2
Application For determination of the specific surface (Blaine value) of powders, especially suitable for the permanent control of manufacturing characteristics in the daily laboratory work. This measuring procedure serves as practice oriented alternative for the standard procedure according to DIN EN 196-6 as well as ASTM C 204. Advantages - an alternating operation with two load cells enables higher test frequency - immediate readout of the specific surface on the apparatus - height adjustable optical sensors for optimal results for the different samples to be tested - density, initial weight and porosity selectable acc. to the specific samples - calibration via reference materials - five different calibration factors savable - runtime indication resolution : 0.01 sec - calculation of the Blaine values incl. standard deviation of up to 99 samples - connection to balance, dial gauge and printer are available as options - improved accuracy as one can select tests as much as desired for one sample; the average value is always calculated by the software - Blaine value calculation is adaptable to the tempera- ture by a built-in sensor - the possibility of „zero“-test avoids measuring errors because of selfcompacting of the powder - connection to PC for data exchange to LIMS and other laboratory data recording systems through the universal file interface software ToniDAT Design The device consists of a microprocessor control unit and a housing with measuring unit. Via magnet valve and membrane pump, a vacuum is achieved by using an U-tube manometer with contactless, adjustable, optical level sensors. This vacuum will be balanced by the air stream through the powder bed in the permeability cell. Working Principle Using the plunger the powder will be compac- ted within the permeability cell to a defined volu- me. (with an optional electronic dial gauge, the Blaine value can be corrected to the actual volume). After placing the filled permeability cell and entering the samples data for the specific test into the control unit the test will be automatically performed and evaluated. The Blaine value is calculated as average value of the values measured in the single cell according to recent standards (e.g. DIN EN 196-6, ASTM C 204). Standard Delivery includes Electronic air permeability apparatus ToniPERM 6565 Round filters 7207.004 1 box, 41 mm Ø, 500 count Dust filters 7207.003 1 box, 13 mm Ø, 100 count Tamper 6565.011 for sample preparation and to eject the sample Product Information Model 6565 Electronic Air Permeability Apparatus ToniPERM PI908.E 02.14 03_908_6565_ToniPERM_PI_E.pdf Electronic Air Permeability Apparatus ToniPERM Technik T oni Building Materials Testing

Toni Technik Building Materials Testinguniversalcalibrations.com/yahoo_site_admin/assets/docs/6565... · Application For determination of the specific surface (Blaine value) of powders,

Embed Size (px)

Citation preview

ApplicationFor determination of the specific surface (Blaine value) of powders, especially suitable for the permanent control of manufacturing characteristics in the daily laboratory work. This measuring procedure serves as practice oriented alternative for the standard procedure according to DIN EN 196-6 as well as ASTM C 204.

Advantages - an alternating operation with two load cells enables

higher test frequency

- immediate readout of the specific surface on the apparatus

- height adjustable optical sensors for optimal results for the different samples to be tested

- density, initial weight and porosity selectable acc. to the specific samples

- calibration via reference materials

- five different calibration factors savable

- runtime indication resolution : 0.01 sec

- calculation of the Blaine values incl. standard deviation of up to 99 samples

- connection to balance, dial gauge and printer are available as options

- improved accuracy as one can select tests as much as desired for one sample; the average value is always calculated by the software

- Blaine value calculation is adaptable to the tempera-ture by a built-in sensor

- the possibility of „zero“-test avoids measuring errors because of selfcompacting of the powder

- connection to PC for data exchange to LIMS and other laboratory data recording systems through the universal file interface software ToniDAT

DesignThe device consists of a microprocessor control unit and a housing with measuring unit. Via magnet valve and membrane pump, a vacuum is achieved by using an U-tube manometer with contactless, adjustable, optical level sensors. This vacuum will be balanced by the air stream through the powder bed in the permeability cell.

Working PrincipleUsing the plunger the powder will be compac-ted within the permeability cell to a defined volu-me. (with an optional electronic dial gauge, the Blaine value can be corrected to the actual volume). After placing the filled permeability cell and entering the samples data for the specific test into the control unit the test will be automatically performed and evaluated. The Blaine value is calculated as average value of the values measured in the single cell according to recent standards (e.g. DIN EN 196-6, ASTM C 204).

Standard Delivery includesElectronic air permeability apparatus ToniPERM 6565

Round filters 7207.0041 box, 41 mm Ø, 500 count

Dust filters 7207.0031 box, 13 mm Ø, 100 count

Tamper 6565.011for sample preparation and to eject the sample

Product Information Model 6565Electronic Air Permeability Apparatus ToniPERM

PI9

08.E

02.

14

03_908_6565_ToniPERM_PI_E.pdf

Electronic Air Permeability Apparatus ToniPERM

TechnikToni Building Materials Testing

Optional Accessories

Software 6565.001for connection to a local printer Software 6565.002for connection to a local PC and LIMS

Software date connection 6565.002

Electronic dial gauge 6565.003for compensation of the permeability cell volume

Anti-twisting device 6565.004for fixing the plunger against turning

Second permeability cell 6565.007to increase the test frequency and for dual testing

Temperature sensor 6565.008for temperature specific test data correction

Software for the connection of a scale 6565.009

Barcode reader 0510.621for sample identification with laboratory information systems

Special voltage (115V) 6565.015

Standard sand sample I 7207.012fineness approx. 2800 cm²/g, officially certified

Standard sand sample II 7207.013fineness approx. 4000 cm²/g, officially certified

For further information please contactToni Technik Baustoffprüfsysteme GmbH Gustav-Meyer-Allee 25 13355 Berlin Tel: +49 (0) 30/464039-21 Fax: +49 (0) 30/464039-22 Email: [email protected]

Technical DataPower supply voltage frequency power

VHzVA

230 (115 optional)50 / 60ca. 15

Volume of permeability cell cm³ 74

Dimensions (measuring device/ control unit) width depth height

mmmmmm

240/220320/330310/380

Weight gross/net kg 25/23

Product Information Model 6565Electronic Air Permeability Apparatus ToniPERM

All data at ambient temperature. Subject to technical modification.

TechnikToni Building Materials Testing