14
Hygiena International Ltd Assessment of a novel ATP monitoring device Executive summary The Hygiena EnSURE luminometer was very simple and easy to use requiring very little instruction. Its hand-held sturdy format allows it to be used in a brewery situation where hygiene assessment may be required. The calibration of the equipment did not show any drift over a 1 month period and the provided standards were also stable over this time. Measurements demonstrated good repeatability. In comparison with a competitor luminometer, the Hygiena unit appeared approximately 10 times more sensitive at the low ATP levels of 1 fmol and lower. This was also demonstrated by the difference in limit of detection which was 0.21 fmol ATP for the EnSURE luminometer and 2.72 fmol ATP for the competitor system. However, at very high ATP concentrations the unit showed signal saturation which is not considered an issue in the context of hygiene monitoring as such samples would be considered ‘fails’ anyway. The EnSURE and SuperSnap swab system satisfactorily detected beer residue of most products down to at least 1:100 dilution. However, on both systems, one beer showed very low readings and so did the two wines and the alcopop tested indicating that this technology is not suitable for residue detection of these particular drinks. In our experiments yeast cells could be detected down to concentrations of 100 cells/ml and there was a linear correlation between ATP bioluminescence and cell numbers up to 10 4 cells/ml, therefore this technique can give an indication of yeast cell numbers in solution. Background The objective of this project was to evaluate Hygiena’s new EnSURE ATP detection system for use in the alcoholic beverage industry as part of a hygiene monitoring programme, specifically as a post cleaning verification test for product residue. In the alcoholic beverage industry the contamination of the process plant and drink food and drink innovation Campden BRI March 2011 Instrument Assessment Report

food and drink innovation Instrument Assessment Hygiena ... · appeared approximately 10 times more sensitive at the low ATP levels of 1 fmol and lower. ... (RLU). The intensity of

Embed Size (px)

Citation preview

Hygiena International LtdAssessment of a novel ATP monitoring device

Executive summaryThe Hygiena EnSURE luminometer was very simple and easy to use requiring very little instruction. Its hand-held sturdy format allows it to be used in a brewery situation where hygiene assessment may be required. The calibration of the equipment did not show any drift over a 1 month period and the provided standards were also stable over this time. Measurements demonstrated good repeatability. In comparison with a competitor luminometer, the Hygiena unit appeared approximately 10 times more sensitive at the low ATP levels of 1 fmol and lower. This was also demonstrated by the difference in limit of detection which was 0.21 fmol ATP for the EnSURE luminometer and 2.72 fmol ATP for the competitor system. However, at very high ATP concentrations the unit showed signal saturation which is not considered an issue in the context of hygiene monitoring as such samples would be considered ‘fails’ anyway. The EnSURE and SuperSnap swab system satisfactorily detected beer residue of most products down to at least 1:100 dilution. However, on both systems, one beer showed very low readings and so did the two wines and the alcopop tested indicating that this technology is not suitable for residue detection of these particular drinks. In our experiments yeast cells could be detected down to concentrations of 100 cells/ml and there was a linear correlation between ATP bioluminescence and cell numbers up to 104 cells/ml, therefore this technique can give an indication of yeast cell numbers in solution.

BackgroundThe objective of this project was to evaluate Hygiena’s new EnSURE ATP detection system for use in the alcoholic beverage industry as part of a hygiene monitoring programme, specifically as a post cleaning verification test for product residue. In the alcoholic beverage industry the contamination of the process plant and drink

food and drink innovation

Campden BRIMarch2011

Instrument Assessment Report

Inst

rum

ent

Ass

essm

ent

Rep

ort with foreign micro-organisms must be prevented, as the potential cost of lost batches and customer

dissatisfaction due to poor hygiene is high. To ensure the microbiological integrity of the process, appropriate tests must be carried out. Although the traditional plate based microbiological method validates the microbiological integrity of the process, several days are usually required before a result is obtained, which is too slow for rapid countermeasures.Adenosine triphosphate (ATP) bioluminescence technology is being used for the rapid assessment of hygiene. ATP is an energy compound in living cells. Methods based on ATP-bioluminescence measure ATP using an enzyme which catalyses an ATP-specific bioluminescence reaction in which ATP is converted to produce light, quantified as Relative Light Units (RLU). The intensity of the emitted light is proportional to the concentration of ATP. Hygiena has developed a sensitive ATP detection system, which comprises of the EnSURE™ instrument, SuperSnap™ surface swabs and AquaSnap™ water testing dippers. Hygiena states that SuperSnap is the most sensitive ATP surface test in its portfolio of all-in-one devices. Whilst other ATP testing devices use freeze-dried enzymes, SuperSnap uses a liquid-stable enzyme, which is said to detect very small amounts of ATP and organic residue, giving consistent true results at low RLUs.The project compared the Hygiena Ensure + SuperSnap/AquaSnap detection system with another luminometer + surface/water sampler system from a leading global competitor company.

Materials and methodsDetection of ATP standardsAn ATP calibration curve was constructed to demonstrate the sensitivity and limit of detection of the systems and the ratio of RLUs to ATP concentration. The ATP standards were provided by Hygiena at 0.1, 1.0, 10, 100 and 1000 femtomoles. Ten repeat measurements were taken at each ATP standard concentration for each instrument. The sample size was 10 µl pipetted onto the swabs and the ATP readings recorded in the two luminometers.The stability of the instrument calibration was tested by using a stable positive and negative control provided by Hygiena, and measuring these standards after one week and one month. The stability of the 2 controls was ensured by freezing them over the month.

Detection of ATP in alcoholic beveragesThe level of ATP bioluminescence was determined for products that represent the alcoholic beverages market (Table 1). The beverages were serially diluted in sterile ATP free water (supplied by Hygiena) down to 10-4 dilution, except for the flavoured alcoholic beverage which was diluted down to 10-1 dilution. The sample size was 10 µl sample pipetted onto the swabs and the ATP reading recorded in the two luminometers. The tests were repeated ten times for each system and each beverage at each dilution.

Ale

Lager

Cider

Red wine

White wine

Flavoured alcoholic drink

Ale A

Ale B

Ale C

Ale D

Ale E

Lager F

Lager G

Lager H

Lager I

Lager J

Cider

Red wine

White wine

Alcopop

Alcoholic beverage type Details

Table 1: Alcoholic beverages used in the study.

Inst

rum

ent

Ass

essm

ent

Rep

ort Detection of low levels of yeast contamination in diluents

An isolate of Saccharomyces was grown aerobically at 25°C in Yeast and Mould (YM) broth for four days. The total yeast count was determined. The isolate was then serially diluted in quarter-strength Ringer's solution down to 10 cells/ml. A 100 µl sample of each dilution was collected using an AquaSnap dipper or the competitor dipper. These tests were repeated ten times for each system at each dilution. A negative control of YM broth diluted in Ringer's solution was also included. A fresh aliquot was used for each sampling to prevent carry-over of extractant from the dipper to the stock solutions.

ResultsEquipment evaluationAt the start of the equipment evaluation project a short training course on the EnSURE system was given by Hygiena. The trainers were friendly and helpful. All questions regarding the use of the system were answered. An instruction booklet was supplied and technical support was promptly provided through a local representative, either in person, over the phone or by email. Instructions were simple and easy to understand.The system is very similar to the competitor system. It requires only minimal setting up making it almost immediately ready for use. It was easy to use, even suitable for an unskilled operator and required little training.The function allowing the user to scroll back through the results, which displays the number, time and date of results, was very useful and was used to double check that the manual recording of the readings was correct.The positive and negative control provided by Hygiena gave confidence to the user that the machine was correctly calibrated. Physically, the machine proved to be well built and strong withstanding the constant use it was subjected to over the course of the experimental days of this project. Its hand-held format allows the device to be conveniently used anywhere as it is easily transported. This is important since the brewery setting calls for a robust portable piece of kit.The EnSURE machine is battery powered and the battery did not require replacing during the length of the study whilst the competitor machine had to be charged on occasions. However, it is to be noted that both machines were in constant use over the course of the equipment evaluation, which is unlikely to reflect the way these machines would be used in a hygiene monitoring programme in a drinks manufacturing situation.A slight malfunction with the EnSURE system occurred on occasions when placing a SuperSnap swab in the machine, rather than measuring, it would ask the operator to remove the swab, close the lid and then replace the swab. The operator would then press the OK button to get the RLU reading. However, at this point the machine would restart.

Detection of ATP standardsThe stability of the calibration of the EnSURE machine was proven by using a stable positive and negative control. The readings for the controls remained very similar throughout the 28 day period (Table 2).

Table 2: EnSURE system’s positive and negative controls over 28 days.Hygiena claims that its new ATP detection system has good linear correlation of ATP concentration to bioluminescence even at low ATP concentrations. This was tested by constructing an ATP calibration curve to demonstrate the sensitivity and limit of detection of the system and the ratio of RLUs to ATP concentration.Table 3 shows the results. A good linear correlation is obtained between RLU and the ATP concentration (fmols) (see Figures 1 and 2). The average ratio of RLUs to ATP concentration (slope) for the EnSURE + SuperSnap system was found to be 1 fmol ATP ≈ 5 RLU.

Days0 7 28

ControlPositive 130 130 130 130 130 130 128 128 128Negative 0 0 0 0 0 0 0 0 0

Inst

rum

ent

Ass

essm

ent

Rep

ort

Table 3: ATP bioluminescence of ATP standards

Figure 1: Regression fit of ATP bioluminescence vs ATP standard concentration (fmols)for the EnSURE system.

Figure 2: Logarithmic plot of ATP bioluminescence vs ATP standard concentration.

Sample

1

2

3

4

5

6

7

8

9

10

Mean

0 fmol

0

0

0

0

0

0

0

0

0

0

0

0.1 fmol

0

0

0

0

0

0

0

1

1

1

0.30

1 fmol

3

3

4

5

2

5

5

5

5

6

4

10 fmol

37

53

53

50

47

49

34

45

17

49

43

100 fmol

328

477

678

626

602

674

503

530

580

382

538

1000 fmol

5694

5390

5130

5103

2865

4674

3923

4335

5388

5360

4786

ATP bioluminescence (RLU) for each ATP standard

Inst

rum

ent

Ass

essm

ent

Rep

ort As all ten readings for the blank sample were zero, to be able to determine the LOD, the smallest

difference in readings (1 RLU) was used. Together with the slope of the calibration curve, the limit of detection (LOD) was calculated as:

Ten blank readings on the competitor’s luminometer were also determined giving values between 10 and 20 RLU. The limit of detection (LOD) for this system was calculated as:Mean blank = 15.3 RLUStandard deviation of blank = 2.63 RLUSlope of calibration curve = 8.401 RLU/fmol

This shows that at low ATP concentrations the EnSURE system is approximately 10 times more sensitive than the competitor system.

Figure 3: The bioluminescence values for ATP standards measured by the EnSURE and competitor system on day 0, 7 and 28.

Analysis of variance (ANOVA) was applied to determine any significant changes to the ATP bioluminescence of the five standards over time. Figure 3 shows the regression plots of ATP bioluminescence for each ATP concentration + luminometer combination. For a few ATP concentration + luminometer combinations there was statistically significant evidence of changes in RLU results over the 28 days (see: 1fmol + EnSURE; 100fmol + Competitor; 1000fmol + EnSURE). However, considering the changes are very small and not associated with a particular standard or device, the judgment is made that the ATP standards are stable over 28 days.Figure 4 shows the calibration curves for the EnSURE and competitor systems measured on the three different days. As one can see the EnSURE system was more sensitive compared to the competitor system at distinguishing between levels of ATP of 0.1 and 1 fmols, showing better linearity at these low ATP concentrations whereas the competitor device shows a tailing off. As observed previously for the ‘positive’ standard, the five ATP standards and the calibration of both devices appeared to be stable over the 28 days.

LOD = = = 0.21 fmol ATPSmallest difference in readings 1 Slope of curve 4.781

LOD = = = 2.72 fmol ATPMean blank + (3 x SD) 23.19 Slope of curve 8.401

Inst

rum

ent

Ass

essm

ent

Rep

ort

Figure 4: ATP calibration curves for the EnSURE and competitor systems on day 0, 7 and 28.

Detection of ATP in alcoholic beveragesFigure 5 shows the average ATP bioluminescence readings for the 14 diluted alcoholic drinks obtained with the EnSURE luminometer. The products appear to fall into 3 groups: those showing the highest readings (Ale A, Ale E, Ale B, Lager G, Ale C and Lager I); those showing somewhat lower values (Ale D, Lager J, Lager H and Cider) and finally those drinks with the lowest ATP values (Lager F, Red and White Wine). The ATP levels of the alcopop were practically undetectable with the EnSURE system. The reason for the differences in products is likely to do with the raw materials used, as the cellular ATP of the plant-matter will contribute to the ATP in the drinks. The flavoured alcoholic drink is a synthetic drink, essentially sugar added to an alcohol solution, whilst beer, cider and wine are derived from agricultural raw materials. The malt, apples and grapes from which beer, cider and wine are made from respectively would also have a microbial load additionally imparting ATP to these alcoholic beverages. Also some products may contain ingredients that exhibit an inhibiting/quenching effect on the bioluminescence reaction. The graphs in Appendix 1 illustrate the ATP bioluminescence of the alcoholic beverages that were tested with each system at each dilution. From these graphs very few outlying individual data points can be identified (one outlier in the 1:10 dilution of white wine data and one outlier in the 1:10 flavoured alcoholic drink data, both measured using the competitor’s system). The competitor luminometer read higher bioluminescense values for all samples. Repeatability of ATP measurements of the neat drinks seemed somewhat better for the EnSURE system than for the competitor machine (wider spread in 10 repeat measurements).

Figure 5: Average ATP bioluminescence readings for the diluted alcoholic beverages.Measurements taken with the EnSURE system.

4

3

2

1

0

1000.0100.010.01.00.1

4

3

2

1

0

EnSURE

ATP standards (fmols)

Mea

n of

log(

RLU

)Competitor

07

28

Days

Inst

rum

ent

Ass

essm

ent

Rep

ort Generally, linearity between ATP bioluminescence and product concentration was achieved on both

systems with linear correlation coefficients mostly lying between 0.92 and 0.97. The readings would be expected to decrease by one log order for each dilution and the bioluminescence measurements showed this for most of the drinks (down to dilution 1:1000 for the beverages with the highest readings and down to 1:100 dilution for beverages with intermediate readings). However, the drinks showing the lowest values did not follow this pattern. There was no indication of signal saturation at the higher concentrations. However, a levelling off at lower concentrations was noticeable and this was more pronounced for the competitor’s system for which ATP bioluminescence levelled off at about 15 RLU for both of the lowest product concentrations tested. The EnSURE system, on the other hand, did show better linearity at these low concentrations indicating that the sensitivity of the EnSURE system at these low ATP concentrations is better than for the competitor system.

Detection of low levels of yeast contamination in diluentsFollowing aerobic growth of Saccharomyces at 25°C in YM broth for four days, the total yeast count was calculated and found to be 1x107 cells/ml. The cell suspension was serially diluted down to 10 cells/ml. The same dilution series was prepared for the YM broth containing no cells to serve as a control.As one can see from Figure 6 the results for the ATP bioluminescence of the neat (1x107 cells/ml) yeast solution measured on the competitor’s system was very high compared to the same samples measured on the EnSURE machine. This is because the EnSURE machine gives a maximum reading of four digits, whilst the competitor machine can give readings of seven digits. The EnSURE machine gave its maximum reading of 9999 for all the neat (1x107 cells/ml) yeast solution aliquots. The bioluminescence signal was saturated and showed a plateau effect at the three highest cell concentrations (105 to 107 cells/ml). This could be interpreted as a lack in accuracy for the Hygiena system at these high ATP concentrations. However, when put into the context of a hygiene monitoring programme, the brewing plant which underwent a post cleaning verification test for product residue would fail if the results reached these high levels anyway.

Figure 6: Logarithmic plot of the ATP bioluminescence of different concentrations of yeast solution measured by the EnSURE and competitor system.

The measurements of the corresponding diluted broth samples gave readings between 15 and 508 RLU for the Hygiena system and between 36 and 2218 RLU for the competitor’s system. The broth contains yeast and malt extract which is the likely cause for the ATP detected. The YM readings were subtracted from the yeast readings and the resulting data is presented in Figure 7. At the yeast cell concentration of 1x104 cells/ml the datapoint for the competitor’s system was negative (this appears to be an outlier). At the lowest cell concentration (101 cells/ml) the values for both systems were

Machine

Dilution

Compe

titor

EnSU

RE1:10

0000

0

1:10

0000

1:10

000

1:10

001:10

01:10

Neat

1:10

0000

0

1:10

0000

1:10

000

1:10

001:10

01:10

Neat

6

5

4

3

2

1

0Log1

0 A

TP b

iolu

min

esce

nce

(RLU

)

Inst

rum

ent

Ass

essm

ent

Rep

ort negative (the 0.1 ml used for the analysis may not have contained any yeast cells). The signal saturation

at the highest concentrations for the EnSURE is still noticeable. In our experiment, the lowest detectable yeast cell concentration for both systems was 100 cells/ml which corresponds to the detection of 10 yeast cells in the analysed samples (0.1 ml sample volumes). Linear regressions were determined for the Hygiena as well as the competitor product. For the EnSURE system only the data points below saturation (up to cell concentration of 1x104 cells/ml) were employed, whereas for the competitor system all data were used.EnSure: ATP (RLU) = 2E+06 x dilution factor (r2 = 0.99)Competitor: ATP (RLU) = 9.5E5 x dilution factor (r2 = 0.99)As all ten readings for the EnSURE blank sample (without any yeast cells) were zero, to be able to determine the LOD, the smallest difference in readings (1 RLU) was used. Together with the slope of the calibration curve, the limit of detection (LOD) was calculated as:

Figure 7: Logarithmic plot of the ATP bioluminescence of the yeast dilution series with the bioluminescence of the corresponding YM broth concentration subtracted.

DiscussionHygiena’s EnSURE system was found to be easy to use and the short training provided would allow even unskilled persons to operate the equipment without a problem. The small hand-held size allows the device to be employed at the manufacturing site and its robust build means that it will not suffer damage easily.The provided positive and negative controls were seen to be stable over 28 days. The “Detection of ATP standards” experiments demonstrated the sensitivity and repeatability of the EnSURE unit + SuperSnap swab to detect ATP over a range of concentrations (0 to 10,000 fmol). The EnSURE machine did not drift following calibration and this was proven through testing the standards over a one month period. Only minor inconsistent instabilities over time were detected for some of the ATP standards, most showed good stability over the 28 days. The average ratio of RLUs to ATP concentration for the EnSURE system was found to be 5 RLU ≈ 1 fmol ATP. Both systems would on occasion give outlier results. The Hygiena system proved to be sensitive at low ATP levels, more so than the competitor system which showed a tailing off at the lowest concentrations. The LOD was determined as 0.21 fmol ATP showing an approximately 10 times higher sensitivity compared to the competitor’s system (LOD = 2.72 fmol ATP).A range of alcoholic beverages (lager, ale, wine, cider, alcopop) were serially diluted and tested on both systems. The products fell into 3 groups with either high, medium or low ATP bioluminescence measurements. The expected one log reduction in readings per dilution was seen for the 2 groups exhibiting the higher light values. Comparison between results of the two systems showed that the

LOD = = = 4 cells/mlSmallest difference in readings 1 Slope of curve 0.248

Inst

rum

ent

Ass

essm

ent

Rep

ort competitor’s luminometer gave higher readings than the EnSURE device. And the sensitivity of the

Hygiena equipment was higher than for the competitor’s at the highest dilutions (lowest ATP levels). The repeatability of measurements seemed slightly better with the EnSURE system.A brewing yeast culture was serially diluted to give cell concentrations ranging between 107 and 101 cells/ml. As a control a matching dilution series with the yeast growth medium only was also tested. A general increase in ATP bioluminescence with rising cell number was detected and this rise followed a linear relationship at cell concentrations up to 104 cells/ml. The slope of the resulting the linear curves on both systems did not strictly follow the expected one log bioluminescence decrease for each dilution step, even with the values obtained for the growth medium subtracted. It is conceivable that some compounds, possibly from the medium, interfere with the reaction. Also the yeast cells may not contribute all equally to the free ATP content in the suspending liquid possibly due to the ATP extraction not being sufficient and/or variation in cell size and ATP content. At the highest cell concentrations > 105 cells/ml the bioluminescence of the Hygiena system was not as sensitive as the competitor’s. A signal saturation plateau was noticeable. This may be because the EnSURE luminometer gives a maximum reading of 9999. The LOD for the EnSURE system was determined as 3 cells/ml – in our experiment 100 cells/ml were reliably detected. This is a low yeast cell concentration determined very rapidly (15 secs) as compared to the standard microbiological methods.In summary, the EnSURE luminometer is an easy to use device which is very portable and sturdy. Its calibration is stable over 28 days and the provided ATP standards were also stable over this period. The limit of detection was very low – below 1 fmol ATP. In comparison with the competitor device the Hygiena luminometer appeared more sensitive at low ATP levels and showed slightly better repeatability. However, at the higher bioluminescence levels it showed signal saturation (EnSURE + AquaSnap) which was not seen with the competitor device. This is not a cause for concern as at these ATP levels of residue the hygiene test would be considered a fail anyway. The unit performed well with most of the alcoholic beverages tested and was able to detect beer and cider residues at a 1:100 dilution. ATP detection was not a suitable method for the detection of wine and alcopop as the bioluminescence values were very low even at the high product concentrations. Interestingly, one of the beers also showed very low readings; therefore it would be recommended to test a beverage’s ATP bioluminescence output before using this technique for residue detection. There was a linear correlation between yeast cell concentration (up to 1x104 cells/ml) and bioluminescence output and yeast cells were reliably detected at 100 cells/ml and above. The system can therefore be used for detection of yeast cell residue e.g. in CIP rinse waters.

Appendix 1

Ales:

ATP bioluminescence of dilution series of Ale C shown in linear and logarithmic plots. Linear regressions have been fitted; in the logarithmic plot the intercept is set at zero.

y = 3928.7x + 30.25R² = 0.9277

y = 1702.1x + 32.944R² = 0.9168

0

1000

2000

3000

4000

5000

6000

0.0001 0.001 0.01 0.1 1

ATP b

iolu

min

esce

nce (

RLU)

Dilution factor

Ale C

Competitor

Ensure

y = 3961.9xR² = 0.9274

y = 1738.3xR² = 0.9151

1

10

100

1000

10000

0.0001 0.001 0.01 0.1 1

ATP b

iolu

min

esce

nce (

RLU)

Dilution factor

Ale C

Competitor

Ensure

Inst

rum

ent

Ass

essm

ent

Rep

ort

ATP bioluminescence of dilution series of Ale A shown in linear and logarithmic plots. Linear regressions have been fitted; in the logarithmic plot the intercept is set at zero.

ATP bioluminescence of dilution series of Ale E shown in linear and logarithmic plots. Linear regressions have been fitted; in the logarithmic plot the intercept is set at zero.

ATP bioluminescence of dilution series of Ale D shown in linear and logarithmic plots. Linear regressions have been fitted; in the logarithmic plot the intercept is set at zero.

y = 2527.6x + 36.872R² = 0.9293

y = 3693x + 53.571R² = 0.9484

0

500

1000

1500

2000

2500

3000

3500

4000

4500

5000

0 0.2 0.4 0.6 0.8 1

ATP

biol

umin

esce

nce(

RLU)

Dilution factor

Ale AEnsure

Competitor

y = 2568.2xR² = 0.9284

y = 3752xR² = 0.9474

1

10

100

1000

10000

0.0001 0.001 0.01 0.1 1

ATP

biol

umin

esce

nce(

RLU)

Dilution factor

Ale AEnsure

Competitor

y = 2033.7x + 9.4608R² = 0.9635

y = 3480.9x + 31.865R² = 0.9326

0

500

1000

1500

2000

2500

3000

3500

4000

4500

5000

0 0.2 0.4 0.6 0.8 1

ATP b

iolu

min

esce

nce (

RLU)

Dilution factor

Ale EEnsure

Competitor

y = 2044.1xR² = 0.9634

y = 3516xR² = 0.9322

1

10

100

1000

10000

0.0001 0.001 0.01 0.1 1

ATP b

iolu

min

esce

nce (

RLU)

Dilution factor

Ale EEnsure

Competitor

y = 171.12x + 0.3728R² = 0.9683

y = 203.18x + 17.55R² = 0.9484

0

50

100

150

200

250

300

0 0.2 0.4 0.6 0.8 1

ATP b

iolu

min

esce

nce (

RLU)

Dilution factor

Ale DEnsure

Competitor

y = 171.53xR² = 0.9683

y = 222.48xR² = 0.9134

1

10

100

1000

0.0001 0.001 0.01 0.1 1

ATP b

iolu

min

esce

nce (

RLU)

Dilution factor

Ale DEnsure

Competitor

Inst

rum

ent

Ass

essm

ent

Rep

ort

ATP bioluminescence of dilution series of Ale B shown in linear and logarithmic plots. Linear regressions have been fitted; in the logarithmic plot the intercept is set at zero.

Lagers:

ATP bioluminescence of dilution series of Lager I shown in linear and logarithmic plots. Linear regressions have been fitted; in the logarithmic plot the intercept is set at zero.

ATP bioluminescence of dilution series of Lager G shown in linear and logarithmic plots. Linear regressions have been fitted; in the logarithmic plot the intercept is set at zero.

y = 2064.4xR² = 0.9516

y = 3349.1xR² = 0.9328

1

10

100

1000

10000

0.0001 0.001 0.01 0.1 1

ATP b

iolu

min

esce

nce (

RLU)

Dilution factor

Ale BEnsure

Competitor

y = 2031.5x + 29.94R² = 0.9526

y = 3312.7x + 33.044R² = 0.9333

0

500

1000

1500

2000

2500

3000

3500

4000

4500

0 0.2 0.4 0.6 0.8 1

ATP b

iolu

min

esce

nce (

RLU)

Dilution factor

Ale BEnsure

Competitor

y = 2342.7x + 35.978R² = 0.738

y = 1702.1x + 32.944R² = 0.9168

0

500

1000

1500

2000

2500

3000

3500

4000

4500

5000

0 0.2 0.4 0.6 0.8 1

ATP b

iolu

min

esce

nce (

RLU)

Dilution factor

Lager I

Competitor

Ensure y = 2382.3xR² = 0.7372

y = 2084.6xR² = 0.9185

1

10

100

1000

10000

0.0001 0.001 0.01 0.1 1

ATP b

iolu

min

esce

nce (

RLU)

Dilution factor

Lager I

Competitor

Ensure

y = 1475.2x + 27.906R² = 0.9703

y = 1337.4x + 31.486R² = 0.9341

0

500

1000

1500

2000

2500

0 0.2 0.4 0.6 0.8 1

ATP b

iolu

min

esce

nce (

RLU)

Dilution factor

Lager GCompetitor

Ensure y = 1505.9xR² = 0.9686

y = 1372xR² = 0.9316

1

10

100

1000

10000

0.0001 0.001 0.01 0.1 1

ATP b

iolu

min

esce

nce (

RLU)

Dilution factor

Lager G

Competitor

Ensure

Inst

rum

ent

Ass

essm

ent

Rep

ort

ATP bioluminescence of dilution series of Lager J shown in linear and logarithmic plots. Linear regressions have been fitted; in the logarithmic plot the intercept is set at zero.

ATP bioluminescence of dilution series of Lager H shown in linear and logarithmic plots. Linear regressions have been fitted; in the logarithmic plot the intercept is set at zero.

ATP bioluminescence of dilution series of Lager F shown in linear and logarithmic plots. Linear regressions have been fitted; in the logarithmic plot the intercept is set at zero.

y = 1475.2x + 27.906R² = 0.9703

y = 292.32x + 1.7805R² = 0.9453

0

100

200

300

400

500

600

700

0 0.2 0.4 0.6 0.8 1

ATP b

iolu

min

esce

nce (

RLU)

Dilution factor

Lager J

Competitor

Ensure

y = 442.33xR² = 0.8806

y = 294.28xR² = 0.9451

1

10

100

1000

10000

0.0001 0.001 0.01 0.1 1

ATP b

iolu

min

esce

nce (

RLU)

Dilution factor

Lager J

Competitor

Ensure

y = 481.98x + 16.515R² = 0.919

y = 369.83x + 0.8754R² = 0.939

0

100

200

300

400

500

600

700

800

0 0.2 0.4 0.6 0.8 1

ATP b

iolu

min

esce

nce (

RLU)

Dilution factor

Lager HCompetitor

Ensure

y = 500.14xR² = 0.9137

y = 370.8xR² = 0.939

1

10

100

1000

10000

0.0001 0.001 0.01 0.1 1

ATP b

iolu

min

esce

nce (

RLU)

Dilution factor

Lager H

Competitor

Ensure

y = 25.658x + 16.498R² = 0.6974

y = 26.432x + 0.7464R² = 0.9065

0

10

20

30

40

50

60

70

0 0.2 0.4 0.6 0.8 1

ATP b

iolu

min

esce

nce (

RLU)

Dilution factor

Lager FCompetitor

Ensure

y = 43.806xR² = -0.73

y = 27.253xR² = 0.9029

1

10

100

1000

10000

0.0001 0.001 0.01 0.1 1

ATP b

iolu

min

esce

nce (

RLU)

Dilution factor

Lager F

Competitor

Ensure

Inst

rum

ent

Ass

essm

ent

Rep

ort Wine:

ATP bioluminescence of dilution series of Red Wine shown in linear and logarithmic plots. Linear regressions have been fitted; in the logarithmic plot the intercept is set at zero.

ATP bioluminescence of dilution series of White Wine shown in linear and logarithmic plots. Linear regressions have been fitted; in the logarithmic plot the intercept is set at zero.

Other alcoholic beverages:

ATP bioluminescence of dilution series of Cider shown in linear and logarithmic plots. Linear regressions have been fitted; in the logarithmic plot the intercept is set at zero.

Inst

rum

ent

Ass

essm

ent

Rep

ort

Campden BRI, Centenary Hall, Coopers Hill Road, Nutfield, Surrey RH1 4HY United Kingdom

Email: [email protected] Tel: +44 (0) 1737 822 272 Web: www.bri-advantage.com

ATP bioluminescence of dilution series of Alcopop shown in linear plot. Linear regressions have been fitted.