Upload
others
View
5
Download
0
Embed Size (px)
Citation preview
XEVO TQ MS PHARMACEUTICAL APPLICATION NOTEBOOK
2
The Changing Face of LC/MS: From Experts to Users .................................................................................3
Improving MS/MS Sensitivity using Xevo TQ MS with ScanWave ........................................................... 11
Simultaneous Confirmation and Quantification using Xevo TQ MS: Product Ion Confirmation (PIC) .............................................................................................................15
Novel Dual-Scan MRM Mode Mass Spectrometry for the Detection of Metabolites during Drug Quantification ...............................................................................................19
Data-Directed Detection and Confirmation of Drug Metabolites in Bioanalytical Studies ...................................................................................................................... 23
Improving Qualitative Confirmation using Xevo TQ MS with Survey Scanning .......................................27
A Novel Method for Monitoring Matrix Interferences in Biological Samples using Dual-Scan MRM Mode Mass Spectrometry ....................................................................................31
Rapid, Simple Impurity Characterization with the Xevo TQ Mass Spectrometer ..................................... 35
XEVO TQ MS: PHARMACEUTICAL AP PLICATIONS
THE CHANgINg FACE OF LC/MS: FROM EXPERTS TO USERSRobert S. Plumb Senior Applications Manager, Pharmaceutical Business Operations, Waters Corporation
Michael P. Balogh Principal Scientist, MS Technology Development, Waters Corporation
3
Researchers and practitioners from various
disciplines and sub-disciplines within chemistry,
biochemistry, and physics regularly depend on
mass spectrometric analysis. Pharmaceutical
industry workers involved in drug discovery and
development rely on the specificity, dynamic
range, and sensitivity of mass spectrometry
(MS). Particularly in drug discovery, where
compound identification and purity from
synthesis and early pharmacokinetics are
determined, MS has proved indispensable.
Biochemists expand the use of MS to protein, peptide,
and oligonucleotide analysis. Using mass spectrometers,
they monitor enzyme reactions, confirm amino acid
sequences, and identify large proteins from databases
that include samples derived from proteolytic fragments.
They also monitor protein folding, carried out by means
of hydrogen-deuterium exchange studies, and important
protein-ligand complex formation under physiological
conditions.
Clinical chemists, too, are adopting MS, replacing the
less-certain results of immunoassays for drug testing and
neonatal screening. So are food safety and environmental
researchers. They and their allied industrial counterparts
have turned to MS for some of the same reasons: PAH
and PCB analysis, water quality studies, and to measure
pesticide residues in foods. Determining oil composition,
a complex and costly prospect, fueled the development of
some of the earliest mass spectrometers and continues to
drive significant advances in the technology.
Today, the MS practitioner can choose among a range
of ionization techniques that have become robust and
trustworthy on a variety of instruments with demon-
strated capabilities.
Two decades ago, mass spectrometry was the preserve
of experts and skilled technicians: the instrumentation
required constant attention and adjustment. At this time
LC/MS was in it infancy and atmospheric pressure ioniza-
tion (API) source interfacing was just beginning. Samples
4
Advances in chromatography
Interfacing Liquid Chromatography with Mass Spectrometry
(LC/MS) allows analytical chemists access to about 80
percent of the chemical universe unreachable by Gas
Chromatography (GC); it is also responsible for the phe-
nomenal growth and interest in mass spectrometry
in recent decades.
A few individuals can be singled out for coupling LC with
MS. Beginning arguably in the 1970s, LC/MS as we know
it today reached maturation in the early 1990s. Many of
the devices and techniques we use today in practice are
drawn directly from that time.
In its simplest form, liquid chromatography relies on the
ability to predict and reproduce – with great precision –
competing interactions between analytes in solution (the
mobile or condensed phase) being passed over a bed of
packed particles (the stationary phase). The development
of columns, packed with a variety of functional moieties
in recent years, and of the solvent delivery systems, able
to precisely deliver the mobile phase, has enabled LC to
become the analytical backbone for many industries.
Continued advances in performance since then, including
development of smaller particles and greater selectivity,
also saw the meaning of the acronym change to high-
performance liquid chromatography (HPLC). In 2004,
further advances in instrumentation and column technol-
ogy achieved significant increases in resolution, speed,
and sensitivity in liquid chromatography. Columns packed
with smaller particles – 1.7 µm – and instrumentation with
specialized capabilities designed to deliver the mobile
phase at pressures up to 15,000 psi (1,000 bar) came to be
known as UltraPerformance® (UPLC®) technology. Much of
what is embodied in this current technology was predicted
by investigators such as Prof. John Knox in the 1970s.
requiring analysis were passed from the requesting scien-
tist to these “experts for analysis,” the samples would be
analyzed, processed, interpreted and the results returned
via a written report.
Two decades later, both the users and the capabilities of
LC/MS have changed significantly. Now mass spectrom-
eters and LC/MS systems are ubiquitous in the analytical
laboratory, especially in the pharmaceutical industry.
These instruments are used by a wide variety of scientists
for a diverse range of tasks, from purity screening in
medicinal chemistry, to the quantification of drugs in
blood, and the identification of proteins for biomarker
discovery.
The usability of current mass spectrometry platforms has
improved dramatically – scientists are now able to oper-
ate the systems remotely via the Internet; they can carry
out complex, data-dependent tasks such as purification
and peptide fragmentation; they are able to use to open
access systems where a non-analytical chemist can queue
samples for analysis and have the results emailed to them
without ever having to know or concern themselves about
the LC/MS process.
Recent reports put the number of LC/MS systems sold per
year in excess of 2500 units. This large number of units
sold each year is also reflected in the increased number of
users. In 1980, the number of scientists attending ASMS
was around 1250; by 2002 this had risen to greater than
4000 with a growth rate of 10 percent per year. This
growth in LC/MS users occurred because of the increase
in the number of samples analyzed each year per user,
creating larger and larger amounts of high-quality data.
More and more, this data is being turned directly into
information or knowledge so that decisions are made in
real-time. Many of these new users have little interest in
becoming expert mass spectroscopists and are instead
looking for the instrumentation itself to decide the appro-
priate experiments to be performed as well as to interpret
the data automatically and recommend a course of action
(pass/fail, pure/impure).
5
Advances in mass spectrometry
Mass spectrometers can be smaller than a coin, or they
can fill very large rooms. Although the various instrument
types serve in vastly different applications, they neverthe-
less share certain operating fundamentals. The unit of
measure has become the dalton (Da), displacing other
terms such as amu. 1 Da = 1/12 of the mass of a single atom
of the isotope of carbon-12 (12C).
Once employed strictly as qualitative devices – adjuncts
in determining compound identity – mass spectrometers
were once considered incapable of rigorous quantification.
But in more recent times, they have proven themselves
as both qualitative and quantitative instruments. A mass
spectrometer can measure the mass of a molecule only
after it converts the molecule to a gas-phase ion. To do so,
it imparts an electrical charge to molecules and converts
the resultant flux of electrically-charged ions into a pro-
portional electrical current that a data system then reads.
The data system converts the current to digital informa-
tion, displaying it as a mass spectrum.
The ions required in mass spectrometry can be created in
a number of ways suited to the target analyte in question:
by laser ablation of a compound dissolved in a matrix on a
planar surface such as by MALDI; by interaction with an
energized particle or electron such as in electron ioniza-
tion (EI); or as part of the transport process itself, as we
have come to know electrospray ionization (ESI), where
the eluent from a liquid chromatograph receives a high
voltage resulting in ions from an aerosol.
The ions are separated, detected, and measured according
to their mass-to-charge ratios (m/z). Relative ion current
(signal) is plotted versus m/z producing a mass spectrum.
Small molecules typically exhibit only a single charge: the
m/z is therefore some mass (m) over 1, with the “1” being
a proton added in the ionization process [represented
by M+H+ or M-H+ if formed by the loss of a proton], or, if
the ion is formed by loss of an electron, it is represented
as the radical cation [M+.]. Larger molecules can capture
charges in more than one location within their structure.
Small peptides typically may have two charges [M+2H+],
while very large molecules have numerous sites, allowing
simple algorithms to deduce the mass of the ion repre-
sented in the spectrum.
The general term atmospheric pressure ionization (API)
includes the most notable technique, electrospray ioniza-
tion (ESI), which itself provides the basis for various
related techniques capable of creating ions at atmospheric
pressure rather than in a vacuum. The sample is dissolved
in a polar solvent (typically less volatile than that used
with GC) and pumped through a stainless steel capillary
that carries between 500 and 4000 V. The liquid forms an
aerosol as it exits the capillary at atmospheric pressure,
and the desolvating droplets shed ions that flow into the
mass spectrometer, induced by the combined effects of
electrostatic attraction and vacuum.
The mechanism by which potential transfers from the
liquid to the analyte, creating ions, remains a topic of con-
troversy. In 1968, Malcolm Dole first proposed the charge
residue mechanism, in which he hypothesized that as a
droplet evaporates, its charge remains unchanged. The
droplet’s surface tension, ultimately unable to oppose the
repulsive forces from the imposed charge, explodes into
many smaller droplets. These Coulombic fissions occur
until droplets containing a single analyte ion remain. As
the solvent evaporates from the last droplet in the reduc-
tion series, a gas-phase ion forms. In 1976, Iribarne and
Thomson proposed a different model, the ion evaporation
mechanism, in which small droplets form by Coulombic
fission, similar to the way they form in Dole’s model. It
is possible that the two mechanisms may actually work
in concert: the charge residue mechanism dominant for
masses higher than 3000 Da while ion evaporation domi-
nant for lower masses.
The mass analyzer is the heart of the instrument and is
a means of separating or differentiating introduced ions.
Both positive and negative ions (as well as uncharged,
neutral species) form in the ion source. However, only one
polarity is recorded at a given moment.
6
The modern mass spectrometer
Modern instruments can switch polarities in milliseconds,
yielding high fidelity records. As well as separating the
ions, modern mass spectrometers can trap and fragment
ions (MS/MS or MSn) to produce a wealth of information
about the molecule’s structure. Other instruments such as
magnetic sector instruments, hybrid quadrupole time-of-
flight (Q-ToF), and ion cyclotron (ICR) mass spectrometers
can record the mass of a compound to 1 ppm, allowing for
the elemental composition of a molecular ion or fragment
ion to be deduced.
The increased sensitivity afforded by modern mass
spectrometry over other forms of detection, such as
UV and fluorescence, comes from the selectivity and
specificity of the MS and MS/MS process. During these
experiments, specific ions are allowed to pass through the
analyzer and reach the detector. During a multiple reac-
tion monitoring (MRM) MS/MS experiment, only ions that
undergo a specific fragmentation are allowed to reach the
detector; while this reduces the number of ions reaching
the detector, it all but eliminates the noise, resulting in
superior signal-to-noise ratio. This dramatically improves
assay sensitivity and specificity. The vast majority of
quantitative experiments are performed on quadrupole-
based instruments; whereas ion traps and accurate mass
instruments are preferred for structural elucidation
experiments.
Single quadrupole mass spectrometers require a clean
matrix to avoid the interference of unwanted ions, and
they exhibit very good sensitivity. Triple quadrupole, or
tandem, mass spectrometers (MS/MS) add to a single
quadrupole instrument an additional quadrupole, which
can act in various ways. One way is simply to separate and
detect the ions of interest in a complex mixture by the
ions’ unique mass-to-charge (m/z) ratio. Another way that
an additional quadrupole proves useful is when used in
conjunction with controlled fragmentation experiments.
Such experiments involve colliding ions of interest with
another molecule (typically a gas like argon). In such an
application, a precursor ion fragments into product ions,
and the MS/MS instrument identifies the compound of
interest by its unique constituent parts.
An ion trap instrument operates on principles similar to
those of a quadrupole instrument. Unlike the quadrupole
instrument, which filters streaming ions, both the ion trap
and more-capable ion cyclotron instruments store ions
in a three-dimensional space. Before saturation occurs,
the trap or cyclotron allows selected ions to be ejected,
according to their masses, for detection. A series of
experiments can be performed within the confines of the
trap, fragmenting an ion of interest to better define the
precursor by its fragments. Dynamic range is sometimes
limited in ion trap instruments and the finite volume/
capacity for ions limits the instrument’s range, especially
for samples in complex matrices.
The tandem quadrupole mass spectrometer
Tandem quadrupole and ion trap instruments have
become the workhorses of modern analytical LC/MS(MS).
The two capabilities have been incorporated into one
instrument platform to produce a linear ion-trap instru-
Figure 1. The Xevo™ TQ Mass Spectrometer.
7
ment that has all the structural characterization benefits
of ion trap mass spectrometers, with the quantitative
capabilities of tandem MS instrumentation.
These instruments have become popular with scientists
who are required to perform more than one type of exper-
iment (quantitative and qualitative) during the course of
their work and require the flexibility to perform it on the
same analytical platform. These tasks include impurity
identification and quantification and discovery DMPK,
where both dosed parent concentration and metabolite
characterization are required.
These instruments, while sounding ideal, do have some
drawbacks especially when using modern high-resolution
chromatography such as UPLC. Here, the chromato-
graphic peak widths are so narrow (1 to 2 seconds)
that there is not sufficient time for these ion trap mass
spectrometers to select the ions for trapping, fragment
the ion, and measure them to produce enough data points
to accurately define the peak. Although the collection of
MS/MS spectra can be performed with a standard tandem
quadrupole MS instrument while still correctly defining
the LC peak, sensitivity is compromised due to the low
duty cycle of the instrument.
A new direction for tandem quadrupole MS: The Xevo TQ
Along with the need to improve the utility and flex-
ibility of tandem quadrupole MS instrumentation are
the requirements to improve its data processing and
monitoring capabilities. The recent introduction of a new
iteration of the tandem quadrupole mass spectrometer,
using traveling-wave technology1, holds the potential to
resolve many of these issues. The Waters® Xevo™ TQ
Mass Spectrometer employs traveling-wave technology
that improves MS capabilities by performing simultane-
ous, multifunctional data acquisition, such as MRM and
product ion acquisition, all within a timescale compatible
with sub-2 µm UPLC. The instrument is equipped with
a modern tool-free source that simplifies the process of
routine maintenance and cleaning; instrument workflow is
also simplified with automated tuning, method generation
wizards, as well as real-time data checking functionality
that prevents sample waste if the analytical run fails for
any reason. The Xevo TQ MS’s software also features an
interactive LC/MS method database, QuanPedia™, that
ensures that the analyst selects and uses the correct
method parameters.
The T-Wave™ collision cell, originally introduced by Waters
in the Quattro Premier™ and used in the SYNAPT™ family
mass spectrometers, is employed to allow functionality
such as rapid MRM switching, fast 20-msec positive ion/
negative ion switching, and minimized crosstalk. This
functionality makes the instrument ideal for rapid method
development or use in a drug discovery environment for
development of multi-component assays.
Figure 3. MS software advancements such as QuanPedia allow users to choose pre-defined tasks for ease of operation.
Figure 2. A tool-free source ensures easy access for any user to perform rapid maintenance.
8
Xevo TQ MS: Benefits for pharmaceutical laboratories
In drug discovery, mass spectrometers often serve a dual
purpose as both a quantitative and a qualitative instrument
in DMPK departments. The Xevo TQ MS provides the
flexibility to not only perform both of these tasks, but also
to achieve them in the same analytical run or even with
the same eluting peak. This ScanWave™ functionality is
achieved by maximizing the duty cycle of the instrument.
In conventional mass spectrometers, the selected ions
enter from the first quadrupole (Q1) into the collision cell
where they are fragmented. The resulting fragmented ions
exit the collision cell and are transferred through the third
quadrupole (Q3) to the detector. This quadrupole (Q3) can
act either as a selective filter, as in MRM mode, only allow-
ing ions with a specific m/z value to pass through to the
detector, or it can scan across the entire m/z value range
providing a full spectrum.
This full-scan mode is particularly useful when performing
structural analysis; unfortunately, conventional instrumen-
tation suffers from poor duty cycle. This is because the ions
exit the collision cell simultaneously regardless of their m/z
value; as the third quadrupole scans, it can only measure or
detect one m/z value at a time. Therefore, for a scan speed
of 1000 m/z per second over a mass range of 1000 Da and
a 2-second-wide peak, the instrument will only spend 2 x
1/1000 second measuring each m/z value.
The Xevo TQ MS uses a novel collision cell design to
improve full scan sensitivity. In the last third of the colli-
sion cell, the fragmented ions are accumulated behind a
DC barrier to effect ion enrichment. These ions are then
released and contained between the DC barrier and an RF
barrier at the end of the collision cell. The RF barrier is
gradually reduced, ejecting the ions from the collision cell
to the third resolving quadrupole. These ions are ejected
according to their m/z ratio, with the heavier ions being
ejected first. To improve the duty cycle of the instrument,
the final quadrupole (Q3) is scanned in synchronization
with the ejection of the ions from the collision cell, thus
increasing the number of ions reaching the detector and
hence increasing sensitivity.
This increased scan sensitivity can be used to address
several business and scientific needs in pharmaceutical
analysis. The acquisition of this high-duty-cycle acquisition
scan can be triggered from a standard MS experiment to
provide structural information on the identity of an LC peak.
n In the field of bioanalytical analysis, the functional-
ity can be employed to confirm identity of a peak by
Product Ion Confirmation (PIC), which is carried out
within an MRM analysis. PIC works by taking one high
quality, high sensitivity spectra after the apex of a peak
and before the “touch-down” of a chromatographic peak.
This does not affect the fidelity or accuracy of the peak
quantification but allows for the acquisition of a product
ion spectra to confirm the identity of the peak.
n In the disciplines of metabolite identification and impu-
rity analysis, MS experiments such as constant neutral
loss or common fragment ion analysis are often em-
ployed to detect ions that are related to the parent API
molecule, or to look for particular metabolites that may
be toxic. Once the peaks of interest have been detected,
a second analytical run is often required to obtain
MS/MS structural information. The Xevo TQ MS can
utilize its ability to perform either constant neutral
loss or common fragment ion analysis and then rapidly
switch to high-sensitivity MS/MS to obtain structural
information. This capability removes the need for a sec-
ond analytical run, and, hence, improves productivity.
Figure 4. Daughter ion scans obtained with ScanWave, top, and without, bottom.
ScanWave DS
DS
O
O O
OO H
OH
O HO
OCH 3
OH
NH 2
C H 3
C l
C l
NH
NH
O
NH
O
O H
NH
O
NHC H 3
CH 3
CH 3
O H
OH O H
NH
O
NH
OO
NH 2
OH
O
CO O H
Fragmentation
0
1306
Daughter Ion Scan
ScanWaveDaughter Ion Scan
O
O O
OO H
OH
O HO
OCH 3
OH
NH 2
C H 3
C l
C l
NH
NH
O
NH
O
O H
NH
O
NHC H 3
CH 3
CH 3
O H
OH O H
NH
O
NH
OO
NH 2
OH
O
CO O H
O
O O
OO H
OH
O HO
OCH 3
OH
NH 2
C H 3
C l
C l
NH
NH
O
NH
O
O H
NH
O
NHC H 3
CH 3
CH 3
O H
OH O H
NH
O
NH
OO
NH 2
OH
O
CO O H
11441332
100
m/z200 400 600 800 1000 1200
%
100
%
0
144
170
725371
144
1306725
99
Recent regulatory guidelines on bioanalysis have placed
greater emphasis on the measurement of ion suppression
and control of the matrix, and measurement of drug-
related metabolites. The functionality of the Xevo TQ
MS facilitates rapid switching between matrix molecules
and analytes of interest. Phospholipids are a class of
endogenous molecules that have been associated with
ion suppression. These molecules can be monitored by
measuring the parents of the common fragment ion m/z
184, which is associated with the choline polar head
group. The Xevo TQ MS allows the simultaneous MRM
monitoring of the compound(s) of interest and precursors
of m/z 184, allowing rapid and reliable method develop-
ment. This capability can also be employed to monitor the
background ions during the course of a clinical trial, to
evaluate any differences between patients due to pheno-
type, gender, age, or diet. This information provides extra
confidence in the results and allows anomalies to
be explained.
Ease-of-use and performance extends the utility of LC/MS
The usability and functionality of mass spectrometers
have improved greatly over the last 15 years. Not only
have these instruments become more sensitive and
capable of performing multiple experiments simultane-
ously, they have also become easier to use – thus
improving instrument up-time and laboratory productiv-
ity. The advent of fast electronics and novel collision cell
designs has allowed high-sensitivity, full-scan MS/MS
experiments to be performed at the same time as high-
sensitivity MRM quantitative experiments. Concurrent
with an increase in MS capabilities has been the move
from analysis by an expert MS scientist to analysis by a
user who is tasked with answering a specific question,
such as to determine whether a product can be shipped or
to monitor food safety. Improvements in ease-of-use and
intelligent software features that help the general user be
successful in their task will continue to push LC/MS adop-
tion even wider into the general analytical community.
Figure 5. Monitoring of a model drug, alprazolam, and matrix effects of phospholipids, in a single analysis.
Acknowledgements
The authors would like to thank Paul Rainville and Marian Twohig for their scientific contributions.
Reference
1. The traveling wave device described here is similar to that described by Kirchner in U.S. Patent 5,206,506 (1993).
10
11
IM P ROV INg M S / M S S ENSIT IV I T y US INg X E V O T Q M S w IT H S C A N wAV E
Marian Twohig, Peter Alden, Gordon Fujimoto, Daniel Kenny, and Robert S. Plumb Waters Corporation, Milford, MA, U.S.
INT RODUCT ION
Tandem quadrupole mass spectrometry (MS) combined with liquid
chromatography (LC) – and, in particular, UltraPerformance LC®
(UPLC®) – has become the technology of choice for high sensitivity
quantitative analyses such as bioanalysis in the pharmaceutical
industry. The high selectivity and specificity of multiple reaction
monitoring (MRM) analysis gives rise to excellent signal-to-noise
ratios for the analysis of compounds in complex matrices. Full-scan
acquisitions are also used to provide useful information for struc-
tural elucidation in MS and MS/MS modes.
Conventional tandem quadrupole MS instruments have limited
sensitivity in full-scan mode due to poor duty cycle. The Waters®
Xevo™ TQ Mass Spectrometer with ScanWave™ functionality
delivers significant duty cycle improvements that provide enhanced
sensitivity in scanning acquisition modes.
ScanWave experiments are performed at up to 10,000 amu/sec,
making it possible to characterize narrow chromatographic peaks
better. This has become a necessity since the advent of sub-2 µm
column particle technology where narrow chromatographic peaks
can be 2 seconds wide or less.
ScanWave defined
The Xevo TQ MS employs a unique concept in collision cell technology.
Based on a novel use of Waters’ proven T-Wave™1 collision cell, the new
ScanWave™ mode of operation enhances both MS scan and product ion
data. ScanWave operation is based upon two concepts (Figure 2).
The first is that the front and back of the collision cell are indepen-
dently controlled, which allows fragmentation and accumulation
of ions to occur in the front of the gas cell while previously-
accumulated ions are simultaneously ejected from the back of the
gas cell. This provides 100 percent sampling efficiency.
Ejection of ions from the gas cell is mass dependent, although low
resolution. This low-resolution behavior allows for high space-
charge capacity without degradation of performance.
The second concept behind ScanWave is that it links the low-
resolution ion ejection from the gas cell with scanning of the
final-resolving quadrupole (MS2). This enables an intelligent ion
delivery where ions are presented to the final quadrupole when
they are actually needed, rather than continuously as in traditional
tandem quadrupole instruments.
This novel ion delivery technique provides significant duty cycle
improvements that in turn result in enhanced signal in scanning
acquisition modes. Since the scanning quadrupole (MS2) is the
device performing the mass analysis, it is not necessary to perform
a separate calibration. Scan rates, mass accuracy, and mass resolu-
tion are all identical to that for operation in traditional scanning
acquisition modes.
Figure 1. Unique T-Wave and ScanWave-enabled collision cell technology for the very best MS/MS data.
12
Figure 2. Schematic depicting a ScanWave experiment, where ions are accumulated before being sequentially ejected.
Significant increases in sensitivity using ScanWave
The data shown in Figure 3A are chromatograms for the conven-
tional product ion scan, DS, and for the enhanced product ion scan,
using ScanWave DS, produced from the UPLC/MS/MS analysis
of vancomycin, a glycopeptide antibiotic, with m/z 725 for the
[M+2H]2+ in positive ion electrospray mode. The chromatograms
have been superimposed and the vertical axes are displayed on
the same scale.
A factor of 6X signal enhancement is observed for the largest
chromatographic peak, number 5, when ScanWave DS is used.
Figure 3. Chromatogram A shows ScanWave product ion scan (ScanWave DS, green trace) versus the regular product ion scan (DS, red trace) of vancomycin, [M+2H]2+ m/z 725. In B, the ScanWave DS chromatogram is shown with the x-axis plotted in scan number.
In the conventional product ion scan mode, peaks 1, 2, 4, and 5 are
detected. When the same sample is analyzed using ScanWave DS,
the resulting signal enhancement improved the level of sensitivity
and the total number of peaks detected. In addition to the peaks that
were found in this sample using the conventional product ion scan,
spectra can be obtained for peaks 3, 6, 7, and 8.
Modern high resolution chromatography using sub-2 µm column
particles produces peaks with widths of 1 to 3 seconds at the base.
To accurately define these peaks, a high duty cycle/scan speed mass
spectrometer is required.
Figure 3B shows the same chromatogram plotted with scan number as
the x-axis. The scan speed of both the ScanWave DS and the conven-
tional product ion scan experiments was 5000 amu/sec. This allowed
more than 10 data points for the mass range 90 to 1455 amu to be
collected across chromatographic peaks which were 3 seconds wide.
Ejection Region
DC Barrier RF Barrier
Storage Region
PotentialTo Scanning Quadrupole
(MS2)
Traveling Wave
Traveling Wave
Traveling Wave
Traveling Wave
Low m/z Ion
High m/z Ion
Intermediate m/z Ion
Time2.80 3.00 3.20 3.40 3.60 3.80
%
0
100
2, 3
4
5
8
71
6
Scan550 600 650 700 750 800 850
%
0
100
ScanWave DS of 726ES+ TIC
1.25e8
2.88
3.73
A B
Waters Corporation 34 Maple Street Milford, MA 01757 U.S.A. T: 1 508 478 2000 F: 1 508 872 1990 www.waters.com
13
Figure 4 shows a mass spectrum of the largest chromatographic peak
(number 5) shown in Figure 3A. ScanWave DS of the doubly-charged
ion m/z 725 resulted in the major singly-charged fragments m/z 100,
m/z 144, and m/z 1306.
The data illustrates that the Xevo TQ MS is capable of acquiring high-
quality spectral data while operating at the high scan speeds required
to characterize narrow UPLC peaks.
Figure 4. ScanWave DS spectrum for vancomycin, [M+2H] 2+ m/z 725.
CONCLUSION
The enhanced sensitivity of the Xevo TQ MS in ScanWave mode allows
users to better characterize low-level components in their samples.
ScanWave technology allows ions to be accumulated, separated,
and ejected according to their m/z. The final quadrupole scanning is
synchronized with ion ejection from the collision cell such that the
ions of a given mass-to-charge ratio are delivered to the quadrupole
when it is ready to scan this m/z value. This results in a more efficient
instrument duty cycle and better sensitivity in scanning acquisitions.
In this application note, ScanWave technology has allowed the
peak detection of the vancomycin sample in MS/MS mode to be
significantly improved. When ScanWave DS mode was used, spectra
could be obtained for chromatographic peaks that were previously
not detected by the conventional product ion scan.
Reference
1. The traveling wave device described here is similar to that described by Kirchner in U.S. Patent 5,206,506 (1993).
100
%
0200 400 600 800 1000 1200 1400
1306
144
217
1143725m/z
ScanWave DS of 725ES+ 3.64e6
Waters, UPLC, and ACQUITY UPLC are registered trademarks of Waters Corporation. Xevo, ScanWave, and The Science of What’s Possible are trademarks of Waters Corporation. All other trademarks are the property of their respective owners.
©2008 Waters Corporation. Printed in the U.S.A.October 2008 720002828EN LB-CP
14
15
S IM U LTA N EOUS CO N F I RMAT IO N A N d QUA N T I F IC AT IO N US INg X E V O T Q M S: P RO dU C T IO N CO N F I RMAT IO N ( P IC )
Marian Twohig, Gordon Fujimoto, Joanne Mather, and Robert S. Plumb Waters Corporation, Milford, MA, U.S.
INT RODUCT ION
Tandem quadrupole mass spectrometers are used extensively in the
pharmaceutical industry for analyte quantification. This is primarily
performed by multiple reaction monitoring (MRM) as the matrices
are complex and the specificity of MRM gives the best signal-to-
noise ratios.
As well as performing quantification, these instruments are often
used for initial qualitative information, with the instrument operated
in scan mode. This information is used to confirm the identity of the
peak of interest that is being quantified.
In complex matrices, situations can arise where closely-related
compounds, e.g., metabolites or matrix interferences, can give rise
to signals even in MRM mode. This can lead to ambiguity and may
require a second qualitative experiment. Product ion confirmation
provides a means of verifying that the signal from the MRM peak is
from the compound of interest.
With conventional instrumentation, these experiments require
separate full-scan analyses. Many conventional tandem quadrupole
MS instruments are unable to perform MRM and scan experiments
simultaneously, in the timeframe of an LC peak, while maintaining
data quality. The Waters® Xevo™ TQ Mass Spectrometer is equipped
with a novel collision cell design. The collision gas is always on,
allowing both quantification (MRM) and characterization to be
performed simultaneously on the peak as it elutes from the LC or
UPLC® column while maintaining good data quality.
The new ScanWave™ mode of operation allows ions within the
collision cell to be accumulated and then separated according to
their mass-to-charge (m/z) ratio. Synchronizing the release of these
ions with the scanning of the second quadrupole mass analyzer
greatly improves duty cycle, which significantly enhances the signal
intensity of full-scan spectra for both MS and product ions.
Figure 1. Xevo TQ Mass Spectrometer with the ACQUITY UPLC® System.
EX PERIMENTAL
Product ion confirmation on Xevo TQ MS
The Xevo TQ MS can simultaneously acquire a product ion con-
firmation (PIC) scan along with an MRM chromatogram to obtain
additional information about an eluting peak. A PIC scan is enabled
in the MRM method, where a scan is used to collect either:
n MS scan
n Enhanced MS scan using ScanWave mode
n Product ion scan
n Enhanced product ion scan using ScanWave DS mode
In PIC mode, the Xevo TQ MS will switch from MRM to scan after
the apex of an LC peak as long as a minimum intensity threshold
is achieved. The trigger to start will occur after four consecutive
downward scans have been detected. If the minimum intensity
criteria is met, an MS or MS/MS spectrum is acquired using the
final resolving quadrupole (MS2) to perform the scan before
switching back to MRM mode (Figure 2). The threshold ensures
that the PIC scan is of sufficient quality to be beneficial to the user.
16
The high data collection rate of the Xevo TQ MS is such that the
area of the MRM peak can still be accurately determined, since PIC
is triggered after the peak top is detected and the definition of the
peak itself is not affected. Consequently, quantitative and qualita-
tive data are acquired simultaneously.
Figure 2. Schematic showing Product Ion Confirmation (PIC) switching after the peak top.
Figure 3 shows an example of an MRM chromatogram (3A) obtained
from the quantification of the corticosteroid fluticasone, m/z 501.
Qualitative confirmation of the peak of interest is provided by the
resulting PIC spectrum operated in ScanWave DS mode (3B).
The scan range for the PIC is selected by the software, in this
case m/z 40 to 511.
Figure 3. Chromatogram from the analysis of fluticasone, with MRM 501 > 293, and an example of the ScanWave DS PIC spectrum.
A PIC spectrum using ScanWave DS is displayed in Figure 4A. Here
it is been compared with a PIC spectrum using conventional product
ion scan (DS), 4B, and a combined spectrum (20 scans) from a
ScanWave DS of fluticasone, 4C. The spectral quality is maintained
when a PIC spectrum in ScanWave DS mode (4A) is compared to a
combined ScanWave DS spectrum (4C).
The data show that a four-fold signal enhancement was observed
when ScanWave DS mode (4A) is used to collect the PIC spectrum
compared to a conventional product ion spectrum (4B). This is due
to the more efficient duty cycle that is achieved in ScanWave mode.
This extra sensitivity available with ScanWave mode allows for high
quality spectra to be obtained even at low levels.
Switches here and acquires PIC Scan
Switches back to MRM data acquisition
MRM Trace
B
m/z40 60 80 100 120 140 160 180 200 220 240 260 280 300 320 340 360 380 400 420 440 460 480 500
%
0
100 293
205
109185155
121147
275
265
251
217235
313
501361333 389 481
Time0.20 0.40 0.60 0.80 1.00 1.20 1.40 1.60 1.80 2.00 2.20 2.40 2.60 2.80
%
0
Flu_1_9_015d 1: MRM of1 Channel ES+ 501.3 > 293.2 (Fluticasone)
4.09e7
MRM forFluticasone
PIC spectrumfrom MRM peakat Rt = 1.80 min
O
O
OH
CH3
H
CH3
F
F
H
CH3
OS
F
O
CH3
A100
17
Figure 4. Spectrum shows a comparison of a PIC spectrum for ScanWave DS, a regular product ion PIC spectrum and a combined spectrum acquired by ScanWave DS for fluticasone m/z 501 (Vertical axis linked).
CONCLUSION
The Xevo TQ MS can be used to perform quantification of fluti-
casone with simultaneous characterization of the MRM peak as it
elutes from the chromatographic system. This eliminates the need
for separate injections when qualitative confirmation of MRM peaks
is required and reduces the total analysis time in these situations.
When used routinely, product ion confirmation increases user
confidence in qualitative results from complex matrixes, and thus
reduces the need for re-analysis.
m/z50 75 100 125 150 175 200 225 250 275 300 325 350 375 400 425 450 475 500
%
0
100
%
0
100
%
0
100Flu_1_9_015d 2 (1.813)
1: Product Ions of 501 ES+501.3 > 293.2 (Fluticasone)
5.26e7
293
205
109185155135
275
251217
313
501361333 389 481
Flu_1_9_014d 2 (1.814)
1: Product Ions of 501 ES+501.3 > 293.2 (Fluticasone)
5.26e7
293
205109 155
275251
313
359333 389
Flu_1_9_017d 961 (1.796)ScanWave DS of 501ES+
5.26e7293
205
109
95155121 185
275
251217
313
501361333 389 481
PIC spectrum ScanWave DS
PIC spectrum DS
Spectrum ScanWave DS
A
B
C
Waters Corporation 34 Maple Street Milford, MA 01757 U.S.A. T: 1 508 478 2000 F: 1 508 872 1990 www.waters.com
Waters, UPLC, and ACQUITY UPLC are registered trademarks of Waters Corporation. Xevo, ScanWave, and The Science of What’s Possible are trademarks of Waters Corporation. All other trademarks are the property of their respective owners.
©2008 Waters Corporation. Printed in the U.S.A.October 2008 720002829EN LB-CP
18
19
NOV E L dUA L-S C A N M RM MO d E MA S S S P EC T ROM E T Ry FO R T H E d E T EC T IO N O F M E TA BO L IT E S dU R INg d RUg QUA N T I F IC AT IO N
Paul D. Rainville, Jose Castro-Perez, Joanne Mather, and Robert S. Plumb Waters Corporation, Milford, MA, U.S.
INT RODUCT ION
The measurement of the levels of circulating drugs and their metab-
olites is important information in the development of new therapies.
Drug levels in biofluids are used to determine the bioavailability of
a drug. Additionally, elucidation of drug metabolite information is
vital due to the fact that they can often be toxic at certain levels,
have a greater pharmacodynamic effect than the parent drug, inter-
fere with concomitant medication, and impact liver function.
These two different pieces of information are normally acquired in
separate analytical experiments, resulting in increased laboratory
workload and reduced efficiency. Therefore the ability to determine
drug concentration and obtain metabolite structural information
during a single analysis is not only faster but more cost effective.
In the case of low sample volumes, e.g., pediatric studies, this
capability is critical for laboratories to obtain required quantitative
and qualitative data.
The Waters® Xevo™ TQ Mass Spectrometer is a tandem quadrupole
system equipped with a novel collision cell design that allows full-
scan MS and quantitative multiple reaction monitoring (MRM) data
to be acquired in a single analytical run.
Here, we present a method whereby full-scan MS and MRM data
can be acquired in a single run to determine the levels of a model
pharmaceutical in urine and utilize the associated full-scan data to
determine its related metabolites.
EX PERIMENTAL
Human urine was collected from volunteer individuals eight hours
after dosing with 400 mg of ibuprofen. The samples were stored
frozen prior to analysis. Samples were prepared by centrifugation
at 13,000 RCF for 5 minutes and diluted with water. Samples were
then injected onto the UPLC®/MS/MS system.
LC conditions
LC system: Waters ACQUITY UPLC® System
Column: ACQUITY UPLC BEH C18 Column
2.1 x 50 mm, 1.7 µm
Column temp.: 40 °C
Flow rate: 600 µL/min
Mobile phase A: 0.1% NH4OH
Mobile phase B: ACN
Gradient: 5% to 95% B/2 min
20
MS conditions
MS system: Waters Xevo TQ MS
Ionization mode: ESI negative
Capillary voltage: 2000 V
Cone voltage: 15 V
Desolvation temp.: 550 °C
Desolvation gas: 1000 L/Hr
Source temp.: 150 °C
Scan range: m/z 100 to 500
Collision energies: MRM data 7 V, full-scan data 3 V
MRM transition: m/z 205 > 161
RESULTS
Determining drug concentration and drug metabolites are both
important aspects in developing a new medicine. This experiment was
designed such that the levels of ibuprofen in urine were measured by
MRM mass spectrometry and full-scan MS data was collected to detect
the associated metabolites during a single injection.
The unique collision cell design of the Xevo TQ MS, which is continu-
ously filled with collision gas, enables it to operate with rapid
switching between MS and MS/MS data acquisition modes. This
occurs in timeframe that is compatible with the fast chromatography
and narrow peaks generated by the ACQUITY UPLC System: the
Xevo TQ MS is capable of operating at up to 10,000 Da/sec and can
correctly define the very sharp peaks produced by UPLC.
In this dataset, greater than 12 scans were acquired for the MRM channel
of ibuprofen while also obtaining full scan MS data. Peaks widths were
on the order of 2.4 seconds measured at peak base (data not shown).
Figure 2 shows the chemical structure of ibuprofen and some of its
major in vivo metabolites. Figure 3 displays the MRM transition
data for ibuprofen in the urine samples and also the simultaneously
acquired full-scan data. The full-scan data were then mined for
potential metabolites resulting from ibuprofen. Figure 4 shows
extracted ion chromatograms (XIC) that were generated relating
to the ketone glucuronide (m/z 411), glucuronide (m/z 381), and
hydroxy glucuronide (m/z 397) metabolites.
Figure 2. Ibuprofen and some of its associated metabolites.
Figure 3. MRM of ibuprofen and full-MS scan data acquired from subject urine.
O
CH2 OH
CH3
CH3
CH3
O
O-Gluc
CH3
CH3
HO
O
O-Gluc
CH3
CH3
CH3
OH
O
O-Gluc
CH3
CH3
CH3
OH
O
O-G luc
CH3
CH3
CH3
1-Hydroxy IbuprofenGlucuronide
3-Hydroxy IbuprofenGlucuronide
2-Hydroxy IbuprofenGlucuronideIbuprofen Glucuronide
Ibuprofen MRM
Diluted urine patient 1Full scan
Diluted urine patient 2Full scan
21
Figure 4. XIC of ibuprofen metabolites and full-scan data.
Further confirmatory product ion MS experiments revealed
several diagnostic fragment ions, such as m/z 193 and 175 for
the glucuronide acid moieties, m/z 221 for the aglycone, and
m/z 113 for ibuprofen itself (Figure 5).1
Figure 5. Product ion spectra of ibuprofen metabolites.
A further advantage to this acquisition approach is that it provides the
scientist with the ability to visualize the differences between subject
matrix using the full-scan MS. These differences may be related to
several factors: diet, sex, age, or the state of an individual’s health.
Thus the full scan data could be additionally utilized for the detection
of biomarkers. Further, the full-scan MS data could be interrogated in
the future if new information is required about the metabolism of the
compound – without the need to re-run the samples.
KetoneXIC m/z 411
GlucuronidesXIC m/z 381
Hydroxy glucuronidesXIC m/z 397
Full scan
Ketone
Glucuronides
Hydroxy glucuronides
Waters Corporation 34 Maple Street Milford, MA 01757 U.S.A. T: 1 508 478 2000 F: 1 508 872 1990 www.waters.com
22
Waters, ACQUITY UPLC, and UPLC are registered trademarks of Waters Corporation. Xevo, ScanWave, and The Science of What’s Possible are trademarks of Waters Corporation. All other trademarks are the property of their respective owners.
©2008 Waters Corporation. Printed in the U.S.A.October 2008 720002832EN LB-CP
CONCLUSION
In this application note, we have demonstrated that the Xevo TQ MS
can acquire full-scan and MRM channel data to determine the level
of a model pharmaceutical compound in urine and its metabolite
information in a single analysis. The speed of the Xevo TQ MS
proves to be highly compatible with the high resolving power of the
ACQUITY UPLC System.
The benefits of this technique are realized in several ways. First,
the ability to gather full-scan data along with MRM channel data
enables scientists to collect multiple dimensions of information
about a sample in a single run – maximizing the resource utilization
of a laboratory that otherwise would have been performing multiple
experiments to gain the same information. Second, coupling this MS
technique with UPLC ensures a faster analysis. Finally, the richness of
the data acquired by full-scan MS allows that information to be mined
in multiple ways, giving researchers more confidence in their deci-
sions as they direct their drug discovery and development studies.
Reference
1. Plumb R, Rainville P., et al. Rapid Communications in Mass Spectrometry. 2007; 21: 4079-85.
23
dATA- d I R EC T E d d E T EC T IO N A N d CO N F I RMAT IO N O F d RUg M E TA BO L IT E S IN B IOA NA Ly T IC A L S T U dI E S
Robert S. Plumb and Paul D. Rainville Waters Corporation, Milford, MA, U.S.
INT RODUCT ION
LC/MS/MS analysis has become the analytical method of choice for
the accurate quantification of pharmaceutical compounds or active
metabolites in biological fluids. The specificity and selectivity
provided by tandem quadrupole MS in multiple reaction monitoring
(MRM) mode allows for rapid high-sensitivity analysis, often in the
pg/mL range. The data produced by LC/MS/MS analysis provides
drug concentration data that is critical to successful drug discovery
and development.
Recent U.S. FDA Guidance, “Industry Safety Testing of Drug
Metabolites,” provides recommendations to industry on when and
how to identify and characterize drug metabolites whose non-
clinical toxicity needs to be evaluated. The aim of these guidelines
is to ensure that variations in metabolic profiles across species are
both quantitatively and qualitatively measured.1
The Waters® Xevo™ TQ Mass Spectrometer is capable of operating
at acquisition speeds up to 10,000 Da/sec, which aids in the
adequate characterization of very sharp chromatographic peaks
produced by the ACQUITY UltraPerformance LC® (UPLC®) System.
The Xevo TQ MS is equipped with a novel collision cell design that
is continuously filled with collision gas, allowing rapid switching
between MS and MS/MS modes in a single analytical run.
This new collision cell is capable of enhanced high-sensitivity
operation in MS/MS mode. In this Scanwave™ mode of operation,
ions are constrained in the final third of the collision cell using both
a DC and RF barrier. These ions are then ejected from the collision
cell, in a controlled manner, from high to low m/z in synchronization
with the scanning of the final resolving quadrupole. This increases
the duty cycle of the instrument.
In this application note, we illustrate the ability of the Xevo TQ MS,
in Survey Scan mode, to detect drug metabolites “on the fly” using
a common diagnostic fragment ion.
EX PERIMENTAL
Rat plasma was spiked with ibuprofen and related major
metabolites. Samples were then precipitated using 2:1 acetonitrile
to sample (v/v). The sample was evaporated to dryness and
reconstituted in 9:1 water/methanol (v/v). The sample was then
injected onto the UPLC/MS/MS system.
LC /MS conditions
LC system: Waters ACQUITY UPLC® System
Column: ACQUITY UPLC BEH C18 Column
2.1 x 50 mm, 1.7 µm
Column temp.: 40.0 °C
Flow rate: 600 µL/min
Mobile phase A: 0.1 % NH4OH
Mobile phase B: Acetonitrile
Gradient: 5% to 95% B/2 min
Figure 1. Xevo TQ Mass Spectrometer.
24
MS system: Waters Xevo TQ MS
Ionization mode: ESI negative
Capillary voltage: 2000 V
Cone voltage: 15 V
Collision energy: 7 eV
RESULTS
The superior efficiency of the ACQUITY UPLC System produces extremely
narrow peaks, 2 seconds or less at the base. These narrow peaks require
a fast data capture rate mass spectrometer to accurately define the peak.
Figure 2 shows the MRM peak for ibuprofen using the transition m/z
205 to 161. The peak is 1.2 seconds wide at the base, and the high data
capture rate of the Xevo TQ MS allows for more than 60 scans across the
peak. This facilitates the accurate definition of the chromatographic peak,
even if several MRM transitions are employed during analysis.
Figure 2. UPLC/MS/MS of ibuprofen using the MRM transition m/z 205 to 161.
Recent FDA guidelines have recommended that, during human
clinical trials, the concentration and identity of any metabolites
with an exposure of greater than 10% of the dosed compound must
be determined. Mass spectrometry can detect and identify drug
metabolites by various means. One method is to utilize Survey
Scan mode. In this mode of operation, the MS is set to monitor a
diagnostic fragment ion from the parent drug compound.
The use of a common fragment ion requires the mass spectrometer
to scan the first quadrupole (Q1) while monitoring for a fixed m/z
with the final resolving quadrupole (Q3). Ibuprofen gives rise to
several distinctive product ions, m/z 113, 133, and 161.2 Figure 3
illustrates Xevo TQ MS operation in Survey Scan mode.
In this example, the common fragment ion of m/z 113 was
monitored by the resolving quadrupole. When a peak, containing
a m/z of 113, was detected the MS switched to collect product ion
data on the precursor ion containing the m/z 113. Peaks that exceed
a user-defined detection threshold are used to trigger the acquisi-
tion of product ion data.
Figure 4 illustrates the MS/MS spectra obtained for the peak detect-
ed at 0.66 minutes. In this example, we can see that the precursor
peak m/z value is 397. The m/z 397 produces major fragment ions
at m/z 113, 175, 193, and 221.
Figure 3. Survey Scan: precursors of m/z 113 switching to product ion scan.
Figure 4. Survey Scan ScanWave DS spectrum of peak eluting at 0.66 minutes.
Time0.25 0.50 0.75 1.00 1.25 1.50 1.75 2.00
%
0
100 0.82
Scan1820 1840 1860 1880 1900 1920 1940 1960
%
0
100 0.81
25
The m/z values and MS fragment pattern confirm the identity of
this peak as the O-glucuronide metabolite of ibuprofen.2 The data
acquired for the peak eluting with a retention time of 0.88 minutes
are shown below in Figure 5.
Figure 5. ScanWave DS of peak eluting at 0.88 minutes with a m/z value of 381.
This peak was determined to have a m/z value of 381. Resulting
fragment ions produced from the product ion MS/MS were
m/z 113, 161, 175, 193, and 205. This data confirmed that this
peak was related to ibuprofen and, with the precursor ion m/z value
of 381, was confirmed as the glucuronide conjugate of ibuprofen.2
Thus with one simple analytical experiment, along with the knowl-
edge of the fragmentation pattern of the ibuprofen, the metabolites
could be detected and the structure confirmed.
CONCLUSION
The quantification of pharmaceutical compounds in biological fluids
is a regulatory requirement as part of any new drug submission,
e.g., IND, CTX. More recently, these regulations have required that
drug metabolites with an exposure greater than 10% of the active
pharmaceutical be quantified and characterized. The Xevo TQ MS
can perform data-directed MS/MS experiments, allowing metabolite
structural confirmation using common fragment ions within a UPLC
peak timeframe.
References
1. U.S.FDA. Guidance for Industry, “Safety Testing for Metabolites.” http://www.fda.gov/CDER/GUIDANCE/6897fnl.pdf
2. Plumb R., Rainville P, et al. Rapid Communications in Mass Spectrometry. 2007; 21: 4079-85.
Waters Corporation 34 Maple Street Milford, MA 01757 U.S.A. T: 1 508 478 2000 F: 1 508 872 1990 www.waters.com
Waters, ACQUITY UltraPerformance LC, ACQUITY UPLC, and UPLC are registered trademarks of Waters Corporation. Xevo, ScanWave, and The Science of What’s Possible are trademarks of Waters Corporation. All other trademarks are the property of their respective owners.
©2008 Waters Corporation. Printed in the U.S.A.October 2008 720002833EN LB-CP
26
27
IM P ROV INg QUA L ITAT IV E CO N F I RMAT IO N US INg X E V O T Q M S w IT H SU RV E y S C A NNINg
Marian Twohig, Andrew Aubin, Michael Jones, and Robert S. Plumb Waters Corporation, Milford, MA, U.S.
INT RODUCT ION
On a conventional tandem quadrupole mass spectrometer, the
search for unknowns generally requires multiple injections: one
injection in full-scan LC/MS mode followed by a second injection
for targeted LC/MS/MS experiments. This results in increased time
required to obtain the necessary data, in addition to the time the
analyst needs to construct the MS/MS methods.
Real-time data-directed switching simplifies this experimental
approach. In data-directed mode, a full spectrum LC/MS run is
collected, with an LC/MS/MS experiment triggered if the signal in
the LC/MS survey meets preset criteria.
Modern Linear Ion Trap (LIT) mass spectrometers allow the col-
lection of MS, MRM, and MS/MS data in the same analytical run,
enabling quantitative and qualitative data to be obtained simultane-
ously. The duty cycle of these instruments when switching between
MS and MS/MS modes is typically 2 to 3 seconds. With modern
high-resolution, sub-2 µm particle chromatography, such as UPLC,®
peak widths of 2 to 3 seconds are now commonplace, thus with
these LIT MS systems this results in just 1 to 2 points across the
peak giving poorly defined peaks and possibly missed components.
The Waters® Xevo™ TQ MS is capable of scan speeds up to 10,000
amu/sec. Consequently, it is possible to employ a number of scan
functions in a single run while still maintaining good peak charac-
terization with no loss in data quality.
EX PERIMENTAL
Survey Scans on Xevo TQ MS
Survey Scans on the Xevo TQ MS allow intelligent switching of
LC/MS and LC/MS/MS data in one run, thus improving productivity.
Conventional MS or ScanWave™ MS scanning experiments can be
used to trigger MS/MS experiments in real time as the peaks are
eluting from the LC column. A more targeted screen can also be
performed using parent ion or neutral loss spectral acquisition,
to screen for compounds that have common structural features.
Conventional product ion or enhanced product ion spectra
(ScanWave) data can be generated for all the components present
in these complex samples. In ScanWave mode, duty cycle improve-
ments result in signal enhancement in scanning acquisition modes,
which facilitates the detection of low-level impurities.
An example of a survey scan for the active pharmaceutical
ingredient (API) quetiapine, an antipsychotic medication, is
shown in Figure 2, where the initial survey function is ScanWave
MS switching to ScanWave DS.
Figure 1. Xevo TQ Mass Spectrometer with ACQUITY UPLC.
28
Figure 2. Shown is an example of a Survey Scan of quetiapine (m/z 384) where the initial ScanWave MS Survey function switches to ScanWave DS mode.
In Figure 3, quetiapine (C21H25N3O2S) was analyzed in survey
scan mode. The structure-characteristic fragments of quetiapine1,2
are m/z 253 (C15H13N2S) and m/z 279 (C17H15N2S). In the above
example, a Precursor Ion Scan (m/z 253) was used to trigger the
acquisition of a ScanWave product ion scan (ScanWave DS),
generating a full product ion spectrum for the compounds
potentially related to quetiapine.
More than 20 compounds were observed to have the fragments
m/z 253 and m/z 279 as well as another signature fragment
m/z 221 (C15H13N2). Shown in Figure 4 are spectra from the
chromatographic peaks at retention times 7.86 min, 9.95 min,
10.13 min, 14.01 min, 15.94 min, and 17.52 min, respectively.
Figure 3. Survey precursor scan of m/z 253 (lower trace) switching to ScanWave DS.
Included are the API, quetiapine, and the previously-characterized1,2
quetiapine carboxylate and bis (dibenzo) piperazine, as well as three
unknown compounds that have similar fragmentation patterns.
This information was obtained from one Survey experiment without
the need for extra confirmatory MS/MS analyses. This allows the
analyst to acquire important structural information in a single run.
SN+
NH
S
N
N N
O
OH
m/z300 350 400 450 500 550 600 650
%
0
100 384
m/z100 200 300 400 500 600
%
0
100 253
158 384
ScanWave MS
ScanWave DS
m/z 384 Quetiapine
Diagnostic Fragment m/z 253
MS/MS of m/z 384
Time1.00 2.00 3.00 4.00 5.00 6.00 7.00 8.00 9.00 10.00 11.00 12.00 13.00 14.00 15.00 16.00 17.00 18.00 19.00 20.00
%
0
100
1.00 2.00 3.00 4.00 5.00 6.00 7.00 8.00 9.00 10.00 11.00 12.00 13.00 14.00 15.00 16.00 17.00 18.00 19.00 20.00
%
0
100 8.41
7.86
6.11
2.68 7.18
9.26
17.15
15.9414.01
11.0412.42 18.27
9.79
9.277.87
8.38
17.51
14.00
Precursors of m/z 253
ScanWave DS
Waters Corporation 34 Maple Street Milford, MA 01757 U.S.A. T: 1 508 478 2000 F: 1 508 872 1990 www.waters.com
29
Figure 4. UPLC/MS/MS spectra in ScanWave DS mode from selected chromatographic peaks shown in Figure 3 (top).
CONCLUSION
The Waters Xevo TQ MS, with its unique collision cell design, where
the collision gas is always on, facilitates the simultaneous acquisition
of MS and MS/MS data in one LC/MS run. Its high scan speed allows
for these experiments to be performed with sufficient points across
the peak to accurately define the narrow peaks produced by UPLC.
This capability facilitates data-directed experiments, where real-time
switching between MS and MS/MS allows more information to be
acquired from a single injection. This reduces the need for separate
experiments and accelerates the process of structural identification
and unknown compound determination.
References
1. Xu H, Wang D, Sun C, Pan Y, Zhou M. Identification of an unknown trace level impurity in bulk drug of Seroquel by high-performance liquid chromatography combined with mass spectrometry. J Pharm Biomed Anal. 2007 Jun 28; 44(2): 414-20. Epub 2007 Mar 7.
2. Bharathi Ch, Prabahar KJ, Prasad ChS, Srinivasa Rao M, Trinadhachary GN, Handa VK, Dandala R, Naidu A. Identification, isolation, synthesis and characterisation of impurities of quetiapine fumarate. Pharmazie. 2008 Jan; 63(1): 14-9.
m/z100 150 200 250 300 350 400 450 500 550 600 650 700 750
%
0
100
%
0
100
%
0
100
%
0
100
%
0
100
%
0
100
02_Sur_QT_3_10_001 2054 (7.856) 2: Auto ScanWave DS 482.46ES+ 8.93e7253
221143
279482
02_Sur_QT_3_10_001 2636 (9.958) 2: Auto ScanWave DS 383.84ES+ 8.93e7253
221158 210 279
02_Sur_QT_3_10_001 2683 (10.132) 2: Auto ScanWave DS 428.56ES+ 5.92e7253
221202
279
02_Sur_QT_3_10_001 3759 (14.007) 2: Auto ScanWave DS 456.48ES+ 7.02e7253
221 279456
02_Sur_QT_3_10_001 4297 (15.946) 2: Auto ScanWave DS 510.38ES+ 8.43e7253
221 279510
02_Sur_QT_3_10_001 4734 (17.524) 2: Auto ScanWave DS 505.53ES+ 3.88e7253
221279
505
Unknown
Quetiapine
Quetiapine Carboxylate
Unknown
Unknown
Bis (dibenzo) Piperazine
Waters and UPLC are registered trademarks of Waters Corporation. Xevo, ScanWave, and The Science of What’s Possible are trademarks of Waters Corporation. All other trademarks are the property of their respective owners.
©2008 Waters Corporation. Printed in the U.S.A.October 2008 720002838EN LB-CP
Waters Corporation 34 Maple Street Milford, MA 01757 U.S.A. T: 1 508 478 2000 F: 1 508 872 1990 www.waters.com
Waters, ACQUITY UPLC, UltraPerformance LC and UPLC are registered trademarks of Waters Corporation. XBridge and The Science of What’s Possible are trademarks of Waters Corporation. All other trademarks are the property of their respective owners.
©2007 Waters Corporation. Printed/Produced in the U.S.A.Xxxxx 2007 72000XXXXEN XX-XX
30
31
A NOV E L M E T HO d FO R MO NIT O R INg MAT R I X IN T E R F E R EN C E S IN B IO LOgIC A L SAM P L E S US INg dUA L-S C A N M RM MO d E MA S S S P EC T ROM E T Ry
Paul D. Rainville, Joanne Mather, and Robert S. Plumb Waters Corporation, Milford, MA, U.S.
INT RODUCT ION
Development of a fast, sensitive, and robust bioanalytical
LC/MS/MS assay is essential for cost-effective and compliant
processing of samples in biological fluids.
However, any bioanalytical assay can be hampered by sample
matrix effects. Components such as drug metabolites, proteins,
and phospholipids within biological matrices frequently interfere
with the robustness and sensitivity of the assay. Furthermore,
regulatory authorities now require that matrix effects be deter-
mined in these assays.
Therefore, it would be very advantageous to actively monitor and
characterize the presence of matrix components coeluting with the
compound of interest during LC/MS/MS assay method development.
Significant time and effort could be saved by ensuring that compo-
nents in the matrix are well resolved from the analyte of interest.
However, conventional tandem quadrupole mass spectrometers
cannot acquire multiple reaction monitoring (MRM) data while
acquiring full-scan data at speeds fast enough for the narrow
chromatographic peaks generated by modern separations techniques
such as UPLC.®
In this application note, we describe the ability of a novel
UPLC/MS/MS platform, the Waters® Xevo™ TQ Mass Spectrometer,
to monitor potential matrix interferences in plasma while monitoring
a pharmaceutical compound of interest. Unique to the Xevo TQ MS
platform is its ability to switch between MS and MS/MS modes in a
UPLC run that typically generates peak widths of 2 to 3 seconds.
EX PERIMENTAL
Alprazolam was spiked into rat plasma at a concentration of
10 ng/mL and then precipitated with acetonitrile using a 2:1
acetonitrile/plasma ratio. The sample was then centrifuged at
13,000 RCF for 5 minutes. The supernatant was removed and
injected onto the UPLC/MS/MS system.
LC conditions
LC system: Waters® ACQUITY UPLC® System
Column: ACQUITY UPLC BEH C18 Column
2.1 x 50 mm, 1.7 µm
Column temp.: 40 °C
Flow rate: 600 µL/min
Mobile phase A: 0.1 % NH4OH
Mobile phase B: MeOH
Gradient: 5% to 95% B/2 min
32
MS conditions
MS System: Waters Xevo TQ MS
Ionization mode: ESI positive
Capillary voltage: 1000 V
Cone voltage: 25 V
Desolvation temp.: 500 °C
Desolvation gas: 1000 L/Hr
Source temp.: 150 °C
Scan range: m/z 100 to 1000
Collision energies: High 20 V, low 3 V
MRM transition: m/z 309 > 281
RESULTS
As previously stated, the large number of components found in
matrices commonly employed in bioanalysis, such as plasma and
urine, can pose a significant problem when developing and validat-
ing a quantitative bioanalytical method. Techniques such as solid
phase extraction (SPE) and high resolution chromatography are often
employed to reduce their effects.1,2
Plasma, for instance, has many endogenous compounds that can
interfere with the pharmaceutical compound undergoing quantifica-
tion. Some of the major interferences in plasma samples that cause
ion suppression or enhancement are phospholipids, specifically
variants containing the choline head group. Consequently, scientists
developing bioanalytical methods often will monitor phospholipids
as they can be a major source of matrix effects.
Parent or precursor ion scanning of the indicative choline fragment
ion (m/z 184, in positive ion mode) is commonly used to monitor
phospholipids (Figure 2).
Figure 2. Fragmentation pattern for a phospholipid containing the choline head group.
CH3
CH3
CH3
N+
O
O
O-
P
O
O
O
O
O
R R'
184
CH3
CH3
CH3
N+
O
O
O-
P
O
O
O
O
O
R R'
33
Figure 3 shows the MRM channel for alprazolam above a scan result
for precursor m/z 184. In this example, we can see that UPLC meth-
odology facilitates excellent resolution of the analyte of interest
from the choline-containing phospholipids.
Figure 3 further illustrates the differences between the two scans
of the two lots of plasma, indicating that each contains different
choline-containing phospholipids. This variance in plasma lot A and
lot B exemplifies the reason six different lots of matrix are required
to be analyzed during the bioanalytical method validation process.3
Figure 3. Monitoring of the model drug alprazolam and matrix effects contributed by choline-containing phospholipids in a single analysis.
Recognizing that there may be other compounds in the matrix that
could potentially interfere, and thus become a source of matrix
effects, a full MS scan was acquired from the same injection as
the MRM channel (Figure 4). Acquiring full-scan data gives the
bioanalytical scientist the tools to observe more potential matrix
interferences in the samples.
Figure 4. Simultaneous monitoring of the model drug alprazolam and matrix in the m/z 100 to 1000 range in a single analysis.
Figure 4 indicates the presence of other coeluting analytes from
the matrix. The coeluting analytes have the potential to introduce
matrix effects to the analysis of alprazolam, and may impact the
quantification of the analyte of interest. The scientist is now in a
position to adjust the conditions of the assay method, taking into
account the interfering components.
The unique capability of the Xevo TQ MS to rapidly switch between
MS and MS/MS modes facilitates simultaneous monitoring of
coeluting components as well measuring the compound of interest,
even when UPLC peaks are typically 2 to 3 seconds wide.
MRM alprazolamplasma lot A
MRM alprazolamplasma lot B
Parents of m/z 184plasma lot A
Parents of m/z 184plasma lot B
MRM alprazolamplasma lot A
MRM alprazolamplasma lot B
MS scan m/z 100 to 1000plasma lot A
MS scan m/z 100 to 1000plasma lot B
Waters Corporation 34 Maple Street Milford, MA 01757 U.S.A. T: 1 508 478 2000 F: 1 508 872 1990 www.waters.com
34
CONCLUSION
In this application note, we have shown the novel ability of the
Waters Xevo TQ MS coupled with an ACQUITY UPLC System to
acquire quality full-scan and MRM data in a single analysis. Using
this technique, we monitored potential matrix interferences present
in protein-precipitated plasma while monitoring the MRM transition
for a model pharmaceutical.
Interferences due to coelution of matrix components can thus be
detected “on the fly” in early method development, which reduces:
n Method development time
n Method variability
n Problematic occurrences during validation
Thus utilizing the Xevo TQ MS enables researchers to increase the
quality of the final MS/MS method while developing a fully-detailed
scan record should reviewing the data be of interest.
References
1. Zhang S, Chen G. Journal of Chromatographic Science. 2008; 46: 220-224.
2. Chambers E, Diehl D, Lu Z, Mazzeo J. Journal of Chromatography B. 2007; 852 (1-2): 22-34.
3. Guidance for Industry, Bioanalytical Method Validation. U.S. FDA. 2001.
Waters, UPLC, and ACQUITY UPLC are registered trademarks of Waters Corporation. Xevo and The Science of What’s Possible are trademarks of Waters Corporation. All other trademarks are the property of their respective owners.
©2008 Waters Corporation. Printed in the U.S.A.October 2008 720002830EN LB-CP
35
R A P I d, S IM P L E IM P U RIT y C HA R AC T E R Iz AT IO N w IT H T H E X E V O T Q MA S S S P EC T ROM E T E R
Robert S. Plumb, Michael D. Jones, and Marian Twohig Waters Corporation, Milford, MA, U.S.
INT RODUCT ION
The detection and characterization of impurities and degradation
products of an active pharmaceutical ingredient (API) are regulatory
filing requirements. The detection and identification of impurities
not only ensures medicine safety but can also be used as a finger-
print for patent protection and counterfeit drug analysis.
Impurity characterization and identification are normally carried
out using information-rich analytical techniques such as NMR and
LC/MS. Analysis by LC/MS provides parent ion mass from full-scan
MS and structural information from the fragments generated in
MS/MS experiments. With traditional tandem quadrupole instrumen-
tation, the generation of this data requires multiple experiments to
obtain MS and MS/MS information.
Modern Linear Ion Trap (LIT) mass spectrometers allow the collection
of MS, multiple reaction monitoring (MRM), and MS/MS data in the
same analytical run, allowing quantitative and qualitative data to
be obtained simultaneously. However, the duty cycle of these instru-
ments when switching between MS and MS/MS modes is typically
2 to 3 seconds. With modern high-resolution, sub-2 µm column par-
ticle chromatography such as UPLC,® peak widths of 2 to 3 seconds
are now commonplace. With these LIT MS systems, this would result in
just 1 to 2 points across the peak, with the peaks either poorly defined
or missed completely; thus slower, lower-resolution LC systems must
be used, resulting in reduced throughput and lower data quality.
The Waters® Xevo™ TQ Mass Spectrometer is equipped with a novel
collision cell design that is continuously filled with collision gas,
allowing rapid switching between MS and MS/MS modes. The Xevo TQ
MS is capable of operating at up to 10,000 Da/sec and can correctly
define the very sharp peaks produced by UPLC, with more than 10
points across a 2-second-wide peak, even on a multi-scan experiment.
This new collision cell is capable of enhanced high-sensitivity
operation in MS/MS mode. In this mode of operation, ions are
constrained in the final third section of the collision cell using both
DC and RF barriers. These ions are then ejected from the collision
cell, in a controlled manner, from high to low m/z in synchronization
with the scanning of the final resolving quadrupole. This increases
the duty cycle of the instrument, resulting in enhanced sensitivity
that is ideal for the detection and characterization of low-concentra-
tion impurities that may result in toxic effects.
EX PERIMENTAL
To evaluate the performance of this system, the impurities of the
common pharmaceutical drug quetiapine, used to treat biopolar
disorder, was investigated using UPLC/MS/MS.
LC /MS conditions
LC system: Waters ACQUITY UPLC® System
Column: ACQUITY UPLC BEH C18 Column
2.1 x 50 mm, 1.7 µm
Column temp.: 65 °C
Flow rate: 800 µL/min
Mobile phase A: 20 mM Ammonium bicarbonate pH 10
Mobile phase B: Acetonitrile
Gradient: 15% to 95% B/18 min
36
MS system: Waters Xevo TQ MS
Ionization mode: ESI positive
Capillary voltage: 30 V
Collision energy: 15 eV
RESULTS
The unique collision cell design allows the Xevo TQ MS to be oper-
ated in several different modes of operation: full scan MS, MRM,
as well as MS/MS mode. As the collision cell is continuously filled
with collision gas, the instrument can rapidly switch between MS
and MS/MS in the same analytical run. This allows MRM and MS
scans to be performed in the same run. Combined with the high scan
rate, this allows for rapid survey scans to be performed, such as MS
neutral loss or parent ion, before switching to MS/MS.
This high data-capture rate allows for the accurate definition of the
peak, even with the very narrow peaks produced by UPLC. Figure 2
shows the UPLC/MS chromatogram produced in the analysis of an
API batch of quetiapine at a concentration of 1 µg/mL. Here we can
see that impurity peaks are 2 to 4 seconds wide at the base. The
data shown in Figure 3 illustrates the number of scans achieved in
MS and MS/MS modes.
Figure 2. UPLC/MS analysis of quetiapine at 1 µg/mL.
Maximizing LC peak definition
In this example, the Xevo TQ MS was operated in ScanWave MS
mode, switching to ScanWave DS (daughter ion scan) mode when a
peak was detected above a user-defined threshold. In this mode of
operation, the instrument selects the most intense peak in the MS
spectrum and acquires MS/MS data on this peak before returning to
MS mode. Since the collision cell is continuous filled with collision
gas, there is a no delay in switching between MS and MS/MS modes.
Figure 3. Rapid data collection is performed simultaneously in both MS and MS/MS modes.
We can see from this data that the instrument has acquired 9 points
across the peak in MS mode, and 15 points across the peak in
MS/MS mode – despite the fact that the peak is only 2 seconds wide
at the base. This high data-capture rate enables the Xevo TQ MS to
perform high quality, data-dependent MS-to-MS/MS experiments in
a UPLC timeframe with sufficient data points to accurately define
the peak. This dual mode of operation can also be used to acquire
full-scan MS data simultaneously with MRM data, or to detect a peak
with precursor ion scanning before switching to MS/MS mode.
Time5.00 10.00 15.00 20.00
%
0
1009.81
9.240.29
8.375.65
17.26
10.90
17.10
15.79
13.88
17.50
17.66
18.07
19.38
Scan2870 2880 2890 2900 2910
%0
100 17.58
17.49
340 360 380 400 420 440 460 4800
%
100
Scan15.61
16.6317.23
17.78
17.57
MS MS/MS
37
Precursor ion scanning
The detection of new impurities, degradation products, or, in a DMPK
study, drug metabolites, is often confounded by the signal from the
matrix. To detect and visualize these analytes, the analytical chem-
ist can use the specificity of the mass spectrometer.
Since compounds can undergo fragmentation as a result of the
degradation or metabolism process, the use of simple, predicted MRM
transitions for common degradation/metabolism pathways may result in
the non-detection of a potentially toxic impurity, degradation product,
or metabolite. A more comprehensive way to detect these compounds
is to monitor for the common fragment ions of the molecule of interest.
The Xevo TQ MS’s Survey Scan functionality utilizes the fast data-
capture rate of the instrument to facilitate the collection of precursor
ion data as well an MS/MS spectrum of the peaks detected.
This functionality was used to evaluate a commercially-purchased API
sample of quetiapine. The MS/MS spectra of quetiapine revealed that
it gave rise to three major product ions having m/z values 221, 253,
and 279. This data was used to detect drug-related impurities in the
API batch by performing a Survey Scan analysis on each of these ions.
Figure 4. Survey Scan UPLC/MS analysis of quetiapine for ion m/z 279 in positive ion mode.
The data collected for the parent ion chromatogram of m/z 279 is
displayed in Figure 4. Here, we can see the presence of seven major
peaks, six impurities and the quetiapine active peak, eluting with a
retention time of 9.86 minutes. A similar analysis using the com-
mon fragment ion m/z 221 to trigger data collection produced the
chromatogram shown in Figure 5. With this fragment ion, a total of
12 peaks were detected and MS/MS data acquired.
Figure 5. Survey Scan UPLC/MS analysis of quetiapine for ion m/z 221 in positive ion mode.
The MS/MS spectra obtained from the peak eluting with a retention time
of 5.6 minutes is displayed in Figure 6. This impurity has a m/z value of
400 amu and has been identified as the S-Oxide impurity of quetiapine.
Figure 6. AutoScanWave MS/MS spectrum of S-Oxide of quetiapine eluting at 5.6 minutes.
Time2.50 5.00 7.50 10.00 12.50 15.00 17.50
%
0
100 9.86
9.29
5.69
8.42
6.46
10.94
15.1017.63
17.91
Time2.50 5.00 7.50 10.00 12.50 15.00 17.50 20.00
%
0
100 8.398.38
5.64
8.40
9.25
17.15
9.84
9.89
9.90
10.12
13.9910.72
12.2817.25
18.28
m/z100 150 200 250 300 350 400
%
0
100 221
221
196
176
158
211247
222
239
279
253
38
ScanWave technology
The detection of low-level impurities is becoming increasingly
important, especially when monitoring potential genotoxins.
Collection of the MS/MS spectrum from Survey Scan experiments,
either precursor ion or common neutral loss, can be performed in
two modes of operation: standard MS/MS or ScanWave MS/MS.
As described previously, ScanWave technology allows for
increased sensitivity in the collection of MS/MS data. This increase
in sensitivity is illustrated by the MS spectra obtained for the
desthanol impurity of quetiapine (Figure 7). The top spectrum is
obtained in standard MS/MS mode, while the lower spectrum is
obtained in ScanWave MS/MS mode.
In this example, we can see that the ScanWave MS/MS data is
13 times more sensitive than that in standard MS/MS mode. This
increase is essential for the correct confirmation or identification
of low-level impurities.
Figure 7. Comparison of standard and ScanWave MS/MS sensitivity.
CONCLUSION
n The Xevo TQ MS provides unrivaled levels of sensitivity
and functionality.
n The high data-capture rates of the instrument, and its unique
collision cell design and ScanWave technology, allows the
maximum amount of data to be collected in one analytical run.
n This reduces the number of experiments needed to make a deci-
sion, allowing impurities to be detected and identified quicker,
and making maximum use of instrumentation.
n The rapid switching between MS and MS/MS possible with the
Xevo TQ MS allows the collection of qualitative data and quanti-
tative data in the same analytical run.
n The instrument’s high data-capture rate ensures that, even with
the narrow peaks of 2 to 3 seconds produced by today’s modern
sub-2 µm particle LC systems, sufficient points can be collected
across for accurate quantification.
n The use of ScanWave technology ensures that even the lowest-
level peaks are detected and MS/MS spectra acquired, ensuring
comprehensive impurity detection.
Waters, ACQUITY UPLC, and UPLC are registered trademarks of Waters Corporation. Xevo, ScanWave, and The Science of What’s Possible are trademarks of Waters Corporation. All other trademarks are the property of their respective owners.
©2008 Waters Corporation. Printed in the U.S.A.October 2008 720002831EN LB-CP
m/z100 125 150 175 200 225 250 275 300 325 350
%
0
100
%
0
100
2: Auto Daughters 340.72ES+ 9.07e5279
221
161177
253
2: Auto ScanWave DS 340.39ES+ 1.21e7221
210206168
253
247228 279
Waters Corporation 34 Maple Street Milford, MA 01757 U.S.A. T: 1 508 478 2000 F: 1 508 872 1990 www.waters.com
PUBS:
AAPSAir, Water & Environment InternationalEuropean BioPharmaceutical ReviewEuropean Pharmaceutical ReviewDrug Discovery WorldRapid Communications in Mass SpectrometryTouch BriefingsWell Characterized BioPharmaceuticalFood Safety EuropeNext Generation PharmaceuticalSpectroscopy Asia
Job # & CC: WATR9752_B
Job Name: WATR9752_B_XevoNew_DBmec.indd Date: 06-13-08Live: 7”x10” Smallest Trim: 8.25”x10.875” Largest Trim: 8.5”x11.75” Bleed: 8.875”x12 Page: 1 Rev: 1Stage: mech Release: 06-13-08 DB AD: MM CW: GK PM: DG Scale: 100%Color: 4c Gutter: — Other: ST: DB
APP
RO
VALS C-LO
MECH PROOF
MK: ______________________ PM: ______________________ AB: ______________________
AD: ______________________ CW: ______________________ RM:______________________
QC: ______________________ CD: ______________________
Initials / Date Initials / Date Initials / Date
Client / Date: ________________________________________________
©2008 Waters Corporation. Waters, The Science of What's Possible, XEVO, IntelliStart, and ACQUITY UPLC are trademarks of Waters Corporation.
[ REACTION ]
Introducing XEVOTM TQ MS, the new tandem quadrupole mass spectrometer from Waters, featuring IntelliStartTM technology for simplified method development. This automated system solution delivers powerful quantitative capability with very low limits of detection —to the widest range of scientists. All this, along with ACQUITY UPLC® compatibility by design, means you get enhanced performance with dramatically improved speed, productivity, and efficiency. Run more samples per day and more compounds in a single analysis. And make critical, time-sensitive decisions with confidence. Find your answer at www.waters.com/xevo
WHEN ANALYZING COMPLEXSAMPLES WITH XEVO TQ MS,WILL YOU BE MORE IMPRESSEDWITH HOW POWERFUL IT IS?
OR HOW ACCESSIBLE IT IS?
Waters Corporation 34 Maple Street Milford, MA 01757 U.S.A. T: 1 508 478 2000 F: 1 508 872 1990 www.waters.com
Waters, ACQUITY UltraPerformance LC, ACQUITY UPLC, UltraPerformance LC, and UPLC are registered trademarks of Waters Corporation. Xevo, Scan-Wave, T-Wave, SYNAPT, Quattro Premier, and The Science of What’s Possible are trademarks of Waters Corporation. All other trademarks are the property of their respective owners.
©2008 Waters Corporation. Printed in the U.S.A.November 2008 720002826EN LB-CP
www.waters.com/xevo