RADIAL HEAD PROSTHESIS SYSTEM For primary and revision joint replacement of the radial head SURGICAL TECHNIQUE

RADIAL HEAD PROSTHESIS SYSTEM

Embed Size (px)

Citation preview

Page 1: RADIAL HEAD PROSTHESIS SYSTEM

RADIAL HEAD PROSTHESIS SYSTEMFor primary and revision joint replacement of the radial head

SURGICAL TECHNIQUE

Page 2: RADIAL HEAD PROSTHESIS SYSTEM

TABLE OF CONTENTS

INTRODUCTION

SURGICAL TECHNIQUE

PRODUCT INFORMATION

Radial Head Prosthesis System 2

Indications and Contraindications 5

Preoperative Planning 6

Surgical Technique 6

Implant Removal 18

Implants 20

Instruments 22

Set List 27

Radial Head Prosthesis Surgical Technique DePuy Synthes Trauma

Page 3: RADIAL HEAD PROSTHESIS SYSTEM

2 DePuy Synthes Trauma Radial Head Prosthesis Surgical Technique

RADIAL HEAD PROSTHESIS SYSTEMFor primary and revision joint replacement of the radial head

Integrated connection screw captivated in head

Articulating Dish– Dish depth grows 0.1 mm with head

diameter from 1.9 mm–2.4 mm– Centered in implant head

The DePuy Synthes Radial Head Prosthesis System is intended to restore joint function.

System features• Modular: – Two-piece construct can accommo-

date a variety of patient anatomies• Side-loading head facilitates in-situ

insertion and assembly• Implant components are locked together

by taper locking and a connection screw

Radial head• Round and symmetrical, allowing for

adaptability and ease of sizing and implantation

• Slight trapezoidal shape features a smooth head-neck transition in order to accommodate the annular ligament

• Facilitates in-situ placement through side loading application and intuitive instrumentation

• Integrated connection screw: – Aids in the construct assembly – Engages the taper locking connection – Indicates when the head is

positively locked to the stem

Taper locking mechanism secures head to stem

Page 4: RADIAL HEAD PROSTHESIS SYSTEM

Radial Head Prosthesis Technique Guide DePuy Synthes Trauma 3

Radial stems• Available in straight or curved option• Fluted design for rotational stability• Straight stem has beveled tip to facili-

tate insertion into the medullary cavity• Curved stem is designed to fi t

the physiological bend of the proximal radius

• Thin collar aids in adjusting height and guiding the head onto the stem

Uniform press fi tThe stems are designed to be uncemented and press-fi t into the radial canal.

The surface of the stems are roughened using a chemical etching process. The roughened surface helps promote bony ongrowth. Chemical etching is a controlled manufacturing process, translating into a uniform 0.25 mm press-fi t.

Straight radial stem

Curved radial stem

Page 5: RADIAL HEAD PROSTHESIS SYSTEM

4 DePuy Synthes Trauma Radial Head Prosthesis Surgical Technique

Head Diameter: Standard +2 mm +4 mm +6 mm

18 mm

20 mm

22 mm

24 mm

26 mm

28 mm

Diameters: 6 mm 7 mm 8 mm 9 mm 10 mm

Short,Straight

Long,Curved

SYSTEM COMPONENTS

24 Head Sizes (6 Diameters x 4 Heights)

10 Stem Sizes

Page 6: RADIAL HEAD PROSTHESIS SYSTEM

Radial Head Prosthesis Surgical Technique DePuy Synthes Trauma 5

INDICATIONS AND CONTRAINDICATIONS

IndicationsThe DePuy Synthes Radial Head Prosthesis System is indicated for• Replacement of the radial head for degenerative or

post-traumatic disabilities presenting pain, crepitation, and decreased motion at the radiohumeral and/or proximal radioulnar joint with:

– Joint destruction and/or subluxation visible on x-ray – Resistance to conservative treatment• Primary replacement after fracture of the radial head• Symptomatic sequelae after radial head resection• Revision following failed radial head arthroplasty

Contraindications• Absolute contraindications include: infection, sepsis,

and osteomyelitis• Relative contraindications include: – Uncooperative patient or patient with neurologic

disorders who is incapable of following directions – Osteoporosis – Metabolic disorders which may impair bone

formation – Osteomalacia – Distant foci of infection which may spread to the

implant site – Rapid joint destruction, marked bone loss or bone

resorption apparent on roentgenogram

Page 7: RADIAL HEAD PROSTHESIS SYSTEM

6 DePuy Synthes Trauma Radial Head Prosthesis Surgical Technique

PREOPERATIVE PLANNING

Position patient

Position the patient supine on the table. Prepare the extrem-ity from the axilla to the hand. This allows rotation of the forearm and flexion and extension of the elbow during the operative replacement of the radial head.

SURGICAL TECHNIQUE1Approach

A lateral approach is most commonly used. Take care to avoid the deep branch of the radial nerve, which runs ante-rior to the capsule and the radial head. To minimize the risk of operative disruption of the lateral collateral ligament, keep the capsular incision in front of the anterior margin of the anconeus muscle and parallel to the fascial limit of the extensor carpi ulnaris. Open the annular ligament, laterally or slightly anteriorly to allow full inspection of the dam-aged radial head. In selected cases, an osteotomy of the lateral epicondyle will allow an extensile approach.

Page 8: RADIAL HEAD PROSTHESIS SYSTEM

Radial Head Prosthesis Surgical Technique DePuy Synthes Trauma 7

2Resect radial head

If the radial head is fractured and is not repairable, remove the fragments to ensure no bone is left in the joint. Using a microsagittal saw, resect any remaining radial head at the level of the radial neck, perpendicular to the axis of the neck (Figures 1 and 2).

In order to determine the cut level, the shortest construct to consider yields a total height of 12.5 mm. The tallest possible construct is 21.0 mm. Be careful to avoid excessive neck resection.

Notes: • The collar on the stem adds 1.0 mm of height to

the selected head. Head heights are shown on pages 21 and 23.

• The shortest construct is obtained with the 18 mm diameter head, standard height (Figure 3).

Head height: 11.5 mm

18 mm diameter head,standard height Collar thickness:

1.0 mm

Totalheight: 12.5 mm

Figure 2

Figure 3

Figure 1

Surgical Technique

Page 9: RADIAL HEAD PROSTHESIS SYSTEM

8 DePuy Synthes Trauma Radial Head Prosthesis Surgical Technique

3Determine radial head diameter and height

Instrument

03.402.701 Sizing Dish

Determine the head diameter by placing the resected head into the circular cutouts in the sizing dish. When there are multiple articular fragments, these can be as-sembled to recreate the native radial head. If the resected head is between sizes, select the smaller diameter. Make note of the diameter determined for use in later steps.

Notes:• Compare the resected head with a trial head to

assess the head height.• When selecting a trial head size, take into account

that the collar on the stem adds 1 mm in height.

Surgical Technique

1 mm

Page 10: RADIAL HEAD PROSTHESIS SYSTEM

Radial Head Prosthesis Surgical Technique DePuy Synthes Trauma 9

Surgical Technique

4Open radial canal

Instrument

03.402.710 Canal Finder

If the elbow is unstable, varus stress and rotation of the forearm into supination allows improved access to the medullary canal. Retraction and elevation of the proximal radial shaft/ neck can allow medullary access. Use the canal finder to initially enter the canal.

Optional technique: Planing

Instrument

03.402.712 Offset Planer

By gently rotating the planer, create a smooth contact surface on the radial neck, perpendicular to the longitudinal axis of the radial neck.

Note: Avoid excessive planing as this may increase the height of the implant head required.

Page 11: RADIAL HEAD PROSTHESIS SYSTEM

10 DePuy Synthes Trauma Radial Head Prosthesis Surgical Technique

Surgical Technique

5Broach radial canal

Instruments

03.402.726– Offset Broach, Straight Tip, 03.402.730 6 mm–10 mm

03.402.731– Offset Broach, Curved Tip, 03.402.735 6 mm–10 mm

359.221 Combined Hammer

It is important to identify the proper axial orientation. The forearm should be in midrotation with the radial tuberosity directed medially. This position is favorable for broaching and implantation as the curved broaches/implant stems will point laterally, away from the radial tuberosity. Starting with the smallest broach, prepare the canal for the stem. Use sequentially larger broaches until a tight fit is achieved. Continuing to increase broach size after a tight fit is achieved, or using excessive force while broaching the canal, may produce a longitudinal radial neck fracture. Utilize special care during this step. Note that the implant stems are slightly larger than the broaches to ensure an adequate press fit.

Notes: • Straight broaching: When broaching with a

straight broach, use an in-line (on-axis) forward and backward technique with a twist motion (1 and 2).

• Curved broaching: When broaching with a curved broach, use ONLY an in-line (on-axis) forward and backward technique (2).

• If necessary, gently hammer to fully seat the straight or curved broach.

1

2

Page 12: RADIAL HEAD PROSTHESIS SYSTEM

Radial Head Prosthesis Surgical Technique DePuy Synthes Trauma 11

Surgical Technique

6Insert trial stem

Instruments

03.402.006 – Straight Trial Radial Stems 03.402.010

03.402.106 – Curved Trial Radial Stems 03.402.110

03.402.741 Offset Stem Inserter/Extractor

Choose the trial stem that corresponds to the last broach size used. Broaches and trial stems are color coded to facilitate the selection process.

Color Stem diameter White 6 mm Purple 7 mm Green 8 mm Yellow 9 mm Blue 10 mm

Secure the trial stem to the stem inserter/extractor by turn-ing the knob and threading the stem onto the instrument. Introduce the trial stem into the canal, making sure it is fully seated. The knob of the stem inserter/extractor should face lateral and opposite the radial tuberosity. This will orient the trial stem properly for in-situ head assembly.

Notes:• The size of the trial stem is the same as the

corresponding broach.• Trial stems are smooth to facilitate insertion. Trial

stems are slightly smaller than the implant stems.• Trial stems will not have a press fit.

Page 13: RADIAL HEAD PROSTHESIS SYSTEM

12 DePuy Synthes Trauma Radial Head Prosthesis Surgical Technique

Surgical Technique

7Attach trial head

Instruments

Trial Radial Heads 03.402.018 – Standard Height, 03.402.028 18 mm–28 mm diameter

03.402.218 – 2 mm Height Extension, 03.402.228 18 mm–28 mm diameter

03.402.418 – 4 mm Height Extension, 03.402.428 18 mm–28 mm diameter

03.402.618 – 6 mm Height Extension,03.402.628 18 mm–28 mm diameter

398.41 Reduction Forceps with Points, Broad, Ratchet

03.113.019 Screwdriver Shaft, T15 StarDrive, 165 mm

311.43 Handle, with quick coupling, small

Trial heads can be attached to the trial or definitive stem for determination of the implant height.

Select a trial head that appears to correctly fill the gap. Guide the trial head onto the trial stem by hand or using the reduction forceps.

Check the head height, positioning the implant flush with the upper edge of the sigmoid notch.

Continue using different trial heads until satisfied with the height.

Implant Trial

Notes:• Trial heads have a matte finish and an aqua-

colored connecting screw. Implant heads have a shiny finish and a gray connecting screw.

• When using the reduction forceps to hold the trial head, invert the reduction forceps (jaws pointed upward) and insert tips of the forceps into the two adjacent holes. Firmly squeeze the forceps to secure the trial head.

• Secure the trial head to the stem using the screwdriver and handle with quick coupling.

Important:• Do not overtighten the screw on the trial head. • Do not use the torque limiting attachment to

tighten the screw on the trial head.

Page 14: RADIAL HEAD PROSTHESIS SYSTEM

Radial Head Prosthesis Surgical Technique DePuy Synthes Trauma 13

Surgical Technique

8Trial reduction

Secure the trial head to the stem using the screwdriver and handle with quick coupling.

Manipulate the elbow through a full range of elbow flexion- extension and forearm pronation-supination positions.

The head should articulate with the capitellum and the sigmoid notch smoothly through a full arc of motion. Some translation of the trial head relative to the capitel-lum is normal with forearm rotation.

The head should articulate up to the most proximal margin of the proximal radioulnar joint.

Taking an image intensification view may be helpful to visualize the ulnohumeral joint.

Important: Avoid unnecessary exposure to image intensification.

Confirm that the medial and lateral joints are close to symmetric and not overstuffed.

Note: A prosthesis that is too large will result in varus alignment with opening of the medial ulnohumeral joint space relative to the lateral ulnohumeral joint space. Overstuffing may have a detrimental effect on motion.

Continue to place different trial heads until satisfied with the size and motion.

Once desired size is determined, remove the trial head and the trial stem.

Loosen the screw on the trial head using the screwdriver shaft and handle with quick coupling to remove the trial head. Use the inserter/extractor to remove the trial stem.

Note: Do not use the torque limiting attachment to remove the trial head.

Page 15: RADIAL HEAD PROSTHESIS SYSTEM

14 DePuy Synthes Trauma Radial Head Prosthesis Surgical Technique

Surgical Technique

9Insert implant stem

Instruments

03.402.741 Offset Stem Inserter/Extractor

359.221 Combined Hammer

Choose the implant stem for insertion that corresponds to the last broach size used. Secure the implant stem to the stem inserter/extractor by turning the knob and threading the stem onto the instrument. Using the stem inserter/ extractor, place the stem into the canal, making sure it is fully seated. The knob of the stem inserter/extractor should face lateral and opposite the radial tuberosity during insertion, which will orient the implant stem prop-erly for in-situ head assembly. The stem inserter/extractor can also be used as an impactor. Gently hammer to fully seat the stem.

Page 16: RADIAL HEAD PROSTHESIS SYSTEM

Radial Head Prosthesis Surgical Technique DePuy Synthes Trauma 15

10Assemble implant head

Instruments

03.110.002 Torque Limiting Attachment, 1.2 Nm

03.110.005 Handle for Torque Limiting Attachment

03.113.019 Screwdriver Shaft, T15 StarDrive, 165 mm

Choose the implant head that corresponds to the appropriate trial head found during trial reduction.

Assemble the implant head to the implant stem by hand or using the reduction forceps.

Manual assemblyAssemble the implant head to the implant stem by hand.

Assembly using reduction forcepsWith the implant head inverted, tighten the jaws around the neck portion of the head.

Important: When applying the reduction forceps to the implant head, make certain the jaws do not interfere with assembly.

Using the 1.2 Nm torque limiting attachment with T15 StarDrive shaft, guide the implant head onto the implant stem by threading the captivated screw into the recess of the implant stem, engaging the taper connection. Main-tain constant compression of the head to the stem during assembly.

Surgical Technique

Page 17: RADIAL HEAD PROSTHESIS SYSTEM

16 DePuy Synthes Trauma Radial Head Prosthesis Surgical Technique

Surgical Technique

Tighten the screw until an audible click of the screwdriver handle indicates the implant head is secure and locked onto the implant stem.

Page 18: RADIAL HEAD PROSTHESIS SYSTEM

Radial Head Prosthesis Surgical Technique DePuy Synthes Trauma 17

Surgical Technique

Optional insertion technique

Instruments

03.110.002 Torque Limiting Attachment

03.110.005 Handle for Torque Limiting Attachment

03.113.019 Screwdriver Shaft, T15 StarDrive, 165 mm

03.402.744 Convex Impactor

359.221 Combined Hammer

Assemble the head to the stem on the table. Guide the implant head onto the implant stem.

Thread the integrated connecting screw into the recess of the implant stem using the 1.2 Nm torque limiting attach-ment and the T15 screwdriver shaft.

Tighten the screw on the head with the TLA until an audible click is heard, which indicates the head is secured to the stem.

Insert the assembled implant using the convex impactor and the combined hammer. Gently hammer to fully seat the stem.

Page 19: RADIAL HEAD PROSTHESIS SYSTEM

18 DePuy Synthes Trauma Radial Head Prosthesis Surgical Technique

IMPLANT REMOVAL

1Remove implant head

Instruments

03.113.019 Screwdriver Shaft, T15 StarDrive, 165 mm

311.43 Handle, with quick coupling, small

Using the handle with quick coupling and the T15 StarDrive screwdriver shaft, begin backing the head off the stem by applying torque in a counterclockwise direction. Once the screw in the head is disengaged from the stem, remove the head.

Notes: • Do not reuse implant heads.• Do not use torque limiting attachment to remove

the implant head.

Page 20: RADIAL HEAD PROSTHESIS SYSTEM

Radial Head Prosthesis Surgical Technique DePuy Synthes Trauma 19

2Remove implant stem

Instruments

03.010.059 Hammer Guide for Slide/Fixed Hammer

03.402.741 Offset Stem Inserter/Extractor

359.221 Combined Hammer

Thread the hammer guide into the top of the stem inserter/ extractor. Attach the stem inserter/extractor to the implant stem, as described in Step 9. The tuning fork section of the hammer slides onto the hammer guide. Extract the implant stem by applying gentle taps away from the proximal radius.

Note: Do not reuse implant stems.

Page 21: RADIAL HEAD PROSTHESIS SYSTEM

20 DePuy Synthes Trauma Radial Head Prosthesis Surgical Technique

IMPLANTS

Titanium Curved Radial Stems, sterile Diameter Length (mm) (mm)

04.402.026S 6 40

04.402.027S 7 42

04.402.028S 8 44

04.402.029S 9 46

04.402.030S 10 48

Titanium Straight Radial Stems, sterile Diameter Length (mm) (mm)

04.402.006S 6 24

04.402.007S 7 26

04.402.008S 8 28

04.402.009S 9 30

04.402.010S 10 32

Length

Length

1 mm collar1 mm collar

1 mm collar

Diameter

Diameter

Page 22: RADIAL HEAD PROSTHESIS SYSTEM

Radial Head Prosthesis Surgical Technique DePuy Synthes Trauma 21

Implants

CoCr Radial Heads, standard height, sterile Diameter Height (mm) (mm)

09.402.018S 18 11.5

09.402.020S 20 12.0

09.402.022S 22 12.5

09.402.024S 24 13.0

09.402.026S 26 13.5

09.402.028S 28 14.0

Height

Diameter

CoCr Radial Heads, 4 mm height extension, sterile Diameter Height (mm) (mm)

09.402.418S 18 15.5

09.402.420S 20 16.0

09.402.422S 22 16.5

09.402.424S 24 17.0

09.402.426S 26 17.5

09.402.428S 28 18.0

CoCr Radial Heads, 2 mm height extension, sterile Diameter Height (mm) (mm)

09.402.218S 18 13.5

09.402.220S 20 14.0

09.402.222S 22 14.5

09.402.224S 24 15.0

09.402.226S 26 15.5

09.402.228S 28 16.0

Height

Diameter

Diameter

CoCr Radial Heads, 6 mm height extension, sterile Diameter Height (mm) (mm)

09.402.618S 18 17.5

09.402.620S 20 18.0

09.402.622S 22 18.5

09.402.624S 24 19.0

09.402.626S 26 19.5

09.402.628S 28 20.0

Height

Diameter

Height

Page 23: RADIAL HEAD PROSTHESIS SYSTEM

22 DePuy Synthes Radial Head Prosthesis Surgical Technique

INSTRUMENTS

Curved Trial Radial Stems Diameter Length (mm) (mm)

03.402.106 6 40

03.402.107 7 42

03.402.108 8 44

03.402.109 9 46

03.402.110 10 48

311.43 Handle with quick coupling, small

359.221 Combined Hammer

398.41 Reduction Forceps with Points, Broad, Ratchet

Straight Trial Radial Stems Diameter Length (mm) (mm)

03.402.006 6 24

03.402.007 7 26

03.402.008 8 28

03.402.009 9 30

03.402.010 10 32

Length

Length

Diameter

Diameter

Page 24: RADIAL HEAD PROSTHESIS SYSTEM

Radial Head Prosthesis Surgical Technique DePuy Synthes 23

Instruments

Trial Radial Heads, standard height Diameter Height (mm) (mm)

03.402.018 18 11.5

03.402.020 20 12.0

03.402.022 22 12.5

03.402.024 24 13.0

03.402.026 26 13.5

03.402.028 28 14.0

Height

Diameter

Trial Radial Heads, 2 mm height extension Diameter Height (mm) (mm)

03.402.218 18 13.5

03.402.220 20 14.0

03.402.222 22 14.5

03.402.224 24 15.0

03.402.226 26 15.5

03.402.228 28 16.0

Height

Diameter

Trial Radial Heads, 4 mm height extension Diameter Height (mm) (mm)

03.402.418 18 15.5

03.402.420 20 16.0

03.402.422 22 16.5

03.402.424 24 17.0

03.402.426 26 17.5

03.402.428 28 18.0

Height

Diameter

Trial Radial Heads, 6 mm height extension Diameter Height (mm) (mm)

03.402.618 18 17.5

03.402.620 20 18.0

03.402.622 22 18.5

03.402.624 24 19.0

03.402.626 26 19.5

03.402.628 28 20.0

Height

Diameter

Page 25: RADIAL HEAD PROSTHESIS SYSTEM

24 DePuy Synthes Trauma Radial Head Prosthesis Surgical Technique

Instruments

03.110.002 Torque Limiting Attachment, 1.2 Nm

03.010.059 Hammer Guide for Slide/Fixed Hammer

03.110.005 Handle for Torque Limiting Attachment

03.113.019 Screwdriver Shaft, T15 StarDrive, 165 mm

03.402.701 Sizing Dish

03.402.710 Canal Finder

Page 26: RADIAL HEAD PROSTHESIS SYSTEM

Radial Head Prosthesis Surgical Technique DePuy Synthes Trauma 25

Instruments

Offset Broach, Curved Tip 03.402.731 6 mm white03.402.732 7 mm purple03.402.733 8 mm green03.402.734 9 mm yellow03.402.735 10 mm blue

Offset Broach, Straight Tip 03.402.726 6 mm white03.402.727 7 mm purple03.402.728 8 mm green03.402.729 9 mm yellow03.402.730 10 mm blue

03.402.712 Offset Planer

03.402.744 Convex Impactor

03.402.741 Offset Stem Inserter/Extractor

Page 27: RADIAL HEAD PROSTHESIS SYSTEM

26 DePuy Synthes Trauma Radial Head Prosthesis Surgical Technique

ALSO AVAILABLE INSTRUMENTS

03.402.740 Stem Inserter/Extractor

03.402.711 Planer

Broach, Straight Tip03.402.716 6 mm white03.402.717 7 mm purple 03.402.718 8 mm green 03.402.719 9 mm yellow03.402.720 10 mm blue

Broach, Curved Tip 03.402.721 6 mm white03.402.722 7 mm purple03.402.723 8 mm green03.402.724 9 mm yellow03.402.725 10 mm blue

Page 28: RADIAL HEAD PROSTHESIS SYSTEM

Radial Head Prosthesis Surgical Technique DePuy Synthes Trauma 27

RADIAL HEAD PROSTHESIS SYSTEM INSTRUMENT AND IMPLANT SET (01.402.001)

Graphic Cases61.402.001 Graphic Case for Radial Head

Prosthesis Instrument Set60.402.002 Implant Carrying Case for Radial Head

Prosthesis

ImplantsTitanium Straight Radial Stems, sterile Diameter Length (mm) (mm)

04.402.006S 6 24

04.402.007S 7 26

04.402.008S 8 28

04.402.009S 9 30

04.402.010S 10 32

Titanium Curved Radial Stems, sterile Diameter Length (mm) (mm)

04.402.026S 6 40

04.402.027S 7 42

04.402.028S 8 44

04.402.029S 9 46

04.402.030S 10 48

CoCr Radial Heads, standard height, sterile Diameter Height (mm) (mm)

09.402.018S 18 11.5

09.402.020S 20 12.0

09.402.022S 22 12.5

09.402.024S 24 13.0

09.402.026S 26 13.5

09.402.028S 28 14.0

Note: For additional information, please refer to package insert. For detailed cleaning and sterilization instructions, please refer to: www.synthes.com/cleaning-sterilization. In Canada, the cleaning and sterilization instructions will be provided with the Loaner shipments.

Page 29: RADIAL HEAD PROSTHESIS SYSTEM

28 DePuy Synthes Trauma Radial Head Prosthesis Surgical Technique

Radial Head Prosthesis System Instrument and Implant Set (01.402.001)

Instruments Straight Trial Radial Stems Diameter Length (mm) (mm)

03.402.006 6 24

03.402.007 7 26

03.402.008 8 28

03.402.009 9 30

03.402.010 10 32 Curved Trial Radial Stems Diameter Length (mm) (mm)

03.402.106 6 40

03.402.107 7 42

03.402.108 8 44

03.402.109 9 46

03.402.110 10 48 Trial Radial Heads, standard height Diameter Height (mm) (mm)

03.402.018 18 11.5

03.402.020 20 12.0

03.402.022 22 12.5

03.402.024 24 13.0

03.402.026 26 13.5

03.402.028 28 14.0 Trial Radial Heads, 2 mm height extension Diameter Height (mm) (mm)

03.402.218 18 13.5

03.402.220 20 14.0

03.402.222 22 14.5

03.402.224 24 15.0

03.402.226 26 15.5

03.402.228 28 16.0

CoCr Radial Heads, 2 mm height extension, sterile Diameter Height (mm) (mm)

09.402.218S 18 13.5

09.402.220S 20 14.0

09.402.222S 22 14.5

09.402.224S 24 15.0

09.402.226S 26 15.5

09.402.228S 28 16.0

CoCr Radial Heads, 4 mm height extension, sterile Diameter Height (mm) (mm)

09.402.418S 18 15.5

09.402.420S 20 16.0

09.402.422S 22 16.5

09.402.424S 24 17.0

09.402.426S 26 17.5

09.402.428S 28 18.0

CoCr Radial Heads, 6 mm height extension, sterile Diameter Height (mm) (mm)

09.402.618S 18 17.5

09.402.620S 20 18.0

09.402.622S 22 18.5

09.402.624S 24 19.0

09.402.626S 26 19.5

09.402.628S 28 20.0

Page 30: RADIAL HEAD PROSTHESIS SYSTEM

Radial Head Prosthesis Surgical Technique DePuy Synthes Trauma 29

Radial Head Prosthesis System Instrument and Implant Set (01.402.001)

Trial Radial Heads, 4 mm height extension Diameter Height (mm) (mm)

03.402.418 18 15.5

03.402.420 20 16.0

03.402.422 22 16.5

03.402.424 24 17.0

03.402.426 26 17.5

03.402.428 28 18.0 Trial Radial Heads, 6 mm height extension Diameter Height (mm) (mm)

03.402.618 18 17.5

03.402.620 20 18.0

03.402.622 22 18.5

03.402.624 24 19.0

03.402.626 26 19.5

03.402.628 28 20.0 03.402.701 Sizing Dish, for Radial Head Prosthesis03.402.710 Canal Finder, for Radial Head Prosthesis03.402.712 Offset Planer, for Radial Head Prosthesis

Offset Broach, Straight Tip, for Radial Head Prosthesis

03.402.726 6 mm03.402.727 7 mm03.402.728 8 mm03.402.729 9 mm03.402.730 10 mm

Offset Broach, Curved Tip, for Radial Head Prosthesis

03.402.731 6 mm03.402.732 7 mm03.402.733 8 mm03.402.734 9 mm03.402.735 10 mm

03.402.741 Offset Stem Inserter/Extractor03.402.744 Convex Impactor03.010.059 Hammer Guide for Slide/Fixed Hammer03.110.002 Torque Limiting Attachment, 1.2 Nm03.110.005 Handle For Torque Limiting Attachment03.113.019 Screwdriver Shaft, StarDrive, 165 mm311.43 Handle with Quick Coupling, small359.221 Combined Hammer 398.41 Reduction Forceps with Points, Broad,

Ratchet

Page 31: RADIAL HEAD PROSTHESIS SYSTEM

Synthes (Canada) Ltd. 2566 Meadowpine Boulevard Mississauga, Ontario L5N 6P9 Telephone: (905) 567-0440 To order: (800) 668-1119 Fax: (905) 567-3185

Synthes, Inc.1302 Wrights Lane EastWest Chester, PA 19380 Telephone: (610) 719-5000 To order: (800) 523-0322

www.depuysynthes.com

© DePuy Synthes Trauma, a division of DOI 2013. All rights reserved.J10882-B 10/13 DV015075