Surgical Technique

The Systematic Solution for Craniomaxillofacial Indications and Displacement Osteotomies

Compact Midface

Page 2: The Systematic Solution for Craniomaxillofacial Indications and …synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes... · 2018-07-07 · implants for craniomaxillofacial indications

Image intensifier control

This description alone does not provide sufficient background for direct use of DePuy Synthes products. Instruction by a surgeon experienced in handling these products is highly recommended.

Processing, Reprocessing, Care and MaintenanceFor general guidelines, function control and dismantling of multi-part instruments, as well as processing guidelines for implants, please contact your local sales representative or refer to:http://emea.depuysynthes.com/hcp/reprocessing-care-maintenanceFor general information about reprocessing, care and maintenance of Synthes reusable devices, instrument trays and cases, as well as processing of Synthes non-sterile implants, please consult the Important Information leaflet (SE_023827) or refer to: http://emea.depuysynthes.com/hcp/reprocessing-care-maintenance

Page 3: The Systematic Solution for Craniomaxillofacial Indications and …synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes... · 2018-07-07 · implants for craniomaxillofacial indications

Compact Midface Surgical Technique DePuy Synthes 1

Table of Contents

Introduction Compact Midface 2

Features and benefits 4

AO Principles 5

Intended Use, Indications, Contraindications, 6 General Adverse Events, Device Specific Adverse Events, Warnings, Precautions, and MRI Information

Surgical Technique Trauma Repair and Reconstruction 8

Orbital Plates 11

Le Fort I Fixation 13

Product Information Implants 15

Instruments 16

Sets 17

Ordering Information 18

Page 4: The Systematic Solution for Craniomaxillofacial Indications and …synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes... · 2018-07-07 · implants for craniomaxillofacial indications

1.0 1.3 1.5 2.0 and 2.0 Lock

2 DePuy Synthes Compact Midface Surgical Technique

Compact Midface is a highly flexible modular system that consists of a variety of surgical instruments and implants for craniomaxillofacial indications.

Key features and benefits• Comprehensive solutions based on indications• Large selection of titanium plates and meshplates to

meet the anatomical needs of the patient• Screw diameters in 1.0, 1.3, 1.5 and 2.0 mm available

in self-tapping and self-drilling • 2.0 mm locking screws for midface• Screw clips for quick and easy restocking of screws• All implants available in sterile and non-sterile

Compact MidfaceThe systematic solution for craniomaxillofacial indications and displacement osteotomies

Introduction

Implant strength

Page 5: The Systematic Solution for Craniomaxillofacial Indications and …synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes... · 2018-07-07 · implants for craniomaxillofacial indications

Compact Midface Surgical Technique DePuy Synthes 3

Available as self-drilling and self-tapping with two different recesses: • PlusDrive• Cruciform

Screws

Screws Overview PlusDrive Cruciform

Self-drilling Self-tapping Self-tapping

1.0 2 2 2

1.3 2 2 2

1.5 2 2 2

2.0 2 2 2

2.0 LOCK 2 2

Lag screw 2

System InformationSystem Screw diameter (mm) Drill bit diameter (mm) Emergency screw diameter (mm) Plate thickness (mm)

1.0 1.0 0.7*, 0.76* 1.2 0.5

1.3 1.3 1.0 1.7 0.5

1.5 1.5 1.1 2.0 0.6 malleable, 0.9

2.0 2.0 1.5 2.4 0.6 malleable, 0.9

2.0 LOCK 2.0 1.5 2.4 0.6 malleable, 0.9

* Drill bit diameter 0.7 mm for screw lengths: 2–3 mm Drill bit diameter 0.76 mm for screw lengths: 4–14 mm

The well thought out design of color-coded screw clips al-lows the modules to be quickly and easily filled with screws.

Page 6: The Systematic Solution for Craniomaxillofacial Indications and …synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes... · 2018-07-07 · implants for craniomaxillofacial indications

4 DePuy Synthes Compact Midface Surgical Technique

Features and benefits

Le Fort I osteotomy• Pre-bent maxilla plate 1.5 facitates intraoperative

contouring steps• Advancements from 3–11 mm• Preoperatively determined advancement reliably

achieved

L, Y, T, and Z Shaped plates• L, Y and T plates in various versions and dimensions• Z plates in various dimensions with asymmetrical

design for secure fixation of the maxilla at the lateral pillars

Page 7: The Systematic Solution for Craniomaxillofacial Indications and …synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes... · 2018-07-07 · implants for craniomaxillofacial indications

Compact Midface Surgical Technique DePuy Synthes 5

AO Principles

In 1958, the AO formulated four basic principles, which have become the guidelines for internal fixation1,2. They are:

Anatomic reductionFracture reduction and fixation to restore anatomical relationships. A comprehensive implant and instrument selection offers the ability to address most simple and complex fixation needs.

Stable fixationStability by rigid fixation or splintage, as the personality of the fracture and the injury requires. The Compact Midface plates and screws are optimized to achieve stable bone fixation.

Preservation of blood supplyPreservation of the blood supply to soft tissue and bone by careful handling and gentle reduction techniques.

Early, active mobilizationEarly and safe mobilization of the part and patient. The Compact Midface system implants, combined with AO technique, provide stable fixation enough to allow a functional aftercare.

1 Müller ME, Allgöwer M, Schneider R, Willenegger H (1995) Manual of Internal Fixation. 3rd, expanded and completely revised ed. 1991. Berlin, Heidelberg, New York: Springer

2 Rüedi TP, Buckley RE, Moran CG (2007) AO Principles of Fracture Management. 2nd expanded ed. 2002. Stuttgart, New York: Thieme

Page 8: The Systematic Solution for Craniomaxillofacial Indications and …synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes... · 2018-07-07 · implants for craniomaxillofacial indications

6 DePuy Synthes Compact Midface Surgical Technique

Intended Use, Indications, Contraindications, General Adverse Events, Device Specific Adverse Events, Warnings, Precautions, and MRI Information

Intended UseThe implants (plates and screws) and their instruments are intended for trauma repair and reconstruction of the craniofacial skeleton.

IndicationsThe Compact systems are indicated for selective trauma of the midface and craniofacial skeleton, craniofacial sur-gery, and orthognathic surgery of the midface.

Orbital plates are indicated for trauma repair and recon-struction of the craniofacial skeleton. Specific Indications are: • orbital floor fractures, • medial orbital wall fractures, and • combined orbital floor and medial wall fractures.

ContraindicationsThe systems are contraindicated for use in areas with active or latent infection or insufficient quantity or quality of bone.

General Adverse EventsAs with all major surgical procedures, risks, side effects and adverse events can occur. While many possible reac-tions may occur, some of the most common include:Problems resulting from anesthesia and patient position-ing (e.g. nausea, vomiting, dental injuries, neurological impairments, etc.), thrombosis, embolism, infection, nerve and/or tooth root damage or injury of other criti-cal structures including blood vessels, excessive bleeding, damage to soft tissues incl. swelling, abnormal scar for-mation, functional impairment of the musculoskeletal system, pain, discomfort or abnormal sensation due to the presence of the device, allergy or hypersensitivity reactions, side effects associated with hardware promi-nence, loosening, bending, or breakage of the device, mal-union, non-union or delayed union which may lead to breakage of the implant, reoperation.

Device Specific Adverse EventsDevice specific adverse events include but are not limited to:• Loosening, bending, or breakage of the devices• Non-union, mal-union, or delayed union that may lead

to breakage of the implants• Pain, discomfort, or abnormal sensation due to the

presence of the devices • Adverse tissue reaction/soft tissue irritation• Local infection/systemic infection• Damage to vital organs, surrounding structures,

and/or soft tissues• Peripheral nerve damage• Bone damage, bone fracture, and/or bone necrosis • User injury

Warnings:• These devices can break during use (when sub-

jected to excessive forces or outside the recom-mended surgical technique).

• While the surgeon must make the final decision on removal of the broken part based on associated risk in doing so, we recommend that whenever possible and practical for the patient, the broken part should be removed.

Precaution: Physicians should inform their patients about the implants’ load restrictions and develop a plan for postoperative behavior and increasing physical loads.

Page 9: The Systematic Solution for Craniomaxillofacial Indications and …synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes... · 2018-07-07 · implants for craniomaxillofacial indications

Compact Midface Surgical Technique DePuy Synthes 7

MRI Information

Magnetic Resonance Environment Torque, Displacement and Image Artifacts according to ASTM F 2213-06, ASTM F 2052-14 and ASTM F 2119-07 Non-clinical testing of worst case scenario in a 3 T MRI system did not reveal any relevant torque or displace-ment of the construct for an experimentally measured local spatial gradient of the magnetic field of 5.4 T/m. The largest image artifact extended approximately 20 mm from the construct when scanned using the Gra-dient Echo (GE). Testing was conducted on a 3 T MRI system.

Radio-Frequency-(RF-)induced heating according to ASTM F 2182-11a Non-clinical electromagnetic and thermal simulations of worst case scenario lead to temperature rises of 9.3 °C (1.5 T) and 6 °C (3 T) under MRI Conditions using RF Coils (whole body averaged specific absorption rate [SAR] of 2 W/kg for 15 minutes).

Precautions: The above mentioned test relies on non-clinical testing. The actual temperature rise in the patient will depend on a variety of factors be-yond the SAR and time of RF application. Thus, it is recommended to pay particular attention to the fol-lowing points: • It is recommended to thoroughly monitor patients

undergoing MR scanning for perceived tempera-ture and/or pain sensations.

• Patients with impaired thermoregulation or tem-perature sensation should be excluded from MR scanning procedures.

• Generally, it is recommended to use an MRI sys-tem with low field strength in the presence of con-ductive implants. The employed specific absorp-tion rate (SAR) should be reduced as far as possible.

• Using the ventilation system may further contrib-ute to reduce temperature increase in the body.

Page 10: The Systematic Solution for Craniomaxillofacial Indications and …synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes... · 2018-07-07 · implants for craniomaxillofacial indications

1 DePuy Synthes Compact Midface Surgical Technique

1. Expose and reduce fracture

After completing the preoperative plan, expose the frac-ture or osteotomy site. In trauma reduce the fracture as required.

2. Select and prepare the implant

Select the appropriate plate to the indication. Orient the plate so that the topside is facing out. Cut to length, if necessary.

Precautions:• In order to determine the appropriate amount of

screws needed to achieve stable construct fi xation, the surgeon should consider the fracture size and shape.

• Take care to protect soft tissue from trimmed plate edges.

• Instrument tips may be sharp, handle with care and dispose sharp cuttings in an approved sharps container.

Surgical TechniqueTrauma Repair and Reconstruction

Surgical Technique

top side bony side

Page 11: The Systematic Solution for Craniomaxillofacial Indications and …synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes... · 2018-07-07 · implants for craniomaxillofacial indications

Compact Midface Surgical Technique DePuy Synthes 9

3. Contour the plate

Contour the plate to fit to the patient anatomy using the plate cutter and bending pliers respectively. Ensure the plate is passively adapted to the bone.

Note: An exact match is not required when using locking plates and screws because plate stability is not dependent on plate-to-bone contact when screws are locked.

Precautions:• If contouring is necessary, the surgeon should

avoid bending the device at a screw hole.• Avoid sharp bends, repetitive and reverse bending

as it increases the risk of implant breakage.• Take care to protect soft tissue from trimmed

edges.

4. Position the plate

Place the plate over the fracture or osteotomy site.

Precaution: Confirm that plate positioning, drill bit, and screw length allow for adequate clearance of nerves, tooth buds and/or tooth roots, the edge of the bone, and any other critical structures.

Page 12: The Systematic Solution for Craniomaxillofacial Indications and …synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes... · 2018-07-07 · implants for craniomaxillofacial indications

11 DePuy Synthes Compact Midface Surgical Technique

5. Pre-drilling and screw insertion

Pre-drilling is recommended in complex fractures of the midface and in mandibular regions with thick cortical bone.

If pilot hole is desired, drill the first hole (for appropriate drill bit diameter, see reference chart on page 3) and in-sert the first screw close to the fracture or osteotomy site, and tighten until secure.

Insert the second screw on the opposite side of the frac-ture or osteotomy site, and then all remaining screws following the outlined procedure.

If the screw is inserted with angulation, verify that the screw is safely retained in the plate hole and that the construct profile is not significantly increased.

Precautions: • Confirm that drill bit length and diameter corre-

spond to selected screw length prior to drilling. • Drill speed rate should never exceed 1,800 rpm,

particularly in dense, hard bone. Higher drill speed rates can result in: – thermal necrosis of the bone, – soft tissue burns, – an oversized hole, which can lead to reduced

pullout force, increased ease of the screws strip-ping in bone, suboptimal fixation, and/or the need for emergency screws.

• Avoid damaging the plate threads with the drill.• Always irrigate during drilling to avoid thermal

damage to the bone and ensure drill bit is concen-tric to plate hole, irrigation ensures the removal of debris potentially generate during implantation.

• Take care while drilling as to not damage, entrap, or tear a patient’s soft tissue or damage critical structures, nerves or tooth roots.

• In order to determine the appropriate amount of screws needed to achieve stable construct fixation, the surgeon should consider the fracture size and shape.

• Confirm screw length prior to implantation. • Tighten screws in a controlled manner. Applying

too much torque to the screws may cause screw/plate deformation or bone stripping. If bone be-comes stripped, remove the screw from the bone and replace with an emergency screw.

For Cranial Region Fixation, the following precautions apply:

In order to determine the appropriate amount of fixation for stability, the surgeon should consider the size and shape of the fracture or osteotomy. DePuy Synthes recommends at least three plates when repairing osteo-tomies. Additional fixation is recommended to ensure stability of large fractures and osteotomies. When using mesh for larger defects, additional screws for fixation are recom-mended.

Surgical Technique Trauma Repair and Reconstruction

Page 13: The Systematic Solution for Craniomaxillofacial Indications and …synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes... · 2018-07-07 · implants for craniomaxillofacial indications

Compact Midface Surgical Technique DePuy Synthes 11

1. Select plate design

Select the appropriate plate shape and thickness that best suits the bony anatomy and treatment objective.

2. Adapt plate to the bone

If required, cut and contour the plate to the patient anatomy using the plate cutter and the bending pliers respectively. Ensure that the plate is placed flush to the bone.

Precautions: • Confirm that plate positioning, drill bit, and screw

length allow for adequate clearance of nerves, the edge of the bone, and any other critical structures.

• Instrument tips may be sharp, handle with care and dispose sharp cuttings in an approved sharps container.

• If contouring is necessary, the surgeon should avoid bending the device at a screw hole.

• Avoid sharp bends, repetitive and reverse bending as it increases the risk of implant breakage.

• Avoid contouring of the implant in situ that may lead to implant malposition and/or posterior canti-lever effect.

• Take care to protect soft tissue from trimmed edges.

Surgical Technique

Orbital Plates

Page 14: The Systematic Solution for Craniomaxillofacial Indications and …synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes... · 2018-07-07 · implants for craniomaxillofacial indications

12 DePuy Synthes Compact Midface Surgical Technique

3. Drill the hole

If pilot hole is desired, select the appropriate diameter drill bit length (see reference chart on page 3) to allow for adequate clearance of nerves, and critical structure.

Precautions: • Drill speed rate should never exceed 1,800 rpm,

particularly in dense, hard bone. Higher drill speed rates can result in: – thermal necrosis of the bone, – soft tissue burns, – an oversized hole, which can lead to reduced

pullout force, increased ease of the screws strip-ping in bone, suboptimal fixation, and/or the need for emergency screws.

• Avoid damaging the plate threads with the drill.• Always irrigate during drilling to avoid thermal

damage to the bone.• Always irrigate and apply suction during drilling

to ensure removal of debris potentially generated during implantation.

Note: Self-tapping and self-drilling screws are avail-able (see reference chart on page 3).

4. Fixate plate to the bone

Stabilize the implant with screws inserted through selected screw holes in the plate. Insert screws of the appropriate diameter and length to secure the plate to the bone.

Precautions: • In order to determine the appropriate amount of

screws needed to achieve stable construct fixation, the surgeon should consider the fracture size and shape.

• Confirm screw length prior to implantation.• Tighten screws in a controlled manner. Applying

too much torque to the screws may cause screw/plate deformation or bone stripping. If bone be-comes stripped, remove the screw from the bone and replace with an emergency screw.

Note: Test for impingement: A forced duction test must be completed to ensure unrestricted lateral and medial movement of the globe.

Surgical Technique Orbital Plates

Page 15: The Systematic Solution for Craniomaxillofacial Indications and …synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes... · 2018-07-07 · implants for craniomaxillofacial indications

Compact Midface Surgical Technique DePuy Synthes 13

Surgical Technique

Le Fort I Fixation

1. Select plate design After the osteotomy has been performed and the new position of the maxilla has been established, select the appropriate plate shape and thickness that best suits the bony anatomy, treatment objective, and the quantity and quality of bone.

2. Adapt plate to the bone

Cut and contour the plate to the bone using the plate cutter and the bending pliers respectively. Ensure that the plate is placed flush to the bone.

Precautions: • If contouring is necessary, the surgeon should

avoid bending the device at a screw hole. • Avoid sharp bends, repetitive and reverse bending

as it increases the risk of implant breakage. • Take care to protect soft tissue from trimmed

edges. • Confirm that plate positioning, drill bit, and screw

length allow for adequate clearance of nerves, tooth buds and/or tooth roots, the edge of the bone, and any other critical structures.

Page 16: The Systematic Solution for Craniomaxillofacial Indications and …synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes... · 2018-07-07 · implants for craniomaxillofacial indications

14 DePuy Synthes Compact Midface Surgical Technique

3. Drill the hole

If pilot hole is desired, select the appropriate diameter drill bit length (see reference chart on page 3) to allow for adequate clearance of nerves, tooth buds, tooth roots, and critical structure.

Note: Self-tapping and self-drilling screws are avail-able (see reference chart on page 3).

Precautions: • Drill speed rate should never exceed 1,800 rpm,

particularly in dense, hard bone. Higher drill speed rates can result in: – thermal necrosis of the bone, – soft tissue burns, – an oversized hole, which can lead to reduced

pullout force, increased ease of the screws strip-ping in bone, suboptimal fixation, and/or the need for emergency screws.

• Avoid damaging the plate threads with the drill. • Always irrigate during drilling to avoid thermal

damage to the bone. • Always irrigate and apply suction during drilling

to ensure removal of debris potentially generated during implantation.

Surgical Technique Le Fort I Fixation

4. Fixate plate to the bone

Insert screws of the appropriate diameter and length to fixate the plate to the underlying bone.

Precaution: In order to determine the appropriate amount of screws needed to achieve stable construct fixation, the surgeon should consider the fracture size and shape.

Page 17: The Systematic Solution for Craniomaxillofacial Indications and …synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes... · 2018-07-07 · implants for craniomaxillofacial indications

Compact Midface Surgical Technique DePuy Synthes 15

malleable

Strut and mesh plates

Orbital floor plates

Midface and cranial plates

Implants

Product Information

Prebent maxilla plates

For more information on the implants, please contact your local representative.

Locking plates

Page 18: The Systematic Solution for Craniomaxillofacial Indications and …synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes... · 2018-07-07 · implants for craniomaxillofacial indications

16 DePuy Synthes Compact Midface Surgical Technique

Instruments

Standard instruments for the Compact MidfaceThe standard solution contains the basic set of instru-ments for craniomaxillofacial surgery. For further informations concerning the burr free plate cutter and the three-prong bending pliers see the corre-sponding flyers (036.000.087 and 036.000.007)

Individual configuration of instrument trays

In addition to the standard model, indivi dually-adapted instrument trays can be assembled.

Appropriate and multifunctional inserts are available for the different instruments.

Page 19: The Systematic Solution for Craniomaxillofacial Indications and …synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes... · 2018-07-07 · implants for craniomaxillofacial indications

Compact Midface Surgical Technique DePuy Synthes 17

Sets

Size 2/3 3/3

Compact 1.0 Midface 2

Compact 1.3 Midface 2

Compact 1.5 Midface 2

Compact 2.0 Midface 2

Compact 2.0 LOCK Midface 2

Compact 2.0 Combi 2

Overview: Standard modules

Size 1/3

Compact 1.0 Midface 2

Compact 1.3 Midface 2

Compact 1.5 Midface 2

Compact 2.0 Midface 2

Size 3/3

Compact 1.5/2.0 Orthognatics 2

Size 2/3

Compact 1.5/2.0 Orthognatics Plates 2

Compact 1.5/2.0 Orthognatics 2

Screws and Instruments

Compact MidfaceThe color-coded solution with inserts for im-plants, screws and instruments. The following pre-configured systems are available:

Compact Midface sterileThe practical solution for using sterile im-plants. The module contains rails for snapping in sterile screw clips, and has space for drill bits and screw drivers. The following standard Midface systems are available:

Compact 1.5/2.0 OrthognaticsThe user-friendly module in the standard configuration with inserts for instruments, screw clips and implants in the dimensions 1.5/2.0. In addition, it has pockets for storing screws up to 40 mm long for the lag screw technique.

Compact 1.5/2.0 Orthognatics Plates Compact 1.5/2.0 Orthognatics Screws and InstrumentsTwo modules that belong together when a wide selection of implants and screws is desired. The Screws and Instruments module is also highly suitable for use with sterile implants.

Page 20: The Systematic Solution for Craniomaxillofacial Indications and …synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes... · 2018-07-07 · implants for craniomaxillofacial indications

11 DePuy Synthes Compact Midface Surgical Technique

Plates

Art. No. Article

04.105.003 Adaption Plate 1.3, 6 holes, Pure Titanium

04.105.004 Adaption Plate 1.3, 8 holes, Pure Titanium

04.105.028 Adaption Plate 1.5, 6 holes, hole spacing 4.5 mm, Pure Titanium

04.105.029 Adaption Plate 1.5, 8 holes, hole spacing 4.5 mm, Pure Titanium

04.105.039 Adaption Plate 1.5, with intersection bars, 4 holes, hole spacing 6.5 mm, Pure Titanium

04.105.040 Adaption Plate 1.5, with intersection bars, 6 holes, hole spacing 6.5 mm, Pure Titanium

04.105.047 Adaption Plate 1.5, with intersection bars, 20 holes, hole spacing 6.5 mm, Pure Titanium

04.105.053 Adaption Plate 2.0, 6 holes, hole spacing 5.0 mm, Pure Titanium

04.105.054 Adaption Plate 2.0, 8 holes, hole spacing 5.0 mm, Pure Titanium

04.105.064 Adaption Plate 2.0, with intersection bars, 4 holes, hole spacing 6.5 mm, Pure Titanium

04.105.065 Adaption Plate 2.0, with intersection bars, 6 holes, hole spacing 6.5 mm, Pure Titanium

04.105.071 Adaption Plate 2.0, with intersection bars, 18 holes, hole spacing 6.5 mm, Pure Titanium

420.010 Strut Plate 1.0, 6 × 30 mm, 2 × 8 holes, thickness 0.5 mm, Pure Titanium

420.020 Strut Plate 1.0, 42 × 38 mm, 10 × 11 holes, thickness 0.4 mm, Pure Titanium

420.040 Strut Plate 1.0, 6 × 42 mm, 2 × 11 holes, thickness 0.5 mm, Pure Titanium

420.060 Strut Plate 1.0, 100 × 100 mm, 26 × 26 holes, thickness 0.4 mm, Pure Titanium

420.070 Screen Plate 1.0, 100 × 100 mm, Pure Titanium

420.100 Adaption Plate 1.0, 34 holes, Pure Titanium

420.210 Orbital Rim Plate 1.0, curved, 11 holes, Pure Titanium

420.302 Orbital Floor Mesh Plate 1.0, thickness 0.2 mm, Pure Titanium

420.303 Orbital Floor Mesh Plate 1.0, thickness 0.3 mm, Pure Titanium

420.304 Orbital Floor Mesh Plate 1.0, thickness 0.4 mm, Pure Titanium

420.401 Anatomic Orbital Floor Plate 1.0, small, Pure Titanium

420.402 Anatomic Orbital Floor Plate 1.0, large, Pure Titanium

420.440 Orbital Floor Plate 1.0, left, Pure Titanium

420.450 Orbital Floor Plate 1.0, right, Pure Titanium

420.540 L-Plate 1.0, left, 9 holes, Pure Titanium

420.550 L-Plate 1.0, right, 9 holes, Pure Titanium

420.620 Y-Plate 1.0, 9 holes, Pure Titanium

420.630 Y-Plate 1.0, 13 holes, Pure Titanium

420.680 Double Y-Plate 1.0, Pure Titanium

420.700 T-Plate 1.0, 10 holes, Pure Titanium

420.710 T-Plate 1.0, 12 holes, Pure Titanium

420.810 X-Plate 1.0, 5 holes, Pure Titanium

420.840 H-Plate 1.0, 6 holes, Pure Titanium

420.850 H-Plate 1.0, 11 holes, Pure Titanium

421.001 Strut Plate 1.3, 38 × 53 mm, 8 × 11 holes, thickness 0.4 mm, Pure Titanium

421.004 Mesh Plate 1.3, 38 × 45 mm, thickness 0.4 mm, Pure Titanium, blue

421.005 Mesh Plate 1.3, 38 × 45 mm, thickness 0.6 mm, Pure Titanium, gold

421.006 Strut Plate 1.3, 100 × 100 mm, 21 × 21 holes, thickness 0.4 mm, Pure Titanium

421.007 Mesh Plate 1.3, 100 × 100 mm, thickness 0.4 mm, Pure Titanium, blue

421.008 Mesh Plate 1.3, 100 × 100 mm, thickness 0.6 mm, Pure Titanium, gold

421.009 Mesh Plate 1.3, 200 × 200 mm, thickness 0.6 mm, Pure Titanium, gold

421.010 Adaption Plate 1.3, 24 holes, Pure Titanium

421.021 Orbital Rim Plate 1.3, curved, 9 holes, Pure Titanium

421.022 Plate 1.3, straight, 2 holes, Pure Titanium

Ordering Information

Page 21: The Systematic Solution for Craniomaxillofacial Indications and …synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes... · 2018-07-07 · implants for craniomaxillofacial indications

Compact Midface Surgical Technique DePuy Synthes 19

421.024 Plate 1.3, straight, 4 holes, Pure Titanium

421.040 Universal Orbital Floor Plate 1.3, Pure Titanium

421.044 Orbital Floor Plate 1.3, left, Pure Titanium

421.045 Orbital Floor Plate 1.3, right, Pure Titanium

421.048 Anatomic Orbital Floor Plate 1.3, small, Pure Titanium

421.049 Anatomic Orbital Floor Plate 1.3, large, Pure Titanium

421.050 L-Plate 1.3, left, 6 holes, Pure Titanium

421.051 L-Plate 1.3, right, 6 holes, Pure Titanium

421.052 L-Plate 1.3, left, 7 holes, Pure Titanium

421.053 L-Plate 1.3, right, 7 holes, Pure Titanium

421.061 Y-Plate 1.3, 5 holes, Pure Titanium

421.068 Double Y-Plate 1.3, short, Pure Titanium

421.069 Double Y-Plate 1.3, long, Pure Titanium

421.071 T-Plate 1.3, 7 holes, Pure Titanium

421.080 Strut Plate 1.3, 8 × 43 mm, 2 × 9 holes, thickness 0.5 mm, Pure Titanium

421.081 X-Plate 1.3, 4 holes, Pure Titanium

421.085 H-Plate 1.3, 11 holes, Pure Titanium

421.091 Burr Hole Cover 1.3, for burr holes up to B 17.0 mm, Pure Titanium

421.095 Frame Plate 1.3, rectangular, 4 holes, Pure Titanium

421.096 Frame Plate 1.3, square, 4 holes, Pure Titanium

421.099 Screen Plate 1.3, 100 × 100 mm, Pure Titanium

421.302 Orbital Floor Mesh Plate 1.3, thickness 0.2 mm, Pure Titanium

421.303 Orbital Floor Mesh Plate 1.3, thickness 0.3 mm, Pure Titanium

421.304 Orbital Floor Mesh Plate 1.3, thickness 0.4 mm, Pure Titanium

443.440 Zygomatic DCP 2.0, 4 holes, Pure Titanium

443.450 Zygomatic DCP 2.0, 5 holes, Pure Titanium

443.451 Zygomatic DCP 2.0, with centre space, 4 holes, Pure Titanium

443.460 Zygomatic DCP 2.0, 6 holes, Pure Titanium

446.024 Cranial Plate 1.5, straight, with centre space, 4 holes, length 25 mm, Pure Titanium

446.042 Maxilla Plate 1.5, right, prebent, lengthening 3 mm, Pure Titanium

446.043 Maxilla Plate 1.5, left, prebent, lengthening 3 mm, Pure Titanium

446.044 Maxilla Plate 1.5, right, prebent, lengthening 5 mm, Pure Titanium

446.045 Maxilla Plate 1.5, left, prebent, lengthening 5 mm, Pure Titanium

446.046 Maxilla Plate 1.5, right, prebent, lengthening 7 mm, Pure Titanium

446.047 Maxilla Plate 1.5, left, prebent, lengthening 7 mm, Pure Titanium

446.048 Maxilla Plate 1.5, right, prebent, lengthening 9 mm, Pure Titanium

446.049 Maxilla Plate 1.5, left, prebent, lengthening 9 mm, Pure Titanium

446.060 Maxilla Plate 1.5, right, prebent, lengthening 11 mm, Pure Titanium

446.061 Maxilla Plate 1.5, left, prebent, lengthening 11 mm, Pure Titanium

446.096 Frame Plate 1.5, square, 4 holes, Pure Titanium

446.100 Adaption Plate 1.5, 20 holes, hole spacing 4.5 mm, Pure Titanium

446.101 Adaption Plate 1.5, 20 holes, thickness 0.6 mm, Pure Titanium

446.180 Orbital Rim Plate 1.5, curved, 8 holes, Pure Titanium

446.200 Orbital Rim Plate 1.5, curved, 10 holes, Pure Titanium

446.220 Orbital Rim Plate 1.5, curved, 12 holes, Pure Titanium

446.400 Universal Orbital Floor Plate 1.5, Pure Titanium

446.500 L-Plate 1.5, left, 5 holes, Pure Titanium

446.504 L-Plate 1.5, right, 6 holes, length 22 × 13 mm, Pure Titanium

446.505 L-Plate 1.5, left, 6 holes, length 22 × 13 mm, Pure Titanium

Page 22: The Systematic Solution for Craniomaxillofacial Indications and …synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes... · 2018-07-07 · implants for craniomaxillofacial indications

21 DePuy Synthes Compact Midface Surgical Technique

446.506 L-Plate 1.5, right, 6 holes, length 24 × 13 mm, Pure Titanium

446.507 L-Plate 1.5, left, 6 holes, length 24 × 13 mm, Pure Titanium

446.508 L-Plate 1.5, right, 6 holes, length 26 × 13 mm, Pure Titanium

446.509 L-Plate 1.5, left, 6 holes, length 26 × 13 mm, Pure Titanium

446.510 L-Plate 1.5, right, 5 holes, Pure Titanium

446.515 L-Plate 1.5, right, 5 holes, length 22 × 9.0 mm, Pure Titanium, malleable

446.516 L-Plate 1.5, left, 5 holes, length 22 × 9.0 mm, Pure Titanium, malleable

446.520 L-Plate 1.5, left, 7 holes, Pure Titanium

446.527 L-Plate 1.5, left, 7 holes, length 27 × 13 mm, Pure Titanium, malleable

446.530 L-Plate 1.5, right, 7 holes, Pure Titanium

446.537 L-Plate 1.5, right, 7 holes, length 27 × 13 mm, Pure Titanium, malleable

447.180 Orbital Rim Plate 2.0, curved, 8 holes, Pure Titanium

447.200 Orbital Rim Plate 2.0, curved, 10 holes, Pure Titanium

447.220 Orbital Rim Plate 2.0, curved, 12 holes, Pure Titanium

447.222 LOCK Plate 2.0, straight, 4 holes, length 20 mm, Pure Titanium

447.223 LOCK Plate 2.0, straight, 6 holes, length 30 mm, Pure Titanium

447.224 LOCK Plate 2.0, small, with centre space, 4 holes, length 28 mm, Pure Titanium

447.225 LOCK Plate 2.0, small, with centre space, 6 holes, length 40 mm, Pure Titanium

447.226 LOCK L-Plate 2.0, right, 7 holes, Pure Titanium

447.227 LOCK L-Plate 2.0, left, 7 holes, Pure Titanium

447.252 LOCK Plate 2.0, straight, with centre space, 2 holes, length 15 mm, Pure Titanium

447.253 LOCK Zygomatic Plate 2.0, straight, with centre space, 4 holes, Pure Titanium

447.254 LOCK Zygomatic Plate 2.0, straight, 4 holes, Pure Titanium

447.255 LOCK Zygomatic Plate 2.0, straight, 5 holes, Pure Titanium

447.256 LOCK Zygomatic Plate 2.0, straight, 6 holes, Pure Titanium

447.257 LOCK Plate 2.0, small, 20 holes, Pure Titanium

447.258 LOCK Orbital Rim Plate 2.0, curved, 8 holes, Pure Titanium

447.260 LOCK Orbital Rim Plate 2.0, curved, 10 holes, Pure Titanium

447.262 LOCK Orbital Rim Plate 2.0, curved, 12 holes, Pure Titanium

447.264 LOCK L-Plate 2.0, right, 5 holes, Pure Titanium

447.265 LOCK L-Plate 2.0, left, 5 holes, Pure Titanium

447.266 LOCK L-Plate 2.0, right, 10 holes, Pure Titanium

447.267 LOCK L-Plate 2.0, left, 10 holes, Pure Titanium

447.268 LOCK Y-Plate 2.0, 5 holes, Pure Titanium

447.269 LOCK Y-Plate 2.0, 8 holes, Pure Titanium

447.271 LOCK Double-Y-Plate 2.0, 8 holes, Pure Titanium

447.272 LOCK T-Plate 2.0, 6 holes, Pure Titanium

447.273 LOCK X-Plate 2.0, 4 holes, Pure Titanium

447.275 LOCK H-Plate 2.0, 9 holes, Pure Titanium

447.276 LOCK T-Plate 2.0, 11 holes, Pure Titanium

447.380 Adaption Plate 2.0, with intersection bars, 30 holes, hole spacing 6.5 mm, Pure Titanium

447.500 L-Plate 2.0, left, 5 holes, Pure Titanium

447.510 L-Plate 2.0, right, 5 holes, Pure Titanium

447.515 L-Plate 2.0, right, 5 holes, length 25 × 10 mm, Pure Titanium, malleable

447.516 L-Plate 2.0, left, 5 holes, length 25 × 10 mm, Pure Titanium, malleable

447.520 L-Plate 2.0, left, 7 holes, Pure Titanium

Ordering Information

Page 23: The Systematic Solution for Craniomaxillofacial Indications and …synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes... · 2018-07-07 · implants for craniomaxillofacial indications

Compact Midface Surgical Technique DePuy Synthes 21

447.527 L-Plate 2.0, left, 7 holes, length 31 × 15 mm, Pure Titanium, malleable

447.530 L-Plate 2.0, right, 7 holes, Pure Titanium

447.537 L-Plate 2.0, right, 7 holes, length 31 × 15 mm, Pure Titanium, malleable

447.540 L-Plate 2.0, left, 10 holes, Pure Titanium

447.548 L-Plate 2.0, long, right, 4 holes, Pure Titanium, malleable

447.549 L-Plate 2.0, long, left, 4 holes, Pure Titanium, malleable

447.550 L-Plate 2.0, right, 10 holes, Pure Titanium

447.557 L-Plate 2.0, left, 4 holes, length 18 × 11 mm, Pure Titanium

447.562 Z-Plate 2.0, right, 4 holes, length 21 × 17 mm, Pure Titanium

447.563 Z-Plate 2.0, left, 4 holes, length 21 × 17 mm, Pure Titanium

447.564 Z-Plate 2.0, right, 4 holes, length 25 × 17 mm, Pure Titanium

447.565 Z-Plate 2.0, left, 4 holes, length 25 × 17 mm, Pure Titanium

447.572 Z-Plate 2.0, short, right, 4 holes, Pure Titanium, malleable

447.573 Z-Plate 2.0, short, left, 4 holes, Pure Titanium, malleable

447.574 Z-Plate 2.0, long, right, 4 holes, Pure Titanium, malleable

447.575 Z-Plate 2.0, long, left, 4 holes, Pure Titanium, malleable

447.610 Y-Plate 2.0, 5 holes, Pure Titanium

447.620 Y-Plate 2.0, 8 holes, Pure Titanium

447.680 Double Y-Plate 2.0, 8 holes, Pure Titanium

447.810 X-Plate 2.0, 4 holes, Pure Titanium

447.850 H-Plate 2.0, 9 holes, Pure Titanium

447.860 H-Plate 2.0, 8 holes, Pure Titanium

450.101 Adaption Plate 1.5, 20 holes, thickness 0.6 mm, Pure Titanium, malleable

450.611 Y-Plate 1.5, 5 holes, thickness 0.6 mm, Pure Titanium, malleable

Screws:Cortex Screw PlusDrive™ B 1.0 mm, self-tapping

Art. No. Article

400.402 Cortex Screw B 1.0 mm, self-tapping, length 2 mm

400.403 Cortex Screw B 1.0 mm, self-tapping, length 3 mm

400.404 Cortex Screw B 1.0 mm, self-tapping, length 4 mm

400.405 Cortex Screw B 1.0 mm, self-tapping, length 5 mm

400.406 Cortex Screw B 1.0 mm, self-tapping, length 6 mm

400.408 Cortex Screw B 1.0 mm, self-tapping, length 8 mm

400.410 Cortex Screw B 1.0 mm, self-tapping, length 10 mm

400.412 Cortex Screw B 1.0 mm, self-tapping, length 12 mm

400.414 Cortex Screw B 1.0 mm, self-tapping, length 14 mm

Cortex Screw PlusDrive™ B 1.0 mm, self-drilling, Titanium Alloy (TAN)

Art. No. Article

400.422 Cortex Screw B 1.0 mm, self-drilling, length 2 mm

400.423 Cortex Screw B 1.0 mm, self-drilling, length 3 mm

Cortex Screw PlusDrive™ B 1.3 mm, self-tapping

Art. No. Article

400.433 Cortex Screw B 1.3 mm, self-tapping, length 3 mm

400.434 Cortex Screw B 1.3 mm, self-tapping, length 4 mm

400.435 Cortex Screw B 1.3 mm, self-tapping, length 5 mm

400.436 Cortex Screw B 1.3 mm, self-tapping, length 6 mm

400.438 Cortex Screw B 1.3 mm, self-tapping, length 8 mm

Page 24: The Systematic Solution for Craniomaxillofacial Indications and …synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes... · 2018-07-07 · implants for craniomaxillofacial indications

22 DePuy Synthes Compact Midface Surgical Technique

400.440 Cortex Screw B 1.3 mm, self-tapping, length 10 mm

400.442 Cortex Screw B 1.3 mm, self-tapping, length 12 mm

400.444 Cortex Screw B 1.3 mm, self-tapping, length 14 mm

400.446 Cortex Screw B 1.3 mm, self-tapping, length 16 mm

400.448 Cortex Screw B 1.3 mm, self-tapping, length 18 mm

Cortex Screw PlusDrive™ B 1.3 mm, self-drilling, Titanium Alloy (TAN)

Art. No. Article

400.454 Cortex Screw B 1.3 mm, self-drilling, length 4 mm

400.455 Cortex Screw B 1.3 mm, self-drilling, length 5 mm

400.456 Cortex Screw B 1.3 mm, self-drilling, length 6 mm

Cortex Screw PlusDrive™ B 1.5 mm, self-tapping, Titanium Alloy (TAN)

Art. No. Article

400.034 Cortex Screw B 1.5 mm, self-tapping, length 4 mm

400.035 Cortex Screw B 1.5 mm, self-tapping, length 5 mm

400.036 Cortex Screw B 1.5 mm, self-tapping, length 6 mm

400.038 Cortex Screw B 1.5 mm, self-tapping, length 8 mm

400.040 Cortex Screw B 1.5 mm, self-tapping, length 10 mm

400.042 Cortex Screw B 1.5 mm, self-tapping, length 12 mm

400.044 Cortex Screw B 1.5 mm, self-tapping, length 14 mm

400.046 Cortex Screw B 1.5 mm, self-tapping, length 16 mm

400.048 Cortex Screw B 1.5 mm, self-tapping, length 18 mm

Cortex Screw PlusDrive™ B 1.5 mm, self-drilling, Titanium Alloy (TAN)

Art. No. Article

400.054 Cortex Screw B 1.5 mm, self-drilling, length 4 mm

400.055 Cortex Screw B 1.5 mm, self-drilling, length 5 mm

400.056 Cortex Screw B 1.5 mm, self-drilling, length 6 mm

400.058 Cortex Screw B 1.5 mm, self-drilling, length 8 mm

Cortex Screw PlusDrive™ B 2.0 mm, self-tapping

Art. No. Article

401.041 Cortex Screw PlusDrive B 2.0 mm, length 4 mm

401.042 Cortex Screw PlusDrive B 2.0 mm, length 5 mm

401.043 Cortex Screw PlusDrive B 2.0 mm, length 6 mm

401.044 Cortex Screw PlusDrive B 2.0 mm, length 8 mm

401.045 Cortex Screw PlusDrive B 2.0 mm, length 10 mm

401.046 Cortex Screw PlusDrive B 2.0 mm, length 12 mm

401.047 Cortex Screw PlusDrive B 2.0 mm, length 14 mm

401.048 Cortex Screw PlusDrive B 2.0 mm, length 16 mm

401.049 Cortex Screw PlusDrive B 2.0 mm, length 18 mm

Cortex Screw PlusDrive™ B 2.0 mm, self-drilling, Titanium Alloy (TAN)

Art. No. Article

401.061 Cortex Screw PlusDrive B 2.0 mm, length 4 mm

401.062 Cortex Screw PlusDrive B 2.0 mm, length 5 mm

401.063 Cortex Screw PlusDrive B 2.0 mm, length 6 mm

Ordering Information

Page 25: The Systematic Solution for Craniomaxillofacial Indications and …synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes... · 2018-07-07 · implants for craniomaxillofacial indications

Compact Midface Surgical Technique DePuy Synthes 23

401.064 Cortex Screw PlusDrive B 2.0 mm, length 7 mm

401.065 Cortex Screw PlusDrive B 2.0 mm, length 8 mm

MF Cortex Screw B 1.0 mm, self-tapping

Art. No. Article

400.502 MF Cortex Screw B 1.0 mm, self-tapping, length 2 mm

400.503 MF Cortex Screw B 1.0 mm, self-tapping, length 3 mm

400.504 MF Cortex Screw B 1.0 mm, self-tapping, length 4 mm

400.505 MF Cortex Screw B 1.0 mm, self-tapping, length 5 mm

400.506 MF Cortex Screw B 1.0 mm, self-tapping, length 6 mm

400.507 MF Cortex Screw B 1.0 mm, self-tapping, length 7 mm

400.508 MF Cortex Screw B 1.0 mm, self-tapping, length 8 mm

MF Cortex Screw B 1.3 mm, self-tapping

Art. No. Article

400.233 MF Cortex Screw B 1.3 mm, self-tapping, length 3 mm

400.234 MF Cortex Screw B 1.3 mm, self-tapping, length 4 mm

400.235 MF Cortex Screw B 1.3 mm, self-tapping, length 5 mm

400.236 MF Cortex Screw B 1.3 mm, self-tapping, length 6 mm

400.238 MF Cortex Screw B 1.3 mm, self-tapping, length 8 mm

MF Cortex Screw B 1.5 mm, self-tapping

Art. No. Article

400.704 MF Cortex Screw B 1.5 mm, self-tapping, length 4 mm

400.706 MF Cortex Screw B 1.5 mm, self-tapping, length 6 mm

400.707 MF Cortex Screw B 1.5 mm, self-tapping, length 7 mm

400.708 MF Cortex Screw B 1.5 mm, self-tapping, length 8 mm

400.709 MF Cortex Screw B 1.5 mm, self-tapping, length 9 mm

400.710 MF Cortex Screw B 1.5 mm, self-tapping, length 10 mm

400.711 MF Cortex Screw B 1.5 mm, self-tapping, length 11 mm

400.712 MF Cortex Screw B 1.5 mm, self-tapping, length 12 mm

400.714 MF Cortex Screw B 1.5 mm, self-tapping, length 14 mm

400.716 MF Cortex Screw B 1.5 mm, self-tapping, length 16 mm

400.718 MF Cortex Screw B 1.5 mm, self-tapping, length 18 mm

400.720 MF Cortex Screw B 1.5 mm, self-tapping, length 20 mm

MF Cortex Screw B 2.0 mm, self-tapping

Art. No. Article

401.104 MF Cortex Screw B 2.0 mm, self-tapping, length 4 mm

401.106 MF Cortex Screw B 2.0 mm, self-tapping, length 6 mm

401.108 MF Cortex Screw B 2.0 mm, self-tapping, length 8 mm

401.110 MF Cortex Screw B 2.0 mm, self-tapping, length 10 mm

401.112 MF Cortex Screw B 2.0 mm, self-tapping, length 12 mm

401.114 MF Cortex Screw B 2.0 mm, self-tapping, length 14 mm

401.116 MF Cortex Screw B 2.0 mm, self-tapping, length 16 mm

401.118 MF Cortex Screw B 2.0 mm, self-tapping, length 18 mm

401.120 MF Cortex Screw B 2.0 mm, self-tapping, length 20 mm

Page 26: The Systematic Solution for Craniomaxillofacial Indications and …synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes... · 2018-07-07 · implants for craniomaxillofacial indications

24 DePuy Synthes Compact Midface Surgical Technique

Emergency Screws: Emergency Screw PlusDrive™ B 1.2 mm, self-tapping

Art. No. Article

400.462 Emergency Screw B 1.2 mm, self-tapping, length 2 mm

400.463 Emergency Screw B 1.2 mm, self-tapping, length 3 mm

400.464 Emergency Screw B 1.2 mm, self-tapping, length 4 mm

400.465 Emergency Screw B 1.2 mm, self-tapping, length 5 mm

400.466 Emergency Screw B 1.2 mm, self-tapping, length 6 mm

400.468 Emergency Screw B 1.2 mm, self-tapping, length 8 mm

400.470 Emergency Screw B 1.2 mm, self-tapping, length 10 mm

400.472 Emergency Screw B 1.2 mm, self-tapping, length 12 mm

400.474 Emergency Screw B 1.2 mm, self-tapping, length 14 mm

Emergency Screw PlusDrive™ B 1.7 mm, self-tapping

Art. No. Article

400.483 Emergency Screw B 1.7 mm, self-tapping, length 3 mm

400.484 Emergency Screw B 1.7 mm, self-tapping, length 4 mm

400.485 Emergency Screw B 1.7 mm, self-tapping, length 5 mm

400.486 Emergency Screw B 1.7 mm, self-tapping, length 6 mm

400.488 Emergency Screw B 1.7 mm, self-tapping, length 8 mm

400.490 Emergency Screw B 1.7 mm, self-tapping, length 10 mm

400.492 Emergency Screw B 1.7 mm, self-tapping, length 12 mm

400.494 Emergency Screw B 1.7 mm, self-tapping, length 14 mm

400.496 Emergency Screw B 1.7 mm, self-tapping, length 16 mm

400.498 Emergency Screw B 1.7 mm, self-tapping, length 18 mm

Emergency Screw PlusDrive™ B 2.0 mm, self-tapping

Art. No. Article

400.274 Emergency Screw B 2.0 mm, self-tapping, length 4 mm

400.275 Emergency Screw B 2.0 mm, self-tapping, length 5 mm

400.276 Emergency Screw B 2.0 mm, self-tapping, length 6 mm

400.278 Emergency Screw B 2.0 mm, self-tapping, length 8 mm

400.280 Emergency Screw B 2.0 mm, self-tapping, length 10 mm

400.282 Emergency Screw B 2.0 mm, self-tapping, length 12 mm

400.284 Emergency Screw B 2.0 mm, self-tapping, length 14 mm

400.286 Emergency Screw B 2.0 mm, self-tapping, length 16 mm

400.288 Emergency Screw B 2.0 mm, self-tapping, length 18 mm

Emergency Screw PlusDrive™ B 2.4 mm, self-tapping

Art. No. Article

401.791 Emergency Screw PlusDrive B 2.4 mm, length 5 mm

401.792 Emergency Screw PlusDrive B 2.4 mm, length 6 mm

401.794 Emergency Screw PlusDrive B 2.4 mm, length 8 mm

401.795 Emergency Screw PlusDrive B 2.4 mm, length 10 mm

401.796 Emergency Screw PlusDrive B 2.4 mm, length 12 mm

401.797 Emergency Screw PlusDrive B 2.4 mm, length 14 mm

401.798 Emergency Screw PlusDrive B 2.4 mm, length 16 mm

401.799 Emergency Screw PlusDrive B 2.4 mm, length 18 mm

Ordering Information

Page 27: The Systematic Solution for Craniomaxillofacial Indications and …synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes... · 2018-07-07 · implants for craniomaxillofacial indications

Compact Midface Surgical Technique DePuy Synthes 25

Emergency Screw Cruciform Recess B 1.2 mm, self-tapping

Art. No. Article

400.602 Emergency Screw B 1.2 mm, self-tapping, length 2 mm

400.603 Emergency Screw B 1.2 mm, self-tapping, length 3 mm

400.604 Emergency Screw B 1.2 mm, self-tapping, length 4 mm

400.605 Emergency Screw B 1.2 mm, self-tapping, length 5 mm

400.606 Emergency Screw B 1.2 mm, self-tapping, length 6 mm

400.607 Emergency Screw B 1.2 mm, self-tapping, length 7 mm

400.608 Emergency Screw B 1.2 mm, self-tapping, length 8 mm

Emergency Screw Cruciform Recess B 1.7 mm, self-tapping

Art. No. Article

400.883 Emergency Screw B 1.7 mm, self-tapping, length 3 mm

400.884 Emergency Screw B 1.7 mm, self-tapping, length 4 mm

400.885 Emergency Screw B 1.7 mm, self-tapping, length 5 mm

400.886 Emergency Screw B 1.7 mm, self-tapping, length 6 mm

400.888 Emergency Screw B 1.7 mm, self-tapping, length 8 mm

Emergency Screw B 2.4 mm, self-tapping

Art. No. Article

401.306 Emergency Screw B 2.4 mm, self-tapping, length 6 mm

401.308 Emergency Screw B 2.4 mm, self-tapping, length 8 mm

401.310 Emergency Screw B 2.4 mm, self-tapping, length 10 mm

401.312 Emergency Screw B 2.4 mm, self-tapping, length 12 mm

LOCK Screw PlusDrive™ B 2.0 mm, self-tapping

Art. No. Article

401.291 LOCK Screw PlusDrive B 2.0 mm, self-tapping, length 5 mm, Titanium Alloy (TAN)

401.292 LOCK Screw PlusDrive B 2.0 mm, self-tapping, length 6 mm, Titanium Alloy (TAN)

401.294 LOCK Screw PlusDrive B 2.0 mm, self-tapping, length 8 mm, Titanium Alloy (TAN)

401.295 LOCK Screw PlusDrive B 2.0 mm, self-tapping, length 10 mm, Titanium Alloy (TAN)

401.296 LOCK Screw PlusDrive B 2.0 mm, self-tapping, length 12 mm, Titanium Alloy (TAN)

401.297 LOCK Screw PlusDrive B 2.0 mm, self-tapping, length 14 mm, Titanium Alloy (TAN)

401.298 LOCK Screw PlusDrive B 2.0 mm, self-tapping, length 16 mm, Titanium Alloy (TAN)

401.299 LOCK Screw PlusDrive B 2.0 mm, self-tapping, length 18 mm, Titanium Alloy (TAN)

LOCK Screw PlusDrive™ B 2.0 mm, self-drilling

Art. No. Article

401.955 LOCK Screw PlusDrive B 2.0 mm, self-drilling, length 5 mm

401.956 LOCK Screw PlusDrive B 2.0 mm, self-drilling, length 6 mm

401.957 LOCK Screw PlusDrive B 2.0 mm, self-drilling, length 8 mm

Page 28: The Systematic Solution for Craniomaxillofacial Indications and …synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes... · 2018-07-07 · implants for craniomaxillofacial indications

26 DePuy Synthes Compact Midface Surgical Technique

Ordering Information* Screws Plates

Packs of 1 unit xxx.xxx xxx.xxx

Packs of 1 unit with clip xxx.xxx.01C n/a

Packs of 4 units with clip xxx.xxx.04C n/a

Packs of 1 unit, sterile n/a xxx.xxxS

Packs of 1 unit with clip, sterile xxx.xxxS n/a

Packs of 4 units with clip, sterile xxx.xxx.04S n/a

* Not all packaging variatons are available for all items.

Page 29: The Systematic Solution for Craniomaxillofacial Indications and …synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes... · 2018-07-07 · implants for craniomaxillofacial indications
Page 30: The Systematic Solution for Craniomaxillofacial Indications and …synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes... · 2018-07-07 · implants for craniomaxillofacial indications

0123

Synthes GmbHEimattstrasse 34436 OberdorfSwitzerlandTel: +41 61 965 61 11Fax: +41 61 965 66 00www.depuysynthes.com

Not all products are currently available in all markets.

This publication is not intended for distribution in the USA.

All surgical techniques are available as PDF files at www.depuysynthes.com/ifu ©

DeP

uy S

ynth

es C

MF,

a d

ivis

ion

of S

ynth

es G

mbH

. 201

8.

All

right

s re

serv

ed.

036.

000.

193

DS

EM

/CM

F/03

16/0

121(

1)

06/1

8