32
Illinois Instruments, Inc (U.S.) 2401 Hiller Ridge Road Johnsburg Illinois IL 60051 Systech Instruments Ltd 17 Thame Park Business Centre Wenman Road Thame Oxfordshire OX9 3XA E: [email protected] E: [email protected] www.systechillinois.com MODEL 8501 OXYGEN PERMEATION ANALYSER OPERATIONS MANUAL Version 2.3 8 th February 2013

MODEL 8501 OXYGEN PERMEATION ANALYSER ... 8501 Operations Manual v2.3 Page 1 1.0 INTRODUCTION Thank you for your purchase. The Model 8501 Oxygen Permeation Analyzer has been designed

  • Upload
    vannhi

  • View
    234

  • Download
    1

Embed Size (px)

Citation preview

Page 1: MODEL 8501 OXYGEN PERMEATION ANALYSER ... 8501 Operations Manual v2.3 Page 1 1.0 INTRODUCTION Thank you for your purchase. The Model 8501 Oxygen Permeation Analyzer has been designed

Illinois Instruments, Inc (U.S.) 2401 Hiller Ridge Road

Johnsburg Illinois IL

60051

Systech Instruments Ltd 17 Thame Park Business Centre

Wenman Road Thame

Oxfordshire OX9 3XA

E: [email protected] E: [email protected]

www.systechillinois.com

MODEL 8501 OXYGEN PERMEATION

ANALYSER

OPERATIONS MANUAL Version 2.3 8th February 2013

Page 2: MODEL 8501 OXYGEN PERMEATION ANALYSER ... 8501 Operations Manual v2.3 Page 1 1.0 INTRODUCTION Thank you for your purchase. The Model 8501 Oxygen Permeation Analyzer has been designed

MODEL 8501 Operations Manual v2.3

www.systechillinois.com

Table of Contents

1.0 INTRODUCTION ........................................................................................................................... 1

2.0 PRINCIPLE OF OPERATION ....................................................................................................... 2

3.0 INSTALLATION AND ADJUSTMENT ......................................................................................... 3 3.1 Voltage Selector/Fuse Cartridge ...................................................................................................... 3 3.2 Connecting Gas Lines ...................................................................................................................... 4

4.0 USING THE INSTRUMENT .......................................................................................................... 5 4.1 Front Panel Operation ...................................................................................................................... 5 4.2 Front Panel Guide ............................................................................................................................ 6 4.3 Data Table Of Programmable Functions .......................................................................................... 7

5.0 SAMPLE PREPARATION AND TESTING ................................................................................... 8 5.1 Film Testing ...................................................................................................................................... 8

6.0 CALIBRATION .............................................................................................................................. 9 6.1 Gas Calibration ................................................................................................................................ 9 6.2 System Validation / Calibration using Calibration films. ................................................................. 11

7.0 RS232 COMMUNICATIONS....................................................................................................... 12 7.1 Installing Software .......................................................................................................................... 12 7.2 Creating a Desktop Icon ................................................................................................................. 15 7.3 Commands for use with the Model 8501 ........................................................................................ 15 7.4 Data Logging .................................................................................................................................. 16

8.0 THE CONTROLLER SOFTWARE ............................................................................................. 17 8.1 Installing the software .................................................................................................................... 17 8.2 Creating a Desktop Shortcut. ......................................................................................................... 20 8.3 Running the software ..................................................................................................................... 20 8.4 Advanced Mode ............................................................................................................................. 23 8.5 Graph Viewer ................................................................................................................................. 26

9.0 ACCESSORIES AND REPLACEMENT PARTS........................................................................ 27

10.0 MAINTENANCE AND WARRANTY ......................................................................................... 28 10.1 Maintenance ................................................................................................................................. 28 10.2 Warranty....................................................................................................................................... 28

11.0 TECHNICAL SPECIFICATIONS .............................................................................................. 29

12.0 RETURN SHIPMENT INSTRUCTIONS .................................................................................... 30 Revision History

Issue Issue Date Changes By

2.1 20-05-2010 Clarified the Warranty statement Nick Stuart

2.2 19-09-2011 Added extra section in the calibration process, Rebranded the manual.

Nick Stuart

2.3 08-02-2013 Removed section 7.0, added section 6.2, changed input gas pressure requirements

Victor Kofman

Page 3: MODEL 8501 OXYGEN PERMEATION ANALYSER ... 8501 Operations Manual v2.3 Page 1 1.0 INTRODUCTION Thank you for your purchase. The Model 8501 Oxygen Permeation Analyzer has been designed

MODEL 8501 Operations Manual v2.3

www.systechillinois.com Page

1

1.0 INTRODUCTION Thank you for your purchase. The Model 8501 Oxygen Permeation Analyzer has been designed to provide accurate and uncomplicated measurements of oxygen transmission rates (OTR) through flat film barriers. The instrument is capable of measuring films with an OTR greater than 1 cc/m

2/day (0.07 cc/100 in

2/day).

We have incorporated into the instrument an electrochemical sensor of our own development and the finest grade electronic components to compliment its performance. This manual is arranged in a manner we feel will provide the greatest ease of reference and use. We are proud of our active program of research and development and wish to know if you encounter any difficulties in your application.

This symbol is known as the ‘Crossed-out Wheelie Bin Symbol’. When this symbol is marked on a product it means that consideration should be given to the disposal of the product, parts or accessories. Only discard electrical/electronic items in separate collection schemes which cater for the recovery and recycling of the materials contained within. Your co-operation is vital to ensure the success of these schemes and for the protection of the environment.

IMPORTANT Please read this manual before attempting to install or operate the equipment. In some variations the sample gas in this equipment is heated to very high temperatures and therefore no combustibles should be present at a limit to cause explosions. The equipment should be electrically connected and grounded in accordance with good standard practice and at the correct voltage. (Refer to the voltage tag label on the back panel.)

No responsibility is accepted by Systech Instruments Limited for accidents resulting from improper use of this equipment.

Page 4: MODEL 8501 OXYGEN PERMEATION ANALYSER ... 8501 Operations Manual v2.3 Page 1 1.0 INTRODUCTION Thank you for your purchase. The Model 8501 Oxygen Permeation Analyzer has been designed

MODEL 8501 Operations Manual v2.3

www.systechillinois.com Page

2

2.0 PRINCIPLE OF OPERATION The Model 8501 uses our proprietary sensor to detect oxygen transmission rates through flat film barriers. Flat film samples are clamped in a diffusion chamber (Figure 1). Pure oxygen (99.9%) is introduced into the upper half of the sample chamber while an oxygen-free carrier gas, such as nitrogen, flows through the lower half of the chamber. Molecules of oxygen diffuse through the film into the lower chamber and are conveyed to the sensor by the carrier gas. This allows for direct measurement of the oxygen without using complex extrapolations. The oxygen transmission rate of the test film is displayed as either cc/100in

2/day or

cc/m2/day.

Figure 1. Cross-sectional view of the diffusion chamber.

Page 5: MODEL 8501 OXYGEN PERMEATION ANALYSER ... 8501 Operations Manual v2.3 Page 1 1.0 INTRODUCTION Thank you for your purchase. The Model 8501 Oxygen Permeation Analyzer has been designed

MODEL 8501 Operations Manual v2.3

www.systechillinois.com Page

3

3.0 INSTALLATION AND ADJUSTMENT

3.1 Voltage Selector/Fuse Cartridge The power entry module on the back of the instrument serves many purposes. It not only allows attachment of the power cord, but also contains the fuses and allows the supply voltage of the instrument to be selected for either 110/120 VAC or 220/240 VAC operation. Changing the supply voltage requires removal of the cartridge located just below the IEC receptacle for the power cord. To determine the set voltage range for the instrument, view the instrument from the back and note the voltage range that reads upright to the bottom edge of the cartridge. This range is the range to which the instrument is set. The small white arrow beneath the voltage statement should also align with the white arrow that is stamped upon the module body. To change the supply voltage range, remove the power cord (if connected) and insert a small, straight screwdriver into the slot on the top edge of the cartridge. Pry the cartridge away from the instrument until it can be grasped by the fingers and removed completely. Then simply flip the cartridge over and press it back into place to alter the voltage range. Refer to Figure 2 below.

110-120 VAC setting 220-240 VAC setting

Figure 2. Selecting the proper supply voltage. Note that the cartridge also contains two fuses. Both of these fuses must be installed to ensure proper operation of the instrument. The fuses are rated 5A @ 250 V and are a slow-blow type. Replace only with fuses of equivalent ratings.

Page 6: MODEL 8501 OXYGEN PERMEATION ANALYSER ... 8501 Operations Manual v2.3 Page 1 1.0 INTRODUCTION Thank you for your purchase. The Model 8501 Oxygen Permeation Analyzer has been designed

MODEL 8501 Operations Manual v2.3

www.systechillinois.com Page

4

3.2 Connecting Gas Lines For best results, oxygen-free nitrogen (<5 ppm oxygen) and pure oxygen (99.9% or better) should be used for analysis. These gases should not contain any hydrogen. Use of gases containing hydrogen will result in sensor damage. An oxygen trap should be used to remove traces of oxygen from the nitrogen. Oxygen traps can be purchased through Systech Instruments. NOTE: The inlet pressure of both gases must be set between 1.4 – 1.6 bar, (20 - 23 PSI). If the supply pressure is not within this range, then erroneous measurements will result. A drawing of the back panel is located below (Figure 3). The back panel consists of four brass fittings for gas connections. There are two fittings for both oxygen and nitrogen. These fittings are marked VENTS and INPUTS. Connect the nitrogen to the NITROGEN INPUT and oxygen to the OXYGEN INPUT. The fittings are

1/8 inch Swagelok type compression fittings. Do not over tighten the nuts when connecting the gas lines to the inlet and outlet fittings. The gas vents should be free of restrictions to flow or any source of back pressure.

Figure 3. Model 8501 back panel.

Oxygen Trap

Nitrogen Input (Note the sealing plug) Oxygen Input

Mains Input

Pipes to Oxygen Trap

Page 7: MODEL 8501 OXYGEN PERMEATION ANALYSER ... 8501 Operations Manual v2.3 Page 1 1.0 INTRODUCTION Thank you for your purchase. The Model 8501 Oxygen Permeation Analyzer has been designed

MODEL 8501 Operations Manual v2.3

www.systechillinois.com Page

5

4.0 USING THE INSTRUMENT

4.1 Front Panel Operation

Figure 4. Model 8501 front panel. The instrument is controlled through the LED display and the six digit keypad or via the RS232 communication port. When the instrument is powered up, it displays “Otr” in the left window and “0.” in the right window. Allow the unit 30 minutes to warm-up prior to running any samples.

To access the programming mode, press the “Program Scroll” button:

When pressed, the LED above it turns on. When the instrument is in the programming mode, if no button is pressed for 5 minutes, the instrument will revert to the normal measurement mode. The following table lists the functions of the buttons and indicator lights.

Page 8: MODEL 8501 OXYGEN PERMEATION ANALYSER ... 8501 Operations Manual v2.3 Page 1 1.0 INTRODUCTION Thank you for your purchase. The Model 8501 Oxygen Permeation Analyzer has been designed

MODEL 8501 Operations Manual v2.3

www.systechillinois.com Page

6

4.2 Front Panel Guide

COMPONENT FUNCTION OPERATION

SCROLL Used to scroll through the programming modes to change the instrument set-up.

Press to select the configuration parameter to be changed. Each press brings up the next parameter.

SELECT Selects display digit to be changed.

Selection is indicated by flashing digit. First press selects the leftmost digit. Successive presses selects all digits, decimal points, and LEDs.

INCREMENT Increments selected display digit.

Each press steps selected digit through complete numerical range (0 – 9).

ENTER Enters new data into memory for that function.

Press once to reset the display. After 1 second the display shows “store” to indicate it has accepted the entry.

CALIBRATE Starts calibration routine.

Press once to begin calibration routine.

Press during test or calibrate to “abort” the routine.

TEST Starts test routine.

Press once to begin test routine.

Press during test or calibrate to “abort” the routine.

PGM LED When on, indicates instrument is in programming mode.

CAL LED Indicates when instrument is in calibration mode.

TEST LED Indicates when instrument is in testing mode.

STATUS LED Indicates status of instrument. Default is “Idle”. Changes to “Purge” and “Test” when entering the purge and test routines.

UNITS LED Indicates the units of measure. Display reads in either cc/100in

2/day

or cc/m2/day.

% LED Display measured valued is percent oxygen by volume.

Active during calibration routine only.

ppm LED Display measured value is parts per million oxygen by volume.

Active during calibration routine only.

Page 9: MODEL 8501 OXYGEN PERMEATION ANALYSER ... 8501 Operations Manual v2.3 Page 1 1.0 INTRODUCTION Thank you for your purchase. The Model 8501 Oxygen Permeation Analyzer has been designed

MODEL 8501 Operations Manual v2.3

www.systechillinois.com Page

7

4.3 Data Table Of Programmable Functions

DISPLAY FUNCTION

Unit Sets the units of measure for the displayed oxygen transmission rate.

“0” selects cc/100in2/day.

“1” selects cc/m2/day.

tESt Sets the length of time for the test routine in minutes.

CAL Sets the calibration of the analyzer to the correct value to agree with a known sample gas or certified cylinder.

LIFE Indicates the life left in the sensor as a percentage.

CtSP Sets temperature for test chamber

Otr Indicates the current OTR value.

Use the following procedure to change the instrument’s configuration. Enter the set-up mode by pressing the “PROGRAM SCROLL” button on the front panel of the instrument. Continue pressing the “PROGRAM SCROLL” button until the parameter you wish to change appears on the display. Set the new value for the parameter using the “DIGIT SELECT” and “DIGIT INCREMENT” buttons. After setting the new value, press the “ENTER” button. The display will show “STORE” to indicate the change has been accepted. In order to prevent accidental changes to instrument parameters, a digit must be flashing before the value is stored. NOTE: Always press the “DIGIT SELECT” button before pressing “ENTER” in order to store changes. Pressing “ENTER” when the display is not flashing will not change the displayed parameter, but will cause you to exit the menu. To exit the configuration mode, press the “PROGRAM SCROLL” button until Otr is displayed.

Page 10: MODEL 8501 OXYGEN PERMEATION ANALYSER ... 8501 Operations Manual v2.3 Page 1 1.0 INTRODUCTION Thank you for your purchase. The Model 8501 Oxygen Permeation Analyzer has been designed

MODEL 8501 Operations Manual v2.3

www.systechillinois.com Page

8

5.0 SAMPLE PREPARATION AND TESTING A representative sample of film should be selected for analysis. Ideally, this sample should be free of wrinkles, creases, or folds. It is recommended that film sample be handled with care. Handle film samples only by the edges. Fingerprints can alter the transmission rates of some films. A circular template is included with the instrument. Before using the circular template, place the supplied cutting pad on a flat, solid surface. Put the film sample on the cutting pad and hold the template on top of the film sample. Carefully cut around the perimeter of the template using a sharp utility knife. Remove the top portion of the sample chamber by unscrewing the chamber knobs. These knobs are constructed such that they cannot be removed completely from the chamber cover. Apply a small amount of grease to the rim of the lower sample chamber. Spread a thin coat evenly around the edge about ½ inch wide. Do not apply any grease to the upper chamber or the o-ring on the chamber cover. Carefully place the film sample on the sample chamber. Be sure not to get grease on the area of the film being tested. Press the film into the grease to create a tight seal. Pull the film tight to remove any wrinkles. Replace the chamber cover and tighten the chamber knobs.

5.1 Film Testing Connect oxygen-free nitrogen and pure oxygen to the INPUT ports on the back of the instrument as described under Section 3.2, Connecting the Gas Lines.

1. Remove the sample chamber cover by turning the three knobs counter clockwise until they are loose. The knobs are constructed such that they cannot be removed from the chamber cover. Therefore, the knobs only need to be loosened in order for the cover to be removed.

2. Prepare the chamber and sample as described above.

3. Replace the sample chamber cover and tighten the knobs a little at a time until each is

uniformly snug.

4. Check the “tESt” parameter setting before starting the test. Many films reach steady-state levels with “tESt” set to 30 minutes.

5. Regulate the pressure between 1.4 – 1.6 Bar, (20 – 23 psi) to set the flow rate to 35cc/min for

both nitrogen and oxygen. Use an external flow meter to measure the flow rates on the nitrogen and oxygen VENTS on the back of the unit. The flow on the “B” VENT usually will be 5-12 cc/min.

6. Press the green “TEST” button to start the test. The status indicator LED will move from the

“Idle” position to the “Purge” position.

NOTE: The test may be stopped at any time by pressing either the “CAL” button or the “TEST” button.

7. After about 30 minutes, the status light will move from “Purge” to “Test”.

The instrument will stay in “Test” mode for the length of time specified by the “tESt” parameter. After this time has expired, the OTR value will flash on and off. This is the final OTR value for the film being tested. The display will continue flashing until the “Enter” button is pressed

Page 11: MODEL 8501 OXYGEN PERMEATION ANALYSER ... 8501 Operations Manual v2.3 Page 1 1.0 INTRODUCTION Thank you for your purchase. The Model 8501 Oxygen Permeation Analyzer has been designed

MODEL 8501 Operations Manual v2.3

www.systechillinois.com Page

9

6.0 CALIBRATION

6.1 Gas Calibration It is recommended that the Model 8501 be calibrated to a certified gas containing approximately 100 ppm oxygen in nitrogen. NOTE: The provided metal plate must be placed in the sample chamber before the calibration routine is started. Systech Instruments suggests a minimum of quarterly calibration. Use the following procedure for calibrating the unit.

1. Locate in the supplied accessory box the following items.

2. Turn off and disconnect all gas lines from the back of the instrument.

3. Referring to the picture below disconnect the pipes leading to the Oxygen Trap.

4. Fit the Oxygen Trap plugs into the open ends of the pipes leading to the Oxygen Trap.

Oxygen Trap Plugs Oxygen Trap Bypass

Disconnect Oxygen Trap Here

Page 12: MODEL 8501 OXYGEN PERMEATION ANALYSER ... 8501 Operations Manual v2.3 Page 1 1.0 INTRODUCTION Thank you for your purchase. The Model 8501 Oxygen Permeation Analyzer has been designed

MODEL 8501 Operations Manual v2.3

www.systechillinois.com Page

10

5. Fit the Oxygen Trap bypass pipe to the Oxygen Trap connections on the rear of the instrument.

6. Connect the calibration gas to the NITROGEN INPUT port on the back of the instrument. Regulate the pressure between 1.4 – 1.6 Bar, (20 - 23 psi) to set the flow rate to 35cc/min. Use an external flow meter to measure the flow rate on the VENT of NITROGEN on the back of the unit.

7. Change the “CAL” parameter in the menu to match the certified level of oxygen in your

calibration gas.

8. Place the provided metal plate in the chamber as described under Section 4.2, Sample Preparation

9. Press the blue “CAL” button on the front panel. The status indicator LED will switch from

“Idle” to “Purge”.

10. After about 30 minutes, the “Purge” light will turn off. At this time the oxygen readings in the left window should be stabilizing.

11. Press the “Enter” button after the readings stabilize to calibrate the instrument. The

instrument is now calibrated.

12. Shut off the calibration gas at the cylinder regulator and remove the pipe from the rear of the instrument.

13. Remove the Oxygen Trap bypass pipe and refit the pipes leading to the Oxygen trap, onto the instrument trap connections.

14. Reconnect the Oxygen and Nitrogen supply pipes. NOTE: Ensure these are connected to the correct place as serious damage can be inflicted on the instrument if the Oxygen is connected to the nitrogen input.

15. The instrument should now be ready for use. NOTE: In order to cancel the calibration routine at any time, press either the “CAL” button or the “TEST” button; “ABORT” will appear on the display.

Page 13: MODEL 8501 OXYGEN PERMEATION ANALYSER ... 8501 Operations Manual v2.3 Page 1 1.0 INTRODUCTION Thank you for your purchase. The Model 8501 Oxygen Permeation Analyzer has been designed

MODEL 8501 Operations Manual v2.3

www.systechillinois.com Page

11

6.2 System Validation / Calibration using Calibration films. Model 8501 is supplied with certified Calibration film (PCF-850101) that could be used for system validation or calibration. To validate the system by using Calibration films follow section 5.1 with test time “tESt” set to 30 min. The system is considered to have passed validation if the measured OTR value is within +/- 5% from the film certified value. Calibration film could also be used to Calibrate the Model 8501 if calibration gas is not available. To calibrate the system by using Calibration set test time “tESt” to 90 min. Follow the below sequence to calibrate your instrument:

1. Place the Calibration film in the chamber as described under Section 4.2, Sample Preparation.

2. Press the green “CAL” button on the front panel. The status indicator LED will switch from

“Idle” to “Purge”.

3. After about 35 minutes, the “Purge” light will turn off and “Test” light will turn on. At this time the OTR readings in the right window should start changing.

4. Wait 50-60 minutes for OTR reading to stabilize and start Illinois Client software.

5. In “Output (to instrument)” window key in “SPAN?1” and push “send”. Read value in “Reply

(from instrument)” window.

6. Calculate the “simulated span” coefficient by dividing certified calibration film OTR value by actual OTR value on display.

7. Calculate a “New SPAN 1” value by multiplying an actual “Span?1” value you have in Illinois

Client by the “simulated span” coefficient.

8. Enter a “New SPAN 1” value in “Output (to instrument)” window in Illinois Client software by keying in “span=1 x.xxx, where x.xxx the “New SPAN 1” value and push “send”, Reply from instrument should be “0”. OTR value on display should change proportionally to the “simulated span” coefficient.

9. Press the “Enter” button to finish the calibration. The instrument is now calibrated.

Example: If your Calibration film has a Certified value of 62.3 cc/m2*day and your actual OTR reading is 65.3 cc/m2*day then your “simulated span” coefficient is 62.3 / 65.3 = 0.95. Read the “span?1” value (usually between 0.02 and 0.04). Lets say it is 0.0312. Multiply 0.0312 by 0.95 (0.0312*0.95 = 0.02964). It will be a “New Span 1” value. Note: Systech Illinois recommends using Gas Calibration to maintain better system accuracy.

Page 14: MODEL 8501 OXYGEN PERMEATION ANALYSER ... 8501 Operations Manual v2.3 Page 1 1.0 INTRODUCTION Thank you for your purchase. The Model 8501 Oxygen Permeation Analyzer has been designed

MODEL 8501 Operations Manual v2.3

www.systechillinois.com Page

12

7.0 RS232 COMMUNICATIONS Systech Client or SYS_Client is a communications program specifically designed to interface using serial RS232 communications with a variety of Systech Instruments Products. It can very useful to diagnose / repair and setup the instruments. However it should be noted that it can also cause damage to the Instrument if used incorrectly. The following details the installation process of the software and the commands available for various instruments. When completed connect the serial port of the computer to the 9 pin D-sub connector on the back panel of the Model 8501.

7.1 Installing Software If the software has been supplied on a Memory stick, insert the device into a spare USB port. The computer should recognise the device as a disk drive; this may take a few moments. When completed you should be able to explore the drive using explorer. Double click on autorun.exe in the root of the disk drive. If your software is supplied on a CD then the disk should autorun as normal, however if it does not then explore the disk as before. You should then get the following screen (below); click on Install Comms Client Utility:-

The installation program then starts with the following screen:-

Page 15: MODEL 8501 OXYGEN PERMEATION ANALYSER ... 8501 Operations Manual v2.3 Page 1 1.0 INTRODUCTION Thank you for your purchase. The Model 8501 Oxygen Permeation Analyzer has been designed

MODEL 8501 Operations Manual v2.3

www.systechillinois.com Page

13

After a short amount of time the following screen is displayed:-

Click on OK, then the screen will change to this:-

Click on the button shown above to install the client software to the specified destination directory. You can at this point select a different folder; however Systech Instruments do not recommend this. You will then be presented with the following screen:-

Page 16: MODEL 8501 OXYGEN PERMEATION ANALYSER ... 8501 Operations Manual v2.3 Page 1 1.0 INTRODUCTION Thank you for your purchase. The Model 8501 Oxygen Permeation Analyzer has been designed

MODEL 8501 Operations Manual v2.3

www.systechillinois.com Page

14

This screen allows you to select or change the program group, Systech recommends that at this point you accept the default setting and click on Continue. Now the software will commence its installation process with the following screen:-

The Blue bar shown indicates the installation progress. When it has reached 100% (which only takes a few minutes) you will be presented with this screen, which indicates that the software has been installed satisfactorily. If during the installation phase you are presented with a screen that informs you that you are trying to replace a file on your computer with one that is an older version than the one you have, always keep the newer file.

Page 17: MODEL 8501 OXYGEN PERMEATION ANALYSER ... 8501 Operations Manual v2.3 Page 1 1.0 INTRODUCTION Thank you for your purchase. The Model 8501 Oxygen Permeation Analyzer has been designed

MODEL 8501 Operations Manual v2.3

www.systechillinois.com Page

15

7.2 Creating a Desktop Icon If you wish to have an icon on the desktop to click in order to start the Sys_Client follow the following steps:- Go to “Start”, “All Programs”, “Systech Client”, and right click on Systech Client. Next click on “Send to” and then click on “Desktop (create shortcut) as shown below. There will then be an icon placed on your desktop to invoke the Sys_Client program

7.3 Commands for use with the Model 8501 The following table lists the commands available for the Model 8501.

Command Function

? Returns OTR reading

UNIT? Returns “0” for cc/100in2/day and “1” for cc/m

2/day

UNIT= Sets the UNIT value

TEMP? Reads sample chamber temperature

VER? Reads VERSION number of software

For example, use the following steps to view the sample chamber temperature. First click on the “Output (to Instrument)” text box. Type temp? and click “Send”. The answer from the instrument will be displayed in the “Reply (from Instrument)” text box.

Page 18: MODEL 8501 OXYGEN PERMEATION ANALYSER ... 8501 Operations Manual v2.3 Page 1 1.0 INTRODUCTION Thank you for your purchase. The Model 8501 Oxygen Permeation Analyzer has been designed

MODEL 8501 Operations Manual v2.3

www.systechillinois.com Page

16

7.4 Data Logging Data logging is possible using the RS232 communications software. The following procedure can be used to save the OTR values from a test routine to a file.

1. Type “?” in the “Output (to Instrument)” text box. This will request the OTR value from the Model 8501.

2. Click on the “Interval” text box and set the rate at which OTR values should be collected. To

sample every minute, type “00:01:00” into the text box.

3. Click on the “Inc Time/Date” box to store the time and date along with the OTR value.

4. Click the “Save Replies to File” box and then click on the “Change File” button. Enter the name of the file to which you want to store the OTR data and click on the “Save” button.

5. Begin testing procedures as explained in Section 4.4 Film Testing. After pressing the “TEST”

button, click on the “Start” button to begin collecting data.

6. Once the test has completed click “Stop” button, the software will save the data to the file specified. This data file will be in a format that can be directly imported into a spreadsheet program.

Note: Process interface software is available directly from Systech Instruments for this analyser. Via RS232 connection, remote set-up and querying of analyzer can be achieved. Also, this software has data acquisition and graphing capabilities. For further information, please contact Systech Instruments’ Sales Department at +44 (0)1844 216838 or email [email protected] or [email protected]

Page 19: MODEL 8501 OXYGEN PERMEATION ANALYSER ... 8501 Operations Manual v2.3 Page 1 1.0 INTRODUCTION Thank you for your purchase. The Model 8501 Oxygen Permeation Analyzer has been designed

MODEL 8501 Operations Manual v2.3

www.systechillinois.com Page

17

8.0 THE CONTROLLER SOFTWARE

8.1 Installing the software This software is designed for your instrument in order to facilitate and automate its use. Please read this documentation carefully as it explains the use of the software and provides troubleshooting tips. The software is certified for use with Windows NT, 98 and XP, but not currently Windows Vista. If the software has been supplied on a Memory stick, insert the device into a spare USB port. The computer should recognise the device as a disk drive; this may take a few moments. When completed you should be able to explore the drive using explorer. Double click on autorun.exe in the root of the disk drive. If your software is supplied on a CD then the disk should autorun as normal, however if it does not then explore the disk as before. You should then get the following screen (below); click on Install M8501 software:-

After a short period of time you should get the following screen appear:-

Page 20: MODEL 8501 OXYGEN PERMEATION ANALYSER ... 8501 Operations Manual v2.3 Page 1 1.0 INTRODUCTION Thank you for your purchase. The Model 8501 Oxygen Permeation Analyzer has been designed

MODEL 8501 Operations Manual v2.3

www.systechillinois.com Page

18

Click on “Install. Again then you should get the following screen appear:-

Systech instruments recommend that in each case you use the default settings, until finally you get the following screen appear:-

Page 21: MODEL 8501 OXYGEN PERMEATION ANALYSER ... 8501 Operations Manual v2.3 Page 1 1.0 INTRODUCTION Thank you for your purchase. The Model 8501 Oxygen Permeation Analyzer has been designed

MODEL 8501 Operations Manual v2.3

www.systechillinois.com Page

19

This indicates that installation was a success. NOTE: After installing the “Controller” software please copy file “8500.ini” from the Software\Software Version 1.0.13\Drivers installation folder to directory C:\Controller\Drivers\, replacing and overwriting the existing 8500.ini file. This can be done via the “Browse Memory Stick / CD” option on the main software screen

Page 22: MODEL 8501 OXYGEN PERMEATION ANALYSER ... 8501 Operations Manual v2.3 Page 1 1.0 INTRODUCTION Thank you for your purchase. The Model 8501 Oxygen Permeation Analyzer has been designed

MODEL 8501 Operations Manual v2.3

www.systechillinois.com Page

20

8.2 Creating a Desktop Shortcut. To facilitate easier access to the software you may like to add a shortcut to the computers desktop. To do this click on Start / Programs / Instrument Controller the right click on Controller and select “send to” and click on Desktop Create Shortcut. As shown below:-

This will the place a “Controller Icon on the desktop. Repeat with Graph Viewer if required.

8.3 Running the software Ensure that the correct communications cable is connected between the computer and the analyser. Please note to which COM port the cable is connected. Turn on your instrument and give it a few seconds to warm up. Start the Controller software by double clicking on the desktop icon created in 9.2. If you have the correct “COM Port” selected and the cables are plugged in correctly, a screen will appear looking similar to this. It shows the current “Properties” of the instrument, as well as, a box for the parameters being monitored i.e. Oxygen Permeation and Chamber Temperature

Page 23: MODEL 8501 OXYGEN PERMEATION ANALYSER ... 8501 Operations Manual v2.3 Page 1 1.0 INTRODUCTION Thank you for your purchase. The Model 8501 Oxygen Permeation Analyzer has been designed

MODEL 8501 Operations Manual v2.3

www.systechillinois.com Page

21

If there is a problem establishing communications, an error message will be displayed similar to that shown on the next page. Press “Yes” in order to select the correct port (if the correct port is selected, check the cables).

The following box will then appear on your screen:-

After selecting the correct port in the window that pops up and/or making sure the cables are connected, press “OK”. The CONTROLLER will now try again to establish a connection. Note the “Baud should always be 9600. Once the connection is established, the CONTROLLER Software will create a “Log. File” for troubleshooting proposes. As shown below.

You can now monitor the values in the parameter window and check the properties of the instrument. (You CANNOT change the properties at this time as this could easily corrupt the calibration. If you need to change them, look in the Advanced Mode section)

Page 24: MODEL 8501 OXYGEN PERMEATION ANALYSER ... 8501 Operations Manual v2.3 Page 1 1.0 INTRODUCTION Thank you for your purchase. The Model 8501 Oxygen Permeation Analyzer has been designed

MODEL 8501 Operations Manual v2.3

www.systechillinois.com Page

22

By pressing the little graph button

(Show/Hide Graph)

on the top right of the “parameter window”, you can open another window that will start graphing the values. (

This window can be minimised, but closing it will stop the graphing).

In the graph window, you can press the "Start Saving to File" button in order to start saving the graph to a file.

The following appears:

Page 25: MODEL 8501 OXYGEN PERMEATION ANALYSER ... 8501 Operations Manual v2.3 Page 1 1.0 INTRODUCTION Thank you for your purchase. The Model 8501 Oxygen Permeation Analyzer has been designed

MODEL 8501 Operations Manual v2.3

www.systechillinois.com Page

23

Enter the folder and file name and press “Save”. You will be asked for delay time between data points on the graph. As shown below:-

NOTE: When using the “Start Saving to File” option above, some problems will occur if the “Delay time” contains a value outside its limits. If the “Cancel” button or any character other than a number is entered the controller software will close down.

Enter desired delay time and press “OK”. The graph will be recorded at the given intervals until you press "Stop Saving To File" button or close the graph.

8.4 Advanced Mode Once you have a working instrument and you need to change values in the “Properties Window”, press the “Advanced” button in the large toolbar at the top.

Page 26: MODEL 8501 OXYGEN PERMEATION ANALYSER ... 8501 Operations Manual v2.3 Page 1 1.0 INTRODUCTION Thank you for your purchase. The Model 8501 Oxygen Permeation Analyzer has been designed

MODEL 8501 Operations Manual v2.3

www.systechillinois.com Page

24

You will be asked for a password. The default is "password".

After entering the correct password you will have access to all the properties. Refer to other sections of this manual regarding properties values. Once you have finished, we recommend that you go back to standard mode, by pushing “Advance Mode” button again, to prevent accidental modification of the properties.

If any properties are changed from the front panel of the analyser, it will be necessary to push the “Stop” then “Start” buttons on the CONTROLLER toolbar to reset the software.

Page 27: MODEL 8501 OXYGEN PERMEATION ANALYSER ... 8501 Operations Manual v2.3 Page 1 1.0 INTRODUCTION Thank you for your purchase. The Model 8501 Oxygen Permeation Analyzer has been designed

MODEL 8501 Operations Manual v2.3

www.systechillinois.com Page

25

You can change the “Advanced Mode” password by using the “Settings” button:

Settings screen will appear: By pushing on “Change Advanced Mode Password” button you will be asked to enter and confirm a new password

To exit CONTROLLER click on “ Exit” button.

Page 28: MODEL 8501 OXYGEN PERMEATION ANALYSER ... 8501 Operations Manual v2.3 Page 1 1.0 INTRODUCTION Thank you for your purchase. The Model 8501 Oxygen Permeation Analyzer has been designed

MODEL 8501 Operations Manual v2.3

www.systechillinois.com Page

26

8.5 Graph Viewer You can view and print saved files using Graph Viewer located in CONTROLLER group in Start menu. Or accessed by the desktop shortcut created in 9.2

Click on “file” and “open” to select a data file previously created. From here again using the “File” menu you can print the graph out to an attached printer.

Page 29: MODEL 8501 OXYGEN PERMEATION ANALYSER ... 8501 Operations Manual v2.3 Page 1 1.0 INTRODUCTION Thank you for your purchase. The Model 8501 Oxygen Permeation Analyzer has been designed

MODEL 8501 Operations Manual v2.3

www.systechillinois.com Page

27

9.0 ACCESSORIES AND REPLACEMENT PARTS Systech Instruments, make a number of useful accessories and replacement parts available to assist in applications and servicing. When ordering, please specify the model and serial number of your instrument and the description of items required as listed below:

Description Part Number

Sensor (fuel cell) 900 021

Sensor (fuel cell) for units build after 2013

851 055

Large Capacity Oxygen Trap 080 050

Film Sealant 085 055

Page 30: MODEL 8501 OXYGEN PERMEATION ANALYSER ... 8501 Operations Manual v2.3 Page 1 1.0 INTRODUCTION Thank you for your purchase. The Model 8501 Oxygen Permeation Analyzer has been designed

MODEL 8501 Operations Manual v2.3

www.systechillinois.com Page

28

10.0 MAINTENANCE AND WARRANTY

10.1 Maintenance The Model 8501 Oxygen Permeation Analyzers are trouble-free and need no routine maintenance. If the message “Cerr” is displayed when you are attempting to calibrate the instrument, then the sensor is in need of replacement. Complete installation instructions are included with each replacement sensor. If you prefer, Systech Illinois can perform this task for a nominal charge.

10.2 Warranty This Instrument is guaranteed for a period of one year from its delivery to the purchaser covering faulty workmanship and replacement of defective parts. This assumes fair wear and tear and usage specified on the data sheet. It does not cover routine calibration and housekeeping. The Warranty is on a Return To Base (RTB) basis. However under exceptional circumstances this may be extended to cover onsite warranty. This may involve extra costs to the end user. We maintain comprehensive after sales facilities and the instrument should be returned to our factory for repairing or servicing if this is necessary. The type and serial number should always be quoted together with full details of any fault. Warranty repair or replacement will be made without charge noting the above comments. Equipment must be shipped prepaid after return authorisation has been obtained. All service and technical enquiries are covered from our factories in Thame, Oxfordshire, and Johnsburg, Illinois where we will endeavour to give a quick and helpful response to all queries. The factory addresses are:

Illinois Instruments, Inc (U.S.) 2401 Hiller Ridge Road, Johnsburg Illinois IL 60051 Tel: +1 815 344 6212

Systech Instruments Ltd 17 Thame Park Business Centre Wenman Road Thame Oxfordshire OX9 3XA Tel: +44 (0) 1844 216838

For customers outside UK or USA, please contact us for details of your local representative. Alternatively check out our website:

www.systechillinois.com

Page 31: MODEL 8501 OXYGEN PERMEATION ANALYSER ... 8501 Operations Manual v2.3 Page 1 1.0 INTRODUCTION Thank you for your purchase. The Model 8501 Oxygen Permeation Analyzer has been designed

MODEL 8501 Operations Manual v2.3

www.systechillinois.com Page

29

11.0 TECHNICAL SPECIFICATIONS

Measurement range: 1 - 99,999 cc/m2/day

(0.07 - 68,000 cc/100in2/day )

Test temperature range: 15°C - 40°C (59°F – 104°F)

Response time: 90% within 20 minutes or better depending on barrier film

Supply pressure: 1.4 – 1.6 Bar, 20 – 23 psi, regulated (oxygen and nitrogen)

Gas fittings: 1/8 in Swagelok (supplied)

Displays: 4 digit LED for calibration and 5 digit LED for measurement

Sample thickness: up to 1.27 mm (0.050 in)

Sample test area: 100 cm2

Operating conditions: standard laboratory environments

Power requirements: 110/220 VAC, 50 – 60 Hz

Enclosure: epoxy coated heavy gauge steel

Dimensions, mm (in): 356 W (14) x 356 D (14) x 279 H (11)

Weight, kg (lb): 18.2 (40)

Page 32: MODEL 8501 OXYGEN PERMEATION ANALYSER ... 8501 Operations Manual v2.3 Page 1 1.0 INTRODUCTION Thank you for your purchase. The Model 8501 Oxygen Permeation Analyzer has been designed

MODEL 8501 Operations Manual v2.3

www.systechillinois.com Page

30

12.0 RETURN SHIPMENT INSTRUCTIONS Before packing the permeation analyzer can you please carry out the following:-

Shutdown the Analyzer software.

Turn OFF the oxygen supply and disconnect the pipes from the Oxygen Source Inlet.

Locate the blanking nut in the white "Start UP" box (This is to protect the oxygen trap from being consumed by air whilst disconnected from the nitrogen supply).

Turn OFF the nitrogen supply and disconnect the pipe from the Nitrogen Source Inlet and fit the blanking nut to the analyzer inlet.

Disconnect the RS232 (if installed) and power cable.

If you have the original packing and crate can you please put the analyzer in it together with its foam protection. If the original shipping container is not available then please ensure that the instrument is packed in a suitable enclosure with a minimum of 50mm (2 inches) of packing material all round and under the instrument. Failure to follow these guidelines could result in extra expense due to incurring shipping damage.

We do not require the pipes, Start Up box or any cables.

The brass plate should be cleaned and placed in the chamber with the covers closed down (we need to check the condition of the plates).

Please then arrange with a suitable carrier for the analyzer to be returned to your local Systech Illinois office.

Illinois Instruments, Inc (U.S.) 2401 Hiller Ridge Road, Johnsburg Illinois IL 60051 Tel: +1 815 344 6212

Systech Instruments Ltd 17 Thame Park Business Centre Wenman Road Thame Oxfordshire OX9 3XA Tel: +44 (0) 1844 216838