QUALITROL IST-104-1 ITM505 HW SW Instructions Rev-28691

Embed Size (px)

Citation preview

  • 8/10/2019 QUALITROL IST-104-1 ITM505 HW SW Instructions Rev-28691

    1/113

    QUALITROL

    505ITM Series

    www.qualitrolcorp.com

    QUALITROL

    505ITM-100Intelligent Transformer MonitorHardware and Software Instructions

    Document ID: IST-104-1 ECN-28691

  • 8/10/2019 QUALITROL IST-104-1 ITM505 HW SW Instructions Rev-28691

    2/113

    Legal

    www.qualitrolcorp.com

    2010 QUALITROLCompany LLC, an ISO 9001 system certified company. All rights reserved. Information subject to change without notice.

    All trademarks are properties of their respective companies, as noted herein. IST-104-1.

    Information in this document is subject to change without notice. This document is provided to purchasers of Qualitrol

    products for use in the installation, operation and servicing of such products. No other use, nor any reproduction,

    distribution or the making of any derivatives of this document is authorized, without the express prior written permission of

    Qualitrol.

    Qualitrol endeavors to ensure the accuracy and quality of its published materials; however, no warranty, expressed or

    implied, is provided. Qualitrol disclaims any responsibility or liability for any direct or indirect damages resulting from the

    use of the information in this manual or products described in it. Mention of any product or brand does not constitute an

    endorsement by Qualitrol of that product or brand.

    This document was originally composed in English and was subsequently translated into other languages. The fidelity of

    subsequent translations cannot be guaranteed. In case of conflict between the English version and another language

    version, the English version takes precedence.

  • 8/10/2019 QUALITROL IST-104-1 ITM505 HW SW Instructions Rev-28691

    3/113

    505ITM Hardware & Software Manual

    www.qualitrolcorp.com

    Contents

    Hardware Instructions ..................................................................................... 1

    Overview ................................................................................................................... 1

    QUALITROL505 Series Intelligent Transformer Monitor ............................................... 1

    About Qualitrol ................................................................................................................ 2

    About These Hardware Instructions ................................................................................ 2

    Controls, Indicators, and Modules .......................................................................... 3

    Front Panel ..................................................................................................................... 3

    Modules .......................................................................................................................... 4

    RTD Input Module ..................................................................................................... 4

    CT Input Module ....................................................................................................... 4

    AC Voltage Input Module .......................................................................................... 4

    Potentiometer Input Module ...................................................................................... 4

    DC Voltage Input Module .......................................................................................... 5

    Current Input Module ................................................................................................ 5

    Dry Contact Closure Module ..................................................................................... 5

    Powered Contact Closure Module ............................................................................. 5

    Instal lat ion ................................................................................................................ 6

    Location and Mounting.................................................................................................... 6

    Installing the Hardware ............................................................................................. 6Mounting Styles ........................................................................................................ 8

    Power Supply ................................................................................................................. 9

    Heater Option ............................................................................................................... 10

    RTD Input Modules ....................................................................................................... 10

    CT Input Module ........................................................................................................... 11

    AC Voltage Input Module .............................................................................................. 12

    Potentiometer Input Module .......................................................................................... 12

    DC Voltage Input Module .............................................................................................. 13

    Current Input Module .................................................................................................... 13

    Dry Contact Closure Input Module ................................................................................ 15

    Powered Contact Closure Input Module ........................................................................ 15

    Output Contacts ............................................................................................................ 16

    Remote Output Signals ................................................................................................. 17

  • 8/10/2019 QUALITROL IST-104-1 ITM505 HW SW Instructions Rev-28691

    4/113

    505ITM Hardware & Software Manual

    www.qualitrolcorp.com

    Communications ........................................................................................................... 18

    Operation ................................................................................................................ 19

    Automatic Operation ..................................................................................................... 19

    Viewing Channels or Viewing Mode Scrolling ......................................................... 19

    MENU Mode ........................................................................................................... 19Program Mode ........................................................................................................ 21

    Testing the System ....................................................................................................... 21

    Functional Specifications ...................................................................................... 22

    Software Instructions .................................................................................... 24

    General Informat ion ............................................................................................... 24

    Connecting to the 505ITM ...................................................................................... 24

    Install ing/Running the Software ............................................................................ 24

    Desktop Communications Window ...................................................................... 25

    File Menu ...................................................................................................................... 25

    Establishing Communications wi th the 505ITM .................................................. 26

    Troubleshooting the Communications Link ........................................................ 28

    Overview of the 505ITM Tools on the Desktop Menu .......................................... 28

    Opening a Tool ....................................................................................................... 30

    Using the Monitor Tool .......................................................................................... 30

    Using the Simulator Tool ....................................................................................... 32

    Using the Factory Calibrator Tool (Advanced users only) ................................. 33

    Using the Setup Tools ........................................................................................... 36

    System Settings ............................................................................................................ 38

    Password Setup ...................................................................................................... 39

    Installed Features ................................................................................................... 40

    Time and Date Setup .............................................................................................. 40

    Display Setup .......................................................................................................... 42

    Heater Setup ........................................................................................................... 43

    Communications ........................................................................................................... 44

    Front USB - Service Port ......................................................................................... 45

    Main 485 - Main Port ............................................................................................... 46

    Aux 485 - Auxiliary Port........................................................................................... 48

    Cancel and Load Buttons ........................................................................................ 48

  • 8/10/2019 QUALITROL IST-104-1 ITM505 HW SW Instructions Rev-28691

    5/113

    505ITM Hardware & Software Manual

    www.qualitrolcorp.com

    Input Modules ............................................................................................................... 49

    Module .................................................................................................................... 50

    Custom Scaling ....................................................................................................... 51

    CT Current Parameters ........................................................................................... 52

    Winding Temperature Setup (Calculated) ............................................................... 53Simple Calculation Setup Method ..................................................................... 53

    Advanced Calculation Setup Method ................................................................. 57

    Output Relays ............................................................................................................... 62

    Relay Tabs.............................................................................................................. 63

    Relay Control .......................................................................................................... 63

    Seasonal Setpoint ................................................................................................... 66

    Ambient Temperature Forecast ............................................................................... 67

    Cooling Equipment Exerciser .................................................................................. 70

    Matrix Controlled ..................................................................................................... 71

    Cancel and Load Buttons ........................................................................................ 72

    Output Loops ................................................................................................................ 73

    Loop Parameters .................................................................................................... 74

    Custom Range ........................................................................................................ 74

    Loop Controlled ...................................................................................................... 76

    Bank Switching ............................................................................................................. 76

    TransLife ................................................................................................................... 78

    TransLifeTMControl ................................................................................................. 79

    Consumed Life ........................................................................................................ 80

    Critical Forecast ...................................................................................................... 81

    Temperature Profiler ............................................................................................... 83

    Import ing and Exporting Conf igurat ion Fi les ...................................................... 85

    Calibrating the 505ITM ........................................................................................... 86

    Calibrating Remote Outputs for Alternate Values .......................................................... 86

    Overview and Requirements ................................................................................... 86

    Calibration Procedure ............................................................................................. 86

    Calibrating a Level Gauge to the 505ITM ...................................................................... 90

    Overview ................................................................................................................. 90

    Before You Begin .................................................................................................... 90

    Calibration Procedure ............................................................................................. 91

  • 8/10/2019 QUALITROL IST-104-1 ITM505 HW SW Instructions Rev-28691

    6/113

    505ITM Hardware & Software Manual

    www.qualitrolcorp.com

    Frequently Asked Questions About the 505ITM .................................................. 97

    Appendix ....................................................................................................... 101

  • 8/10/2019 QUALITROL IST-104-1 ITM505 HW SW Instructions Rev-28691

    7/113

    505ITM Hardware & Software Manual

    1www.qualitrolcorp.com

    Hardware Instructions

    Overview

    QUALITROL

    505 Series Intell igent Transformer Monitor

    The QUALITROL505 Series Intelligent Transformer Monitor (505ITM) for Liquid-FilledTransformers offers microprocessor technology and advanced digital signal processing toaccurately assess the health and performance of oil-filled transformers.

    The 505ITM can monitor up to four total parameters and can be adapted to suit the requiredapplication.

    Input 1: Liquid Temperatures (Main Tank Top or Bottom, Load Tap Changer) using a100 ohm Resistance Temperature Detector (RTD)

    Input 2: Winding Temperature using a Clamp-on Current Transformer (0-5, 0-10, 0-20, 0-50, 0-100 amps)

    Inputs 3 & 4: Variable from the list below (factory configured)

    o Liquid Temperature

    o Winding Temperature

    o Winding Current

    o Cooling Bank Current

    o Liquid Level

    o Switch Contact

    o Inputs from other transducers such as Tank Pressure or Dissolved Gas in Oil

    Required 505ITM inputs include:

    Resistance Temperature Detectors (100 ohm RTDs)

    Clamp-on Current Transformers (0-5, 0-10, 0-20, 0-50, 0-100 amps)

    AC Voltage (0 - 140 VAC or 0 - 320 VAC)

    Potentiometer (5K - 15K)

    Current Loops (0 - 1 or 4 - 20 mA DC)

    DC Voltage (0 - 100 mV or 0 - 10 V)

    Switch Contact Closure (powered or dry)

  • 8/10/2019 QUALITROL IST-104-1 ITM505 HW SW Instructions Rev-28691

    8/113

    505ITM Hardware & Software Manual

    2www.qualitrolcorp.com

    The 505ITM offers:

    Advanced thermal modeling of winding temperatures

    Superior temperature control for higher loads

    Standalone or networked substation monitoring Diagnostic tools for condition based maintenance

    Four, Form C adjustable relays to operate cooling equipment, signal alarms, andprovide trip functions, depending on transformer conditions

    Two, 0 - 1 or 4 - 20 mA DC loops for use with SCADA systems

    Digital communication ports RS-232/USB and RS-485

    The device is easy to install and use. It is designed to be mounted outdoors, with a heateravailable inside the enclosure, in a transformer control cabinet, and it is powered by auniversal power supply of 90 - 264 VAC 47 - 63 Hz single-phase power or 40 - 290 VDC. Itcomes with a user-friendly, computer-aided setup software that can be supplied with non-

    intrusive, easy-to-retrofit sensors.

    About Qual itrol

    Qualitrol has been a leader in supplying pressure, liquid level, and temperature controlssince 1945. As an ISO 9001 System-certified company, we are committed to providingquality and reliability, both in our products and in our service.

    About These Hardware Instructions

    These Hardware Instructions provide all the information you will need to configure, install,and operate your 505ITM. They are organized into the following sections:

    Controls, Indicators , and Modules-Diagrams and brief descriptions to help you locateand understand basic functions of each of the controls, indicators, and modules that makeup the 505ITM.

    Installation-Illustrated instructions to guide you through mounting, supplying power to, andwiring your transformer monitor.

    Operation-A description of automatic operation, as well as easy instructions on viewingparameters and set points, resetting memories, programming the 505ITM using the Keypad,and testing the system.

    Functional Specifications-A detailed, functional specification of the monitor.

    Appendix-A functional wiring diagram for your reference, an example of the front panel

    menu diagram, and a rear cover view.

  • 8/10/2019 QUALITROL IST-104-1 ITM505 HW SW Instructions Rev-28691

    9/113

    505ITM Hardware & Software Manual

    3www.qualitrolcorp.com

    Controls, Indicators, and Modules

    Front Panel

    The functions of the display and the keypad on the front panel are described below. For

    operating instructions, refer to the Operationsection.DISPLAY -Every two seconds, the display automatically scrolls through the channels anddisplays the channel name and the value of the parameter. It also displays the selectedindications of the keypad control.

    KEYPAD -The ENTERkey is used to change menu modes or accept changes to values;the TESTkey displays the set point source and value on the DISPLAY; the MENUkey isused to access the programming menu; and the RESETkey is used to reset the maximumand minimum values, while in the program menu, or unlatch any latched relays while in thenormal scrolling mode. The ARROW keys are used to move through the various menus. Fora full operation description, refer to the Operationsection.

    SET POINT INDICATORS -A red LED illuminates to indicate that the set point for the relay,

    indicated on the overlay, has been exceeded.

    RS-232 PORT or USB PORT -When used with the Qualitrol software, this port allows forthe setup of all of the parameters of the monitor. See the Software Instructionssection for afull description of this setup procedure.

    50 5 ITMINTELLIGEN T TRANSFORM ER M O NITOR

    RELAY 1

    RELAY 2

    RELAY 3

    RELAY 4

    MENU

    TEST

    RESET

    EN TER

    RS-232

    KEYPAD

    DISPLAY

    SET POINT

    INDICATORS

    USB PORT

    RS-232

    PORT

    505ITM Front Panel

  • 8/10/2019 QUALITROL IST-104-1 ITM505 HW SW Instructions Rev-28691

    10/113

    505ITM Hardware & Software Manual

    4www.qualitrolcorp.com

    Modules

    The modules comprising the 505ITM include:

    RTD Input Module

    CT Input Module

    AC Voltage Input Module

    Potentiometer Input Module

    DC Voltage Input Module

    Current Input Module

    Dry Contact Closure Module

    Powered Contact Closure Module

    RTD Input Module

    RTD Input Modules are used for monitoring liquid, ambient, or if used in a heated well,

    winding temperatures. The input to the module is either a 100 ohm platinum RTD or 10 ohmcopper RTD.

    For measuring liquid or ambient temperatures, the module is scaled over the range of -40 -120 C. If it is used in a heated well, the scale is -40 - 200 C. When ordering the 505ITMfrom the factory, the temperature requirements need to be specified.

    CT Input Module

    The CT Input Module is used to monitor winding temperatures, winding currents, motorcurrents of LTCs, or fan currents. The input to the module is a clamp on the CT, provided byQualitrol, which has an input range of 0 - 5 A, 0 - 10 A, or 0 - 20 A, other ranges areavailable, and an output of 0 - 0.333 VAC, can be used as the input to the CT Input Module.

    If this module is used for winding temperatures, the scaling is in the range of -40 - 200 C;for anything other than temperature, the range is scalable to a maximum of 99999.9 C.

    AC Voltage Input Module

    The AC Voltage Input Module is used to monitor any AC voltages under 320 VAC. The inputmodule works in conjunction with the Qualitrol Voltage Sensor, TRA-600-1, and measuresranges of either 0 - 140 VAC or 0 - 320 VAC.

    The range is scalable to a maximum of 99999.9 VAC.

    Potentiometer Input Module

    The Potentiometer Input Module is used for monitoring liquid level, flow gauges, or gasaccumulation. The input to the module is a potentiometer that is typically supplied byQualitrol. The range of the potentiometer input is up to 15,000 ohm. Typically, the 25 markon a liquid level dial is set to the center of the potentiometer range, and the high and lowreadings are simply ratios of the angular deflections from the normal setting. Qualitrolgauges use a 340 angular degree potentiometer on a 2:1 gear ratio from the dial travel tothe potentiometer travel. The Potentiometer Input Module must be calibrated on site. For

  • 8/10/2019 QUALITROL IST-104-1 ITM505 HW SW Instructions Rev-28691

    11/113

    505ITM Hardware & Software Manual

    5www.qualitrolcorp.com

    more information, see the Using the Factory Calibrator Tool (Advanced users, only)sectionin the SoftwareInstructions.

    The scaling of the 505ITM for this module is 0 - 100%.

    DC Voltage Input Module

    The DC Voltage Input Module is used for monitoring transducer outputs with an outputsignal in either the 0 - 100 mVDC or 0 - 10 VDC range.

    The range for this module is scalable to a maximum of 99999.9 VDC.

    Current Input Module

    The Current Input Module is used for monitoring transducer output with an output signal ineither the 0 - 1 mA DC or 4 - 20 mA DC range. These can be devices such as pressuretransducers for measuring main tank pressure, bushing pressure, or dissolved gas in oiltransducers.

    The input module can be configured to operate with either two-wire loop transducers, which

    are powered by the module, or three-wire transducers that are powered by the module orfrom the transducer.

    The range for this module is scalable to a maximum of 9999.99 mA DC.

    Dry Contact Closure Module

    The Dry Contact Closure Module is used for monitoring a contact and indicating on a displayif it is open or closed. An open contact means there is no connection between the contacts;a closed contact signifies a shorted connection across the contacts. This can be used, forexample, to monitor pumps or fans. The system internally treats closed and opened contactsas either a 0 or 100 numeric value. Therefore, if the module is used to control a relay, therelay set point should be set at midrange, e.g., 50.

    This module can monitor a non-powered, dry contact.

    Powered Contact Closure Module

    The Powered Contact Closure Module is used to monitor a powered contact, and it indicateson the display whether it is opened or closed. An open means there is no connectionbetween the contacts, and closed indicates a shorted connection across the contacts. Thiscan be used to monitor alarms or active fans, for example. The system internally treatsclosed and open contacts as either a 0 or 100 numeric value. Therefore, if the module isused to control a relay, the relay set point should be set at midrange, e.g., 50.

    This module can monitor a powered contact of >80 VAC/DC or >130 VAC/DC jumperselectable.

  • 8/10/2019 QUALITROL IST-104-1 ITM505 HW SW Instructions Rev-28691

    12/113

    505ITM Hardware & Software Manual

    6www.qualitrolcorp.com

    Installation

    Location and Mounting

    Installing the Hardware1. Unpack the 505ITM.

    2. Check that all of the materials on the packing slip are included.

    3. Make four holes on the transformer panel for installation of the monitor.

    Important:

    Note the mounting differences between the enclosure style and the panel-only style.

    Allow space below the monitor for the wiring harness.

    4. Install the monitor.

    Important: Ensure that the sensor type matches the input designator labels on the505ITM dust cover.

    3

    2

    1

    3

    1

    2

    3

    1

    2

    TB3INPUT MODULE 1

    3

    2

    1

    22

    26

    24

    25

    23

    18

    20

    21

    19

    TB117

    INPUT POWER: (UNIVERSAL)

    90-264 VAC, 47-63 Hz

    40-290 VDC

    WATTAGE: < 30 WATTS AC IN

    < 15 WATTS DC IN

    LINE FUSE - 3.15 AMPS

    RELAY CONTACTS:

    10 A @ 115/230 VAC

    10 A @ 30 VDC

    TB4INPUT MODULE 2

    TB5INPUT MODULE 3

    TB6INPUT MODULE 4

    SERIAL NO.

    MFG. DATE

    MODEL NO.

    28

    32

    30

    31

    29

    27

    6

    10

    8

    9

    7

    2

    4

    5

    3

    1

    12

    16

    14

    15

    13

    11

    TB1 TB1

    RELAY 210 A 6

    10

    8

    9

    7

    2RELAY 1

    10 A

    4

    5

    3

    1

    22

    25

    26

    23

    24

    21

    19

    20

    18

    17

    T/RXD-(A)

    SIGNAL GND

    RS-485

    SIGNAL GND

    T/RXD+(B)

    RXD+(B)

    RXD-(A)

    TXD-(A)

    TXD+(B)

    16

    15

    13

    14

    12

    11

    32

    31

    29

    30

    28

    27

    RELAY410 A

    RELAY 310 A

    SYSTEMSTATE

    RELAY

    10A

    CURRENT

    LOOP 2

    (+)

    (-)

    CURRENT

    LOOP 1

    (+)

    (-)

    TB25

    4

    3

    2

    1

    2

    4

    5

    3

    1

    VAC LINEINPUT(+)

    EARTH GROUND

    VACC NEUTI NPUT(-)

    VAC LINEHEATER

    VACC NEUTHEATER

    Input Modules

    Input Power

    Optional Heater

    TB1-1,17

    RS-485Interface

    Current OutputLoops

    Alarm Relays

    505ITM Back Panel

  • 8/10/2019 QUALITROL IST-104-1 ITM505 HW SW Instructions Rev-28691

    13/113

    505ITM Hardware & Software Manual

    7www.qualitrolcorp.com

    5. Wire input sensors to the correct inputs, TB3 through TB6.

    6. Connect the correct alarm control circuitry, TB1-2 through TB1-16. See the wiringschematics or engineering drawings for your specific transformer.

    Note: The optional VAC heater input is TB1-1 and TB1-17.

    Relay contact ratings: 10 A @ 115/230 VAC, 10 A @ 30 VDC

    CAUTION: The loops must be connected to the isolated analog input to avoidpotential damage to the 505ITM.

    7. If required, connect the current output loops, TB-29 through TB1-32.

    8. Connect the RS-485 communications to TB1-19 through TB2-23.

    4-wire:TB1-19 through TB1-22; signal ground on TB1-23, not an earth ground

    2-wire:TB1-25, 26; signal ground on TB1-27, not an earth ground

    9. Connect the input power, TB2-3, 5, and earth ground to TB2-1.

    Universal power supply:90 - 264 VAC, 47 - 63 Hz40 - 290 VDC

    CAUTION: Installing power outside of the recommended input range will damagethe unit.

    10. Energize the 505ITM.

    The monitor briefly displays Qualitrol and the version of the board firmware.

    Important: The following operation can cause the alarm output state to change.

    11. On the front panel, press TEST.

    Relays 1 through 4 actuate.

    Note: Relays that have been configured with the test lockout feature enabled will notoperate, but the corresponding LED will light.

    12. Upon commissioning the 505ITM, do the following steps to reset the maximum andminimum values:

    (a) Press MENU.

    (b) Scroll to the View Maximum Value column (see menu flowchart).

    (c) Press RESET.

    (d) Scroll to the View Minimum Value column.

    (e) Press RESET.(f) Press MENU.

    When the test is complete, the monitor defaults back to the scrolling mode, cyclingconsecutively through the display of each active input. You are now ready to use the ITMManager softwarewhich is found on the documentation CD-ROM, to view, change, ormodify the settings of the 505ITM to meet your requirements.

  • 8/10/2019 QUALITROL IST-104-1 ITM505 HW SW Instructions Rev-28691

    14/113

    505ITM Hardware & Software Manual

    8www.qualitrolcorp.com

    Mounting Styles

    Enclosure Style

    The 505ITM is a 3R enclosure style, designed to be back-mounted on any smooth surface.Four 0.31 0.50 in. oval slots are provided for use with fasteners. The monitor requires an

    approximate area that is 11.5 inches in width, 13.0 inches in height, and 6.5 inches in depth.An additional area below the monitor should be allowed for the wiring harness. SeeMounting the 505ITM figure for mounting centers and dimensions.

    Panel Mount Style

    The 505ITM is designed to be flush-mounted inside a transformer control cabinet on anysmooth surface. Four 0.28 in. square holes are provided for use with the 0.25 in. diametercarriage bolts. The monitor requires anapproximate area of 8.50 inches in width, 10.50inches in height, and 3.00 inches in depth. An additional area below the monitor should beallowed for the wiring harness.

    Captive Thumb Screw

    (2) Places

    3.00[76.3]

    Suggested Panel Cutout = 7.8 WD x 9.3 HI

    Mounti ng the 505ITM

  • 8/10/2019 QUALITROL IST-104-1 ITM505 HW SW Instructions Rev-28691

    15/113

    505ITM Hardware & Software Manual

    9www.qualitrolcorp.com

    Power Supply

    Supply power is applied to Terminal Block TB2 terminals 3 and 5. For DC input, positivevoltage is applied to terminal 3, and ground or negative voltage is applied to terminal 5.Chassis ground is applied to terminal 1, as shown in the Power Supply Connectionfigure.This non-metallic enclosure does not automatically provide a common grounding lug.

    Grounding must be provided as part of the installation, and it must be in accordance with therequirements of the National Electrical Code (NEC) or any international code that is in effect.

    TB25

    4

    3

    2

    1

    2

    4

    5

    3

    1

    VAC LINE INPUT (+)

    EARTH GROUND

    VACC NEUT INPUT (-)

    EARTH

    GROUND

    90-264 VAC, 47-63 Hz

    40-290 VDC

    Power Supply Connection

  • 8/10/2019 QUALITROL IST-104-1 ITM505 HW SW Instructions Rev-28691

    16/113

    505ITM Hardware & Software Manual

    10www.qualitrolcorp.com

    Heater Option

    If the Enclosure or Panel Mount monitor is supplied with the Heater Option, the supplypower, 120 or 240 VAC, as specified when ordering, is applied to Terminal Block TB1terminals 1 and 17, as shown in the Heater Option Connection figure.

    Heater Option Connection

    Qualitrol recommends that the heater circuit be installed with a fuse for safety and protectionpurposes. The fuse is the responsibility of the installer. The rating of the heater is 75 watts at120 VAC. A 240 VAC heater uses two of these heaters in a series, for a total of 150 watts.Qualitrol recommends installing a slow-blow fuse rated at 1.00 A, at 600 V.

    RTD Input ModulesFor each RTD Input Module, one standard 100 ohm RTD (or 10 ohm RTD), with a three-conductor, shielded cable is required to measure liquid and ambient temperature, or if in aheated well, winding temperature. The three conductors and the shield drain lead areconnected to the monitor, as shown in the RTD Input Wiring Connections figure.

    CAUTION: RTD wiring should be continuous, and the shield drain lead must begrounded only at the instrument.

    The monitor uses a compensation scheme for the RTD cable connection. The three-wireconnection compensates for the resistance of the connecting wires.

    RTD Input Wiring Connections

    If you are upgrading an existing Qualitrol device, or if the RTD you have installed is a four-wire RTD, DO NOT connect the fourth wire in parallel with the third wire, this will adverselyaffect the compensation scheme.

    If you have a fourth wire on the RTD connection, cut the fourth wire flush with the jacket.Qualitrol recommends a maximum length of 75 feet for RTD cabling.

  • 8/10/2019 QUALITROL IST-104-1 ITM505 HW SW Instructions Rev-28691

    17/113

    505ITM Hardware & Software Manual

    11www.qualitrolcorp.com

    Note: Due to dimensional and mounting variations, the RTD, well, and connector must beordered separately and are available from Qualitrol.

    CT Input Module

    For each CT Input Module, one standard clamp on the CT, TRA-017-X, is required to

    calculate winding temperature or currents. Input the current range as specified.

    Pull the supplied Qualitrol clamp on CT Sensor apart and place the wire which carries thecurrent to be sensed, into the CT opening. Push the CT sensor back together and check thatit is fully engaged. As shown in the CT Input Connections figure, connect the output wires ofthe clamp on CT to the terminal board.

    CT Input Connections

    The Qualitrol clamp-on CT sensor comes with 20 feet of cable. While it is not recommendedto run the wire longer than 20 feet, if the application requires a longer distance, splice thelonger cable as close to the clamp on sensor as possible. This replacement cabling shouldconsist of shielded twisted pair wire.

    Note: Because the CT sensor is isolated from the transformer, the burden is so small it canonly be measured in a laboratory. For all practical purposes, the burden sensed bythe transformer is zero.

    Caution: The 505ITM measures current using the clamp on CT sensors. Wiringcurrent directly to the 505ITM will damage the unit.

  • 8/10/2019 QUALITROL IST-104-1 ITM505 HW SW Instructions Rev-28691

    18/113

    505ITM Hardware & Software Manual

    12www.qualitrolcorp.com

    AC Voltage Input Module

    For each AC Voltage Input Module, one AC Voltage Sensor, TRA-600-1, is required tomeasure the AC voltage. The module can measure either 0 - 140 VAC or 0 - 320 VAC,50/60 Hz, as specified at the factory. Connect the voltage to be measured to the sensorinput Terminal Block TB1 of the voltage sensor (TRA-600-1). Then connect the sensor

    output to the terminal block of the AC Voltage Input Module, as shown in the AC VoltageInput Connections figure.

    AC Vol tage Input Connections

    Caution: The 505ITM measures voltage using the Voltage sensor. Wiring direct ly tothe Voltage Input Module wi ll damage the unit.

    Potentiometer Input Module

    For each Potentiometer Input Module, one standard potentiometer in the range of 1500 -15000 ohm, as supplied by various Qualitrol case assemblies, is required to measure liquidlevel, flow, or gas accumulation. The module must be calibrated on site. See the Using theFactory Calibrator Tool (Advanced users, only)section in the Software Instructions.

    The three conductors and the shield drain lead are connected to the monitor, as shown inthe Potentiometer Input Connections figure.

    CAUTION: Wiring should be continuous, and the shield drain lead must begrounded only at the instrument.

    Potentiometer Input Connections

  • 8/10/2019 QUALITROL IST-104-1 ITM505 HW SW Instructions Rev-28691

    19/113

    505ITM Hardware & Software Manual

    13www.qualitrolcorp.com

    DC Voltage Input Module

    For each Voltage Input Module, one voltage transducer, 0 - 100 mVDC or 0 - 10 VDC, asspecified, is required.

    Connect the output wires of the transducer to the terminal block as shown in the DC VoltageInput Wiring figure.

    DC Voltage Input Wiring

    Current Input Module

    The Current Input Module is configurable to accept either a 0 - 1 mA DC or a 4 - 20 mA DC(SCADA type) signal. Either of these options can be a transducer that is:

    2-wire, self-powered from the transducer

    2-wire, powered from the input module

    3-wire, powered from the input module

    For each Current Input Module, one current transducer (0 - 1 mA DC or

    4 - 20 mA DC) is required.

    Connect the output wires of the transducer to the terminal block as shown in the CurrentInput Wiring figure.

  • 8/10/2019 QUALITROL IST-104-1 ITM505 HW SW Instructions Rev-28691

    20/113

    505ITM Hardware & Software Manual

    14www.qualitrolcorp.com

    Current Input Wiring

  • 8/10/2019 QUALITROL IST-104-1 ITM505 HW SW Instructions Rev-28691

    21/113

    505ITM Hardware & Software Manual

    15www.qualitrolcorp.com

    Dry Contact Closure Input Module

    The Dry Contact (Switch) Closure Input Module is designed to detect the actuation or de-actuation of one dry, non-powered switch contact. The system internally treats closed andopen contacts as either a 0 or 100 numeric value. Therefore, if the module is used to controla relay, the relay set point should be set at midrange, e.g., 50.

    See the Contact Closure Wiring figure for the correct circuit wiring.

    Contact Closure Wiring

    Powered Contact Closure Input Module

    The Powered Contact (Switch) Closure Input Module is designed to detect the actuation orde-actuation of a powered switch contact. This is accomplished by measuring a potentialacross the switch contact of 0 volts when closed and a potential greater than 80 volts AC/DCcontact is open. For applications where a higher potential is being measured, e.g., 240 V

    AC, 250 V DC) there is a jumper on the board that can be moved that will raise the sensevoltage to 130 V, if required. The system internally treats closed and open contacts as eithera 0 or 100 numeric value. Therefore, if the module is used to control a relay the relay setpoint should be set at midrange, e.g., 50.

    See the Powered Contact Wiring figure for correct circuit wiring.

    Powered Contact Wiring

  • 8/10/2019 QUALITROL IST-104-1 ITM505 HW SW Instructions Rev-28691

    22/113

    505ITM Hardware & Software Manual

    16www.qualitrolcorp.com

    Output Contacts

    Five isolated sets of normally open/normally closed output contacts, for Output Relays 1 - 4and the System State Output Relay, are provided for controlling cooling equipment,sounding an alarm, or operating remote breaker coils. Each set is capable of switching115/230 VAC, 30 VDC @ 10 A, and the sets are accessed on Terminal Block TB1, see the

    Relay Wiring figure for the Enclosure and Panel Mount monitors.Refer to the Software Instructionsfor configuring the relays controlling signal, asfailsafe/non-failsafe, time delay, and other features.

    Refer to the back cover of the 505ITM for configuration and identification of the outputcontacts for each relay. Note that each relay is shown in the non-powered state.

    Caution:App lying h igh DC vol tage to the ou tput relays wil l damage the relays.

    22

    26

    24

    25

    23

    18

    20

    21

    19

    TB117

    28

    32

    30

    31

    29

    27

    6

    10

    8

    9

    7

    2

    4

    5

    3

    1

    12

    16

    14

    15

    13

    11

    TB1 TB1

    RELAY 210 A

    6

    10

    8

    9

    7

    2RELAY 1

    10 A

    4

    5

    3

    1

    22

    25

    26

    23

    24

    21

    19

    20

    18

    17

    16

    15

    13

    14

    12

    11

    32

    31

    29

    30

    28

    27RELAY 4

    10 A

    RELAY 310 A

    SYSTEM

    STATE

    RELAY

    10A

    Relay Wiring

  • 8/10/2019 QUALITROL IST-104-1 ITM505 HW SW Instructions Rev-28691

    23/113

    505ITM Hardware & Software Manual

    17www.qualitrolcorp.com

    Remote Output Signals

    The monitor is configured to supply two milliamp (mA) current outputs for remote indicationor use with SCADA devices. This output is proportional to the full scale of the parameterselected in the configuration. Refer to Functional Specificationssection, for maximum load.

    The output signals are accessed on Terminal Block TB1 terminals 29 through 32, See the

    Remote Output Signal Terminals figure for the Enclosure and Panel Mount monitors.

    CAUTION: All remote output signals are independent, variable current loops, drivenby the system power and designed to d rive an isolated load, such as acurrent meter. They cannot be grounded or tied together.

    22

    26

    24

    25

    23

    18

    20

    21

    19

    TB117

    28

    32

    30

    31

    29

    27

    6

    10

    8

    9

    7

    2

    4

    5

    3

    1

    12

    16

    14

    15

    13

    11

    TB1 TB1

    6

    10

    8

    9

    7

    2

    4

    5

    3

    1

    22

    25

    26

    23

    24

    21

    19

    20

    18

    17

    16

    15

    13

    14

    1211

    32

    31

    29

    30

    2827

    CURRENT

    LOOP 2

    (+)

    (-)

    CURRENT

    LOOP 1

    (+)

    (-)

    Remote Output Signal Terminals

  • 8/10/2019 QUALITROL IST-104-1 ITM505 HW SW Instructions Rev-28691

    24/113

    505ITM Hardware & Software Manual

    18www.qualitrolcorp.com

    Communications

    The monitor comes equipped with digital communication capabilities. A four-wire, RS-485connection is located on the back of the monitor. See the Four-Wire RS-485 Connectionfigure. If the operator is using a two-wire, RS-485 connection with the 505ITM, shortTXD+(B) with RXD+(B), and TXD-(A) with RXD-(A), then connect the two-wire, RS-485 as

    an A/B system. The RS-485 connection may be used with the Qualitrol Remote SoftwarePackage and other available protocols. Refer to the Software Instructionsand the 505ITMCommunications Manual, IST-105-1, for a complete description of these features. Forspecific options and/or options for a special communications protocol, contact Qualitrol orour authorized representative.

    22

    26

    24

    25

    23

    18

    20

    21

    19

    TB117

    28

    32

    30

    31

    29

    27

    6

    10

    8

    9

    7

    2

    4

    5

    3

    1

    12

    16

    14

    15

    13

    11

    TB1 TB1

    6

    10

    8

    9

    7

    2

    4

    5

    3

    1

    22

    25

    26

    23

    24

    21

    19

    20

    18

    17

    16

    15

    13

    14

    12

    11

    32

    31

    29

    30

    28

    27

    RS-485

    SIGNAL GND

    RXD+(B)

    RXD-(A)

    TXD-(A)

    TXD+(B)

    Four-Wire RS-485 Connection

  • 8/10/2019 QUALITROL IST-104-1 ITM505 HW SW Instructions Rev-28691

    25/113

    505ITM Hardware & Software Manual

    19www.qualitrolcorp.com

    Operation

    Automatic Operation

    Viewing Channels or Viewing Mode Scrol ling

    Power Up

    When power is applied to the 505ITM, it automatically begins to monitor the input andcontrols the output, based on the factory configuration that was specified by the customer.

    At power up, the display will Auto Scroll through all of the parameters being monitored at afrequency of one every two seconds.

    MENU Mode

    Viewing Channels

    (See theAppendixfor an example of the menu structure.)

    Auto Scroll, MENU, View Mode Scrolling

    Press MENU to move the display out of Auto Scrolland into Viewing Mode. The Up andDown Arrows (or ENTER) will move you individually through each parameter beingmonitored. If no key activity takes place for 60 seconds, the monitor automatically reverts to

    Auto Scroll, or press MENUagain to revert back to ViewingMode Scrolling.

    Viewing Max Readings

    MENU, View Mode Scrolling, Right Arrow

    From Viewing Mode, press the Right Arrow twice. The Up and Down Arrows (or ENTER) willmove you through the maximum values, as well as the time and date of each parameter

    being monitored. Press RESETto reset all of the maximum readings. If no activity takesplace for 60 seconds, the monitor automatically reverts to Auto Scroll, or press MENUagainto revert the system back to Viewing Mode Scrolling.

    Note: To avoid any erroneous readings caused by the installation, Qualitrol recommendsresetting the maximum values immediately after setting up the monitor.

  • 8/10/2019 QUALITROL IST-104-1 ITM505 HW SW Instructions Rev-28691

    26/113

    505ITM Hardware & Software Manual

    20www.qualitrolcorp.com

    Viewing Min Readings

    MENU, View Mode Scrolling, Right Arrow, Right Arrow

    From Viewing Mode, press the Right Arrow twice. The Up and Down Arrows (or ENTER) willmove you through the minimum values, as well as the time and date of each parameterbeing monitored. Press RESETto reset allof the minimum readings. If no key activity takes

    place for 60 seconds, the monitor automatically reverts to Auto Scroll,or press MENUagainto revert the system back to Viewing Mode Scrolling.

    Note: To avoid any erroneous readings caused by the installation, Qualitrol recommendsresetting the minimum values immediately after setting up the monitor.

    Viewing TransLifeTMValues

    MENU, View Mode Scrolling, Right Arrow, Right Arrow, Right Arrow

    From Viewing Mode, press the Right Arrow three times. The Up and Down Arrows (orENTER) will move you through the TransLifeTMvalues, time to critical point, consumed life

    and temperature profiler. If no key activity takes place for 60 seconds, the monitorautomatically reverts to Auto Scroll, or press MENUagain to revert the system back toViewing Mode Scrolling.

  • 8/10/2019 QUALITROL IST-104-1 ITM505 HW SW Instructions Rev-28691

    27/113

    505ITM Hardware & Software Manual

    21www.qualitrolcorp.com

    Program Mode

    MENU, View Mode Scrolling, Right Arrows

    From Viewing Mode, keep pressing the Right Arrow until you reach the Change Settingscolumn. Press ENTERto go to Change Settings Mode. This is the location where you canchange some of the settings of the monitor. This mode is protected. Before you are allowedto change any parameters, you must first enter a program user ID #. This is the samepassword that is used to enter the Customer Setup in the remote software. If you are notauthorized to change these settings, press MENUto go back to Viewing Mode Scrolling.

    The Up and Down Arrows allow you to change the value of the character highlighted by thedisplay, while the Right and Left Arrows move the highlighted character. Press ENTERtomove to the next setting. At the end of the Program Mode menu, press ENTERto reboot thesystem. If no activity takes place for 60 seconds, the monitor automatically reverts to AutoScroll and discards any changes that were made, or press MENUagain to revert back to

    Auto Scroll.

    50 5 ITMINTELLIGEN T TRANSFORM ER M O NITOR

    RELAY 1

    RELAY 2

    RELAY 3

    RELAY 4

    MENU

    TEST

    RESET

    EN TER

    RS-232

    KEYPAD

    DISPLAY

    SET POINT

    INDICATORS

    USB PORT

    RS-232

    PORT

    Front Panel Control

    Testing the System

    To perform a system confidence test, press and hold TEST. Press TESTto actuate all non-Test Lockout relays (see the Output Relayssection in the Software Instructions) in one-second intervals.

  • 8/10/2019 QUALITROL IST-104-1 ITM505 HW SW Instructions Rev-28691

    28/113

    505ITM Hardware & Software Manual

    22www.qualitrolcorp.com

    Functional Specif ications

    Power Supply Universal; 90 - 264 VAC, 47 - 63 Hz and 40 - 290VDC; (< 30 watts, AC in; < 15 watts, DC in)

    Front Panel

    and Interface

    Display: One easy-to-read 2-line, 16-character alphanumeric

    Liquid Crystal Display (LCD). Character size: 0.38"(9.66 mm) high x 0.19" (4.84 mm) wide.

    Status Indicators: Four Light Emitting Diodes (LED)

    Controls: Eight large keys for programmable settings and userinteraction

    Front Panel Membrane: UV stabilized polyester

    InputParameters

    Accuracy: 0.5% full scale input range

    Temperature: 100 ohm platinum (Pt100), 10 ohm copper (Cu10)RTD; simulated winding or fiber optic probe;

    Liquid/ambient range: -40 - 120 CWinding range: -40 - 200 C

    Current: Clamp-on CT, 0 - 5 A, 10 A, 20 A, 100 A and othersavailable

    DC Current Loops: 0 - 1 and 4 - 20 mA DC

    DC Voltage: 0 - 100 mVDC and 0 - 10 VDC

    AC Voltage: 0 - 140 VAC and 0 - 320 VAC; 50/60 Hz

    Potentiometer: 1500 - 15,000 ohm

    Switch Contact (dry): Open/Closed

    Switch Contact(powered):

    >80 V or >130 V open, jumper selectable; opticallyisolated

  • 8/10/2019 QUALITROL IST-104-1 ITM505 HW SW Instructions Rev-28691

    29/113

    505ITM Hardware & Software Manual

    23www.qualitrolcorp.com

    Outputs Control/Alarm Contacts: Four Form C relay contacts for cooling control,alarms, trip, etc.

    Diagnostics Alarm: One Form C relay for power, sensors and internalcircuitry diagnostics

    Heater Control: One relay

    Relay Contact Ratings: 10A @ 115/230 VAC, 10A @ 30 VDC

    Remote (SCADA)Outputs:

    Two loops; 0 - 1 mA (max resistive load 10,000 ), 4- 20 mA (max resistive load 450 ), other optionsavailable

    Data Communications: RS-232, RS-485 (monolithic transformer isolation)

    Protocols: DNP 3.0 (Level 3 slave), Modbus, ASCII

    Immunity Dielectric Isolation(High-Pot):

    2500 V AC, 60 seconds to ground

    Surge WithstandCapability: IEEE C37.90.1

    Conducted/RadiatedEmissions:

    IEC 61000-6-1

    Conducted/RadiatedRadio FrequencyImmunity:

    IEC 61000-6-2

    Safety: IEC 61010-1

    Environment Temperature Range: -40 - +72 C

    Storage Temperature: -50 - +85 C

    Humidity: 90% non-condensing

    Vibration: 60/120 Hz @ .004 inch displacement

    Shock: 10 Gs half-sine, in 3-orthogonal planes

    WeatherproofEnclosure:

    NEMA #3R (vented), UV stabilized, corrosion-prooffiberglass

    Weight Approximately: Weatherproof enclosure version: 8.7 lb (4.0 kg)

    Panel Mount version: 4.0 lb (1.82 kg)

    Terminal Connectors: Accepts wire sizes from 24 - 12 AWG

  • 8/10/2019 QUALITROL IST-104-1 ITM505 HW SW Instructions Rev-28691

    30/113

    505ITM Hardware & Software Manual

    24www.qualitrolcorp.com

    Software Instructions

    General InformationThe software for the QUALITROL505ITM Series Monitor runs on a personal computer (PC)that connects to the 505ITM with an RS232/USB or RS485 cable for monitoring, simulation,calibration, and setup. You must do the initial setup through the RS232/USB connection onthe front panel of the monitor. After that, you can use either the RS232/USB or the RS485

    connection. Qualitrol recommends that you use a RS232 male-to-female DB-9 extensionstraight-thru (not a null) shielded cable or a USB-to-USB cable suitable for substationenvironments.

    Note: Throughout these instructions, the software for the QUALITROL505ITM SeriesMonitor is also referred to as the remote software, because it allows you tocommunicate remotely with the 505ITM.

    Connecting to the 505ITM1. Check that the 505ITM is connected to the main power and is energized.

    2. Connect a USB cable between the USB port on your PC and the port on the 505ITM.

    - or -Connect an RS232 cable between the RS232 port on your PC and the port on the505ITM.

    Installing/Running the SoftwareThe software runs on Microsoft WindowsNT, 98, 2000, XP, and Vista operating systems.You can run the software directly from the CD provided by Qualitrol or from files installed onyour PC.

    To install the software from the CD:

    1. Place the CD into the CD drive of your PC.

    2. In the CD drive directory, double-click the file named Setup.exe. This will automaticallystep through the procedure for loading the 505ITM software in a Qualitrol folder in theprogram directory and install a shortcut on the desktop.

    3. Clicking on the shortcut the Qualitrol ITM Manager windowdisplays.

  • 8/10/2019 QUALITROL IST-104-1 ITM505 HW SW Instructions Rev-28691

    31/113

    505ITM Hardware & Software Manual

    25www.qualitrolcorp.com

    Desktop Communications WindowThe Desktop Communications window provides the communications interface with the505ITM and your PC.

    The software automatically loads a default communications data file nameddefault505ITM.qco on your desktop when it opens the Desktop Communicationswindow.

    File Menu

    The Filepull-down menu provides several options for selecting a communications file.

    Open Opens an existing communications file (*.qco).

    Save Saves the communications data under the current communications filename.

    Save As... Saves the communications data with a new communications file name.

    Close Closes all applications and communications with the 505ITM, andcloses the Desktop Communicationswindow. The SaveCommunication Setupwindow displays if any communications

    parameters have been changed and not saved before the Closecommand.

  • 8/10/2019 QUALITROL IST-104-1 ITM505 HW SW Instructions Rev-28691

    32/113

    505ITM Hardware & Software Manual

    26www.qualitrolcorp.com

    Establishing Communications with the 505ITMFrom theDesktop Communicationswindow, select, set, or observe the following items asrequired:

    Address Each 505ITM has an assigned address that allows multiple 505ITMs tobe used with a common communication link. The factory defaultconfiguration has an address of 00. Select the correct address from thepull-down menu.

    Baud Rate The baud rate on your PC must be the same as that on the 505ITMmonitor with which you are communicating. The range is 1,200 -38,400. The default is 19,200. (If using the USB port the baud rate goesfrom 1200 to 230,400). Select the correct baud rate from the pull-downmenu. If the baud rate is unknown, it is possible to view it on the LCDdisplay of the 505ITM. Enter the MENU Mode and go to the ChangeSettings column (see the MENU Modesection).

    COM Allows you to select the serial communications port on the PC fromCOM1 COM255. The default is COM1. Select the communication portthat the serial port is connected to on your PC.

    Off line You can run the software offline without being connected to the505ITM.

    1. With no communications to the 505ITM established, click the Offlinecheck box.

    2. Click on the Connect button.

    3. Select the 505ITM icon and hit the Connect button.

    4. Type demointo the Passwordfield of either the Setup - Customeror Setup - Factoryapplication section.

    The Setupwindow opens and displays the available features. Thisfeature is useful during training or to learn how the product functionswhen there is no 505ITM available.

    Important: Changes made in the Demo mode are not permanent.These modifications cannot be saved or downloaded intoa device.

    Connect Button Click to start communications with a 505ITM.

    TX and RX Displays transmission activity.

    Note: Short transmissions will often occur too fast to be viewed by theoperator.

    Close AllButton

    Closes all open 505ITM applications, terminates communications, andcloses the Desktop Communicationswindow. The SaveCommunication Setupwindow displays if any communicationsparameters have been changed and not saved before clicking [Close

    All].

  • 8/10/2019 QUALITROL IST-104-1 ITM505 HW SW Instructions Rev-28691

    33/113

    505ITM Hardware & Software Manual

    27www.qualitrolcorp.com

    A successfu l communication s l ink is ind icated when:

    No error message displays.

    An unsuccessfu l communications l ink is ind icated when:

    A Communications link failure message displays.

    If this error message displays, troubleshoot the communications link.

  • 8/10/2019 QUALITROL IST-104-1 ITM505 HW SW Instructions Rev-28691

    34/113

    505ITM Hardware & Software Manual

    28www.qualitrolcorp.com

    Troubleshooting the Communications LinkIf a problem occurs in establishing a communications link between the PC and the 509 ITM,check the following:

    Is the cable a straight-through cable (not a null) and is it securely connected to the PC

    and the 505ITM? Is the 505ITM energized?

    Is the cable connected to the same COM port displayed in the Desktop MenuCommunications?

    Does the 505ITM device address match the selection in the Desktop MenuCommunications Addressfield?

    Does the 505ITM service port baud rate match the baud rate in the Desktop MenuCommunications? The default baud rate is 19,200 baud.

    Important: If the 505ITM service port was reconfigured to another baud rate, the DesktopMenu Communications baud rate must match the new configuration.

    RETRY Button Click this button to attempt to establish communications with the505ITM again.

    IGNORE Button Click this button to ignore the communications error. However, forcorrect operation of the system, it is essential that communication to the505ITM is established and functioning correctly.

    ABORT Button Click this button to discontinue further attempts to establishcommunications with the 505ITM.

    Overview of the 505ITM Tools on the Desktop Menu

    There are five 505ITM software tools available to run from theDesktop Menu window:

    Monitor Allows remote operation of the 505ITM.

    Simulator Allows simulation of field input signals/sensors.

    FactoryCalibrator

    Allows calibration of 505ITM inputs and outputs. This tool isconditionally blocked to the user.

    Customer Allows you to configure the 505ITM, including setup of cooling control,

  • 8/10/2019 QUALITROL IST-104-1 ITM505 HW SW Instructions Rev-28691

    35/113

    505ITM Hardware & Software Manual

    29www.qualitrolcorp.com

    Setup alarms, communications, and other features.

    Factory Setup Allows the factory to configure the 505ITM. This tool is not accessible tothe user.

    Password Each tool requires a password. Passwords are set at the factory, butyou can change them when you are in the tool. The factory setpasswords are:

    Monitor 111111

    Simulator 111111

    Calibrator If required, call Qualitrol ApplicationEngineering at 585-586-1515 x 410.

    Customer Setup 111111

    The Factory Setup tool can only be entered by Qualitrol factory

    personnel. This section is reserved for the factory except in the Demomode(offline).

    Remember MeCheck Box

    By checking in this check box, the password is stored in thecommunications file so that you do not have to re-enter it every timeyou start the tool.

    Important: Enabling this feature will disable security protection andallow anyone using this computer to gain access to thesoftware.

    Start Button Activates the selected tool.

  • 8/10/2019 QUALITROL IST-104-1 ITM505 HW SW Instructions Rev-28691

    36/113

    505ITM Hardware & Software Manual

    30www.qualitrolcorp.com

    Opening a ToolTo open any of the 505ITM software tools, do the following:

    1. Establish communications with the 505ITM.

    2. From the Desktop Menuwindow, enter the correct password for the tool you wish tostart.

    Note: If the Remember Mefeature is enabled, you do not have to enter a password.

    3. Click [Start].

    Using the Monitor ToolThe Monitor tool provides remote monitoring and control of the 505ITM by simulating thecontents on the 505ITM LCD, and relay LEDs appear on your PC screen. The monitorrequests an update from the 505ITM three times every second.

  • 8/10/2019 QUALITROL IST-104-1 ITM505 HW SW Instructions Rev-28691

    37/113

    505ITM Hardware & Software Manual

    31www.qualitrolcorp.com

    There are eight control buttons on your PC screen: Test, Reset, Menu, Enter, Up, Down,Left, and Right. These buttons are activated by clicking on them with the left mouse button.

    Activating a button in this way has the same effect as pressing the same button on the505ITM monitor itself. See the Operationsection of the Hardware Instructionsfor specificinformation on the control and operation of the 505ITM.

    Click [Close] to exit the tool.

    Note: The actual 505ITM continues to operate when the Monitor tool is closed.

  • 8/10/2019 QUALITROL IST-104-1 ITM505 HW SW Instructions Rev-28691

    38/113

    505ITM Hardware & Software Manual

    32www.qualitrolcorp.com

    Using the Simulator ToolThe Simulator tool allows you to simulate values regardless of the actual inputs that thesensors register. The 505ITM responds to the input signal changes from the Simulatorwindow as if the signals originated in the field. This mode is useful for testing items such asrelay connections and wiring.

    When you click [Start Simulation]:

    The field inputs disconnect from the actual field signals and connect to the signals from theSimulator tool. You can now change the simulated signals remotely as you desire. The

    505ITM responds to the simulated signals as if they were the actual field signals.

    With the exception of the CT, Temperature signals, the transfer from the field to thesimulation is almost undetectable because the initial simulation value is set to the last valueof the actual field signal. The winding temperature (CT, Temperature) is a calculated valueusing load currents, oil temperatures, and time constants. All of these are bypassed in thesimulation mode where the calculated values are replaced with the simulated values.

    Important: While the simulation is active, the actual field signals are not being monitored.If you have Trip functions connected to the 505ITM, you can activate them insimulation mode.

    When you click [End Simulation]:

    The field signals reconnect to the input of the 505ITM. The 505ITM responds immediately tothe actual field signals.

    Note: If you make no changes to any of the input signals for more than 3 minutes, thesimulation ends automatically. Whenever any button is pressed, the 3-minute timerresets.

  • 8/10/2019 QUALITROL IST-104-1 ITM505 HW SW Instructions Rev-28691

    39/113

    505ITM Hardware & Software Manual

    33www.qualitrolcorp.com

    Using the Factory Calibrator Tool (Advanced users only)The Factory Calibrator tool allows you to calibrate most inputs. When you f irst enter theCalibrator - Signal Selectionwindow, the Input 1channel is selected by default.

    Important: All channels, except for the Potentiometer Input module, are calibrated at the

    factory. Under normal operation, they will not need to be calibrated again. Youmust have a separate password for this application for the ITM manager toallow you to use this tool.

  • 8/10/2019 QUALITROL IST-104-1 ITM505 HW SW Instructions Rev-28691

    40/113

    505ITM Hardware & Software Manual

    34www.qualitrolcorp.com

    Calibrating the Input Channels

    1. Open the Factory Calibrator tool.

    2. Select the input channel to calibrate.

    3. Click [Calibrate]. The Input Channel Calibrationwindow displays.

    4. In the terminal block that corresponds to the input module being calibrated, enter thesignal value displayed in the Set input signal tofield.

    Important: The Set input signal tofield displays the value of the external signal thatmust be applied to the module terminals. For the RTD Input Modules, theinput signal is expressed in terms of degrees C. If you continue the calibrationprocedure without inputting the correct excitation (stimulus) signal, you willoverwrite the calibration with invalid data. This will cause the monitor tomalfunction.

    5. After you have entered the input signal on the terminal block and it is stabilized, click[Next] to accept the value in the Set input signal tofield.

    Note: All channels require at least a two-step calibration procedure. The [Next] buttonchanges to [Finish] for the second step in the procedure. If you want to return to theprevious step, click [Back].

    6. Enter the value for the second input and click [Finish]. The calibration of this channel iscomplete.

  • 8/10/2019 QUALITROL IST-104-1 ITM505 HW SW Instructions Rev-28691

    41/113

    505ITM Hardware & Software Manual

    35www.qualitrolcorp.com

    Calibrating the Output Channels

    The analog outputs can be adjusted in a four-point calibration; see the Output Calibrationsection of the manual for more details. However, this calibration should not be necessarybecause it is done at the factory.

    Saving Calibration s in FilesYou can save calibration values in files using the Filepull-down menu. These files have thedefault file extension .qca3.

    Import Allows you to import a valid calibration file stored in the PC.

    Export Allows you to save the current calibration to a previously specified filefor future use.

    Export As... Allows you to save the current calibration in a newly specified file forfuture use.

  • 8/10/2019 QUALITROL IST-104-1 ITM505 HW SW Instructions Rev-28691

    42/113

    505ITM Hardware & Software Manual

    36www.qualitrolcorp.com

    Using the Setup ToolsThe User Setup tool allows you to set all the adjustable parameters of the 505ITM. TheFactory Setup tool is used to set the hardware configurations of the monitor. The CustomerSetup limits access to some of the 505ITM parameters, such as serial number,

    manufacturing date, and configuration date.

    Note: You can save configurations as f iles, which you can use to restore settings or to pre-configure before connecting to the 505ITM. See Importing and ExportingConfiguration Files.

    When you open the Customer Setup tool, the Customer - Setupwindow displays. When youopen the Factory Setup tool, the Factory - Setupwindow appears. The windows for bothtools are the same and allow for a specific subsystem to be configured.

  • 8/10/2019 QUALITROL IST-104-1 ITM505 HW SW Instructions Rev-28691

    43/113

    505ITM Hardware & Software Manual

    37www.qualitrolcorp.com

    The Setup windows have three main sections:

    ModelInformation

    Displays the fields for Serial number, Manufacturing date, and the lastConfiguration dateof the 505ITM. The Configuration datefielddisplays when the 505ITM was last updated.

    Model Type Displays that the monitor is a Transformer Monitor.

    System Index Displays the system buttons required to set up and view the parametersof the 505ITM.

    Click the buttons to begin setting up the configurations.

    [System Settings]

    [Communications]

    [Input Modules]

    [Output Relays]

    [Output Loops]

    [Bank Switching]

    [TransLifeTM]

  • 8/10/2019 QUALITROL IST-104-1 ITM505 HW SW Instructions Rev-28691

    44/113

    505ITM Hardware & Software Manual

    38www.qualitrolcorp.com

    System Settings

    The System Settingswindow opens when you click [System Settings] from the Setupwindow. This window allows you to set the configuration for the items displayed in thedifferent sections of the window.

    More Information

    Password Setup

    Installed Features

    Time and Date Setup

    Display Setup

    Heater Setup

  • 8/10/2019 QUALITROL IST-104-1 ITM505 HW SW Instructions Rev-28691

    45/113

    505ITM Hardware & Software Manual

    39www.qualitrolcorp.com

    Password Setup

    This section of the System Settingswindowallows you to configure the passwords thatprovide access for the different software tools. The passwords are stored in the 505ITM andhave the following characteristics and requirements:

    Passwords consist of one to six alphanumeric characters, and they are case-sensitive.

    The password for the User Setup tool is also the password used to enter themodification section of the MENU Mode from the front panel of the 505ITM.

    The password for the Calibrator tool must be different from the other tool passwords.This is a precaution to prevent users from accidentally entering the calibration sectionof the monitor.

    There is a hierarchy to the password system. The password for the Customer Setuptool will also work in the Monitor and Simulatortools, even if they have differentpasswords. This is because all of the passwords can be changed in the CustomerSetup tool, so it is treated as the master password.

    When passwords are changed in the Password Setup tool, the user must enter thenew password values the next time the tools are opened.

    There are five Passwordfields and five corresponding Verifyfields. The Passwordfieldsaccept the password values for the associated tools. The Verifyfields are used to check thecorresponding password.

    Note: The Factory Setup password is not available to the customer.

    To enter a new password:

    1. Type the new password value into the Passwordfield. Use one to six alphanumericcharacters.

    2. Type the same password value into the Verifyfield. Use the same alphanumericcharacters and the same case as you did in the Passwordfield.

    3. Click [Load].

    The software compares the values in the Passwordand Verifyfields when they areloaded into the 505ITM. If the values do not match, an error message displays.

  • 8/10/2019 QUALITROL IST-104-1 ITM505 HW SW Instructions Rev-28691

    46/113

    505ITM Hardware & Software Manual

    40www.qualitrolcorp.com

    Installed Features

    This section of the System Settingswindow displays the installed features available in the505ITM. If a feature is installed, there will be a check mark in the check box next to thelegend.

    Time and Date Setup

    This section of the System Settingswindowallows you to configure the time and date valuesstored in the 505ITM.

    Note: The time and date values can also be changed along with the front panel operations.

  • 8/10/2019 QUALITROL IST-104-1 ITM505 HW SW Instructions Rev-28691

    47/113

    505ITM Hardware & Software Manual

    41www.qualitrolcorp.com

    Current PCTime

    Displays the current time of the PC that is connected to the 505ITM whenthe Time and Date Setupwindow is opened. This value is notautomatically updated. The current time of this PC is loaded into the505ITM when you click [Load]. If the Current PC Timevalue is notcorrect:

    1. Adjust the PC clock to the correct time.

    2. Click [Load] on the System Settingswindow.

    3. Continue with the setup procedure.

  • 8/10/2019 QUALITROL IST-104-1 ITM505 HW SW Instructions Rev-28691

    48/113

    505ITM Hardware & Software Manual

    42www.qualitrolcorp.com

    Time ZoneDifference

    Allows you to adjust for a difference in time zones between the location ofsetup and the location of the installed 505ITM. If the two locations are thesame, set the value to 0. Otherwise, select the appropriate number ofhours, positively or negatively, to provide for a correct time at theinstallation location. For example, if the setup is done in New York andthe installation is in California, set the Time Zone Difference to -3 hours.

    Current PCdate

    Displays the current date of the PC that is connected to the 505ITM whenthe Time and Date Setupwindow is opened.

    505 Time anddate

    Displays the current values for the date, time, and time zone in the505ITM.

    Display Setup

    This section of the System Settingswindowallows you to adjust the contrast and intensity ofthe 505ITM Display.

    Brightness The slider adjusts the intensity of the back light of the display. You alsohave the option to disable the back light.

    Contrast The slider adjusts the display contrast to an appropriate viewing level.

  • 8/10/2019 QUALITROL IST-104-1 ITM505 HW SW Instructions Rev-28691

    49/113

    505ITM Hardware & Software Manual

    43www.qualitrolcorp.com

    Heater Setup

    This section of the System Settingswindowallows you to configure the 505ITM Heateroption. The Heater is not a standard feature and has to be ordered. It is used in extremelycold or moist environments. The Heater is controlled by a 505ITM internal temperaturemeasurement. If the Heater option is installed, the following fields are available for setup.

    HeaterFunction

    Allows you to enable or disable this feature.

    HeaterSetpoint

    Allows you to set the temperature at which the Heater is energized. TheHeater energizes when the internal temperature of the 505ITM dropsbelow the setpoint value.

    HeaterHysteresis

    Allows you to set the temperature at which the Heater is de-energized.The Heater de-energizes when the internal temperature of the 505ITMrises above the sum of the setpoint and the hysteresis values.

  • 8/10/2019 QUALITROL IST-104-1 ITM505 HW SW Instructions Rev-28691

    50/113

    505ITM Hardware & Software Manual

    44www.qualitrolcorp.com

    Communications

    The Communications Setupwindow opens when you click [Communications] from the Setupwindow. This window allows you to configure the 505ITM communications ports displayed inthe different tabs of the window.

    More Information

    Front USB - Service Port

    Front Serial Service Port

    Main 485 - Main Port

    Aux 485 - Auxiliary Port

    Cancel and Load Buttons

  • 8/10/2019 QUALITROL IST-104-1 ITM505 HW SW Instructions Rev-28691

    51/113

    505ITM Hardware & Software Manual

    45www.qualitrolcorp.com

    Front USB - Service Port

    This section of the Communications Setupwindowincludes the fields described below.

    Type Indicates the type of signaling hardware used for the Service Port. Thisfield is for informational purposes only.

    Baud Rate Configures the baud rate that is used to communicate over the USB

    Service Port at the front panel. The default value is 19,200 baud. TheDesktop Menu software automatically configures the PC's COM port withthe correct parameters for communicating with the Remote Software.

  • 8/10/2019 QUALITROL IST-104-1 ITM505 HW SW Instructions Rev-28691

    52/113

    505ITM Hardware & Software Manual

    46www.qualitrolcorp.com

    Front Serial - Service Port

    This section of the Communications Setupwindowincludes the fields described below.

    Type Indicates the type of signaling hardware used for the Service Port. Thisfield is for informational purposes only.

    Baud Rate Configures the baud rate that is used to communicate over the RS232Service Port at the front panel. The default value is 19,200 baud. TheDesktop Menu software automatically configures the PC's COM port withthe correct parameters for communicating with the Remote Software.

  • 8/10/2019 QUALITROL IST-104-1 ITM505 HW SW Instructions Rev-28691

    53/113

    505ITM Hardware & Software Manual

    47www.qualitrolcorp.com

    Main 485 - Main Port

    This section of the Communications Setupwindowincludes the fields described below.

    Type Indicates the type of signaling hardware used for the Main Port. This isan information-only field.

    InstalledProtocol

    The Qualitrol ITM Manager software comes standard with every 505ITMunit.

    OptionalProtocols

    Note: All of these options are upgradeable by the factory.

    DNP Indicates if the DNP V3.00 Level 3 communications protocol isinstalled.

    Modbus Indicates if the Modbus communications protocol isinstalled.

    QASCII Indicates if the Qualitrol ASCII communications protocol is

  • 8/10/2019 QUALITROL IST-104-1 ITM505 HW SW Instructions Rev-28691

    54/113

    505ITM Hardware & Software Manual

    48www.qualitrolcorp.com

    installed.

    Protocol Selects between the available communications protocols to be used bythe Main Port communications.

    Address Indicates the communication address of the 505ITM used by thespecified protocol.

    Baud Rate Selects the desired baud rate for the Main Port communications protocol.The default value is 9,600.

    Number ofData Bits

    Selects the desired number of data bits for the Main Portcommunications protocol. The default value is 8.

    Parity Selects the desired type of parity for the Main Port communicationsprotocol. The default is none.

    Number ofStop Bits

    Selects the desired number of stop bits for the Main Port communicationsprotocol. The default value is 1.

    Aux 485 - Auxi liary Port

    This port is unused at the present time.

    Cancel and Load Buttons

    This section of the Communications Setupwindowincludes the command buttons describedbelow.

    Cancel Button Cancels all changes made in this window and returns you to the Setupwindow.

    Load Button Loads all communications parameters from the PC into the 505ITM. All

    new communications parameters take effect only after the 505ITMhardware has been reset.

  • 8/10/2019 QUALITROL IST-104-1 ITM505 HW SW Instructions Rev-28691

    55/113

    505ITM Hardware & Software Manual

    49www.qualitrolcorp.com

    Input Modules

    The Input Module Setupwindow opens when you click [Input Modules] from the Setupwindow. The 505ITM has up to four input modules. The types of available modules aredescribed in the Modulessection of the Hardware Instructions.

    To begin, click the tab of the input module that you want to configure. The Input ModuleSetupwindow displays for the selected input module.

    More Information

    Module

    Custom Scaling

    CT Current Parameters

    Winding Temperature Setup (Calculated)

    Low-End and High-End Scaling Values

  • 8/10/2019 QUALITROL IST-104-1 ITM505 HW SW Instructions Rev-28691

    56/113

    505ITM Hardware & Software Manual

    50www.qualitrolcorp.com

    Module

    This section of the Input Module Setupwindowincludes the fields described below.

    Input ModuleNumber

    Indicates the input module number that is being configured. This field isfor informational purposes only.

    Input ModuleType

    Indicates what module type was installed in this corresponding site of the505ITM. This is designated at the factory according to the input modulethat was ordered and installed. If the module type indicates "undefined,"there is no physical module installed in that location.

    Input ModuleFunction

    Enables and disables the module function. If the input module isdisabled:

    It cannot be viewed on the LCD display located on the front panel.

    It cannot control any relays.

    Input SignalName

    Identifies the input module name. The name can be changed and canconsist of up to seven alphanumeric characters. This name will alsoappear on the 505ITM LCD display, along with its input signal value.

  • 8/10/2019 QUALITROL IST-104-1 ITM505 HW SW Instructions Rev-28691

    57/113

    505ITM Hardware & Software Manual

    51www.qualitrolcorp.com

    Custom Scaling

    The configuration parameters in this section of the Input Module Setupwindowallow you tochange the scaling of the process input signal. Custom Scaling is not available if the InputModule Typefieldis set to a "Logic Switch", "AC/DC Input" or Temperature input module,for example, RTDs, Ambient, and CT, Temperature. The fields in this section aredescribed below.

    Use CustomScaling

    Indicates that custom scaling is defined for the input signal. If the checkbox is checked, you must specify three additional parameters for theinput:

    Custom Units Labelfield allows the user up to three characters for aunit of measure, for example, ppm, RH, %.

    Low-End Valuefield is the scaled minimum value for thecorresponding minimum input value.

    High-End Valuefield is the scaled maximum value for thecorresponding maximum input value.

    Custom UnitsLabel

    If the Use Custom Scalingfield is checked, the 505ITM uses this labelinstead of the default label defined by theInput Module Typefield. Thislabel can consist of up to three alphanumeric characters.

    Low-End andHigh-EndScalingValues

    Provide the custom values corresponding to the indicated low-end andhigh-end values for the selected module type. The 505ITM translates theinput signal into the alternate monitors using the low- and high-endvalues provided.

    Example:You have a Dissolved Gas Monitor with 0 - 1 mA representinga 0 - 2000 PPM scale. Using a 0 - 1 mA input module, set:

    Low-End Scaling Valuefield to 0 mA = 0

    High-End Scaling Valuefield to 1 ma = 2000

    Custom Units Labelfield to PPM

    An input of 0.4 mA would appear as 800 PPM on the display.

    Low-End and High-End Scaling Values

  • 8/10/2019 QUALITROL IST-104-1 ITM505 HW SW Instructions Rev-28691

    58/113

    505ITM Hardware & Software Manual

    52www.qualitrolcorp.com

    CT Current Parameters

    If the Input Module Typefieldis set to "CT, Current" or CT, Temperature, this section ofthe Input Module Setupwindowprovides linear scaling between the 505ITM output and theactual current sensed by the Power Transformer CT. The fields in this section are describedbelow.

    CT clamp-onsensor Rating

    Indicates the maximum rating of the CT clamp-on sensor used with thismodule, for example, 10 Amp, 30 Amp). You can read this rating from theclamp-on sensor. When selecting the sensor value, make sure that:

    the maximum rating of the sensor value is greater than the current it

    will be sensing. This will prevent saturation. the sensor value is as close to the maximum current reading as

    possible. This will optimize the resolution.

    CT ratio(2 fields)

    Provides the parameters needed for CT Current Scaling. The left fieldidentifies the true current, such as winding or bushing, being sensed bythe instrumentation or metering CT. The right field identifies thesecondary or output of the instrumentation or metering CT that will beused with the 505ITM Clamp-On CT Sensor.

    Example:If the Instrument CT on the transformer has a ratio of 1200 : 5Amps, the two CR ratiofields are set to CT Ratio = 1200 : 5. The clamp-on sensor is on the 5 Amp circuit and now acknowledges that the 5 Amp

    are passing through the CT to scale the current to the 1200 Amp range.

  • 8/10/2019 QUALITROL IST-104-1 ITM505 HW SW Instructions Rev-28691

    59/113

    505ITM Hardware & Software Manual

    53www.qualitrolcorp.com

    Winding Temperature Setup (Calculated)

    This section of the Input Module Setupwindowallows you to configure the "CT,Temperature" input module type.

    The checked box indicates which winding calculation method is active. Click either methodto open the corresponding setup window.

    Simple Calculation Setup Method

    Advanced Calculation Setup Method

    Simple Calculation Setup Method

    This method allows you to configure the simple scaling parameters for input module typesthat are set to CT, Temperature. The winding calculation method presented here

    corresponds to the process used by the IEEE Guide for Loading Mineral-Oil-ImmersedTransformers (IEEE Std C57.91).

    The Simple CT Temperature Scaling Setupwindow opens when you click [SimpleCalculation Setup] from the Winding Temperature Setupsectionof the Input Module Setupwindow.

  • 8/10/2019 QUALITROL IST-104-1 ITM505 HW SW Instructions Rev-28691

    60/113

    505ITM Hardware & Software Manual

    54www.qualitrolcorp.com

    More Information

    Input Module Number and Input Signal Name Fields

    Winding Rise Parameters

    Oil Rise Parameters

  • 8/10/2019 QUALITROL IST-104-1 ITM505 HW SW Instructions Rev-28691

    61/113

    505ITM Hardware & Software Manual

    55www.qualitrolcorp.com

    Input Module Number and Input Signal Name Fields

    This section of the Simple CT Temperature Scaling Setupwindowincludes the fieldsdescribed below.

    Input modulenumber

    Displays the input module number for which the simple CT,Temperature scaling parameters are being configured.

    Input signalname

    Displays the user-assigned name of the signal for which the simple CT,Temperature scaling parameters are being configured.

    Winding Rise Parameters

    This section of the Simple CT Temperature Scaling Setupwindowincludes the fieldsdescribed below.

    OilTemperature

    Adder

    Allows you to select one of the available RTD Input Modules. Theselected RTD Input Module will be the liquid temperature measurementthat is added to the winding curve simulation, for example, top oiltemperature.

    Winding RiseTemperature

    Allows you to set the number of degrees that the winding temperature willbe over the oil temperature (delta T) at nominal load for the transformer.You can calculate this parameter from the transformer heat run testreports by subtracting the oil temperature at nominal loading from thewinding temperature at nominal loading. If you cannot obtain the heat run

    report, contact the transformer manufacturer for these values.CurrentFactor

    Displays the value measured at the clamp-on CT sensor of the 505ITM atnominal loading of the transformer.

  • 8/10/2019 QUALITROL IST-104-1 ITM505 HW SW Instructions Rev-28691

    62/113

    505ITM Hardware & Software Manual

    56www.qualitrolcorp.com

    Winding TimeConstant

    Allows you to set the time constant for the temperature rise of thiswinding. The time constant is defined as the time it takes the windingtemperature to reach 63.2% of its final value for a given current input.

    Note: It takes approximately five time constants to elapse for thetemperature to reach full value for a given load. If the time

    constant is unknown, the IEEE Std C57.91-1995 Annex Grecommends a value between 3 and 7 minutes. Qualitrol uses adefault value of 6 minutes.

    Oil Rise Parameters

    This section of the Simple CT Temperature Scaling Setupwindowincludes the parametersthat are required if you want to use the Critical Forecast feature of the TransLifeTMSetup. Ifyou are not using the Critical Forecast feature, leave these parameters at the default values.

    The fields in this section are described below.

    Oil risetemperature

    Allows you to set the temperature rise value of the transformer oil atnominal load. This is derived from the Transformer Heat Run Report bytaking the value of the Oil Rise Temperaturefield over ambient atnominal load, and subtracting the ambient test temperature.

    Oil Rise Temp = (Oil Rise Temp over Ambient - Ambient Temp)

    Oil time

    constant

    Allows you to set the time constant for the temperature rise of this

    winding. The time constant is defined as the time it takes the windingtemperature to reach 63.2% of its final value for a given current input.

    Note: It takes approximately five time constants to elapse for thetemperature to reach full value for a given load. If the Oil TimeConstant is not provided, the IEC 354 Loading Guide for Oil-Immersed Power Transformers recommends:

    For distribution transformers:

    All c