Magnetic PMC

Embed Size (px)

DESCRIPTION

Polymer composite with magnetic particle

Citation preview

  • 1

    University Of Glasgow

    School of Engineering

    Department of Mechanical Engineering

    Final Year Project

    Manufacture and Characterisation of Magneto-Rheological Elastomers

    By Mr Bruce Miller

    Supervised by Dr Philip Harrison and Ms Gerlind Schubert

    1 Abstract

    This report deals with the manufacturing, testing and characterisation of Magneto

    Rheological Elastomers (MREs). MREs are a set of smart composite materials consisting of

    an elastomeric matrix with magnetic particles dispersed in said matrix. When a magnetic

    field is applied the stiffness of the material changes instantaneously and reversibly. This is

    due to the interaction of the magnetic particles dispersed in the matrix. This report deals

    with the manufacturing processes and how to optimize them for the best mechanical

    properties and most time efficient methods. Furthermore the method by which these

    materials are compression tested while a uniform magnetic field is applied will be dealt

    with. Finally the results from experiments will be used to compare the mechanical

    properties of MREs with different volume percentages of magnetic particles, samples with

    Isotropic and Anisotropic particle structures and finally samples with applied magnetic

    fields. These results will be used in future work to generate a general constitutive model for

    these materials.

    List of Objecticves:

    Design a test setup to create uniform magnetic field on samples while compression

    testing

    Compression tests on samples with different curing times to optimize manufacturing

    technique

    Compression Tests on anisotropic and isotropic samples with different Volume

    percentage of Carbonyl magnetic particles (10%,20%,30%)

    Compression tests on anisotropic and isotropic samples under the influence of a

    magnetic field

    Model material behaviour using Abaqus FEA

  • 2

    Table of Contents

    List of Figures ........................................................................................................................................ 3

    List of Equations.................................................................................................................................... 3

    List of Tables ......................................................................................................................................... 4

    Glossary of Nomenclature .................................................................................................................... 4

    Acknoledgements ................................................................................................................................. 5

    Introduction .......................................................................................................................................... 5

    Background Information ....................................................................................................................... 5

    Manufacturing Process ......................................................................................................................... 6

    Test Method ......................................................................................................................................... 9

    Test Setup ........................................................................................................................................... 10

    Test Results ......................................................................................................................................... 13

    Pure Rubber, 10%, 20%, 30% Isotropic samples cured for 24 Hours at 25oC .................................. 14

    Compression of Pure Rubber with different curing conditions ....................................................... 17

    Compression of 10% volume Isotropic samples with different curing conditions .......................... 18

    Compression of 10%, 20% and 30% Anisotropic samples ............................................................... 19

    Compression of 10%, 20% and 30%, Anisotropic and Isotropic Samples with applied magnetic

    fields ............................................................................................................................................... 22

    10% Isotropic Samples ................................................................................................................ 23

    20% Isotropic Samples ................................................................................................................ 24

    30% Isotropic Samples ................................................................................................................ 25

    10% Anisotrpic Samples .............................................................................................................. 26

    20% Anisotropic Samples ............................................................................................................ 27

    30% Anisotropic Samples ............................................................................................................ 28

    Modelling of MREs .............................................................................................................................. 29

    Modelling of 10% Isotropic Samples ........................................................................................... 30

    Modelling of 20% Isotropic Samples ........................................................................................... 34

    Modelling of 30% Isotropic Samples ........................................................................................... 38

    Modelling of Pure Rubber ........................................................................................................... 42

    Conclusions ......................................................................................................................................... 43

    Bibliography ........................................................................................................................................ 44

    Appendix A : Calculation for Solenoids ............................................................................................... 45

    Appendix B: Faulty test setup results .................................................................................................. 47

  • 3

    List of Figures

    Figure 1 Picture of Heating Plates ......................................................................................................... 8

    Figure 2 Picture of Electromagnet with mould in between poles ......................................................... 9

    Figure 3 Picture of Electromagnet ........................................................................................................ 9

    Figure 4 Original Magnetic Setup ........................................................................................................ 11

    Figure 5 Revised Magnetic Setup ........................................................................................................ 11

    Figure 6 Rendered Drawing of Test Setup........................................................................................... 12

    Figure 7 Picture of Test Setup ............................................................................................................. 12

    Figure 8 ............................................................................................................................................... 14

    Figure 9 Microscopic Picture of 10% Isotropic Sample ....................................................................... 15

    Figure 10 Microscopic Picture of 20% Isotropic Sample ..................................................................... 15

    Figure 11 Graph of all Isotropic Sample Configurations ...................................................................... 16

    Figure 12 Graph of Pure Rubber with different Curing Conditions ..................................................... 17

    Figure 13 Graph of 10%v CIP samples with different curing conditions.............................................. 18

    Figure 14 Microscopic Picture of 10% Anisotropic Sample ................................................................. 19

    Figure 15 Buckling Samples ................................................................................................................. 20

    Figure 16 Graph of all volume percentages of CIP for Anisotropic and Isotropic structures ............... 21

    Figure 17 Graph of 10% Isotropic CIP Samples with different applied magnetic fields ....................... 23

    Figure 18 Graph of 20% Isotropic CIP Samples with different applied magnetic fields ....................... 24

    Figure 19 Graph of 30% Isotropic Samples with different applied magnetic fields ............................. 25

    Figure 20 Graph of 30% Anisotropic Samples with different applied magnetic fields ........................ 26

    Figure 21Graph of 20% Anisotropic samples with different applied magnetic fields .......................... 27

    Figure 22 Graph of 30% Anisotropic samples with different applied magnetic fields ......................... 28

    Figure 23 10% Isotropic Samples with 400mT applied magnetic field ................................................ 30

    Figure 24 10% Isotropic Samples with 270mT applied magnetic field ................................................ 31

    Figure 25 10% Isotropic samples no applied magnetic field ............................................................... 32

    Figure 26 20% Isotropic Sample with 400mT applied magnetic field .................................................. 34

    Figure 27 20% Isotropic Samples with 270mT applied magnetic field ................................................ 35

    Figure 28 20% Isotropic Samples with no applied magnetic field ....................................................... 36

    Figure 29 30% Isotropic Samples with 400mT applied magnetic field ................................................ 38

    Figure 30 30% Isotropic Samples with 270mT applied magnetic field ................................................ 39

    Figure 31 30% Isotropic Samples with no applied field ....................................................................... 40

    Figure 32 Pure Rubber Modelling Curves............................................................................................ 42

    List of Equations

    Equation 1 Equation for Mass of CIP[] ................................................................................................... 8

    Equation 2 Standard Ogden Model ..................................................................................................... 29

    Equation 3 Standard Mooney Rivlin Model ........................................................................................ 29

    Equation 4 Standard Neo Hookean Model ......................................................................................... 29

  • 4

    List of Tables

    Table 1 Rubber Poperties ...................................................................................................................... 6

    Table 2 CIP properties ........................................................................................................................... 7

    Table 3 Calculation of Masses ............................................................................................................... 7

    Table 4 Youngs Moduli for all Isotropic Sample Configurations .......................................................... 16

    Table 5 Youngs Moduli of Pure Rubber with different Curing Times .................................................. 17

    Table 6 Youngs Moduli of 10%v CIP samples with different curing times ........................................... 18

    Table 7 Youngs Moduli for Anisotropic Samples ................................................................................. 21

    Table 8 Table of Youngs Moduli for 10% Isotropic Samples ................................................................ 23

    Table 9 Table of Youngs Moduli for 20% Isotropic Samples ................................................................ 24

    Table 10 Table of Youngs Moduli for 30% Isotropic Samples .............................................................. 25

    Table 11 Table of Youngs Moduli for 10% Anisotropic samples .......................................................... 26

    Table 12 Table of Youngs Moduli for 20% Anisotropic Samples ........................................................ 27

    Table 13 Table of Youngs Moduli for 30% Anisotropic samples .......................................................... 28

    Table 14 Table of Coefficients for Mooney-Rivlin Model .................................................................... 33

    Table 15 Table of Coefficients for Neo-Hookean Model ..................................................................... 33

    Table 16 Table of Coefficients for Neo-Hookean Model ..................................................................... 37

    Table 17 Table of Coefficients for Mooney-Rivlin Model .................................................................... 37

    Table 18 Table of R2 values for 30% Isotropic Samples ...................................................................... 41

    Table 19 Youngs Modulus for 30% Isotropic Samples for Neo-Hookean Model ................................. 41

    Table 20 Youngs Modulus for 30% Isotropic Samples for Mooney-Rivlin Model ................................ 41

    Table 21 Youngs Moduli for Pure Rubber for Mooney Rivlin Model ................................................... 43

    Table 22 Youngs Moduli for Pure Rubber Neo-Hookean model ......................................................... 43

    Glossary of Nomenclature

    MR effect- Magneto-Rheological effect whereby the stiffness of material increases with an

    applied magnetic field

    CIP- Carbonyl Iron Powder

    Mullins Effect- The instantaneous softening of rubbers when the all time maximum applied

    stress is reached

    FEA- Finite Element Analysis

    Magnetic Flux Density- This is the amount of magnetic field passing through a surface

    measured in Teslas

  • 5

    Acknoledgements

    I would like to thank Dr Harrison for helping me out during this project.He has always kept me

    organisedand wa always giving me new suggestions and some really good ideas to work on. I would

    also like to thank Gerlind Schubert for helping with all backround information on this topic as well as

    the analysis parts of the project. I would also like to thank John Davidson the materials lab

    technician. He really helped with the testing and the test setup.

    Introduction

    Magneto Rheological Elastomers are a set of smart composites whose mechanical

    properties can be reversibly and instantaneously changed when a magnetic field is applied.

    These materials can be used in systems where the ability to vary the stiffness of a

    component is required, such as vibration control systems and variable suspension systems

    in automobiles. Currently there is no available general constitutive model for these

    materials and the aim of Ms Schuberts PhD research project is to develop a constitutive

    model for these materials. The objective of this project is to undertake the preliminary

    compression testing of these materials to determine their stress strain behaviour.

    Furthermore the manufacturing process and testing procedures will be developed and

    refined throughout. The objective is to test all the relevant specimens and all configurations

    possible in order to assist with the development of the constitutive model.

    Background Information1

    Magneto Rheological Elastomers were first studied in 1995 by Toyota Central Research and

    Development Laboratories. They tested silicone gels with the magnetic particles aligned

    through dynamic shear experiments with small deformations. They noted the change in

    moduli due to the effects of the applied external magnetic field. Further research was

    carried out by Jolly and Carlsen in 1996 at the Thomas Lord Research Centre where they also

    tested silicon gels under small deformation shear with and without magnetic field and noted

    the change in storage modulus due to the applied magnetic field. The first applications for

    this kind of material were brought forth by the Ford Motor Company. They suggested and

    implemented the material as a suspension bushing thats stiffness could be altered to

    change ride comfort or handling quality of their motor cars. Kallio2 carried out many

    experiments in 2003 in order to discover which materials were best to create an MR effect.

    Kallio conducted mainly small strain experiments and found that the best materials are

    Silicone Rubber Matrix with CIP magnetic particles. Farshad3 in 2003 conducted

    compression testing up to 30% strain, using anistropic samples and isotropic samples for

    testing.

    1 All Backround information has been researched through Ms Gerlind Schuberts 1st Year Literature Revue 2 M. Kallio Preliminary tests on an MRE device 3 M. Farshad Magnetoactive elastomer composites

  • 6

    Furthermore testing with an applied magnetic field were conducted, although moving

    magnets were used which means further steps are needed in calculations to allow for the

    attractive forces of the magnets to be removed from the force measurements. Varga4 in

    2005 conducted similar compression testing but chose to use a solenoid coil to apply

    magnet fields to the samples. Applications for this material mainly involve vibration control

    as the forced response can be controlled by changing the stiffness components in springs.

    Manufacturing Process

    For this project a silicone rubber has been chosen as the matrix material along with

    Carbonyl Iron powder acting as the magnetic particle filler. The silicone rubber was chosen

    since has been the most widely used matrix material for MREs in past research.

    Furthermore it has good mechanical properties and chemical resistance over a large range

    of temperatures. The rubber which is being used is defined by the manufacturer as

    MM240TV. Like most silicon rubbers it consists of two parts. Part A and Part B where part B

    acts as the hardener. The components are mixed at a ratio of 10:1. MM240TV has the

    properties defined in Table 1.

    Table 1 Rubber Poperties

    Figures from data sheet provided by ACC silicones Ltd

    Viscosity 96000 mPa.s Tensile Strength 5.4MPa

    Elongation at Break 330%

    Youngs Modulus 1.88MPa

    Hardness 40o Shore A

    The magntic particle filler that has been chosen is Carbonyl Iron powder. This is a typical

    magnetic filler as it has high magnetic permeability, low remnant magnetisation and high

    saturation.5 The CIP powder being used has the properties shown below in table 2

    4 Z. Varga Magnetic Field Sensitive functional elastomers with tunable modulus 5Dr Philip Harrison - Magneto-Rheological Elastomers: Manufacture, characterisation and modelling Presentation

  • 7

    Table 2 CIP properties

    These values are based on the information provided by the provider BASF Ltd

    *Values are based on BASFs SQ standard

    There is a third component which has been added to the MRE that is being used for this

    project. It is defined by the manufacturer as ACC34 thinner. The reason that this is being

    used is that the rubber on its own has a high viscosity and this leads to the agglomeration of

    particles in the matrix. This in turn produces a non uniform distribution of particles. This can

    makes modelling much harder. Therefore this solvent thinner has been added at 30%

    weight to reduce the viscosity. This in turn allows for an even distribution of particles and

    also allows the particles to align with the magnetic fields easily. However adding the thinner

    will reduce the modulus of samples while no magnetic field is applied, this may lead to a

    higher MR effect.

    The amount of CIP that is required for testing is 10, 20 and 30 percent volume. The

    equivalent mass was calculated using the Table 3 and Equation 1 shown below.

    Parameter Unit Specification Test method

    Iron Content g/100g Min 99.5 Calculated Carbon Content g/100g Max 0.05 IRS (RCA/Q-C-296)

    Permeability (i)* % 98-110 RCA/Q C 302

    Q-Value * a 100kHz % 95-120 RCA/Q C 302

    Particle Size Distribution

    D50 m 3.7-4.7 Microtrac X 100

    Using 50g of Rubber Component A

    Therefore: Mass of part B = 0.1 x 50 = 5g

    Using 30 w% of Solvent therefore:

    Mass of solvent = 0.3 x (50+5) = 16.5g

    Table 3 Calculation of Masses

  • 8

    Equation 1 Equation for Mass of CIP[6]

    Where v = relative volume of CIP and x = required mass, 71.5g is the total mass of pure

    rubber 1.06g/cm3 is density of rubber and 7.874 g/cm3 is density of CIP.

    The next stage of the process was to mix the mixture thoroughly using a hand mixer for a

    minimum time of three minutes. The mixture is then degassed in a vacuum chamber for ten

    minutes in order to avoid air bubbles when the rubber is cured. The mixture is then poured

    into moulds. These moulds are designed according to the standard sample size for test

    method A of BS ISO 77427. The sample size required is 29mm 0.5mm diameter and

    12.5mm 0.5mm height.

    After the rubber has been moulded it requires time to cure, 24 hrs at 25oC or 1 hrs at 100oC

    as defined by the manufacturer. In order to fast cure the rubber at 100oC the heating plates

    shown in Figure 2 are used.

    Figure 1 Picture of Heating Plates

    6 Equations and Calculations are from Guide for rubber mixing process by Ms Gerlind Schubert 7 BS ISO 7742 Determination of Compression stress strain properties for Vulcanized Rubbers

  • 9

    One of the objectives for this report is to optimize the manufacturing process therefore

    samples will be fast and slow cured to find the optimal curing time.

    In order to create anisotropy within the samples they will be cured while under the

    influence of a magnetic field produced by an electromagnet. As shown in figures 3 and 4.

    This will create chains of magnetic particles within the samples. For this project a magnetic

    field density of 400mT (milli Teslas) was used during curing. This value was measured using a

    gauss meter at the centre of the poles with no medium present. Unfortunately the amplifier

    tends to get hot and becomes overloaded after time, this meant that the higher magnetic

    field densities that were measured (max of 1 Tesla) could not be used. 400mT was found to

    be a safe value in order to cure the samples for the necessary time without overloading the

    amplifier. Furthermore other research projects have successfully used similar field densities

    for creating anisotropy for example Varga 2006 [4].

    Test Method

    For this project the British Standard for Determination of compression stress-strain

    properties for Rubber, vulcanized or thermoplastic (BS ISO 7743:2008) is being used. Test

    method A from the standard has been selected and requires that the compression plates be

    lightly coated in a film of lubricant. Furthermore the test requires four cycles at a speed of

    10mm/min. From the recorded force displacement data the stresses and strains can be

    calculated. The standard also requires a minimum of three samples for each sample

    configuration.

    Figure 3 Picture of Electromagnet

    Figure 2 Picture of Electromagnet with mould in between poles

  • 10

    Test Setup

    In order to test these materials with an applied magnetic field it is necessary to create a

    device or alternative setup to allow the magnetic field to pass through the samples while

    they are compressed. A paper by Varga [4] indicated that a solenoid coil was used to

    implement a uniform magnetic field during testing. The aim of this project was to use quite

    a high field density of 400mT, and after some calculations (see appendix A) it was decided

    that the required coil would be expensive and impractical for use. Another paper by Farshad

    [3] indicated that a pair of permanent magnets could be used to generate a uniform field

    through the sample. The test setup used by Farshad [3] indicated that the plates be wedged

    between two aluminium plates while the compression took place, thereby the magnets

    would move up and down with the compression. It was decided that this method was

    inaccurate due to the fact that the attractive force between the magnets would increase as

    the magnets were brought closer together. This would add false readings to the load cell

    and would mean that additional steps would have to be taken to interpret the results

    correctly. Therefore a new test setup was designed and assembled using neodymium

    permanent magnets. The technical drawing is shown in figure 4. Unfortunately when this

    setup was used there were some problems with the test results (see appendix B). Therefore

    the setup was reworked again the revised setup is shown in figure 5. The magnetic field

    density can be altered simply by making the distance between the magnets larger or

    smaller. During testing a magnetic field density of 400mT was measured between the poles

    at a separation of 36mm and this decreased to around 270mT at a distance of 47mm.

    From the drawings provide you will see that the magnets are held in place as the

    compression plates are free to move.

  • 11

    Figure 4 Original Magnetic Setup

    Figure 5 Revised Magnetic Setup

  • 12

    Figure 7 Picture of Test Setup

    Figure 6 Rendered Drawing of Test Setup

    Wooden Blocks

    Magnets

    Clamps

  • 13

    Test Results

    The tests that have been completed are:

    Compression of Pure rubber, 10%, 20%, 30% volume CIP cured for 24 hours at 25oC. Purely Isotropic samples.

    Compression of Pure Rubber cured for 1 hour and 1.5 hours at 100 oC.

    Compression of 10% volume cured for 1 hour and 1.5 hours at 100 oC. Purely Isotropic samples.

    Compression of 10%, 20% and 30% Anisotropic samples cured for 1 hour at 100 oC

    under a magnetic field of 400mT.

    Compression with applied magnetic fields using samples of 10%, 20%, 30% volume CIP cured for 1 hour at 100 oC. Purely Isotropic samples.

    Compression with applied magnetic field using samples of 10%, 20%, 30% volume CIP cured for 1 hour at 100 oC under field of 400mT magnetic field. Purely Anisotropic Samples

    During the testing it was noticed that the first cycle of loading showed larger force than the subsequent three cycles. This is effect is known as the Mullins effect and is typical for rubbers. This effect is defined as:

    The Mullins Effect can be idealized for many purposes as an instantaneous and irreversible

    softening of the stress-strain curve that occurs whenever the load increases beyond its prior

    all-time maximum value. At times when the load is less than a prior maximum, nonlinear

    elastic behaviour prevails.[8]

    The main theory for why this occurs says that it is caused by the breaking of cross links in

    the matrix materials, thereby reducing the overall cross-linking density and therefore the

    stiffness of the material. This effect is illustrated by the full compression cycle test shown in

    figure 8. As you can see the first cycle is higher than the others and all the others are equal.

    8 Reference from (http://en.wikipedia.org/wiki/Mullins_effect)

  • 14

    Figure 8

    Due to this effect for the comparisons of all tests the upload part of the third cycle will be

    used. A mean curve will then be calculated from the third cycles of the four samples tested.

    This ensures that there will be no softening after this loading cycle and therefore provides

    more accurate data for creating a constitutive model.

    This of course means that if these materials are ever used in a practical application there

    will need to be a degree of conditioning before entering service. This would require the

    material to be loaded to a higher stress than that which would be expected to encounter

    during service.

    Pure Rubber, 10%, 20%, 30% Isotropic samples cured for 24 Hours at 25oC

    The samples mixed for this testing regime were cured for 24 hours at 25oC as prescribed by

    the manufacturer. The samples produced the stress-strain data shown in figures 11 and

    table 4. Figure 8 below shows the structure of an isotropic sample loaded with 10% volume

    CIP at a magnification factor of x20 using a microscope, the white flecks are CIP clusters.

    This photograph shows that there is no order to the distribution of particles. Figure 9 shows

    a 20% volume isotropic sample of the same magnification. From figure 9 it can be seen that

    there is a higher density of white flecks and therefore CIP.

    Cycle 1 Upload

    Cycle 2-4 Upload

    All Unload Parts

  • 15

    Figure 9 Microscopic Picture of 10% Isotropic Sample

    Figure 10 Microscopic Picture of 20% Isotropic Sample

  • 16

    Figure 11 Graph of all Isotropic Sample Configurations

    Table 4 Youngs Moduli for all Isotropic Sample Configurations

    Sample

    Composition

    Cure Time

    (hrs)

    Mean Youngs

    Modulus (Mpa)for

    0-10% strain

    Percentage

    Increase compared

    to pure rubber

    Pure 24 0.4856

    10% 24 0.9505 95.737

    20% 24 1.6845 246.890

    30% 24 2.4553 405.622

    Note that for comparing Youngs moduli we are considering only the 0-10% strain region

    due to the fact that this region has very linear stress-strain behaviour and therefore a

    relatively constant gradient, i.e. youngs modulus, unlike the higher strain regions where the

    behaviour is very much non-linear.

    As expected the higher volume percentage of CIP produces higher stiffnesss. The choice of

    material for a specific application will depend on the required stiffness but also the limit of

    weight.

  • 17

    Compression of Pure Rubber with different curing conditions

    The samples for this set of tests were cured at 100oC for 1 hour, 100oC for 1.5 hours and

    25oC for 24 hours. This is to check to see if there is a significant difference in stiffness for

    different curing times in order to optimize the manufacturing process.

    Figure 12 Graph of Pure Rubber with different Curing Conditions

    Table 5 Youngs Moduli of Pure Rubber with different Curing Times

    Sample

    Composition

    Cure Time

    (hrs)

    Mean Youngs Modulus

    (Mpa)for 10% strain

    % increase compared

    to 24 hrs

    Increase

    Factor

    Pure 24 0.4856

    Pure 1 0.67 37.974 1.3797

    Pure 1.5 0.7073 45.655 1.4565

    From these results the fast curing process produces higher stiffnesss. This may be due to

    additional energy being provided by the heating process, creating more intermolecular

    bonding and therefore producing higher stiffness. Furthermore the 1.5 hours cured samples

    have slightly higher stiffness than the 1 hour cured since the additional heat energy is

    provided for a longer period.

  • 18

    Compression of 10% volume Isotropic samples with different curing conditions

    This set of test has been done to see if the stiffness of filled samples is altered by the

    additional heat energy provided as shown in the pure rubber samples shown before. The

    samples as before have been cured for 1.5 hours, 1 hour and 24 hours. These results are

    very important to optimize the manufacturing process.

    Figure 13 Graph of 10%v CIP samples with different curing conditions

    Table 6 Youngs Moduli of 10%v CIP samples with different curing times

    Sample

    Composition

    Cure Time

    (hrs)

    Mean Youngs Modulus

    (Mpa)for 10% strain

    Percentage increase

    compared to 24 hrs

    Increase

    Factor

    10% 24 0.9505

    10% 1 0.9996 5.166 1.051657

    10% 1.5 1.0052 5.755 1.057548

  • 19

    From these results it can be seen that the addition of heat has very little affect on the

    stiffness of the samples, there is only a 5% increase in stiffness. Therefore unlike the pure

    rubber there is only a small amount of additional bonding caused by the additional energy

    provided. This means that the interaction between the filler and the matrix does not allow

    for strengthening of the matrix, it can therefore be assumed that most of the heat energy is

    absorbed by the CIP particles. This may be due to a higher thermal conductivity. From these

    results all future samples will be cured for 1 hour at 100oC.

    Compression of 10%, 20% and 30% Anisotropic samples

    These samples have been cured for 1 hour at 100oC under a magnetic field of 400mT. This

    field strength was used mainly because the amplifier being used overloads after time while

    high currents are used. The value of 400mT was a safe magnetic field density that could be

    achieved for the full curing time needed. Furthermore other studies (VARGA [4]) have

    shown that 400mT is adequate to create anisotropy. Figure 14 shows the structure created

    by applying the magnetic field during curing. As you can see there are distinct chains of

    particles as opposed to random orientation shown in figure 8 and 9.

    Figure 14 Microscopic Picture of 10% Anisotropic Sample

    Direction

    of

    Alignment

  • 20

    There have been some problems with the compression tests for this anisotropic structure.

    The rubber samples when compressed actually buckled which is not supposed to happen

    the rubber should remain at a constant volume and therefore the diameter of the sample

    should increase with decreasing height. Figure 15and 16 shows one of the samples as it

    buckles.

    The result from this buckling is a very steep upload at the beginning of the cycle then it

    flattens out. Interestingly this behaviour is very consisitent for all the samples. At first it was

    suspected that because the moulds between the poles of the electromagnet were not

    completely covered by the pole and it was thought that the flux density would be weaker at

    the edge of the poles.

    To stop this only one mould at a time was cured between the magnetic poles. Unfortunately

    this did not solve the problem. It was then discovered that the moulds that were being used

    had deformed due to the heat and pressure of the curing process. New top plates were

    manufactured with additional screw holes for extra security. Unfortunately this still did not

    stop the buckling. It may be that the shear modulus is much less than the compressive

    modulus. In order to stop the buckling occurring, the film of oil on the sample during testing

    was removed. This did stop the buckling but unfortunately this causes additional friction on

    the sample during testing which means that the force displacement will not be comparable

    with the other tests. The test results are shown in figure 14 and table 7, but it must be

    stated that these results will not be an accurate representation of the anisotropic samples.

    Figure 15 Buckling Samples

  • 21

    Table 7 Youngs Moduli for Anisotropic Samples

    Sample

    Composition

    Particle

    Orientation

    Mean Youngs Modulus

    (Mpa)for 10% strain

    % increase compared to

    equivalent isotropic

    30% Anisotropic 4.625 88.368

    20% Anisotropic 2.5068 48.816

    10% Anisotropic 1.2233 28.701

    30% Isotropic 2.4553

    20% Isotropic 1.6845

    10% Isotropic 0.9505

    Figure 16 Graph of all volume percentages of CIP for Anisotropic and Isotropic structures

  • 22

    Interestingly it can be seen that the higher percentages of CIP produce larger increases in

    the Youngs modulus. This will be due to denser particle chains created by the magnetic

    field. The higher concentration of ferromagnetic elements will create a higher magnetic flux

    density within the samples when curing, meaning that the individual particles will be

    magnetically attracted towards each other creating thicker and longer chains.

    Due to the fact that buckling is a factor in the testing of Anisotropic samples, it may be more

    poignant to perform tensile tests instead of compression to discover the mechanical

    properties that Anisotropy creates.

    Compression of 10%, 20% and 30%, Anisotropic and Isotropic Samples with applied

    magnetic fields

    This set of tests will deal with the effect of applying a magnetic field through the samples

    while being compressed. The magnetic field should increase the stiffness of the samples and

    the higher volume percentages should produce a bigger MR effect. Furthermore the

    Anisotropic samples should produce a bigger increase compared to the Isotropic samples.

    The increase in stiffness is caused by the interaction of the magnetic particles in the matrix.

    The magnetic flux causes a magnetic dipole interaction between every particle in the matrix.

    In the Anisotropic samples the magnetic particles will be closer together, thereby increasing

    the attractive force between them as attractive force is a function of distance.

  • 23

    10% Isotropic Samples

    Table 8 Table of Youngs Moduli for 10% Isotropic Samples

    Sample Composition Flux Density

    0-10% Youngs Modulus

    10% 400 1.2453

    10% 270 1.2697

    10% 0 1.4567

    As we can see the results are not as expected. The applied magnetic field should increase

    stiffness not decrease it. The reason for this is currently unknown. It is a completely

    unexpected result. The curves indicate that the samples with the 400mT applied field have

    higher stress values than the 270mT but the gradient of these lines i.e. Youngs Modulus

    indicate that the stiffness is lower for the higher Flux Density. This is contrary to work that

    has been carried out previously. Furthermore the zero field samples, especially at higher

    strains, have much higher stresses than those samples with an applied field. The only thing

    that changed between the magnetically tested samples and the no field samples was that

    for the no field samples a standard compression setup for the Zwick Z250 test machine was

    used. That particular setup involved less pinned joints and fittings. These fittings may have

    been slightly loose at certain points and may have resulted in a slightly decrease in the

    measured force.

    Figure 17 Graph of 10% Isotropic CIP Samples with different applied magnetic fields

  • 24

    20% Isotropic Samples

    Table 9 Table of Youngs Moduli for 20% Isotropic Samples

    Sample Composition

    Flux Density (mT)

    0-10% Youngs Modulus (MPa)

    20% 400 1.6624

    20% 270 1.6478

    20% 0 1.806

    From figure 18 and table 9, we can see a similar result from the 10% isotropic samples in

    that the samples with the applied magnetic fields have lower stiffnesss than the no field

    samples. This would support the theory that the magnetic test setup is slightly flawed in

    some way. This sample however shows that the Youngs modulus did increase slightly for a

    higher magnetic field. From these results I would assume that the 10% samples have

    become magnetically saturated and this may have had a detrimental effect on the

    mechanical properties. The 20% samples, having a higher Ferro-Magnetic content can have

    higher flux densities applied to them without saturating and therefore maintain an increase

    in stiffness.

    Figure 18 Graph of 20% Isotropic CIP Samples with different applied magnetic fields

  • 25

    30% Isotropic Samples

    Table 10 Table of Youngs Moduli for 30% Isotropic Samples

    From the curves and values shown in figure 19 and table 10 respectively, we can see that

    the problem of the zero field samples having a higher stiffness than those with the applied

    magnetic field is still partly true as the 270mT field samples still have lower stiffness.

    However we are starting to see a larger MR effect as the stiffness of the 400mT sample is

    higher than the other two configurations. This shows that for higher magnetic particle

    loading we are getting a larger MR effect. If the problem with the magnetic test setup had

    not been present, we would have seen that the 270mT field samples curve would fall

    inbetween the other two configurations.

    Sample Composition

    Flux Density

    (mT)

    0-10% Youngs

    Modulus (MPa)

    30% 400 2.7491

    30% 270 2.519

    30% 0 2.5882

    Figure 19 Graph of 30% Isotropic Samples with different applied magnetic fields

  • 26

    10% Anisotrpic Samples

    Table 11 Table of Youngs Moduli for 10% Anisotropic samples

    Sample Composition Flux Density (mT) Youngs Mod for 0-10% Strain MR Effect

    10% 400 2.763363 0.123328795

    10% 270 2.627874 0.068251451

    10% 0 2.459977

    From the results in figure 20 and table 11 we can see that compared to the Isotropic

    samples we are seeing a larger MR effect, this is due, as stated previously, to the magnetic

    particles being closer together. Still present is the fact that the zero field samples are

    producing higher stresses than the 270mT applied field samples. From the curves we can

    also see that there is a steep gradient in the 0.05 to 0.1 strain region followed by a levelling

    off of the curve after this region. This may have been caused by buckling but it is unclear at

    this point.

    Figure 20 Graph of 30% Anisotropic Samples with different applied magnetic fields

  • 27

    20% Anisotropic Samples

    Table 12 Table of Youngs Moduli for 20% Anisotropic Samples

    Sample Composition Flux Density (mT) Youngs Mod for 0-10% Strain (MPa) MR Effect

    20% 400 4.009061 0.042071333

    20% 270 3.873636 0.006870444

    20% 0 3.847204

    As with 10% Anisotropic Samples we are seeing that the zero field samples have a similair

    than the 270mT field samples. Again it is suspected that a loose fitting in the magnetic test

    setup is to blame. We are also seeing that at higher strains for the zero field samples have a

    slightly higher stresses than the 400mT samples. The upload curves however show that all

    samples have very similar upload behaviour up to around 30% strain. This is unexpected as

    we would have expected larger differences in all curves for this sample configuration.

    Figure 21Graph of 20% Anisotropic samples with different applied magnetic fields

  • 28

    30% Anisotropic Samples

    Table 13 Table of Youngs Moduli for 30% Anisotropic samples

    Sample Composition Flux Density (mT) Youngs Mod for 0-10% Strain (MPa) MR Effect

    30% 400 5.947736 0.588323435

    30% 270 5.146382 0.374324472

    30% 0 3.744663

    The curves shown in figure 22 are like what was expected, we can see that the lowest curve

    is the zero field samples, with the red line representing the 270mT field samples and just

    above that the 400mT field samples. There is very little difference in the stresses off the two

    magnetic samples. This graph is a good indicator of the sort of MR effect we would hope to

    achieve in practice with these materials. The Youngs moduli figures shown in table indicate

    a larger MR effect than previous samples configurations.

    Figure 22 Graph of 30% Anisotropic samples with different applied magnetic fields

  • 29

    Modelling of MREs

    In this report the modelling of the Isotropic samples will be concentrated on. The main

    reason is the experimental problems with the anisotropic samples which need to be solved

    first. I am using for analysis the Abaqus 6.8 Student Edition. Already implemented models

    for rubber-like materials are the Neo-Hookean, the Mooney-Rivlin and the Ogden Model

    which are all models for isotropic materials. Modelling an anisotropic material would

    require a user defined constitutive equation using UMAT. This would go beyond the scope

    of this report. A further assumption of the said implemented models is that the materials

    are fully incompressible i.e. the volume does not change during deformation and the

    poisson ratio is 0.5. Furthermore G the shear modulus is related to the Elastic modulus E by

    the equation, , but this is only valid on the linear theory ,i.e. in the small strain

    region.

    The standard hyper elastic models which will be used are the, Neo-Hookean and Mooney

    Rivlin. All these models are derived from the standard Ogden model shown in Equation 2.

    This model is based on the relationship between the Strain Energy Function9 and the

    principal stretch ratios.

    Equation 2 Standard Ogden Model

    Where is the strain energy and , are the principal stretch ratios for each

    direction. is defined as the shear modulus, and is a material constant.

    The Mooney Rivlin model is obtained by setting N=2, = 2, = -2. This produces Equation

    3 shown below.

    Equation 3 Standard Mooney Rivlin Model

    Where c10

    , c01

    , and the shear modulus is =

    The Neo-Hookean model is derived from the standard Ogden model by setting N=1, = 2,

    this results in Equation 4.

    Equation 4 Standard Neo Hookean Model

    9 Equations from Non-Linear Solid Mechanics by Gerhard A Holzapfel

  • 30

    Where c1

    and the shear modulus is

    The Youngs modulus can be calculated from the coefficients calculated by Abaqus shown by

    these equations:

    These constitutive model need to be fitted to experimental data to calculate the model

    parameters. This fitting will be done by Abaqus using a least square method. To get an idea

    of the suitability of the fit a coefficient of determination R2 will be calculated. This

    coefficient is often used in statistics. It approaches 1 the better the fit, a value of 0 would

    mean the fit is useless.

    exp fitres

    22

    2

    2

    1 ... nresresresnorm

    i

    itotS2

    exp

    _

    exp, )(

    tot

    tot

    S

    normSR

    2

    2

    Where res are the residuals, norm is the 2-Norm also known as the Euclidian length of the residuals, Stot is the sum of squares and finally R

    2 is the coefficient of determination.

    Modelling of 10% Isotropic Samples

    Figure 23 10% Isotropic Samples with 400mT applied magnetic field

  • 31

    Coefficients for 10% Isotropic samples with 400mT applied magnetic field

    HYPERELASTICITY - MOONEY-RIVLIN STRAIN ENERGY

    D1 C10 C01

    0.000000 84061.8510 34812.0823

    HYPERELASTICITY - NEO-HOOKEAN STRAIN ENERGY

    D1 C10 C01

    0.00000 83567.7120 0.00000000

    Coefficients for 10% Isotropic Samples with 270mT applied magnetic field

    HYPERELASTICITY - MOONEY-RIVLIN STRAIN ENERGY

    D1 C10 C01

    0.00000000 120041.678 18855.2399

    HYPERELASTICITY - NEO-HOOKEAN STRAIN ENERGY

    D1 C10 C01

    0.00000000 145569.612 0.00000000

    Figure 24 10% Isotropic Samples with 270mT applied magnetic field

  • 32

    Coefficients for 10% Isotropic Samples with no applied magnetic field

    HYPERELASTICITY - MOONEY-RIVLIN STRAIN ENERGY

    D1 C10 C01

    0.00000000 62757.9238 78950.0356

    HYPERELASTICITY - NEO-HOOKEAN STRAIN ENERGY

    D1 C10 C01

    0.00000000 172278.547 0.00000000

    Figure 25 10% Isotropic samples no applied magnetic field

  • 33

    From Figure 23 we can see that the best model fit is the Mooney-Rivlin. This was confirmed

    by the R2 coefficient which was calculated as 0.9988 for the Mooney-Rivlin and 0.5558 for

    the Neo-Hookean.

    From Figure 24 we can see that both models are quite good for most strain levels. The R2

    coefficient for the Mooney Rivlin model the R2 value was 0.9987 and for the Neo-Hookean

    model R2 is 0.9902. this shows that the Mooney-Rivlin is still the better fit for this material

    configuration with and without magnetic fields.

    From Figure 25 we can also see that the Mooney Rivlin model is still the best fit, and the R2

    values confirm this as they are 0.9917 for the Mooney-Rivlin and 0.9231 for the Neo-

    Hookean. Finally the Mooney-Rivlin Model is the best fit for this material tested with and

    without magnetic field. Using the coefficients shown previously Table 14 and Table 15 were

    constructed to show the differences in the Youngs moduli from test data and coefficients.

    Table 14 Table of Coefficients for Mooney-Rivlin Model

    Samples Composition

    Applied Field (mT)

    C10 Coefficient

    C01 Coefficient

    Shear Modulus (MPa)

    Youngs Modulus (MPa)

    Youngs Modulus form test data (MPa)

    10% 400 84061.851 34812.0823 0.2377 0.7132 1.2453

    10% 270 120041.67 18855.2399 0.2778 0.8334 1.2697

    10% 0 62757.923 78950.0356 0.2834 0.8502 1.4567

    Table 15 Table of Coefficients for Neo-Hookean Model

    Samples Composition

    Applied Field (mT)

    C10 Coefficient

    Shear Modulus (MPa)

    Youngs Modulus (MPa)

    Youngs Modulus form test data (MPa)

    10% 400 83567.7120 0.1671 0.5014 1.2453

    10% 270 145569.6120 0.2911 0.8734 1.2697

    10% 0 172278.5470 0.3446 1.0337 1.4567

  • 34

    Modelling of 20% Isotropic Samples

    Figure 26 20% Isotropic Sample with 400mT applied magnetic field

    Coefficients for 20% Isotropic Samples with 400mT applied magnetic field

    HYPERELASTICITY - MOONEY-RIVLIN STRAIN ENERGY

    D1 C10 C01

    0.00000000 95259.1322 61905.0798

    HYPERELASTICITY - NEO-HOOKEAN STRAIN ENERGY

    D1 C10 C01

    0.00000000 89553.1134 0.00000

  • 35

    Coefficients for 20% Isotropic Samples with 270mT applied magnetic field

    HYPERELASTICITY - MOONEY-RIVLIN STRAIN ENERGY

    D1 C10 C01

    0.00000000 128736.268 48048.7121

    HYPERELASTICITY - NEO-HOOKEAN STRAIN ENERGY

    D1 C10 C01

    0.00000000 192142.047 0.00000000

    Figure 27 20% Isotropic Samples with 270mT applied magnetic field

  • 36

    Figure 28 20% Isotropic Samples with no applied magnetic field

    Coefficients for 20% Isotropic Sample with no applied magnetic field

    HYPERELASTICITY - MOONEY-RIVLIN STRAIN ENERGY

    D1 C10 C01

    0.00000000 -20993.9161 162880.115

    UNIAXIAL TENSION: UNSTABLE AT A NOMINAL STRAIN LARGER THAN 2.9800

    UNIAXIAL COMPRESSION: UNSTABLE AT A NOMINAL STRAIN LESS THAN -0.6414

    BIAXIAL TENSION: UNSTABLE AT A NOMINAL STRAIN LARGER THAN 0.6700

    BIAXIAL COMPRESSION: UNSTABLE AT A NOMINAL STRAIN LESS THAN -0.4987

    PLANAR TENSION: UNSTABLE AT A NOMINAL STRAIN LARGER THAN 1.6700

    PLANAR COMPRESSION: UNSTABLE AT A NOMINAL STRAIN LESS THAN -0.6255

    VOLUMETRIC TENSION: STABLE FOR ALL VOLUME RATIOS

    VOLUMETRIC COMPRESSION: STABLE FOR ALL VOLUME RATIOS

  • 37

    HYPERELASTICITY - NEO-HOOKEAN STRAIN ENERGY

    D1 C10 C01

    0.00000000 174917.522 0.00000000

    From Figure 26 we can see that again the Mooney-Rivlin Model is the best fit for this sample

    configuration. This is confirmed with the R2 coefficient being 0.9987 for the Mooney-Rivlin

    and 0.2354 for the Neo-Hookean.

    Figure 27 shows that for lower strains both models are quite accurate but as strain increases

    the Neo-Hookean curve starts to move away from the test data. The Mooney-Rivlin is still

    the best model for this material. Furthermore this is confirmed with the R2 coefficient, it is

    0.9970 for the Mooney-Rivlin and 0.9671 for the Neo-Hookean.

    In Figure 28 we are starting to see a few discrepancies between the Mooney-Rivlin model

    and the test data especially at higher strains. Even so the Mooney-Rivlin is still the best

    fitting model for this data as shown again by the R2 coefficient. For the Mooney-Rivlin R2 is

    0.9827 and for the Neo-Hookean R2 is 0.7041. Unusually the Abaqus has stated the

    Mooney-Rivlin Model as unstable.

    Table 16 Table of Coefficients for Neo-Hookean Model

    Samples Composition

    Applied Field (mT)

    C10 Coefficient

    Shear Modulus (MPa)

    Youngs Modulus (MPa)

    Youngs Modulus form test data (MPa)

    20% 400 89553.1134 0.1791 0.5373 1.6624

    20% 270 192142.0470 0.3843 1.1529 1.6478

    20% 0 174917.5220 0.3498 1.0495 1.806

    Table 17 Table of Coefficients for Mooney-Rivlin Model

    Samples Composition

    Applied Field (mT)

    C10 Coefficient

    C01 Coefficient

    Shear Modulus (MPa)

    Youngs Modulus (MPa)

    Youngs Modulus form test data (MPa)

    20% 400 95259.

    1322 61905.07

    98 0.3143 0.9430 1.6624

    20% 270 128736

    .2680 48048.71

    21 0.3536 1.0607 1.6478

    20% 0 1.806

  • 38

    Modelling of 30% Isotropic Samples

    Coefficients for 30% Isotropic Samples with 400mT applied magnetic field

    HYPERELASTICITY - MOONEY-RIVLIN STRAIN ENERGY

    D1 C10 C01

    0.00000000 265354.405 59029.5065

    HYPERELASTICITY - NEO-HOOKEAN STRAIN ENERGY

    D1 C10 C01

    0.00000000 352369.597 0.00000000

    Figure 29 30% Isotropic Samples with 400mT applied magnetic field

  • 39

    Coefficients for 30% Isotropic Samples with 270mT Magnetic Field

    HYPERELASTICITY - MOONEY-RIVLIN STRAIN ENERGY

    D1 C10 C01

    0.00000000 162187.941 94669.0067

    HYPERELASTICITY - NEO-HOOKEAN STRAIN ENERGY

    D1 C10 C01

    0.00000000 238740.306 0.00000000

    Figure 30 30% Isotropic Samples with 270mT applied magnetic field

  • 40

    Coefficients for 30% Isotropic Samples with no applied magnetic field

    HYPERELASTICITY - MOONEY-RIVLIN STRAIN ENERGY

    D1 C10 C01

    0.00000000 -93284.8951 304559.773

    ***WARNING: UNSTABLE HYPERELASTIC MATERIAL

    UNIAXIAL TENSION: UNSTABLE AT A NOMINAL STRAIN LARGER THAN 0.9500

    UNIAXIAL COMPRESSION: UNSTABLE AT A NOMINAL STRAIN LESS THAN -0.4513

    BIAXIAL TENSION: UNSTABLE AT A NOMINAL STRAIN LARGER THAN 0.3500

    BIAXIAL COMPRESSION: UNSTABLE AT A NOMINAL STRAIN LESS THAN -0.2839

    PLANAR TENSION: UNSTABLE AT A NOMINAL STRAIN LARGER THAN 0.7200

    PLANAR COMPRESSION: UNSTABLE AT A NOMINAL STRAIN LESS THAN -0.4186

    VOLUMETRIC TENSION: STABLE FOR ALL VOLUME RATIOS

    VOLUMETRIC COMPRESSION: STABLE FOR ALL VOLUME RATIOS

    Figure 31 30% Isotropic Samples with no applied field

  • 41

    HYPERELASTICITY - NEO-HOOKEAN STRAIN ENERGY

    D1 C10 C01

    0.00000000 325749.014 0.00000000

    From these simulations we can see that the Mooney-Rivlin Model is a good model for the

    samples where a magnetic field is present. However when there is no magnetic field present

    this model becomes unstable. The R2 values for all the cycles are shown in Table 18. This

    means that this model will have to be altered in order to be used for simulations. Table 19

    and Table 20 show a comparison of the simulated youngs moduli to the calculated Modulus

    from the test data.

    Table 18 Table of R2 values for 30% Isotropic Samples

    Samples Composition

    Applied Field Strength (mT)

    R2 for Mooney Rivlin

    R2 for Neo-Hookean

    30% 400 0.9964 0.9854

    30% 270 0.9977 0.8139

    30% 0 0.9496 0.3185

    Table 19 Youngs Modulus for 30% Isotropic Samples for Neo-Hookean Model

    Samples Composition

    Applied Field (mT)

    C10 Coefficient

    Shear Modulus (MPa)

    Youngs Modulus (MPa)

    Youngs Modulus form test data (MPa)

    30% 400 352369.5970 0.7047 2.1142 2.7491

    30% 270 238740.3060 0.4775 1.4324 2.519

    30% 0 325749.0140 0.6515 1.9545 2.5882

    Table 20 Youngs Modulus for 30% Isotropic Samples for Mooney-Rivlin Model

    Samples Composition

    Applied Field (mT)

    C10 Coefficient

    C01 Coefficient

    Shear Modulus (MPa)

    Youngs Modulus (MPa)

    Youngs Modulus from test data (MPa)

    30% 400 265354.40

    50 59029.506

    5 0.6488 1.9463 2.7491

    30% 270 162187.94

    10 94669.006

    7 0.5137 1.5411 2.519

    30% 0 2.5882

  • 42

    Modelling of Pure Rubber

    Figure 32 Pure Rubber Modelling Curves

    Coefficients for Pure Rubber Testing

    HYPERELASTICITY - MOONEY-RIVLIN STRAIN ENERGY

    D1 C10 C01

    0.00000000 73412.7398 23617.1556

    HYPERELASTICITY - NEO-HOOKEAN STRAIN ENERGY

    D1 C10 C01

    0.00000000 79240.8245 0.00000000

  • 43

    Table 21 Youngs Moduli for Pure Rubber for Mooney Rivlin Model

    Samples Composition

    Applied Field (mT)

    C10 Coefficient

    C01 Coefficient

    Shear Modulus (MPa)

    Youngs Modulus (MPa)

    Youngs Modulus form test data (MPa)

    Pure 0 73412.7398 23617.155 0.1941 0.5822 0.67

    Table 22 Youngs Moduli for Pure Rubber Neo-Hookean model

    Samples Composition

    Applied Field (mT)

    C10 Coefficient

    Shear Modulus (MPa)

    Youngs Modulus (MPa)

    Youngs Modulus form test data (MPa)

    Pure 0 79240.8245 0.1585 0.4754 0.67

    The pure rubber is like all the rest of the samples that have been modelled in that the best fit is with

    the Mooney-Rivlin model. The R2 values are 0.9995 for the Mooney-Rivlin model and 0.7656 for the

    Neo-Hookean.

    In conclusion it would appear that the best model for this material is the Mooney-Rivlin model, it has

    consistently close to 1 R2 value providing a good fit.

    Conclusions

    The main conclusions of this report are that the manufacturing process has been optimized and now

    all future samples will be cured for 1hrs at 100oC. Furthermore it is now possible to successfully

    create Anisotropy within the MREs. Another main point from these set of test is that a method

    whereby a magnetic field is passed through the sample while being compressed has been achieved.

    Some of the problems that have occurred unexpectedly during these tests were the buckling of

    anisotropic samples. This means that in future it would be best to test the anisotropic samples in

    tension rather than compression. Unfortunately applying a magnetic field through the sample will be

    difficult and a new test setup will have to be created.

    Future work on this material will be carried out to create a general constitutive model, and to

    achieve this task many more tests will have to be undertaken to fully define the MREs mechanical

    properties.

  • 44

    Bibliography

    [1] Gerlind Schuberts 1st Year Report

    [2] M. Kallio Preliminary tests on an MRE device

    [3] M. Farshad Magnetoactive elastomer composites [4] Z. Varga Magnetic Field Sensitive functional elastomers with tunable modulus

    [5]Dr Philip Harrison, MRE Presentation

    [6] Equations and Calculations are from Guide for rubber mixing process by Ms Gerlind Schubert [7 ] BS ISO 7742 Determination of Compression stress strain properties for Vulcanized

    Rubbers

    [8] Reference from (http://en.wikipedia.org/wiki/Mullins_effect) [9] Equations from Non-Linear Solid Mechanics by Gerhard A Holzapfel

    [10]Picture from http://en.wikipedia.org/wiki/Solenoid

    [11] Referenced from http://www.powerstream.com/Wire_Size.htm (AWG 23)

  • 45

    Appendix A : Calculation for Solenoids

    First a target Magnetic Flux Density of B = 400 mTeslas was set. This is a relatively high magnetic

    field but will be suitable for all experiments that need to be done. Furthermore the flux density can

    be changed by altering the current supply. The max current that the power supply can safely

    produce is 5A.

    Figure A1[10]: A typical solenoid field. Lines represent magnetic field lines. Dots and Xs are coil

    cross-section.

    Figure A1 shows a wire coil in cross-section and the arrowed lines represent the magnetic field lines.

    This shows that the field through a solenoid is very uniform and will be ideal for use when

    compressing samples.

    The basic equation for calculating the magnetic flux density at a inside the coil at a point away from

    the ends is given as:

    Where: B = magnetic flux density Unit is Tesla

    = K0 where k = relative permeability of core material and

    0=magnetic constant and is the permeability of a vacuum

    0= 4 x 10-7 unit is Tesla Metre per Ampere

    N =number of turns L = length of solenoid

    I = Current Unit is Amperes

    10 Picture from http://en.wikipedia.org/wiki/Solenoid

  • 46

    Therefore what is required to build this solenoid is the turns density N/L. With this the appropriate

    coil can be designed and manufactured.

    The core material is going to be air which has K=1

    B (Tesla) I (Amperes) K (T.m/Ampere)

    0.4 5 1 1.25664e-006

    Therefore:

    =1.5915e+006 turns/metre

    As you can see this number is completely impractical. 1.5 million turns in one metre would

    require a long amount of time to wind. Furthermore the wire diameter would be required to

    be 6.2832e-004mm diameter and this diameter would not support the necessary current

    required as specified by the American Wire Gauge. This gauge specifies that a wire will

    carry a maximum current of 4.7 Amperes with a wire diameter of 0.57404mm[11].

    For all the above reasons the idea of a solenoid to apply the magnetic field was rejected.

    11 Referenced from http://www.powerstream.com/Wire_Size.htm (AWG 23)

  • 47

    Appendix B: Faulty test setup results

    The results shown in figure B1 are from the original magnetic test setup which had been

    intended to be used. Unfortunately there were a few problems with the test setup. The

    problem was that the bottom test plates, which held a magnet between them (see figure 4

    from report), had a loose connection between them and the test machine. The force of

    attraction between the top and bottom magnets caused the whole bottom part of the setup

    to rise up a little. Once the samples were compressed and the applied force became larger

    than the force of attraction a rigid body movement took place moving the compression

    plates down creating a plateau in the force displacement curve as seen in figure B1.

    The problems that were caused by this were solved by decoupling the bottom magnet from

    the plates and allowing the attractive force of the magnets to hold the magnets in place

    while being separated by two wooden blocks as shown in figure 5.

    Figure B1: Results from faulty tests

    Unload Part for all cycles

    Upload Part for first cycle

    Upload Part for cycles 2,3,4

    Plateau caused by loose

    fitting