STLCC_CPLS;Morrison 4/7/2014 Page 1
Hot Plates, Block Heaters,
Plate Incubator,
Water Baths, Bead Bath
Standard Operating Procedures
Prepared by: Bob Morrison
STLCC, Instrumentation Specialist
Mar08, Last revision Mar 11
Typical Digi-Block Jr.
@FV
Precision Analog
Water Bath@FV
Hot Plate &
Stirrer@FV
Precision Digital
Shaker @FV
BioRad-IPS Microplate
Heater/Incubator@FV Bead Bath
@ FV
Block Heater
@BRDG, Digital
VWR 18LShaker
Digital @ BRDG
VWR 18L Analog
Non-shaker
@ BRDG
Heater; Hot Plate, Stirrer
STLCC_CPLS;Morrison 4/7/2014 Page 2
Corning PC-420
Temp. Range: 25-550°C
Stir Range: 60-1100
Size: 5" x 7" Pyroceram© top
Corning PC-351
6"x7.5" ceramic top
• Stirring range: 250 - 1000 rpm
• Temperature range: 25 - 520°C
Corning PC-35
Temp. Range: 25-500°C
Size: 5" x 7”
Analog Stir on/off and
rpm Control Analog On/off and Heat
Control
“On” indicator lights
Heater: Dry Block, Digi-block Jr.
STLCC_CPLS;Morrison 4/7/2014 Page 3
On/Off
Use Up/Down Arrows
to set desired temp. Actual Block temperature
from probe
Probe: Insert in small
diameter hole in block
to monitor actual
temperature, displayed
above set value.
Error Conditions:
Er 2 – probe not working
Er 7 – no probe connected
Recommended: Insert
100C thermometer in
one hole/cell to
validate temperature
and probe display.
STLCC_CPLS;Morrison 4/7/2014 Page 4
Dry Block Incubator/Heater
Low and High Temperature Control Settings
(data sources: tech specs corrected by measured FVCC data (Tolerance: +5C, -10C)
0
20
40
60
80
100
120
140
160
Low Temp High Temp
Low Temp 22 22 22 22 22 25 30 35 40 50 60
High Temp 50 55 60 70 80 90 100 110 120 130 135
0 1 2 3 4 5 6 7 8 9 10
Low
Temp
Setting
High
Temp
Setting
On/Off
Still
Heating if
on
Recommended Procedure:
- Use the chart below to get initial settings for either the low or high dials
Note: they are not additive, use only one dial with the other set a “0”
- Add a thermometer to an open well hole and monitor the temperature
- Well should reach desired temperature in <10 min. for most models
- Adjust setting dial as needed to get closer to desired temperature
Heater: Dry Block, Digital, BRDG, Labnet, Dual block
STLCC_CPLS;Morrison 4/7/2014 Page 5
Product Specifications , Model D1200 (dual block) Temperature range Ambient +5º to 150ºC
Temp. resolution 0.1ºC
Temp. uniformity ±0.2ºC
Temp. accuracy ±0.3ºC
Block chamber Stainless steel
Block capacity
Single block unit One standard block
Dual block unit Two standard blocks or one dual block
Dimensions (WxDxH) 7.9 x 10.4 x 3.3 in./20 x 26.5 x 8.3 cm
Weight 5.5 lb/2.2 kg
Electrical 120V~, 50/60 Hz (also available in 230V)
Dry Bath Blocks
Construction High grade, nonporous aluminum with anodized
surface
1) Use Up/Down arrows to set desired temperature
2) Press Start/Stop to begin heating cycle, note green light
3) Press Start/Stop to turn off heating, light now off
4) To change temperature settings, Press
START/STOP once, use Up/Down arrows to enter
the new set point, then press START to resume
the new heating cycle.
5) Remain cautious of block after heat is turned off. 6) Toggle to OFF using switch in rear
On/Off toggle in Rear.
(4) Blocks for 24 x 0.2ml PCR Tubes
(4) Blocks for 24 x 1.5ml Tubes
(4) Blocks for 12 x 15ml Tubes
(4) Blocks for 5 x 50ml Tubes
D1102
D1105A
D1115-TALL
D1150-TALL
Baths: VWR, 18L, Shaking, Digital
STLCC_CPLS;Morrison 4/7/2014 Page 6
Temperature Range: 5 to 99°C (41 to 210°F)
Shaking Speed Range: 40–400 strokes/min.
Stroke Length: 18 mm Electrical: 115V, 50/60Hz
18.4L ,50.5 x 30 x 15 cm, VWR Cat # 89032-226
On/Off
toggle
(rear
Lower)
Themo
Protect
Set @ high
safe value
Setting Desired Temperature:
1. Turn the sample protection thermostat knob to maximum
2. Press Left arrow control key until “C” appears, press again to show current set temperature
3. Press ^UP or Down arrows to set the desired temperature
4. Press Left arrow control key again to store this set value
5. Reset the sample protection thermostat knob to a safe value
6. Temperature will revert to actual water temperature and move toward set value
7. Use a thermometer to verify that set temperature is accurate
Setting Shaking Speed:
1. Press Left arrow control key until “C” appears
2. Press the ^UP arrow until the display changes to a moving “[___]” symbol
3. Press Left arrow control key again to show the current shaking speed
4. Press UP arrow to set the desired speed between 40-400 or DOWN arrow to set to 0/zero
5. Press Left arrow control Key to lock in the desired speed
SPC-Single Point Calibration:
1. Enter this mode to finely tune the deviation from the set temperature to a closure tolerance.
2. Press Left arrow control key until “C” appears, then UP arrow until SPC displays
3. Use Up/Down arrows to set new tolerance, then Left arrow control key to save
Hotlink to VWR Water Bath User Manual …pdf
Baths: VWR, 18L, Analog, non-shaking
STLCC_CPLS;Morrison 4/7/2014 Page 7
Analog Water Bath 18 L
50.5L x 30W x 15D cm (1929/32 x 1113/16 x 529/32")
CAT# 89032-204
On/Off
toggle
(rear
Lower)
Themo
Protect
Set @ high
safe value
Hotlink to VWR Water Bath User Manual …pdf
Baths: Fuses, 12-15A, 250V required for most bath
STLCC_CPLS;Morrison 4/7/2014 Page 8
STLCC_CPLS;Morrison 4/7/2014 Page 9
STLCC_CPLS;Morrison 4/7/2014 Page 10
Water Bath: Precision Model 185 (static)
Specifications:
Volume 18 liters, chamber size 11.5"x19.5"x6"
Max temperature 99.9°C (65 deg;C without a cover)
Temperature uniformity 0.2°C at 37°C, sensitivity 0.1°C at 37°C
Operating voltage 120V
Desired Bath
Temp Setting
5 ~40C
Max High
Temp Setting
On/Off
Insert Thermometer through ring into
water to get actual temperature of bath
Safety Notes:
1) Make sure that the thermocouple probe in the
bottom of the tank is under the snap lever
and in contact with the tank bottom.
2) Turn off when not in use.
STLCC_CPLS;Morrison 4/7/2014 Page 11
Water Bath: Reciprocal Shaking - Precision Digital
Microprocessor control with digital LED display and shaker control. Push button convenience for entry of
temperature and shaking speed, Over-temperature protection capability, Automatic motor shut down
capability, One-piece, easy-to-remove tray for easy use and maintenance. Bath provides the gentle,
reciprocating motion required for many common biological, tissue culture, and biotechnology
applications. Suitable for use in applications which require a controlled atmosphere.
Chamber Capacity 3.9 gallons / 14.5 liters
Maximum Temperature 99.9oC
Temperature Display
Temperature Uniformity ±0.05oC at 37oC
Temperature Sensitivity ±0.05oC
Oscillation per Minute 30 to 200
Stroke Length 0.5 to 1.5 in.
Bath: Bead, Lab Companion (Jeiotech), 20L
STLCC_CPLS;Morrison 4/7/2014 Page 12
Operating features for BW-20B (Analog model)
• 7℃ above room temperature to 100℃
• PID temperature control
• Over-temperature limiter, alarm indication
• Analog dialed indication with fine adjustment
Dimensions: 19.6” x 11.8 x 6.5 (bathsize)
22.2” x 15.4 x 11.6 (cabinet) 18.5lbs without beads
Constructional features
• Low-profile baths with front slope-panel controls provide easier access and safe operation
• Additional inner circulating system by equipped agitator (BW-E/ D series)
• Dual bath models features individually temperature controller chambers (BW-D series)
• Various accessories gable and flat covers, test tube and spring wire racks, half-shelf adjusters etc. (optional)
* All accessories for dual baths (BW-D series) refer to BW-E/ B series accessories options per each bath size.
(Ex. Each of gable covers for BW-1005D are same as one cover for BW-05E/ B and the other cover for BW-10E/ B)
Dry, metallic beads that replace water in water baths, aluminum
blocks in dry baths, or ice in ice buckets. Bath Armor™ Beads
are convenient, versatile, and can maintain hot or cold sample
temperatures in just about any vessel type.
MIDSCI.com
42370-016 Bath Armor
Beads 16L $900.00 Set Temp
Maximum
(screwdriver) Set Time
For Bath
On/Off
Toggle and
Fuses
Specifications:
Operating range 30-40°C
Temperature variations ±0.5°C
Dimensions 240 x 360 x 160 cm (9,4 x 14,1 x 6,2 in)
Weight 7.2 kg (15,8 lb)
Capacity 1 to 4 x 96-well micro plates
Heater: Plate Incubator; BioRad-IPS, Digital, Type B
STLCC_CPLS;Morrison 4/7/2014 Page 13
The IPS microplate incubator is a dry incubator
designed to heat and maintain a constant and
homogeneous temperature inside the wells of a
microplate.
• Usable for all microplate techniques
• No water or maintenance required
• Incubation temperature regulated to within 0.5°C
• Four plate pads
Note: Allow 5 minutes to reach set
temperature of 37C degrees. 1. On/Off toggle in rear
near voltage plug,
2. Red light “on”
indicator
4. Push this button to get
a current “read” of the
plate temperature
3. Displays approximate
temperature of plates,
use “read” button to
update.
Calibration “screw”, do
not tamper with unless
you have accurate
devices to measure plate
temperatures inside.
110 or 220V set screw on
rear panel. Leave at
110V position.