Upload
marcelo-gomes
View
158
Download
5
Embed Size (px)
Citation preview
The following article is a translation of parts of the original
publication of Karl-Ludwig Bath in the german astronomical
magazine:
“Sterne und Weltraum 1973/6, p.177-180”.
The publication of this translation on the interferometer wiki
(http://starryridge.com/mediawiki-1.9.1) is with kind permission
of the author and the copyright holders of the original article at
Sterne und Weltraum, represented by Uwe Reichert, SuW
Heidelberg, Germany. Translation by Andreas Derwahl.
fig. 1: schematic representation of the interferometer, working principle see text, R is the autocollimation focus.
KARL-LUDWIG BATH
A simple interferometer for testing astronomical optics
Fellow stargazers have access to
a whole range of qualitative and
quantitative methods for the
analysis of astronomical optics.
[1]. There are different reasons,
however, that interferometric
methods are virtually non existent
among amateur telescope
makers. The following pages
describe an interferometer that
can be easily assembled and
adjusted with little practice, but
still gives accurate quantitative
results and is made from parts
that can be found in every cheap
binoculars.
To sum up a few of the specifics:
- The interferometer is suitable
for all usual f-ratios and focal
lengths down to about 25 cm.
- It can be used with white,
unpolarized light.
- The interferometer yields
bright interferograms. When
using a laser the efficiency at
exit port A2 (see fig. 1) is up
to 50%.
- The fringe contrast is high
and always 100% at exit A1
(translator remark:
independent on the splitting
ratio of the beam splitter).
- The interferogram is free of
false light and all secondary
reflections can be suppressed.
- Both exits (A1 and A2) are
complementary, meaning at
exit A1 the zero order fringe
is bright, at exit A2 it is dark.
In order to make the instrument
usable for people that are not
acquainted with interferometric
work, it shall be described in
maybe greater detail than seems
necessary at first sight.
Working principle:
Let us have a look at fig. 1: the
assembly from lamp Q to baffle
B2 is used to produce a
collimated beam, which is already
given when using a laser as a
source. The beam is divided by
the beam splitter cube into two
coherent parts, beam 1 and 2,
that are capable of interfering.
The beam that is reflected from
the splitting surface inside the
cube, beam 1, creates an image
Q’ of the baffle B1 on the concave
test mirror, which for simplicity is
drawn as spherical.
After reflection from the mirror
surface and transmission through
the small symmetrical biconvex
lens L3, beam 1 forms a spherical
wave with its centre at P1. The
quality of this “reference wave”,
as we might call it, is not
influenced by the defects of the
test mirror, because the reflection
at Q’ is generated by only a small
part of the surface, which can be
regarded as defect free.
Beam 2 passes through lens L3
on the way to the mirror and
illuminates the complete diameter
as a spherical wave. The defects
of the mirror are imprinted on
this wave, forming a second light
wave with origin at P2. The two
wave fronts with centres P1 and
P2 are now mixed by cube W1
and their interference can be
observed at exit A2 or - with a
second beam splitter cube W2 -
at exit A1.
The interferometer does not work
in strict auto collimation, i.e. the
reflected beam 2’ does not
coincide with beam 2. Therefore
the diameter of the usable field of
view of the test mirror/objective
must be bigger than the P1-P2
distance, otherwise it will show
astigmatism and coma, defects
that should not be present on axis
of the test specimen.
Optical parts:
1) The light source Q: we can
use for example a halogen
incandescent lamp with
cylindrical coil, or if not
available even a flashlight /
torch will do. Best of course is
a laser. In order to avoid
extraneous interference all
surfaces should be cleaned
carefully and after adjustment
the parts in direct illumination
by the collimated beam
should be made free of dust
with a soft brush. If the laser
beam is too small for a given
f-ratio, it can be expanded
some without disadvantage
by a single negative lens in
front of the beam splitter.
2) A camera lens of focal length
around 50-mm or a binocular
eyepiece serves as
OBJECTIVE L1.
3) The pinhole B1 is made for
example from a piece of tin
foil with different sized holes
to choose.
4) OBJECTIVE L2 is stepped
down to 5 to 10 mm; its focal
length is chosen such that the
image Q’ of baffle B1 is
smaller than 1/10 of the test
mirror diameter.
5) The beam splitter cube W2 is
only necessary for exit A1 and
should be removed when
using exit A2 (A2 provides
four times the image
brightness of A1). If we don’t
have a cube if necessary we
can cement two Porro prisms
with water or use a thickish
glass plate as beam splitter.
6) The beam splitter cube W1
should have a minimum edge
length of 25 mm. If need be
we can build one by
cementing two suitable Porro
prisms with largely any oil,
for example sun flower oil. If
we used the oil sparingly the
cube is mechanically stable
and compared with
commercial beam splitter
cubes it has the advantage of
an accessible fourth face and
hence the exit A2 becomes
available.
7) The focal length of the
symmetrical biconvex lens L3
is less than 1/20 of the test
mirror’s focal length and
under 15 mm in diameter. Its
defects are compensated
automatically even if it is
tilted in the beam path, hence
a corrected system would not
have any advantage. It
should be mentioned that an
asymmetrical lens (e.g.
plano-convex) can be used.
In this case, however, the
lens has to be adjusted
carefully and the test mirror
f-ratio should be at least 10.
If, because of a small usable
field of view of the test
mirror, the P1-P2 distance
(i.e. beam separation, remark
of the translator) has to be
very small, the lens L3 can be
ground close to the lens
centre.
[Comment KLB: with half
moon beam 2 is cut from
above so that the lower half
of the mirror is NOT
illuminated.]
Assembly:
For lens L3 and test mirror PR we
require adjustment in all three
coordinates. At least one of these
two elements has to be fine
adjustable in height
(perpendicular to the drawing
plane). Further, the centers of all
elements must be adjustable to
the same height.
We begin assembly with the
lamp. Its coil is tilted a small
amount from the interferometer
axis. The diaphragm B1 is placed
at a distance from the lamp that
equals approximately 4.5 times
the focal length of lens L1. We
center the emerging light cone on
the test mirror. Next we insert
the objective L1 and project the
image of the inside wall of the
lamp coil onto baffle B1.
This in turn is projected with L2
onto the centre of the test mirror.
If the objective does not possess
a built in diaphragm we insert
another baffle B2 into the beam
that is adjustable in both
directions perpendicular to the
interferometer axis.
The assembly of both beam
splitter cubes is not critical. Cube
W1 acts on beam 2 just like a
plane parallel plate, which here
can only cause a lateral beam
shift. Beam 1 however, which is
reflected off the splitting surface,
has to be adjusted on the mirror
centre by rotating the cube.
Because we want to keep the
requirements for the field of view
of the test mirror small, we will
choose a small distance between
both beams leaving the cube, for
example 10 mm. The position of
beam 2 is set by moving baffle
B2. The beam separation is
adjusted by moving cube W1
along the direction indicated by
the arrow in Fig. 1.
Finally we insert lens L3 in beam
2 at a distance from W1 equalling
its focal length. Only when testing
fast mirrors we place the beam
splitter cube closer to the lens.
In order to avoid light passing the
lens on its side we may need to
reduce baffle B2 to a smaller
beam diameter. It should also be
small enough so that the light
cone created by L3 does not over-
illuminate the test mirror by
much.
The distance of the test mirror
from L3 equals its radius of
curvature (or its focal length if it
is an astronomical objective with
a plane mirror). Now the
assembly of the interferometer is
finished and we can take on the
alignment.
Fig. 2: Elimination of objectionable reflections
Alignment
As a small tool we need some
strips of stiff paper. We place one
strip under each beam splitter
cube to be able to turn them in
small amounts. Now we move
lens L3 or baffle B2 in the plane
normal to the beam and center
the light cone leaving the lens on
the test specimen. If this an
objective we cover the flat mirror
(Fig. 4) with black paper and
adjust the objective axis with the
help of the lens reflections to lie
on the interferometer axis.
Following this we intersect beam
1 with a white paper strip at P2
and adjust the image formed by
the test mirror/lens of the lens
focus F exactly opposite P1 into
the beam center. This is achieved
by tipping, tilting and moving the
mirror or objective along the
optical axis.
Now we reach the final phase of
adjusting. We look into either exit
A1 or A2 into the interferometer
(take care when using a laser,
translator remark), and we will
see all kinds of reflections: one is
from cube W2 -- it can be moved
by rotating the cube. Another is
from cube W1 (see Fig.2). Finally
there can be reflections from lens
L3; they can be removed by
tilting the lens.
If we were careful during
assembly and adjustment so far,
we can now see the two
diffraction disks from P1 and P2.
P2 is only visible inside the test
objects outline and can be
recognized more easily by moving
our head back 20 to 30 cm.
The following fine adjustment
positions the interferometer or
the test object so that both
diffraction disks have equal size
and coincide. The size is
controlled by changing the L3-PR
distance. The discs are made to
coincide by tipping and tilting the
mirror (or plane mirror if an
objective is being tested) by small
amounts. Another method is to
adjust the lens height of L3 and
rotate cube W1. Doing this the
beam will move on the test
object, but it should not leave it.
This second method is more
convenient but affects the
illumination of the test object.
If we managed to get the two
spots to the same size and to
superimpose we can now see the
interference fringes. To get a
feeling for this phenomenon we
now alter the previous
adjustment steps by very small
amounts. The focus of the beams
that pass the central area of the
test specimen - which is of
biggest interest to us – is called
paraxial focus. We can find it
easily if we bring the center of the
visible ring system to the centre
of the fringe pattern (i.e. get the
bulls eye, translator remark), and
only then adjust the L2-PR
distance. If by accident we lose
the fringe pattern by large
misalignment, it is not advisable
to search for the diffraction disks
at the interferometer exit, but
rather start again from the
beginning of this paragraph.
Fringe pattern
Let us assume the test object is a
perfectly spherical mirror. In this
case with the interferometer at
the centre of curvature the fringe
pattern shows an evenly bright
(exit A1) or dark (A2) surface. If
we now move P1 or P2 sideways
or in height (translator remark:
X,Y coordinates,), straight,
parallel and structure-less fringes
appear (see also the numbered
front cover pictures) regardless of
their orientation. If the test
specimen has surface defects it
clearly shows in the fringes
(Picture 1 and 2).
Pictures 1 and 2
We have to examine these
interferences in greater detail
because we want to know if
maybe the usable field of view is
smaller than P1-P2, and what
defects the test specimen shows
for on axis and tilted rays. The
interferograms are so manifold,
that we can only deal with
spherical aberration, astigmatism
and coma here, and have to
restrict ourselves in most cases to
the optical axis.
Irregular variations of the
wavefront are reflected by
irregular variations in fringe width
and directions. We see an
example in picture 2, the
interferogram of a Fraunhofer
objective of unknown origin. It
shows 1 lambda aberration for
single pass (see pictures 3 and
4).
Pictures 3 and 4
The maximum aberration
following picture 1 is lambda/15.
Of special interest to us is of
course the interferogram of the
spherical aberration for the
central rays, the paraxial focus
(picture 3) and next to it (picture
4 and also picture 2).
Additional parts of the article
mostly dealing with fringe
analysis are omitted here because
electronic data reduction has now
taken the place of visual
geometric analysis. Appropriate
and powerful software is available
today at:
“http://starryridge.com/mediawiki-
1.9.1/index.php?title=Interferogram_Analysis”.
Two final remarks on fringe
photography. To avoid vignetting
a small Kepler telescope with a
power of 2 to 4 between the
interferometer and the camera is
advisable. Further the camera has
to be focused onto the rim of the
mirror under test, otherwise the
fringes get frayed out at their
ends.
Literature
[1] K.-L. Bath, Ein einfaches
Common Path Interferometer,
Optik 36 (172) 349.
[2] Original article in Sterne und
Weltraum, 1973/6, 177 –
180. Translation from German
by Andreas Derwahl with kind
permission of Sterne und
Weltraum, Heidelberg.