Reliable, Representative & Relevant
The 3 R’s for online waste water measurement
Innovation in Environmental & Process Technology
www.envitech.co.ukwww.envitech.co.uk
Two very different analyser design approaches to
ensuring reliable, representative and relevant dataA. PBS Insitu auto-calibrating ammonia monitor
B. Lar Quick Cod / Toc Ultra High temp oxidation
PBS
Insitu
Ammonia
Lar Quick COD Ultra
Innovation in Environmental & Process Technology
www.envitech.co.ukwww.envitech.co.uk
Typical causes of sampling errors –
taken from discussion of Water & Steam sampling issue – Jonas, Power
engineering May 2005
Causes of Sampling Errors (in Order of Priority/Impact)
• Sample withdrawal...
• - not representing sampled stream (wall effects, stratification, not
isokinetic, mixing, etc.)
• - not representing all phases (solid, liquid, gas)
• Deposition in sample line (could also result in plugging)
• Deposit release (leading to spikes)
• High pressure drop resulting in insufficient sample flow
• Changes in sample flow rate
• High sample temperature (causing pH and conductivity errors)
• Chemical reactions in sample lines or coolers (oxygen reduction, etc.)
• Deposition in the pressure reducing device
Innovation in Environmental & Process Technology
www.envitech.co.ukwww.envitech.co.uk
Typical causes of sampling errors
Causes of Sampling Errors (in Order of Priority/Impact)
• Dissociation of water in the pressure reducing device - forming O2 +
H2 (high pressure and temperature water)
• Leakage of cooling water into the sample - leaking sample cooler
• Corrosion of the sampling system - generation of corrosion products
• Filters in the sampling system interfere with sampling suspended
solids
• Sorption on sample tubing and suspended oxides removes a portion of
monitored chemical species
• Leakage of air into grab sample container increases O2 and
conductivity, reduces pH, possible introduction of bacteria)
• Dissolution or precipitation and plating-out of metal species in the
sample tubing and containers
Innovation in Environmental & Process Technology
www.envitech.co.ukwww.envitech.co.uk
Representative• Flowsampler – example 1
• High Flow – Iso 5667-14 – Isokinetic sampling
• Aprox 1 m/sec
• No requirement of filtration
Innovation in Environmental & Process Technology
www.envitech.co.uk
IsoKinetic Sampling
Innovation in Environmental & Process Technology
www.envitech.co.ukwww.envitech.co.uk
Representative• Flowsampler
• High Flow
• 1 m/sec
• No requirement of filtration
Innovation in Environmental & Process Technology
www.envitech.co.ukwww.envitech.co.uk
Representative• Flowsampler
• Influence of filtration
Sample Analyser
with filter
Lar Quick
TOC Ultra
1 53.4 110
2 ➢ 50 122
Airport Site TOC
measurement
Ultrafiltration for Uv TOC analyser
Lar Quick TOC Ultra gave results
double the Uv TOC
Innovation in Environmental & Process Technology
www.envitech.co.ukwww.envitech.co.uk
RepresentativePBS , or Process Buoy, In-situ – example 2
• Measuring settled activated sludge
Innovation in Environmental & Process Technology
www.envitech.co.ukwww.envitech.co.uk
RepresentativePBS In-situ – example 2
• Measuring settled activated sludge
Innovation in Environmental & Process Technology
www.envitech.co.uk
Solenoid valve opens
permitting sample to fill
settling chamber
Solenoid valve
reopens permitting
supernatant to fill
measurement
chamber
Ammonia is measured with
gas-selective probe
Sludge /solids settles for pre
programmed time period
Reagent is
added to
increase
pH>11.5 and
convert NH4
(Ammonium) to
NH3 (Ammonia)
Sample is mixed
Innovation in Environmental & Process Technology
www.envitech.co.uk
Relevant – Example 1• Ammonia - PBS
• Uses a GSE –Minimal effect of interferences
• Only volatile amines effect the GSE probe
• The Ammonium electrode is effected by interferences such as
Potassium
• Electrode membrane is prone to drift and fouling
ISE
Ion Selective
Electrode
GSE
Gas Sensing
Electrode
Innovation in Environmental & Process Technology
www.envitech.co.ukwww.envitech.co.uk
Relevant – Example 1• Ammonia - PBS
• GSE - Use of caustic dosing with the GSE to release ammonia from
solution to gas phase
• Minimal interferences and probe is kept clean
ISE GSE
Innovation in Environmental & Process Technology
www.envitech.co.ukwww.envitech.co.uk
Relevant Filtration installed to protect the analyser from fouling often results in
changing the nature of sample being measured
• Ammonia can be reduced by slime build up resulting in local
nitrification / de-nitrification
Innovation in Environmental & Process Technology
www.envitech.co.uk
Relevant- Example 2• Measuring throughout the WWTP process with the Lar Quick COD analyser
Innovation in Environmental & Process Technology
www.envitech.co.ukwww.envitech.co.uk
Relevant• COD – Lar Quick COD Ultra
• Oxygen demand method close to dichromate oxygen demand method
• Speed of response
Typical Ratio of TOC v COD in
waste water in various industries
Innovation in Environmental & Process Technology
www.envitech.co.uk
RelevantUsing an analysis technique developed for clean water for waste water often
doesn’t perform as well as expected – e.g. Uv TOC oxidation
• Incomplete measurement
• Require lots more chemicals
• Slows the response so that spills can be missed
Innovation in Environmental & Process Technology
www.envitech.co.uk
RelevantUsing an analysis technique developed for clean water for waste water often
doesn’t perform as well as expected
• Poor recovery efficiencies – effects of high salt
• Can require significant dilution – slows response and leads to
inaccuracies
Innovation in Environmental & Process Technology
www.envitech.co.uk
How hot is 1200 oC
Innovation in Environmental & Process Technology
www.envitech.co.uk
Quick COD Ultra Method –Oxygen Demand rather than CO2 Production
Innovation in Environmental & Process Technology
www.envitech.co.ukwww.envitech.co.uk
RelevantCOD – Lar Quick COD Ultra
• Speed of response
Innovation in Environmental & Process Technology
www.envitech.co.ukwww.envitech.co.uk
Reliable
Others - Achieved ? With add on’s :• Filters
• Ultrasonics
• Air cleaning
• Chemical cleaning
• Wipers
Innovation in Environmental & Process Technology
www.envitech.co.ukwww.envitech.co.uk
Reliable• A design approach - Minimise contact between the mechanics of the
analyser and the waste water material itself
• PBS insitu – use of air to fill and remove the sample
• Lar Quick COD Ultra – use of XY robot
• Autocheck / Autocalibrate
• Calibration is always needed when measuring waste water.
PBS
Insitu
Ammonia
Quick COD Ultra
Innovation in Environmental & Process Technology
www.envitech.co.uk
Typical installations
Sample Handling Capability Crucial !
Zoogloea Biological slime is typical in final effluent
Innovation in Environmental & Process Technology
www.envitech.co.uk
Solenoid valve opens
permitting sample to fill
settling chamber
Solenoid valve
reopens permitting
supernatant to fill
measurement
chamber
Ammonia is measured with
gas-selective probe
Sludge /solids settles for pre
programmed time period
Reagent is
added to
increase
pH>11.5 and
convert NH4
(Ammonium) to
NH3 (Ammonia)
Sample is mixed
Innovation in Environmental & Process Technology
www.envitech.co.uk
Innovation in Environmental & Process Technology
www.envitech.co.uk
Reliable• Complicated sample paths through small tubes bring reliability issues
Innovation in Environmental & Process Technology
www.envitech.co.uk
Reliable• Complicated sample paths through small tubes bring reliability issues
Innovation in Environmental & Process Technology
www.envitech.co.uk
Lar Quick COD Ultra :
Innovation in Environmental & Process Technology
www.envitech.co.ukwww.envitech.co.uk
Summary:• Understanding the nature of the waste water itself and designing the
analyser to take this into account can result in Relevant,
Representative and Reliable data
• The PBS and the Quick COD Ultra are both very similar and very
different in the way this is achieved
PBS
Insitu
Ammonia
Quick COD Ultra