27
1 Spectroscopic ANALYSIS Part 5 – Spectroscopic Analysis using UV-Visible Absorption Chulalongkorn University, Bangkok, Thailand January 2012 Dr Ron Beckett Water Studies Centre & School of Chemistry Monash University, Melbourne, Australia Email: [email protected] Water Studies Centre

1 Spectroscopic ANALYSIS Part 5 – Spectroscopic Analysis using UV-Visible Absorption Chulalongkorn University, Bangkok, Thailand January 2012 Dr Ron Beckett

Embed Size (px)

Citation preview

1

Spectroscopic ANALYSISPart 5 – Spectroscopic Analysis

using UV-Visible Absorption

Chulalongkorn University, Bangkok, Thailand January 2012

Dr Ron Beckett

Water Studies Centre & School of ChemistryMonash University, Melbourne, Australia

Email: [email protected]

2

UV-Visible Absorption Spectroscopy

1020 1018 1016 1014 1012 108

Cosm

ic

rays

-rays X-rays UV Vis

ible

Infrared Microwave

Electronic excitation

Bond breaking and ionization Vibration Rotation

Visible Spectrum

400 500 600 700

Absorption of UV and visible light by a molecule causes electronic excitation

3

UV-Visible spectral peaks result from electronic-vibrational transitions

Case (b) in the diagram is most common which gives the typical symmetric peak shape

4

Molecular Orbitals

• Bonding in organic molecules is based on overlap between s and p atomic orbitals.

• This can give rise to bonding and molecular orbitals, nonbonding n molecular orbitals antibonding * and * molecular orbitals

Two p atomic orbitals overlapping to give a and a* molecular orbital

5

Molecular Orbitals

+

px px

*

Two p atomic orbitals overlapping to give a bonding molecular orbital and anonbonding*molecular orbital

A

A B

A

B

B

6Electronic energy levels of polyatomic molecules

* (antibonding)

* (antibonding)

n (non-bonding)

(bonding)

(bonding)

Molecular Orbitals and Electronic Jumps

*

*

n * n *

7

Peak Position and the Type of Electronic Jump

Conjugated bonds

8

Peak Position for Molecules containing Double and Triple Bonds

9

Effect of Conjugation on Peak Position

The greater the number of conjugated double bonds the lower the energy jump and higher the wavelength of the UV-visible peak

*

10

Effect of Conjugation on Peak Position

Highly conjugated molecules may be coloured if the absorption peak moves into the visible region

11

Question Time !

Fanta has red and green colours !

Will red light pass through each of these solutions or will it be absorbed ?

(a) (b) (c) (d)

12

Question Time !

Fanta has red and green colours !

Will green light pass through each of these solutions or will it be absorbed ?

(a) (b) (c) (d)

13

Complementary Colours

ofmaximumabsorption

Colour Absorbed Colour Observed

380-440 violet-blue green-yellow

440-500 blue-green orange-red

500-580 green-yellow violet-blue

580-680 orange-red blue-green

680-780 purple green

When white light is absorbed by a chromophore, the eye detects the colours that are notnot absorbed. This is called the complementary colour to the colour absorbed.

V I B G Y O R

14

Colorimetric Analysis

Used for determination of the concentration of analytes in solution when:

1. The analyte is a coloured compound

2. The analyte produces a coloured species when a suitable reagent is added

15

Colorimetric Analysis

Photometric measurement

(a) visual comparison using colour standards

P Po

Determination of concentration depends on detection of change in colour intensity (absorption) at a particular wavelength.

Eye

16

Colorimetric Analysis(b) Colorimeter/Photometer

• Filters used to select a wavelength range

• Detection with photosensing device

PPo

Filter

wheel

Photodetector

17

Spectrophotometric Analysis(c) Spectrophotometer

– Spectral bandwidth ≤ 1 nm, i.e very monochromatic light.

– can operate in both the visible and UV ranges

– Colorimetry and spectrophotometry provide sensitivesensitive methods of analysis, i.e. ppm to ppb ranges.

PPo

PhotodetectorMonochromator

Prism or Grating Phototube, photomultiplier or photodiode

18

Single Beam Colorimeter•

Single beam spectrometerSingle beam spectrometer

19

Quantifying Light Absorption

Incident Light Intensity (PI) (sometimes Ii is used)

PI = Pr + Pa(solvent) + Pa(solute) + P

Pa(solvent) & Pa(solute) are absorbed light intensities

Pr ≈ 4% for air-glass interface

PPI

b

Absorbing solution of concentration,c.

Reflected beam

PrPa(solvent)

Pa(solute)Incident beam Transmitted beam

20

Quantifying Light Absorption

Transmitted Light (P0)

PI = Pr + Pa(solvent) + P0

P0 = PI - Pr - Pa(solvent)

P0PI

b

Absorbing solvent

Reflected beam

Pr

Pa(solvent)

Incident beam Transmitted beam

Intensity lost due to reflection and solvent absorption are removed by measuring the transmitted intensity of a blank containing only solvent

21

Quantifying Light Absorption

A = log P0

P

AbsorbanceAbsorbance is defined as

A = log (1/T) = log(100/%T)

TransmittanceTransmittance defined as

T = PP0

Thus

22

Relationship between Absorbance and Concentration

Beer-Lambert Law

A = l c

Where:• l is the path length in cmpath length in cm• c is the concentration concentration in mol/L • is the molarmolar absorptivityabsorptivity

23

Applications of the Beer-Lambert Law

Analysis of a single analyte1. Measure absorbance of a series of standard solutions

2. Plot a standard curve (should be a straight line ?)

3. Measure absorbance of unknown samples

4. Use standard curve to measure concentrations

Assumptions

– At fixed and l, is constant for a given solute

– the chemical matrix of the standards is the same as the sample.

Ax

Cx

4

3

2

1

0 2 3

A

A

A

A

A0

C C 1 C C C 4

Concentration

A = l c

24

Standard Addition Method

Used for samples with complex matrix &chemicalinterferences.

1. Measure A of sample

2. Repeat with known additions of standard to the sample.

Applications of the Beer-Lambert Law

Concentration of Standard added (mL)

0CAdd

Sample

Sample plus standard additions

Sample Concentration

25

Limitations of the Beer-Lambert Law

Concentration effects– B-L law applies to dilute solutions (negligible interaction

between solute ions).

– Higher concentrations of analyte (i.e. > 10-2 M) or high electrolyte concentrations, may produce molecular/ionic interactions which result inreduced light absorption at somewavelengths.

Concentration

Deviation from B-L law (loss of sensitivity)

Adherence to B-L law

26

Experimental Considerations

Wavelength selection• Choose where A is large to obtain best sensitivity.

• Choose where dA/d = 0 or is small.A

bso

rban

ce

3

Wavelength

27

Experimental Considerations

Choice of reagents for colorimetric analysis

– Should be stable and pure

– Should not absorb at of measurement

– Should react rapidly with analyte to give a stable

coloured compound (chromophore).

– Absorptivity, should not be sensitive to minor

changes in pH, Temp., electrolyte changes, etc.

– Should be selective for the analyte of interest.