106
STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi

STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi

Embed Size (px)

Citation preview

Page 1: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi

STEREOTACTIC BREAST BIOPSY

Smriti Hari Associate Professor of Radiology

All India Institute of Medical Sciences New Delhi

Page 2: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi

Percutaneous Breast Biopsy

Ultrasound guided biopsy

Mammographic (Stereotactic) biopsy

MRI guided biopsy

Modality selection depends on

Visibility of the lesion

Operator preference

Page 3: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi

Principle of Stereotactic Imaging

Page 4: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi

Stereotactic imaging - Requirements

Stereo pair

Reference point

Computer

Page 5: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi

Scout image

Page 6: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi

Stereo Pair

+ 15 degrees - 15 degrees

Page 7: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi

Parallax- apparent shift of the lesion

+ 15 degrees - 15 degrees

Scout (0 degree)

Page 8: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi

Stereotactic Biopsy- Equipment

Add-on erect unit

Advantages

Cost

Space

Efficiency

Disadvantages

Access to some calcifications limited

Vasovagal episodes

Page 9: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi

Advantages

Improved access to microcalcifications in all locations

Patient comfort

Disadvantages

Cost

Space

Efficiency

Stereotactic Biopsy-Dedicated Prone Table

Page 10: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi

Stereotactic biopsy-

Careful mammographic interpretation

Planning the procedure

Informed consent

Performing the biopsy

Specimen radiography

Clip placement

Post procedural care

Assessment of histopathological concordance

Follow up imaging

Page 11: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi

Careful mammographic interpretation

Is there an indication for biopsy BIRADS 4/5 lesions

Selected BIRADS 3 lesions

Is the lesion only visible in mammography Microcalcifications

Subtle masses

Architectural distortion

Is the lesion amenable to stereotactic biopsy

Page 12: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi

Stereotactic biopsy- Limitations

Very thin breast(< 3cm compressed thickness)

Lesions close to skin

Subareolar lesions

Posterior lesions

Page 13: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi

Planning the procedure

What to target

Target on the worse looking calcifications

What approach

Shortest distance from skin to lesion

Page 14: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi

Informed consent

Performing the biopsy

Understanding the equipment

Positioning

Targetting

Needle placement errors

Acquiring samples

Page 15: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi

Hologic MultiCare Platinum and Atec Biopsy System

Page 16: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi

•360°Access

Cranio Caudal Approach

Caudo-Cranial

Approach

Lateral Approach

Page 17: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi

Reference Point

Where 3 axes (x, y, & z) intersect at right angles

0 0

50

0

+25 -25

Z axis

Page 18: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi

Positioning

Page 19: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi

Breast should be in the centre of the hole

Page 20: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi

Only one breast otherwise loss of lateral tissue

Page 21: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi

Before rotating C-arm, move breast away from the breast platform to avoid painful twisting

Page 22: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi

Positioning

Pull in lateral breast tissue

Sweep in superior breast tissue

Stabilize with both hands while bringing in the compression paddle

Page 23: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi

Positioning

There should be no gap b/w the breast and breast platform –loss of posterior tissue

Elevated inframammary fold

Page 24: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi

Positioning

Pt positioned with area of interest in the Bx window

Be alert for pt motion Mark the corners of targeted

area

Take scout image

Page 25: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi

Scout view – Straight On

Reposition if the lesion is not imaged in the center of the window

Page 26: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi
Page 27: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi
Page 28: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi
Page 29: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi
Page 30: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi
Page 31: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi
Page 32: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi
Page 33: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi
Page 34: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi
Page 35: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi
Page 36: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi
Page 37: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi
Page 38: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi
Page 39: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi
Page 40: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi

Targeted Inferior

Page 41: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi

Mulit-Pass

Page 42: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi
Page 43: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi
Page 44: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi
Page 45: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi

Targeted off the line

Page 46: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi

Targeted inferior

Page 47: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi
Page 48: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi
Page 49: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi

Scout view – Stereo Pair

Page 50: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi

Confirm reference point

Page 51: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi

Target the lesion

Page 52: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi

Transmit the target coordinates

Page 53: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi
Page 54: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi

Targeted Inferior

Page 55: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi

Stroke vs. Stroke Margin

The distance the probe moves forward starting in the cocked position to the fired position.

STROKE

The distance from the probe tip to the detector/breast platform once fired.

STROKE MARGIN

Page 56: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi

STROKE MARGIN FORMULA

T (compression thickness)

+12mm (paddle thickness)

-Stroke (biopsy device)

-Pre-fire (Z value – pull back of instrument)

= Stroke Margin

Page 57: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi

Stroke and Stroke Margin

Compression Thickness

Compression Plate Detector or Breast Platform

12 mm

Target

Page 58: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi

Compression Thickness

Compression Plate

Detector or Breast Platform

12 mm

Target

Pre-Fire

Stroke and Stroke Margin

Page 59: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi

Compression Thickness

Compression Plate

Detector or Breast Platform

12 mm

Target

Pre-Fire Post-Fire

Stroke Stroke Margin

Stroke and Stroke Margin

Stroke Margin should always be positive

Page 60: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi

Correcting Negative Stroke Margin

Compression Thickness

Compression Plate

Detector or Breast Platform

12 mm

Target

Reposition from opposite angle

Pre-Fire

Page 61: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi

Correcting Negative Stroke Margin

Compression Thickness

Compression Plate

Detector or Breast Platform

12 mm

Target

Pre-Fire Post-Fire

Page 62: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi

Correcting Negative Stroke Margin

Compression Thickness

Compression Plate

Detector or Breast Platform

12 mm

Target

Don’t dial in as far for Pre-Fire

Adjusted Pre-Fire e.g., -4

Original Pre-Fire

-2

Page 63: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi

Correcting Negative Stroke Margin

Compression Thickness

Compression Plate

Detector or Breast Platform

12 mm

Target

Pre-Fire Post-Fire

Page 64: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi

SHORT CUT FORMULA FOR

STROKE MARGIN CALCULATION

Z + (variable) must be less than or equal to compression

Page 65: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi

Set up the biopsy device

Disinfect the skin

Attach the biopsy device

Move to Z position

Align needle tip to reference point

Z-zero the needle

Page 66: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi

Vacuum assisted Biopsy Device

Page 67: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi

Needle insertion

Move to target Move the needle close to

the skin Inject LA and make a skin

nick Advance the needle into

the breast using the Z dial until the Z differential has been met

Take pre-fire images Evaluate needle position

Page 68: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi

Prefire images

Page 69: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi

Pre-fire Position

-15 +15

x x

Pre-fire Pullback -2

Page 70: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi

Is the Needle in the Correct Position?

• Pre-fire & Post-fire Needle Position – Is the Needle Aligned to the Lesion?

– Is the Needle in the Correct Position?

– Has the Lesion Moved?

– How Do I Correct the Situation?

Page 71: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi

How Do We Know We Have A Needle Error ?

Select view stage cursors to

determine lesion/patient movement

Retarget if concerned about the position of the lesion

Page 72: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi
Page 73: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi
Page 74: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi
Page 75: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi
Page 76: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi

X or Horizontal Error

Sampling should be increased between 12 and 6 o’clock going through 3 o’clock

Page 77: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi

X or Horizontal Error

Sampling should be increased between 12 and 6 o’clock going through 9 o’clock

Page 78: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi

Y or Vertical Error

Sampling should be increased between 9 and 3 o’clock going through 12 o’clock

Page 79: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi

Y or Vertical Error

Sampling should be increased between 9 and 3 o’clock going through 6 o’clock

Page 80: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi

Z or Depth Error

The depth must be increased by advancing the probe

Page 81: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi

Z or Depth Error

The depth must be decreased by pulling back the probe

Page 82: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi

Fire the biopsy gun

Take post-fire images

Evaluate needle position

Page 83: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi

Post Fire

Page 84: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi

Post Exam

Page 85: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi

Post-fire Position

-15 +15

x x

Page 86: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi

Obtain multiple tissue cores

Take post-biopsy stereo images

Page 87: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi
Page 88: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi

Specimen Radiography

Confirms calcification retrieval

Caner can be missed, if calcification is not demonstrated* Miss rate 1% with retrieval

11% without retrieval

Recommended for US guided biopsies also

*Radiology 2004;233:251-54

Page 89: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi

VAB: Marker Clip Deployment

Deployed at the end of VAB If complete removal of lesion is

expected

If biopsy diagnosis is cancer, metallic clip facilitates hook wire localization

Also recommended prior to chemotherapy in LABC For any size of mass

Page 90: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi

Clip placement

Page 91: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi
Page 92: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi

Images for Documentation

Scout (straight on)

Scout (stereotactic pair)

Pre-fire (stereotactic pair)

Post-fire (stereotactic pair)

Specimen radiograph

Post clip placement

Post procedure mammogram (two view)

Page 93: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi

Post Procedure

Light compression at biopsy site (not at the puncture site) for 10 minutes

Patient to avoid strenuous ipsilateral arm movements for 24 hours (lifting, pushing)

Observe for local bleed/ breast enlargement

Complications of breast biopsy are rare Bleed, hematoma, infection Vasovagal attack

Page 94: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi

Misdiagnoses can be minimized Accuracy depends on showing representative

microcalcifications on specimen X-ray

Histology must be correlated with imaging

If biopsy results do not match expectation (imaging histology discordance) Re-biopsy

Malignancy in up to 33%*

Surgical excision, if high risk lesion on biopsy Malignancy in up to 31%#

Two years follow-up of benign biopsy results

*Breast Cancer Res Treat 2007;101:291-97

#Am J Surg 2006;192:534-37

Page 95: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi

Take Home Message

Accurate and reliable breast biopsy can be performed using stereotactic technique

All biopsy results must be correlated with imaging. If results are not concordant - follow up!

Radiologist should be responsible for initiating, performing and auditing results of image guided breast

Page 96: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi

Problem lesions

Lesions near the chest wall

Page 97: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi

Lesions near chest wall

Pt relaxed

Vigorous breast traction and firm compression

Arm through hole

Helps with lateral lesions

Remove the table pad

Helps with medial lesions

Arm through hole with special angled compression paddle

Page 98: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi

Special Positioning Techniques

Page 99: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi

Drop the shoulder in to view axillary tissue

Page 100: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi

Compression ~55 mmCompression ~55 mm

Compression ~55 mmCompression ~55 mm

Superficial lesion Pre-fire

Superficial lesion Post-fire

Page 101: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi

Thin breast – petite needle Pre-fire

Thin breast - petite needle Post -fire

Compression

~25 mm

Compression

~25 mm

Compression

~25 mm

Compression

~25 mm

Page 102: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi

Bolstering to increase compression thickness

Page 103: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi

Rolling towards breast platform while applying compression

Page 104: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi
Page 105: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi

Easing the breast to bring in posterior tissue

Page 106: STEREOTACTIC BREAST BIOPSY - Human Health Campus · STEREOTACTIC BREAST BIOPSY Smriti Hari Associate Professor of Radiology All India Institute of Medical Sciences New Delhi

Release compression slightly,pull breast platform towrds th breast. It will relax the posthe breast and allow more posterior tissue to be pulled in before recompression is reapplied