Design for patient safetyA guide to the design of infusion devices

Edition 12009

Page 2: A guide to the design of infusion devices · 2008-07-24 · Design for patient safety: infusion devices is the first guidance published that focuses on the safe design of infusion

DRAFTa

Contents

Foreword 4

Introduction 6

Advisory Panel 8

Background information 10

User Testing 12

1 Hardware design 15

2 Software design 51

3 Disposables 77

4 Specialist pumps 83

Information and training for users 90

References 92

Further information 94

About this publication

This booklet is one of a series of design publications produced by the National Patient Safety Agency (NPSA). Other publications in the Design for patient safety series:

NPSA in collaboration with the Royal College of Art Helen Hamlyn Centre:A guide to labelling and packaging of injectable medicines (2008)A guide to the graphic design of medication packaging (second edition) (2007)Future ambulances (2007)The design of infusion devices (in progress)

NPSA in collaboration with Lucid Design: A guide to the design of dispensed medicines (2007)A guide to the design of the dispensing environment (2007)

Research and Methodology

This publication is based on the results of a design research collaboration between the NPSA and the Royal College of Art Helen Hamlyn Centre (HHC), London.

The study was carried out over a one-year period by Sally Halls, a postgraduate specialist in medical design, working to a brief set out by the NPSA and the HHC. Existing design guidance was reviewed and consultations were undertaken with experts in graphic and information design, and design for patient safety. Technical support was provided throughout the project by David Cousins, Head of Safe Medication Practice, NPSA.

A wide range of stakeholders contributed to the research, including patients, healthcare professionals, NHS organisations, the Medicines and Healthcare products Regulatory Agency (MHRA) and industry personnel.

Observational research was undertaken in clinical environments, such as critical care areas, wards and pharmacies. The outcome was a design rationale to enhance patient safety and a fully illustrated set of design considerations with both good and bad examples.

DRAFT COPY Contents

Page 3: A guide to the design of infusion devices · 2008-07-24 · Design for patient safety: infusion devices is the first guidance published that focuses on the safe design of infusion

4 | A guide to the design of infusion devices

DRAFT

A guide to the design of infusion devices | 5

DRAFT

Foreword

ForewordForeword

To be added later.

Page 4: A guide to the design of infusion devices · 2008-07-24 · Design for patient safety: infusion devices is the first guidance published that focuses on the safe design of infusion

6 | A guide to the design of infusion devices

DRAFT

A guide to the design of infusion devices | 7

DRAFT

Introduction

Fifteen million infusions are performed in the NHS every year. The vast majority are delivered safely. However, at least 700 unsafe incidents are reported each year, of which 19 per cent are attributed to user error.1,2 In the 10 years between 1990 and 2000 there were 1495 incidents involving infusion pumps alone in the UK.3

Introduction Introduction

In 2004 the National Patient Safety Agency published a patient safety notice on infusion devices to reduce the risk of patient safety incidents involving infusion devices.1 The Safety Notice made recommendations concerning:

• how purchasing decisions on infusion pumps should be made in NHS organisations; • the importance of evaluating the necessity for an infusion device before it is purchased;• the need to reduce the range of infusion device types in use and, within each type, have agreed default configurations;• the benefits of establishing a centralised equipment library.

The NPSA published a toolkit to help NHS organisations review their existing device management systems, as well as assess the potential for significant cost benefits and improved patient safety and an evaluation report on this safety initiative piloted in six hospitals. 4 -6

Design for patient safety: infusion devices is the first guidance published that focuses on the safe design of infusion devices including infusion pumps and syringe drivers used in hospital and ambulatory care. This publication illustrates how design can be used to change and make safer the use of

Human beings usually make mistakes because the systems, tasks and products they work within are poorly designed. Effective design can deliver systems, products that are intuitive, simple to understand, simple to use and mistake proofed.

Mistake-proofing is the use of process design to facilitate correct actions, make wrong actions more difficult, make it easier to discover errors that occur, and make it possible to reverse or undo incorrect actions. Mistake-proofing tends to be inexpensive, very effective, and based on simplicity and ingenuity.8

Review of Infusion Device Technology and Evidence Based Purchasing

The Centre for Evidence-based Purchasing (CEP) is part of the Policy and Innovation Directorate of NHS PASA. CEP underpins purchasing decisions by providing objective evidence to support the uptake of useful, safe, innovative products and related procedures in health and social care.9

CEP Buyer’s guides include one or more of the following elements:• technology overview• market information• decision trees• comparative specifications• whole life costing• cost/benefit analysis• adoption guidance• sustainable procurement• reliability analysis.

CEP have published buyers guide reviews on some infusion devices and more are planned in the future:• Buyers guide, Portable pumps for home parenteral nutrition CEP 08021, May 2008.• Buyers guide, Insulin pumps, CEP 08004, February 2008. • Evaluation report, Hospira Plum A+ volumetric infusion pump, CEP 07014, October 2007

Planned Publications:• Buyers guide. Medication software for infusion pumps. Summer 2008• Buyers Guide. Portable syringe drivers. Summer 2008.The Medicines and Healthcare Product Regulatory Agency (MHRA) in collaboration with the Bath Institute of Medical Engineering publish evaluations of infusion devices.10

The MHRA also publishes medical device alerts concerning infusion devices.11

infusion devices in practice. It is intended as a best practice guide to be used by infusion device developers in medical devices companies as well as a reference guide for those involved in the procurement of medicines in the NHS.

The Department of Health, Design Council report published in 2003, Design for Patient Safety, acknowledged that the use of design in other safety critical industries had produced significant improvements in safety, quality and efficiency.7 The report recommended that a similar approach be taken within healthcare.

Page 5: A guide to the design of infusion devices · 2008-07-24 · Design for patient safety: infusion devices is the first guidance published that focuses on the safe design of infusion

8 | A guide to the design of infusion devices

DRAFT

A guide to the design of infusion devices | 9

DRAFT

Advisory Panel

Head of Clinical Engineering, Imperial College Healthcare NHS Trust

Head of Safe Medication Practice, National Patient Safety Agency

Centre Manager, BIME Evaluation Laboratory

MESU, Nottingham University Hospitals NHS Trust Clinical Engineering, Imperial College Healthcare NHS Trust

Clinical Engineering, Cardiff and Vale Hospital NHS Trust Medicine and Healthcare products Regulatory Agency Head of Design for Patient Safety, Helen Hamlyn Centre

Equipment Library Manager, Newcastle upon Tyne Hospitals

Clinical Engineering, Cardiff and Vale Hospital NHS Trust

ITU Nurse Educator, Imperial College Healthcare NHS Trust

Nick Abraham

Prof David Cousins

Craig Davey

Beth Griffiths

Mona Habib Roger Jones Jim Lefever Ed Matthews

Michelle Sanderson Marina Stock

Shakira Watts

Advisory panel Advisory panel

Page 6: A guide to the design of infusion devices · 2008-07-24 · Design for patient safety: infusion devices is the first guidance published that focuses on the safe design of infusion

10 | A guide to the design of infusion devices

DRAFT

A guide to the design of infusion devices | 11

DRAFT

Background information

To be added later.

Background information Background information

Page 7: A guide to the design of infusion devices · 2008-07-24 · Design for patient safety: infusion devices is the first guidance published that focuses on the safe design of infusion

12 | A guide to the design of infusion devices

DRAFT

A guide to the design of infusion devices | 13

DRAFT

User Testing

Stakeholders are invited to provide details of user testing for infusion devices.

User testing User testing

Page 8: A guide to the design of infusion devices · 2008-07-24 · Design for patient safety: infusion devices is the first guidance published that focuses on the safe design of infusion

Hardware design

1

DRAFT COPY

15

Page 9: A guide to the design of infusion devices · 2008-07-24 · Design for patient safety: infusion devices is the first guidance published that focuses on the safe design of infusion

16 | A guide to the design of infusion devices

DRAFT

A guide to the design of infusion devices | 17

DRAFT

1.1 Physical dimensions

Consideration is not given to the manoeuverability of the pumps, leading to bulky and heavy equipment.

The pump size/shape is not suitable for it’s application.

Pump are awkward to hold and handle.

Attaching devices to stands can be very cumbersome, particularly if there is more than one device already loaded onto the stands. Pole clamps may also be too small to fit around the stand.

Products should be developed with users, to ensure that they are of an appropriate size, shape and weight.

Different sizes and shapes may be more suited to different applications. For example, an ambulatory device needs to be small and portable, but may be too small for hospital use.

Pumps should have handles positioned to aid lifting and transport.

Design the device to be able to be attached to the dripstand as easily as possible. Consider the ergonomics and logistics of holding the device whilst operating attachment mechanisms.

Consider the addition of mini fluid stands to aid transportation.

Recommendations•

Issues•

10/200806K07BY

Do not reconnect partially used bags

Store out of reach of children

Do not use unless solution is clear and containerThere use unless solution is clear and cosilly

cant believe i ‘m hadding alth detalse unless solutisilteni gjtotke today is stily y is stily

se unless solutisilteni gjtotke se unless solutisilteni gjtotke today is stily

se unless solutisilteni gjtotke today is stilyy is stilyse unless soluti

se unless solutisilteni gjtotke

se unless soluti

today is stily y is stily

y is stily This is such a waste of

will anyone notice thissf hotnelstly

Healthcare Ltd Lektv

1

2

3

5

6

7

8

9

Th kicitsv lv

Wherehe

tell mel abt oei tidiff abt oei titgive me te ic hu nder

me the sonthem like a i

would tththe at make my head sorgte a i

Primary

rate

Secon-dary rate

Primary

rate

Secon-dary rate

Primary

rate

Secon-dary rate

Primary

rate

Secon-dary rate

Stop

Start

on /off

Primary

rate

Secon-dary rate

Primary

rate

Secon-dary rate

Primary

rate

Secon-dary rate

Primary

rate

Secon-dary rate

Stop

Start

10/200806K07BY

Do not reconnect partially used bags

Store out of reach of children

Do not use unless solution is clear and containerThere use unless solution is clear and cosilly

cant believe i ‘m hadding alth detalse unless solutisilteni gjtotke today is stily y is stily

se unless solutisilteni gjtotke se unless solutisilteni gjtotke today is stily

se unless solutisilteni gjtotke today is stilyy is stilyse unless soluti

se unless solutisilteni gjtotke

se unless soluti

today is stily y is stily

y is stily This is such a waste of

will anyone notice thissf hotnelstly

Healthcare Ltd Lektv

1

2

3

5

6

7

8

9

Th kicitsv lv

Wherehe

tell mel abt oei tidiff abt oei titgive me te ic hu nder

me the sonthem like a i

would tththe at make my head sorgte a i

Primary

rate

Secon-dary rate

Primary

rate

Secon-dary rate

Primary

rate

Secon-dary rate

Primary

rate

Secon-dary rate

Stop

Start

on /off

Primary

rate

Secon-dary rate

Primary

rate

Secon-dary rate

Primary

rate

Secon-dary rate

Primary

rate

Secon-dary rate

Stop

Start

Hardware design Hardware design

Page 10: A guide to the design of infusion devices · 2008-07-24 · Design for patient safety: infusion devices is the first guidance published that focuses on the safe design of infusion

18 | A guide to the design of infusion devices

DRAFT

A guide to the design of infusion devices | 19

DRAFT

1.2 Durability

Pumps are not designed for the rough treatment they receive in hospitals, where they are often dropped.

Pump surfaces are not sealed against fluid ingress.

Products should be tested for durability. A pump should be able to be dropped from bed height without significant damage occurring. Manufacturers should consider ways of alerting the user that the pump has previously been dropped and may malfunction.

Surfaces should be sealed such that they are able to be wiped with a wet cloth without undue harm occurring to the pump.

RATE

PRI

SEC

LIM

ITPR

ES

REA

D

PRES

CRI

S

TOT

VOL

INF

VOL

TO B

E IN

F

POW

ER

CLE

AR

SILE

NC

E

STA

RTST

OP

10/200806K07BY

Do not reconnect partially used bags

Store below 25CStore out of reach of childrenDiscard any unused portion

Do not use unless solution is clear and container

Do not use unless solution is clear and container There use unless solution is clear and cosilly

cant believe i ‘m hadding alth detal

se unless solutisilteni gjtotke today is stilyy is stily

se unless solutisilteni gjtotke

se unless solutisilteni gjtotke today is stily

se unless solutisilteni gjtotke today is stily y is stily se unless soluti

se unless solutisilteni gjtotke

se unless solutisilteni gjtotke

today is stilytoday is stilyy is stily

y is stilyThis is such a waste of

will anyone notice this

se unless solutisf hotnelstly

Healthcare Ltd Lektv

1

2

3

54

6

7

8

9

Th kicitsv lv

Wherehe

tell mel abt oei tidiff abt oei ti tgive me te ic hu nder

me the sonthem like a i

would tththe at make my head sorgte a i

RATE

PRI

SEC

LIM

ITPR

ES

REA

D

PRES

CRI

S

TOT

VOL

INF

VOL

TO B

E IN

F

POW

ER

CLE

AR

SILE

NC

E

STA

RTST

OP

10/200806K07BY

Do not reconnect partially used bags

Store below 25CStore out of reach of childrenDiscard any unused portion

Do not use unless solution is clear and container

Do not use unless solution is clear and container There use unless solution is clear and cosilly

cant believe i ‘m hadding alth detal

se unless solutisilteni gjtotke today is stilyy is stily

se unless solutisilteni gjtotke

se unless solutisilteni gjtotke today is stily

se unless solutisilteni gjtotke today is stily y is stily se unless soluti

se unless solutisilteni gjtotke

se unless solutisilteni gjtotke

today is stilytoday is stilyy is stily

y is stilyThis is such a waste of

will anyone notice this

se unless solutisf hotnelstly

Healthcare Ltd Lektv

1

2

3

54

6

7

8

9

Th kicitsv lv

Wherehe

tell mel abt oei tidiff abt oei ti tgive me te ic hu nder

me the sonthem like a i

would tththe at make my head sorgte a i

Hardware design Hardware design

Recommendations•

Issues•

Page 11: A guide to the design of infusion devices · 2008-07-24 · Design for patient safety: infusion devices is the first guidance published that focuses on the safe design of infusion

20 | A guide to the design of infusion devices

DRAFT

A guide to the design of infusion devices | 21

DRAFT

1.3 Pump stand

Pump stands are not designed to hold stands and can be unsteady and inclined to topple over.

Consideration is not given to the manoeuverability of the stands. and makes movement very awkward for patients.

Base units can make pump stands heavy and cumbersome.

Stands should consider the additional weight that pumps add to the stand, and be designed accordingly. They should also specify how many pumps can be safely mounted. Refer to the British Standards for guidelines on stability.12

Manufacturers should consider ways of aiding the pump attachment process. Docking stations may be a suitable option.

Consider the incorporation of power sockets to facilitate battery management.

Add handles to the stands to aid patient movement. Larger wheels can also help navigate over uneven floor surfaces.

Consideration should be given to the manoeuverability of stands, particularly with regard to the needs of mobile patients.

This stand This stand can hold can hold

up toup to

3 3 pumpspumps

This stand This stand can hold can hold

up toup to

3 3 pumpspumps

Hardware design Hardware design

Recommendations•

Issues•

Page 12: A guide to the design of infusion devices · 2008-07-24 · Design for patient safety: infusion devices is the first guidance published that focuses on the safe design of infusion

22 | A guide to the design of infusion devices

DRAFT

A guide to the design of infusion devices | 23

DRAFT

beep beep

beep beep

Epidural Epidural Epidural

1.4 Use of colour and labelling of pumps

Colour is used as an identifier without additional visual indicators.

Poor use of colour and design can increase the risk of miselection of infusion devices.

The overuse of colour creates overwhelming interfaces.

Colour application has not considered the needs of colour blind users.

If colour is used as an identifier, there should be an additional differentiator such as text, to emphasise this difference.

Use the colour yellow to denote spinal & epidural infusions. Text should be added to emphasise this difference.

Colour should be used judiciously to highlight key information.

Consider the requirements of colour blind users when applying colour.

Hardware design Hardware design

Recommendations•

Issues•

Page 13: A guide to the design of infusion devices · 2008-07-24 · Design for patient safety: infusion devices is the first guidance published that focuses on the safe design of infusion

24 | A guide to the design of infusion devices

DRAFT

A guide to the design of infusion devices | 25

DRAFT

2h 43mins remaining

• Infusing •

Sodium Chloride

250ml/h

Vol infused: 143mlLine pressure: Normal

Running on battery !!!

Press ok to confirm

1.5 Power

It can be unclear whether a pump is plugged into the main supply.

The level of battery charge is often ambiguous. This may lead to pumps cutting out mid-transfer, with the consequential delay to patient treatment.

Power cords may be inadvertantly pulled out when the pump in use.

Pumps should notify the user when they are not plugged into the mains supply. Depending upon the application of the pump, for example in ICU wards, it should require the user to confirm this state.

Pumps should give a good indication of battery life. This should be expressed in hours and minutes.

The battery indicator light should flash if the pump is running on the battery power.

Consider how to prevent power cords being pulled out. Take care to provide for the needs of the patient as well as the user.

2h 43mins remaining

• Infusing •

Sodium Chloride

250ml/h

Vol infused: 143mlLine pressure: Normal

Running on battery !!!

Press ok to confirm

Hardware design Hardware design

Recommendations•

Issues•

Page 14: A guide to the design of infusion devices · 2008-07-24 · Design for patient safety: infusion devices is the first guidance published that focuses on the safe design of infusion

26 | A guide to the design of infusion devices

DRAFT

A guide to the design of infusion devices | 27

DRAFT

1.6a Loading and priming infusion pumps

It can be unclear if an administration set has been correctly loaded into the pump.

Lines may not be primed before beginning the infusion.

Lineloading can be an awkward, complex process with very little signposting to aid the user. Horizontal feeds can be particularly confusing, as the line may go from right to left, or left to right.

In-line cassettes can complicate lineloading processes.

Fitting lines into multi-channel pumps may cause confusion, particularly as to where each line goes.

Fluid may leak from the bags into the pump, causing the pump to fail.

It can be ambiguous as to which way horizontally loading administrations sets should be loaded into the pump.

Loading mechanisms should be designed such that incorrect installation is extremely difficult. It should be immediately apparent when a line has been misloaded, with a corresponding alarm message on screen.

Clear instructions should be visible to the user for the lineloading process. The vertical loading of administration sets is more instinctive than horizontal loading. There should always be an indication as to the direction of the fluid flow to prevent confusion.

Cassettes should be easy to install in the pump. Good design should indicate the orientation of the cassette, and aid its installation.

Multi-channel pumps should clearly indicate which channel operates which medicine. Lineloading processes should be as simple that of single channel pumps.

Pumps should be sealed to prevent fluid contamination and aid cleaning.

Pumps should prompt the user to prime the line and provide functionality to aid this process.

10/200806K07BY

Do not reconnect partially used bags

Store below 25CStore out of reach of childrenDiscard any unused portion

Do not use unless solution is clear and container

Do not use unless solution is clear and containerThere use unless solution is clear and cosilly

cant believe i ‘m hadding alth detalse unless solutisilteni gjtotke today is stily y is stily

se unless solutisilteni gjtotke se unless solutisilteni gjtotke today is stily

se unless solutisilteni gjtotke today is stilyy is stilyse unless soluti

se unless solutisilteni gjtotke

se unless solutisilteni gjtotke

today is stily today is stilyy is stily se unless soluti

y is stily This is such a waste of

will anyone notice this

se unless solutisf hotnelstly

Healthcare LtdLektv

1

2

3

54

6

7

8

9

Th kicitsv lv

Wherehe

tell mel abt oei tidiff abt oei titgive me te ic hu nder

me the sonthem like a i

would tththe at make my head sorgte a i

Infusing

Press PrimaryOptions

Rate:

AREA 1

mL/hr

Volume remaining 500ml

Time remaining

2002:30 hr:min

Primary

Infusing

Press PrimaryOptions

Rate:

AREA 1

mL/hr

Volume remaining 500ml

Time remaining

2002:30 hr:min

Primary

10/200806K07BY

Do not reconnect partially used bags

Store below 25CStore out of reach of childrenDiscard any unused portion

Do not use unless solution is clear and container

Do not use unless solution is clear and containerThere use unless solution is clear and cosilly

cant believe i ‘m hadding alth detalse unless solutisilteni gjtotke today is stily y is stily

se unless solutisilteni gjtotke se unless solutisilteni gjtotke today is stily

se unless solutisilteni gjtotke today is stilyy is stilyse unless soluti

se unless solutisilteni gjtotke

se unless solutisilteni gjtotke

today is stily today is stilyy is stily se unless soluti

y is stily This is such a waste of

will anyone notice this

se unless solutisf hotnelstly

Healthcare LtdLektv

1

2

3

54

6

7

8

9

Th kicitsv lv

Wherehe

tell mel abt oei tidiff abt oei titgive me te ic hu nder

me the sonthem like a i

would tththe at make my head sorgte a i

Infusing

Press PrimaryOptions

Rate:

AREA 1

mL/hr

Volume remaining 500ml

Time remaining

2002:30 hr:min

Primary

Infusing

Press PrimaryOptions

Rate:

AREA 1

mL/hr

Volume remaining 500ml

Time remaining

2002:30 hr:min

Primary

Hardware design Hardware design

Recommendations•

Issues•

Page 15: A guide to the design of infusion devices · 2008-07-24 · Design for patient safety: infusion devices is the first guidance published that focuses on the safe design of infusion

28 | A guide to the design of infusion devices

DRAFT

A guide to the design of infusion devices | 29

DRAFT

1.6b Loading and priming syringe pumps

Fitting the syringe into the pump can be a complex and unwieldy process.

The syringe and plunger may not be secured into the pump, potentially allowing the entire contents of the container to freeflow into the patient.

The syringe driver may not be calibrated for the make and size of syringe used.

The syringe and giving set may not be adequately purged, with the consequential mechanical backlash delaying treatment.

The syringe loading process should be simple and able to be performed by users with limited dexterity. Simultaneous actions should be avoided where possible.

The pump should be designed to ensure that the syringe and plunger cannot disengage. It should also have corresponding alarms to alert the user to potential free flow.

The driver should allow for easy calibration of the make and size of syringe. It should also be simple to change the settings if the auto-sensing detects the wronge syringe.

The pump should prompt the user to prime the line and reduce any mechanical backlash.

Hardware design Hardware design

Recommendations•

Issues•

Page 16: A guide to the design of infusion devices · 2008-07-24 · Design for patient safety: infusion devices is the first guidance published that focuses on the safe design of infusion

30 | A guide to the design of infusion devices

DRAFT

A guide to the design of infusion devices | 31

DRAFT

1.7a Device controls - numeric keypads

1 2 3

4 5 6

7 8 9

* 0 #

Memory

Redial

1571

1 2 3 4 5 6 7 8 9 0 -_

=+

Q W E R 7 8 9 -

= / *

4 5 6 +

1 2 3Enter

Numlock

T Y U I O P [ ] 1

A S D F G H J K L:

;“‘

|

\

~

`Z X C V B N M >

.<,

?/

|

\

0 .|

\Alt GrAlt Ctrl

Caps Lock

±§

1 2 3

4 5 6

7 8 9

. 0 clear

1 2 3

4 5 6

7 8 9

. 0 clear

Stopped

Press PrimaryOptions

Rate:

AREA 1

mL/hr

Volume remaining 500ml

Time remaining

2002:30 hr:min

Primary

1 2 3

4 5 6

7 8 9

. 0 clear

+-mrmcm-m+

AC

x

1/x

=.

x

+

-

987

654

2 31

0

2

Infusion pumps use different layouts for numerical keypads, causing confusion amonst users.

The keypad layout may be rearranged, placing the ‘0’ and ‘ . ’ in unexpected places.

Tests found that staff are more familiar with the telephone layout, where the number ‘1’ is in the top left corner. This layout should be used on all devices.

The numerical layout should not be altered. The ‘0’ and ‘ . ’ should always be positioned below the rest of the numbers.

1 2 3

4 5 6

7 8 9

* 0 #

Memory

Redial

1571

1 2 3 4 5 6 7 8 9 0 -_

=+

Q W E R 7 8 9 -

= / *

4 5 6 +

1 2 3Enter

Numlock

T Y U I O P [ ] 1

A S D F G H J K L:

;“‘

|

\

~

`Z X C V B N M >

.<,

?/

|

\

0 .|

\Alt GrAlt Ctrl

Caps Lock

±§

1 2 3

4 5 6

7 8 9

. 0 clear

1 2 3

4 5 6

7 8 9

. 0 clear

Stopped

Press PrimaryOptions

Rate:

AREA 1

mL/hr

Volume remaining 500ml

Time remaining

2002:30 hr:min

Primary

1 2 3

4 5 6

7 8 9

. 0 clear

+-mrmcm-m+

AC

x

1/x

=.

x

+

-

987

654

2 31

0

2

Hardware design Hardware design

Recommendations•

Issues•

Page 17: A guide to the design of infusion devices · 2008-07-24 · Design for patient safety: infusion devices is the first guidance published that focuses on the safe design of infusion

32 | A guide to the design of infusion devices

DRAFT

A guide to the design of infusion devices | 33

DRAFT

1.7b Device controls - other buttons

It can be unclear what function soft keys perform or what figures they correspond to.

It can be ambiguous which figures arrow keys refer to.

The start and stop functions may be provided by one button, which may cause ambiguity as to which operation is being invoked.

Numerous keypresses may be required in order to reach a particular desired function.

It should be clear what action controllers perform.

Space controls intelligently and ensure that there is sufficient space between them.

Start and stop should be two distinct buttons. Key buttons such as the bolus, start, stop and on/off buttons should be positioned away from the main group of buttons.

Consider grouping buttons of similar functions together.

Minimise the number of key presses required to perform any function.

OK

P

OPT EXIT

MODE

OPTSTOP OK

RATEPRI SEC

LIMITPRES

READ PRESCRIS

TOTVOLINF

VOLTO BE

INF

POWER

CLEAR

SILENCE

START

STOP

7 8 9

4 5 6

1 2 3

0 .BACKPOWER

Bolus Start Stop

MAIN MENU CHG RATE OPTIONS CLEAR TOTALS

INJECTOSOL 60ML TVD 0.31ML

PVD 0.81ML

RATE 25ML/HR

Primary rate

Secon-dary rate

Pressure limit

Read pressure

VolumeInfused

VolumeTo be

Infused

Clear

StopStart

1 2 3

4 5 6

7 8 9

0 .Power

Bolus Start Stop

MAIN MENU CHG RATE OPTIONS CLEAR TOTALS

Infusing TVD 0.31ml

PVD 0.81ml

Rate 25ml/hr

Back

OK

P

OPT EXIT

MODE

OPTSTOP OK

RATEPRI SEC

LIMITPRES

READ PRESCRIS

TOTVOLINF

VOLTO BE

INF

POWER

CLEAR

SILENCE

START

STOP

7 8 9

4 5 6

1 2 3

0 .BACKPOWER

Bolus Start Stop

MAIN MENU CHG RATE OPTIONS CLEAR TOTALS

INJECTOSOL 60ML TVD 0.31ML

PVD 0.81ML

RATE 25ML/HR

Primary rate

Secon-dary rate

Pressure limit

Read pressure

VolumeInfused

VolumeTo be

Infused

Clear

StopStart

1 2 3

4 5 6

7 8 9

0 .Power

Bolus Start Stop

MAIN MENU CHG RATE OPTIONS CLEAR TOTALS

Infusing TVD 0.31ml

PVD 0.81ml

Rate 25ml/hr

Back

Hardware design Hardware design

Recommendations•

Issues•

Page 18: A guide to the design of infusion devices · 2008-07-24 · Design for patient safety: infusion devices is the first guidance published that focuses on the safe design of infusion

34 | A guide to the design of infusion devices

DRAFT

A guide to the design of infusion devices | 35

DRAFT

Manufacturers are increasingly replacing text with symbols on their device controls. This enables one product to be marketed to many countries and overcomes many of the different language requirements.

Issue•

1.8a Symbols - ISO standards

The use of symbols introduces new risks as symbols may be poorly understood by users.13 Some symbols have been included in ISO 60878. However, the meaning of these symbols is not always intuitive or well known by heathcare users.

It is recommended that manufacturers add text labels to symbols until the meaning of the symbols are universally recognised by users.

OK

P

OPT EXIT

MODE

OK OPT EXITMODE

P

OPTSTOP OK

Slow run: slow speed

Caution

Contrast

Brightness

Printer

3

3N

AP APG

“On” (power)

“Off” (power)

“On/Off” (push-push)

“On/Off” (push button)

Stand-by

Stand-by or preparatory state for a part of equipment

“Off” for a part of equipment

“On” for a part of equipment

Start (of action)

Stop (of action)

Pause

Handheld switch

Manual control

Variability

Variability in steps

Movement in one direction

Movement in both directions

Battery check

Power plug

Lamp; lighting; illumination

Bell

Bell cancel

Loudspeaker

Sound muting temporary

Eject

Patient

Nurse

Injection syringe

Infusion status

Locking (locked status)

Unlocking

Infusion bottle

Trend information

OffOn Start Stop

Handheldswitch

Manual control

Variability Variability in steps

Power plugBattery check Bell cancel Sound muting

NursePatient Lighting Do not reuse

Sound muting

Injection syringe

Infusion status

Infusion bottle

Trend information

ISO symbols

Symbols in use

History

mmHg

PAUSESTOP

POWER ON

RATE START

CLRCLEAR

SILENCEP/C

MODESEC

INFUSIONVOLUME INFUSED

PRISEC/

CLEARVOL

READVOL

VOLTO BEINF

VOLUME TO BE

INFUSED

TOTVOLINF

VOLTO BE

INFPRI SEC CRIS

LIMITPRES

READPRES

EXITCLEAR

RUNHOLD

ENTERSET

BOLUSPURGE

STARTSTOP

Main Display

Volume History

start stop

22

Injection syringe

Infusion status

Infusion bottle

Trend information

OffOn Start Stop

Handheldswitch

Manual control

Variability Variability in steps

Power plugBattery check Bell cancel Sound muting

NursePatient Lighting Do not reuse

2

OK

P

OPT EXIT

MODE

OK OPT EXITMODE

P

OPTSTOP OK

Slow run: slow speed

Caution

Contrast

Brightness

Printer

3

3N

AP APG

“On” (power)

“Off” (power)

“On/Off” (push-push)

“On/Off” (push button)

Stand-by

Stand-by or preparatory state for a part of equipment

“Off” for a part of equipment

“On” for a part of equipment

Start (of action)

Stop (of action)

Pause

Handheld switch

Manual control

Variability

Variability in steps

Movement in one direction

Movement in both directions

Battery check

Power plug

Lamp; lighting; illumination

Bell

Bell cancel

Loudspeaker

Sound muting temporary

Eject

Patient

Nurse

Injection syringe

Infusion status

Locking (locked status)

Unlocking

Infusion bottle

Trend information

OffOn Start Stop

Handheldswitch

Manual control

Variability Variability in steps

Power plugBattery check Bell cancel Sound muting

NursePatient Lighting Do not reuse

Sound muting

Injection syringe

Infusion status

Infusion bottle

Trend information

ISO symbols

Symbols in use

History

mmHg

PAUSESTOP

POWER ON

RATE START

CLRCLEAR

SILENCEP/C

MODESEC

INFUSIONVOLUME INFUSED

PRISEC/

CLEARVOL

READVOL

VOLTO BEINF

VOLUME TO BE

INFUSED

TOTVOLINF

VOLTO BE

INFPRI SEC CRIS

LIMITPRES

READPRES

EXITCLEAR

RUNHOLD

ENTERSET

BOLUSPURGE

STARTSTOP

Main Display

Volume History

start stop

22

Injection syringe

Infusion status

Infusion bottle

Trend information

OffOn Start Stop

Handheldswitch

Manual control

Variability Variability in steps

Power plugBattery check Bell cancel Sound muting

NursePatient Lighting Do not reuse

2

Hardware design Hardware design

Recommendation•

Page 19: A guide to the design of infusion devices · 2008-07-24 · Design for patient safety: infusion devices is the first guidance published that focuses on the safe design of infusion

36 | A guide to the design of infusion devices

DRAFT

A guide to the design of infusion devices | 37

DRAFT

1.8b Symbols - manufacturer variations

Manufacturers devise their own symbols for device controls in addition to using ISO standards symbols. Non standard symbols may be misunderstood by users.

Symbols need to be tested with users in context. There is an ISO standard for developing and testing symbols with users.14,15

It is recommended that manufacturers add text labels to symbols until the meaning of the symbols are universally recognised by users.

OK

P

OPT EXIT

MODE

OK OPT EXITMODE

P

P

OPTSTOP OK

Injection syringe

Infusion status

Infusion bottle Trend information

OffOn Start Stop

Handheldswitch

Manual control

Variability Variability in steps

Power plugBattery check Bell cancelSound muting

NursePatient Lighting Do not reuse

Symbols in use

History

History

mmHg

mmHg

PAUSESTOP

POWER ON

RATE START

CLRCLEAR

SILENCEP/C

MODESEC

INFUSIONVOLUME INFUSED

PRISEC/

CLEARVOL

READVOL

VOLTO BEINF

VOLUME TO BE

INFUSED

TOTVOLINF

VOLTO BE

INFPRI SEC CRIS

LIMITPRES

READPRES

EXITCLEAR

RUNHOLD

ENTERSET

BOLUSPURGE

STARTSTOP

Main Display

Main Display

Volume History

Volume History

Use green as an additional indicator for the start button

The on/off button should be the only button with 2 functions. In line with MDA requirements, turning off the pump should be a 2 step process requiring confirmation from the user.

Use red as an additional indicator for the stop button

Use 3 arrows to indicate the bolus function and avoid confusion with rewind / fast forward icons.

Keep icons simple and minimalist. Remove any unnecessary graphics where possible.

Where abstract icons are used, use labels to help identify the button.

22

start

stop

OK

P

OPT EXIT

MODE

OK OPT EXITMODE

P

P

OPTSTOP OK

Injection syringe

Infusion status

Infusion bottle Trend information

OffOn Start Stop

Handheldswitch

Manual control

Variability Variability in steps

Power plugBattery check Bell cancelSound muting

NursePatient Lighting Do not reuse

Symbols in use

History

History

mmHg

mmHg

PAUSESTOP

POWER ON

RATE START

CLRCLEAR

SILENCEP/C

MODESEC

INFUSIONVOLUME INFUSED

PRISEC/

CLEARVOL

READVOL

VOLTO BEINF

VOLUME TO BE

INFUSED

TOTVOLINF

VOLTO BE

INFPRI SEC CRIS

LIMITPRES

READPRES

EXITCLEAR

RUNHOLD

ENTERSET

BOLUSPURGE

STARTSTOP

Main Display

Main Display

Volume History

Volume History

Use green as an additional indicator for the start button

The on/off button should be the only button with 2 functions. In line with MDA requirements, turning off the pump should be a 2 step process requiring confirmation from the user.

Use red as an additional indicator for the stop button

Use 3 arrows to indicate the bolus function and avoid confusion with rewind / fast forward icons.

Keep icons simple and minimalist. Remove any unnecessary graphics where possible.

Where abstract icons are used, use labels to help identify the button.

22

start

stop

Hardware design Hardware design

Recommendations•

Issue•

Page 20: A guide to the design of infusion devices · 2008-07-24 · Design for patient safety: infusion devices is the first guidance published that focuses on the safe design of infusion

38 | A guide to the design of infusion devices

DRAFT

A guide to the design of infusion devices | 39

DRAFT

1.9 Display screen

Screens are too small to be able to display the key information properly.

Screens cannot be seen when interacting with pumps due to the screen angle.

Light reflections can impede legibility of information.

Screens cannot be read in dark lighting conditions.

Screens should be large enough to display the key information without causing user confusion.

Consideration should be given to the visibility of screens when pumps are stacked. Tilt the screen towards the user where possible.

Use non-reflective finishes to reduce reflections.

Screens should have variable backlighting settings to maximise usability and patient comfort.

Consider the use of screens with a black background and white numerics. These can be more visible in all lighting conditions than other types of screens.

c

history menuInfusing... Adrenaline 1mg/ml Infusing... Adre

• Infusing •

Sodium Chloride

250ml/h

Vol infused: 143mlLine pressure: Normal

3.40

c

history menuInfusing... Adrenaline 1mg/ml Infusing... Adre

• Infusing •

Sodium Chloride

250ml/h

Vol infused: 143mlLine pressure: Normal

3.40

Hardware design Hardware design

Recommendations•

Issues•

Page 21: A guide to the design of infusion devices · 2008-07-24 · Design for patient safety: infusion devices is the first guidance published that focuses on the safe design of infusion

40 | A guide to the design of infusion devices

DRAFT

A guide to the design of infusion devices | 41

DRAFT

1.10 Touchscreens

Touchscreens are usually operated with a finger. Touchscreen controls may be too small to be easily selected and operated using a finger.

Keying in many numbers or letters by finger pointing is tiring.

Buttons are touched to provide information, provide options and initiate action. Unintended action may be initiated whilst reviewing information and selection options.

Points of interaction should be clearly identifiable without the need to clarify these by activation.

Points of interaction should be large enough to be pushed by a finger. The minimum size of vital buttons should be 2cm wide with 1mm spacing in between. Buttons should not be less than 12mm wide.16,17

Create invisible active areas around buttons in case the finger is not correctly positioned. Leave at least 1mm of inactive space between buttons.

Group buttons of similar function together to streamline hand movment.

Action buttons should be positioned at the bottom of the screen so that consequences are not obscured by the hand or arm.

Some touchscreen technologies are less reliable at screen edges. Position buttons in from the edge or make them larger to compensate for this.

The software should be error tolerant and allow the user to backtrack easily. Using a lift-off tapping strategy instead of a land-on strategy allows for greater accuracy.

Use a list selection option where possible to avoid many button presses and user fatigue.

Create tactile feedback for the user. Non-audible methods are preferred.

Downstream occlusion values

2

4

6

5

8

11

9

12

15

<32

Flowrate ml/h

32-150

>150

32-150 <150

<32 32-150 <150

<32 32-150 <150

Rate range / mL/hr

home options

Alarm sensitivity

Low

Longer to alarm

Quicker to alarm

High74

confirmcancel

home options

Alarm sensitivity

Low

High

cancel confirm

Poor examples of

Size

Grid

Grouping

Feedback and

state of button

exit

ok

mmHg

Minimum Moderate Maximum

Options

more details

more details

Downstream occlusion values

2

4

6

5

8

11

9

12

15

<32

Flowrate ml/h

32-150

>150

32-150 <150

<32 32-150 <150

<32 32-150 <150

Rate range / mL/hr

home options

Alarm sensitivity

Low

Longer to alarm

Quicker to alarm

High74

confirmcancel

home options

Alarm sensitivity

Low

High

cancel confirm

Poor examples of

Size

Grid

Grouping

Feedback and

state of button

exit

ok

mmHg

Minimum Moderate Maximum

Options

more details

more details

Hardware design Hardware design

Recommendations•

Issues•

Page 22: A guide to the design of infusion devices · 2008-07-24 · Design for patient safety: infusion devices is the first guidance published that focuses on the safe design of infusion

42 | A guide to the design of infusion devices

DRAFT

A guide to the design of infusion devices | 43

DRAFT

1.11 Alarms

It can be hard to identify which piece of equipment is alarming. Particularly in intensive care wards, where equipment may be stacked up.

Patients in isolated rooms with alarming pumps may not be noticed by nursing staff.

Abbreviated warning messages lead to confusion in how to rectify the situation.

Alarms don’t indicate the urgency of the situation.

It is possible to silence the alarm situation by pressing keys other than the alarm silence key.

beep

beep

beep

beep

beep

beep

bzzz

bzzzzz

bzzz

Check

Mr P Smith’s

Morphine

Alarms should be clearly visible as well as audible.

Consider the possibility of alarms signalling to pagers or radio signals. This would allow for a quieter alarm at the patient’s bedside, and a louder alarm with the user.

Alarm messages should be in clear simple English (or the language of the country).

Differentiate between an alert and an alarm, through the use of different colours and audio rhythms.18

When drug libraries are in use, consideration should be given to the different toxicity levels of drugs being dispensed so that a saline infusion receives a lower alarm status than a cytotoxic.

Users should only be able to silence the alarm by pressing the alarm silence key and/or rectifying the cause of the alarm.

Hardware design Hardware design

Recommendations•

Issues•

Page 23: A guide to the design of infusion devices · 2008-07-24 · Design for patient safety: infusion devices is the first guidance published that focuses on the safe design of infusion

44 | A guide to the design of infusion devices

DRAFT

A guide to the design of infusion devices | 45

DRAFT

1.12 Cleaning

Pumps are cleaned daily, but poorly designed pumps take longer to clean.

Moisture may get into the pumps and contaminate the working mechanics.

Sensors become dirty but poor device design prevents them from being cleaned.

Incidence of super viruses are increasing in hospitals, leading to a higher risk of cross infection between patients.

Pumps may have to be cleaned whilst in operation, but controls may be knocked accidentally.

Pumps should be designed to have smooth surfaces, with filled corners and no crevices. IT connections should be recessed into the pump. Where grooves are necessary (for example with heat sinks) make these large enough to be easily cleaned.

Pump surfaces should be impermeable to moisture and be fitted with watertight seals.

Sensors should be designed to enable access for cleaning purposes.

Consider the use of bacteria-resistant coatings.

Incorporating keypad locks into devices can help prevent accidental knocks from occurring.

Hardware design Hardware design

Recommendations•

Issues•

Page 24: A guide to the design of infusion devices · 2008-07-24 · Design for patient safety: infusion devices is the first guidance published that focuses on the safe design of infusion

46 | A guide to the design of infusion devices

DRAFT

A guide to the design of infusion devices | 47

DRAFT

1.13 Maintenance

Maintenance prompts are provided by stickers, which get worn away and lost in the visual clutter of the pump.

Items that require regular access such as batteries, cannot be easily accessed by maintenance staff.

When pumps are due for maintenance they should prompt the user with an onscreen warning. Pumps should not function when they require servicing.

Pump designs should facilitate simple maintenance procedures such as changing batteries and cleaning sensors. Simple designs create simple maintenance procedures, and minimising part numbers can also make repairing easier.

On screen maintenance message.

USE ONLY WITH APPROVED ADMINISTRATION SETS

RENLY WITH STPROVED RESETSDELET LTINE 20W

NOWMACX THIS IS A

VOLUMETRIC TN PUMP FOR USE

MAXI FUNTIONTSANT EIGGNE

157411 OETOWM

220/240 VAC ~T 0.2A50/60 HZ

SERVICE Jun 2011

CLI

NICAL ENGINEERING

QUALITY CONTROL

SERVICE DATE

SERVICE

More text written here to make it

Broomfield NHS trust

ELECTRICAL TEST

RETEST DUE

CONTACT CLINICAL ENGINEERING

Equipment of this typecomplies withProduct NoCertificate No

g2857

1 2 3

4 5 6

7 8 9

. 0 clear

Warning!

Do not use.This pump is now due for

maintenance (Oct 08).

Start 1 2 3

4 5 6

7 8 9

. 0 CStop

Menu

History

Bolus Infusace

Warning!Do not use.

This pump is now due for maintenance (Oct 08).

On screen maintenance message.

USE ONLY WITH APPROVED ADMINISTRATION SETS

RENLY WITH STPROVED RESETSDELET LTINE 20W

NOWMACX THIS IS A

VOLUMETRIC TN PUMP FOR USE

MAXI FUNTIONTSANT EIGGNE

157411 OETOWM

220/240 VAC ~T 0.2A50/60 HZ

SERVICE Jun 2011

CLI

NICAL ENGINEERING

QUALITY CONTROL

SERVICE DATE

SERVICE

More text written here to make it

Broomfield NHS trust

ELECTRICAL TEST

RETEST DUE

CONTACT CLINICAL ENGINEERING

Equipment of this typecomplies withProduct NoCertificate No

g2857

1 2 3

4 5 6

7 8 9

. 0 clear

Warning!

Do not use.This pump is now due for

maintenance (Oct 08).

Start 1 2 3

4 5 6

7 8 9

. 0 CStop

Menu

History

Bolus Infusace

Warning!Do not use.

This pump is now due for maintenance (Oct 08).

Hardware design Hardware design

Recommendations•

Issues•

Page 25: A guide to the design of infusion devices · 2008-07-24 · Design for patient safety: infusion devices is the first guidance published that focuses on the safe design of infusion

48 | A guide to the design of infusion devices

DRAFT

A guide to the design of infusion devices | 49

DRAFT

1.14 Pump storage

Pump storage is disorganised and chaotic, leading to device mismanagement.

Pumps are left unplugged when not in use, leading to flat batteries.

Hospital trusts should have central storage units implemented, where pumps are sent to be serviced, maintained and stored.

Consider ways of ensuring devices are plugged in whilst in storage. Potential solutions may be alarms of docking stations.

Consider ways of tracking devices throughout a trust. For example the incorporation of RFID technology can help to easily locate a device, and even prompt the pump to change its personality settings with location.

RATEPRI SEC

LIMITPRES

READ PRESCRIS

TOTVOLINF

VOLTO BE

INF

POWER

CLEAR

SILENCE

START

STOP

RATEPRI SEC

LIMITPRES

READ PRESCRIS

TOTVOLINF

VOLTO BE

INF

POWER

CLEAR

SILENCE

START

STOP

RATEPRI SEC

LIMITPRES

READ PRESCRIS

TOTVOLINF

VOLTO BE

INF

POWER

CLEAR

SILENCE

START

STOP

RATEPRI SEC

LIMITPRES

READ PRESCRIS

TOTVOLINF

VOLTO BE

INF

POWER

CLEAR

SILENCE

START

STOP

7 8 9

4 5 6

1 2 3

0 .BACKPOWER

Bolus Start Stop

7 8 9

4 5 6

1 2 3

0 .BACKPOWER

Bolus Start Stop

7 8 9

4 5 6

1 2 3

0 .BACKPOWER

Bolus Start Stop

Hardware design Hardware design

Recommendations•

Issues•

Page 26: A guide to the design of infusion devices · 2008-07-24 · Design for patient safety: infusion devices is the first guidance published that focuses on the safe design of infusion

DRAFT a

Software design

2

DRAFT COPY

A guide to the design of infusion devices | 51

Page 27: A guide to the design of infusion devices · 2008-07-24 · Design for patient safety: infusion devices is the first guidance published that focuses on the safe design of infusion

52 | A guide to the design of infusion devices

DRAFT

A guide to the design of infusion devices | 53

DRAFT

2.1 Layout - essential information

Users cannot readily find the information they require as emphasis is given to less important data.

It can be difficult to ascertain what a pump is doing.

Screen layouts are cluttered with information clustered around the edges of screens.

Channel stopped

Press Primary

Options

AREA 1

Rate 200 mL/hrVolume to be infused 500 mlVolume Remaining 500 mlTime Remaining 2:30 hr:min

Primary

Drug name 1mg/ml

StoppedOptions

Rate:

AREA 1

mL/hr

Volume to be infused 500mlVolume infused 0ml

Time remaining

2002:30 hr:min

Primary

Highlight the information most readily required by the users of the pump. They need to be able to see: – Drug name and concentration (where applicable) – Flowrate – Device status (infusing/stopped/standby) – Time remaining/Volume to be infused – Volume infused / Total volume infused – Mode of delivery - continuous, PCA, KVO – Device status (Infusing, stopped, standby) – Power supply (mains/battery, battery life) – Occlusion pressure levels

There should be a clear visual indicator of whether a pump is on and of what it is doing.

Key information should be positioned in the centre of the screen.

Channel stopped

Press Primary

Options

AREA 1

Rate 200 mL/hrVolume to be infused 500 mlVolume Remaining 500 mlTime Remaining 2:30 hr:min

Primary

Drug name 1mg/ml

StoppedOptions

Rate:

AREA 1

mL/hr

Volume to be infused 500mlVolume infused 0ml

Time remaining

2002:30 hr:min

Primary

Software design Software design

Recommendations•

Issues•

Page 28: A guide to the design of infusion devices · 2008-07-24 · Design for patient safety: infusion devices is the first guidance published that focuses on the safe design of infusion

54 | A guide to the design of infusion devices

DRAFT

A guide to the design of infusion devices | 55

DRAFT

2.2 Fonts and text

Text is written in small font sizes that are hard to read. Text can be hard to read when written in capital letters.

Similar words/drugnames may be confused with each other.

Words may be abbreviated into unintelligible words.

Terms are used that may be misunderstood or confused e.g. ‘proximal and distal’.

1 REVIEW PROGRAM2 HISTORIES3 KEYPAD LOCK4 PUMP SETTINGS

Review programHistoriesKeypad lockPump settings

DOPamine 4mg/ml

StoppedOptions

Rate:

AREA 1

mL/hr

Volume infused 0mlVolume to be infused 100ml

Time remaining

20.52:30 hr:min

Primary

Use clear legible anti-aliased fonts, in large fontsizes. Numbers after a decimal point should be in a smaller font size.

Use sentence case.

Use tallman lettering to distinguish between drugs with similar names.

Use clear, simple English that can be easily understood.

Avoid using terms such as proximal and distal that may not be understood. Terms like downstream and upstream are better.

Software design Software design

Recommendations•

Issues•

NBDistortion occurs when high resolution graphics are displayed on a low resolution screen. Anti-aliasing is the technique of minimising this distortion and thereby maintaining legibility.

1 REVIEW PROGRAM2 HISTORIES3 KEYPAD LOCK4 PUMP SETTINGS

Review programHistoriesKeypad lockPump settings

DOPamine 4mg/ml

StoppedOptions

Rate:

AREA 1

mL/hr

Volume infused 0mlVolume to be infused 100ml

Time remaining

20.52:30 hr:min

Primary

Page 29: A guide to the design of infusion devices · 2008-07-24 · Design for patient safety: infusion devices is the first guidance published that focuses on the safe design of infusion

56 | A guide to the design of infusion devices

DRAFT

A guide to the design of infusion devices | 57

DRAFT

Pump configuration or personalties enable default rate, volume, air and occlusion detection, drug settings and displays to be set for different patient types and clinical areas of use. Pumps may also be configured for continuous, patient control, intermittent and background infusion.

From time to time device manufacturers release new versions of device software to correct software faults and provide new functionality. These settings assist the safe and effective use of these devices for the application for which they are intended.

Issues•

2.3 Pump configuration and personalities

It is often not clearly displayed which configuration, personality or software release is operating in the device.

The default settings of each personality are not always displayed or understood by users.

Infusion devices may move between clinical areas and be used for different patient groups. The configuration, personality and software version that helps to ensure safe use in one clinical area and group of patients, may increase risk in another area with different patients. Pumps are often moved to different patients and areas that require other personalities.

Software versions, configuration and personality setups should be clearly displayed.

The default settings of the different personalities should be made clear to the user.

Hospital trusts should give great consideration to the way that pump personalities are managed. If pumps are frequently moving between different care areas they may want to consider not using personalities and having one standard setting. Alternatively, a trust may wish to allow authorised users as well as clinical engineering staff to be able to change the personality settings. Manufacturers should discuss the different options with the trusts to best provide for their varying requirements.

In future these configurations may be controlled by wireless signals. For example, a pump would automatically configure to its paediatric settings when brought into a paediatric area. Or a critical care configuration would change to a general ward area set-up if the device is moved from the critical care environment. This will help to ensure that the pump is always operating with the correct personality, and help to minimise risk to the patient.

Recommendations

Draw multiple identical pumps with very suitable display differences

Paediatrics - 4.1ml/hourOcclusion alarm pressure lowAir in line detection - high

Critical care - 4ml/hourOcclusion alarm pressure lowAir in line detection - high

General ward - 4ml/hourOcclusion alarm pressure mediumAir in line detection - medium

Right hand page - same again - with clearer labelling

Critical care

Flowrate

fusing.. Infusing.. In

Volume infusedTotal volume to be infused

21.3ml500ml

4 ml/h

General ward

Flowrate

fusing.. Infusing.. In

Volume infusedTotal volume to be infused

21.3ml500ml

4 ml/h

Paediatrics

Flowrate

fusing.. Infusing.. In

Volume infusedTotal volume to be infused

21.3ml500ml

4.1 ml/h

General ward

Flowrate

nfusing Morphine 1mg/ml..

fusing.. Infusing.. In

Volume infusedTotal volume to be infused

1.3ml10ml

4 ml/h

P

S

M

L

S

MM

Critical care

Flowrate

nfusing Morphine 1mg/ml..

Volume infusedTotal volume to be infused

1.3ml10ml

4 ml/h

P

M

L

L

Paediatrics

Flowrate

nfusing Morphine 1mg/ml..

Volume infusedTotal volume to be infused

1.3ml10ml

4.1 ml/h

P

S

S

Draw multiple identical pumps with very suitable display differences

Paediatrics - 4.1ml/hourOcclusion alarm pressure lowAir in line detection - high

Critical care - 4ml/hourOcclusion alarm pressure lowAir in line detection - high

General ward - 4ml/hourOcclusion alarm pressure mediumAir in line detection - medium

Right hand page - same again - with clearer labelling

Critical care

Flowrate

fusing.. Infusing.. In

Volume infusedTotal volume to be infused

21.3ml500ml

4 ml/h

General ward

Flowrate

fusing.. Infusing.. In

Volume infusedTotal volume to be infused

21.3ml500ml

4 ml/h

Paediatrics

Flowrate

fusing.. Infusing.. In

Volume infusedTotal volume to be infused

21.3ml500ml

4.1 ml/h

General ward

Flowrate

nfusing Morphine 1mg/ml..

fusing.. Infusing.. In

Volume infusedTotal volume to be infused

1.3ml10ml

4 ml/h

P

S

M

L

S

MM

Critical care

Flowrate

nfusing Morphine 1mg/ml..

Volume infusedTotal volume to be infused

1.3ml10ml

4 ml/h

P

M

L

L

Paediatrics

Flowrate

nfusing Morphine 1mg/ml..

Volume infusedTotal volume to be infused

1.3ml10ml

4.1 ml/h

P

S

S

Software design Software design

Page 30: A guide to the design of infusion devices · 2008-07-24 · Design for patient safety: infusion devices is the first guidance published that focuses on the safe design of infusion

58 | A guide to the design of infusion devices

DRAFT

A guide to the design of infusion devices | 59

DRAFT

2.4 Setting up an infusion

Order of basic information entry - 1. Select mode (where appropriate) 2. Select drug (label AND dosage checking) 3. Enter volume 4. Enter rate of administration, ml/h, or time to be delivered, hr:min.

Instructions should be clearly highlighted on screen, particularly those requiring the user to press start or stop.

The number of steps required should be kept to a minimum. Indicate to users how many steps there are, and how far through the programming process they are.

Use prompts and menus to cue the user about important steps.

Show only the minimum options necessary.

There should be clear indicators as to what mode the pump is operating in. It should be very clear how to exit or change the mode.

Instructions can be unclear or hidden.

Programming may be a lengthy and unwieldy process, with very little information conveyed to the user as to the end of the process.

Prompts and roadmaps are absent, leaving the user lost and confused.

The user may become overwhelmed by too many options being displayed.

It can be unclear what mode a pump is operating in, which may have completely different settings assigned to it. Modes can also be very difficult to exit or change.

SET RATE

0.00 mL/hrENTER WHEN DONE

1

PROGRAM A LOADING DOSE?

YES OR NO?

2

SET LOADING DOSE0.00 MG

ENTER WHEN DONE

3Select mode:

1 2

3 4

5 6

CONTAINER SIZE

0.00 mLENTER WHEN DONE

4

PROGRAM REVIEW

5

PAIN MANAGEMENTCONTINUOUS ONLYPROGRAM IN mLs

6

RATE10mL/hr

LOADING DOSE1mL

7

CONTAINER SIZE250 mL

8

AIR ALARM 2mL

SAVING PROGRAM

REVIEW COMPLETEPRESS ENTER

9

13

10

INFUSED 0mLAT 10mL/hr

PRESS STARTTO INFUSE

INFUSED

LOADING DOSE

0.02mL

C

AT 10mL/hr

11

DELIVER THELOADING DOSE

NOW?YES OR NO

12

14

INFUSED 2.01mL

C

AT 10mL/hr

SET LOADING DOSE0.00 mg

ENTER WHEN DONE

3

CONCENTRATION1 mg/mL

LOADING DOSE4 mg

11

SET 4 HOUR LIMIT0.00 mg

7

ENTER WHEN DONE

15

AIR ALARM ON

REVIEW COMPLETEPRESS ENTER

19

INFUSED

LOADING DOSE

0.02mL

C

AT 10mL/hr

SET BOLUS DOSE0.00 mg

ENTER WHEN DONE

4

12

BOLUS DOSE2 mg

BOLUS LOCKOUT5 MINUTES

CONTAINER SIZE0.00 mg

in 0.00 mL

8

ENTER WHEN DONE

16

SAVING PROGRAM

20

INFUSED 3.01mL

C

AT 10mL/hr

CONCENTRATION0.00 mg/mL

ENTER WHEN DONE

1

BOLUS LOCKOUT0 MINUTES

ENTER WHEN DONE

5

PROGRAM REVIEW

9

13

4 HOUR LIMIT25 mg

17

INFUSED 0mLOR 0mg

PRESS STARTTO INFUSE

PROGRAM A LOADING DOSE?

YES OR NO?

2

1 4 HOUR LIMIT2 1 HOUR LIMIT3 # BOLUSES/HOUR4 NO LIMIT SEL.

6

PAIN MANAGEMENTBOLUS ONLY

10

14

CONTAINER SIZE250 mg 250 mLIN

18

DELIVER THELOADING DOSE

NOW?YES OR NO

ICU ward - PCA Mode

Program saved

Enter total volume50 mL

ICU adult

Press start

Time: Volume:

Flowrate: 25 ml/hr

Do you want to use Drugsure?

C OK

OK

Heparin SodiumVolume: 50mlTime: 12hr 30minRate: 4000 units/hr

Changemode OKC

Dobutamine 5mg/ml

Dopamine 4mg/ml

Heparin Sodium 1000units/ml

Midazolam 1mg/mlDobutamine 5mg/ml

Dopamine 4mg/ml

Heparin Sodium 1000u/ml

Midazolam 1mg/ml

Heparin Sodium 1000units/ml

50ml in 2hr00min

25ml/hr

Press start

50 ml

2 hr 00 min

Dose checkingDrug labelRate/time mode

1 1 2 3

Software design Software design

Recommendations•

Issues•

SET RATE

0.00 mL/hrENTER WHEN DONE

1

PROGRAM A LOADING DOSE?

YES OR NO?

2

SET LOADING DOSE0.00 MG

ENTER WHEN DONE

3Select mode:

1 2

3 4

5 6

CONTAINER SIZE

0.00 mLENTER WHEN DONE

4

PROGRAM REVIEW

5

PAIN MANAGEMENTCONTINUOUS ONLYPROGRAM IN mLs

6

RATE10mL/hr

LOADING DOSE1mL

7

CONTAINER SIZE250 mL

8

AIR ALARM 2mL

SAVING PROGRAM

REVIEW COMPLETEPRESS ENTER

9

13

10

INFUSED 0mLAT 10mL/hr

PRESS STARTTO INFUSE

INFUSED

LOADING DOSE

0.02mL

C

AT 10mL/hr

11

DELIVER THELOADING DOSE

NOW?YES OR NO

12

14

INFUSED 2.01mL

C

AT 10mL/hr

SET LOADING DOSE0.00 mg

ENTER WHEN DONE

3

CONCENTRATION1 mg/mL

LOADING DOSE4 mg

11

SET 4 HOUR LIMIT0.00 mg

7

ENTER WHEN DONE

15

AIR ALARM ON

REVIEW COMPLETEPRESS ENTER

19

INFUSED

LOADING DOSE

0.02mL

C

AT 10mL/hr

SET BOLUS DOSE0.00 mg

ENTER WHEN DONE

4

12

BOLUS DOSE2 mg

BOLUS LOCKOUT5 MINUTES

CONTAINER SIZE0.00 mg

in 0.00 mL

8

ENTER WHEN DONE

16

SAVING PROGRAM

20

INFUSED 3.01mL

C

AT 10mL/hr

CONCENTRATION0.00 mg/mL

ENTER WHEN DONE

1

BOLUS LOCKOUT0 MINUTES

ENTER WHEN DONE

5

PROGRAM REVIEW

9

13

4 HOUR LIMIT25 mg

17

INFUSED 0mLOR 0mg

PRESS STARTTO INFUSE

PROGRAM A LOADING DOSE?

YES OR NO?

2

1 4 HOUR LIMIT2 1 HOUR LIMIT3 # BOLUSES/HOUR4 NO LIMIT SEL.

6

PAIN MANAGEMENTBOLUS ONLY

10

14

CONTAINER SIZE250 mg 250 mLIN

18

DELIVER THELOADING DOSE

NOW?YES OR NO

ICU ward - PCA Mode

Program saved

Enter total volume50 mL

ICU adult

Press start

Time: Volume:

Flowrate: 25 ml/hr

Do you want to use Drugsure?

C OK

OK

Heparin SodiumVolume: 50mlTime: 12hr 30minRate: 4000 units/hr

Changemode OKC

Dobutamine 5mg/ml

Dopamine 4mg/ml

Heparin Sodium 1000units/ml

Midazolam 1mg/mlDobutamine 5mg/ml

Dopamine 4mg/ml

Heparin Sodium 1000u/ml

Midazolam 1mg/ml

Heparin Sodium 1000units/ml

50ml in 2hr00min

25ml/hr

Press start

50 ml

2 hr 00 min

Dose checkingDrug labelRate/time mode

1 1 2 3

Page 31: A guide to the design of infusion devices · 2008-07-24 · Design for patient safety: infusion devices is the first guidance published that focuses on the safe design of infusion

60 | A guide to the design of infusion devices

DRAFT

A guide to the design of infusion devices | 61

DRAFT

2.5 Feedback and communication

Ambiguous displays may lead users to believe that the pump is performing an action that it’s not.

Users may become confused about whether an action has been carried out by the pump.

It may be unclear as to whether a button press has been registered by the pump.

SET LOADING DOSE0.00 mg

ENTER WHEN DONE

3

CONCENTRATION1 mg/mL

LOADING DOSE4 mg

11

SET 4 HOUR LIMIT0.00 mg

7

ENTER WHEN DONE

15

AIR ALARM ON

REVIEW COMPLETEPRESS ENTER

19

INFUSED

LOADING DOSE

0.02mL

C

AT 10mL/hr

CONCENTRATION0.00 mg/mL

ENTER WHEN DONE

1

BOLUS LOCKOUT0 MINUTES

ENTER WHEN DONE

5

PROGRAM REVIEW

9

13

4 HOUR LIMIT25 mg

17

INFUSED 0mLOR 0mg

PRESS STARTTO INFUSE

Main screen

Main screen

OCCLUSION ALARM- HIGH- MEDIUM- LOW

PROGRAM A LOADING DOSE?

YES OR NO?

2

1 4 HOUR LIMIT2 1 HOUR LIMIT3 # BOLUSES/HOUR4 NO LIMIT SEL.

6

PAIN MANAGEMENTBOLUS ONLY

10

14

CONTAINER SIZE250 mg 250 mLIN

18

DELIVER THELOADING DOSE

NOW?YES OR NO

Choosing reverts to main screen. No indication of alarm level selected.

Distal occlusion 1 High2 Medium3 Low

Medium alarm level selected

Occlusion alarm1 High2 Medium3 Low

Alarm response: - Slow- Normal- Fast

Normal alarm response time

selected.

Users should be given continual feedback as to what the pump is doing.

Users want to know that a command is being carried out. If it can’t be carried out then they should be informed why.

Buttons should give tactile feedback to the user, to confirm that the button action has taken place.

SET LOADING DOSE0.00 mg

ENTER WHEN DONE

3

CONCENTRATION1 mg/mL

LOADING DOSE4 mg

11

SET 4 HOUR LIMIT0.00 mg

7

ENTER WHEN DONE

15

AIR ALARM ON

REVIEW COMPLETEPRESS ENTER

19

INFUSED

LOADING DOSE

0.02mL

C

AT 10mL/hr

CONCENTRATION0.00 mg/mL

ENTER WHEN DONE

1

BOLUS LOCKOUT0 MINUTES

ENTER WHEN DONE

5

PROGRAM REVIEW

9

13

4 HOUR LIMIT25 mg

17

INFUSED 0mLOR 0mg

PRESS STARTTO INFUSE

Main screen

Main screen

OCCLUSION ALARM- HIGH- MEDIUM- LOW

PROGRAM A LOADING DOSE?

YES OR NO?

2

1 4 HOUR LIMIT2 1 HOUR LIMIT3 # BOLUSES/HOUR4 NO LIMIT SEL.

6

PAIN MANAGEMENTBOLUS ONLY

10

14

CONTAINER SIZE250 mg 250 mLIN

18

DELIVER THELOADING DOSE

NOW?YES OR NO

Choosing reverts to main screen. No indication of alarm level selected.

Distal occlusion 1 High2 Medium3 Low

Medium alarm level selected

Occlusion alarm1 High2 Medium3 Low

Alarm response: - Slow- Normal- Fast

Normal alarm response time

selected.

SET LOADING DOSE0.00 mg

ENTER WHEN DONE

3

CONCENTRATION1 mg/mL

LOADING DOSE4 mg

11

SET 4 HOUR LIMIT0.00 mg

7

ENTER WHEN DONE

15

AIR ALARM ON

REVIEW COMPLETEPRESS ENTER

19

INFUSED

LOADING DOSE

0.02mL

C

AT 10mL/hr

CONCENTRATION0.00 mg/mL

ENTER WHEN DONE

1

BOLUS LOCKOUT0 MINUTES

ENTER WHEN DONE

5

PROGRAM REVIEW

9

13

4 HOUR LIMIT25 mg

17

INFUSED 0mLOR 0mg

PRESS STARTTO INFUSE

Main screen

Main screen

OCCLUSION ALARM- HIGH- MEDIUM- LOW

PROGRAM A LOADING DOSE?

YES OR NO?

2

1 4 HOUR LIMIT2 1 HOUR LIMIT3 # BOLUSES/HOUR4 NO LIMIT SEL.

6

PAIN MANAGEMENTBOLUS ONLY

10

14

CONTAINER SIZE250 mg 250 mLIN

18

DELIVER THELOADING DOSE

NOW?YES OR NO

Choosing reverts to main screen. No indication of alarm level selected.

Distal occlusion 1 High2 Medium3 Low

Medium alarm level selected

Occlusion alarm1 High2 Medium3 Low

Alarm response: - Slow- Normal- Fast

Normal alarm response time

selected.

Software design Software design

Recommendations•

Issues•

Page 32: A guide to the design of infusion devices · 2008-07-24 · Design for patient safety: infusion devices is the first guidance published that focuses on the safe design of infusion

62 | A guide to the design of infusion devices

DRAFT

A guide to the design of infusion devices | 63

DRAFT

2.6 Error recovery

Users may not be able to exit from an error message, resulting in them turning the pump off and on again.

Users cannot rectify a mistaken keypress.

Users cannot exit back to the main screen without changing settings.

Data can be easily deleted in error.

There should be clear instructions on screen as to how to rectify an error.

It should be clear to users how to rectify any mistaken keypresses and return to previous screens.

There should be a way of returning to the main screen without changing any settings. This will encourage users to explore the system and understand the system. If any settings have been changed the user should be clearly notified.

Where an action has potentially dangerous consequences, such as deleting data like the patient history, the default action should be the cancelling option.

Clearing history...

History cleared...

Main screen

- History- Volume Totals- Configuration- Clear History

Choosing clears history with no warning.

Automatically reverts back to main screen

Clear history?

- History- Volume Totals- Configuration- Clear History

- History- Volume Totals- Configuration- Clear History

ConfirmCancel

Clearing history...

History cleared...

Main screen

- History- Volume Totals- Configuration- Clear History

Choosing clears history with no warning.

Automatically reverts back to main screen

Clear history?

- History- Volume Totals- Configuration- Clear History

- History- Volume Totals- Configuration- Clear History

ConfirmCancel

Software design Software design

Recommendations•

Issues•

Page 33: A guide to the design of infusion devices · 2008-07-24 · Design for patient safety: infusion devices is the first guidance published that focuses on the safe design of infusion

64 | A guide to the design of infusion devices

DRAFT

A guide to the design of infusion devices | 65

DRAFT

2.7 Shortcuts & multi-function buttons

Shortcuts may create confusion amongst users who are not familiar with them. It also discourages users from exploring the pump features, for fear of causing undesirable consequences.

Hidden shortcuts may be inadvertantly activated without the user’s knowledge and cause changes to the infusion.

Buttons that are labelled with two symbols can be confusing and hard to identify.

1 2 3

4 5 6

7 8STANDBY

START

STOP

BOL

ml

ml

CC

9

.C F

F

0

7e.g. Bolus function

Start 1 2 3

4 5 6

7 8 9

. 0 C

StopMenu

History

Bolus

Shortcuts should be avoided where possible.

Where shortcuts must be used, the function should be confirmed by the user before operating. It should also be clear how to perform an action through the longer route.

Buttons should have one function only and be clearly marked and identifiable.

Software design Software design

Recommendations•

Issues•

1 2 3

4 5 6

7 8STANDBY

START

STOP

BOL

ml

ml

CC

9

.C F

F

0

7e.g. Bolus function

Start 1 2 3

4 5 6

7 8 9

. 0 C

StopMenu

History

Bolus

Page 34: A guide to the design of infusion devices · 2008-07-24 · Design for patient safety: infusion devices is the first guidance published that focuses on the safe design of infusion

66 | A guide to the design of infusion devices

DRAFT

A guide to the design of infusion devices | 67

DRAFT

2.8 Interface consistency

The pump may behave differently in different areas of the software.

Pumps from one manufacturer may have differing interfaces, requiring two different interfaces to be learnt by the user. This may cause confusion and error in operation and may also lead to additional time being spent training users. Primary

rateSecon-dary rate

PressureVolumeInfused

1 2 3

4 5 6

7 8 9

.

0

C

StartStop

Start 1 2 3

4 5 6

7 8 9

. 0 C

StopMenu

History

Bolus

Infusace

Infusace

Start 1 2 3

4 5 6

7 8 9

. 0 C

Stop

Menu

History

Bolus

Infusace

Infusace

Start 1 2 3

4 5 6

7 8 9

. 0 C

StopMenu

History

Bolus

There should be the same user interface and operation in all areas of the software. This means that the same set of actions should produce the same response from the pump throughout the software.

By applying consistent interfaces across a range of products, users are able to transfer a set of skills from one pump to another.

Primary rate

Secon-dary rate

PressureVolumeInfused

1 2 3

4 5 6

7 8 9

.

0

C

StartStop

Start 1 2 3

4 5 6

7 8 9

. 0 C

StopMenu

History

Bolus

Infusace

Infusace

Start 1 2 3

4 5 6

7 8 9

. 0 C

Stop

Menu

History

Bolus

Infusace

Infusace

Start 1 2 3

4 5 6

7 8 9

. 0 C

StopMenu

History

Bolus

Software design Software design

Recommendations•

Issues•

Page 35: A guide to the design of infusion devices · 2008-07-24 · Design for patient safety: infusion devices is the first guidance published that focuses on the safe design of infusion

68 | A guide to the design of infusion devices

DRAFT

A guide to the design of infusion devices | 69

DRAFT

2.9 Medicine labels

There is risk of pump mis-selection in a multiple device infusion system.

Sometimes paper labels are attached to devices to identify individual infusions, but these paper labels must be removed at the end of each infusion treatment and are not always easily read.

Medicine labels software (sometimes confusingly called drug library software) enables medicine names to be clearly displayed on the infusion device screens to help minimise device mis-selection.

It is important for users to be made aware of the difference between medicine label software and dose error reduction software (see next page).

Software design Software design

Recommendations•

Issues•

Page 36: A guide to the design of infusion devices · 2008-07-24 · Design for patient safety: infusion devices is the first guidance published that focuses on the safe design of infusion

70 | A guide to the design of infusion devices

DRAFT

A guide to the design of infusion devices | 71

DRAFT

DERS software alerts the user if they try to over or under infuse a specified infusion.

The user will be alerted to potential over or under infusion and will have the opportunity to select a safer rate of administration for the patient.

Issues•

2.10 Dose Error Reduction Software (DERS)

Users may think that the pump is using dose error reduction software when it is displaying a medicine label.

Devices with DERS software usually have default configurations with the DERS software switched off. Consequently the DERS may not be used on these pumps in practice.

Users are presented with traditional rate and volume of infusion screens at start-up and have to take separate actions to switch DERS on.

Programming and configuring DERS can be a time consuming and tedious process.

C

Safeguards are automatically selected. User must actively change this.

To remove dosage checking Continues with setup

Dose checking activated

ModesDose safeguards

No drug software

Syringe type50cc Codan

Start up screen

C OK

C OK

Default setting does not use dose checking software, which has been given an ambiguous name. User must actively select to engage the dose error reduction software e.g. ‘Drugsure’.

Medicine can then be chosen from a long list.

Do you want to use Drugsure?

Dobutamine 5mg/ml

Dopamine 4mg/ml

Heparin Sodium 1000units/ml

Midazolam 1mg/ml

C OK

C OK

Rate

Volume

100 mL/Hr

250 mL

Changemode

Drug-sure

Do you want to use Drugsure?

C OK

OK

Heparin SodiumVolume: 50mlTime: 12hr 30minRate: 4000 units/hr

Changemode

Self testing..

OK

OKC

Dobutamine 5mg/ml

Dopamine 4mg/ml

Heparin Sodium 1000units/ml

Midazolam 1mg/ml

The pump should clearly indicate whether it is using DERS or a medicine label.

The default configuration for infusion devices with DERS should be for the software to be on. When the device starts up, the user should be presented with the DERS medicine selection screen. Users should have to turn DERS off to obtain the traditional rate and volume to be infused screens.

The device should have comprehensive log-analysis software, where it monitors the number of occasions where DERS alerts have been activated, and what percentage of these have resulted in a revised rate of administration. It should also monitor the percentage of infusions where DERS is switched on and off.

It should be possible to develop and edit drug libraries and DERS on a spreadsheet based system and transfer this information to the device.

C

Safeguards are automatically selected. User must actively change this.

To remove dosage checking Continues with setup

Dose checking activated

ModesDose safeguards

No drug software

Syringe type50cc Codan

Start up screen

C OK

C OK

Default setting does not use dose checking software, which has been given an ambiguous name. User must actively select to engage the dose error reduction software e.g. ‘Drugsure’.

Medicine can then be chosen from a long list.

Do you want to use Drugsure?

Dobutamine 5mg/ml

Dopamine 4mg/ml

Heparin Sodium 1000units/ml

Midazolam 1mg/ml

C OK

C OK

Rate

Volume

100 mL/Hr

250 mL

Changemode

Drug-sure

Do you want to use Drugsure?

C OK

OK

Heparin SodiumVolume: 50mlTime: 12hr 30minRate: 4000 units/hr

Changemode

Self testing..

OK

OKC

Dobutamine 5mg/ml

Dopamine 4mg/ml

Heparin Sodium 1000units/ml

Midazolam 1mg/ml

Software design Software design

Recommendations•

Page 37: A guide to the design of infusion devices · 2008-07-24 · Design for patient safety: infusion devices is the first guidance published that focuses on the safe design of infusion

72 | A guide to the design of infusion devices

DRAFT

A guide to the design of infusion devices | 73

DRAFT

2.11 Data logging

The recent ECRI report requested a minimum data log of 1000 actions, and praised pumps that could store 10,000.19

When data is being deleted a confirmation message should come up to confirm the deleting action.

Data should be easily downloaded to a computer and be easy to read and access.

There should be a dedicated DERS log, which can store a minimum of 256 entries of alerts /alarms.

Events Log

Start 1 2 3

4 5 6

7 8 9

. 0 C

StopMenu

History

Bolus

Events Log

Start 1 2 3

4 5 6

7 8 9

. 0 C

StopMenu

History

Bolus

Pumps may store insufficient amounts of data.

Data may be inadvertantly lost by pressing the wrong button.

Data can be hard to access and awkward to view.

Keypress data is not separated from DERS logging data.

Software design Software design

Recommendations•

IssuesData logging records individual key presses and other operating events of the infusion device.

Review of this data is helpful in the event of a patient safety incident to determine how the pump was operated up to and beyond the incident.

Page 38: A guide to the design of infusion devices · 2008-07-24 · Design for patient safety: infusion devices is the first guidance published that focuses on the safe design of infusion

74 | A guide to the design of infusion devices

DRAFT

A guide to the design of infusion devices | 75

DRAFT

Infusace

Infusace

Infusace

Infusace

Start 1 2 3

4 5 6

7 8 9

. 0 C

StopMenu

History

Bolus

Start 1 2 3

4 5 6

7 8 9

. 0 C

StopMenu

History

Bolus

Morphine1mg/ml 1ml/h

Start 1 2 3

4 5 6

7 8 9

. 0 C

StopMenu

History

Bolus

Propofol1mg/ml 3ml/h

Start 1 2 3

4 5 6

7 8 9

. 0 C

StopMenu

History

Bolus

Noradrenaline1mg/ml 10ml/h

Infusace

Start 1 2 3

4 5 6

7 8 9

. 0 C

StopMenu

History

Bolus

Noradrenaline1mg/ml 10ml/h

Start 1 2 3

4 5 6

7 8 9

. 0 C

Stop

Menu

History

Bolus

InfusaceMr M SmithNoradrenaline 1mg/ml50ml over 4 hours to be givenintravenously

100

200

300

400

Code B3253 Big Logo

Super container XIntravenous Infusion BP

Generic name B

Lot

Manufacturer name Manufacturer number

Licence number Address line 1Address line 2

Each 500ml contains:

Infusion

Additive

Rate

Route

Aminophylline 1000mg/1000ml

100 ml/hr

ICU Ward

Volume infused: 0 mlTime Remaining: 4:50 hr:min

Battery remaining: 8 hours 25 min

2.12 Future technology

Infusions may be prepared in ward areas or supplied ready to administer. Users have to correctly select and connect the correct medicine, infusion device and administration set to the correct patient and programme the device accurately.

Use of bar code technology in the future will enable infusion pumps to check and record the correct selection of the medicine, administration set and patient, and consequently set-up the appropriate dose error prevention software.

Pumps should be enabled for wireless connectivity in order to facilitate data downloads and automatic personality profile activation. Bi-directional connectivity will be expected in the near future to enable smart software upgrades.

Infusace

Infusace

Infusace

Infusace

Start 1 2 3

4 5 6

7 8 9

. 0 C

StopMenu

History

Bolus

Start 1 2 3

4 5 6

7 8 9

. 0 C

StopMenu

History

Bolus

Morphine1mg/ml 1ml/h

Start 1 2 3

4 5 6

7 8 9

. 0 C

StopMenu

History

Bolus

Propofol1mg/ml 3ml/h

Start 1 2 3

4 5 6

7 8 9

. 0 C

StopMenu

History

Bolus

Noradrenaline1mg/ml 10ml/h

Infusace

Start 1 2 3

4 5 6

7 8 9

. 0 C

StopMenu

History

Bolus

Noradrenaline1mg/ml 10ml/h

Start 1 2 3

4 5 6

7 8 9

. 0 C

Stop

Menu

History

Bolus

InfusaceMr M SmithNoradrenaline 1mg/ml50ml over 4 hours to be givenintravenously

100

200

300

400

Code B3253 Big Logo

Super container XIntravenous Infusion BP

Generic name B

Lot

Manufacturer name Manufacturer number

Licence number Address line 1Address line 2

Each 500ml contains:

Infusion

Additive

Rate

Route

Aminophylline 1000mg/1000ml

100 ml/hr

ICU Ward

Volume infused: 0 mlTime Remaining: 4:50 hr:min

Battery remaining: 8 hours 25 min

Software design Software design

Recommendations•

Issue•

Page 39: A guide to the design of infusion devices · 2008-07-24 · Design for patient safety: infusion devices is the first guidance published that focuses on the safe design of infusion

A guide to the design of infusion devices | 77

DRAFT a

Administration sets

3

DRAFT COPY

A guide to the design of infusion devices | 77

Page 40: A guide to the design of infusion devices · 2008-07-24 · Design for patient safety: infusion devices is the first guidance published that focuses on the safe design of infusion

78 | A guide to the design of infusion devices

DRAFT

A guide to the design of infusion devices | 79

DRAFT

Recommendations•

3.1 Administration sets

Infusion systems may be very complex in practice. It is often difficult to identify the infusion conatiner and infusion line with the corresponding infusion catheter.

It can be easy to confuse administration sets. This can be especially problematic where the sets are for different routes of administration.

Consider the development of different coloured lines for different routes of administration. Spinal /epidural infusions should use the colour yellow, and arterial lines should use red. Manufacturers should consider the use of different connectors on their epidural administration sets.

Manufacturers may wish to consider the inclusion of labels with the administrations sets to promote good practice.

mg/ml

ANOL

mg/ml

ANOLDATE CONNECTED

mg/ml

bupivacaine

Sodium Chloride 0.9%

lp

hate

mg/ml

ANOL

mg/ml

ANOLDATE CONNECTED

mg/ml

bupivacaine

Sodium Chloride 0.9%

lp

hate

Administration sets Administration sets

Issues•

Page 41: A guide to the design of infusion devices · 2008-07-24 · Design for patient safety: infusion devices is the first guidance published that focuses on the safe design of infusion

80 | A guide to the design of infusion devices

DRAFT

A guide to the design of infusion devices | 81

DRAFT

10/200806K07BY

Do not reconnect partially used bags

Store below 25CStore out of reach of childrenDiscard any unused portion

Do not use unless solution is clear and container

Do not use unless solution is clear and containerThere use unless solution is clear and cosilly

se unless solutisilteni gjtotke today is stily y is stily

se unless solutise unless soluti

silteni gjtotke today is stilyy is stily

se unless solutise unless soluti

today is stily today is stily

y is stily

This is such a waste of

will anyone notice this

se unless solutisf hotnelstly

Healthcare Ltd

1

2

3

54

6

7

8

9

me the sonthem like a i

would tththe at make my head sorgte a i

InfusingInformatine 100ml/h

Misloaded

Recommendations•

3.2 Anti-freeflow systems

There can be a risk of uncontrolled ‘free flow’ of infusions in infusion devices.

Administrations sets can be easily removed from infusion pumps.

Syringe plungers are not retained or may become disengaged in syringe drivers, leading to the occurrence of free flow.

Anti-freeflow technology should be employed in infusion devices.

Volumetric pumps should automatically engage a valve or clamp when the administration set is removed from the pump.

Syringe drivers should ensure that the pump does not operate when the syringe driver is disengaged from the retaining clips.

Administration sets Administration sets

Issues•

Page 42: A guide to the design of infusion devices · 2008-07-24 · Design for patient safety: infusion devices is the first guidance published that focuses on the safe design of infusion

A guide to the design of infusion devices | 83

DRAFT a

Specialist pumps

4

DRAFT COPY

A guide to the design of infusion devices | 83

Page 43: A guide to the design of infusion devices · 2008-07-24 · Design for patient safety: infusion devices is the first guidance published that focuses on the safe design of infusion

84 | A guide to the design of infusion devices

DRAFT

A guide to the design of infusion devices | 85

DRAFT

4.1 Patient controlled analgesia devices

Setting up an infusion can be a long and complex procedure.

It may be unclear as to what mode the pump is operating in.

Setting up an infusion should be a clear and simple procedure. Manufacturers should consider the provision of a default setup for PCA pumps.

The pump should clearly indicate whether it is operating in a ‘continous’, ‘PCA’ or ‘continous&PCA’ mode.

Specialist pumps Specialist pumps

Recommendations•

Issues•

SET LOADING DOSE0.00 mg

ENTER WHEN DONE

3

CONCENTRATION1 mg/mL

LOADING DOSE4 mg

11

SET 4 HOUR LIMIT0.00 mg

7

ENTER WHEN DONE

15

AIR ALARM ON

REVIEW COMPLETEPRESS ENTER

19

INFUSED

LOADING DOSE

0.02mL

C

AT 10mL/hr

SET BOLUS DOSE0.00 mg

ENTER WHEN DONE

4

12

BOLUS DOSE2 mg

BOLUS LOCKOUT5 MINUTES

CONTAINER SIZE0.00 mg

in 0.00 mL

8

ENTER WHEN DONE

16

SAVING PROGRAM

20

INFUSED 3.01mL

C

AT 10mL/hr

CONCENTRATION0.00 mg/mL

ENTER WHEN DONE

1

BOLUS LOCKOUT0 MINUTES

ENTER WHEN DONE

5

PROGRAM REVIEW

9

13

4 HOUR LIMIT25 mg

17

INFUSED 0mLOR 0mg

PRESS STARTTO INFUSE

PROGRAM A LOADING DOSE?

YES OR NO?

2

1 4 HOUR LIMIT2 1 HOUR LIMIT3 # BOLUSES/HOUR4 NO LIMIT SEL.

6

PAIN MANAGEMENTBOLUS ONLY

10

14

CONTAINER SIZE250 mg 250 mLIN

18

DELIVER THELOADING DOSE

NOW?YES OR NO

ICU adult- PCA Mode

Programme saved

Loading dose: 4mgBolus dose: 2mgLock-out time: 5 min Limit: none

Confirm Change

Concentration: 1mg/mlContainer size: 250mg in 250ml

ICU adult - PCA mode

Press start

An example of a PCA setup process

SET LOADING DOSE0.00 mg

ENTER WHEN DONE

3

CONCENTRATION1 mg/mL

LOADING DOSE4 mg

11

SET 4 HOUR LIMIT0.00 mg

7

ENTER WHEN DONE

15

AIR ALARM ON

REVIEW COMPLETEPRESS ENTER

19

INFUSED

LOADING DOSE

0.02mL

C

AT 10mL/hr

SET BOLUS DOSE0.00 mg

ENTER WHEN DONE

4

12

BOLUS DOSE2 mg

BOLUS LOCKOUT5 MINUTES

CONTAINER SIZE0.00 mg

in 0.00 mL

8

ENTER WHEN DONE

16

SAVING PROGRAM

20

INFUSED 3.01mL

C

AT 10mL/hr

CONCENTRATION0.00 mg/mL

ENTER WHEN DONE

1

BOLUS LOCKOUT0 MINUTES

ENTER WHEN DONE

5

PROGRAM REVIEW

9

13

4 HOUR LIMIT25 mg

17

INFUSED 0mLOR 0mg

PRESS STARTTO INFUSE

PROGRAM A LOADING DOSE?

YES OR NO?

2

1 4 HOUR LIMIT2 1 HOUR LIMIT3 # BOLUSES/HOUR4 NO LIMIT SEL.

6

PAIN MANAGEMENTBOLUS ONLY

10

14

CONTAINER SIZE250 mg 250 mLIN

18

DELIVER THELOADING DOSE

NOW?YES OR NO

ICU adult- PCA Mode

Programme saved

Loading dose: 4mgBolus dose: 2mgLock-out time: 5 min Limit: none

Confirm Change

Concentration: 1mg/mlContainer size: 250mg in 250ml

ICU adult - PCA mode

Press start

Page 44: A guide to the design of infusion devices · 2008-07-24 · Design for patient safety: infusion devices is the first guidance published that focuses on the safe design of infusion

86 | A guide to the design of infusion devices

DRAFT

A guide to the design of infusion devices | 87

DRAFT

4.2 Devices for epidural use

Users are easily confused between pumps that are being used for epidural infusions and those that are being used for intravenous infusions.

Identical, unlabelled infusion administration sets may be used for both epidural and intravenous administration, and can be connected to the wrong route and the wrong infusion container.

The label, colour and design of epidural infusion pumps, syringe drivers, and the infusion systems they are attached to, should clearly identify them as for epidural use only.

The administration sets for epidural infusions should be clearly indentifiable through the use of labels or colour. Manufacturers should consider the use of different connectors on their epidural administration sets.

Specialist pumps Specialist pumps

Recommendations•

Issues•

Page 45: A guide to the design of infusion devices · 2008-07-24 · Design for patient safety: infusion devices is the first guidance published that focuses on the safe design of infusion

88 | A guide to the design of infusion devices

DRAFT

A guide to the design of infusion devices | 89

DRAFT Specialist pumpsSpecialist pumps

4.3 Devices for ambulatory care

Pumps may be large and cumbersome, making them awkward and embarrassing for patients to carry around.

Outpatients require discrete pumps, but flashing lights may cause distress by attracting attention.20

It may be complicated to set up an infusion and be awkward to load the administration set.

Ambulatory pumps should be small and light for maximum patient comfort. This may mean that the pump is too small for on-ward use.

Flashing lights should be concealed where possible. Consider vibratory alerts to alert the user to potential problems.

It should be easy to set up an infusion and load the administration set.

Recommendations•

Issues•

Some pumps are programmed in mm of syringe plunger travel per hour or per 24 hours. This can cause confusion and is prone to changes in the infusion rate.

It can be hard to determine battery life and know when the battery will run out.

1. 2123456789

onoff

All infusions should be calculated and programmed in ml/hr.

Pumps should give an indication of battery life and alert the user to waning battery power.

Consideration should be given to the incorporation of a keylock into the pump.

1. 2

123456789

onoff

Page 46: A guide to the design of infusion devices · 2008-07-24 · Design for patient safety: infusion devices is the first guidance published that focuses on the safe design of infusion

90 | A guide to the design of infusion devices

DRAFT

A guide to the design of infusion devices | 91

DRAFT Training Training

Information and training for users

Stakeholders are invited to identify issues concerning the use of user manuals in practice and possible design solutions to improve safer user of these devices.

Page 47: A guide to the design of infusion devices · 2008-07-24 · Design for patient safety: infusion devices is the first guidance published that focuses on the safe design of infusion

92 | A guide to the design of infusion devices

DRAFT

A guide to the design of infusion devices | 93

DRAFT

References

References

National Patient Safety Agency. Improving infusion device safety. Patient Safety Notice 01. 2004. Available at www.npsa.nhs.uk/patientsafety/alerts-and-directives/notices/infusion-devices

Anon. Infusion devices. Understanding the patients perspective in order to avoid errors. Professional Nurse. 1st October 2003. 5-9.

Medicines and Healthcare Products Regulatory Agency (MHRA): Device Bulletin 2003 (02) ‘Infusion Systems’ 2003. Available at http://www.mhra.gov.uk/Publications/Safetyguidance/DeviceBulletins/CON007321

National Patient Safety Agency. The Infusion Device Toolkit. 2004. Available at www.npsa.nhs.uk/patientsafety/alerts-and-directives/notices/infusion-devices.

National Patient Safety Agency. Evaluation report on standardising and centralising infusion devices – a project to develop solutions for NHS Trusts. Full evaluation report.2004. Available at www.npsa.nhs.uk/patientsafety/alerts-and-directives/notices/infusion-devices.

Quinn C, Stevenson E, Glenister H. NPSA Infusion device toolkit: a cost-saving way to improve patient safety. Clinical Governance 2004; 9:195-199

Department of Health and Design Council. Design for patient safety. A system wide design led approach to tackling safety in the NHS 2003. Available at www.edc.eng.cam.uk/medical/reports.html Shingo, S. Mistake-proofing for operators: The ZQC System. Productivity Press, Incorporated. (1997).

1.

2.

3.

4.

5.

6.

7.

8.

References

Centre for Evidence based Purchasing.Available at www.pasa.nhs.uk/pasaweb/NHSprcourement/CEP

Bath Institute For Medical Engineering. MHRA Infusion Device Evaluations. Infusion systems and Pumps. Available http://staff.bath.ac.uk/mpscd/pumpevaluation/index.htm

Medicines and Healthcare Products Regulatory Agency. Available at http://www.mhra.gov.uk/Safetyinformation/Generalsafetyinformationandadvice/Product-specificinformationandadvice/Infusionsystemsandpumps/index.htm

Stability in normal use, Section 24 BRITISH STANDARD BS EN 60601-1:1990, BS 5724-1: 1989

9.

10.

11.

12.

Liu L, Holescher U & Gruchmann T, Symbol Comprehension in Different Countries: Experience Gained from Medical Device Area in A.Auinger (Hrsg.): Workshops-Proceedings der 5 fachübergreifenden Konferenz Mensch und Computer Wien: Oesterriechische Computer Gesellschaft, 2005, S 81-87

ISO7239 Development and principles for application of public information symbols

ISO/DIS 9186 Procedures for the development and testing of public information symbols

Colle & Hiszem (2004) Standing at a kiosk: Effects of key size and spacing on touch screen numeric keypad performance and user perference, Ergonom-ics, 47(13), 1406-1423

13.

14.

15.

16.

17.

18.

19.

20.

21.

Jin, Z. X., Plocher, T. & Kiff, L. (2007) Touch screen user interfaces for older adults: but-ton size and spacing. Universal access in human computer in-teraction. Coping with diversity, 4554/2007, 933-941, August 24.

Edworthy J & Hellier E. (2006) Alarms and human behaviour: implications for medical alarms, BJA 97 (1): 12-17

General Purpose Infusion Pumps, Health Devices October 2007, www.ecri.org

Quinn C. (2003) Infusion devices: understanding the patient perspective to avoid errors. Prof Nurse.Oct;19(2):79-83.

NHS Core Learning Unit. Infusion devices training programme. Available at: www.Corelearningunit.com/index.php?id=2.11

Page 48: A guide to the design of infusion devices · 2008-07-24 · Design for patient safety: infusion devices is the first guidance published that focuses on the safe design of infusion

94 | A guide to the design of infusion devices

DRAFT

A guide to the design of infusion devices | 95

DRAFT

Further reading

Further reading

Stakeholders are invited to recommend further reading to provide further understanding and background on this topic.

Further reading