Upload
others
View
1
Download
0
Embed Size (px)
Citation preview
09/11/2017
1
Tube current modulation and dose reduction :
How TCM works
Dean PekarovičUMC Ljubljana,
Institute of Radiology
Quality and Safety office
IAEA RER/9/135COURSE ON OPTIMIZATION IN COMPUTED TOMOGRAPHY
Sofia, Bulgaria, 2017
Pekarovic, Sofia, 2017
Plain Radiography
one kV valueone mAs value
What should be on Image ?Pulmo, Fat, Air, Bones …
Do we have same attenuation in all parts of the Image ?
09/11/2017
2
Pekarovic, Sofia, 2017
kV increase will
Decrease the mAs required to achieve a constant detector exposure (AEC on).
Decrease ESD and effective dose. Decrease image contrast. Increase scattered radiation.
Ref: ICRP 121, Radiological protection in paediatric diagnostic and interventional radiology
79 Kv0,71 mAsDAP 0,128 dGy cm2
EI: 250,9DI: -0,4524,98 µGy
67 Kv0,71mAsDAP 0,085 dGy cm2
EI 191,5
DI -0,7134,57 µGy
Radiologist to decide – contrast ribs/pulmo
Pekarovic, Sofia, 2017
mAs increase willIncrease detector exposure (without AEC) Increase radiation dose to the patient.
Improve image quality – increased CNR and SNR ?.
75 Kv1,2 mAsDAP 0,224 dGy cm2
DI 3,23NO AEC
75 Kv0,63 mAsDAP 0,084 dGy cm2
DI -0,9NO AEC
noise / higher mAs
09/11/2017
3
Pekarovic, Sofia, 2017
Lookin for proper ImQ DX
3 weeks, girl, 2760 g
65 kV DAP: 0,021 dGycm2
0,71 mAs DI : 0,35
63 kV DAP: 0,032 dGycm2
0,9 mAs DI : 1,1558 kV DAP: 0,053 dGycm2
1 mAs DI : 2,26
62 kV DAP: 0,054 dGycm2
1,21 mAs DI : 3,04, + Cu Filter70 kV DAP: 0,021 dGycm2
0,56 mAs DI : - 0,42, + Cu Filter
newborn girl 7 days2750 g
Pekarovic, Sofia, 2017
What we need is how to control theImQ and Dose
kV mAs EI DI DAP µGy mGy/s
90 1 231,53 -0,32 0,861
90 1,1 256,4 0,12 0,958 71,9 16,6
90 1,2 284,8 0,57 1,045
80 2 268,5 0,32 1,483
80 1,8 1246,5 -0,03 1,343 101 15,15
80 1,6 217,9 -0,6 1,181
72 3,6 238,5 -0,2 2,215 166 16,6
kV mAs EI DI DAP
90 1 233 -0,28 0,872
92 1 261 0,26 0,41691 1 245,7 -0,07 0,883
DI %3 1002 581 260 0-1 -21-2 -37-3 -50
How to improve DI1mAs cca 0,4 DI1 Kv cca 0,3 DI
CsJ Nastavljeni protokoli
Size range kv mAs focal spot target EI
OTROK PC 1 KG SMALL 68 0,45 1 226,22MEDIUM 70 0,56 1 226,22
LARGE 70 0,63 1 226,22range kv mAs focal spot target EI
OTROK PC 2 KG SMALL 70 0,45 1 226,22
MEDIUM 72 0,56 1 226,22LARGE 72 0,63 1 226,22
range kv mAs focal spot target EI
OTROK PC 3 KG SMALL 71 0,5 1 226,22MEDIUM 73 0,56 1 226,22
LARGE 73 0,63 1 226,22range kv mAs focal spot target EI
OTROK PC 4 KG SMALL 72 0,56 1 226,22
MEDIUM 74 0,63 1 226,22
LARGE 74 0,63 1 226,22range kv mAs focal spot target EI
OTROK PC 5-10 KG SMALL 75 0,63 1 226,22
MEDIUM 75 0,71 1 226,22
LARGE 75 0,8 1 226,22range kv mAs focal spot target EI
OTROK PC 10-15 KG SMALL 75 0,8 1 226,22
MEDIUM 77 0,8 1 226,22
LARGE 77 0,9 1 226,22range kv mAs focal spot target EI
OTROK PC 15-20 KG SMALL 80 0,71 1 226,22
MEDIUM 80 0,8 1 226,22
LARGE 80 0,9 1 226,22range kv mAs focal spot target EI
OTROK PC 20-30 KG SMALL 80 0,71 1 226,22
MEDIUM 85 0,8 1 226,22
LARGE 85 0,9 1 226,22range kv mAs focal spot target EI
OTROK PC 30-50 KG SMALL 85 0,9 1 226,22MEDIUM 90 0,9 1 226,22
LARGE 95 1,1 1 226,22
09/11/2017
4
Pekarovic, Sofia, 2017
From plain to CT …
Pekarovic, Sofia, 2017
CT -problem
09/11/2017
5
Pekarovic, Sofia, 2017
Tube Current Modulation
Purpose is to ensure the same picture quality regardless of the characteristics of the patient.
= same CNR in all images
Pekarovic, Sofia, 2017
Tube Current
• Determines the number of electrons accelerated across the x-ray tube per unit time
• Units: milliAmperes (mA)
• CTDIvol is directly proportional to Tube Current
CTDIvol Tube Current
REF : AAPM Computed Tomography Radiation Dose Education Slides
09/11/2017
6
Pekarovic, Sofia, 2017
TCM – along z axis mAs changes automatically
55mAs
130mAs110mAs
140mAs
Pekarovic, Sofia, 2017
Tube Current Modulation (TCM) / Automatic Exposure Control (AEC)
• Automatically adapts the Tube Current or Tube Potential according to patient
attenuation to achieve a specified image quality
– Automatic adjustment of Tube Current may not occur when Tube
Potential is changed
– Centering the patient in the gantry is VITAL for most AEC systems
• AEC aims to deliver a specified image quality across a range of patient sizes.
• It tends to increase CTDIvol for large patients and decrease it for small patients
relative to a reference patient size.
The use of Automatic Exposure Control may decrease or increase CTDIvol depending on the patient size and body area imaged and
image quality requested
REF : AAPM Computed Tomography Radiation Dose Education Slides
09/11/2017
7
Pekarovic, Sofia, 2017
TCM - Principe
• Reference data is collected from attenuation of anatomical structures during scanned projection radiograph -SPR
(one or two)
• Data collected from previous rotation.
Pekarovic, Sofia, 2017
TCM -principle
• TCM could be defined as a CT technique that performs automatic modulation of tube current in the x, y plane (angular modulation), or along the scanning direction, z-axis, (longitudinal modulation), or both (combined modulation).
Is really total „Automatic“ ? • The modification is done according to each patient’s size,
shape and attenuation of body parts being scanned.
• The operator must select a required image quality level and then the system can adjust the tube current to obtain the predetermined image quality with improved radiation efficiency.
Ref :Kalra MK et all, Computed tomography radiation dose optimization: scanning protocols and clinical applications of automatic exposure control.Curr Probl Diagn Radiol 2005; 34:171-181
09/11/2017
8
Pekarovic, Sofia, 2017
Start info..
• TCM needs input WHAT is needed Image quality = radiologist input, but team should be
involved; + radiographer + medical physicist
• Different vendor solutions :– Noise = SD CT-numbers on the Image
• (GE, Toshiba)
– Reference mA or mAs for a standard patient for specified quality
• (Philips, Siemens)
Pekarovic, Sofia, 2017
TCM idea
AP diameter : 36 cm
LAT diameter : 28 cm
09/11/2017
9
Pekarovic, Sofia, 2017
Scan where Tube Current Modulation was used
Blue Curve Represents actual instantaneous mARed Curve Represents avg mA for each imageYellow Curve Represents avg mA over entire scanOverall avg is used for CTDIvol reported in Dose Report
Pekarovic, Sofia, 2017
Dose Modulation -AEC
• Many CT scanners automatically adjust the technique parameters (and as a result the CTDIvol) to achieve adesired level of image quality and/or to reduce dose
• Dose Modulation and Reduction techniques vary by scanner manufacturer, model and software version.
Tip : read the manual and then test the AEC.
09/11/2017
10
Pekarovic, Sofia, 2017
Variations in mA modulation
Spatial mA modulation
– Longitudinal -Z axis modulation
– Angular / x-y axis / transverse modulation
– Combined / x-y-z modulation
Temporal mA modulation
– mA adapted at different time points using ECG gating (Cardiac CTA);not included in this PPT
Pekarovic, Sofia, 2017
TCM Modulations
Longitudinal (Z direction )
• Is an TCM feature that adjusts the Tube Current as patient size and attenuation changes of the anatomic region in the longitudinal direction.
• The CT Localizer Radiograph is used to estimate patient attenuation.
REF : AAPM Computed Tomography Radiation Dose Education SlideS
09/11/2017
11
Pekarovic, Sofia, 2017
TCM Modulations
Angular (X,Y direction)
• Is an TCM feature that adjusts the Tube Current as the x-ray tube rotates around the patient according to the size, shape, and attenuation of body region being scanned to compensate for attenuation changes with view angle.
• Angular Tube Current Modulation is used to adjust the Tube Current to attempt to deliver similar dose to the detector at all view angles.
REF : AAPM Computed Tomography Radiation Dose Education SlideS
Ref: M.K.Karla, Automatic Exposure Control in Multidetector-row CT
180°-mirror
Pekarovic, Sofia, 2017
TCM Modulations
Angular + Longitudinal (x, y, z direction)
• Is an AEC feature that incorporates the properties of both Angular and Longitudinal Tube Current Modulation to Adjust the Tube Current based on the patient’s overall attenuation.
• Modulate the Tube Current in the angular (X-Y) and longitudinal (Z) dimensions to adapt to the patient’s shape.
Ref: M.K.Karla, Automatic Exposure Control in Multidetector-row CT
REF : AAPM Computed Tomography Radiation Dose Education SlideS
09/11/2017
12
Pekarovic, Sofia, 2017
All in one
Ref : Marcus Söderberg; Automatic exposure control in CT
Pekarovic, Sofia, 2017
Principles are similar, but ..?
How it works on my CT modality
09/11/2017
13
Pekarovic, Sofia, 2017
Siemens : Care Dose 4D
• Implies need for image quality equal to that obtained with the use of specified reference mAs in a standard adult (70 kg -80 kg) or child (20 kg).
• Head Protocols –only x,y modulation
• Quality Reference mAs is NOT the max or min.
AEC System Operator-Controlled Z-Axis Angular X,Y
Vendor Name Parameter Parameter Explanation Principles Technique Modulation Modulation
Angular modulation of
tube current in the x,
Image quality y, and z axes on the
Siemens CAREDose4D reference mAs mAs that would be used for basis of patient size x-y-z/combined Yes Yes
an average-sized patient relative to the mAs
specified by the user
for a standard-sized
reference patient
Pekarovic, Sofia, 2017
Siemens : Care Dose 4D
• The degree to which the tube current is adjusted for patient size can be selected, using ‘very weak’ to‘very strong in 5 steps (high degree of mA adjustment)’, compensation settings.
Ref : Siemens and Sodberg, Automatic exposure control in CT
09/11/2017
14
Pekarovic, Sofia, 2017
AEC System Operator-Controlled Z-Axis Angular X,Y
Vendor Name Parameter Parameter Explanation Principles Technique Modulation Modulation
Modulation of tube
current on the basis of Z-DOM D-DOM
Reference image Image quality expressed in patient size to achieve with ACS
Phillips Dose Right Ref mAs/slice terms of noise level of an the same image noise Yes Yes
ACS existing optimal clinical level as in a previously
image defined reference
image
Philips : Dose Right
• “Baseline mAs” is used as a reference to obtain constant image noise along the z axis.
• DoseRight ACS (automatic current selector) provides patients based AEC(object size), by use of a reference image
• DoseRight DOM (Dose Modulation)– Z-Dom: Z-axis AEC– D-Dom: Tube current is set so that 90% of images will have
equal or lower noise than the reference image, with remaining 10% of images in a series having equal or higher noise than the reference image.
• No ACS for Head CT protocols.
*miss new technique with x, y and z axis modulation
Pekarovic, Sofia, 2017
Philips : Dose Right*
DoseRight 3D-DOM (three dimensional dose modulation) combines angular and longitudinal patient information to modulate dose in three dimensions (x-y-z-axis). It incorporates modulation of tube current time product (mAs) according to changes in individual patient’s size and shape in the transverse (x-y-axis; angular) direction during helical scans, in addition to changes in the
craniocaudal or caudocranial (z-axis; longitudinal) direction, as the tube rotates.
Additional Parameter: Reference Noise Index
09/11/2017
15
Pekarovic, Sofia, 2017
GE : AutomA 3D
AEC System Operator-Controlled Z-Axis Angular X,Y
Vendor Name Parameter Parameter Explanation Principles Technique Modulation Modulation
Measure of image quality/ Modulation of tube
noise level defined current only in the
Auto mA Noise index relative to uniform water longitudinal direction z axis/longitudinal Yes No
phantom to maintain a constant NI
Measure of image quality/ Modulation of tube
GE noise level defined current in the x, y, and
Smart mA Noise index relative to uniform water z axes to maintain a x-y axis/angular Yes Yes
phantom constant noise index
Auto mA 3D* Noise index x-y-z/combined Yes Yes
• GE : defined with NI (Noise Index)• NI : Std.Dev. of HU number in water phantom with standard algorithm used • NI CTDI vol
Auto mA use last scout if more then one were performed.
Smart mA MUST works with Auto mA
Pekarovic, Sofia, 2017
GE : AutomA 3D
Noise Index is Image Quality Parameter which sets the image noise in the image. Scout is used to determine patient attenuation characteristics and size and along with Noise Index the mA per rotation for the acquisition is determined. Noise Index will vary based on the slice thickness selected due to the difference in image noise relative to slice thickness. The same NI should never be used across all slice thicknesses.
Minimum and maximum mA - Range of allowed mA to achieve desired noise index
09/11/2017
16
Pekarovic, Sofia, 2017
Toshiba – SureExposure 3D
• Sure Exposure 3D is Longitudinal Modulation• XY – Modulation is Angular Modulation
• 2 possibilities for radiographer :• Standard Deviation –SD• Image Quality level –High, Standard..
SD of pixel values measured in a patient-equivalent water phantom.
AEC System Operator-Controlled Z-Axis Angular X,Y
Vendor Name Parameter Parameter Explanation Principles Technique Modulation Modulation
User prespecifies image
quality on the basis of
Standard deviation of pixel a patient-equivalent
Sure Exposure Target Image quality values in an image water phantom, and x-y-z/combined Yes Yes
3D level Higher SD= higher noise mAs is modulated on Always ON
the basis of patient
Toshiba size to maintain image
quality
Sure Exposure z axis/longitudinal
Pekarovic, Sofia, 2017
LL (50cm), L (40cm), M (32cm), S (24cm), SS (18cm)
AEC setup allows tube current to be limited by max. and min. values.
Be careful with reconstructed slice thickness and reconstruction kernel.
Toshiba – SureExposure 3D
09/11/2017
17
Pekarovic, Sofia, 2017
Siemens -ref. mAs
Image Quality Parameters -TCM
Ref :M.P.Supanich,Tube Current Modulation,3rd CT Dose Summit
GE NI (Noise Index) Philips – mAs/slice
Pekarovic, Sofia, 2017
Scan and Reconstruction parameters
Which parameter effect TCM ?
Ref :
CARE Dose 4D * Yes Yes Yes
Dose Right Yes Yes Yes Yes Yes
AutomA 3D Yes Yes Yes Yes Yes
SureExposure 3d Yes Yes Yes Yes Yes Yes
*New Versions of CARE Dose 4D, a change in the tube volatage will result in a change in tube current by the AEC
** add to original Table
Localizer**AEC systemTube
Voltage
Rotation
timePitch Slice Thickness
Reconstruction
kernel
09/11/2017
18
Pekarovic, Sofia, 2017
Localizer -SPR
• Start and end of scan range.• Used for appropriate tube current modulation
(AEC in use).• Incorrect position in iso center leads to
inadequate tube current modulation (higher or lower).
• Incorrect position lead to incorrect geometry and size of scanned object.
Pekarovic, Sofia, 2017
6 cm from Center
TH 0 kV 120mAs 190AEC ON
LocalizerBoth has same exposure parameters
• kV 120• mAs 35• Length 256 mm• CTDI vol (32cm) 0,13mGy• DLP 3 mGycm• Table Height Δ 6cm
Positioned above isocenter (+6 cm)
TH +6 cm kV 120mAs 230AEC ON
1. Spiral mode• kV 120• mAs/ref mAs 204/180• Length 117 mm• Slice thickness 5 mm• CTDI vol (32cm) 13.6 mGy• DLP 145 mGycm
2. Spiral mode• kV 120• mAs/ref mAs 219/180• Length 117 mm• Slice thickness 5 mm• CTDI vol (32cm) 16.7 mGy• DLP 178 mGycm
22 % difference in dose from 6 cm table height difference.
09/11/2017
19
Pekarovic, Sofia, 2017
Isocenter – noise and CR
noise increase
isocenter
8
10
12
14
16
18
20
0 50 100 150
No
ise
(1
SD H
U)
y-coordinate (mm)centered 60 mm below
Ref : Eurosafe, M. Kortesniemi, D. Pekarovic, D.Sheppard
• Cylindrical 16 cm CTDI-head phantom scanned in center position and while lowered by 60 mm.
• Noise (1SD HU) increases vertically across the phantom in lower position as the beam shape is non-optimally targeted in the scan.
60
70
80
90
100
110
0 50 100 150
Co
ntr
ast
(HU
)
y-coordinate (mm)centered 60 mm below
isocenter
slight contrast change
Pekarovic, Sofia, 2017
Spr DIRECTION
09/11/2017
20
Pekarovic, Sofia, 2017
Scanning Outside The Localizer
• Localizer is basic source for TCM
• If scan region extends outside of localizer „we have a problem“.
Four possible outcomes in scan region not covered by localizer :
• Tube Current goes to maximum
• Tube Current goes to minimum
• Tube Current stays what it was at edge of localizer
• Tube Current goes to manual settingRef : TCM, M.Supanich,3rd CT Dose Summit
Pekarovic, Sofia, 2017
One or two and direction of thelocalizer
09/11/2017
21
Pekarovic, Sofia, 2017
TCM performance on QA phantom
Ref :Iball: A QA test for CT AEC systems , Journal Of Applied Clinical Medical Physics
Pekarovic, Sofia, 2017
0
25
50
75
100
125
150
175
200
225
250
122 43 64 85 10
612
714
816
919
021
123
225
327
4
mA
mA
120 kV53 mAs
120 kV146 mAs
mA mA mA mA mA mA
215 56,5 57 119,5 54,5 62
211 56 57,5 112,5 54,5 62
208,5 56 57,5 109 54,5 62
206 55,5 58 106,5 54,5 62
200 55,5 58,5 101,5 54,5 62
197 55,5 59 99,5 54,5 62
194 55,5 59 96 54 62
188 55,5 59,5 91 54 62
185 55,5 59,5 90 54 62
181,5 55,5 59,5 89 54 62
175,5 56 59,5 87 54 62
174 56 59,5 85 54 62
172,5 56 59,5 83,5 53,5 62
169 56 59,5 80,5 53,5 62
167 56,5 60 79,5 53,5 62
164 56 60 79 53,5 62
157 55,5 60 76,5 53,5 62
153,5 55,5 60,5 76 53,5 62
150 55 60,5 75 53,5 62
142,5 55 60,5 74 53,5 62
138,5 55 61 74 53,5 62,5
134,5 54,5 61,5 73 53,5 62,5
127 54,5 61,5 70 53,5 62,5
123 54,5 62 69 53,5 62,5
Example – how to start
09/11/2017
22
Pekarovic, Sofia, 2017
0
20
40
60
80
63
3
64
6.3
65
9.6
67
2.9
68
6.2
69
9.5
71
2.8
72
6.1
73
9.4
75
2.7
76
6
77
9.3
mA
26 Months
Thorax wo CM
Topogram Cr/Ca100kV 35 mAs
Helical100 Kv30 mAs AEC off
g/rot : 0,5 sCTDIvol: 1.23 mGyDLP : 23 mGy cm
Example
kV 100mAs 30
Pekarovic, Sofia, 2017
19 Months
Thorax +Adomen wo CM
Topogram Cr/Ca100kV 100 mAs
Helical100 Kv100 mA AEC off
g/rot 0,5 sCTDIvol : 4.50 mGyDLP : 182 mGy cm (+Abd)
kV 100mAs 100
0
100
200
1 27 53 79 105
131
mA
mA
Example
kV 100mAs 100
09/11/2017
23
Pekarovic, Sofia, 2017
18 Months
Thorax wo CM
Topogram Cr/Ca100kV
Helical100 Kv30 ref/ 27 mAs AEC on
g/rot 0,5 s
CTDIvol 4.50 mGyDLP 182 mGy cm
kV 100mAs 18
0
20
40
60
80
1 16
31
46
61
76
91
10
6
12
1
13
6
mA
mA
30 % mAs deviation in Z plane
Example
Pekarovic, Sofia, 2017
26 Months
Thorax wo CMTopogram Cr/Ca
120kV Helical
100 Kv20 ref/ 15 mAs AEC on
g/rot 0,5 sCTDIvol : 0.80 mGyDLP : 14 mGy cm
kV 100mAs 11
Example
09/11/2017
24
Pekarovic, Sofia, 2017
When should it be avoided?
A great tool, but sometimes ..
• Head exams? – Brain
– Orbits
– Sinuses
• CT Perfusion – Little or no table motion
– No chance to change cross section shape
– low manual technique
Pekarovic, Sofia, 2017
Team
• Therefore, it is paramount that radiologists know how to use the AEC systems in their own scanners. AEC systems do not reduce radiation dose per se; rather, they control radiation exposure relative to the requiredimage quality.
Only Radiologist ?
09/11/2017
25
Pekarovic, Sofia, 2017
EMAN - WP 8 :Training & Education
1.The medical practitioners requesting a CT examination. This group requires knowledge about indications for CT, its alternatives and the associated risks and benefits.
2. The core CT team that defines and optimizes the set of standard scan protocols on a specific scanner (radiographer, medical physicist and radiologist).This team will usually start with a standard set of protocols provided by the manufacturer and adapt it to the local needs. This team requires in-depth knowledge of scan parameters and how to optimize them.
3. The professionals (radiologists, radiographers) define the CT protocols. This group has to have knowledge when not to use CT, but another image technique, according to patient clinical indication. They are ultimately responsible for the individual choice of the correct protocol associated with each of the set of available standard protocols at a specific scanner / institution.
4. The radiographers that actually perform the examination. This group requires knowledge about individual routine adaptations required for each patient, such as centring of patients, adapting scan range, adapting protocol to patient size, optimizing modality performance in order to obtain the best diagnostic image at the lowest possible dose.
Pekarovic, Sofia, 2017
How to evaluate
• Understanding CTDIvol
• Dose Reference Levels ( will be disuseed..)
09/11/2017
26
Pekarovic, Sofia, 2017
CTDIvol – DLP – Effective Dose
CT “Dose Numbers”
Pekarovic, Sofia, 2017
Why CTDIvol ?
• CTDIvol provides information about the amount of radiation used to perform the study.
• CTDIvol is a useful index to track across patients and protocols for quality assurance purposes .
• CTDIvol can be used as a metric to compare protocols across different practices and scanners when related variables, such as resultant image quality, are also taken in account.
09/11/2017
27
Pekarovic, Sofia, 2017
DLP
• The Dose Length Product (DLP) is also calculated by the scanner.
• DLP is the product of the length of the irradiated scan volume and the average CTDIvol over that distance.
• DLP has units of mGy*cm.
Pekarovic, Sofia, 2017
CTDI
Absorbed dose in the slice.
Absorbed dose including scattercontributions from outside the slice
(CTDI).
The CTDI is calculated as the integral of the absorbed dose along the z axis, divided by the nominal slice thickness S.
Ref : Easy Guide to low dose,Siemens
09/11/2017
28
Pekarovic, Sofia, 2017
CTDIw
There are different ways to calculate the CTDI. One of them is to consider the differences between the absorbed dose in the periphery and in the centre of the patient’s body by a weighted sum of the central and peripheral CTDI values.
Ref : Easy Guide to low dose,Siemens
Pekarovic, Sofia, 2017
CTDIvol -DLP
The Dose Length Product (DLP) is the product of CTDIvol and the examination range.
Ref : Easy Guide to low dose,Siemens
The Dose Length Product (DLP) is also calculated by the scanner.
DLP is the product of the length of the irradiated scan volume and the average CTDIvol over that distanceDLP has units of mGy*cm
09/11/2017
29
Pekarovic, Sofia, 2017
Effective dose
Normalised values of effective dose per dose-length product (DLP) over various body regions.
Region of body
Normalised effective dose, EDLP
(mSv mGy-1 cm-1)
Head 0.0023
Neck 0.0054
Chest 0.017
Abdomen 0.015
Pelvis 0.019
E = EDLP · DLP
Effective dose E -measured in Sievert (Sv) units, includes the sensitivity to radiation of the different organs. It is the
sum of the equivalent doses in all irradiated organs multiplied by the respective tissue weighting factors wi.
Pekarovic, Sofia, 2017
ICRP 60 – ICRP 103
The Recommendations of the International Commission on Radiological Protection of 2007 (ICRP 103) has different coefficients than that of 1990 (ICRP 60).
In particular, gonads are less radiosensitive and the breast is more radiosensitive than previously assumed.
The effective dose E is an approximate measure that was introduced to compare the stochastic risk of a nonuniform exposure of ionizing radiation with the risk caused by a
uniform exposure of the whole body.
09/11/2017
30
Pekarovic, Sofia, 2017
Conclusions
Users must possess a good understanding of the concepts of noise index, standard deviation, reference images, and reference mA(mAs) as they relate to each AEC system.
Use User manual and ask vendor for additional papers.
Evaluate your own AEC system, for each software and scanner type.
Don`t accept one solution. Understanding will improve reports and CT Dose will become controllable.
Pekarovic, Sofia, 2017
Thank you