Upload
colman
View
26
Download
0
Tags:
Embed Size (px)
DESCRIPTION
Experiences of using the Philips CardioMD Camera University Hospital Birmingham Foundation NHS Trust. Why choose a dedicated Cardiac camera?. Three cameras were routinely used before the CardioMD was installed: Dual Head (ADAC Vertex) used for MPI - PowerPoint PPT Presentation
Citation preview
Experiences of using the Philips CardioMD Camera University Hospital Birmingham Foundation NHS Trust
Why choose a dedicated Cardiac camera?
Three cameras were routinely used before the CardioMD was installed:
Dual Head (ADAC Vertex) used for MPISingle Head Large FOV (Toshiba)Single Head Small FOV (Siemens)
Plan was to replace the small FOV camera with a dedicated Cardiac Camera because:
MPI requests were growing A dedicated cardiac camera has a smaller
footprint than a modern SPECT-CT system Money was available from the PCT for a
dedicated Cardiac Camera (£190k)!
Why Choose the CardioMD?
Three dedicated Cardiac cameras were considered: Philips CardioMD (AC available) Siemens C-cam (AC available in
future) Pulse IS2 (AC available in future)
Specifications similar between systems.
The Philips CardioMD was chosen primarily due to the availability of AC. The department had performed AC on MPI scans since 1996.
The CardioMD was also the cheapest!
Changes to Service in Department
Change in workflow for each camera Dual Head (ADAC Vertex)
Free for work other than MPI Single Head Large FOV (Toshiba)
Used for studies performed on the old small FOV camera replaced with CardioMD
CardioMD (Philips) Used for MPI and MUGA studies
The CardioMD has doubled the MPI scanning capacity and a full 5 day service can now be provided. (~550 ~1100patients/Yr)
Setting up a Study
Easy and quick to set-up an acquisition on the system.
However, since the introduction of the DICOM work list, there have been a few problems. Repetition of patient
files. Handling of attenuation
correction data
Positioning the Patient
Patients have difficulty both in getting on and off the bed.
Staff complain about bed not moving up or down.
Positioning has to be precise as the bed translates by a small amount.
Easy to set-up once patient is in position
However, range limits on very small or very large patients.
Image Reconstruction/Processing
Fast reconstruction/ processing software.
Motion correction software is quick and effective.
Software could be more intuitive. Some technologists report that they have to read manual/protocol on some occasions!
Routine QC (57Co Daily Flood)
Routine QC (99mTc Intrinsic Flood)
30M counts acquired using ~30MBq 99mTc (~15min/head)
Integral uniformity results calculated and plotted
Weekly Intrinsic Uniformity Results
Intrinsic Integral Uniformity
0.0
0.5
1.0
1.5
2.0
2.5
3.0
3.5
4.0
4.5
5.0
5.5
6.0
6.5
7.0
7.5
17-Jul-05 16-Jan-06 18-Jul-06 17-Jan-07 19-Jul-07
Date
Inte
gra
l Un
iform
ity (%
)
Detector 1 Uniformity (%)Detector 2 Uniformity (%)Series8Series9Less than 3% - excellent
3%-4% - good4%-6% - acceptable
>6% - needs calibration
PMT's Retuned on: 13-Jul 05
30-Aug-05 31-May-0614-Nov-06 12-Dec-06 09-Jan-07 08-May-07
Routine QC (COR)COR Check : X & Y Ranges
0.0
0.5
1.0
1.5
2.0
2.5
3.0
3.5
4.0
4.5
5.0
28-May-05 26-Nov-05 27-May-06 25-Nov-06 26-May-07Date
Ra
ng
e (
mm
)
X range (mm)
Y range (mm)
Must be below red line
Easy to set-up. X range values considered to
be satisfactory. Y range values considered to
be good.
Summary Pros
Camera easy to use and reliable (since installation of UPS)
System is relatively easy to maintain (e.g. PM tube retune)
Quick and easy to use acquisition and reconstruction software
Good motion correction software Cons
Most significant problem with the system is regarding ergonomics of the bed.
Several problems with acquiring data following introduction of the DICOM work list.
Set-up would be easier if an ‘OK’ button was present on the handset.
Manual collimator exchange can be awkward