Upload
roshield-randasan
View
387
Download
0
Embed Size (px)
Citation preview
INDUSTRIAL TRAINING PROJECTMAQUET: VENTILATOR SYSYEM
SERVO – I V3.1
INSTITUION : UNIVERSITY OF MALAYA
NAME : ROSHIELD RANDASAN
TEAM LEADER : MR JASNI HJ SAPAR
SUPERVISOR : MR MARTIN LEE SZE EN
SITE : BEMS, HOSPITAL DUCHESS OF KENT, SANDAKAN
DATE OF SUBMISSION: 22 JULY 2011
TMD REF NO: REGION REF NO:
Table of Contents
Contents Page No
Abstract…………………………………………………………………….. 1
Acknowledgement…………………………………………………………. 2
Objectives………………………………………………………………….. 3
Introduction…………………………………………………………………
i. Human respiratory system…………………………………….. 4 - 5
ii. Ventilator history………………………………………………. 6 - 7
iii. Modes of mechanical ventilation……………………………… 8 - 9
iv. Waveform in mechanical ventilation………………………….. 10-11
v. Important parameters in ventilation……………………………. 12-14
Ventilator System Servo – I V3.1
i. Description………………………………………………………. 15-18
ii. Modes……………………………………………………………. 19-20
iii. Set – ups and preparations………………………………………. 21
iv. Pre – use check…………………………………………………... 22-25
v. User Interface and Keys…………………………………………. 26-27
vi. Starting Ventilation…...…………………………………………. 28-29
vii. Alarm Types and Alarm Handling………………………………. 30-31
viii. Cleaning and Maintenance……………………………………….. 32-34
ix. Troubleshooting and Servicing…………………………………… 35-37
x. Planned Preventive Maintenance………………………………… 38-40
xi. Asset management……………………………………………….. 41
Conclusions………………………………………………………………….. 42
References………………………………………………………………….... 43
Appendices…………………………………………………………………... 44-51
Prepared by:
Name: Roshield Randasan
Date: 22 July 2011
Institution: University of Malaya, Kuala Lumpur
Signature
Checked by:
Name: Mr. Martin Lee Sze En
Position: Technician, HDOK
Date: 22 July 2011
Signature
Acknowledged by:
Name: Mr. Jasni Hj Sapa
Position: Team Leader, Sandakan
Date: 22 July 2011
Signature
Approved Rejected
Name: Mr. Herman Abu Saini
Position: Regional Head, Sabah
Signature
Abstract
Human respiratory system is consist of two lungs and trachea that function as a tube
that will be the path of air or gaseous when it is inhale or exhale from the lung. In lung, there
is site called gas exchange site that consist of bronchioles, alveolar ducts and alveoli.
However, the primary site for gaseous exchange is the alveoli where exchange of oxygen and
carbon dioxide from and into the blood will occur. Since human body cells need oxygen to
live and run their function, respiratory system is considered one of the vital human systems in
the body aside from the heart. However, respiratory system can also infected by several
diseases that may interrupt its function. Diseases such as bronchitis may cause the exchange
of gas to failure. Patient with polio will also have problem with their respiratory system as
they may not be able to breathe thus failed to exchange gaseous in their blood. To overcome
this problem, ventilator machine was introduced. Firstly, this report will focus on the Maquet
Ventilator System Servo – I V3.1. So, in this report, details on the history of the ventilator
machine will be explained. The modes of ventilator that used for treatment such as pressure
or volume control will also be explained in detailed. Besides that, procedure of Planned
Preventive Maintenance, User Training and Testing and Commissioning will also be
explained later. Documentation for ventilator management as one of the assets in hospital will
also be included and explained in this report especially on Maquet Ventilator System Servo –
I V3.1.
1
Acknowledgement
After several weeks of research and training, the report on Maquet Ventilator System
Servo – I V3.1 can finally be done. Firstly I want to thanks University of Malaya especially
Department of Biomedical Engineering and also CITRA for giving me the chance to undergo
industrial training which is an essential opportunity for me and other students to exactly see
the situation in the job world and also gained experience from people who is actually
working. Also big thanks to all my lecturers for the knowledge that they have given as it help
me the most during my training and in doing this report.
Lots of thank and gratitude also to Healthronics (M) Sdn Bhd for giving me the
chance to undergo training in their company. Also to Mr Herman Abu Saini, the Regional
Head of Sabah region Healthronics for giving advice in my project proposal form, to Mr Jasni
Hj Sapar whom is the Team Leader of Healthronics Sandakan for giving lots of knowledge
and advice along my journey in doing this report and also in my industrial training. Big
thanks also for the Healthronics Sandakan staff for their support and knowledge that they
have shared me during the making of this report. To Mr Martin Lee Tze En, my supervisor in
doing my report, for his knowledge and guidance that extremely helping me. Once again big
thank and gratitude from my heart to all of you for helping me in making this report and
while I undergo my industrial training. Last but not least, to my family that always got my
back and supported me while I am making this report and undergo training, thank you very
much. Also to other people that contribute directly or indirectly in making in this report, a big
thanks to all of you. May this report will be beneficial to people who read it and most likely
give if not much, a bit information on ventilator machine on how important it is in treating
patient with respiratory failure.
2
Objectives
There are several objectives that need to be achieved through this report. Those objectives
are:
i. To know and understand the history of ventilator machine and how it has
developed throughout the years before.
ii. To learn and understand the function of ventilator machine.
iii. To learn and understand the mode of treatment used in ventilator for respiratory
failure treatment.
iv. To learn and understand on how to do Planned Preventive Maintenance for
ventilator machine.
v. To learn and understand how to use equipment needed for doing Planned
Preventive Maintenance such as safety test analyser.
vi. To learn and understand about the documentation needed for registering and
managing ventilator machine as one of the assets in the hospital.
3
Introduction
I. Human respiratory system
Human body consists of several system that run their own function such as skeletal
system, digestive system, muscular system and also circulatory system. Each system run their
own function in order so that human do their daily activity. Every day in human life, there is
one most crucial and certain thing that must they done which is breathing. Breathing is
important as it enable human body to receive oxygen and remove carbon dioxide that not
used by the cells. In order for human to breath and exchange gaseous, respiratory system is
needed. This system consists of several organs such as nose, lung, trachea and bronchi. These
organs take part in the process of gaseous exchanging as some of them function as passage
for the airway such as nose and trachea while some function as gas exchanger such as lung
and bronchi.
Figure 1. Human Respiratory System
Taken from: http://bcscience8.wikispaces.com/The+Respiratory+System
4
When air is inhaled, it will first enter the nose passage. So, nose is important in here
as it become the medium of air to go into the lungs. However, air can also enter from the
mouth as nose and mouth are connected in the passage called the trachea. After the air have
gone through the passage, it will then arrive at the lungs. There are actually two lungs in
human body, which is the right and left lung. Each of them is actually covered by thin
membrane called plural membrane. Inside the lung there is bronchi where there is bronchioles
and alveoli (Jakab, 2007). All these are like tree branches that spread out in the lungs. The
most important site or part in the lungs is actually the alveoli. This is the main site for the
exchanges of oxygen and also carbon dioxide. Alveoli as mentioned before is the exchange
site of gaseous, but how does it happen will be explained here.
Alveoli is a sac like shape and have an elastic characteristic. It is able to expand and
deflate due to its elasticity. It is also surrounded by many blood vessels around its wall. It is
very tiny and given lots of empty space in the lungs due to its size. It walls is very thin and
about the same thickness as one cell. Due to its thin wall, oxygen that goes in from the nose
can diffused into it and goes into the blood. The same process happen to carbon dioxide
where it also diffuse into alveoli and into the air and follow the air out from the lungs.
Figure 2. Exchange of gas in alveoli
Taken from:
http://peer.tamu.edu/curriculum_modules/organsystems/module_4/whatweknow_lungs2.htm
5
II. Ventilator history
Ventilator machine is a machine used mechanically move air or gaseous mixture into
and out of the lung. It is also delivers either controlled or supported breaths with either
constant volume or pressure with a set of oxygen concentration (Maquet, 2004). Ventilator
was first introduced during the 20th century when polio diseases spread during that time.
However, the first form of ventilator introduced was using negative pressure and non -
invasive (Geddes, 2007). The improvement was then made by the person named John Haven
in 1931. There are also other forms of non – invasive ventilators such as rocking bed,
biphasic cuirass ventilation and iron lung and these types of ventilators were used widely for
polio patient (Geddes, 2007). Later in 1950, ventilator machine began to be used in
anaesthesia and intensive care due to the development of mechanical assister by John Haven
with the cooperation of Harvard University in 1949. The first positive pressure ventilator was
introduced back in year 1952 by Roger Manley and the ventilator was named Mark I. It then
developed become Manley Mark II after a collaboration of Roger Manley with Blease
company was made. The ventilator then widely used around Europe and was the starting
point of the used of positive pressure ventilation around the continent.
In 1955, the Americans introduced a ventilator that widely known as “Bird” or Bird
Mark 7. This ventilator was a pneumatic device thus not required any electrical power source
to operate. By the time of 1971, the intensive care environments revolutionized due to the
introduction of the SERVO 900 ventilator machine. This ventilator is small, silent and
effective in operating. It also has SERVO feedback system that enable user to regulate and set
the delivery of gaseous to patient. Later in 1991, SERVO 300 ventilator series was
introduced. This version enable the used of ventilator for all range or categories of patient,
6
form neonate to adult in one single unit of ventilator. It also has rapid flow – triggering
response.
The latest model, which used mostly nowadays, the SERVO – i was introduced in
2001. This model used modular concept, a concept that enable user to choose different mode
of mechanical ventilation from single ventilator. This type of ventilator is extremely easy and
saved more space in certain ward.
7
III. Modes of Ventilation
As mentioned earlier, ventilator is used as a machine that will help carried air and
oxygen into and out of the lung so that gas exchange in the alveoli can occur. So basically,
patient that need ventilator for in their treatment is usually patient that have respiratory
diseases or patient that not able to breath themselves such as patient that is on comma or
having stroke and paralyzed on the upper body parts. Respiratory problem mentioned here is
including acute respiratory distress syndrome, pneumonia, heart failure and complications
from surgery and trauma. COPD or chronic obstructive pulmonary disease and
neuromuscular disorder are also the factor that will lead to respiratory failure (Carbery,
2008).
To determine the right treatment for certain treatment, clinician or doctor must
considered several factors such as patient’s history and appropriate lung’s compliance. But
the most important assessment can be obtained from the ABG test. ABG test or arterial blood
gas test is important because it can give information such as blood pH, partial pressure of
CO2, and also partial pressure of O2. From this test, an acute respiratory can be determined if
the partial pressure of oxygen is less than 60 mmHg, partial pressure of carbon dioxide is
greater than 50 mmHg, and the blood pH is lower than 7.25 and falling (Carbery, 2008).
Respiratory failure if not treated can lead to more chronic diseases such as CNS (Central
Nervous System) disorder.
In mechanical ventilation, there are several modes that used in delivering breath or air
into patient. Those modes are volume control, volume support, pressure control, pressure
support, SIMV (Synchronized Intermittent Mandatory Ventilation), and CPAP (Continuous
Positive Airway Pressure). Basically, there are three basic of breath delivering system which
is controlled, support and continuous. In control mode such as volume or pressure control, the
8
whole patient breath is supported by the ventilator machine. The first mode in controlled
breath delivering system is the volume control. In volume control, the volume or Tidal
Volume is set by the user and the pressure will be dependent on patient’s Tidal Volume,
inspiration time and the resistance and compliance of the patient’s respiratory system
(Maquet, 2002). The other mode is the pressure control mode. In pressure controlled, the
parameter that set as constant is the pressure meanwhile the volume (Tidal volume) will be
independent on pressure above PEEP, lung compliance and also the resistance in the patient’s
circuit. Volume and pressure control mode can also be called as the control – assist mode.
Usually in many books or reference, the word control – assist mode is used instead of control
mode, but these two carried the same meaning.
Besides control mode, there is also support mode in mechanical ventilation. Support
mode happen whenever a patient triggered the ventilator. After being triggered, ventilator
will give positive pressure ventilation to the patient. This mode is also divided into two which
is pressure and volume support. Support mode that need to be used is determined by the
doctor or clinician. The next mode is the spontaneous mode or continuous positive airway
pressure. In this mode, patient is breathing on their own, but trigger function will also be
provided in this mode if patient effort is not sufficient to be said as spontaneous breathing.
Another mode is the Synchronize Intermittent Ventilation mode. In this mode, patient
breathing is synchronized with the ventilation given by ventilator itself. SIMV in Maquet
Servo – I is used with other mode which is the control mode and also support mode.
Triggered also available in this mode too. Triggered function will be explained more in the
Servo – I V3.1 section, on how to use the ventilator.
9
IV. Waveform in Mechanical Ventilation
There are several waveforms that need to be observed during different mechanical
ventilation. These waveforms will indicate several parameters that will be important in
assessments of lung condition or the flow of air that travel through the lungs. Three main
important waveforms that usually observed in mechanical ventilation are the pressure, flow
and volume waveform. In a mode where pressure is the variable that being set as a constant,
the waveform will be looked like in the Figure 2a below.
Figure 2a. Pressure controlled waveform
Taken from: http://www.frca.co.uk/article.aspx?articleid=100421
Volume also will be constant in a mode where volume is delivered constantly in a same value
throughout the ventilation. The waveform when volume is being fixed is pictured as in the
Figure 2b.
10
Figure 2b. Volume controlled waveform
Taken from: http://fn.bmj.com/content/77/3/F202.abstract
As can be seen from these two types of waveform, it can be said that pressure and volume
waveform changed it shapes as the mode changes. The flow waveform also change shape but
still same process happen in both waveform which is inspiration and expiration. There is also
one similarity between the flow waveform in pressure and volume control modes, which both
of the flow waveform have negative value. The negative phase in the flow volume is actually
referring to the expiration phase of a breathing (Chatburn, 2003). Meanwhile the positive
value is indicating the inspiration phase of certain breathing.
11
Inspiration phase
Expiration phase
Volume
Pressure
Flow
V. Important Parameters in Ventilation
There are many parameters in mechanical ventilation that need to be focus on while
on treatment. It is also called as breathing parameters since it is all important to maintain
sufficient or efficient breath that delivered to the patient. This report will mentioned some of
the important parameters such as tidal volum, peak expiratory end pressure (PEEP) and
breath rate. First parameter would be the oxygen concentration (O2 concentration). This
parameter is important and need to be observed during ventilation so that patient will receive
enough oxygen to their lungs and blood. The range of oxygen given to patient is usually 20 –
100% of concentration. The second one is the respiratory rate or also called as breath rate or
frequency. Normally, a person should breath total of 12 cycles of breath in one minute
(Rakhimov, 2011). But this value can be higher in person that is sick or in an infant. By
determining respiratory rate, the total of tidal volume per minute can also be calculated. Next
is the tidal volume which is volume that delivered in one breath. Some also called tidal
volume as the target volume that need to be achieved during ventilation. Then there is PEEP
or peak expiratory end pressure. PEEP is actually the pressure that left behind in the lungs
after the expiration is done. The need of PEEP in mechanical ventilation is to ensure that
human lungs do not flattened completely. This is important as lungs that contained PEEP will
be accepting more oxygen during the next breath due to increase in mean airway pressure.
Usually PEEP is set to be 5cmH2O. The next parameters would be inspiratory rise time which
is the time for a person to inhaled air into their lung. Longer inspiratory rise time will
increase the oxygen absorption and also make the patient more comfortable. Other parameter
that also important is shown in the Table 1.
12
Parameters Definition and Range
Inspiratory cycle-off (%)Fraction of maximum flow at which
inspiration should switch to expiration.
Minute volume (Vmin)
Volume per minute or target Minute volume (ml/min or l/min).
Note: Presentation can be configured to either tidal or minute volume.
Trigger sensitivity
Determines the level of patient effort required to trigger inspiration. The sensitivity is set as high as possible without self-triggering. This ensures that triggering is patient initiated and avoids autocycling by the ventilator.
There are two types of triggering:
1) Pressure triggering—This is the pressurebelow PEEP which the patient must create toinitiate an inspiration. The allowed range is:• -20 to 0 cmH2O.
2) Flow triggering—As the dial is advanced to the right (step wise from the green into thered area) the trigger sensitivity increases sothat the inhaled fraction of the bias flow leading to triggering is reduced. The allowedrange is:• 100% to 0% of the bias flow.
Note: You can’t set trigger sensitivity in NIVmode.
PC above PEEPInspiratory pressure level for each breath (cmH2O) in Pressure Control.Important: In all pressure controlled modes,it is important to set alarm limits to adequatelevels.
Pause time (Tpause) Time for no flow or pressure delivery (% or
13
s).
I:E ratio (I:E) Ratio of Inspiration time + Pause time to Expiration time. Usually set as 1:2.
Table 1. Other parameters in mechanical ventilation
14
Maquet Servo – I V3.0 Ventiilator
I. Description
Maquet Servo – I V3.0 is a ventilator from Maquet Getinge Group which is a
company from Sweden. This machine is a Class 1 and Type B machine which means that it
has protected ground and it also have long conductivity contact to the person, but not directly
to the patient’s heart. This product wa launched in the year 2001, which is 10 years ago.
Servo – I as mentioned earlier is a ventilator which comprises of many different mode within
a ventilator. This is a major advantage of this type of ventilator as it is made easy for user
plus save cost or space in getting many ventilator with different function each. Based on
Figure 3 below, it can be seen that Servo – I V3.0 has three main sections which is the user
interface, patient unit and also the ari compressor below the patient unit. There are also other
accessories that comes with this ventilator which will be explained in detail later.
Figure 3: Maquet Servo – I V3.0 Ventilator
15
User Interface
Patient Unit
Air Compressor
Patient Breathing System
As seen from Figure 3, there is one section called the User Interface, which is the
place for user such as doctor or nurse to operate the system of the machine or setting any
values that correlated to any mechanical ventilation. This User Interface is also the place for
data to be displayed so that clinician will be able to monitor patient’s condition. All the
alarms related to patient or even the ventilator itself will also indicated in the User Interface.
There are also several keys and buttons that important in managing the settings for
mechanical ventilation. All these keys will be displayed and explained further in the User
Interface section in this chapter.
Next part is the patient unit where all the gas from the module will flow in together
before flowing into the inspiratory circuit. Servo – I V3.0 can support different modules such
as air module, oxygen module and carbon dioxide module. But in this report, only two
module will be included which is the air and the O2 module. After gaseous flow from this two
modules, it will then enter patient unit where the gas can be said to be mixed before it is
proceed into the inspiratory circuit. The next section is the patient breathing system which
comprises of patient circuits. There are actually two patient circuit which is the inspiratory
and the expiratory circuit. The inspiratory circuit or tubing carry the breath that goes into a
patient’s lung while the expiratory circuit carry the air that exhaled by the patient. These two
patient circuit is important and any leakage in any of these two circuit will bring complication
to the patient. To find out whether there is any leakage in the tubing system, a Pre – Use
check must be done on the machine. Tubing compensation must also be done to make sure
that the breath that delivered to the patient is same as the one that is being set in the system.
Compensation is done also during the Pre – Use check. Pre – Use check procedure will be
explained later in detailed.
16
Other than these three main components, there are also others accessories that used
together with Servo – I ventilator such as humidifier and nebulizer. Humidifier in mechanical
ventilation is very important as it helps prevent patient from inhaling dry air. Without the
usage of humidifier, the epithelium cell of a patient will undergo destruction. Patient can also
suffered hypothermia which is a condition of low in body temperature. When using
humidifier, the temperature is set to be maximum of 37°C, and using sterile water to be
vapored (AARC, 1992). The humidifier used in this Servo – I V3.1 ventilator is the Fisher &
Paykel humidifier with model MR340E.
Figure 4. Fisher & Paykel Heated Humidifier MR340E
Other than humidifier, heat and moist exchanger can also be used to provide humidity to
patient’s breath. Heat and moist exchanger is also called as HME. The other accessory is the
nebulizer or Servo Ultra Nebulizer. Same as basic nebulizer, Servo Ultra Nebulizer also
deliver medication through mist that produced by vibration. When using Servo Ultra
17
Nebulizer, the HME or humidifier cannot be used to prevent any blockage to any of these
two.
18
II. Modes
There are several modes that can be run on Servo – I V3.1. Modes that provided in
this ventilator is almost same as the basic mechanical ventilation that already explained in the
previous chapter. But there is an addition of mode which is the Pressure Regulated Volume
Control or PRVC. PRVC mode means that pressure is also one of the key factor in giving
controlled volume breath to patient. For example, if a tidal volume is set to be sent at certain
level, the pressure that comes along must suitable with lung compliance as high pressure that
may come with high volume may burst the lung. If this case happen, PRVC mode will cut off
the tidal volume value that need to delivered if the pressure that comes with it exceeded the
limit set in the system or not compatible with lung compliance.
In Servo – I V3.1, there is actually many modes which is the pressure and volume
control, pressure and volume support, SIMV mode, PRVC mode, spontaneous mode and also
the automode. Automode in this machine provides unique mechanisms that enable the
controlled and supported mode to exchange whenever patient initiated breath or not. There is
also backup ventilation is support mode except in controlled and Automode. Backup
ventilation is important because if patient suddenly having apnea, this backup ventilation can
replaced the patient initiated breath.
19
Figure 5. Ventilation Modes in Servo – I V3.1
20
III. Sets – up and Preparation
Before starting using Servo – I Ventilator, several set – ups and preparation need to be
done to make sure that the ventilator is safe to use. The most important thing would the Pre –
Use check that must be done each time a new patient circuit or expiratory cassette are
changed. The detail procedure of Pre – Use check will be explained later in the next section
of this chapter. Battery module must also be inserted at least two to support the machine if
any blackout happen during ventilation. There are actually six slots for battery module, and if
all of it are inserted, these batteries can support the machine for at least 3 hours of operation
time. Figure 6 below shows the battery module location on Servo – I V3.1.
Figure 6. Battery module
21
IV. Pre – Use Check
Pre – Use Check is actually the most important process before starting of using the
Servo – I. Other than making sure that all pressure transducer is working fine, it also able to
compensate the volume in the patient circuit that will be useful in delivering breath later. Pre
– Use check also test the ability of the machine to transform into battery support if the mains
is not connected to the machine. Any leakage will also be tested and detected (if any) in this
procedure. Calibration process for new O2 cell and expiratory cassette membrane will also be
done in this process even though the process is not actually appear on the screen during the
test. Servo – I that installed with new patient circuit must also undergo this test to make sure
that the volume compensation for patient circuit can be done. To run Pre – Use check,
equipment that will be needed is the “blue test tube” that provided by the manufacturer that
will be used later to connect the inspiratory and the expiratory outlet.
Figure 7. Pre – Use check flow22
While doing the Pre – Use check, some instructions will appear on the screen of the Servo –
i. First will be the “ Connect the inspiratory and the expiratory outlet”, which means the “blue
test tube” should be used to connect inspiratory and expiratory outlet. Other type of tube
cannot be used in this process as it will failed the test. The next instruction will be to unplug
and plugged the power source to test the battery operation. Then the patient circuit is
connected and Y – piece is blocked. Compensation will be done in this section. If there is Y –
sensor module on the machine, Y – sensor test will also be done.
Figure 7. Y – Sensor module
After all test is done and pass, the Pre – Use checked can be completed. If any error message
appear, action needed for be taken will be explained in the Table 2 and Table 3.
23
Message Description Remedy if test fails
Cancelled Test cancelled by user Users are recommended to perform Pre – Use Check before connecting ventilator to patient
Failed Test did not pass Check all connections and expiratory cassette.
Not Completed Test was not completed Test pass with some limitation.i. Battery capacity less than 10 minutesii.Gas missing while Pre –
Use check.Passed Test has passed -Running Test in progress -
Table 2. Pre – Use check status
Display message Description Remedy if test fails
Alarm state test Checks that no Technical error alarms are active during Pre-Use check.
Refer to Service Manual
Barometer test Check the barometric pressure measured by the internal barometer
Check barometer pressure in
Status in touch panel.
Gas supply test Check whether gas supply measured by transducer are within the specified range.
Check whether gas connected is within the range for air and oxygen.
Internal leakage test Check patient circuit or any other tube connected to machine.
Make sure blue test tube is connected properly and patient circuit connection is clean and connected properly.
Pressure transducer test Calibrates and check inspiratory and expiratory transducer.
Check/replace transducer for inspiratory/expiratory section or make sure expiratory cassette is cleaned and dried thoroughly.
Safety valve test Check or adjust safety valve Check whether safety valve
24
opening pressure to 117 ± 3 cmH2O.
membrane is placed and closed properly during Pre – Use check. Also check if the inspiratory pipe is mounted properly.
O2 cell/sensor test Check O2 lifespan and calibrates it at 21% and 100% of oxygen.
Check/replace O2 sensor or gas modules.
Flow transducer test Check/calibrate inspiratory and expiratory flow transducers.
Check whether gas supplied is within range or expiratory cassette is properly mounted in its place.
Battery switch test Checks if the power supply change to battery if mains is disconnected and vice versa.
Check battery status. Battery status must be more than 10 minutes.
Patient circuit test Check patient circuit leakage Check/replace patient circuit
Y sensor test Check the pressure and flow
measurement of Y sensor
Check Y module and Y
sensor and replaced if
necessary.
Alarm state test Checks that no Technical error alarms are active during Pre-Use check.
-
Table 3. Pre – Use check error
V. User Interface and Keys
25
The User Interface of Servo – I is a touchscreen interface. Thus it made easier for the
user as all can be seen and access directly on the screen. Figure below shows how the screen
appear on the Standby mode after the Pre – Use check. All the entry such as ventilation
mode, patient data, and status of battery or oxygen cell can be observed from the User
Interface.
Figure 8. Screen (User interface) of Servo – I V3.1
On the top left corner, there is mode appear and user can just press it to enter the ventilation
mode, beside it is the patient category and now it shows infant category for this ventilator.
There is also Admit Patient column specially for entering patient data and the Status column
for the general status of machine such as battery, oxygen cell and expiratory membrane life
span. There is also Additional Setting column for Additional Value parameter setting. Other
than User Interface in the touchscreen panel, there are also few keys or button that also
function as parameter setter for the ventilator. There are also some rotary dial that will be
26
used to set the values of any parameter on the ventilator. All of these keys and rotary dial are
shown in the figures below.
Figure 9. Fixed keys, Quick Access Keys and Main Rotary Dial
Figure 10. Direct Access Knob and Special Function Keys
All these knob and keys have their own function. It will be explained in detail in the
Appendices section.
27
VI. Starting Ventilation
To start a ventilation, first a mode need to be selected. The mode should be selected
from the modes provided on the Touchscreen panel.
Figure 11. Selecting ventilation mode
After mode is being selected such as Pressure Control like in the figure above, parameters
such as PEEP, respiratory rate, and oxygen concentration can be set using the Direct Access
Knob. As a reminder, the changes made using Direct Access mode will be automatically
applied to the next cycle of breath without any confirmation. After all setting is done the
Standby key beside the Direct Access Knob is pressed to start the ventilation.
28
Select mode from here
Figure 12. Standby button for start ventilation
29
Press to start ventilation
VII. Alarm Types and Alarm Handling
In Servo – I V3.1, there are actually three types of alarm. These alarm are high
priority alarm, medium priority alarm and low priority alarm. High priority alarm is indicated
by red colour while medium and low priority alarms are indicated by yellow colour. The high
priority alarm will turn into yellow colour after it is fixed or reduced. Details of several
important high, medium and low alarms and action needed to be taken to fixed the alarm will
be explained in the table below.
Alarm messages Possible Cause Remedy
Apnea Preset or default alarm exceeded. Time between two consecutive inspiratory exceeded the limit.
Check ventilator setting, patient and breathing system.
Gas supply pressures: Low Air and O2 is below 2.0kPa Check gas connection for any leakage
Expiratory Minute Volume: Low
Preset or default alarm exceeded.
Check breathing system and patient. Change alarm limit settings
No battery capacity Less than 3 minutes of battery capacity
Change power supply to mains and recharge battery.
Paw high Airway pressure exceed Upper pressure limit
Check patient and also breathing system. Check also ventilator and alarm settings.
Restart ventilator! Software related error Restart ventilator and perform Pre-Use check.
Table 4. High priority alarms
30
Alarm messages Possible Cause Remedy
Air supply pressure: Low Air supply below 2kPa Check gas supply lines and perform Pre-Use check
Battery operation Mains voltage disappears Check mains connectionCheck Y sensor Y sensor in not working
properly or not connected to patient breathing system.
Check sensor connection or replace Y sensor module
Inspiratory flow over range Combination of settings exceed allowable inspiration flow range
Check ventilator settings or increase the gas inlet pressure.
Nebulizer disconnected Cable connection problem or nebulizer disconnected during nebulization.
Connect nebulizer or change cable connection.
Respiratory rate: High Respiratory frequency too high. Auto triggering.
Attend patient and check trigger settings.
Table 5. Medium priority alarms
Alarm messages Possible Cause Remedy
Touch screen or knob press time exceeded.
Screen or knob has been pressed for more than 1 min. Screen or knob hardware time out.
Check screen and knobs. Contact technician if problem persist.
31
VIII. Cleaning and maintenance
Servo – I ventilator must be clean and maintain regularly to make sure that it perform
to the fullest and to make sure that any bacteria or harmful virus did not spread into patient’s
body through the patient circuit. For cleanliness of the machine, any blood or other form of
liquid should not left dry on the machine. This is to prevent any bacteria to build up at the
machine. To clean the body of the machine, soap or bacteria disinfector should be used. The
patient unit must also be cleaned especially in the ventilation area such as the fan to prevent
any dirt from built up in the patient unit. This is to make sure that in the future, the dirt does
not interrupt the function and ventilation of the Servo – I itself.
Figure 13. Fan of the patient unit that need cleaning for clear ventilation of the Servo
– I itself
Next feature of cleaning would be the cleaning of the expiratory cassette. It is
recommended that the expiratory cassette is cleaned with soap or medical dish disinfector.
But the use of alcohol is also can and isopropyl is used. The use of glutaraldeyhde however is
not recommended. After cleaning and rinsing the expiratory cassette, it should be dried in the
32
dryer and not by using any pressure or force as it may damage the expiratory membrane.
However, the expiratory cassette can be shake a little bit to let out all the liquid from the
disinfectant earlier. The best is to hang the expiratory cassette vertically in the dryer. The
expiratory cassette should be dry or it will fail the Pre – Use check. Sterilization can also be
done on the expiratory cassette but it is not recommended as autoclaving can reduce the
lifetime of the expiratory cassette itself.
To make sure that any bacteria from patient is not exposed to the surrounding
environment, Servo Guard bacteria filter is mounted in the expiratory outlet. The inspiratory
inlet is also mounted with filter to prevent any bacteria from outside environment to enter the
patient circuit and patient’s lungs.
Maintenance of Servo – I is including the maintenance kit of the Servo – I itself and
the replacement of other parts such as battery and expiratory membrane. Maintenance kit
mentioned earlier is including filter of the gas module, nozzle unit for gas module, bacteria
filter for inspiratory circuit and bacteria filter for oxygen cell. This kit actually need to be
changed every 5000 hours. The oxygen cell must also be replaced if its status shown in the
Status menu in the User Interface is less than 10%. After replacement, a Pre – Use check
must be performed. The membrane of the expiratory cassette is also recommended for
exchanged if its status is shown as 0% in the Status menu. Even though the status shows that
the status is already 0%, the membrane still can actually be used, but the manufacturer
suggest that it should be replaced. The handling of expiratory should be done as in the Figure
15.
33
Figure 15. Handling of expiratory membrane
Figure 16. Parts of the expiratory membrane
After replacing the expiratory cassette, it lifespan value should be reset back to 100%
in Biomed menu right after Pre – Use check is done.
34
IX. Troubleshooting
In troubleshooting, there is actually two parts which is action needed for Pre – Use
check error and also the Technical errors. This report will explained some of them and the
full troubleshooting procedure can be refer from the Service Manual of the Servo – I V3.1.
Pre – use Check
35
Technical Error Codes
36
37
X. Planned Preventive Maintenance
This section will explained about the procedure and tools used in doing Planned
Preventive Maintenance by Biomedical Engineer Maintenance and Service staff. In the
planned preventive maintenance, the most important will be the safety test and also the test
that included in the checklist provide by Ministry of Health. Basically there are two
equipment needed which is the ventilator analyser and also the safety test analyser. The
model used for ventilator analyser in this PPM is Certifier FA Plus which is manufactured by
TSI company. Some of the parts of the ventilator analyser is shown in the figures below.
Figure 17. Certifier FA Plus kit
Figure 18. High Flow Module
38
Figure 19. Oxygen sensor tester
To perform PPM, the ventilator should be connected to the analyser as shown in Figure 20.
After connection is done, parameters such as Tidal Volume, Breath Rate, FiO2 and Peak
Inspiratory Pressure are set in a certain value according to the checklist. The value set should
be display in the Certifier FA Plus and must be within the range given.
39
For safety test, safety test analyser is used. The voltage, current and the resistance are
measured in this test. Earth leakage current and leakage current is also included in the
procedure. Details about the value will be shown in the Appendix section.
40
XI. Asset Management
Asset management is severely important in every BEMS department and also
hospitals. Document such as KEW – PA is important as it is considered as a birth certificate
for a machine. It will later useful for variation order or beyond economic procedure which is
very important for a machine. Testing and commissioning form is also important as it will be
the from that needed for a machine to be registered to the Central Management System.
Other than that, there is also Condition Appraisal form that needed for beyond
economic repair procedure. This form is actually not available for Servo – I since there is no
Servo - I ventilator that need to undergo BER. Other form such as Material Request form is
also important as it is the form used for ordering any parts that need to be replaced. All these
forms will be shown in the Appendix section.
41
Conclusions
From this report it can be conclude that Servo – I is actually important in treating
patient with respiratory problem. Other than easy to use, it also give much user friendly
function such as touchscreen interface. The asset management form such as Training and
Commissioning and KEW-PA is also important for any asset especially Servo – I so that
proper disposal of the machine when it undergo Beyond Economic Repair can be done.
42
References
1. American Association of Respiratory Care. (1992). Humidification during Mechanical
Ventilation. Journal of Respiration Care. 37. pp 887 – 890.
2. Artour Rakhimov. (2011). Normal Respiratory Rate and Ideal Breathing. Retrieved
July 20, 2011 from Normal Breathing website:
http://www.normalbreathing.com/index-rate.php
3. Catherine Carbery. (2008). Basic Concept in Mechanical Ventilation. Retrieved June
29, 2011 from Bnet website:
findarticles.com/p/articles/mi_m074B/is_3_18/ai_n31345616
4. Cheryl Jakab. (2006). Respiratory System. Smart Apple Media. Minnesota.
5. Geddes LA. (2007). The History of Artificial Repiration. IEEE Engineering in
Medicine and Biology Magazine. 26(6). pp 38 – 41
6. Maquet. (2004). Servo – I V3.1 Service Manual. Maquet Getinge Group. Sweden.
7. Robert Chartburn. (2003). Fundamentals of Mechanical Ventilation: A Short Course
in The Theory and Application of Mechanical Ventilators. Mandu Press Ltd.
Cleveland.
43
Appendices
44
45
46
47
48
49
50
51