Please read this Manual / Instructions carefully prior to using the product for your own safety and proper use of the product. If not followed the manual, it may result in an accident or malfunctioning of the product. Caution APC-30D / APM -30D APC–30DR / APM-30DR APC-50D / APM -50D APC–50DR / APM-50DR ASTEC CO., LTD User Manual - Penguin AQ

Astec - Penguin AQ (APC/APM) Incubator User Manual - Supplied by RI UK and Ireland

Embed Size (px)

DESCRIPTION

Astec - Penguin AQ (APC/APM) Incubator User Manual - Supplied by RI UK and Ireland.

Citation preview

Page 1: Astec - Penguin AQ (APC/APM) Incubator User Manual - Supplied by RI UK and Ireland

Please read this Manual / Instructions carefully prior to using the

product for your own safety and proper use of the product. If not

followed the manual, it may result in an accident or malfunctioning

of the product.

Caution

APC-30D / APM-30D APC–30DR / APM-30DR

APC-50D / APM-50D APC–50DR / APM-50DR

ASTEC CO., LTD

User Manual - Penguin AQ

Page 2: Astec - Penguin AQ (APC/APM) Incubator User Manual - Supplied by RI UK and Ireland

2

1. General Description

1-1 About this manual

1-2 Warranty and Liability

1-3 Warnings

1-4 Warning Labels

1-5 Earth Connection

1-6 CE Mark

1-6-1 Electromagnetic Compatibility

4

4

4

4

5

8

8

8

2. Bio Hazard 12

3. Important Notice 13

4. Copy Rights 13

5. Product overview 14

6. Specifications 15

7. Accessories 17

8. Dimensions 17

9. Diagram 18

10. Control Panel 19

11. Set up the Incubator

11-1 Location

11-2 Power

11-3 Power Code

11-4 Two Stacked Chambers

11-5 Earth Connection

11-6 Solenoid Valve

21

21

21

21

22

23

23

12. Shelves and Supports assembling 24

13. Water Jacket 26

14. Set Up Mode 27

15. Humidity 28

16. Gas Connections 29

17. Operations

17-1 Basic Settings

17-1-1 Temperature Setting

17-1-2 Over Temp Setting

17-1-3 CO2 Concentration Setting

17-1-4 Gas Selection Setting

17-1-5 O2 Concentration Setting

30

30

30

30

30

30

31

INDEX of Contents

Page 3: Astec - Penguin AQ (APC/APM) Incubator User Manual - Supplied by RI UK and Ireland

3

17-2 Settings for Suspending Gas Controls and Alarms

17-2-1 CO2 ON/OFF Setting

17-2-2 O2 ON/OFF Setting

17-2-3 CO2 / O2 Alarm ON/OFF Setting

31

31

31

31

18. Alarms

18-1 ADD WATER Alarm

18-2 OVER TEMP. Alarm

18-3 CO2 Alarm

18-4 O2 Alarm

32

32

32

32

32

19. CO2 Concentration Check

19-1 CO2 Check prior to supplying CO2 gas to Incubator (CO2 ZERO)

19-2 CO2 Check

33

33

33

20. Sampling Mode 33

Flow Chart of Setting Operations 34

21. Adjustment and Calibration

21-1 Temperature Adjustment (In-Chamber Temperature)

21-2 Displayed Temperature Adjustment

21-3 Door Temperature Adjustment

21-4 CO2 Zero Adjustment for AP series with T/C sensor

21-5 O2 Zero Adjustment

21-6 CO2 Adjustment

21-7 CO2 Adjustment Procedure

35

35

35

36

36

36

36

36

22. O2 Zero Adjustment and Calibration

22-1 Automatic CO2 Calibration

22-2 Manual CO2 Calibration

22-3 O2 Sensor failure

Flow Chart of Adjustment and Settings

38

38

38

38

39

23. Calibration Settings

23-1 All format

23-2 CO2 span format

23-3 Cleaning and Disinfections

24-3-1 Cleaning

24-3-2 Disinfections

40

40

40

41

41

41

24. Trouble Shooting 42

25. Regulator

26. Air Dust Filter Replacement Instruction

45

47

Factory Settings

48

Page 4: Astec - Penguin AQ (APC/APM) Incubator User Manual - Supplied by RI UK and Ireland

4

1-1 About This manual

1-2 Warranty and Liability

We warrant the product you have purchased for one calendar year after the date of

purchase. In case of any malfunctions caused by the manufacturer during this period of

time, ASTEC will be responsible for repair or replacement of failed parts or even the

instrument itself as circumstance allows for free of charge.

However, this warranty excludes the following conditions.

♦ Any damages due to fire, earthquake, thunderstorm, and other catastrophic events and

pollution and abnormal electrical supply.

♦ Any damages due to unduly repair, adjustment, calibration and modification.

♦ Any damages due to unduly usage or mishandling.

♦ Any damages caused by moving, dropping, transportation of the instrument.

♦ Replacement and repair of expendables and consumables.

For service, please contact our agent that you have purchased the instrument from.

1-3 Warnings 1. Connect the instrument to a line plug according to local or county standards and to the

specifications. 2. Do not use the instrument for different purposes from those shown by the manufacturer.

(See Chapter 1 for General Descriptions) 3. Do not switch on the instrument without having waited at least 20 seconds after

switching off. 4. Do not open the instrument before reading the servicing specific conditions described in

the manual. 5. Follow instruction contained in the manual in case of changing any parts 6. To move the instrument, at least two persons are required. We suggest to contact the

distributor or directly to the manufacturer for servicing.

1. General Description

Thank you for purchasing ASTEC AP series incubator (Penguin AQ). The following instructions are for your own protection and safety use. Please read carefully before start using the instrument. Failure to observe these precautions can result in property damage or personal injuries. Make sure to keep this manual handy so that you can read this whenever you may need it.

Warranty is Valid only if the installation is done by people trained, in conformity to instructions provided in this manual.

Page 5: Astec - Penguin AQ (APC/APM) Incubator User Manual - Supplied by RI UK and Ireland

5

1-4 Warning Labels Water Jacketed CO2/Multigas Incubator has an intrinsic danger due to the structure

designed for the purpose. For the safe use of the instrument, the following mark is labeled

on the instrument. Please use this instrument under a caution indicated on the labels.

Power inlet on the back of the incubator is labeled with the Warning sticker.

The following symbols are used in this manual.

Warning, Caution

Prohibited

Important Note

Ventilate the room in which you use carbon dioxide with the instrument. It can harm your

body without adequate air ventilation. Do not breath carbon dioxide directly.

Do not place flammable objects in the incubator, such as alcohol, benzene, ethanol or

propane gas. Especially when you use oxygen, do not use matches or lighters or

cigarettes into the laboratory.

Do not use this instrument outdoors. It could cause an electrical short or a lethal shock.

Ask the manufacturer or your sales agent for the setting up. This will prevent from the

events such as gas leaks, water leaks or any other unexpected problems caused by

mis-handling.

Warning This may cause serious or lethal injury.

Caution This may cause serious injury or damage

Warning

Make sure of ground wire of

power cord properly earthed.

Warning

Page 6: Astec - Penguin AQ (APC/APM) Incubator User Manual - Supplied by RI UK and Ireland

6

The location where the instrument will be placed must be steady, firm and leveled.

Do not leave this instrument in humid or wet conditions. It could cause an electrical short.

When double-stack the incubators, use the optional brackets. The top incubator may fall

by earthquake or other unexpected cause or accidents.

Connect only to a suitable power source specifically assigned for the instrument(s)

Using the instrument outside of Japan may require a transformer. Please ask the

manufacturer or your sales agent.

The power plug should be inserted all the way into the power outlet. Make sure that the

power plug is clean and dust-free when used.

Make sure that the instrument is securely earth-grounded.

Do not earth the instrument to the gas pipe, water pipe and lighting rod. This will cause

an electrical shock.

Never use the flammable objects with the instrument.

Make sure the type of gas and appropriate gas pressure are used with the instrument. All

the gas connections with gas tube(s) must be firmly connected and tightened with

cable-tie.

Do not plug any wire or pin into gas ports, sampling port.

Toxic, pathogenic and radioactive substances should be kept away from the instrument.

During maintenance and cleaning, make sure that power off the instrument and unplug the

power code.

Do not take apart or modify the instrument without first-hand supervising.

In case of malfunctions, turn off the power switch in the back of the instrument, and then,

unplug the power cord.

When the incubator is disposed, please be concerned for environment safety.

Page 7: Astec - Penguin AQ (APC/APM) Incubator User Manual - Supplied by RI UK and Ireland

7

Pedestal, bench, desk or table where the instrument will be set up MUST be solid and

leveled even to prevent leakage of the water in the Jacket.

Using the correct power source specifically assigned for the instrument is extremely

required.

Do not unplug the power cable with soaked hands.

In case any cuts or worn-out was found on the power cable of the instrument, stop using

the cable and replace it with brand-new power cable.

Do not operate the instrument with soaked hands.

Do not put heavy objects or bottles that contain liquid.

Do not get on the instrument.

When close the inner and outer doors please hold the knob and handle in appropriate

manner to avoid pinching your hands or fingers.

Do not intentionally step on, bend or twist the power cable.

Do not lean against the instrument.

Do not lean against the door of the instrument, especially when it is open.

Use appropriate gloves when cleaning the instrument, especially for cleaning the chamber.

Do not inhale any chemicals, organic compounds, dust or moisture when cleaning.

Power off and unplug the power cable when moving the instrument to different location.

Remove the water from the water pan when moving the incubator to the different location.

In case the instrument will be unused for a long period of time, remove the water in the

water pan and dry out the chamber completely.

Do not splash the water to the outside of the instrument especially on the top of it.

The receiver of O2 sensor should be kept away from condensation.

Used O2 sensor should be properly discarded in order to avoid contamination of the natural

environment.

Cautions

Page 8: Astec - Penguin AQ (APC/APM) Incubator User Manual - Supplied by RI UK and Ireland

8

1-5 Earth Connection

This symbol shows at earth protection connections.

1-6 CE Mark

Representative for CE: CGA Strumenti Scientifici Via Luca Giordano 7b 50132 Firenze – FI Tel. +39055571476 Fax +390555000889 E-Mail – [email protected]

1-6-1 Electromagnetic Compatibility

1-6-1-1 Requirements relating to electromagnetic compatibility

The instrument must be installed and put in service according to the EMC information contained

in this manual and any accompanying documents.

You need to pay attention at the fact that other devices (particularly those portable radio and

mobilei) can affect its comportment. The incubator should not be used near or above other

apparatus.

Page 9: Astec - Penguin AQ (APC/APM) Incubator User Manual - Supplied by RI UK and Ireland

9

1-6-1-2 Electromagnetic Emissions

The incubator is designed to work in the electromagnetic ambient here below specified. The

incubator user must warranty the instrument to work in that ambient.

Emission test Conformity Electromagnetic ambient - guide

Emissions RF CISPR 11 Group 1

Class B

uses RF energy only for its internal functioning.

For this its RF emissions are very low and

probably do not cause any interference in

apparatus located near it.

Armonic Emissions

IEC 61000-3-2

Class A

Voltage fluctuations/flicker

IEC 61000-3-3

Conform

The incubator is suitable to be used in every

ambient, including domestic and those connected

directly to a public net supply with low voltage

intended to supply buildings used for domestic

scopes.

1-6-1-3 Electromagnetic Immunity

The incubator is designed to work in the electromagnetic ambient here below described.

The user must warrantee to use the incubator in such ambient.

Immunity test Test level IEC 60601 Conformity level Electromagnetic ambient – guide

Electrostatic discharge

(ESD) IEC 61000-4-2

±6kV in contact

± 8kV in air

± 2kV in contact

(± 2; ±4; ± 8)kV in air

Floors must be in wood, concrete

or ceramics. If floors are covered

with synthetic material, relative

humidity should be at least 30%

A possible electrostatic discharge

can cause only a display error, no

significant error is generated.

Transitory/fast electrical

subsequent spikes

IEC 61000-4-4

± 2 kV for power supply lines

± 1 kV for I/O lines

± 2 kV for power supply lines

± 1 kV for I/O lines

The quality of the line voltage

should be that typical of a

commercial or hospital ambient.

Overvoltages

IEC 61000-4-5

± 1 kV differential mode

± 2 kV common mode

± 1 kV differential mode

± 2 kV common mode

The quality of the line voltage

should be that typical of a

commercial or hospital ambient

Page 10: Astec - Penguin AQ (APC/APM) Incubator User Manual - Supplied by RI UK and Ireland

10

Immunity test Test level IEC 60601 Conformity level Electromagnetic ambient – guide

Voltage holes, short

voltage breaks on input

supplì lines.

IEC 61000-4-11

<5% UT

(>95% hole in UT ) for 0,5

cycles

40% UT

(60% hole in UV) for 5 cycles

70% UT

(30% hole in UT)

for 25 cycles

<5% UT

(>95% hole in UT ) for 5 sec

<5% UT

(>95% hole in UT ) for 0,5

cycles

40% UT

(60% hole in UV) for 5 cycles

70% UT

(30% hole in UT)

for 25 cycles

<5% UT

(>95% hole in UT ) for 5 sec

The quality of the line voltage

should be that typical o fan

commercial or hospital ambient. If

the incubator user require a

continuous functioning also during

the line voltage break, we

recommend to supply the incubator

through a continuity group (UPS) or

with batteries.

Magnetic field at line

frequency (50/60 Hz)

IEC 61000-4-8

3 A/m 3 A/m Magnetic fields at line frequency

should have levels characteristic of

a typical place of a commercial or

hospital ambient

Page 11: Astec - Penguin AQ (APC/APM) Incubator User Manual - Supplied by RI UK and Ireland

11

Immunity test Test level IEC 60601 Conformity level Electromagnetic ambient – guide

Conducted RF

IEC 61000-4-6

3 Veff

from 150 kHz to 80 MHz out of

ISM(b) bands

10 Veff

from 150 kHz to 80 MHz

inside ISM(b) bands

3 Veff

from 150 kHz to 80 MHz out of

ISM(b) bands

10 Veff

from 150 kHz to 80 MHz inside

ISM(b) bands

The RF communication equipment

and mobile handsets should not be

used longer anywhere near the

incubator, cables, the distance

separating recommended

calculated using equation

applicable to the frequency of the

transmitter.

Recommended separating

distance

Pd 2,1= from 80 MHz to

800 MHz

Pd 3,2= from 800 MHz to

2,5 GHz

where P is the maximum transmitter

output power in Watt (W) as the

transmitter manufacturer and d is

the recommended distance in

meters (m).

The fixed RF transmitters field

intensity, as determined by

electromagnetic survey of the site

could be less than the level of

compliance in each frequency

range.

You can check interference in the

vicinity of equipment marked with

the following symbol:

NOTE:

Recommended separation distances between APM/APC incubators and radio

communications equipment and mobile handsets

The incubator is expected to operate in an electromagnetic environment in which irradiated

RF noises are under control. The User of the incubator can help prevent electromagnetic

Page 12: Astec - Penguin AQ (APC/APM) Incubator User Manual - Supplied by RI UK and Ireland

12

interference ensuring a minimum distance between appliances and mobile communications

portable RF (transmitters) and the incubator as below recommended, in connection with a

maximum power output of radio communication equipment.

Distance Separation at the transmitter frequency (m) Maximum power output of

transmitter specified (W)

from 80 MHz to 800 MHz from 800 MHz to 2.5 GHz

0,01 0,12 0,23

0,1 0,38 0,73

1 1,20 2,30

10 3,79 7,27

100 12,00 23,00

NOTE:

• At 80 MHz and 800 MHz the higher frequency interval is applied.

• These guidelines may not apply to all situations. The electromagnetic propagation is affected

dall'assorbimento and reflection on the part of structures, objects and people.

The data reported in this chapter apply for the equipment used in the manner and with

accessories specified in these instructions. The use of accessories other than those specified

may result in increased emissions, or a decrease of immunity.

This instrument may be used with radioactive products or pathogenic organism. If so,

please be careful not to be exposed or infected by these substances. If the instrument have

been used with radioactive products or pathogenic organism or any other substances which

may be hazardous to human health, and it needs to be sent back for maintenance, it must

be completely and thoroughly decontaminated before being sent back.

Also, in case the instrument will be used with radioactive products or pathogenic organism

or any other hazardous substances, please make sure to label the instrument as Biohazard.

2. Bio Hazard

Pd 2,1= Pd 3,2=

Page 13: Astec - Penguin AQ (APC/APM) Incubator User Manual - Supplied by RI UK and Ireland

13

★ The AP series incubator requires separate CO2 and N2 (for Multi-Gas type - APM) cylinder.

Premixed gas (ex. CO2: 5% O2: 5% N2: 90%) is prohibited.

★ A power outlet should be solely assigned to the incubator. Use of power strip or same

kind of connections is not recommended.

★ Acid, benzene, soap, toothpaste are not recommended to use for cleaning.

★ Use of brush to any part of the incubator for cleaning is not recommended.

★ Do not wipe the plastic and rubber area with volatile liquid such as benzene. Use

neutral detergent and wipe off completely with water.

★ Do not clean the incubator by poring water (especially on top panel) and wiping it off.

★ Pure distilled or even cleaner water is recommended for the humidification in the

chamber in order to avoid any chance of contamination. Check the volume of remaining

water at least once in a week.

★ Water for the humidification must be replaced or refilled as needed. The water should

be pre-warmed to the same temperature as the set temperature of the incubator.

★ Pedestal, bench, desk or table where the incubator will be set up MUST be solid and

leveled even to prevent leakage of the water in the Jacket.

★ Do not place flammable substances and volatile compounds in the incubator, such as

alcohol, benzene, ethanol, adhesive or propane gas. Especially when you use oxygen,

do not use matches or lighters and/or smoke cigarettes into the laboratory.

★ CO2 cylinder, which is equipped with siphon inside, cannot be used with the incubator.

★ Gas pressure to supply into the incubator such as CO2/N2/O2 should be maintained at

0.03~0.05MPa (4.35~7.25PSI).

★ In case the incubator will be kept powered off, remove the water for the humidification

and dry out the chamber completely.

★ The incubator MUST be properly earthed to ground.

★ The AP series incubator is double structured with stainless-steal water tank jacket inside.

Never attempt to make a hole or any kind of modification to the stainless tank.

★ Used O2 sensor should be properly discarded in order to avoid contamination of the

natural environment.

All rights are reserved by ASTEC Co., Ltd.

Revision, modification of the contents of this manual prior to permission is strictly prohibited.

3. Important Notice *Read this section to the end prior to use

4. Copy Rights

Page 14: Astec - Penguin AQ (APC/APM) Incubator User Manual - Supplied by RI UK and Ireland

14

Design Having display and control panel on the front door, ASTEC AP series incubator has enables

the small footprint while having water-jacket.

Chamber Structure Main chamber is made out of Stainless steel, and is double insulated with water jacket and

fiberglass insulation for secure temperature stability.

Temperature Control PID control independently works for the water jacket main heater located on the bottom of

the water jacket and door heater.

Door Heater

Ambient temperature significantly affects to the temperature distribution inside the incubator.

Heater on the front door is adjustable to cope with the fluctuation of ambient temperature.

Reducing the risk of Contamination

The air circulation in the chamber is vertical flow, which is believed to be more

anti-contamination. Stainless steel shelves, brackets and supports are removable without

tools for easy cleaning. Stainless steel chamber with completely coved corners is easy to

clean which enables the incubator to reduce the risk of contamination.

Optional built-in HEPA filter, which can be replaced without tools, optimizes the air

environment in the incubator and be preventive to contamination.

Humidifier

AP series utilizes Natural evaporation.

CO2 Sensor

IR sensor or TC sensor can be selected. Automatic and manual calibrations are available

with AP series as standard.

O2 Sensor (Multi-Gas: APM only)

O2 sensor operates on galvanic battery. Automatic and manual calibrations are available

with AP series as standard.

Safety Functions

Thermostat for temperature cut-off in case of deviation of water jacket temperature, Add

water alarm for decreased water in the water jacket, and alarm functions for deviation of

CO2, O2 are available to minimize the product loss.

5. Product overview

Page 15: Astec - Penguin AQ (APC/APM) Incubator User Manual - Supplied by RI UK and Ireland

15

APC-30D APC-30DR APC-50D APC-50DR

Temperature Controller PID Independent electronic control

Temperature Range 5C above ambient to 50C (122F)

Control Accuracy ±0.1C

Uniformity ±0.1C @37.0C (98.6F)

Humidity Method Natural evaporation with humidity pan (Standard)

RH Ambient to 95% @37.0C (98.6F)

CO2 Control CO2 Sensor T/C Infra Red T/C Infra Red

CO2 Range 0~19.9%

Readability 0.1%

CO2 Control ±0.1%

O2 Control O2 Sensor N/A

O2 Range N/A

Readability N/A

O2 Control N/A

Constructions and Electrical Exterior Dimensions W410 x D400 x H500 (mm)

16.1” x 15.7” x 19.7” (inch)

W450 x D472 x H605 (mm)

17.7” x 18.6” x 23.8” (inch)

Interior Dimensions W320 x D300 x H330 (mm)

12.6” x 11.8” x 19.0” (inch)

W360 x D370 x H430 (mm)

12.6” x 11.8” x 19.0” (inch)

Weight 28kg / 61.7lbs

(No water in the jacket)

46kg / 101.4lbs

(No water in the jacket)

Insulation Water jacket and fiberglass insulation

Interior Volume 32 Liters/1.13 ft3 57 Liters/2.01 ft3

Water Jacket Volume 11 Liters 14 Liters

Exterior Construction Cold- Rolled, Bonderized Steel, Powder coated

Interior Construction SUS304, 100% coved, round corners

Inner Door Tempered glass (5mm Thickness)

Shelves SUS304, W280 x D265 x H7 (mm),

3 shelves standard, 7 maximum

SUS304, W338 x D332 x H7 (mm),

3 shelves standard, 10 maximum

Air Flow Direct airflow by Plastic wheel fan

Operational Temp 5C~30C / 41F~86F

Operational Humidity 30~80% RH

Storage Temp -20C~50C / -3.9F~122F

Power Voltage 110V 60Hz /230V 50Hz

6. Specifications / CO2 Incubator

Page 16: Astec - Penguin AQ (APC/APM) Incubator User Manual - Supplied by RI UK and Ireland

16

APM-30D APM-30DR APM-50D APM-50DR

Temperature Controller PID Independent electronic control

Temperature Range 5C above ambient to 50C (122F)

Control Accuracy ±0.1C

Uniformity ±0.1C @37.0C (98.6F)

Humidity Method Natural evaporation with humidity pan (Standard)

RH Ambient to 95% @37.0C (98.6F)

CO2 Control CO2 Sensor T/C Infra Red T/C Infra Red

CO2 Range 0~19.9%

Readability 0.1%

CO2 Control ±0.1%

O2 Control O2 Sensor Galvanic Battery

O2 Range 1.0~89.0%

Readability 0.1%

O2 Control ±0.1%

Constructions and Electrical Exterior Dimensions W410 x D400 x H500 (mm)

16.1” x 15.7” x 19.7” (inch)

W450 x D472 x H605 (mm)

17.7” x 18.6” x 23.8” (inch)

Interior Dimensions W320 x D300 x H330 (mm)

12.6” x 11.8” x 19.0” (inch)

W360 x D370 x H430 (mm)

12.6” x 11.8” x 19.0” (inch)

Weight 28kg / 61.7lbs

(No water in water jacket)

46kg / 101.4lbs

(No water in water jacket)

Insulation Water jacket and fiberglass insulation

Interior Volume 32 Liters/1.13 ft3 50 Liters/2.01 ft3

Water Jacket Volume 11 Liters 14 Liters

Exterior Construction Cold- Rolled, Bonderized Steel, Powder coated

Interior Construction SUS304, 100% coved, round corners

Inner Door Tempered glass (5mm Thickness)

Shelves SUS304, W280 x D265 x H7 (mm),

3 shelves standard, 7 maximum

SUS304, W338 x D332 x H7 (mm),

3 shelves standard, 10 maximum

Air Flow Direct airflow by Plastic wheel fan

Operational Temp 5C~30C / 41F~86F

Operational Humidity 30~80% RH

Storage Temp -20C~50C / -3.9F~122F

Power Voltage 110V 60Hz(US)/230V 50Hz (EU)

6. Specifications / Triple Gas (Multi Gas) Incubator

Page 17: Astec - Penguin AQ (APC/APM) Incubator User Manual - Supplied by RI UK and Ireland

17

List of standard accessories associated with AP series. antiseptic

Side panel (Shelf support for right) 1 Manual 1

Side panel (Shelf support for left) 1 Plastic gas tube (3m) 1

Shelf brackets 6 Water filling silicon tube 1

Humidity Pan 1 Funnel 1

Ventilation ceiling 1 Anticorrosive solution 1

Shelves 3 Hexagonal wrench 1

3P adapter 1 Fill port 1

Knobbed screws for Ventilation Ceiling 2

Figure 1 Dimensions

7. Accessories

8. Dimensions

Page 18: Astec - Penguin AQ (APC/APM) Incubator User Manual - Supplied by RI UK and Ireland

18

Figure 2 Names of Components

1. Control Panel Displays the conditions of incubator and used for settings.

2. Adjustable Feet Adjust the physical level of incubator.

3. Outer Door Heated and insulated with fiberglass. Magnetic gasket.

4. Inner Door Inner door with sealed door trim.

5. Power inlet 110V 60Hz(US)/230V 50Hz (EU)

6. Circuit Breaker Protection for excess currency.

7. Recessed Power Switch

8. N2/O2 Gas Port

9. CO2 Gas Port

10. Gas Sample Port

11. Front Fill Port Used to fill the water jacket.

12. Air Inlet Port Used to calibrate IR CO2 (DR) and O2 (APM) sensor with ambient air.

13. Access Port *Optional chamber thru-wall access port with rubber plug for 30L type

Standard feature for 50L type

Figure 3

Components in Chamber

9. Diagram

CO2 Sensor Sampling Port

Fan

Monitor Temp Sensor

N2 Gas Port* CO2 Gas Port

O2 Sensor*

Screw

O2 Cal Tube*

* Multi-gas type only

Page 19: Astec - Penguin AQ (APC/APM) Incubator User Manual - Supplied by RI UK and Ireland

19

(1) (2) (3) (4)

APC-30DR/50DR (CO2 incubator)

Figure 4 Control panel for APC

1. Temperature Display

2. CO2 Display

3. SET key Mode select, Set

4. Alarm Silence key

5. Heating indicator Blinks when main heater being activated.

6. CO2 Inject indicator Blinks when CO2 being injected.

7. CO2 Setpoint Lights on when CO2 setting being modified.

8. Over Temp Setpoint Lights on when Over Temp setting being modified.

9. Temp Setpoint Lights on when Temperature setting being modified.

10. Programming Button Down cursor used when set parameter needs to be decreased.

11. Programming Button Up cursor used when set parameter needs to be increased.

12. Add Water Indicator Visual Alarm. Lights on when water in the water jacket needs to be refilled.

13. Over Temp Indicator Visual Alarm. Lights on when Over Temp alarm is activated.

14. CO2 Alarm Indicator Visual Alarm. Lights on when CO2 alarm is activated.

10. Control Panel

(5) (6) (7) (8) (9) (10) (11) (12)(13)(14)

Page 20: Astec - Penguin AQ (APC/APM) Incubator User Manual - Supplied by RI UK and Ireland

20

APM-30DR/50DR

Figure 5 Control Panel for APM

1. Temperature Display

2. CO2 Display

3. O2 Display

4. SET key Mode select, Set

5. Alarm Silence key

6. Heating indicator Blinks when main heater being activated.

7. N2/O2 Inject indicator Blinds when CO2 being injected.

8. CO2 Inject indicator Blinds when CO2 being injected.

9. CO2 Setpoint Lights on when CO2 setting being modified.

10. Over Temp Setpoint Lights on when Over Temp setting being modified.

11. Temp Setpoint Lights on when Temperature setting being modified.

12. O2 Setpoint Lights on when O2 setting being modified.

13. N2/O2 indicator Selectable Gas indicator. Lights on when N2/ O2 selectable.

14. Programming Button Down Key used when set parameter needs to be decreased.

15. Programming Button Up key used when set parameter needs to be increased.

16. Add Water Indicator Visual Alarm. Lights on when water in the water jacket needs to

be refilled.

17. Over Temp Indicator Visual Alarm. Lights on when Over Temp alarm is activated.

18. CO2 Alarm Indicator Visual Alarm. Lights on when CO2 alarm is activated.

19. O2 Alarm Indicator Visual Alarm. Lights on when O2 alarm is activated.

(1) (2) (3) (4) (5)

(6) (7) (8) (9) (10) (11) (12) (13) (14) (15) (16)(17)(18)(19)

10 . Contro l Pane l -Continued-

Page 21: Astec - Penguin AQ (APC/APM) Incubator User Manual - Supplied by RI UK and Ireland

21

11-1 Location

l The place (bench, table or stand) where the incubator will be set up MUST be steady,

firm and level. Make sure the surface that incubator will be placed by using adjustable

feet as needed. To move the instrument, two or more persons are needed.

l Please keep 25cm or more of clearance in the back and both side of the incubator.

l Avoid placing an incubator where cold air drifts directly to the incubator such as air

conditioning or air vent. It may cause condensation in the chamber.

l All the selves, brackets, supports and surface of the chamber are thoroughly cleaned

and sterilized with ethanol or autoclaves to make sure there is no contaminant on all

these accessories. The laboratory itself introduces common source of contamination,

incubator is designed to promote growth, through and constant cleaning are

recommended.

l Choose the place in the laboratory where there is no drastic change in temperature is

expected to avoid condensation in the chamber. Preferably, incubator should be

placed in a clean room environment such as biological clean room.

11-2 Power

This incubator is designed to be operated with 110V 60Hz(US)/230V 50Hz (EU). Insert

the supply cable in a plug with correct 110V 60Hz(US)/230V 50Hz (EU). The power plug

connection must be independent, and provided with adequate earth connection.

11-3 Power Code

If power outlet has 2 poles, avoid multiple connections using power strips.

11. Set up the Incubator

Page 22: Astec - Penguin AQ (APC/APM) Incubator User Manual - Supplied by RI UK and Ireland

22

11-4 Two Stacked Chambers

For space saving configuration, an incubator could be piled on the other. The optional

hardware (a set of 2 brackets ③ and ④ in Figure 6) is needed to safely and steadily stack

2 incubators. ASTEC does not recommend stacking more than 3 incubators for safety

reason.

How to stack two incubators

A. Remove the top panel [②] of the incubator which will be placed at the bottom by

unscrewing the screw [①] in Figure 1.

B. Securely mount the stacking hardware [③ and ④ in Figure 6] to the top frame of the

bottom incubator with optional four of flat headed screws [⑤].

C. Remove the adjustable feet [⑥] from the incubator that will be placed on the top of the

other.

D. Slowly stack the top incubator onto the bottom incubator, and securely fix the incubator

with optional doom headed screws [⑦] as shown in Figure 6.

Figure 6 Stacking Incubators

⑤ Optional Flat headed screw

⑦ Optional Doom headed screw

Page 23: Astec - Penguin AQ (APC/APM) Incubator User Manual - Supplied by RI UK and Ireland

23

11-5 Earth Connection

The incubator must be earthed to the ground in order to prevent electrical shock in case of

deviation in electronic insulation.

Do not earth the incubator to a water pipe or plastic make pipe.

Do not earth the incubator to a gas pipe, telephone line or lighting conductor.

11-6 Solenoid valve

The solenoid valve (s) will not start functioning for the first one minute after powering the

incubator on in order to stabilize and initialize itself.

Page 24: Astec - Penguin AQ (APC/APM) Incubator User Manual - Supplied by RI UK and Ireland

24

Firmly built stainless steel shelves and supports can be assembled and removed without

using any tools for easy cleaning and adjustment.

Clean the shelves, humidity pan and supports with 70% ethanol or inverted soap.

All the interiors (shelves, supports, brackets) should be built in BEFORE the water

jacket is filled with appropriate water. The water jacket expands when it is filled up.

1. Mount the shelve support inner walls to both right and left side as indicated in Figure 7

and 8.

2. Mount the top part of the shelve system (ceiling) to the top part in the chamber, and hold

it up with your hand as indicated in Figure 9.

3. Screw the ceiling onto the top inside of the chamber with knobbed screws as shown in

Figure 10.

Any part of the edge of the hole in the ceiling for the Fan must NOT touch the Fan.

12. Shelves and Supports assembling

Figure 7 Figure 8

Figure 9 Figure 10

Page 25: Astec - Penguin AQ (APC/APM) Incubator User Manual - Supplied by RI UK and Ireland

25

Mount the Brackets for the shelves to the holes on the support inner wall with the shelf

assignment you need for the incubator. 3 shelves are come as standard.

1. A bracket comes with two hooks on each. Mount the bracket to the support inner wall

to the place you wish to mount a shelf as shown in Figure 11.

2. Slide the bracket into the slits that are made to the further holes as shown in Figure 12.

3. Hook the bracket down to the corresponding hole as shown in Figure 13.

Figure 11 Figure 12

Figure 13

12. Shelves and Supports assembling -Continued-

Page 26: Astec - Penguin AQ (APC/APM) Incubator User Manual - Supplied by RI UK and Ireland

26

All the interiors (shelves, supports, brackets) should be built in BEFORE the water

jacket is filled with appropriate water. The water jacket expands when it is filled up.

1. Take the plug out of the water port with supplied hexagon wrench as shown in Figure 14,

and plug in the supplied connector to the water port as shown in Figure 15.

2. Turn the power on and make sure the “ADD Water” indicator blinks with beeps. Wait

for about 5 seconds until the incubator detects “No water in the jacket” to activate the

indicator and beeps.

3. Press and hold the SET key for about 10 seconds, the incubator will be in “Set Up

mode”. ? Please see the page 26 for Set Up Mode

4. Supply water into the water jacket from the water port with supplied hose or any other

alternative hose until the beeping stops.

5. Supplied Anticorrosive solution must be supplied to the water jacket before the water

jacket is full.

13. Water Jacket

Figure 14 Figure 15

Use RO water or cleaner water for the water jacket, however, water with high resistance such as ultra pure water would cause malfunction of the water level sensor in the jacket.

If “Set Up mode” is not used when water being supplied, CO2 /O2 alarm would be activated.

Appropriate resistance is needed for the water level alarm to work properly, make sure to provide Anticorrosive solution to the water jacket to gain resistance in the water.

Page 27: Astec - Penguin AQ (APC/APM) Incubator User Manual - Supplied by RI UK and Ireland

27

The Set Up Mode is recommended for setting up the incubator for the first time, or

setting it up again after long time unused period.

Set Up Mode: Press and hold the SET key on the control panel for about 10

seconds.

During the Set Up Mode, the Temp display shows the actual temperature and “SET” by turn.

The Set Up mode enables the incubator to properly set itself up automatically. It

automatically sets the pre-fixed parameters shown below after 12 hours from the time that

Set Up mode is activated. During the 12 hours, incubator will shut CO2/O2 solenoid valves

but warm up the water jacket in order to bring the temperature in the chamber to the pre-set

temperature (37C). After 12 hours, it starts controlling the CO2/O2 to achieve the pre-set

concentrations.

Set Up Mode / Pre-fixed parameters Temperature 37 Celsius

Over Temperature Set 38 Celsius

CO2 Concentration 5.0 %

Gas used with CO2 (APM only) N2

O2 Concentration (APM only) 5.0 %

* If you wish to change these settings, change them after Set Up mode is done after 12

hours.

14. Set Up Mode

Before using Set Up mode, make sure there is NO CO2 gas in the chamber. The air in the chamber must be completely replaced with the ambient air.

Humidity Water MUST be placed in the chamber when Set Up mode is in effective. ? See page 22.

Page 28: Astec - Penguin AQ (APC/APM) Incubator User Manual - Supplied by RI UK and Ireland

28

Water in the humidity water pan should be distilled or pure water and supplied in the pan to

1~2 cm from the bottom of the pan. Humidity water, because it is stagnant, can be

common source of contamination within the incubator. So we strongly recommend that the

humidity water should be replaced weekly or more often, and the water pan should be

autoclaved or wipe down with a 70% of isopropyl solution or 70% ethanol.

It is also recommended to warm up the humidity water to about 37C before it is pored into

the pan for faster evaporation regain.

15. Humidity

When incubator is powered off more than 30 minutes, make sure to remove the water from the chamber. When incubator is powered back up, make sure there is NO humidity in the chamber. Humidity can harm CO2 and O2 sensor when incubator is powered off.

Page 29: Astec - Penguin AQ (APC/APM) Incubator User Manual - Supplied by RI UK and Ireland

29

Gas port(s) is located on the back of the incubator as shown in Figure 16. Connect plastic

hose(s) (supplied with the incubator) and firmly tie them up with cable ties. The other ends

of the hose(s) needs to be connected to gas regulator of gas supply tank(s).

*APM series only

16-1 Gas Pressure

The optimal gas pressure is 0.03~0.05 Mpa (0.3~0.5 bar / 4.35~7.25 psi)

16. Gas Connections

Figure 16

APC series (CO2 incubator) can only be connected with CO2 gas, and APM series (Multi-Gas type) can be connected with N2 or O2 gas of the user ’s choice. In case O2 gas is connected to the incubator, make sure to be cautious of the usageof fire as high concentration of O2 gas increases the combustibility of flammable substances. Excess gas pressure may cause the malfunction of the incubator. Be sure to select the correct gas type to your application of the incubator. ? See page 24 for gas selection for N2 and O2. Humidity and partial pressure of CO2 may cause the unstable control of high concentration O2 gas at the approximate range of 18~23%.

Gas Sampling port CO2 gas port N2 gas port*

Power cord Circuit Breaker Main Power Switch

Page 30: Astec - Penguin AQ (APC/APM) Incubator User Manual - Supplied by RI UK and Ireland

30

17-1 Basic Settings Every operation and parameter can be changed by using <SET> key on the front

control panel. Using <UP> and <Down> cursors allows changing the value for the

settings or changing the parameters. After changing the value or parameter, if the

operation is left unchanged about 7 seconds, it goes back to the normal display

which indicates the current condition of the incubator.

? For detailed descriptions of the control panel, see the Control panel section

on Page 12 and 13.

17-1-1 Temperature setting Press <SET> key once, the TEMP SET indicator lights on. Use <UP> and <DOWN>

cursors to set the temperature to the value you wish to set.

Press <SET> key once more, the control panel brings you to Over Temp Setting. If the

control is left unchanged for about 7 seconds, it returns to normal display.

17-1-2 Over Temp Setting Press <SET> key until the Over temp set point lights on. Use <UP> and <DOWN> cursors

to set the temperature to the value you wish to set. In case the incubator indicates the

temperature in the chamber reaches the temperature you set here, it activates the Over

Temp alarm.

? For detailed descriptions for Over Temp Alarm, see the Alarms section on

Page 26.

Press <SET> key once more, the control panel brings you to CO2 concentration Setting. If

the control is left unchanged for about 7 seconds, it returns to normal display.

17-1-3 CO2 Concentration Setting Press <SET> key until the CO2 set point lights on. Use <UP> and <DOWN> cursors to set

the CO2 concentration you wish to set.

For APM series: Press <SET> key once more, the control panel brings you to the Gas

selection (N2 or O2) Setting.

For APC series: Press <SET> key once more, the control panel brings you back to the

Temperature Setting. There is no Gas selection (N2 or O2) Setting on APC series.

If the control is left unchanged for about 7 seconds, it returns to normal display.

17. Operations

Page 31: Astec - Penguin AQ (APC/APM) Incubator User Manual - Supplied by RI UK and Ireland

31

17-1-4 Gas Selection Setting For APM series only Press <SET> key until the N2/O2 indicator lights on. Use <UP> cursor to choose N2 and

<DOWN> cursors to choose O2. Pressing <SET> key once more leads you to the next

setting O2 concentration setting (APM series only). If the control is left unchanged for

about 7 seconds, it returns to normal display.

17-1-5 O2 Concentration Setting Press <SET> key until the O2 set point lights on. Use <UP> and <DOWN> cursors to set

the O2 concentration you wish to set. Pressing <SET> key once more leads you back to

the Temperature setting. If the control is left unchanged for about 7 seconds, it returns to

normal display.

17-2 Settings for Suspending Gas controls and Alarms By pressing <SET> key with <SILENCE> key pressed, the incubator enters the setting

where Gas controls and Alarm function can be suspended by selecting ON or OFF.

Function Suspending: SILENCE + SET

17-2-1 CO2 ON/OFF Setting In Function Suspending mode, Temperature display shows co2 and CO2 display show ON or

OFF. The setting can be changed to ON or OFF by using <UP> and <DOWN> key.

When the setting is set on OFF, the solenoid valve completely closes itself, and the

incubator stops supplying CO2 gas.

17-2-2 O2 ON/OFF Setting For APM series only Pressing <SET> key once more when you are on O2 ON/OFF setting leads you to the O2

ON/OFF setting for APM series. Temperature display shows co2 and O2 display show ON or

OFF. The setting can be changed to ON or OFF by using <UP> and <DOWN> key.

When the setting is set on OFF, the solenoid valve completely closes itself, and the

incubator stops supplying O2 gas.

17-2-3 CO2 / O2 Alarm ON/OFF Setting Press <SET> key in Function suspending mode until Temperature display shows ALA.

Then, CO2 display shows ON or OFF. Use <UP> and <DOWN> key to change the setting.

This function only suspends CO2 and O2 alarm but the ADD Water alarm for water jacket

and Over Temp alarm are kept activated.

17. O p e r a t i o n s -Continued-

Page 32: Astec - Penguin AQ (APC/APM) Incubator User Manual - Supplied by RI UK and Ireland

32

18-1 ADD WATER Alarm ADD WATER Alarm activates when the water in the water jacket decreased. The alarm

gives out beeping sound that can be shut off by pressing <SILENCE> key. However, ADD

WATER indicator keeps blinking until the water is actually refilled.

? For refilling water in the water jacket, see P20 Section Water Jacket

18-2 OVER TEMP. Alarm OVER TEMP. Alarm activates when the temperature inside the chamber reaches to the

temperature that is set on OVER TEMP. Setting. The alarm gives out beeping sound that

can be shut off by pressing <SILENCE> key. However, OVER TEMP setpoint indicator

keeps blinking until the temperature is actually decreases. While the alarm is being

activated, incubator partially shuts down the power to all the heaters for safety reason.

18-3 CO2 Alarm CO2 Alarm activates when CO2 concentration is off ±1.0% or more from the concentration

that is set on CO2 concentration setting for 5 consecutive minutes. The alarm gives out

beeping sound and CO2 Alarm indicator blinks. By pressing <SILENCE> key, beeping

shuts down and the CO2 indicator turns off. The time delay for the alarm to activate is 5

minutes as a default configuration, however it can be modified at Adjustment B mode.

? See

18-3 O2 Alarm For APM series only O2 Alarm activates when O2 concentration is off ±5.0% more or less from the concentration

that is set for O2 concentration setting for 60 minutes. The alarm gives out beeping sound

and O2 Alarm indicator blinks. By pressing <SILENCE> key, beeping shuts down and the

O2 indicator turns off.

18. A l a r m s

Page 33: Astec - Penguin AQ (APC/APM) Incubator User Manual - Supplied by RI UK and Ireland

33

19-1 CO2 Check prior to supplying CO2 gas to Incubator (CO2 Zero) Temperature and Humidity must be adequately prepared prior to supplying CO2 gas to

incubator. CO2 concentration must be 0.0% (ambient air).

19-2 CO2 Check CO2 gas concentration must be checked frequently by CO2 analyzer for optimized use. If

the displayed CO2 concentration differs from the concentration measured with calibrated

CO2 analyzer, modify the concentration on the display of the incubator to the concentration

measured with the analyzer.

Before the Gas in the incubator is sampled out for concentration check, set the incubator to

“Sampling Mode”. The Sampling Mode prevent CO2, N2 or O2 gas from coming into the

chamber from the gas tanks in order to measure more precise concentration in the chamber.

? See the section 22 for Sampling Mode.

? See the section 23 for Adjustment and Calibration of Gas concentration.

For more accurate measurement of CO2 and O2 concentration, there should be none of

such gas supplied during the measurement. The sampling mode allows the user to shut

down the supply of gas for 3 minutes for reason above.

Sampling Mode : <UP> key pressed for 5 seconds

During the Sampling Mode, Temperature display shows SP and the current temperature by

turns for 3 minutes. Concentration check must be done during this period. After 3

minutes, the Sampling Mode automatically cancels itself and goes back to normal function.

19. CO2 Concentration Check

When you samples out some amount of Gas in the incubator, it sometimes activates false ADD WATER alarm. This is provoked by the gas sample port is capped with the Gas analyzer, and the analyzer pumps some gas out of the incubator and this causes negative pressure in the chamber. The false ADD WATER alarm will be deactivated when the pressure in the chamber regains positive pressure when gas sample port is open after sampling gas.

20. Sampling Mode

Page 34: Astec - Penguin AQ (APC/APM) Incubator User Manual - Supplied by RI UK and Ireland

34

Flow Chart of Setting Operations

Settings

Normal Function

Temperature Setting

OVER TEMP Setting

CO2 Concentration Setting

Gas Selection Setting

For APM series only

O2 Concentration Setting

SET

Pressed for 0.5 sec

SET

SET

SET

SET

For APC series

Settings for Suspending

Gas controls and Alarm

Normal Function

SET

CO2 ON/OFF Setting

O2 ON/OFF Setting

Alarm ON/OFF Setting

Silence

+

Pressed for 0.5 sec

SET

SET

SET

SET

Page 35: Astec - Penguin AQ (APC/APM) Incubator User Manual - Supplied by RI UK and Ireland

35

Adjustment / Calibration Mode is utilized for modifications to Temperature, CO2 and O2

concentrations by changing factory pre-set parameters as needed. The modifications

should be made accordingly to adjust the environment in which the incubator is used.

Some of the pre-set parameters should never be modified, as they are critical to overall

performance of the incubator. These parameters are also indicated below here in this

section.

By pressing <SET> key for about 0.5 sec with <UP> and <DOWN> keys pressed and held,

Adjustment / Calibration Mode will be activated.

Adjustment / Calibration Mode:

<UP> + <DOWN> Pressed and Held + SET pressed for 0.5 sec

In Adjustment / Calibration Mode, the temperature display shows ST1~ST6

respectively as <SILENCE> key is pressed. However, ST2, ST3, ST4 and ST5 are all

set by the factory and should not be changed or modified at all, except for C-5 under

ST2 and E-4 under ST5.

? See the flow chart of Adjustment / Calibration Mode on page 37 for further details.

21-1 Temperature Adjustment (In-chamber temperature) When Temperature display shows ST1 in Adjustment / Calibration Mode, press SET key to

go to section A-1. You will see a value in this section. Increasing or decreasing the value

here allows you to adjust the actual temperature in the chamber adjusted to the actual

temperature.

For example, when you need to increase the actual temperature by 0.2 C, increase the

value shown by 0.2 using <UP> and <DOWN> keys.

To achieve a precise adjustment of temperature, you need to use a calibrated thermometer

to measure the actual temperature correctly.

21-2 Displayed Temperature Adjustment (In-chamber temperature) When Temperature display shows ST1 in Adjustment / Calibration Mode, press SET key

until you see section A-2. Increasing or decreasing the value here allows you to adjust the

displayed temperature on the control panel.

For example, when you need to increase the displayed temperature by 0.2 C, increase

the value shown by 0.2 using <UP> and <DOWN> keys.

21. Adjustment and Calibration

Page 36: Astec - Penguin AQ (APC/APM) Incubator User Manual - Supplied by RI UK and Ireland

36

21-3 Door Temperature Adjustment When Temperature display shows ST1 in Adjustment / Calibration Mode, press SET key

until you see section A-3. Increasing or decreasing the value here allows you to adjust the

For example, when you need to increase the displayed temperature by 0.2 C, increase

the value shown by 0.2 using <UP> and <DOWN> keys.

21-4 CO2 Zero Adjustment for AP series with T/C sensor This section is used for AP series incubator equipped with T/C sensor.

Unused for APC/M-30DR series equipped with IR CO2 sensor.

21-5 O2 Zero Adjustment APM only When Temperature display shows ST1 in Adjustment / Calibration Mode, press SET key

until you see section A-5 on the control panel. Increasing or decreasing the value here

allows you to adjust the O2 ZERO on the control panel.

APM incubator has a function called automatic O2 sensor ZERO calibration, so that this

section will also be usually unused. However, the O2 sensor used for AP series is operated

by a galvanic battery and it does have a life span of several years. When the battery

deteriorates its out-voltage, this O2 Zero Adjustment can help up the voltage by gaining a

few percentages (by increasing the value here).

21-6 CO2 Adjustment When Temperature display shows ST1 in Adjustment / Calibration Mode, press SILENCE

key 5 times until you see section ST6 on the control panel. Pressing SET key allows you

to go to section F1 ~ F8, however, F4 ~ F8 are used only for factory settings and should

not be changed or modified by the users.

21-6-1 CO2 SPAN Adjustment F-1

When measured CO2 concentration is different from pre-set CO2 concentration

(see page 24, Section 19-1-3 for CO2 Concentration setting), and it needs to be

adjusted, change the value on F-1 to actual CO2 concentration by using <UP> and

<DOWN> cursors. It will allow the incubator to automatically adjust the value to

the measured value.

21-6-2 CO2 ZERO Manual Calibration F-2

This function allows the incubator to carry out manual ZERO Calibration. As a

default, F-2 shows ON on the display. Change this to OFF with <UP> and

<DOWN> cursors and wait for 10 second until the incubator starts manual

calibration.

Page 37: Astec - Penguin AQ (APC/APM) Incubator User Manual - Supplied by RI UK and Ireland

37

21-6-3 CO2 ZERO Auto Calibration Interval F-3

The interval (days) for Auto Calibration can be changed here by modifying the

number pre-set on F-3. Use <UP> and <DOWN> cursors to change the number

(days). The Auto Calibration interval is set 10 days.

21-7 CO2 Adjustment Procedure In case there is variation between displayed CO2 concentration and actually measured

value, go through the below adjustment procedure to calibrate for the concentrations to

match.

When Temperature display shows ST1 in Adjustment / Calibration Mode, press <SILENCE>

key 5 times until you see section ST6 on the control panel.

On ST6, Press SET key until F-2 shows up on the Temperature display.

On CO2 display, change displayed OFF to ON, Manual CO2 calibration will starts.

When Manual CO2 calibration is finished, Press <UP> key for 5 second, Sampling mode will

start and SP will show on the temperature display.

Analyze the actual CO2 concentration with CO2 analyzer.

Go to ST6 again and press SET key once to show F-1 on the temperature display.

Change the value on F-1 to the actual CO2 concentration you have just measured.

Leave the incubator for appropriate period until the CO2 concentration and humidity are

Do not change the F-4 ~ F-8 as these setting are only used by the manufacturer.

Page 38: Astec - Penguin AQ (APC/APM) Incubator User Manual - Supplied by RI UK and Ireland

38

stabilized. And check the CO2 concentration again for double check.

As described in the section 22-5, the O2 sensor used on APM operates on galvanic battery

and it has life span. It needs to be adjusted the voltage that effect the O2 concentration

control. In case the battery is extremely run out, it must be replaced.

There are two ways to calibrate the O2 sensor on APM series as described below.

22-1 Automatic O2 Calibration Automatic calibration works when the incubator is powered on and O2 setting is ON (see

Page 29 for section 18-2-2: O2 ON/OFF setting). The automatic calibration performs no

later than 12 hours after the incubator is first turned on. And it will perform once in every

week.

During the calibration, the O2 display shows the actual concentration of O2 and CAL by turn.

22-2 Manual O2 Calibration Manual O2 calibration should be carried out when O2 sensor is replaced, or the incubator

has been not used for long period of time.

22-2-1 Manual Calibration

Go to ST5 by pressing SET, <UP> and <DOWN> keys at the same time, go to E-4.

You will see OFF on the display. Change it to ON by press <UP> cursor and leave it

for a few seconds. The incubator starts manual calibration on O2 sensor.

22-3 O2 Sensor failure When the battery in O2 sensor wears out, O2 display shows the actual concentration and

“---“ (bars) by turn, which indicates O2 sensor failure.

The O2 sensor still controls the concentration even if the sensor failure is shown, but it is

recommended to replace the sensor with new one at earliest possible.

After replacing the sensor, make sure to do the Manual Calibration of the sensor.

? Please refer to the section 23-2-1 for Manual O2 sensor calibration.

22. O2 Adjustment and Calibration -APM only-

Page 39: Astec - Penguin AQ (APC/APM) Incubator User Manual - Supplied by RI UK and Ireland

39

Flow Chart of Adjustment and setting

Normal Function

Press <UP> + <Down> + SET all at once for few seconds

A-1 Chamber Temp

Adjustment

Adjustment Mode ST1

A-2 Displayed Temp

Adjustment

A-3 Door Temp Adjustment

A-4 CO2 Zero

Adjustment

A-5 O2 Zero

Adjustment

A-6 Ambient Temp Adjustment

SET

SET

SET

SET

SET

SET

SET

SILENCE

Calibration Mode ST2

Calibration Mode ST3

Parameter Mode ST4

Calibration Mode ST5

Calibration Mode ST6

SILENCE SILENCE SILENCE SILENCE SILENCE

B-1 Water Jacket Temp

Adjustment

B-2 Displayed Temp

Adjustment

B-3 Door Temp Sensor

Adjustment

C-1 CO2 Span

Adjustment

C-2 CO2 Solenoid valve

Control (Open)

C-3 CO2 Solenoid valve

Control (Close)

C-4 CO2 Solenoid valve

Control ON/OFF

C-5 CO2 Alarm Time

Setting

D-1 Water Temp

Control Setting

D-2 Water Temp

Control Setting

D-3 Water Temp

Control Setting

D-4 Water Temp

Anti-Reset Setting

D-5 Door Temp

Control Setting

D-6 Door Temp

Control Setting

D-7 Door Temp

Control Setting

D-8 Door Temp

Control Setting

SET

SET

SET

SET

SET

SET

SET

SET

SET

SET

SET

SET

SET

SET

SET

SET

E-1 O2 Solenoid valve

Control (Open)

E-2 O2 Solenoid valve

Control (Close)

E-3 O2 Solenoid valve

Control ON/OFF

E-4 O2 Sensor Manual

CAL Setting

E-5 Off-Set Voltage

Adjustment

SET

SET

SET

SET

SET

SET SET SET SET

F-1 CO2 Span

Adjustment

F-2 CO2 Manual

Zero CAL

F-3 CO2 Auto

Zero Interval

F-4 CO2 Solenoid

Proportional Cycle Setting

F-5 CO2 Solenoid

Start Cycle Setting

F-6 CO2 Solenoid

Filter HIGH Setting

F-7 CO2 Solenoid

Filter LOW Setting

F-8 CO2 Zero

CAL Value

SET

SET

SET

SET

SET

SET

SET

SET

Page 40: Astec - Penguin AQ (APC/APM) Incubator User Manual - Supplied by RI UK and Ireland

40

Calibration setting

ST 1 ST 2 ST 3 ST 4 ST 5 *3

A-1 0.00 B-1 XX.X C-1 X.X D-1 90.0 E-1 0.00

A-2 0.00 B-2 XX.X C-2 0.00 D-2 0.3 E-2 0.00

A-3 1.00 B-3 XX.X C-3 0.00 D-3 20.0 E-3 ON

A-4 0.00 C-4 ON D-4 0.5 E-4 OFF

A-5 -0.04 D-5 15.0 E-5*4 OFF

A-6 0.00

D-6 1.00

D-7 4.0

D-8 0.5

ST 6

F-1 XX.X

F-2 OFF

F-3 10

F-4 5

F-5 -1.0

F-6 15

F-7 4

F-8 0.0

23-1 All format (Do not change it for daily operation) Power on while pressing UP/DOWN keys till it beeps 3 times for all formats. It will clear all setting

of temperature, Overtemp, CO2 concentration, O2/N2 gas, O2 concentration, each mode setting.

All the values will be displayed as shown the “Calibration setting” table above. XX shown in the

table will be displayed 00.

23-2 CO2 span format (Do not change it for daily operation) Power on while pressing UP/SCIRENCE keys till it beeps 3 times for format CO2 Span.

Once you set CO2 span adjustment value, no need to reset even on replacement of the CO2

sensor.

SET

XXX represents that the values are unique to each individual incubator.

23. Calibration Settings

Page 41: Astec - Penguin AQ (APC/APM) Incubator User Manual - Supplied by RI UK and Ireland

41

23-3 Cleaning and Disinfections

24-3-1 Cleaning

The outside of the incubator should be kept clean through with a soft cloth, water

and soap.

24-3-2 Disinfections

The product recommended for disinfecting is the Biocidal that can be used with the

state in which we find ourselves following the instructions provided in the package.

NOTE: You must use gloves as a routine precaution.

Do not forget to turn off the incubator and disconnected the plug outlet from the

network.

Remove all shelves, media and the container of water to maintain humidity.

Rub with a soft cloth soaked with Biocidal and dry carefully all surfaces. Be careful

that no liquid comes in contact with the electrical parts.

Reassemble in reverse order before turning on the incubator.

Under no circumstances use following substances to clean The parts of stainless

steel and others may be damaged:

Sodium Nitride

Acqua Regia

Iodine

Iron Chloride

Sulphuric Acid

Page 42: Astec - Penguin AQ (APC/APM) Incubator User Manual - Supplied by RI UK and Ireland

42

Problem Check point

Cannot power on. Is the power plug surely connected to the inlet?

• Connect the plug into the inlet.

Is the inlet supplied with electric power?

• Check the breaker in your room or outside of

the building.

Is the breaker pulled out from the body?

• Put the breaker in place and power on again.

Warning! There are two fuses restorable: One is in the box on the back,

and a second on the back next the main switch. Check both.

Is the power switch on?

• Check the switch and press ON of it.

Breaker of the device

is turned off even

though power is on.

Is any liquid inside the device?

• Pull the power plug out immediately in the

case of water splashing or pouring on the body,

Contact Astec or any other distributors as soon as

possible. If you keep using it, it may result in fire.

Is the incubator placed in a rusty place?

• Rust may stick to the power line to make it

short circuit. Pull out the power plug immediately

and contact Astec or distributors. It may cause fire

if you keep using it.

Breaker on the side of

building goes down.

Is the same power line divided into many devices?

• Move each device to single lines. It may

cause fire if you continue using without taking

measure.

Is the incubator connected into outlet with other machine that needs

much power?

• Connect the line of incubator to the single

power inlet.

The temperature

inside the chamber

cannot be stabilized.

Is the fan inside the chamber moving properly?

• Check if the fan is not touched to the duct or

the sidewall of the chamber.

24. Trouble Shooting

Page 43: Astec - Penguin AQ (APC/APM) Incubator User Manual - Supplied by RI UK and Ireland

43

Is the glass window closed well?

• Close the window properly.

Inside temperature

cannot reach the set

temperature.

Is the set temperature low?

• Confirm the set temperature and set it to be

[room temp + 5℃] if it is low.

Is the room temperature too high?

• The set temperature can be effective only it is

not lower than [room temp + 5℃ ]. If the room

temperature is too high, adjust it lower.

CO2 concentration

cannot be stabilized.

Is the fan inside the chamber moving properly?

• Check if the fan is not touched to the duct or

the sidewall of the chamber.

Are shelves positioned where they should be?

• Disposition of the shelves causes bad circuit

of the air inside and CO2 concentration cannot be

stabilized. Put shelves in position. See p.10.

Is the humidifying water enough?

• Pour water into a tray immediately if there is

not enough water.

CO2 concentration

does not become the

set value.

Is the set CO2 concentration value low?

• Confirm the set value. See p. 13.

Is the rest of gas enough?

• Exchange with a new gas cylinder

immediately if there is no gas in the cylinder.

Is the regulator of CO2 gas cylinder adjusted to be 0.03 to 0.05 MPa?

• Adjust the second pressure of the regulator if

the pressure is too low. See p. 10.

Is the exit valve of the regulator full open?

• Open the exit valve fully on use.

Is the CO2 gas line tube connected to the provided CO2 gas supply?

• Confirm the tube connected to CO2 gas

supply on use.

Page 44: Astec - Penguin AQ (APC/APM) Incubator User Manual - Supplied by RI UK and Ireland

44

Is the gas line tube bended or squashed?

• Repair the bended or squashed parts on use.

Contact our distributor or us if the repair seems to

be too hard or the tube is torn.

CO2 alarm is beeping. Is the rest of gas enough?

• Exchange with a new gas cylinder

immediately if there is no gas in the cylinder.

Is the regulator of CO2 gas cylinder adjusted to be 0.03 to 0.05 MPa?

• Adjust the second pressure of the regulator if

the pressure is too low. See p. 10.

Is the exit valve of the regulator full open?

• Open the exit valve fully on use.

Is the CO2 gas line tube connected to the provided CO2 gas supply?

• Confirm the tube connected to CO2 gas

supply on use.

Is the gas line tube bended or squashed?

Repair the bended or squashed parts on use. Contact our distributor

or us if the repair seems to be too hard or the tube is torn.

ADD WATER alarm is

beeping.

The water level of the water jacket is low. Supply water from the

water supply port at the front of the body. Refer the “Supplying

water” on p. 11.

ADD WATER alarm

does not stop during

supplying water.

Is the water supplied Super pure water or Ion exchange water?

• The water level us detected by electrical

conduction. The sensor cannot response the water

with high resist value (14MΩ) such as super pure

water or ion exchange water. Supply antirust

solution attached before the water full because it

decreases resist value.

Page 45: Astec - Penguin AQ (APC/APM) Incubator User Manual - Supplied by RI UK and Ireland

45

General information This device adjusts high-pressure gas to low pressure to release the gas with regular pressured

quantity.

There is a high-pressure gas inside the regulator. Use the pressure-resistant device and inspect

it regularly to keep normal condition.

Preparation 1. The rust in regulator will cause troubles (gas leak, closure of gas path,

etc). When installing to gas cylinder, clean the cap of the cylinder

completely. On new plumbing, remove all of the rust, water, and lotion

first.

2. When connecting regulator and gas cylinder, check that the gasket is

inserted in position. If there is no gasket or the gasket is damaged, use

preparatory one.

3. On installing regulator to gas cylinder, use right sized spanner and

tighten firmly. Check gas leak after the installation.

Operation 1. Turn the pressure-adjusting handle to “CLOSE” or “SHUT” (counterclockwise) before gas

supply. Shut the valve on the exhaust to stop gas flow.

Do not turn handle to “OPEN” by force when there is no gas in the chamber or the cylinder is

not connected.

2. Open the valve of gas cylinder slowly and keep it to small amount until the gas fills with the

chamber. After the gas full, open the valve fully.

Avoid standing right in front of the pressure meter when opening or closing

valve.

3. The regulator with high lateral pressure meter indicates the pressure at intake.

4. On the use of exhaustive pressure changeable regulator, the second pressure will be

increased if the from pressure adjusting handle or screw turned to the direction of “OPEN”.

The range of adjustable pressure is 60 to 70% of the maximum value of constant and lateral

meter.

5. Open the valve of the exhaust to release gas.

25. Regulator

Page 46: Astec - Penguin AQ (APC/APM) Incubator User Manual - Supplied by RI UK and Ireland

46

Attention 1. Safety valve is preset at the safe pressure.

2. Check gas leak with soapy water or leak detective liquid. On the case of exhaust or

damage of assembly, request our distributor or us for repair.

3. Our warranty excludes the following conditions even during the warranty period.

l Damage due to fire, earthquake, thunderstorm, and other catastrophic events, and

pollution and abnormal electricity supply.

l Damage due to unduly repair, adjustment, and alteration.

l Damage due to unduly usage or handling.

l Damage due to the operation not mentioned in this manual.

Page 47: Astec - Penguin AQ (APC/APM) Incubator User Manual - Supplied by RI UK and Ireland

47

Air Filter Replacement Instruction

Please read through this instruction prior to replacing air dust filter.

♦ Air dust filters are NOT autoclavable.

♦ Sterilization with ethanol or method is not recommended since such methods do not

remove dusts trapped in the filter.

♦ Incubators must be kept clean regularly even when air dust filter is in place.

♦ Do not touch air dust filters with bare fingers/hands. Aseptic gloves should be used.

♦ Air dust filters must be handled with care since it is somewhat fragile.

♦ Air dust filters must be stored protected from high temperature, humidity and dust.

♦ Do not immerse air dust filter in any kind of liquid.

♦ Air dust filters are disposable products, not reusable.

[How to install Air dust filter]

1. Hook on the right side of the Air dust filters to the right steel

clip as shown in Figure 1.

2. Gently push up the filter until the other side of the filter clips

on the left steel clip as shown in Figure 1.

3. Make sure the filter is firmly placed and centered.

[How to remove Air dust filter]

Unclip the left side of the filter by gently pushing up the left steel

clip.

Air dust filters should be replaced at least once in every year. However, every laboratory must define and optimize the replacement interval for its own risk management according to the environment of laboratories.

Figure 1

Figure 2 After replacing Air dust filter, you may need CO2

calibration due to the slight airflow change.

Page 48: Astec - Penguin AQ (APC/APM) Incubator User Manual - Supplied by RI UK and Ireland

48

Factory Settings

ST 1 ST 2 ST 3 ST 4 ST 5 *3

A-1 0.00 B-1 C-1 D-1 90.0 E-1 0.00

A-2 0.00 B-2 C-2 0.00 D-2 0.3 E-2 0.00

A-3 1.00 B-3 C-3 0.00 D-3 20.0 E-3 ON

A-4 0.00 C-4 ON D-4 0.5 E-4 OFF

A-5 -0.04 D-5 15.0 E-5 OFF

A-6 0.00

D-6 1.00

D-7 4.0

D-8 0.5

ST 6

F-1

F-2 OFF

F-3 10

F-4 5

F-5 -1.0

F-6 35

F-7 12

F-8 0.0