1Vo Planung – Medizinische Informatik
Richard PÖTTER
Treatment planning
principles
Dept. of Radiotherapy & Radiobiology
Medical University of Vienna
2Vo Planung – Medizinische Informatik
Soon after the discovery of x rays by Roentgen in 1895 and of natural radioactivity by Becquerel in 1896 it became apparent that ionizing radiation was not only useful for diagnosis and treatment of disease but also harmful to human tissues in terms of both somatic and genetic effects.
Londons Radiotherapy department 1905
Isocentric linac + MLC
3Vo Planung – Medizinische Informatik
Radiotherapy is localized cancer treatment where a maximum dose shall be delivered to the tumor while the dose to the surrounding normal tissue must be minimized.
Curative and palliative treatments
Imaging and dose calculation in RT - 2 D versus 3 D
Justification of technology
4Vo Planung – Medizinische Informatik
Treatment techniques…
D ose
p
0
1
TC P
N TC P
Dose response curves
External beam therapy vs. brachytherapy
Different radiation “sources”
5Vo Planung – Medizinische Informatik
0
10
20
30
40
50
60
70
80
90
100
0 5 10 15 20 25WASSERTIEFE [cm]
REL
ATI
VE T
IEFE
ND
OSI
S [%
]
Röntgen 150keV
Co - 60 (Gamma)
Elektronen 6 MeV
Elektronen 18 MeV
Photonen 6 MV
Photonen 18 MV
Protonen 185 MeV
Radiation qualities
6Vo Planung – Medizinische Informatik
ISODOSEN VERTEILUNG
90 %
80 %
70 %
60 %
50 %
10 %
90 %80 %70 %60 %
50 %
10 %
30 %
30 %
90 %
30 %10 %
60 %
80 %70 %
50 %5
10
15
20
30
25
5
10
15
20
30
25
10
20
15
5
Co - 60 6 MV Photonen 6 MeV Elektronen
Tiefendosisverlauf
Halbschatten
Profile
7Vo Planung – Medizinische Informatik
Innsbruck: 4Feldkirch: 2
Salzburg: 4
Klagenfurt: 3
Graz: 3
Wr. Neustadt: 3
Linz: 4
WienAKH: 5Lainz: 2WSP: 2SMZO: 1KFJ: 2
IST-STAND IN ÖSTERREICH:13 Abteilungen für Strahlentherapie37 Hochvolttherapiegeräte + 2 Gamma Knife
Krems: 2
8Vo Planung – Medizinische Informatik
The treatment technique itself depends on the underlying basis for treatment planning or technique for patient data acquisition.
QUALITY ASSURANCE throughout all processes in the RT chain
collimationdose calculation
Treatment plan selectiontreatment
geometryBiological
considerations
verification
Imaging and target
definition
9Vo Planung – Medizinische Informatik
TREATMENT PLANNINGExternal beam radiotherapy: usually carried out with more than one beam
Exceptions for single field treatments is therapy utilising conventional X rays and electron beam therapy.
Modern RT is carried out with a variety of beam energies and field sizes under one of two setup conventions
In an SSD setup, the distance from the source to the surface of the patient is kept constant for all beams,
For an SAD setup the center of the target volume is placed at the machine isocenter.
Field sizes: from small circular fields used in radiosurgerythrough standard rectangular and irregular fields to very large fields used for total body irradiations.
10Vo Planung – Medizinische Informatik
TREATMENT PLANNINGAn optimal dose distribution can only be achieved if different technical therapy options are utilized. The following parameters have an influence on dose distributions:
Beam quality (x-rays, high energy photon or electron beams, neutrons, heavy charged particles) and beam energyBeam incidence (single field, multiple fields, rotational therapy, conformal radiotherapy, stereotactic RT, intensity modulated RT)Beam modifiers (wedges, bolus, multileaf collimators, individually shaped blocks, …Dose modifications (boost, inhomogeneous dose distribution,..)
The treatment technique itself depends on the underlying basis for treatment planning or technique for patient data acquisition