11
1 VA1908i VA1907i VA1905i TP200 Rev. A © 2014 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 | Applies to Models: VT Special Tools: none Language of origin: English 003-2776-00 Rev. A (1/5/15) Attention: Midmark recommends two installers and using a licensed professional to install required plumbing. Wet or Dry Treatment Table Installation Wall Mount (page 5) Recessed Mount (page 7) Note: This install guide covers multiple mounting options. Refer to appropriate pages for installation steps. Flush Mount (page 4) Chase Mount (page 6) Free Standing Mount (page 3) VA1929i VA1906i

Wet or Dry Treatment Table Installation

  • Upload
    others

  • View
    3

  • Download
    0

Embed Size (px)

Citation preview

1

VA1908iVA1907i

VA1905i

TP20

0 R

ev. A

© 2014 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |

Applies to Models:VT

Special Tools:none Language of origin: English

003-2776-00 Rev. A (1/5/15)

Attention:Midmark recommends two installers and using a licensed professional to install required plumbing.

Wet or Dry Treatment Table Installation

Wall Mount (page 5)

Recessed Mount (page 7)

Note:This install guide covers multiple mounting options. Refer to appropriate pages for installation steps.

Flush Mount (page 4)

Chase Mount (page 6)

Free Standing Mount (page 3)

VA1929i VA1906i

English-2

TP20

0 R

ev. A

© 2014 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |

Table Dimensions

StandardWet Table Dimensions Dry Table Dimensions

VA1903

ShallowWet Table Dimensions

English-3

TP20

0 R

ev. A

© 2014 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |

VA1928i

Free Standing Mount Installation

Equipment Alert Midmark requires anchoring table to the floor when using in the middle of office.

Note:Refer to later page for Table Plumbing Installations.

Step 1: Remove table from skidA) Pull tabs on drawer outward to release from slides.B) Drill four holes into floor through mounting holes in bottom of table. Use appropriate drill for floor materials. C) Prior to installing mounting fasteners, table may need to be leveled. Shims may be required. D) Install mounting hardware at four locations.

Dry Island

Shallow Wet Table

English-4

TP20

0 R

ev. A

© 2014 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |

VA1904i

Flush Mount Installation

Step 2: Secure Table to BracketsA) Open door on table and secure to corner brackets using hex screws provided.

Step 1: Install BracketA) Mount corner brackets to base unit using hex screws provided. Note: Door not shown on cabinet for clarity.

Dry Table

Note:Refer to later page for Table Plumbing Installations.

English-5

TP20

0 R

ev. A

© 2014 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |

VA1909i

Wall Mount Installation

Step 1: Mount to wall with studsA) Find center of each stud to be used for mounting. B) Slide table into position.C) Secure to wall using washers and screws being sure to hit studs.

Step 2: Mount to wall without studsA) Slide table into position and mark hole locations on wall.B) Slide table away from wall and drill holes using 3/16” bit.C) Insert anchors into wall. D) Slide cabinet back to wall and tighten screws and washer to anchors.

Note:Refer to later page for Table Plumbing Installations.

3/16” Drill Bit

English-6

TP20

0 R

ev. A

© 2014 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |

Chase Mount Installation

Note:Refer to chase installation manual for mounting hole locations.

Rectangular Chase

Pentagonal Chase

Note:Refer to later page for Table Plumbing Installations.

Step 1: Secure Table to ChaseA) Secure Table to Chase using mounting hardware provided with the Chase (#10 screws, washers and nuts).

Note: Door not shown on table for clarity.

English-7

TP20

0 R

ev. A

© 2014 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |

VA1922i

Recessed Mount Installation

7/32” Drill Bit

Step 1: Drill for BracketA) Use dimensions shown to mark top and bottom hole locations on back panel of table. B) Drill 7/32” holes in marked locations.

Back Panel

StandardWet Table Dimensions

ShallowWet Table Dimensions

English-8

TP20

0 R

ev. A

© 2014 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |

VA1910i

Recessed Mount Installation (cont’d)

Note:Verify plumbing connections will reach proper location before performing Step 3.

Step 2: Install BracketsA) Secure brackets to base units using hex screws provided.

Note: Door not shown on cabinet for clarity.

Note:Refer to later page for Table Plumbing Installations.

Step 3: Secure Table to BracketA) Move table into mounting position.B) Install table to brackets through door opening.

Note: Install table bracket to cabinet bracket on door/drawer side. Install through drilled holes on back panel to cabinet bracket.

English-9

TP20

0 R

ev. A

© 2014 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |

VA1911i

1 1/2" (max.)(3.8 cm)

LOWERTABLE TABLE

RAISE

Step 1: Level Table (if necessary)A) Pull tabs on drawer outward to release from slides.B) Level Table as shown.C) Reinstall drawer

Step 2: Install Faucet and SprayerA) Follow the manufacturer’s instructions to install faucet and sprayer.B) If sprayer is not applicable, use hole cap to plug the hole.

Step 3: Install Drain, Tail Pipe, and Hair Trap

Note: The debris trap is not intended to replace a P-trap. The debris trap is a plumbing device that has a removable strainer intended for installation between the tub drain and before the standard P-trap.

Plumbing InstallationDrain

P-TrapNOT Supplied

English-10

TP20

0 R

ev. A

© 2014 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |

VA1913i

Dental Suite Installation

VetPro™Solo

Step 3: Install VetPro™ Solo unit or Junction Box

Note: Refer to Solo Installation Manual 003-1996-00, or Junction Box installation manual 003-1914-00.

Junction Box

Step 1: Install Wet Peninsula Dental Table

Step 2: Install VetPro™ 5000 Table Mount

Note: Refer to VetPro™ 5000 Install Manual 003-2000-00.

VA1912i

English-11

TP20

0 R

ev. A

© 2014 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |

VA1915i

Wet Table Grate Installation

Grate

Step 1: To Install GrateA) Place grate on top of grooves in table.