TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE

(PPVI)

BASIL D. THANOPOULOS MD, PhD

Director Interventional Cardiology of CHD

“Euroclinic”

ATHENS - GREECE

Page 2: TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE …static.livemedia.gr/livemedia/documents/al17365_us... · restoring pulmonic function Option to intervene earlier, ... disfunctioning

NOTHING TO DECLARE

Page 3: TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE …static.livemedia.gr/livemedia/documents/al17365_us... · restoring pulmonic function Option to intervene earlier, ... disfunctioning

Surgical implantation of valved aortic homografts or heterografts is the initial procedure of choice for patients with tetralogy of Fallot and severe pulmonary stenosis or atresia. This surgical procedure can be performed with low mortality and rate of complications in experienced centers, but valved conduits have limited lifespan, less than 10 years. As a result, the majority of patients with right ventricular outflow tract conduits will undergo multiple re-operations with increased complexity and surgical risk as a result of conduit stenosis and/or insufficiency, particularly, if the initial surgery was performed early in life.

TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE

70000 worldwide

Conduit failure

Page 4: TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE …static.livemedia.gr/livemedia/documents/al17365_us... · restoring pulmonic function Option to intervene earlier, ... disfunctioning

TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE

RATIONALE

Help to reduce the total number of surgeries over the patient’s lifetime by postponing time to surgery while

restoring pulmonic function

Option to intervene earlier, providing better outcomes for patients while avoiding

surgical complications

Page 5: TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE …static.livemedia.gr/livemedia/documents/al17365_us... · restoring pulmonic function Option to intervene earlier, ... disfunctioning

TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE

Available pulmonary valves

Melody® valve

Edwards – Sapien THV

Page 6: TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE …static.livemedia.gr/livemedia/documents/al17365_us... · restoring pulmonic function Option to intervene earlier, ... disfunctioning

Melody ® transcatheter pulmonary valve

TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE

Page 7: TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE …static.livemedia.gr/livemedia/documents/al17365_us... · restoring pulmonic function Option to intervene earlier, ... disfunctioning

The percutaneous implantation of pulmonary valve in patients with dysfunctioning RV to pulmonary artery conduits is considered to be the most exciting advancement in interventional pediatric cardiology the last 5 years.

TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE

BONHOEFER ET AL 2000

12-Year-old boy with stenosis

and insufficiency of a RV-PA

conduit

Page 8: TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE …static.livemedia.gr/livemedia/documents/al17365_us... · restoring pulmonic function Option to intervene earlier, ... disfunctioning

TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE

CONDUIT FAILURE (Stenosis + insufficiency)

Progressive RV dilation can lead to eventual heart failure

Enlarged RV can be arrhythmogenic-AF

RV dysfunction can ultimately lead to LV dysfunction

RV failure can lead to early mortality

Timely intervention can save RV function and regress dilatation

Page 9: TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE …static.livemedia.gr/livemedia/documents/al17365_us... · restoring pulmonic function Option to intervene earlier, ... disfunctioning

TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE

INDICATIONS

A. Patients with stenotic and/or regurgitant prosthetic right ventricular outflow tract (RVOT) conduits with a clinical indication for invasive or surgical intervention.

1. Doppler gr 40 mm Hg

2. Moderate to Severe PR

3. RVEDV 150-170 ml/m²

4.RV Fractional area 40%

B. Existence of a full RVOT conduit ≥ 16 mm/ 22 mm when originally implanted.

Page 10: TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE …static.livemedia.gr/livemedia/documents/al17365_us... · restoring pulmonic function Option to intervene earlier, ... disfunctioning

TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE

RESULTS (Melody valve implanters: MS Vienna 2015) *

1000 pts-FU 2006-2013 (694:1Y FU)

RVOT obstruction: 40%

PR: 20%

Mixed: 30% *= 8000 Pts

Page 11: TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE …static.livemedia.gr/livemedia/documents/al17365_us... · restoring pulmonic function Option to intervene earlier, ... disfunctioning

TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE

RESULTS (Melody valve implanters: MSR Vienna 2015) *

No significant residual gradient

PR: p 0.01 (+-++/ 72 months)

RVEDV + RVESV

Exercise capacity

Page 12: TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE …static.livemedia.gr/livemedia/documents/al17365_us... · restoring pulmonic function Option to intervene earlier, ... disfunctioning

TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE

COMPLICATIONS (Melody valve implanters: MSR Vienna 2015)

Mortality

Early: 1 pt Late: 6 pts

Homograft rupture: 3 pts

Dislodgment of the stented valve: 2 pts

Stent fracture: 8-20%-<5%

Bacterial endocarditis: 5% (2.4% AR)

Coronary compression: 2 pts (4.7-6%)

Rare P.Edema PA injury

Page 13: TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE …static.livemedia.gr/livemedia/documents/al17365_us... · restoring pulmonic function Option to intervene earlier, ... disfunctioning

TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE

Stent fracture: 8-20%

Page 14: TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE …static.livemedia.gr/livemedia/documents/al17365_us... · restoring pulmonic function Option to intervene earlier, ... disfunctioning

Mortality

Early: 1 pt Late: 4 pts

Page 15: TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE …static.livemedia.gr/livemedia/documents/al17365_us... · restoring pulmonic function Option to intervene earlier, ... disfunctioning

AEPC 2011

TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE

PPVI Versus SI/R

Page 16: TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE …static.livemedia.gr/livemedia/documents/al17365_us... · restoring pulmonic function Option to intervene earlier, ... disfunctioning

26 mm 23 mm

TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE

22-24F DS

Page 17: TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE …static.livemedia.gr/livemedia/documents/al17365_us... · restoring pulmonic function Option to intervene earlier, ... disfunctioning

TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE

EDWARDS – SAPIEN VALVE

GLOBAL EXPERIENCE:36 PTS Kenny et al (JACC 2011)

Results

Success: 31/36 (86%) Dgr=15-20 mmHg No significant PR

Complications: Embolization: 3 pts

VF: 2 pts Follow-up: 6 m No valve failure (1 pt)

No stent fractures

Page 18: TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE …static.livemedia.gr/livemedia/documents/al17365_us... · restoring pulmonic function Option to intervene earlier, ... disfunctioning

GREEK EXPERIENCE

25 patients

(Stenotic – regurgitant)

Native RVOT: 2 pt

Dysfunctioning aortic homograph: 19 pts

TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE

Gore-tex® conduit: 1pt

RVOT patch: 1 pt

Magna Ease valve : 1 pt

Hancock conduit: 1 pt

Age 8-31 years

Page 19: TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE …static.livemedia.gr/livemedia/documents/al17365_us... · restoring pulmonic function Option to intervene earlier, ... disfunctioning

TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE

EDWARDS – SAPIEN VALVE

Comparative outcome of the Edwards-Sapien and Melody valve

Faza et al (CCI 2013)

Melody valve (13 pts) ES valve 20 pts)

Men RSG =11.2 mm Hg Men RSG =11.2 mm Hg

Stent migration 1 pt Stent migration -

No deaths

Page 20: TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE …static.livemedia.gr/livemedia/documents/al17365_us... · restoring pulmonic function Option to intervene earlier, ... disfunctioning

TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE

GREEK EXPERIENCE

Surgical history

TGA + PS: 3 pts

Ross procedure: 4 pts

Previous surgical conduit replacements 2-3 : 8 pts

TOF + APV: 2 pts

TOF + PA: 16 pts

Page 21: TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE …static.livemedia.gr/livemedia/documents/al17365_us... · restoring pulmonic function Option to intervene earlier, ... disfunctioning

TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE

GREEK EXPERIENCE

RVOT obstruction: 14 pts

Mixed (PR): 9 pts

PR : 2 pts Hemodynamic data

PGr: 40-110 mm Hg PR: + -+++

Dysfunctioning RV 12 pts: (EF<50%)

Page 22: TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE …static.livemedia.gr/livemedia/documents/al17365_us... · restoring pulmonic function Option to intervene earlier, ... disfunctioning

TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE

FOLLOW-UP

Clinical examination

2-D + Doppler echocardiography

ECG – Holter monitoring

Chest x-ray

Cardiac MRI (MSCT)

Page 23: TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE …static.livemedia.gr/livemedia/documents/al17365_us... · restoring pulmonic function Option to intervene earlier, ... disfunctioning

TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE

GREEK EXPERIENCE

Page 24: TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE …static.livemedia.gr/livemedia/documents/al17365_us... · restoring pulmonic function Option to intervene earlier, ... disfunctioning

TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE

GREEK EXPERIENCE

Page 25: TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE …static.livemedia.gr/livemedia/documents/al17365_us... · restoring pulmonic function Option to intervene earlier, ... disfunctioning

TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE

GREEK EXPERIENCE

Page 26: TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE …static.livemedia.gr/livemedia/documents/al17365_us... · restoring pulmonic function Option to intervene earlier, ... disfunctioning

TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE

GREEK EXPERIENCE

Results

PGr: 0 – 25 mm Hg

PR: 0 (23 pts) – Trivial (2pts)

Complications Prestent embolization: 2 pts

Page 27: TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE …static.livemedia.gr/livemedia/documents/al17365_us... · restoring pulmonic function Option to intervene earlier, ... disfunctioning

TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE

Greek experience

Page 28: TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE …static.livemedia.gr/livemedia/documents/al17365_us... · restoring pulmonic function Option to intervene earlier, ... disfunctioning

TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE

GREEK EXPERIENCE

Page 29: TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE …static.livemedia.gr/livemedia/documents/al17365_us... · restoring pulmonic function Option to intervene earlier, ... disfunctioning

TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE

GREEK EXPERIENCE

Page 30: TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE …static.livemedia.gr/livemedia/documents/al17365_us... · restoring pulmonic function Option to intervene earlier, ... disfunctioning

TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE

GREEK EXPERIENCE

Page 31: TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE …static.livemedia.gr/livemedia/documents/al17365_us... · restoring pulmonic function Option to intervene earlier, ... disfunctioning

TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE

GREEK EXPERIENCE

Page 32: TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE …static.livemedia.gr/livemedia/documents/al17365_us... · restoring pulmonic function Option to intervene earlier, ... disfunctioning

TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE

GREEK EXPERIENCE

Page 33: TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE …static.livemedia.gr/livemedia/documents/al17365_us... · restoring pulmonic function Option to intervene earlier, ... disfunctioning

TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE

GREEK EXPERIENCE

Page 34: TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE …static.livemedia.gr/livemedia/documents/al17365_us... · restoring pulmonic function Option to intervene earlier, ... disfunctioning

TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE

GREEK EXPERIENCE

Page 35: TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE …static.livemedia.gr/livemedia/documents/al17365_us... · restoring pulmonic function Option to intervene earlier, ... disfunctioning
Page 36: TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE …static.livemedia.gr/livemedia/documents/al17365_us... · restoring pulmonic function Option to intervene earlier, ... disfunctioning

TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE

GREEK EXPERIENCE

Native RVOT-TGA

Page 37: TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE …static.livemedia.gr/livemedia/documents/al17365_us... · restoring pulmonic function Option to intervene earlier, ... disfunctioning

TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE

HANCOCK VALVE CONDUIT

Page 38: TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE …static.livemedia.gr/livemedia/documents/al17365_us... · restoring pulmonic function Option to intervene earlier, ... disfunctioning

TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE

EDWARDS – SAPIEN VALVE

Page 39: TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE …static.livemedia.gr/livemedia/documents/al17365_us... · restoring pulmonic function Option to intervene earlier, ... disfunctioning

TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE

GREEK EXPERIENCE

STENT EXPANSION

Page 40: TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE …static.livemedia.gr/livemedia/documents/al17365_us... · restoring pulmonic function Option to intervene earlier, ... disfunctioning

TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE

HANCOCK VALVE CONDUIT

Page 41: TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE …static.livemedia.gr/livemedia/documents/al17365_us... · restoring pulmonic function Option to intervene earlier, ... disfunctioning

GREEK EXPERIENCE

Complications

TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE

Page 42: TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE …static.livemedia.gr/livemedia/documents/al17365_us... · restoring pulmonic function Option to intervene earlier, ... disfunctioning

TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE

GREEK EXPERIENCE

Page 43: TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE …static.livemedia.gr/livemedia/documents/al17365_us... · restoring pulmonic function Option to intervene earlier, ... disfunctioning

TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE

Page 44: TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE …static.livemedia.gr/livemedia/documents/al17365_us... · restoring pulmonic function Option to intervene earlier, ... disfunctioning

TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE

Medronic native outflow tract device

Page 45: TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE …static.livemedia.gr/livemedia/documents/al17365_us... · restoring pulmonic function Option to intervene earlier, ... disfunctioning

TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE

Page 46: TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE …static.livemedia.gr/livemedia/documents/al17365_us... · restoring pulmonic function Option to intervene earlier, ... disfunctioning

TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE

Venous-P Valve

Page 47: TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE …static.livemedia.gr/livemedia/documents/al17365_us... · restoring pulmonic function Option to intervene earlier, ... disfunctioning

TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE

Melody ® transcatheter pulmonary valve

CONCLUSIONS I

Transcatheter replacement of pulmonary valve using

the Medronic® pulmonary valve implant or in some patients the Endwards sapien valve is a safe and effective alternative to open heart surgery for the treatment of selected patients with dysfunctioning valve conduits. Close collaboration between Congenital Cardiac Surgery and Cardiology teams are key to long-term success!!

Page 48: TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE …static.livemedia.gr/livemedia/documents/al17365_us... · restoring pulmonic function Option to intervene earlier, ... disfunctioning

TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE

CONCLUSIONS II

RV to PA conduits are currently a great first step to pulmonary blood-flow repair.

•The Melody™ in its current design is not intended to replace the surgically placed conduit – but is instead intended to extend the life of that conduit and reduce the number of operations over the total lifetime of patients . Further research is required to improve the implantation techniques and outcome of

•percutaneous pulmonary valve replacement therapy and to extend it to all patients with a clinical indication to

•delay or avoid open heart surgery.

Page 49: TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE …static.livemedia.gr/livemedia/documents/al17365_us... · restoring pulmonic function Option to intervene earlier, ... disfunctioning

TRANSCATHETER TRICUSPID VALVE IN VALVE IMPLANTATION

Godart et al TTVI : A multicenter French study Archives of Cardiovascular Disease 2014

Complications Embolization Endocarditis CAVB Valve failure Death

Page 50: TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE …static.livemedia.gr/livemedia/documents/al17365_us... · restoring pulmonic function Option to intervene earlier, ... disfunctioning

TRANSCATHETER REPLACEMENT OF THE TRICUSPID VALVE

CONCLUSIONS II

Transcatheter tricuspid valve in valve implantation

using the Medronic® pulmonary valve or the Edwards Sapien valve implants is a new techique that can be used as an effective alternative to surgical valve replacemt in selective high risk patients with disfunctioning biological tricuspid valves.However, further studies are required to document its efficasy, safety and long-term results in a larger patient population.

Page 51: TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE …static.livemedia.gr/livemedia/documents/al17365_us... · restoring pulmonic function Option to intervene earlier, ... disfunctioning

TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE

RESULTS (Melody valve-Bonhoeffer: PICS 2008)

Page 52: TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE …static.livemedia.gr/livemedia/documents/al17365_us... · restoring pulmonic function Option to intervene earlier, ... disfunctioning

TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE

Page 53: TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE …static.livemedia.gr/livemedia/documents/al17365_us... · restoring pulmonic function Option to intervene earlier, ... disfunctioning
Page 54: TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE …static.livemedia.gr/livemedia/documents/al17365_us... · restoring pulmonic function Option to intervene earlier, ... disfunctioning
Page 55: TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE …static.livemedia.gr/livemedia/documents/al17365_us... · restoring pulmonic function Option to intervene earlier, ... disfunctioning

TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE

EDWARDS – SAPIEN VALVE

20F DS? Cobalt stent

Page 56: TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE …static.livemedia.gr/livemedia/documents/al17365_us... · restoring pulmonic function Option to intervene earlier, ... disfunctioning

TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE

CONCLUSIONS II

RV to PA conduits are currently a great first step to pulmonary blood-flow repair. The Melody™ in its current design is not intended to replace the surgically placed conduit – but is instead intended to extend the life of that conduit and reduce the number of operations over the total lifetime of patients . Close collaboration between Congenital Cardiac Surgery and Cardiology teams are key to long-term success!!

Page 57: TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE …static.livemedia.gr/livemedia/documents/al17365_us... · restoring pulmonic function Option to intervene earlier, ... disfunctioning

TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE

COMPLICATIONS (Melody valve-Bonhoeffer: PICS 2008)

RPA obstruction: 1pt

CA compression: 1pt

Guide wire perforation: 1pt

Page 58: TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE …static.livemedia.gr/livemedia/documents/al17365_us... · restoring pulmonic function Option to intervene earlier, ... disfunctioning

GREEK EXPERIENCE

TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE

Stent fractures=7% Endocarditis=3% Conduit tears=5%

Page 59: TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE …static.livemedia.gr/livemedia/documents/al17365_us... · restoring pulmonic function Option to intervene earlier, ... disfunctioning
Page 60: TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE …static.livemedia.gr/livemedia/documents/al17365_us... · restoring pulmonic function Option to intervene earlier, ... disfunctioning

TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE

Page 61: TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE …static.livemedia.gr/livemedia/documents/al17365_us... · restoring pulmonic function Option to intervene earlier, ... disfunctioning

TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE

Future

Page 62: TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE …static.livemedia.gr/livemedia/documents/al17365_us... · restoring pulmonic function Option to intervene earlier, ... disfunctioning

TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE

Melody valve

Page 63: TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE …static.livemedia.gr/livemedia/documents/al17365_us... · restoring pulmonic function Option to intervene earlier, ... disfunctioning

TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE

GREEK EXPERIENCE

Gore-Tex® conduit; 1pt

Page 64: TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE …static.livemedia.gr/livemedia/documents/al17365_us... · restoring pulmonic function Option to intervene earlier, ... disfunctioning
Page 65: TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE …static.livemedia.gr/livemedia/documents/al17365_us... · restoring pulmonic function Option to intervene earlier, ... disfunctioning

TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE

PATIENT POPULATION

Pulmonary stenosis

TOF + PA

Truncus arteriosus

TOF physiology TGA DORV

70000 worldwide

Conduit failure

Page 66: TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE …static.livemedia.gr/livemedia/documents/al17365_us... · restoring pulmonic function Option to intervene earlier, ... disfunctioning

TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE

The lifetime for the Melody® device is 2 years.

Page 67: TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE …static.livemedia.gr/livemedia/documents/al17365_us... · restoring pulmonic function Option to intervene earlier, ... disfunctioning

TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE

PRE-CATHETER ASSESMENT

History

ECG

Holter monitoring

Echocardiography 2/3 D + Doppler

MRI Cine Flow studies

Page 68: TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE …static.livemedia.gr/livemedia/documents/al17365_us... · restoring pulmonic function Option to intervene earlier, ... disfunctioning

TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE

IMPLANTATION TECHIQUE

BIB catheter 22 F delivery

sheath

Diagnostic cardiac cath

Femoral venous approach

Page 69: TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE …static.livemedia.gr/livemedia/documents/al17365_us... · restoring pulmonic function Option to intervene earlier, ... disfunctioning

TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE

IMPLANTATION TECHIQUE

Prestenting

Redilation (HPB) – Residual gradient

Balloon inflation in the RVOT (exclude CA compression)

Pulmonary valve-in-valve implantation (Residual stenosis-stent fracture)

Page 70: TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE …static.livemedia.gr/livemedia/documents/al17365_us... · restoring pulmonic function Option to intervene earlier, ... disfunctioning

TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE

POST-PROCEDURAL EVALUATION

Hemodynamic evaluation

Biplane cineangiography

2-D + Doppler echocardiography

Page 71: TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE …static.livemedia.gr/livemedia/documents/al17365_us... · restoring pulmonic function Option to intervene earlier, ... disfunctioning

TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE

GREEK EXPERIENCE

Page 72: TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE …static.livemedia.gr/livemedia/documents/al17365_us... · restoring pulmonic function Option to intervene earlier, ... disfunctioning

TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE

GREEK EXPERIENCE

Gore-Tex® conduit; 1pt

Page 73: TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE …static.livemedia.gr/livemedia/documents/al17365_us... · restoring pulmonic function Option to intervene earlier, ... disfunctioning

TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE

GREEK EXPERIENCE

Page 74: TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE …static.livemedia.gr/livemedia/documents/al17365_us... · restoring pulmonic function Option to intervene earlier, ... disfunctioning
Page 75: TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE …static.livemedia.gr/livemedia/documents/al17365_us... · restoring pulmonic function Option to intervene earlier, ... disfunctioning

EDWARDS – SAPIEN VALVE

Implantation technique

Retroflex I delivery system

Previously placement of stent For accurate positioning

Page 76: TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE …static.livemedia.gr/livemedia/documents/al17365_us... · restoring pulmonic function Option to intervene earlier, ... disfunctioning

TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE

GREEK EXPERIENCE

Page 77: TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE …static.livemedia.gr/livemedia/documents/al17365_us... · restoring pulmonic function Option to intervene earlier, ... disfunctioning
Page 78: TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE …static.livemedia.gr/livemedia/documents/al17365_us... · restoring pulmonic function Option to intervene earlier, ... disfunctioning
Page 79: TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE …static.livemedia.gr/livemedia/documents/al17365_us... · restoring pulmonic function Option to intervene earlier, ... disfunctioning

TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE

18mm Contegra modified-bovine jugular vein with

valve segment

Melody ® transcatheter pulmonary valve

Mounted on a NuMed Platinum Iridium Stent 28 mm length, Crimped down to 6mm-re-

expanded 18mm up to 22mm

Page 80: TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE …static.livemedia.gr/livemedia/documents/al17365_us... · restoring pulmonic function Option to intervene earlier, ... disfunctioning

EnsembleTM Delivery System

Sheath Balloons (currently

covered by the sheath)

Catheter

Shaft

Hemostasis

Adaptor Access

Site

Guidewire Port Outer Balloon Hub

Inner Balloon Hub

Stopcock

Balloon

size

Indicator

Tip

Marker – Sheath

Uncovered

Marker – Sheath

Covered

Sheath-Hemostasis Valve

TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE

Page 81: TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE …static.livemedia.gr/livemedia/documents/al17365_us... · restoring pulmonic function Option to intervene earlier, ... disfunctioning
Page 82: TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE …static.livemedia.gr/livemedia/documents/al17365_us... · restoring pulmonic function Option to intervene earlier, ... disfunctioning

TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE

GREEK EXPERIENCE

22 patients

Age: 8 – 31 years

Weight: 18 – 58 Kg

Dysfunctioning aortic homograph: 17 pts (Stenotic – regurgitant)

Native RVOT: 3 pt

Gore-Tex® conduit: 1pt

RVOT patch : 1pt

Page 83: TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE …static.livemedia.gr/livemedia/documents/al17365_us... · restoring pulmonic function Option to intervene earlier, ... disfunctioning
Page 84: TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE …static.livemedia.gr/livemedia/documents/al17365_us... · restoring pulmonic function Option to intervene earlier, ... disfunctioning

TRANSCATHETER REPLACEMENT OF THE PULMONARY VALVE

CONTANDICATIONS

1.Venous anatomy unable to accommodate 22 F introducer sheath

2. Implantation in the left heart

3. Unfavorable RVOT for good stent anchorage Large/Severely stenotic

4. Active infection/endocarditis

5. Pregnancy? 6. Allergy to aspirin or heparin