P BP BTT
CALIBRATION OFAC-DC CURRENT TRANSFER STANDARDS
BASED ONCALCUABLE THERMAL CONVERTERS ON
QUARTZ SUBSTRATE
Torsten Funck
22
P BP BTT
EM Day 2007 ACEM Day 2007 AC--DC current transfer standardsDC current transfer standards
Contents
• Introduction
• Quartz-PMJTCs for current transfer
• Calculation of the current transfer difference
• Validation of the model used for calculation
• Measurement setup
• AC-DC current transfer standards
• Step-up chains
• Improved measurement uncertainties
• Conclusions
33
P BP BTT
EM Day 2007 ACEM Day 2007 AC--DC current transfer standardsDC current transfer standards
Introduction
Electrical quantities are defined, realized an maintained at direct current (dc).
For alternating current (ac) calibrations, traceability is given due to ac-dc transfer.
For current, the measured quantity is δ i , the ac-dc current transfer difference:
dc
dcaci I
II=δ −
44
P BP BTT
EM Day 2007 ACEM Day 2007 AC--DC current transfer standardsDC current transfer standards
Quartz-PMJTCs
Planar multijunction thermal converters (PMJTCs) on a quartz substrate are used as a calcuable standard
+ Very low conductivity of the substrate
+ Low dielectric constant εr
+ Low, calculable transfer difference
- Difficult to manufacture
- Low thermal time-constant
- Low sensitivity
55
P BP BTT
EM Day 2007 ACEM Day 2007 AC--DC current transfer standardsDC current transfer standards
Calculation of δi due to heater impedance
{ } 1Re
Hi1 −
ZR=δ
LH /10
LH /10
RH/10
RH
/10
C2/10
C2/10
C1/10
0
0
0
0
G 2vs (f)
G2vs (f)
G1vs (f)
66
P BP BTT
EM Day 2007 ACEM Day 2007 AC--DC current transfer standardsDC current transfer standards
Calculation of δi due to stray capacitances
LH
LH
LCu
LCu
RCu
RCu
LAu
LAu
RAu
RAu
RHCwAuGquartz
CcarrierGcarrierI If, 0
| || | 1
2 213
432i2 −
⋅⋅
⋅
)Z+Z(Z)Z+Z(Z
=δ
77
P BP BTT
EM Day 2007 ACEM Day 2007 AC--DC current transfer standardsDC current transfer standards
Current transfer differences
0 ,1 1-5
0
5
1 0
1 5
2 0
2 5
3 0
δ i
f re q u e n c y
R H = 7 0 0 Ω R H = 3 5 0 Ω R H = 2 4 0 Ω R H = 1 1 7 Ω
M H z
μ A /A
88
P BP BTT
EM Day 2007 ACEM Day 2007 AC--DC current transfer standardsDC current transfer standards
Validation of the calculation
0,1 1-16
-14
-12
-10
-8
-6
-4
-2
0
2μA/A
MHz
117 Ω - 700 Ω measured calculated
Δδ i
frequency
99
P BP BTT
EM Day 2007 ACEM Day 2007 AC--DC current transfer standardsDC current transfer standards
Uncertainties using built-in Tee
0,1 1
0,2
0,3
0,4
0,5
0,6
0,7
u
frequency MHz
μA/A
1010
P BP BTT
EM Day 2007 ACEM Day 2007 AC--DC current transfer standardsDC current transfer standards
Measurement setup using driven guards
• Symmetrical setup• Upper standard connected „upside down“• Constant voltage across parasitic capacitances =>
equal transfer differences in “upper position” and “lower position”
1111
P BP BTT
EM Day 2007 ACEM Day 2007 AC--DC current transfer standardsDC current transfer standards
AC-DC current transfer standards
• Low currents: PMJTC with low heaterresistance
Problem: Technology limits 90 Ω < RH < 900 Ω
• Medium currents: Shunt + PMJTCProblems: Shunt inductance
Current dependence
• High currents: Cooled shunt + PMJTCProblems: Shunt inductance
Current dependenceCompliance voltage of current source
1212
P BP BTT
EM Day 2007 ACEM Day 2007 AC--DC current transfer standardsDC current transfer standards
Shunts used at PTB
Up to 300 mA: PTB design (top)
500 mA to 5 A: Holt design (middle)
10 A and 20 A: Fluke design (bottom)
1313
P BP BTT
EM Day 2007 ACEM Day 2007 AC--DC current transfer standardsDC current transfer standards
Shunts in Justervesenet design
Resistors
Current input
Output toTCC
+ Low cost materials
+ Large number of resistors ensure equal current distribution
+ Very small area of current path yields low inductive coupling
- No caseJustervesenet
1414
P BP BTT
EM Day 2007 ACEM Day 2007 AC--DC current transfer standardsDC current transfer standards
High current shunts
Set of four shunts:• 10 A to 30 A (25 mΩ)
• 30 A to 50 A (10 mΩ)
• 50 A to 80 A (5 mΩ)
• 80 A to 100 A (3 mΩ)
EL-9800
(NIST Design)
1515
P BP BTT
EM Day 2007 ACEM Day 2007 AC--DC current transfer standardsDC current transfer standards
New high current shunts
+ Precision resistors provide also dc accuracy
+ Large number of resistors ensure equal current distribution
+ Small area of current path yields low inductive coupling
- No case
- Quite expensive
Proven design scaled for higher currents
800 mV nominal output voltage
Available up to 100 A
1616
P BP BTT
EM Day 2007 ACEM Day 2007 AC--DC current transfer standardsDC current transfer standards
Step-up chains using different standards
5 A72 A6
5 A1 A53 A5 A500 mA41 A1 A200 mA3
300 mA500 mA100 mA2100 mA100 mA50 mA130 mA30 mA20 mAstart
CBAStep \ Chain
Chains A & B: Self-made and Holt shuntsChain C: Justervesenet-Shunts (two sets)Comparison of the chains: 1 A at 1 MHz: Δ < 10 µA/A
5 A at 100 kHz: Δ < 5 µA/A
1717
P BP BTT
EM Day 2007 ACEM Day 2007 AC--DC current transfer standardsDC current transfer standards
Measurement of the level dependence
Assumption:Current level dependence of shunts
results from self-heating
Measurement approach:
Shunt with low voltage drop used as a standard
• => Low power => low self-heating => low current level dependence
• Low voltage drop => voltage-amplifier necessary to operate PMJTC
1818
P BP BTT
EM Day 2007 ACEM Day 2007 AC--DC current transfer standardsDC current transfer standards
Measurement setup: Shunt with Amplifier
DCVoltage
ACVoltage
AC-DC Switch
x11
Amplifierwith PMJTCShunt
TransconductanceAmplifier
PMJTC
nV nV
Shunt
Hi
Low
Hi
Low
Device under test Standard
1919
P BP BTT
EM Day 2007 ACEM Day 2007 AC--DC current transfer standardsDC current transfer standards
Uncertainty of measurement
The following sources of uncertainty of the step-up procedure can be identified for each step:
u (δstd) Standard uncertainty of the (calculable) transfer difference of the standard used, i.e. uncertainty of previous step
u (δC) Standard uncertainty of the comparison procedure
u (δA) Type A standard uncertainty, i.e. the standard deviation of the mean
u (δLev) Standard uncertainty due to the current level effects in the shunt
u (δLF) Standard uncertainty due to PMJTC low frequency effects, which are level dependent
2020
P BP BTT
EM Day 2007 ACEM Day 2007 AC--DC current transfer standardsDC current transfer standards
Improvements in measurement uncertainties
Expanded uncertainty of measurement in µA/A at the frequencies Current 10 Hz 1 kHz 10 kHz 100 kHz 500 kHz 1 MHz
5 5 5 10 - - 30 mA 1 1 1 1 1 2 5 5 5 10 - - 100 mA 2 2 2 2 2 5 10 10 10 20 - - 300 mA 2 2 2 2 3 5 10 10 10 30 - - 1 A 4 3 3 4 6 10 20 20 20 60 3 A 4 4 4 8 25 25 25 80 10 A 12 12 12 25 - - - - 30 A 30 30 60 120 - - - - 100 A 40 40 80 150
Red : old (2002)Blue: (2006)
2121
P BP BTT
EM Day 2007 ACEM Day 2007 AC--DC current transfer standardsDC current transfer standards
Conclusions
• PMJTCs on quartz substrate developed at PTB are well suited for current transfer at low current up to 1 MHz
• Potential driven guarding reduces the uncertainty of the comparison measurements
• New shunts with low current level dependence and low transfer difference allow further reduction of the step-up procedure's uncertainty
• New approaches for the determination of the current level dependence allow for corrections and therefore for further reduction of the uncertainty