6
  • Slide 7
  • Slide 8
  • Slide 9
  • Slide 10
  • Slide 11
  • Slide 12
  • Slide 13
  • Slide 14
  • Slide 15
  • Slide 16
  • Slide 17
  • Slide 18
  • Slide 19
  • Slide 20
  • Slide 21
  • Slide 22
  • Slide 23
  • Slide 24
  • Slide 25
  • Slide 26
  • Slide 27
  • Slide 28
  • Slide 29
  • Slide 30
  • Slide 31
  • Slide 32
  • Slide 33
  • Slide 34
  • Slide 35
  • Slide 36
  • Slide 37
  • Slide 38
  • Slide 39
  • Slide 40
Page 46: P igh erformance L iquid C hromatography  HPLC is a form of liquid chromatography used to separate compounds that are dissolved in solution. HPLC instruments

Supelcosil LC-PAH Columns A BConditions A 150mm x 46mm 5micro

Flow Rate 15 mLmin

Conditions B 50mm x 46mm 3micro

Flow Rate 30 mLmin

  • Slide 5
  • Slide 6
  • Slide 7
  • Slide 8
  • Slide 9
  • Slide 10
  • Slide 11
  • Slide 12
  • Slide 13
  • Slide 14
  • Slide 15
  • Slide 16
  • Slide 17
  • Slide 18
  • Slide 19
  • Slide 20
  • Slide 21
  • Slide 22
  • Slide 23
  • Slide 24
  • Slide 25
  • Slide 26
  • Slide 27
  • Slide 28
  • Slide 29
  • Slide 30
  • Slide 31
  • Slide 32
  • Slide 33
  • Slide 34
  • Slide 35
  • Slide 36
  • Slide 37
  • Slide 38
  • Slide 39
  • Slide 40