Upload
emmeline-brown
View
212
Download
0
Embed Size (px)
Citation preview
RAD 254 Chapter 14Control of Scatter
Break down into: Those that reduce patient dose and those that are
geometrical in nature and those not
3 factors affecting scatter (primary)
• Increased kVp
• Increased field size
• Increased patient thickness
Spatial Resolution & Contrast Resolution
• Spatial resolution may be thought of as geometric in nature (focal spot size, emission spectrum, etc. dealing with geometric image formation
• Contrast resolution – driven by scatter and other sources of “noise”
Scatter
• Increased field sizes = more scatter – collimation is the most readily available and easiest thing to lower the amount of scatter
• Patient thickness also increases scatter – compression may be used to help avoid this (IVP’s and mammo are examples)
Beam restricting deviceslimit the radiation to the patient
• Aperature diaphram (size and resultant field size are a direct proportion – draw the damn picture and figure the problems)
• Cones and cylinders – great for absorbing scatter, but are circular shaped = great for improving contrast and removing scatter BUT requires much more mAs as a result
Variable aperture diaphram
• Mandated in 1974 by the US Dept of Food and Drug Administration (mandate removed later)– Positive Beam Limitation Devices (PBL’s)
• Automatically collimate to the size of cassette/receptor in the bucky and CANNOT be a bigger size than the film/receptor
Filtration
• Filtration also will decrease the low energy rays and limit patient dose and some scatter
The GRID
Only “FORWARD” scatter is of any benefit to the radiographic image – all
other scatter degrades the image!
Scatter = LOWER contrast
• Using a grid (alternating strips of fine leaded strips with alternating radiolucent interspace material) can effectively reduce the amount of ANGELED scatter from reaching the film/recepter
Grid terms
• Grid ratio = height of the lead lines divided by the interspace WIDTH
• Grid frequency/lines per inch = the more lines per inch, the more clean up
• Grid clean up = scatter w/o a grid vs scatter reaching film with a grid AKA “Contrast Improvement Factor”
• Grid function = improve image contrast
Bucky Factor
• Refers to the AMOUNT of radiation to the patient with a grid vs W/O a grid.– Higher the grid ratio the higher the “bucky
factor”– The higher the kVp, the higher the “bucky
factor”
• Grid weight refers to how heavy it is – duh – the more lead, the heavier it is
Grid Types
• Parallel
• Crossed (cross-hatch)
• Focused– Focused - crossed
Grid Problems
• Grid cut-off = short SID’s result in the vertical, parallel strips absorbing the “diverging” beam at the outer margins of the grid/film/rescepter MOST pronounced at short SID’s
• Most grid problems are “positioning” related– Uneven grid/off level grid– Off centered (lateral decentering)– Off focus grid– Upside down focused grid
Focused Grid Misalignment
• Off level = grid cutoff across image; underexposed image (light)
• Off center = ditto
• Off focus = CR centered to one side or the other of a focused grid
• Upside down grid = severe grid cut-off (no density) at both sides of the image
Grid Ratio Selection
• 8:1 grid is the most widely used
• Grid ratio is kVp driven– Higher kVp’s warrant higher grid ratios– Higher grid ratios = higher patient dose (more
radiation needed to produce an image)– As kVp increases past MAXIMUM OPTIMUM
kVp, patient dose INCREASES
mAs – Grid considerationsAs grid ratio increases, so much
mAs
• 5:1 grid = 2 X mAs
• 8:1 grid = 4 X mAs
• 12:1 grid = 5 X mAs
• 16:1 grid = 6 X mAs
Air gap technique
• By allowing the scatter radiation to “diffuse” in the atmosphere after the patient but BEFORE the film results in a higher contrast image as the scatter diffuses and does NOT reach the film
– C-spine lateral is a good example of this