Scientific deep dive: clinical results from in-vivo and in- vitro competitor analysis Ramin Khoramnia, Grzegorz Łabuz, Gerd U. Auffarth [email protected] www.ivcrc.com www.djapplelab.com International Vision Correction Research Centre (IVCRC) The David J. Apple International Laboratory for Ocular Pathology Department of Ophthalmology Ruprecht-Karls-University of Heidelberg Chairman: G. U. Auffarth, MD, PhD, FEBO

Scientific deep dive: clinical results from in-vivo and in ... · Scientific deep dive: clinical results from in-vivo and in-vitro competitor analysis Ramin Khoramnia, Grzegorz Łabuz,

  • Upload
    others

  • View
    2

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Scientific deep dive: clinical results from in-vivo and in ... · Scientific deep dive: clinical results from in-vivo and in-vitro competitor analysis Ramin Khoramnia, Grzegorz Łabuz,

Scientific deep dive: clinical results from in-vivo and in-vitro competitor analysis

Ramin Khoramnia, Grzegorz Łabuz, Gerd U. Auffarth

[email protected] www.ivcrc.com www.djapplelab.com

International Vision Correction Research Centre (IVCRC)

The David J. Apple International Laboratory for Ocular Pathology

Department of Ophthalmology

Ruprecht-Karls-University of Heidelberg

Chairman: G. U. Auffarth, MD, PhD, FEBO

Page 2: Scientific deep dive: clinical results from in-vivo and in ... · Scientific deep dive: clinical results from in-vivo and in-vitro competitor analysis Ramin Khoramnia, Grzegorz Łabuz,

University Eye Clinic Heidelberg

1 = Research Grants; 2 = Travel Expenses; 3 = Lecture Fees; 4 = Consulting

1stQ1,3

Acufocus1,3

Alcon1,2,3,4

Alimera1,2,3

Allergan2,3

AMO/Johnson&Johnson1,2,3,4

Anew1

Bausch+Lomb2,3

Bayer1,2,3

Biotech1,3

BVI3

Carl Zeiss Meditec1,2,3

Contamac1

Glaukos1

Hoya1,2,3

Kowa1,2,3

Novartis1,2,3

Oculentis1,2,3

Oculus1,2,3

Ophtec3

Physiol1,3

Presbia2,3,4

Rayner1,2,3

Roche1,2,3

Santen1,2,3

SIFI1,2,3

Ursapharm1,2,3

IVCRC / DJ Apple Laboratory: Financial Disclosure 2018/2019

Page 3: Scientific deep dive: clinical results from in-vivo and in ... · Scientific deep dive: clinical results from in-vivo and in-vitro competitor analysis Ramin Khoramnia, Grzegorz Łabuz,

University Eye Clinic Heidelberg

Reversible Multifocality: Duet Procedure with bifocal Sulcoflex*

*Yildirim TM, Auffarth GU, Son HS, Mayer CS, Tandogan T, Khoramnia R. [Duet Procedure in High Myopia to Achieve Reversible Multifocality]. KlinMonbl Augenheilkd. 2019 [Epub ahead of print]

Page 4: Scientific deep dive: clinical results from in-vivo and in ... · Scientific deep dive: clinical results from in-vivo and in-vitro competitor analysis Ramin Khoramnia, Grzegorz Łabuz,

University Eye Clinic Heidelberg

Optical quality of trifocal IOL

Khoramnia R, Yildirim TM, Tandogan T, et al. [Optical quality of three trifocal intraocular lensmodels : An optical bench comparison]. Ophthalmologe. 2017.

• The implantation of

trifocal IOL into the

capsular bag has

become standard of care

in the treatment of

presbyopia.

• Trifocal supplementary

IOLs can now also be

used.

Page 5: Scientific deep dive: clinical results from in-vivo and in ... · Scientific deep dive: clinical results from in-vivo and in-vitro competitor analysis Ramin Khoramnia, Grzegorz Łabuz,

University Eye Clinic Heidelberg

Reversible Trifocality: DUET Procedure with Sulcoflex Trifocal

Page 6: Scientific deep dive: clinical results from in-vivo and in ... · Scientific deep dive: clinical results from in-vivo and in-vitro competitor analysis Ramin Khoramnia, Grzegorz Łabuz,

University Eye Clinic Heidelberg

➢ 20 D PanOptix (Alcon, USA)

➢ 20 D FineVision POD F (PhysIOL, Belgium)

➢ 20 D AT Lisa Tri (Carl Zeiss, Germany)

Optical quality measurements

OptiSpheric IOL PRO 2,

Trioptics, Germany

Assessment of the optical quality of study IOLs

➢ 0 D Sulcoflex Trifocal (IOL703F, Rayner) and

20 D RayOne Aspheric (RAO600C, Rayner)

➢ 20 D RayOne Trifocal (RA0603F, Rayner)

2 against 1

Competitor analysis

Page 7: Scientific deep dive: clinical results from in-vivo and in ... · Scientific deep dive: clinical results from in-vivo and in-vitro competitor analysis Ramin Khoramnia, Grzegorz Łabuz,

University Eye Clinic Heidelberg

Light transmission: 2 against 1

The light attenuation from surface reflections was assessed by

theoretical calculations of the reflection coefficients (R) based on Fresnel

equations:

Rs = 𝑛1cosθ𝑖 − n2 1 −

𝑛1n2

sinθ𝑖 2

𝑛1cosθ𝑖 + n2 1 − 𝑛1n2

sinθ𝑖 2

2

Rp = n1 1 −

𝑛1n2

sinθ𝑖 2− 𝑛2cosθ𝑖

n1 1 − 𝑛1n2

sinθ𝑖 2

+ 𝑛2cosθ𝑖

2

R =Rs + Rp

2

s = perpendicular polarization

p = parallel polarization

n1 = 1.336; n2 = 1.46

𝜃𝑖 = 0 to 15 degrees

Page 8: Scientific deep dive: clinical results from in-vivo and in ... · Scientific deep dive: clinical results from in-vivo and in-vitro competitor analysis Ramin Khoramnia, Grzegorz Łabuz,

University Eye Clinic Heidelberg †Sulcoflex = Sulcoflex trifocal and RayOne monofocal

MTF: 2 against 1

Page 9: Scientific deep dive: clinical results from in-vivo and in ... · Scientific deep dive: clinical results from in-vivo and in-vitro competitor analysis Ramin Khoramnia, Grzegorz Łabuz,

University Eye Clinic Heidelberg †Sulcoflex = Sulcoflex trifocal and RayOne monofocal

Through focus MTF: 2 against 1

Page 10: Scientific deep dive: clinical results from in-vivo and in ... · Scientific deep dive: clinical results from in-vivo and in-vitro competitor analysis Ramin Khoramnia, Grzegorz Łabuz,

University Eye Clinic Heidelberg

†Sulcoflex = Sulcoflex trifocal and RayOne monofocal

Strehl Ratio: Competitor analysis

Page 11: Scientific deep dive: clinical results from in-vivo and in ... · Scientific deep dive: clinical results from in-vivo and in-vitro competitor analysis Ramin Khoramnia, Grzegorz Łabuz,

University Eye Clinic Heidelberg

PanOptix Pod F AT Lisa RayOne Sulcoflex F

ar

Inte

rme

dia

te

Ne

ar

†Sulcoflex = Sulcoflex Trifocal and RayOne monofocal

USAF-Target:Competitor analysis

Page 12: Scientific deep dive: clinical results from in-vivo and in ... · Scientific deep dive: clinical results from in-vivo and in-vitro competitor analysis Ramin Khoramnia, Grzegorz Łabuz,

University Eye Clinic Heidelberg

For comparison: 1 AcrySof IOL with a refractive index of 1.55

Reflectance = 1.1%

Transmission = 98.9%

Reflectance

[%]

Transmission

[%]*

1 IOL± 0.4 99.6

2 IOLs† 0.8 99.2

*Transmission = 100 – Reflectance

±RayOne monofocal

†RayOne monofocal and Sulcoflex

Light transmission: 2 against 1

Page 13: Scientific deep dive: clinical results from in-vivo and in ... · Scientific deep dive: clinical results from in-vivo and in-vitro competitor analysis Ramin Khoramnia, Grzegorz Łabuz,

University Eye Clinic Heidelberg

Conclusions

• Combination of Sulcoflex Trifocal and RayOne Aspheric monofocal :

− Good performance regarding the MTF in the far, intermediate and near

focal point

− comparable results to “standard” trifocal IOLs for capsular bag

implantation

− No disadvantages due to additional interfaces (e.g., no light loss)

• The polypseudophakic approach can be used as a reversible procedure for

presbyopia treatment without affecting the optical quality.