Upload
others
View
7
Download
0
Embed Size (px)
Citation preview
Calibration Discussion
Ed Fortner, John Jayne
There are 3 fundamental calibrations for the
AMS
• Flow rate
• Ionization efficiency
• Particle size calibration
g
m3
mass
volume
From Mass Spectrometer
Ionization Efficiency calibration
From volumetric flow rate
Particle Mass Loading
g
m3
g
m3
mass
volume
mass
volume
From Mass Spectrometer
Ionization Efficiency calibration
From volumetric flow rate
Particle Mass Loading
Particle Size
pTOF to Dva
IE Calibration
• We have two approaches
– Single particle
– Total mass based method
Single Particle Based IE Calibration
Time
Sig
na
l
Single particle pulses
Particle threshold set
above single ion level
Single ions above electronic noise level
Time
Am
ps (
Co
ulo
mb
s/t
ime)
Average single ion pulse
Average single particle pulse = Ions per particle (IPP)
Ionization Efficiency = IPP/Molecules per Particle
• In principal single particle approach is
good.
– Avoids complications associated with DMA-
Q2s and lens transmission.
– Avoids errors that the CPC might introduce
However in practice there are details which if not followed WILL result in an inaccurate IE determination You need to have a low # conc of AN particles usually 200 or
less in order to ensure that you are only counting a single particle per extraction should have less than 5% hit rate
– Count at least 300 particles to get good counting statistics
– Selection of threshold is critical too low you count noise too high you bias toward higher signal particles
– Use as large a particle as practical to enhance IPP.
– Important to maximize time resolution to better resolve vaporization event. This means shorter pulser period. Pulser time effects IE. Need to correct.
More BFSP Considerations
– Get the RIE for NH4 using MS data. Don’t rely
on BFSP derived RIE for NH4
– Use a HEPA filter to ensure no false counts;
Also a tool to use for determining how low
threshold can go
– This is only a single point calibration
AMS DMA
CPC
Atomizer
Aerosol
Diluter
Setup for CPC-Mass Based IE Determination
By-pass
Input Mass = x Volume(size) x Number Instrument Response
filter
Plot Instrument Response vs Input Mass
Drier
Mixing Tube
Very Important! Very Important!!
Mass Based IE Determination
12
10
8
6
4
2
0
Mass r
eport
ed b
y V
1.5
ToF
DA
Q (
ug/m
3)
2520151050
Calculated/Input AN mass (based on CPC, ug/m3)
Dmob 300 nm
250 nm
intercept = 0.068 ± 0.041slope = 0.493 ± 0.003
Fast_IE_Cal.pxp
Slope is the IE correction factor
CPC-Mass Based IE Determination
Lens transmission limitation
Multiple charged diameters 12
10
8
6
4
2
0
Mass r
eport
ed b
y V
1.5
ToF
DA
Q (
ug/m
3)
2520151050
Calculated/Input AN mass (based on CPC, ug/m3)
Dmob 200 nm
250 nm
300 nm
400 nm
intercept = 0.068 ± 0.041slope = 0.493 ± 0.003
Fast_IE_Cal.pxp
Nit
rate
Sig
nal
Calculated Mass
Recommendations
• If Comparing BFSP IE to CPC based IE it is important to
consider differences in particle size between methods
BFSP optimum is 400 nm and CPC based IE optimum is
300 nm.
Do Both Calibrations
Velocity Calibration
12
8
4
0
Sig
na
l (b
its)
0.0060.0050.0040.0030.002
p-TOF (s)
NH4NO3 Dmob 100 nm 250 nm 450 nm 350 nm
d600/ship/U_Tokyo/velocity.pxp
5
6
7
8
9100
2
3
Ve
locity (
m/s
)
101
2 4 6 8
102
2 4 6 8
103
2 4 6 8
104
Aerodynamic Diameter (nm)
PSL, F=1.43 cc/s
NO3, F=1.41 cc/s
p_0 = 658 ± 0 fixedp_1 = 4.4553 ± 0.353p_2 = 0.43986 ± 0.00864p_3 = 16.449 ± 2.2
d600/ship/U_Tokyo/velocity.pxp
Particle Velocity Calibration
Daero = Dgeo * ρ* Shape Fac
Sample known size particles
and calculate a velocity…
Velocity = flight path / TOF
What’s wrong with this calibration?
Velocity calibration data should span maximum
measurable range
Extrapolation of fit beyond data points can introduce errors
Critical data points
Size Calibration Summary
• Use caution extrapolating velocity curve.
• Document lens pressure (flow) during a
calibration.
• If your lens pressure changes check if orifice is
clogged don’t take different sizes at different
flow rates.
• Watch for a lens pressure change when
connecting the AMS inlet to a sampling system.
Flow Rate Calibration
Flow Meter
Low pressure drop
flow meter
Gilibrator™ or
equivalent soap film
meter 10 Torr lens
pressure gauge
P
Aerodynamic Lens
Fine metering
valve
AMS Flow Rate Calibration Setup
Poiseuille equation relates volumetric
laminar flow (F) to a pressure drop
(ΔP) across the tube
Flow into
vacuum
Replace 100 um pin hole
with fine metering valve
SS-SS4.
Flow Calibration
Depends on Pressure and Temperature
3.0
2.5
2.0
1.5
1.0
0.5
0.0
Flo
w (
cc/s
)
2.52.01.51.00.50.0
Pressure (torr)
U. Tokyo Cal at ARI
NOAA Cal at ARI
NOAA Cal at Boulder
NOAA Scaled to 760 torr
Always document ambient pressure and temperature
100 µm leak
Flow Rate Calibration Summary
• In principal you only need to do this once as long as
you document ambient pressure and temperature
condition when you do the calibration (also assumes
you are always sampling air).
• What you are measuring/calibrating here is a
property of the aerodynamic lens via the Poiseuille
relation ship.
• When reporting µg m-3, always reference the T&P
conditions you use in the “m3”
• The flow rate calibration is independent of critical
orifice. Critical orifice is removed during calibration
Lens Alignment
• This is often not checked.
• Generally, this does not need to be adjusted but if its off and not checked your data may not be quantified correctly.
• Recommend checking this after shipping the AMS.
• Don’t be afraid to do this (Turn TPS off to be cautious)
Documentation
• Recommend save all of your cals in a cal
folder on the AMS computer.
• So when the next person comes along
they have a record of how this AMS is
running
END PRESENTATION
Converting your CPC number
concentration into an IE
• an_molec = numconc
* pi/6 * (300e-7)^3 *
1.72 * 0.8*1.35 / 80 *
6.022e23