PROVENATM
®POWERPICC
CATHETER
3Fr single lumen
3 ml/sec max flow rate
4Fr dual lumen
5 ml/sec max flow rate
5Fr triple lumen
5 ml/sec max flow rate
nursing guidelines RECOMMEND THE USE OF SMALLER- DIAMETER CATHETERS**
By combining advances in polyurethane material technology and manufacturing processes, Bard Access Systems has created smaller-diameter* PICCs that maintain high performance for infusion therapy.
Available in standard adult trays and in small patient trays with smaller* procedural components
* Comparison in relation to Bard® PowerPICC® catheters
**Infusion Nurses Society. “Infusion Therapy Standards of Practice.” Journal of Infusion Nursing. Volume 39. Jan/Feb 2016. ^Estimates based on a pump alarm setting of 500 mmHg and a full length catheter at 55 cm.
BARD ACCESS SYSTEMS, INC. - 605 North 5600 West - Salt Lake City, UT 84116 USA - BAS/PPPR/1016/0016 customer service: 800-545-0890 - clinical information: 800-443-3385 - bardaccess.com
Bard, PowerPICC, Provena, and Sherlock 3CG are trademarks and/or registered trademarks of C. R. Bard, Inc. © 2016 C. R. Bard, Inc. All rights reserved.
PROVENATM
®POWERPICC
CATHETER
PowerPICC® Provena™ Catheter - 3 F single lumen and 4 F dual lumen cathetersThe PowerPICC® Provena™ Catheters are indicated for short or long-term peripheral access to the central venous system for intravenous therapy, blood sampling, power injection of contrast media, and allows for central venous pressure monitoring. For central venous pressure monitoring, it is recommended that a catheter lumen of 20 gauge or larger be used.
PowerPICC® Provena™ Catheter - 5 F triple lumen cathetersThe PowerPICC® Provena™ catheter is indicated for short or long term peripheral access to the central venous system for intravenous therapy, power injection of contrast media, and allows for central venous pressure monitoring. For blood sampling, infusion, or therapy use a 4 French or larger catheter. The maximum recommended infusion rate is 5 mL/sec for power injection of contrast media. For central venous pressure monitoring, it is recommended that a catheter lumen of 20 gauge or larger be used.
Please consult product labels and inserts for any indications, contraindications, hazards, warnings, precautions, and directions for use.
CATHETER SIZE MAXIMUM FLOW RATE
3 F Single Lumen 3 mL/sec
4 F Dual Lumen 5 mL/sec
INDICATIONS FOR USE
3 FR SL POWERPICC® PROVENA™ CATHETER
4 FR DLPOWERPICC® PROVENA™ CATHETER
5 FR TLPOWERPICC® PROVENA™ CATHETER
Outer Diameter (Fr) 3 4 5
Lumen Size (Ga) 18 18/18 18 19/19
Max Power Injection Rate (ml/sec)
3 5 5 N/A
Max Achievable PumpFlow Rate^ (ml/hr)
Saline (1.0 cP) >999TPN (2.0 cP) >999 Blood (3.5 cP) 621
Saline (1.0 cP) >999TPN (2.0 cP) 812 Blood (3.5 cP) 464
Saline (1.0 cP) >999 TPN (2.0 cP) >999 Blood (3.5 cP) 649
Saline (1.0 cP) >999TPN (2.0 cP) 536 Blood (3.5 cP) 307
Gravity Flow Rate (ml/hr) 645 531 824 260
Priming Volume (ml) 0.53 0.52 0.67 0.43
ALSO AVAILABLE WITH SHERLOCK 3CG™ TIP CONFIRMATION SYSTEM
please contact your BARD® Access Systems representative call toll-free at 1-800-545-0890 or visit us on the web at www.bardaccess.com