INSTRUMENTASI CT SCAN KONVENSIONAL

PRAMONO, S.ST

Apakah Konvensional Itu?

Kuno – usang …………jadul ? Bagaimana dengan CT?

Axial slice by Slice

CONVENTIONAL AND SPIRAL/HELICAL CT

Generasi BaruJadul

Instumentasi CT Konvensional

1.TUBE Geometri berkas sinar pencil/ paralel

untuk (generasi I),dan narrow beam (generasi II)........... Waktu scan = lama

Geometri berkas sinar dan Fan beam (generasi III)

Heat capacity 1 MHU - 2 MHU Target angle 120

Instumentasi CT Konvensional

2. Meja CT/ COUCH - MOVEMENT Berdasarkan indek setting 2,3,5,10,20 mm3.Detektor - dua buah /satu baris detektor/linear one detector array; - gas/ kristal detektor4.Gerakan Tube dan Detektor - Translasi dan rotasi, - Rotasi-Rotasi ( Generasi III) - Diam - diam5. Stationary partcomputer = analog komputer

Proses Akuisisi Data Single Slice CT/ Slice-by-slice CT

1.Tabung sinar-x harus bergerak dengan kecepatan rotasi konstan 3600 (tube dan detektor berputar di sekitar pasien dengan mengeluarkan X-Ray)

2. Detektor CT (baris-tunggal) menangkap x-ray / DATA POTONGAN TUNGGAL DARI JARINGAN yg DIIKUTI

Computer akan mengumpulkan dan mengukur data dari nilai atenuasi dari X-Ray yg sampai ke detektor, dan hasil yg didapatkan akan dicitrakan.

3. Tube dan Detektor berhenti berputar (Inter Scan Delay (ISD); Crawford dan King , 1990) 4. Meja CT kmd bergerak bertahap ke lokasi lokasi slice

berikutnya dengan index yg telah ditetapkan( Index 10 x 10 berarti meja akan melakukan perjalanan 10 mm dan setiap irisan dari 10 mm data yg akan dipindai berikutnya)

Proses selanjutnya berulang sampai selesai( Seram, 2001 capter13)

AKUISISI DATA PADA SETIAP GENERASI

FOUR STEP PROCESS OF CONVENTIONAL CT

TUBE AND DETECTORS ARE ACCELERATED TO CONST. SPEED

X-RAYTUBE ENERGIZEDDATA COLLECTED

FOR 360DEG.

TUBE & DETECTORS

SLOW DOWN & STOP

TABLE & PATIENT INDEXED TO THE

NEXT SCANNING POSITION

STOP

START 1 2

34

PARAMETER CONVENTIONAL CT 1. SECTION (Slice ) THICKNES 5. Scan Time 2. INDEX ? 6. fov 3. KV4. mAS

INDEX INDEXSCAN SCAN

Lanjutan..

1.Index = Gerakan meja CT dari slice ke slice berikutnya Nilainya biasanya digunkan= 1 mm- 20 mm Umumny digunakan: 3 mm, 5 mm dan 10 – 20 mm

2. SECTION (Slice ) THICKNES = tebal irisan scan Nilainya sebaiknya dibuat sama denan index

INDEX vs SECTION THICKNESS-

5 MM

5MM

5 MM 5 MM 5 MM

5MM 5MM 5MM

Index

Thickness 1 Thickness IV

INDEX vs SECTION THICKNESS

5MM SLICE –5 MM = NO GAP5MM SLICE –10 MM = GAP = MISREGISTRATION

5 MM

10MM

5 MM 5 MM

10MM 10MM

5 MM

INDEX vs SECTION THICKNESS50% OVERLAP SLICES5MM SLICE 2.5 MM INDEXScan Time meningkat

2.5

5MM 5MM

2.5 2.5 2.5

5MM 5MM 5MM

DATA ACQUSITION CONVENTIONAL CT

TIME

BREATHHOLD

BREATHHOLD

BREATHHOLD

CONVENTIONAL CTINTERSCAN DELAY TUBE COOLS OFF

TIME

ISDBREATH

HOLDBREATH

HOLDBREATH

HOLDISD ISD

Keuntungan ?

Dosis Radiasi <Detil gambar cukup baikCukup akurat untuk guiding biopsi

LIMITATIONS OF CONVENTIONAL CT

LONGER EXAM TIMEPERTINENTANATOMY OMISSION

(MISREGISTRATION) (Regauts et al, 1990).

INACCURATE GENERATION OF 3-D IMAGES (Kalender et al, 1990a).

DIFFICULTY IN MAINTAINING HIGH CONTRAST ENHANCEMENT

MISREGISTRATIONCONVENTIONAL CT

SPIRAL CT

No MISREGISTRATION

MPR IN CONVENTIONAL CT

Stair Step Artefact