1 Vo Planung – Medizinische Informatik Richard PÖTTER Treatment planning principles Dept. of Radiotherapy & Radiobiology Medical University of Vienna

Treatment planning principles · ÎIn an SSD setup, the distance from the source to the surface of the patient is kept constant for all beams, ÎFor an SAD setup the center of the

  • Upload
    others

  • View
    1

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Treatment planning principles · ÎIn an SSD setup, the distance from the source to the surface of the patient is kept constant for all beams, ÎFor an SAD setup the center of the

1Vo Planung – Medizinische Informatik

Richard PÖTTER

Treatment planning

principles

Dept. of Radiotherapy & Radiobiology

Medical University of Vienna

Page 2: Treatment planning principles · ÎIn an SSD setup, the distance from the source to the surface of the patient is kept constant for all beams, ÎFor an SAD setup the center of the

2Vo Planung – Medizinische Informatik

Soon after the discovery of x rays by Roentgen in 1895 and of natural radioactivity by Becquerel in 1896 it became apparent that ionizing radiation was not only useful for diagnosis and treatment of disease but also harmful to human tissues in terms of both somatic and genetic effects.

Londons Radiotherapy department 1905

Isocentric linac + MLC

Page 3: Treatment planning principles · ÎIn an SSD setup, the distance from the source to the surface of the patient is kept constant for all beams, ÎFor an SAD setup the center of the

3Vo Planung – Medizinische Informatik

Radiotherapy is localized cancer treatment where a maximum dose shall be delivered to the tumor while the dose to the surrounding normal tissue must be minimized.

Curative and palliative treatments

Imaging and dose calculation in RT - 2 D versus 3 D

Justification of technology

Page 4: Treatment planning principles · ÎIn an SSD setup, the distance from the source to the surface of the patient is kept constant for all beams, ÎFor an SAD setup the center of the

4Vo Planung – Medizinische Informatik

Treatment techniques…

D ose

p

0

1

TC P

N TC P

Dose response curves

External beam therapy vs. brachytherapy

Different radiation “sources”

Page 5: Treatment planning principles · ÎIn an SSD setup, the distance from the source to the surface of the patient is kept constant for all beams, ÎFor an SAD setup the center of the

5Vo Planung – Medizinische Informatik

0

10

20

30

40

50

60

70

80

90

100

0 5 10 15 20 25WASSERTIEFE [cm]

REL

ATI

VE T

IEFE

ND

OSI

S [%

]

Röntgen 150keV

Co - 60 (Gamma)

Elektronen 6 MeV

Elektronen 18 MeV

Photonen 6 MV

Photonen 18 MV

Protonen 185 MeV

Radiation qualities

Page 6: Treatment planning principles · ÎIn an SSD setup, the distance from the source to the surface of the patient is kept constant for all beams, ÎFor an SAD setup the center of the

6Vo Planung – Medizinische Informatik

ISODOSEN VERTEILUNG

90 %

80 %

70 %

60 %

50 %

10 %

90 %80 %70 %60 %

50 %

10 %

30 %

30 %

90 %

30 %10 %

60 %

80 %70 %

50 %5

10

15

20

30

25

5

10

15

20

30

25

10

20

15

5

Co - 60 6 MV Photonen 6 MeV Elektronen

Tiefendosisverlauf

Halbschatten

Profile

Page 7: Treatment planning principles · ÎIn an SSD setup, the distance from the source to the surface of the patient is kept constant for all beams, ÎFor an SAD setup the center of the

7Vo Planung – Medizinische Informatik

Innsbruck: 4Feldkirch: 2

Salzburg: 4

Klagenfurt: 3

Graz: 3

Wr. Neustadt: 3

Linz: 4

WienAKH: 5Lainz: 2WSP: 2SMZO: 1KFJ: 2

IST-STAND IN ÖSTERREICH:13 Abteilungen für Strahlentherapie37 Hochvolttherapiegeräte + 2 Gamma Knife

Krems: 2

Page 8: Treatment planning principles · ÎIn an SSD setup, the distance from the source to the surface of the patient is kept constant for all beams, ÎFor an SAD setup the center of the

8Vo Planung – Medizinische Informatik

The treatment technique itself depends on the underlying basis for treatment planning or technique for patient data acquisition.

QUALITY ASSURANCE throughout all processes in the RT chain

collimationdose calculation

Treatment plan selectiontreatment

geometryBiological

considerations

verification

Imaging and target

definition

Page 9: Treatment planning principles · ÎIn an SSD setup, the distance from the source to the surface of the patient is kept constant for all beams, ÎFor an SAD setup the center of the

9Vo Planung – Medizinische Informatik

TREATMENT PLANNINGExternal beam radiotherapy: usually carried out with more than one beam

Exceptions for single field treatments is therapy utilising conventional X rays and electron beam therapy.

Modern RT is carried out with a variety of beam energies and field sizes under one of two setup conventions

In an SSD setup, the distance from the source to the surface of the patient is kept constant for all beams,

For an SAD setup the center of the target volume is placed at the machine isocenter.

Field sizes: from small circular fields used in radiosurgerythrough standard rectangular and irregular fields to very large fields used for total body irradiations.

Page 10: Treatment planning principles · ÎIn an SSD setup, the distance from the source to the surface of the patient is kept constant for all beams, ÎFor an SAD setup the center of the

10Vo Planung – Medizinische Informatik

TREATMENT PLANNINGAn optimal dose distribution can only be achieved if different technical therapy options are utilized. The following parameters have an influence on dose distributions:

Beam quality (x-rays, high energy photon or electron beams, neutrons, heavy charged particles) and beam energyBeam incidence (single field, multiple fields, rotational therapy, conformal radiotherapy, stereotactic RT, intensity modulated RT)Beam modifiers (wedges, bolus, multileaf collimators, individually shaped blocks, …Dose modifications (boost, inhomogeneous dose distribution,..)

The treatment technique itself depends on the underlying basis for treatment planning or technique for patient data acquisition