Chapter 6 Lecture Spin Echo Imaging Methods Mark D. Herbst, MD, PhD • Two Main Types of MR methods – Spin Echo –uses RF pulse to get an echo – Gradient Echo –uses gradient pulse to get an echo

Chapter 6 Lecture Spin Echo Imaging Methods Mark D. Herbst, MD, PhD Two Main Types of MR methods –Spin Echo –uses RF pulse to get an echo –Gradient Echo

Embed Size (px)

Citation preview

Page 1: Chapter 6 Lecture Spin Echo Imaging Methods Mark D. Herbst, MD, PhD Two Main Types of MR methods –Spin Echo –uses RF pulse to get an echo –Gradient Echo

Chapter 6 LectureSpin Echo Imaging Methods

Mark D. Herbst, MD, PhD

• Two Main Types of MR methods– Spin Echo –uses RF pulse to get an echo– Gradient Echo –uses gradient pulse to get an

echo

Page 2: Chapter 6 Lecture Spin Echo Imaging Methods Mark D. Herbst, MD, PhD Two Main Types of MR methods –Spin Echo –uses RF pulse to get an echo –Gradient Echo

The spin echo process

• Net magnetization vector starts at equilibrium, pointing up along the +z direction.

• 90 degree RF pulse M now in xy plane• Spins in xy plane begin dephasing, and when

totally dephased, M=0.• 180 degree refocusing RF pulse is applied• Spins in xy plane begin to rephase, and echo

forms. This echo is captured and stored in k-space.

Page 3: Chapter 6 Lecture Spin Echo Imaging Methods Mark D. Herbst, MD, PhD Two Main Types of MR methods –Spin Echo –uses RF pulse to get an echo –Gradient Echo

The Spin Echo Process

Page 4: Chapter 6 Lecture Spin Echo Imaging Methods Mark D. Herbst, MD, PhD Two Main Types of MR methods –Spin Echo –uses RF pulse to get an echo –Gradient Echo

Three basic spin echo methods

• PD-weighted (proton density) – long TR, short TE

• T2-weighted – long TR, long TE

• T1-weighted – short TR, short TE

Page 5: Chapter 6 Lecture Spin Echo Imaging Methods Mark D. Herbst, MD, PhD Two Main Types of MR methods –Spin Echo –uses RF pulse to get an echo –Gradient Echo

Causes of Spin Dephasing

• Spin-spin interactions – energy exchange among the spinning nuclei causes relaxation (dephasing)

• Magnetic field inhomogeneity – cases spins to dephase, reversed by 180 degree RF refocusing pulse

Page 6: Chapter 6 Lecture Spin Echo Imaging Methods Mark D. Herbst, MD, PhD Two Main Types of MR methods –Spin Echo –uses RF pulse to get an echo –Gradient Echo

Multiple Spin Echo

• Done by using more than one 180 degree RF refocusing pulse

• Commonly used to get both PDW and T2W images at the same time

• Used in FSE (fast spin echo, turbo spin echo) to get images faster by filling k-space faster than conventional spin echo (CSE)

Page 7: Chapter 6 Lecture Spin Echo Imaging Methods Mark D. Herbst, MD, PhD Two Main Types of MR methods –Spin Echo –uses RF pulse to get an echo –Gradient Echo

Inversion Recovery

• Gives high level of T1 contrast• Used to suppress either water or fat (STIR

suppresses fat, FLAIR suppresses water)• Starts with a 180 degree RF inversion pulse,

placing M in the –z direction• Then, wait TI (inversion time) before applying

the 90 degree RF pulse that puts M in the xy plane

• Finally, apply the 180 degree RF refocusing pulse to form the echo for data collection, put it in k-space.

Page 8: Chapter 6 Lecture Spin Echo Imaging Methods Mark D. Herbst, MD, PhD Two Main Types of MR methods –Spin Echo –uses RF pulse to get an echo –Gradient Echo

TI: inversion time

• The time of inversion

• The time that M is allowed to stay inverted (-z)

• The time between the initial 180 degree inversion pulse and the 90 degree pulse that puts M in the xy plane.

Page 9: Chapter 6 Lecture Spin Echo Imaging Methods Mark D. Herbst, MD, PhD Two Main Types of MR methods –Spin Echo –uses RF pulse to get an echo –Gradient Echo

180 degree pulse

Page 10: Chapter 6 Lecture Spin Echo Imaging Methods Mark D. Herbst, MD, PhD Two Main Types of MR methods –Spin Echo –uses RF pulse to get an echo –Gradient Echo

IR pulse diagram (STIR and FLAIR)

Page 11: Chapter 6 Lecture Spin Echo Imaging Methods Mark D. Herbst, MD, PhD Two Main Types of MR methods –Spin Echo –uses RF pulse to get an echo –Gradient Echo

STIR – fat at null point

Page 12: Chapter 6 Lecture Spin Echo Imaging Methods Mark D. Herbst, MD, PhD Two Main Types of MR methods –Spin Echo –uses RF pulse to get an echo –Gradient Echo

FLAIR -- water at null point