Sales Manual B.Braun SpaceStrategic Marketing ICU&OR SystemsNov 13, 2003
Chapter 5.2 - 1
Chapter 5.2
Overview
Variants in Application
Safety features
Therapy diversity
Anesthesia
Further Options
Chapter 5.2: B. Braun SpaceControl – Product details
Sales Manual B.Braun SpaceStrategic Marketing ICU&OR SystemsNov 13, 2003
Chapter 5.2 - 2
SpaceControl is making the pump universal…
SpaceControl is an expanded programming station for B. Braun
Space Pumps.
SpaceControl increases the features of a single pump acc. to the therapeutic needs in special therapies.
SpaceControl offers a unique user interface structure in the field of infusion technique.
The colored touch-screen display enables all users to program all Space Pumps via its excellently ergonomic handling structure.
With the integrated barcode software all relevant therapy data can be transmitted into the Space Pumps in a very safe and cost-efficient way.
SpaceControl makes it possible that one type of pump can be used in all areas of the hospital and for all necessary forms of therapies.
Sales Manual B.Braun SpaceStrategic Marketing ICU&OR SystemsNov 13, 2003
Chapter 5.2 - 3
Expanded Programming Station Connection to pump
Protected display through slide-in
technology for safe transports
Barcode reader socket
Status LED and Infrared interface
Press here, then
pull to open
Optional input for central
pressure measurem
ent
Big coloured touch screen display for a
simple and easy programming of a single pump
Pen
SpaceControl – General Overview
Sales Manual B.Braun SpaceStrategic Marketing ICU&OR SystemsNov 13, 2003
Chapter 5.2 - 4
SpaceStation – How does it work?
SpaceControl only can be used together with a connected pump.
SpaceControl can be used with a single pump as one stand-alone unit.
SpaceControl also can be used together with SpaceStation as a central user interface for all pumps.
SpaceControl does not have an integrated power source, it receives the power via the pump also if it is used with SpaceStation.
SpaceControl adapts itself to the data of the connected pump.
The „brain“ is located in the pumps.
Sales Manual B.Braun SpaceStrategic Marketing ICU&OR SystemsNov 13, 2003
Chapter 5.2 - 5
SpaceControl with SpaceStation
SpaceControl with SpaceStation
SpaceControl with Perfusor Space
SpaceControl with Perfusor Space
SpaceControl with Infusomat Space
SpaceControl with Infusomat Space
SpaceControl - Variants of Applications
Sales Manual B.Braun SpaceStrategic Marketing ICU&OR SystemsNov 13, 2003
Chapter 5.2 - 6
SpaceControl – Safety Features
Colored user surface with clearly structured menus for a safe operation
The Main MenuThe Main Menu
Sales Manual B.Braun SpaceStrategic Marketing ICU&OR SystemsNov 13, 2003
Chapter 5.2 - 7
SpaceControl – Safety Features
Comfortable and safe data input via barcode reader
Pre-set medications for a safe input of drug-related parameters.
The barcoded drug parameters can be identical with the parameters in the drug library.
Barcode- and druglist editor as a common software tool.
The barcode data include all drug parameters as well as patient ID and colored-related drug illuminations.
Sales Manual B.Braun SpaceStrategic Marketing ICU&OR SystemsNov 13, 2003
Chapter 5.2 - 8
SpaceControl – Safety Features
Safe central data input per pump
Selection of the pump directly at the pump assures that the right data will be transmitted to the right pump
1
2
Choose your pump to be modified
A log-in key at the pump adapts the required pump to SpaceControl.
Then set parameters at SpaceControl
No mix up of pumps
Sales Manual B.Braun SpaceStrategic Marketing ICU&OR SystemsNov 13, 2003
Chapter 5.2 - 9
SpaceControl – Therapy Diversity
Therapy Profiles: Standard Mode (continuous
infusion, also at Pump available)
Ramping/Tapering Profile Step- /Circadian Profile PCA OTCI
Piggy Back (also at Infusomat available)
After the Profiles have been programmed the pumps continue the infusion also without SpaceControl.
More therapy flexibility to Perfusor- and Infusomat Space with SpaceControl.
Profiles in red only programmable with SpaceControl.
Sales Manual B.Braun SpaceStrategic Marketing ICU&OR SystemsNov 13, 2003
Chapter 5.2 - 10
Ramping-/Tapering
Slide increasing and decreasing of the Infusion rate, mainly used for Total Parenteral Nutrition.
SpaceControl – Therapy Diversity
Sales Manual B.Braun SpaceStrategic Marketing ICU&OR SystemsNov 13, 2003
Chapter 5.2 - 11
Setting chronobiological profiles For oncological drugs The profile can be repeated with a choosed amount of cycles
Time
Ra
te
One Loop
Step time
Step 1
Step 2
Step 3
Step 4
Step 5
Step 6
Loop time
Step 6
Loop 1 Loop 2 … Loop n
Time
Ra
te
One Loop
Step time
Step 1
Step 2
Step 3
Step 4
Step 5
Step 6
Loop time
Step 6
Loop 1 Loop 2 … Loop n
Step-/Circadian Profile
SpaceControl – Therapy Diversity
Sales Manual B.Braun SpaceStrategic Marketing ICU&OR SystemsNov 13, 2003
Chapter 5.2 - 12
Anesthesia OperationsAnesthesia Operations
PCA & Target Controlled InfusionPCA & Target Controlled Infusion
SpaceControl – Therapy Diversity
Sales Manual B.Braun SpaceStrategic Marketing ICU&OR SystemsNov 13, 2003
Chapter 5.2 - 13
Perfusor- And Infusomat Space can be loaded with a PCA profile.
No modification of the pump housing for therapy lock necessary.
A special container compartment for the Infusomat Space offers an outstanding feature for comfortable mobility.
Three Safety levels for parameter adjustments via the service menu available.
Time
Ra
te
Observation interval
Lockout time
Basal rate
Bolus rate
Successful bolus demand
Unsuccessful bolus demand
Initial Bolus
Time
Ra
te
Observation interval
Lockout time
Basal rate
Bolus rate
Successful bolus demand
Unsuccessful bolus demand
Initial Bolus
Patient Controlled Analgesia (PCA)
SpaceControl – Therapy Diversity
Sales Manual B.Braun SpaceStrategic Marketing ICU&OR SystemsNov 13, 2003
Chapter 5.2 - 14
Safety
Locking mechanism Perfusor Space
Via a PCA key the syringe holder can be locked to prevent unauthorized remove of the syringe
Container 250 ml for adoption to Infusomat Space
Bag compartment, easily to be pushed underneath
the pump.
Patient Controlled Analgesia (PCA)
SpaceControl – Therapy Diversity
Sales Manual B.Braun SpaceStrategic Marketing ICU&OR SystemsNov 13, 2003
Chapter 5.2 - 15
B.Braun SpaceStation with 3 Perfusor Space and B. Braun SpaceControl is a combination of components
of the B.Braun Space System, optimized for
Total Intra Venous Anesthesia.
The application of certain drugs may be performed as OTCI.
Target Controlled Infusion
SpaceControl – Therapy Diversity
Sales Manual B.Braun SpaceStrategic Marketing ICU&OR SystemsNov 13, 2003
Chapter 5.2 - 16
Smallest and lightest TIVA workstation Integrated battery pack for a safe and easy transport from
induction room to OR. Easy transport of the TCI pump because OTCI is incorporated in
the pumps Lower costs due to the use of generic Propofol Higher user comfort by
- Touch Screen- Graphical color display of therapy courses
Higher patient safety by- Showing effect site concentrations- Using drug library with predefined rate-limits
Target Controlled Infusion – Features and Benefits
SpaceControl – Therapy Diversity
Sales Manual B.Braun SpaceStrategic Marketing ICU&OR SystemsNov 13, 2003
Chapter 5.2 - 17
The Workflow in Anesthesia with B.Braun Space
ICU
Induction Room Operation Theatre Recovery Room
Ward
with PCA button
Sales Manual B.Braun SpaceStrategic Marketing ICU&OR SystemsNov 13, 2003
Chapter 5.2 - 18
Features and Benefits
Cost efficiency by using OTCI instead of TCI (LOCK function) Prevention of Propofol overdose (OTCI) Reduced wake-up-times (due to lower levels in OTCI) Reduced recovery times (due to lower levels in OTCI) Better usage due to touch screen Safe operation due to the innovative simple menu structure
Target Controlled Infusion
SpaceControl – Therapy Diversity
Sales Manual B.Braun SpaceStrategic Marketing ICU&OR SystemsNov 13, 2003
Chapter 5.2 - 19
Where to sell
Completion of existing balanced anesthesia devices (highest potential)
Promotion of B.Braun SpaceAnesthesia exclusively for TIVA (has the advantage that no gas-scavenging is needed and a normal ventilator can be used)
In hospitals, in ambulatory operation centers, for general practitioners
Sales Manual B.Braun SpaceStrategic Marketing ICU&OR SystemsNov 13, 2003
Chapter 5.2 - 20
Features of SpaceControl for Anesthesia Color screen with comprehensive information Touch screen usage Structure
- Basic informantion area on top- Navigation area on the right side- Information and controlling area in the
middle PCA-button SP for Lock-Function Bolus possibility
SpaceControl – Therapy Diversity
Sales Manual B.Braun SpaceStrategic Marketing ICU&OR SystemsNov 13, 2003
Chapter 5.2 - 21
The OTCI Screen
Sales Manual B.Braun SpaceStrategic Marketing ICU&OR SystemsNov 13, 2003
Chapter 5.2 - 22
SpaceControl – Further Options: Occlusion Pressure
Automatic Pressure limit
Always having the optimal alarm triggering time
Pressure setting in steps Pressure lost alarm For a better identification of disconnected cannulae
Sales Manual B.Braun SpaceStrategic Marketing ICU&OR SystemsNov 13, 2003
Chapter 5.2 - 23
SpaceControl – Further Options: Patient Data
The complete data file of the patient at a glance.
Sales Manual B.Braun SpaceStrategic Marketing ICU&OR SystemsNov 13, 2003
Chapter 5.2 - 24
SpaceControl – Further Options: Data Lock
Data Lock in 2 levels for a single pump or centrally for all pumps in one bed-side system.
Sales Manual B.Braun SpaceStrategic Marketing ICU&OR SystemsNov 13, 2003
Chapter 5.2 - 25
SpaceControl – Further Options: History
Memory of 1000 events in each pump. SpaceControl picks up these pump data and is able to visualize these data
Sales Manual B.Braun SpaceStrategic Marketing ICU&OR SystemsNov 13, 2003
Chapter 5.2 - 26
SpaceControl – Further Options: Configuration of the Display
Configuration of the display for individual needs: Run graphic Display (Brightness, contrast Country-specific values for pressure (mmHG, bar etc.), Date, clock adjustments
Sales Manual B.Braun SpaceStrategic Marketing ICU&OR SystemsNov 13, 2003
Chapter 5.2 - 27
SpaceControl – Further Options: System Data
This function enables the user to control functions of SpaceCom (see Chapter SpaceCom)
Sales Manual B.Braun SpaceStrategic Marketing ICU&OR SystemsNov 13, 2003
Chapter 5.2 - 28
SpaceControl – Further Options: Standby
Standby offers 2 options Standby of the single pump Central Standby of all pumps in the system, e.g. for therapy interruption