Quick Guide Supra R22 / Supra R12Cat. No. : SU-R22 / SU-R12
Display & Control Panel
Pulse
PULSE
xg
VIEWSETTINGS
Start Stop
Display ScreenStatus of RPM / RCF / Time / Temperature / Rotor ID / ACC / DEC / Program
Setting SpeedPress once : Setting RPMPress twice : Setting RCF
Setting TimeUp to < 100 hr Continuous run (00:00)
Lid Open
Program Save / Call Saving a program (up to 100)
Calling up a program
Setting the Rotor ID manually When rotor id is not identified automatically, enter the rotor ID. manually
Max. RPM : 22,000 rpm
Max. RCF : 57,358 xg
Max. RPM : 12,000 rpm
Max. RCF : 22,265 xg Supra R12
Supra R22
Press the Pulse button
Enter the desired speed (See ‘Setting Run
Parameters’)
Serial No. Installation Date
Saving / Calling Up a Program
Press a parameterto be changed
Press theEnter button
Enter the desired values on the keypad
Enter the desired program number on
the keypad
Press theSave or call button
Setting Run Parameters(RPM / RCF / Time / Temperature / ACC / DEC)
Press theEnter button
Press theTemp. limit button
Press theEnter button
Temperature range that stops operation when the temperature is lower or higher than the set temperature.
Enter the desired temperature range
on the keypad
Setting Temp. Limit
View SettingsVerifying the setting parameters during centrifugation
PulseAfter reaching the set speed, the centrifugation is decelated and stopped
Setting ACC / DEC Rate ACC (0 ~ 9 steps, Max.9)
DEC (0 ~ 9 steps, Natural DEC.0)
Setting Temperature Setting temperature ( -10℃ ~ 40℃)
Setting temp. limit
2 31
5 64
8 97
0±
Connect the device to voltage
sources, which correspond to the
electrical requirements on the label
attached to the device.
Check the wall outlet is earth-
grounded, surge-protected and
can supply enough power.
Connect the power cable to the
distribution box, or install the plug
and socket that are packed with
the device.
Switch off the device after use.
Unplug the power plug before
cleaning or unused for a long
period of time.
Check the Line Voltage Grounded Outlet Power Connection Switch Off the Device
Quick Guide Supra R22 / Supra R12Cat. No. : SU-R22 / SU-R12
Maintenance
Electrical safety
A2.0-36
A10-12
A15-12
A15c-12
A50-6
A50c-6
36 x 1.5/2.0 mL / Max. 18k
12 x 10 mL / Max. 22k
12x 15 mL / Max. 22k
12 x 15 mL conical / Max. 17k
6 x 50 mL / Max. 22k
6 x 50 mLconical / Max. 17k
6 x 250 mL / Max. 12k
6 x 500 mL / Max. 10k
4 x 1,000 mL / Max. 8k
6 x 1,000 mL / Max. 7k
A250-6
A500-6
A1000-4
A1000-6
Rotors
Allow a safety clearance of 30 cm around the centrifuge during operation.
Remove Dust Empty the Condensation WaterAllow a Safety Clearance of 30 cmEmergency Lid Open
30 cm
30 cm30 cm
If the centrifuge lid cannot be opened, insert the provided T-wrench into the emergency release hole on the right side of the device and press it. (Before activating emergency open, switch off the device and wait for the rotor to stop.)
Remove any dust from the ventilation slits on the right sideof the device.
Empty the condensation water regularly by removing the drain cap on the left side of the device.
8 x 50 mL / Max. 20k
6 x 85 mL / Max. 20k
6 x 250 mL / Max. 12k
6 x 500 mL / Max. 9k
4 x 1,000 mL / Max. 8k
A50-8
A85-6
A250-6
A500-6
A1000-4
Supra R22 Supra R12
Cautions for Use
Inspect the rotor, chamber, rotor & accessories for foreign objects.
Check the Chamber & Accessories
Wipe up the moisture in the chamber with a dry cloth after use. Keep the lid open when not in use.
Clean the chamber and rotor by wiping with a cloth soaked in mild detergent and dry completely.
Keep the Chamber Dry
Cleaning
Load tubes symmetrically into the rotor with identical amount.
Load Tubes Symmetrically
Check the rotor, buckets and tubes are free of scratch and crack before operation.
Check the Accessories for Abnormalities
After relocating the device or loading the rotor, balance them by using a leveler. If necessary, level the device by adjusting the ‘Leveling foot’ on the bottom of the device.
Balancing the Centrifuge and Rotor
3.4KVA (Supra R22)
3.0KVA (Supra R12)
220~240V 50/60Hz
Rev.1